outcomes of transscleral sulcus fixation of intraocular lenses through a 2.4-mm incision with an...
TRANSCRIPT
![Page 1: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/1.jpg)
Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up
Akiko Masai, MD, Tomoichiro Ogawa, MD Takuya Shiba, MD, Hiroshi Tsuneoka, MD
Department of OphthalmologyThe Jikei University School of Medicine
Tokyo, Japan
The authors have no financial interest in the subject matter of this e-poster.
![Page 2: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/2.jpg)
Objective
• To evaluate postoperative outcomes of scleral-sutured ciliary sulcus fixation of foldable intraocular lens (IOL) through a 2.4-mm incision
![Page 3: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/3.jpg)
Methods
• Subjects: 9 eyes (9 patients)– Lens subluxation: 1 eye– Aphakia: 5 eyes– IOL dislocation 3 eyes
• Variables examined– Best-corrected visual acuity (BCVA)
• Before and after surgery (1 week; 1, 3, 6, 12 months)– Surgically induced astigmatism (SIA) (Cravy method)
• After surgery (1 week; 1, 3, 6, 12 months)– Endothelial cell density
• Before and after surgery
![Page 4: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/4.jpg)
Surgical Technique
1. Preparation for IOL setting of PC-9 hooking• 10-0 double polypropylene thread (PC-9, Alcon) was
inserted through the cartridge (E1, HOYA).• PC-9 was looped around the haptics of the IOL
(VA70AD or VA65BB, HOYA) by using a cow hitch knot.
![Page 5: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/5.jpg)
Surgical Technique
2. Scleral pockets created at the 2 and 8-o’clock positions
3. Anterior vitrectomy performed4. 2.4-mm scleral corneal incision created at
12-o’clock position
![Page 6: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/6.jpg)
5. IOL setting: IOL and hooked PC-9 loaded into cartridge
![Page 7: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/7.jpg)
6. IOL inserted through incision with injector (ISH-001, HOYA), leaving opposite haptics out of anterior chamber
7. PC-9 looped around haptics8. IOL fixed to ciliary sulcus
![Page 8: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/8.jpg)
BCVA
Surgically induced astigmatism was minimal
Early improvement in BCVA maintained for 1 year
SIA
Result 1
![Page 9: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/9.jpg)
![Page 10: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/10.jpg)
Result 2
• Mean endothelial cell density reduced 9.80%• Complications
– IOL iris capture and pupillary block in 1 case– Retinal detachment in 1 case– Cystoid macular edema in 1 case.
![Page 11: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/11.jpg)
Treatments for Complications
• IOL capture– IOL repositioned– Peripheral iridectomy at 6 and 12 o’clock
• Retinal detachment– Scleral buckling surgery– Retina reattached
• Cystoid macular edema – Improvement without medication
![Page 12: Outcomes of Transscleral Sulcus Fixation of Intraocular Lenses through a 2.4-mm Incision with an Injector System: 1-Year Follow-Up Akiko Masai, MD, Tomoichiro](https://reader036.vdocuments.us/reader036/viewer/2022062309/5697c02f1a28abf838cda5cb/html5/thumbnails/12.jpg)
Conclusion
• With this technique, visual acuity improved early postoperatively and was maintained for 1 year