overview and alignment results of the bpm button collimator mock-up
DESCRIPTION
Overview and alignment results of the BPM button collimator mock-up. D. Wollmann, on behalf of - PowerPoint PPT PresentationTRANSCRIPT
CWG 02.04.2012
Overview and alignment results of the BPM button
collimator mock-up
D. Wollmann, on behalf of
O. Aberle, R.W. Assmann A. Bertarelli, C. Boccard, F. Burkart, R. Bruce, M. Cauchi, A. Dallochio, D.
Deboy, M. Gasior, R. Jones, V. Kain, L. Lari, A. Masi, A. Nosych, A. Rossi, S. Redaelli, B.
Salvachua, G. Valentino, E. Veyrunes
Outline
• Introduction
• Why Collimators with Beam Diagnostics Functionality?
• Design overview
• Results from Beam Measurement
• ConclusionDaniel Wollmann 2
PRESENT: Beam based setup and qualification of collimation
system• Centre collimator jaws around beam (by touching
the beam halo create losses BLM based method)
• Determine local beam size at collimators• Set up system with agreed collimator settings
~7 mins per collimator & state (two beams in parallel)
Destructive in halo. Overhead: special fills and ramps!• Qualify system by measuring the cleaning efficiency
– β-tron losses– Momentum lossesDaniel Wollmann 3
History of Collimator Jaws with BPM Buttons
• First discussions early on in 2004 during design of the phase 1 collimators: F. Caspers and R. Assmann.
• Due to time pressure: Decision to delay this improvement to the phase 2 collimator design.
• Given as design requirement for any new LHC collimators:– Concerns on impact of shower initially strong. However:
• Phase 1 collimator tests in SPS were not able to generate shower-induced changes in BPM’s in SPS good news that showers do not impact performance.
• Also no shower effects so far reported in LHC for LHC BPM’s in IR3/7.
– EN/MME detailed design work to integrate BPM buttons ongoing. – SLAC pursuing design with buttons in flanges.– BE/BI designing BPM button solution and electronics.
• First phase 2 mockup collimator with buttons installed in SPS early 2010. 4Daniel Wollmann
Conceptual Strengths of Design with Buttons in Movable Jaws
• Advantages:– Centering of beam done by minimizing difference signal
(L-R), down to zero independent of absolute calibration of BPM signals, electronics offsets, … for same cable length.
– The now movable buttons are placed much closer to the circulating beam much better resolution than achievable with standard LHC BPM’s.
– No intercepting of beam halo, so fully non-destructive. No special fills and intensity constraints!
• Disadvantages:– If buttons are inside loss-induced showers then signals
could be disturbed? Setup with buttons might only be possible during low-loss periods (e.g. just before colliding, during stable beams, …). Look at difference up- (no shower) and downstream (max shower).
Daniel Wollmann 5
Operational Gains with in-jaw BPM buttons?
• Drastic reduction of setup time of collimation system (gain time for physics). For example 7 min 1 second Factor > 100!
• Continuous monitoring of beam offsets at collimators. Measurement of jaw angle w.r.t. the closed orbit. Increased passive machine protection as orbit drifts are quickly detected (watch dog).
• Collimators can follow without overhead long-term orbit drifts.
• More flexibility for local orbit changes in the experimental IPs (crossing angle, separation for luminosity leveling, etc.). Relaxed restrictions for luminosity optimization in the experimental IPs.
• Allows reduction of margins between collimator families, as collimators can follow orbit drifts tighter collimator settings possible better cleaning and lower beta* possible.
Daniel Wollmann 6
First CERN mock-up collimator with integrated BPM buttons
(Jan 2010)• BPM mock-up produced at CERN
(EN-MMI, BE-BI, Collimation Team)• Installed into SPS in 2010
Daniel Wollmann 7
BPM buttons
Distance from jaw face (Up-, Downstream buttons): 10mm
Courtesy A. Bertarelli, A. Dallocchio, O. Aberle, et. al
First CERN mock-up collimator with integrated BPM buttons
(Jan 2010)• BPM mock-up produced at CERN
(EN-MMI, BE-BI, Collimation Team)• Installed into SPS in 2010
Daniel Wollmann 8
BPM buttons
Distance from jaw face (Up-, Downstream buttons): 10mm
Courtesy A. Bertarelli, A. Dallocchio, O. Aberle, et. al
Buttons at extremities proved most useful (no shower effects seen). Therefore here focus on results of these buttons. No beam results from central two buttons!
Daniel Wollmann 9
8th January 2010
From the lab into the SPS tunnel
First CERN mock-up collimator with integrated BPM buttons
(Jan 2010)
Tank
Jaws
Courtesy A. Bertarelli, A. Dallocchio, O. Aberle et. al
Measurements in the SPS 2011:4 MDs (Mai/June,
September/October)• Test new BPM electronics (details see talk by M.
Gasior).• Test new BPM based alignment method.• Compare to classic BLM based alignment method.• Measure non-linearity of BPM buttons (see talk by
A. Nosych)• Vacuum: measure out-gazing during movement and
beam impacts (not discussed here).• Test BPMs with standard LHC electronics to
measure beam offsets in single pass / turn by turn application of button collimators in transfer line collimators?Daniel Wollmann 10
SPS Beam Tests: Compare present BLM and new BPM method for jaw alignment
• Create a four corrector orbit bump at collimator (steps of 500 mm).
• New method: Align collimator jaws around beam with in-jaw BPM buttons (~few seconds).
• Present method: Align collimator jaws around beam with BLM based method (~7min per collimator).
Accuracy:• BLM-method: max. error ± 50mm due to used step
size of 50mm.• BPM-method: assumed max. error ± 50mm as
electronic channels were not yet calibrated for gain and cable length.Daniel Wollmann 11
Measurements: Beam offsets with 2 methods
Daniel Wollmann 12
NEW (few seconds)
OLD(~7mins)
Good agreement between orbit bump, BLM and BPM centers.
Correlation: Bump versus BPM-method,
BLM-method
Daniel Wollmann 13
Correlation: BPM-method versus BLM-
method
Daniel Wollmann 14
Correlation: BPM-method versus BLM-
method
Daniel Wollmann 15
Takes 7 min per collimator setup. Requires special low intensity fills.
Correlation: BPM-method versus BLM-
method
Daniel Wollmann 16
Takes a few s per collimator setup. Can be done all the time: no overhead.
Correlation: BPM-method versus BLM-
method
Daniel Wollmann 17
Excellent correlation between the two methods.
Measured deviation between: bump & BLM / BPM
Daniel Wollmann 18
Maximum deviation to bump :
[-50mm, +140mm]
HIGHLIGHT summary result
Measured deviation between: BLM & BPM
Daniel Wollmann 19
Maximum deviation between BLM and BPM :
[-50mm, +63mm]
HIGHLIGHT summary result
Measured deviation to bump: BLM & BPM
Daniel Wollmann 20
Orbit drifted
Deviation to bump:
≤ ±40mm
HIGHLIGHT summary result
Measured deviation between: BLM & BPM
Daniel Wollmann 21
Orbit drifted
Deviation between BLM and BPM :
≤ ±25mm
HIGHLIGHT summary result
Measurement: Influence of radiation on BPM signals
Daniel Wollmann 22
• Variation of BPM signal (<35mm)
• Drift of signal due to non-linearities in the BPM electronics at low beam intensities
Collimator gap: 21mm
Measurement with secondary halo created by an upstream SPS collimator
Low intensity point
SPS beam tests with standard LHC electronics: Measure turn
by turn accuracy
Daniel Wollmann 23
• Measurement performed over 10mins.
• Move jaw around beam at constant gap (0.5mm steps).
• Average beam position measured by BPMs.
• Calibrating slope and offset to bump settings.
SPS Beam Tests with standard LHC electronics: Measure turn
by turn accuracy
Daniel Wollmann 24
Agreement between bump and BPM measurement not as good as with other
electronics.
SPS Beam Tests with standard LHC electronics: Measure turn
by turn accuracy
Daniel Wollmann 25
• RMS of variation: ~82mm.• b-tron oscillations to bet
taken into account.
Conclusion• Collimators with integrated BPMs have been shown
to:– Decrease setup time: ~ factor > 100 per collimator
without overheads.– Continuous monitoring of beam in collimator: Passive
machine protection.• Measurements show:
– Excellent agreement between the two methods.– Average discrepancy between BPM and BLM method
better than 25mm, limited by the step size used. Present LHC setup accuracy shown with BPM’s.
– No disturbance in BPM signal due to primary protons or secondary showers seen so far.
– RMS variation for turn by turn measurements ~82mm. Further measurements for single pass application needed.
• Better cleaning and lower beta* possible.Daniel Wollmann 26
END
Thank you for your attention!
Daniel Wollmann 27
Backup slides
Daniel Wollmann 28
Design: BPM Mockup
• Simplified jaw design: only to support BPM buttons• Almost no cooling• Thin jaw material
Daniel Wollmann 29
Cross section of Mockup jaw
Cross section of phase 1 secondary collimator jaw