Orion Star Series Meter User’s Guide Analyze Detect Measure Control

Upload: others

Post on 08-Mar-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Orion Star Series MeterUser’s Guide

Analyze • Detect • Measure • Control™

Page 2: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

ROSS and the COIL trade dress are trademarks of Thermo Electron Corporation.

AQUAfast, Cahn, EZ Flash, Ionalyzer, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS Ultra, Sure-Flow, TEA Analyzer, Titrator PLUS, TURBO2 and Wine Master are registered trademarks of Thermo.

1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoE Z, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, Flash Titration, Flash Titrator, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Constellation Software, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo.

Guaranteed Success and The Technical Edge are service marks of Thermo.

PerpHecT meters are protected by U.S. patent 6,168,707.

PerpHecT ROSS are protected by U.S. patent 6,168,707.

ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753.

ionplus electrodes and Optimum Results solutions are protected by US Patent 5,830,338.

ROSS Ultra electrodes are protected by US patents pending 6,793,787.

Orion ORP Standard is protected by US Patent 6,350,367.

Orion NoCal electrodes have patent pending.

© Copyright 2005, Thermo Electron Corporation. All rights reserved. Question everything, and Analyze.Detect.Measure.Control are trademarks of Thermo Electron Corporation.

The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.

This publication supersedes all previous publications on this subject.

Page 3: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IOrion Star Series MeterThermo Electron Corporation

Table of Contents

Chapter IGeneral Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-1 How to use the manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-1 Orion Star Series Quick Start Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-2

Chapter IIIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II-1 Unpacking the meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II-1 Installing the power adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II-1 Connecting the electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II-2 Turning on the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II-3 Replacing the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II-4

Chapter IIIOperational Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-1 Display definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-1 Keypad functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-3 Set up Menus navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-4 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-8 Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-9

Chapter IVpH. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-1 Set up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-1

Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-3 Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-5

Chapter VISE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . V-1 Set up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . V-1

Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . V-4 Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . V-7

Chapter VImV and ORP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI-1 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI-1

Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI-2

Page 4: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationII Orion Star Series Meter

Chapter VIIConductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VII-1 Set up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VII-1

Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VII-7 Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VII-8

Chapter VIIIDissolved Oxygen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VIII-1 Set up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VIII-1

Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VIII-7

Chapter IXKey Meter Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-1 Audible signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-1 Automatic shutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-1 Display backlight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-1 AUTO-READ™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-1 Stirrer control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-2 Methods and passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-2 Calibration alarm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-3 Datalog setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-7 General instrument setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-9 GLP instrument setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-12 Time and Date setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-15 RS-232 instrument setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-19 Multiparameter printout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-21 Log View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-22 Error Codes and Debugging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX-23

Chapter XSpecifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . X-1

Chapter XITerms and Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XI-1

Chapter XIIDeclaration of Conformity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . XII-1

Page 5: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

I-1Orion Star Series MeterThermo Electron Corporation

Chapter I General InformationHow to Use the ManualFor a quick and easy start up begin with the Quick Start Guide. If more details are required start with more detailed instructions beginning on Chapter II, Introduction. This operator’s manual offers more detailed information on meter features, calibration and measurement procedures and options that will optimize your daily activities. The table of contents and index can be used to find specific topics.

Visit www.thermo.com/water for advanced information such as operation of meter by remote control, 21 CFR software package manual, and other advanced capabilities. Our website also has an updated list of application notes developed specifically for the Star series meter line.

Orion Star Series Quick Start GuideStar series Quick Start guide is designed to get user’s up and running in a few minutes with basic calibration and measurement information.

You can contact the Thermo Water Analysis Instruments Technical EdgeSM for a method development request form and fee schedule to have a method developed for your specific sample. Within the United States call 1.800.225.1480, outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit www.thermo.com/water.

Page 6: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationI-2 Orion Star Series Meter

Intended use of the Star Series MetersThermo Electron Corporation provides specific pH, ISE (ion selective electrode), DO (dissolved oxygen) and Conductivity meters for laboratory and field use. All meters are microprocessor controlled.

These systems are ideal for busy, multi-user laboratories or plant environments. The bench top meters are designed for use in a laboratory setting according to the instructions in this manual. The benchtop meters carry a IP54 dustproof and splash proof rating and can easily be cleaned using a damp cloth, water with a mild detergent. The portable meters are constructed for use outdoors with a IP67 waterproof rating; they can float and be submersed for a short period of time without any negative effect on their operation. Any use outside of these instructions could invalidate your warranty cause permanent damage to the meter.

The Orion Star Series Meters offer basic RS232 interface connections to a printer or computer. Other features include all required features for GLP documentation to create printouts of data and reports fully compliant with “Good Laboratory Practice.”

Page 7: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Keypad Functions• Turns meter on, if shut off.

• Toggles backlight on & off, if meter is on.

• If meter is on, holding down will turn off the meter.

• Changes measurement mode of the selected line.

• Changes selected line in Setup, Methods & Log View.

• Edits the value of the flashing digit for Setup, Password entry, and Calibration.

Switches arrow on left of the screen between 3 display lines to select and edit.

Changes selected digit to edit and move decimal point when changing values in Setup, Password entry, & Calibration.

Starts calibration for currently selected line and measurement mode.

• If arrow is pointed to the top line and the current units are pH, will start a pH calibration.

• Each time is pressed in the calibration it will accept the current point value and move to the next cal pt until the max # of cal pts are performed, then it will return to measurement mode.

• Print & Log a measurement in continuous or timed measurement mode.

• Prints, Logs, & Freezes screen when the reading becomes stable.

• Exits SETUP and returns to measurement mode.

• Accepts Cal pt and returns to measurement mode.

Enters SETUP menu starting with selected line and measurement mode:

If arrow is pointed to top line and current units are ISE, then will enter the ISE setup screen.

Enters LOG View & Download Screens.

Turns stirrer on & off.

Keypad Layout

Stirrer On/Off

MeasureSave/Print

View Log

Power

Line SelectDigits

Setup

Calibrate

Back Light

Page 8: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Instrument Setup Mode• pH (PH)

• ISE (ISE)

• Cond (COnd)

• DO (dO)

• DUE — Calibration frequency alarm (dUE)

• Date — Date & Time (dAtE)

• READ — Type “Auto/Cont”, Time “Timed readings” (rEAd)

• Sets RS232 Baudrate (r232)

• Log — log delete on downloading or log roll over (LOg)

• General Instrument Setup (gEn)

• GLP — Turns ON/OFF Methods & Password Protection (gLP)

To select the General Instrument setup option line (Middle Line).

Returns to measurement mode.

Stirrer jackstirrer or DO

Auto stir probe

Reference pH or ISE

CondATC8 pin

DO9 pin

GroundBNC

PowerRS232

Rear View of Meter - Electrode Connections

Page 9: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Method Select and Edit

If methods have been turned on in gLP then this screen will appear when setup is pressed.

To edit the selected method.

& To select the method to be used or edited.

To edit the selected method.

To implement the selected method and returns to measurement mode.

Calibration Setup and Measurement

pH, ISE, Conductivity Calibration Screen for 1-5 Points.

Accepts Autobuffer recognition as a 1 pt calibration & starts the 2nd cal pt if pressed again.

, & To edit the value for a manual buffer entry.

Accepts Autobuffer recognition and returns to measurement mode.

DO Calibration

& • AIr = Water Saturated Air

• H2O = Air Saturated Water

• mAn = Winkler (Manual)

• SEt0 = Zero

Auto Read

pH/ISE/ORP units

Conductivity units

DO units

LogTemperature

Sensor Error

Calibration

Meter Display

Page 10: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Password Entry

This screen will appear if password has been setup in gLP, enter password to enter and edit method.

& Changes the value of the flashing digit.

Changes the digit to be scrolled.

Changes display lines.

Returns to measurement mode.

pH Setup Menu Screen

Selects pH setup.

• rES = Resolution

• bUF = Auto-Buffer Set (USA or EUrO)

Selects the pH Resolution entry line (Bottom Line).

Returns to measurement mode.

ISE Setup Menu ScreenSelects ISE setup.

• rES = Resolution

• UnIt = ISE Measurement units

• rAng = ISE Measurement range

• nLIn = Non-linear Blank Correction

Selects ISE Resolution entry line (Bottom Line).

Returns to measurement mode.

Page 11: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Conductivity Setup Menu ScreenSelects Conductivity setup.

• tC = Temperature Compensation (Linear or nLF)

• COEF = Temperature coefficient for Linear temp compensation - Screen only shown if Linear Temp Comp Selected

• tdSF = TDS Factor for Total Dissolved Solids measurement

• CELL = Conductivity Cell Constant

• trEf = Reference temperature for temperature compensation

• tyPE = Conductivity cell type (Planar or Conventional)

Selects Cond temperature compensation entry line (Bottom Line).

Returns to measurement mode.

DO Setup Menu Screen.Selects DO setup.

• rES = Percent Saturation Resolution, % SAt

• rES = Concentration Resolution, mg/L

• bAr = Barometric pressure compensation type (Auto/Manual)

• PrES = Manual Barometric Pressure Entry - Only shown if manual pressure compensation is selected

• SAL = Salinity correction factor type (Auto/Manual)

• SALF = Manual salinity correction factor (Only shown if manual salinity correction factor)

• CALt = DO calibration type (Air, Water, Manual, SEtO)

Selects Percent Saturation Resolution entry line (Bottom Line).

Returns to measurement mode.

General Instrument Setup Menu Screen

Selects General Instrument setup:

• AUtO = Auto-Shutoff of the instrument after 20 Minutes without use.

• dEgC = Set Manual Temperature

• StIr = Set stirrer speed

Selects Auto-Shutoff entry line (Bottom Line).

Returns to measurement mode.

Page 12: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Log Setup Menu Screen

Selects Reading setup:

• dEL = Deletes LOG after downloading or roll over readings

Selects Delete Log after downloading entry line (Bottom Line).

Returns to measurement mode.

GLP Setup Menu ScreenSelects GLP setup:

• SEt = Turn SET (Methods) ON/OFF.

• PASS = Enter a Password to secure Methods & Setup

Selects methods to be on or off (Bottom Line).

Returns to measurement mode.

Date & Time Setup Menu ScreenSelects the Date & Time setup:

• HOUr = Enter the Hour in 24 hr format

• mIn = Enter minutes 0 - 59

• tyPE = US or European Style Month/Day/Year or Day/Month/Year

• yEAr = Enter the Year

• dAtE = Enter the Month

• dAy = Enter the day

Selects Hour entry line (Bottom Line).

Returns to measurement mode.

RS232 Setup Menu ScreenSelects RS232 setup:

• bAUd = Baud Rate 1200, 2400, 4800, 9600

• OUtF = Output Format Printer (Prnt) or Computer (COmP)

Selects Baud rate or output entry line.

Returns to measurement mode.

Page 13: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter II Introduction Unpacking the Instrument Carefully remove the meter box from the shipping box. Place the meter box on a bench or a flat, stable work surface. Open the meter box and place the meter on the bench.

Remove the boxes from the shipping box. The number of items present will depend on which meter configuration was ordered.

Note: The line adapter for the benchtop meters inside the meter box in a tray below the meter. ▲

Place all items with the meter on the bench.

Installing the Universal Power Adapter (for benchtop meters only)Carefully, remove the components of white power adapter box. The universal power supply has three plugs to be use for 110V, 220V and 240V wall plugs. Select the plug that is appropriate for your power source and slide it into the grooves on the adapter. A click will be heard when the plug is properly in place in the adapter.

The external electrical power supplies are rated to be operated at 100-240 VAC, 0.5A, 50/60 Hz.

This power supply is the ONLY power supply recommended for use with this unit. Use of any other power supply will void your meter warranty.

Gently remove the protective plastic over the meter display.

II-1Orion Star Series MeterThermo Electron Corporation

Page 14: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationII-2 Orion Star Series Meter

Introduction

Connecting the ElectrodesThe electrodes and probes must be connected to the correct locations on the meter. Review the attached diagram. Depending on the functions on the meter you purchased, some of these connectors may not be on your meter.

Benchtop Meter - Electrode Connections

Portable Meter - Electrode Connections

Stirrer jackstirrer or DO

Auto stir probe

Reference pH or ISE

CondATC8 pin

DO9 pin

GroundBNC

PowerRS232

RS232

BNC

CondATC 8 pin

DO 9 pin

Page 15: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

II-3Orion Star Series MeterThermo Electron Corporation

Introduction

• The pH, ISE and ORP combination and sensing electrodes have a waterproof BNC connector. These are connected to the metal BNC post.

• Automatic Temperature Compensation (ATC) probes are inserted into the waterproof 8 pin MiniDIN Connector.

• Benchtop meters have a pin tip jack for a separate reference electrode. These require an appropriate BNC sensing electrode for measurement.

• Conductivity probes use the waterproof 8 pin MiniDIN connector. All Orion Star series 2- and 4- electrode cell conductivity probes have built-in temperature sensors.

• The DO autostir probe uses the waterproof 9 pin MiniDIN connector and the stirrer jack for the smaller probe connector.

Note: The 9708 DO electrode can be used with the BNC connector. ▲

Non-glass ISFET pHuture® MMS™ sensors for pH, ORP and temperature use the waterproof 9 pin MiniDIN connector. The pH/ORP,conductivity and temperature Pentrode™ (6175MD) uses both the waterproof 9 pin MiniDIN and waterproof 8 pin MiniDIN connectors.

Turning On the InstrumentWith the batteries installed in the portable meters or either the batteries installed or line power attached to the benchtop meter, press the Power Key to turn on the instrument. A quick press of the Power Key after the meter has powered up will toggle the backlight on & off. If the meter is on battery power then the backlight will automatically turn off after 2 minutes to save battery life. If the low Bat is displayed you will no longer be able to turn on the backlight until the batteries are replaced. If you benchtop is on line power then the backlight will stay on until turned off with the power/backlight key. To turn off the meter press and hold the power button for 3 seconds.

Page 16: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationII-4 Orion Star Series Meter

Introduction

Replacing the BatteriesWhen battery power is running low, The screen will periodically flash Lo Bat. If the batteries are low then the meter will no longer allow you to turn on the backlight to preserve battery life. The batteries are getting low and should be changed within 20 hours.

Note: The instrument uses four AA Alkaline Batteries. Do not use lithium or rechargeable batteries. Improper installation of non-alkaline batteries could create a hazard. ▲

To access the Benchtop Meter Battery Door

Note: Stored data, calibrations and methods will remain in nonvolatile memory, so data is saved while batteries are changed. ▲

1. Turn the meter off.

2. Place it upside down gently on a clean lint-free wipe or cloth so the display is not scratched.

3. Remove the battery case cover on the benchtop meter on the back of the meter.

Page 17: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

II-5Orion Star Series MeterThermo Electron Corporation

Introduction

4. Remove all batteries from the meter.

5. Insert all new batteries in the orientation depicted in the diagram in the battery.

Note: One of the batteries has the positive top inserted at the spring end of the battery compartment. This is the correct orientation. ▲

6. Replace the cover.

7. Turn the meter power on.

8. If the batteries took more than a minute to replace, you may want to update the time & date by pressing the setup key to reset the time and date.

9. Dispose of the used alkaline batteries in accordance with local regulations.

Page 18: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationII-6 Orion Star Series Meter

Introduction

To access the Portable Meter Battery Door Unscrew the 2 screws in the center of the back of the meter. One is located at the bottom and one is in the middle of the back. These are captive screws and will not come out. With these loose lift the bottom back of the meter to expose the batteries. Install the batteries in the orientation shown in the graphics of the battery compartment.

Note: The instrument uses four AA Alkaline Batteries. Do not use lithium or rechargeable batteries. Improper installation of non-alkaline batteries could create a hazard.

Note: After a battery change the real time clock will be off by however much time is used to replace the batteries. ▲

Page 19: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

II-7Orion Star Series MeterThermo Electron Corporation

Introduction

Meter Cleaning Instructions The bench top meter is IP54 dustproof and splashproof. Regular maintenance should include dusting and wiping the unit down with a damp cloth. A warm water or a mild water based detergent may be used if necessary

Spills should immediately be removed using the proper procedure for the type of spill.

The portable meters are IP67 and may also be maintained with regular cleaning with a water-based detergent and a cloth. These meters are submersible and will float for short periods of time. After being submersed the meters should be cleaned with a mild detergent solution for the protection of future users.

Approved Operating Conditions The Star Series Meters are designed to operate according to their specifications under the following conditions:

Portable and Bench top Meters Environmental Conditions

• Ambient Operating Temperature 5 to 45°C

• Relative Humidity 5 to 85% non-condensing

• Storage Temperature –20 to 60 °C

• Storage Humidity 5 to 85% non-condensing

Wall Adapter Environmental Conditions

• Ambient Operating Temperature 0 to 40 °C

• Relative Humidity 20 to 90%

• Operating Altitude Sea Level to 10,000 feet

• Storage Temperature –20 to 85°C

• Storage Humidity 20 to 90%

• Storage Altitude Sea level to 10,000 feet

Page 20: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationII-8 Orion Star Series Meter

Page 21: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter III Operational OverviewDisplay Definitions

Display - the screen has 4 measurement lines.

• At the top left a line containing symbols begins with the temperature reading.

• The top main line for pH, mV, Rel mV, & ISE

• The middle main line for Conductivity, TDS, Salinity, & Resistivity

• The bottom main line for DO % Sat, DO Concentration, and Barometric Pressure

a. On the left of the screen are arrows that indicate the currently selected line.

b. The right of the screen shows the measurement units for the selected measurement modes. The units will flash until the reading is stable.

c. nLF icon indicates that non-linear temp comp is active for conductivity.

d. The graph indicates Calibration mode or calibration setup.

Auto Read

pH/ISE/ORP units

Conductivity units

DO units

LogTemperature

Sensor Error

Calibration

5 Star Meter

III-1Orion Star Series MeterThermo Electron Corporation

Page 22: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIII-2 Orion Star Series Meter

Operational Overview

e. The SETUP icon indicates Setup mode.

f. The MAN icon indicates manual pH calibration was performed. Cal icon is only lit in Cal mode. If any pH points were manual then the man icon is lit.

g. The numbers on the top indicate which pH Auto buffers were used in the last calibration.

h. The Caution symbol indicates an error condition or when shown with the probe that a sensor quality issue exists.

i. The AR icon indicates the meter is in AUTO-READ mode.

j. The printer icon indicates the meter is printing, exporting or logging data.

3 & 4 Star Meter Display

Auto Read

pH/ISE/ORP units

Conductivity units

DO units

LogTemperature

Sensor Error

Calibration

Page 23: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

III-3Orion Star Series MeterThermo Electron Corporation

Operational Overview

KeypadsPortable Mater

The portable Meters have 9 Keys:

1. The POWER key:

a. If the meter is off a quick press of the power key will turn the meter on.

b. If the meter is on a quick press of the power key will toggle the backlight on & off.

c. If the meter is on, holding down the power key will turn off the meter.

2. The LINE SELECTION key switches the arrow on the left of the screen between the 3 display lines to select the line to edit.

Power Back Light

MeasureSave/Print

View Log

Line SelectDigits

Setup

Calibrate

Scroll Up

Scroll Down

Page 24: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIII-4 Orion Star Series Meter

Operational Overview

3. The UP & DOWN arrow keys:

d. Changes the measurement mode within the selected line in measurement. (such as from TDS to Resistivity measurement on the selected Conductivity line)

e. Changes the options for the selected line in Setup, Methods & Log View.

f. Edits the value of the flashing digit of numbers in Setup, Password entry, and Calibration.

4. The DECIMAL/DIGIT key changes the selected digit to edit and moves the decimal point when editing numbers in Setup, Password entry, & Calibration.

5. The LOG key enters the LOG View & Download Screens.

6. The CALIBRATE key starts a Calibration for the currently selected line and measurement mode.

g. So if the Arrow is pointed to line 1 (Top Line) and the current units on that line is pH, then pressing the CALIBRATE key will start a pH Calibration.

h. Each time the CALIBRATE key is pressed in the calibration it will accept the current point calibration value and move to the next calibration point until the maximum number of calibration points is performed, then it will return to measurement mode.

7. The SETUP key will enter the SETUP menu starting with the SETUP for the currently selected line and measurement mode.

i. So if the Arrow is pointed to line 1 (Top Line) and the current units on that line is for ISE or no units are displayed, then pressing the SETUP key will enter the ISE setup screen.

Page 25: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

III-5Orion Star Series MeterThermo Electron Corporation

Operational Overview

8. The MEASURE key will:

j. Print & Log a measurement in Continuous measurement mode.

k. Print, Log, & Freeze the screen when the reading becomes stable in AUTO-READ mode.

l. Will exit SETUP and return to measurement mode when in setup.

m. Will accept the Calibration point and return to measurement mode when in Calibration.

Benchtop Meters

The benchtop meters have an additional key

On all Benchtop meters there is a 10th key for turning on & off the STIRRER. In AUTO-READ mode the MEASURE key will also turn the stirrer on & off automatically during the measurement. This feature is not available on portable meters.

Stirrer On/Off

MeasureSave/Print

View Log

Power

Line SelectDigits

Setup

Calibrate

Back Light

Page 26: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIII-6 Orion Star Series Meter

Operational Overview

SetupHow to Access the Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

c. The SETUP key will enter the SETUP menu starting with the SETUP for the currently selected line and measurement mode. So if the Arrow is pointed to line 1 (Top Line) and the current units on that line is ISE, then pressing the SETUP key will enter the ISE setup screen.

How to Navigate the Setup Menus:

• Press the UP & DOWN arrow to scroll through the Setup Menu’s until the Setup Menu you want (Grayed Text in Center Column) is displayed on the Top line.

• Press the RIGHT ARROW key to select the setup option line (Middle Line).

• Press the UP & DOWN arrow to scroll through the Setup options until the Setup option you want (Non-Grayed Text in Center Column) is displayed on the Bottom line.

• Press the RIGHT ARROW key to select the setup option line (Bottom Line).

• Use the UP & DOWN arrows and the DECIMAL/DIGIT key to edit the value/selection (Value Range or Selections shown in Options Column; Defaults in Bold).

• Press the RIGHT ARROW to accept the value and move back to the top line. Press the MEASURE exit to measurement.

Page 27: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

III-7Orion Star Series MeterThermo Electron Corporation

Operational Overview

Setup Options for each Parameter

Note: All default values are in bold text. ▲

pH PH Options Measurement rEAd Options

Resolution rES 0.1,0.01,0.001 Measurement Type tyPE Cont,AUtO,tImE

Buffer Set bUF USA, EUrO Timed Reading tInE 00:05 - 99:61, 01:00

ISE ISE Options Data Logging LOg Options

Resolution rES 1,2,3 Delete on Download dEL nO,yES

Units UnIt M,mG/L,per,PPb,nOnE General gEn Options

Concentration Range rAng LOw,HIgH Manual Temp dEgC -5.0 to 105, 25.0

Auto-Blank Correction nLIn AUtO,OFF Stirrer Speed StIr OFF,1,2,3,4,5,6,7

Conductivity COnd Options Auto-Shut Off AUtO OFF, On

Temp Comp Selection tC OFF,LIn,nLF GLP gEn Options

Temp Comp Coefficient COEF 0.0 - 10.0, 2.1 Enable Methods Set OFF,On

TDS Factor tdSF 0.00 - 10.0, 0.49 Enable/Set Password PASS 0000 - 9999

AutoCal Cell Constant CELL 0.001 - 199.0, 0.475 Time & Date dAtE Options

Reference Temp trEF 15,20,25 Hour Setting HOUr 00 - 23, 12

Cell Type & Ranging tyPE PLnr,Std,1,2,3,4,5,6,7 Minute Setting tInE 00 - 59

Dissolved Oxygen DO Options Date Format tyPE mdy,dmy

%Sat Resolution rES 1,0.1 Year Setting yEAr 00 - 99, 04

mg/L Resolution rES 0.1,0.01 Month Setting dAtE 01 - 12

Pressure Comp bAr AUtO,mAn Day of the Month dAy 01 - 31

Manual Pressure PrES 450.0 - 850.0, 760.0 RS-232 RS232 Options

Salinity Correction SAL AUtO,mAn Baud Rate bAUd 1200,2400,4800,9600

Salinity Factor SALF 0 & 2 - 45 Printout Format OUtF Prnt,COmP

Calibration Type CALt Air,H2O,mAn,SEt0

Cal Alarms dUE Options

pH Cal Alarm PH 0000 - 9999

ORP Cal Alarm OrP 0000 - 9999

ISE Alarm ISE 0000 - 9999

Cond Cal Alarm COnd 0000 - 9999

DO Cal Alarm DO 0000 - 9999

Page 28: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIII-8 Orion Star Series Meter

Operational Overview

CalibrationsBefore Beginning a Calibration

Prior to calibration, the electrode or probe should be prepared according to the instructions received with the electrode. The calibration standards or buffers or cal sleeve should also be prepared prior to the calibration steps below. The electrode(s) or probes should be properly connected to the meter. Some basic information on proper calibration technique and calibration options for each measurement is found in the appropriate section.

General Navigation for Selecting Calibration to be Performed

1. In measurement mode press the RIGHT ARROW key until the arrow on the left of the screen is pointed toward the measurement line associated with the measurement your calibrating (pH, ORP, ISE – Top Line, Conductivity – Middle Line, DO – Bottom Line)

2. Press the UP or DOWN scroll key until the appropriate measurement icon is lit for the measurement you wish to calibrate.

a. Top Line: pH icon – pH Calibration

b. Top Line: RelmV icon – ORP Calibration

c. Top Line: ISE icon or no icon– ISE Calibration

d. Middle Line: µS/cm, mS/cm, mg/L, MW-cm, or ppt icon – Conductivity Calibration

e. Bottom Line: %Sat or mg/L icon – DO Calibration

f. Bottom Line: No icon – Barometer Calibration

3. Then press the CALIBRATE key to being the selected calibration.

Page 29: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

III-9Orion Star Series MeterThermo Electron Corporation

Operational Overview

Measurements Selecting the Measurement Modes

The arrow on the left of the screen indicates which line is selected for mode selection, one button calibration or setup. Press the UP or DOWN arrow keys to change the measurement mode of the selected line. This can also be used to turn off this measurement line. Press the RIGHT arrow key to move the arrow selecting the line to the next line.

Mode Selections:

Top Line: pH mV Rel mV ISE Off

Middle Line: µS/cm or mS/cm for conductivity mg/L for TDS ppt for Salinity MegOhms-cm for resistivity Off

Bottom Line: %Sat for DO percent saturation mg/L for DO concentration Barometric Pressure Off

Page 30: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIII-10 Orion Star Series Meter

Operational Overview

Measurement Setup Menu

The following table describes the Measurement Setup Options:

Measurement Setup

Description Display Range of Values Default Value

Method Specific

Continuous, Timed, or Auto-Read Measurement Selection

rEAdtyPECont

Cont, AUtO, timETimed, Continuous, or Auto-Read; Auto Prints and Datalogs in AUTO & Timed Modes

Auto Yes

Timed Reading Setting rEAdtInE00:00

00:05 - 99:61Timed Readings in Minutes& Seconds

01:00 Yes

Continuous, Timed, or AUTO-READ™ Selection: This defines the measurement operation of the instrument.

• In Continuous mode the instrument is constantly making measurements and updating the screen. To print & Log the measurement in this mode press the MEASURE key.

• In Timed mode the instrument is constantly making measurements and updating the display but it prints & logs the data at the selected time interval.

• In AUTO-READ mode the instrument starts a measurement whenever the MEASURE key is pressed. Once the reading becomes stable it auatomatically prints & logs the measurement and freezes the display. On the Benchtop meter it will also automatically operate the stirrer if a stirrer speed is set. It will start the stirrer when the MEASURE key is pressed and stop the stirrer when the reading becomes stable.

Timed Reading Interval Setting: This screen will only appear if Timed Readings has been selected. This is where you enter the minutes and seconds between readings. The minimum time interval is 5 seconds.

Page 31: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

III-11Orion Star Series MeterThermo Electron Corporation

Operational Overview

How to Change the Measurement Setup Options

To access the Measurement Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

• Press the UP & DOWN arrow to scroll through the Setup Menu until rEAd is displayed on the Top line.

• Press the RIGHT ARROW key to select the rEAd setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

Measurement Setup Option Menu Screen.

Press the UP & DOWN arrow to select the rEAd setup Options as listed below:

tyPE = Type of measurement, Continuous, Timed, or AUTO-READ tImE = The Timed Measurement Interval

• Press the RIGHT ARROW key to select the option editing line (Bottom Line). Press the MEASURE key to return to measurement mode.

• Meaurement Type Selection Menu Screen.

Press the UP & DOWN arrow to select the Measurement Type options as listed below:

Cont = Continuous Measurements AUtO = AUTO-READ Measurements tImE = Readings at timed intervals

Page 32: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIII-12 Orion Star Series Meter

Operational Overview

• Press the RIGHT ARROW key to accept the selection and select the Measurement Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Timed Readings Setting Menu Screen.

This screen will only appear when timed readings have been selected for the measurement type. Enter the Minutes & Seconds between timed readings by Pressing the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

00.05 to 99.59

• Press the RIGHT ARROW key to accept the selection and select the Measurement Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 33: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter IV pHSetup MenuThe following table describes the pH Setup Options:

pH Setup

Description Display Range of Values Default Value

Method Specific

pH Resolution rES O,1, 0.01, 0.001 0.01 Yes

pH Buffer Set bUF USA, EUrO USA Yes

How to Change the pH Setup OptionsTo access the pH Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

IV-1Orion Star Series MeterThermo Electron Corporation

Page 34: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIV-2 Orion Star Series Meter

pH

• Press the UP & DOWN arrow to scroll through the Setup Menu until PH is displayed on the Top line.

• Press the RIGHT ARROW key to select the pH setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

pH Setup Option Menu Screen.Press the UP & DOWN arrow to select the pH setup Options as listed below:

rES = Resolution bUF = Auto-Buffer Set

• Press the RIGHT ARROW key to select the option editing line (Bottom Line). Press the MEASURE key to return to measurement mode.

pH Resolution Selection Menu Screen.

Press the UP & DOWN arrow to select the pH resolution options as listed below:

0.001, 0.01 or 0.1

• Press the RIGHT ARROW key to accept the selection and select the pH Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• pH Auto-Buffer Set Selection Menu Screen.

Press the UP & DOWN arrow to select the pH resolution options as listed below:

USA = USA (USA/Orion Buffers) 1.68, 4.01, 7.01, 10.01, 12.46 EUrO = EUrO (Euro Buffers) 1.68, 4.01, 6.86, 9.15,

• Press the RIGHT ARROW key to accept the selection and select the pH Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 35: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IV-3Orion Star Series MeterThermo Electron Corporation

pH

pH Calibration for 1-5 Points1. Prepare the electrode for use according to the instructions received with the

electrode.

2. Select the buffer set of NIST or DIN being used for the proper buffer recognition to occur.

Note: the calibration points should bracket the expected pH range of the samples to be measured. These should be not less that 1 pH unit apart and not more than 4 pH units apart. ▲

GLP indicates that at least a two-point calibration should be performed before the first samples of the day are analyzed and a single point calibration should be performed every two hours.

Fresh buffers should be used each time you calibrate.

Care should be taken when rinsing the electrode with deionized water and GENTLY BLOTTING all remaining water drops with a lint-free tissue prior to moving the electrode to the next beaker. DO NOT wipe a glass or epoxy electrode body with a tissue as static build up will result.

3. Place the electrode and ATC probe into the first buffer. The neutral buffer (pH 7.01 or 6.86) is usually recommended to be used first followed from lowest to highest pH.

Page 36: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIV-4 Orion Star Series Meter

pH

When the pH icon stops flashing 1. Press MEASURE key to accept the Autobuffer recognition temperature-corrected

value for the buffer and return to measurement mode.

a. Before returning to measurement mode the slope will be displayed. SLP appears in the lower field while the actual electrode slope, in percent, appears in the main field.

b. For a 2 or more point calibration it will automatically proceed to measurement mode after 3 seconds. If only a 1 point calibration was performed then the slope will be displayed and flash to allow you to edit the slope. After editing the slope press the MEASURE key to proceed to measurement. The probe icon is displayed above the main field. then press the MEASURE key to proceed to measurement mode.

2. Press the CALIBRATE key to accept Autobuffer recognition as the 1st calibration point & start the 2nd calibration point when prompted for the next buffer.

3. Use the UP & DOWN arrow keys and the DIGIT/DECIMAL keys to edit the value for a manual buffer entry. Then follow step 1 or 2.

4. Continue measuring for each calibration point, rinsing the electrode between each buffer until the desired calibration points are reached. Press MEASURE key to escape from the calibration mode and start sample readings.

After calibration, the electrode and meter are ready to start taking sample measurements. Make sure that the instrument is in measurement mode and that the pH icon is lit. If the probe icon is not lit press the MEASURE key to return to measurement mode. If the pH icon is not lit, press the RIGHT ARROW key until the arrow on the left of the display is pointing at the top line. Then press the UP arrow key until the pH icon lights. You are now ready to start measuring.

Page 37: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IV-5Orion Star Series MeterThermo Electron Corporation

pH

pH Measurements1. Rinse the electrode, as well as ATC probe and stirrer if being used, in deionized

water. Blot dry with a lint-free tissue. Do not wipe glass electrodes as static charge will build up on the sensor and need to discharge prior to the measurement.

2. Place the electrode in your sample.

a. If you are in continuous measurement mode, the instrument will start measuring immediately. If you are using the benchtop meter & the stirrer control is enabled then pressing the STIRRER button will start the stirrer. The measurement icon will flash until the reading is stable. Once the reading is stable you can Log & Print the measurement by pressing the MEASURE key. If you are recording the data into a notebook, record the pH as well a the temperature at which the pH reading is taken. If the stirrer is on, press the stirrer button to turn off the stirrer before removing the stirrer from the sample.

b. If you are in AUTO-READ mode then press the MEASURE key to start the measurement. Once the ready is stable the meter will automatically Log & Print the results and freeze the display. If the stirrer is enabled, then the stirrer will turn on when the MEASURE key is pressed and turn off once the reading has stabilized.

With the AutoStir BOD Probe pressing the button on the probe will start the AUTO-READ measurement.

c. If you are in timed measurement mode then the meter will start making measurements as soon as it goes into measurement mode at the frequency selected in setup. It will automatically Log & Print each measurement. If you are using the benchtop meter & the stirrer control is enabled then pressing the STIRRER button will start the stirrer. Pressing the stirrer button again will turn off the stirrer.

3. Remove the electrode from the sample and rinse with deionized water, then place it in your next sample and repeat step 2.

4. Once your finished measuring samples, rinse the electrode with deionized water and blot dry. Consult your electrode manual for proper electrode storage instructions.

Page 38: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIV-6 Orion Star Series Meter

Page 39: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter V ISESetup MenuThe following table describes the ISE Setup Options:

ISE Setup

Description Display Range of Values Default Value

Method Specific

ISE Resolution ISErES1

1, 2, 3ISE Resolution inSignificant Digits

1 Yes

ISE Units ISEUnItPPb

M, mG/L, per, PPb, nOnEISE Units Displayed

PPb Yes

ISE Calibration StandardConcentration Range

ISErAngHIgH

LOw, HIgHStability criteria usedduring ISE Calibration

HIgH Yes

ISE Auto-Blank Correction ISEnLInAUtO

AUtO, OFFEnables or DisablesCalibration Auto-Blank

AUtO Yes

ISE Resolutions: This selects the number of significant digits displayed in measurement mode. In addition the resolution selection is used by SMART STABILITY & SMART AVERAGING algorithms to optimize the accuracy, precision, and response of the ISE measurement.

ISE Units: This selects the ISE Units to be displayed in Measurement & Calibration. Both measurement and calibration must be performed in the same units.

V-1Orion Star Series MeterThermo Electron Corporation

Page 40: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationV-2 Orion Star Series Meter

ISE

ISE Calibration Standard Concentration Range: This selects the stability criteria to be used in the ISE calibration. For most calibrations the HIGH setting is the best option. But to maximize the accuracy of low concentration measurements using low concentration calibration standards the LOW setting can improve performance by giving the sensor longer to stabilize in the calibration standard.

ISE Auto-Blank Correction: This meter uses Thermo’s patented Auto-Blank algorithm to compensate for sensor non-linearities in low-level measurements. But since these require the solution of a set of non-linear equations that can only be calculated numerically, the user cannot analytically verify the calibration. In applications were that is a requirement the user can choose the turn the Auto-blank correction OFF.

How to Change the ISE Setup OptionsTo access the ISE Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

• Press the UP & DOWN arrow to scroll through the Setup Menu until ISE is displayed on the Top line.

• Press the RIGHT ARROW key to select the pH setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

Page 41: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

V-3Orion Star Series MeterThermo Electron Corporation

ISE

ISE Setup Option Menu Screen.Press the UP & DOWN arrow to select the ISE setup Options as listed below:

rES = Resolution UnIt = ISE Measurement units rAng = ISE Measurement range nLIn = Non-linear Blank Correction

• Press the RIGHT ARROW key to select the option editing line (Bottom Line). Press the MEASURE key to return to measurement mode.

• ISE Resolution Selection Menu Screen.

• Press the UP & DOWN arrow to select the ISE resolution (Sig. Digits) Options as listed below:

1, 2 or 3

• Press the RIGHT ARROW key to accept the selection and select the ISE Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• ISE Measurement Units Selection Menu Screen.

• Press the UP & DOWN arrow to select the ISE measurement units as listed below:

m = Molar mgL = mg/L PEr = Percent PPb = Parts per Billion nOnE = No units

• Press the RIGHT ARROW key to accept the selection and select the ISE Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 42: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationV-4 Orion Star Series Meter

ISE

• ISE Measurement Range Selection Menu Screen.

• Press the UP & DOWN arrow to select the ISE measurement Range as listed below:

LOw = Low Range HIgH = High Range

• Press the RIGHT ARROW key to accept the selection and select the ISE Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• ISE Calibration Blank Selection Menu Screen.

• Press the UP & DOWN arrow to select the ISE measurement Range as listed below:

AUTO = Auto blank correction OFF = No Blank Correction

• Press the RIGHT ARROW key to accept the selection and select the ISE Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

ISE Calibration for 1-5 Points1. Prepare the electrode and standards for use according to the instructions

received with the electrode.

2. The standards should be prepared in the same ISE units as required by the sample results. It is preferable to use serial dilutions with volumetric glassware to obtain the different concentration levels.

Note: Any reagents such as ionic strength adjustors, should be added in equal amounts to each sample and standard. ▲

The calibration points should bracket the expected concentration range of the samples and must cover a range of at least ten times greater in concentration.

Page 43: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

V-5Orion Star Series MeterThermo Electron Corporation

ISE

Fresh aliquots of sample should be used at each calibration.

Prior to each day’s measurements, a two or more point calibration should be performed and a minimum one point calibration should be performed every two hours for the greatest accuracy.

Recommended slopes are given with electrode instructions.

It is recommended that the calibration and sample measurements be taken at the same temperature for greater accuracy and reproducibility.

3. Place the electrode and ATC probe into the least concentrated standard for the first calibration point.

4. Rinse well with deionized water and gently blot the electrode with a lint-free tissue to remove water drops before moving the electrode to the next beaker. Continue from the next lowest concentration up to the highest point to reduce the potential effects of carryover.

Note: For greater low-level measurement accuracy, a new low-level ISE calibration option is available. An additional rinse of the electrode with the next standard or sample solution is recommended for any low level measurements. ▲

Autoblank CorrectionBlank correction occurs automatically when calibrating with three or more standards. This feature automatically calculates and corrects for background levels of the species of interest.

Auto-Blank correction can be deactivated in ISE setup as described the the ISE setup instructions.

1. Press the UP or DOWN scroll key until the ISE icon is lit.

a. Make sure the correct ISE units are displayed. If not go to the ISE setup menu and set the correct ISE units.

Page 44: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationV-6 Orion Star Series Meter

ISE

When the ISE icon stops flashing:1. Press the CALIBRATE key.

2. Rinse the electrode(s) and place into the standard.

3. When the electrode is stable, the ISE icon will stop flashing and the ISE value based on the previous calibration will be displayed.

4. Use the UP & DOWN arrow and the DECIMAL/DIGIT keys to edit the displayed concentration to the correct value of your standard then press CALIBRATE.

Note: The standard value is stored and the meter automatically switches to standard two as indicated by the “CAL.2” on the display in the lower field. ▲

5. If doing fewer than five calibration points, when you are entering your last calibration point you should press the MEASURE key instead of the CAL key to accept the value and return to the measurement mode.

If you are doing a manual calibration, go above to step #4 to edit the value of your manual standard entry.

Note: Before returning to measurement mode the slope will be displayed. For a 2 or more point calibration it will automatically proceed to measurement mode after 3 seconds. On a 1 point calibration it will allow you to edit the slope then press the MEASURE key to proceed to measurement mode. ▲

6. Press the CALIBRATE key to accept the value as the 1st calibration point & start the 2nd calibration point.

7. Continue measuring for each calibration point, rinsing the electrode between each standard until the desired calibration points are reached. Press measure to escape from the calibration mode and start sample readings.

8. Connect the electrode(s) to meter. Choose standards that will bracket the expected sample concentration.

9. In measurement mode press the RIGHT ARROW key until the arrow on the left of the screen is pointed toward the top display line.

Page 45: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

V-7Orion Star Series MeterThermo Electron Corporation

ISE

10. Repeat steps 5 and 6 for each standard.

11. After the standard value for the last standard has been entered, the electrode slope will be displayed automatically. SLP appears in the lower field while the actual electrode slope, in percent, appears in the main field. If this was a 2 or more point calibration then the meter automatically advances to the measure mode. If only a 1 point calibration was performed then the slope will be displayed and flashing allowing the user to edit the slope. After editing the slope press the MEASURE key to proceed to measurement. The probe icon is displayed above the main field.

12. Rinse electrode(s) and place into sample. Record ISE directly from the main meter display when the ISE icon stops flashing indicating a stable value. Temperature is displayed in the top left corner of the display.

ISE MeasurementAfter calibrating the electrode the meter is ready to start making measurements. Make sure that the instrument is in measurement mode (the probe icon is lit) and that the ISE icon is lit. If the probe icon is not lit press the MEASURE key to return to measurement mode. If the ISE icon is not lit, press the RIGHT ARROW key until the arrow on the left of the display is pointing at the top line. Then press the UP arrow key until the ISE icon lights. You are now ready to start measuring.

1. Rinse the electrode in deionized water.

2. Place the electrode in your sample.

a. If you are in continuous measurement mode, the instrument will start measuring immediately. If you are using the benchtop meter & the stirrer control is enabled then pressing the STIRRER button will start the stirrer. The ISE icon will flash until the reading is stable. Once the reading is stable you can Log & Print the measurement by pressing the MEASURE key. If the sitirrer is on press the stirrer button to turn off the stirrer before removing the stirrer from the sample.

Page 46: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationV-8 Orion Star Series Meter

ISE

b. If you are in AUTO-READ mode then press the MEASURE key to start the measurement. Once the ready is stable the meter will automatically Log & Print the results and freeze the display. If the stirrer is enabled then the stirrer will turn on when the MEASURE key is pressed and turn off once the reading has stabilized.

c. If you are in timed measurement mode then the meter will start making measurements as soon as it goes into measurement mode at the frequency set in setup. It will automatically Log & Print each measurement. If you are using the benchtop meter & the stirrer control is enabled then pressing the STIRRER button will start the stirrer. Pressing the stirrer button again will turn off the stirrer.

3. Remove the electrode from the sample and rinse with deionized water, then place it in your next sample and repeat step 2.

4. Once your finished measuring samples, rinse the electrode with deionized water and blot dry. Then consult your electrode manual for proper electrode storage instructions.

Page 47: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter VI mV/ORPRelative mV and ORP(EH) Calibration Note: mV is a raw input reading and therefore does not have a user calibration. ▲

Prepare the electrode and standards for use according to the instructions received with the electrode.

1. Connect the electrode to meter.

2. Place the Orion ORP Triode™ with built-in temperature probe or ORP electrode and ATC probe into the calibration standard. 3. Make sure that the instrument is in measurement mode and that the arrow on the leftof the display is pointing at the top line. Then press the UP arrow key until the RmV icon lights

4. Press CALIBRATE key.

5. When the electrode is stable, the RmV icon will stop flashing and the temperature corrected value for the Standard Hydrogen Electrode value for Orion ORP standard (967901 or 967961) will be displayed.

6. Press CALIBRATE or MEASURE key to accept the standard value and return to measurement mode.

If performing a manual calibration, use the UP & DOWN arrow keys and the DIGIT/DECIMAL keys to change the value, then press CALIBRATE or MEASURE key to accept the standard value and return to measurement mode. Use the temperature chart on the standard bottle or instruction sheet to enter the appropriate value for the temperature and reference type being used. Then follow step 1.

VI-1Orion Star Series MeterThermo Electron Corporation

Page 48: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVI-2 Orion Star Series Meter

mV/ORP

mV MeasurementThe instrument can measure the absolute millivolts from the pH, ISE, or ORP sensors. Make sure that the instrument is in measurement mode (the probe icon is lit) and that the arrow on the left of the display is pointing at the top line. Then press the UP arrow key until the mV icon lights. The mV will now be displayed. This mode is very helpful in troubleshooting because it displays the input signal being received from the electrode.

If the probe icon is not lit press the MEASURE key to return to measurement mode. If the mV icon is not lit, press the RIGHT ARROW key until the arrow on the left of the display is pointing at the top line. Then press the UP arrow key until the mV icon lights. You are now ready to make measurements.

1. Rinse the electrode in deionized water.

2. Place the electrode in your sample.

a. If you are in continuous measurement mode, the instrument will start measuring immediately. If you are using the benchtop meter & the stirrer control is enabled then pressing the STIRRER button will start the stirrer. The mV icon will flash until the ready is stable. Once the ready is stable you can Log & Print the measurement by pressing the MEASURE key. If the sitirrer is on press the stirrer button to turn off the stirrer before removing the stirrer from the sample.

b. If you are in AUTO-READ mode then press the MEASURE key to start the measurement. Once the ready is stable the meter will automatically Log & Print the results and freeze the display. If the stirrer is enabled then the stirrer will turn on when the MEASURE key is pressed and turn off once the reading has stabilized.

c. If you are in timed measurement mode then the meter will start making measurements as soon as it goes into measurement mode at the frequency set in setup. It will automatically Log & Print each measurement. If you are using the benchtop meter & the stirrer control is enabled then pressing the STIRRER button will start the stirrer. Pressing the stirrer button again will turn off the stirrer.

Page 49: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

VI-3Orion Star Series MeterThermo Electron Corporation

mV/ORP

3. Remove the electrode from the sample and rinse with deionized water, then place it in your next sample and repeat step 2.

4. Once your finished measuring samples, rinse the electrode with deionized water and blot dry. Then consult your electrode manual for proper electrode storage instructions.

Relative mV & ORP (EH) MeasurementAfter calibrating the electrode the meter is ready to start making measurements. Make sure that the instrument is in measurement mode (the probe icon is lit) and that the Rel mV icon is lit. If the probe icon is not lit press the MEASURE key to return to measurement mode. If the Rel mV icon is not lit, press the RIGHT ARROW key until the arrow on the left of the display is pointing at the top line. Then press the UP arrow key until the Rel mV icon lights. You are now ready to start measuring.

1. Rinse the electrode in deionized water.

2. Place the electrode in your sample.

a. If you are in continuous measurement mode, the instrument will start measuring immediately. If you are using the benchtop meter & the stirrer control is enabled then pressing the STIRRER button will start the stirrer. The Rel mV icon will flash until the ready is stable. Once the ready is stable you can Log & Print the measurement by pressing the MEASURE key. If the sitirrer is on press the stirrer button to turn off the stirrer before removing the stirrer from the sample.

b. If you are in AUTO-READ mode then press the MEASURE key to start the measurement. Once the ready is stable the meter will automatically Log & Print the results and freeze the display. If the stirrer is enabled then the stirrer will turn on when the MEASURE key is pressed and turn off once the reading has stabilized.

Page 50: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVI-4 Orion Star Series Meter

mV/ORP

c. If you are in timed measurement mode then the meter will start making measurements as soon as it goes into measurement mode at the frequency set in setup. It will automatically Log & Print each measurement. If you are using the benchtop meter & the stirrer control is enabled then pressing the STIRRER button will start the stirrer. Pressing the stirrer button again will turn off the stirrer.

3. Remove the electrode from the sample and rinse with deionized water, then place it in your next sample and repeat step 2.

4. Once your finished measuring samples, rinse the electrode with deionized water and blot dry. Then consult your electrode manual for proper electrode storage instructions.

Page 51: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter VII ConductivitySetup MenuThe following table describes the Conductivity Setup Options:

Conductivity Setup

Description Display Range of Values Default Value

Method Specific

Conductivity TemperatureCompensation Selection

CondtCLin

OFF, Lin, nLFSelects Temp Comp OFF,Linear, or Ultra-pure Water

Lin Yes

Conductivity LinearTemperature CompensationCoefficient Setting

CondCOEF2 . 1

0.0 - 10.0Linear Temp CompensationCoefficient in %/C

2.1 Yes

Conductivity TDS FactorSetting

CondtdSF0 . 49

0.00 - 10.0TDS factor

0.49 Yes

Conductivity Auto-Calibration Default CellConstant Setting

CondCELL0 . 475

0.001 - 199.0Cell Constant used forCond Auto-Standard

0.475 Yes

Conductivity TemperatureReference Selection

CondtrEF25

15, 20, 25Conductivity ReferenceTemperature

25 Yes

Conductivity Cell Type &Manual Ranging Selection

CondtyPEPLnr

PLnr, Std, 1, 2, 3, 4, 5, 6, 7Standard, Planer Cond CellOr sets Manual range 1-7

Std Yes

Conductivity Temperature Compensation: This menu allows you to select between turning temperature compensation OFF, selecting a linear temperature compensation, or a non-linear temperature compensation for natural and ultra-pure water.

VII-1Orion Star Series MeterThermo Electron Corporation

Page 52: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVII-2 Orion Star Series Meter

Conductivity

Conductivity Linear Temperature Compensation Coefficient: This menu will only appear when Linear temperature compensation has been selected.

The conductivity of a solution with a specific electrolyte concentration changes with temperature. The relationship of the change in conductivity as a function of temperature is described by a solution’s temperature coefficient. Temperature coefficients vary with each solution. Some examples are shown in Table 1. This Meter uses 2.1%/˚C as the default linear temperature coefficient, which is representative of many aqueous samples, to compensate for temperature changes. The meter has an adjustable temperature coefficient of 0.0 to 10.0%/˚C

Table 1 Typical Temperature Coefficients Between 25 and 50 ûC (Percent change of Conductivity per ûC)

Solution %/˚CUltrapure Water 4.55Salt (NaCl) Solutions 2.125% NaOH 1.72Dilute Ammonia 1.8810% HCl 1.325% Sulfuric Acid 0.9698% Sulfuric Acid 2.84Sugar Syrup 5.64

Conductivity TDS Factor: Total Dissolved Solids (TDS) refers to the dissolved inorganics in a solution. These dissolved inorganics will carry a current and will be reflected as a conductivity reading. Since a relationship can be shown between conductivity and TDS, a conductivity reading can be used as an estimation of the presence of inorganics.

The standard method of determining TDS (Total Dissolved Solids) is by evaporating the sample to dryness at 180 ˚C and weighing the residue. After evaporating the sample, one would have to calculate a specific “factor” for the solution by taking the TDS value obtained in weighing and dividing it by the conductivity of the sample. This factor would then be multiplied by the subsequent conductivity readings to obtain the TDS value of the sample as referenced in Standard Methods for Water and Wastewater.

Page 53: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

VII-3Orion Star Series MeterThermo Electron Corporation

Conductivity

This Meter allows the operators to enter their own TDS factor for the calculation of the TDS value in mg/L, with a selectable range from 0.01 to 10.00 and a default value of 0.49. The default value for TDS factor on the meter is 0.49, which is a reasonable estimate of a TDS factor for natural water applications.

Conductivity Auto-Calibration Default Cell Constant: This allows to determine the Auto-standard used. In order for the instrument to Auto-recognize a conductivity standard it must know the cell constant within a factor of 10. So it’s needs am approximate starting value. If the real cell constant was 1.0 then having an estimate of any value between 0.5 and 5.0 would be sufficient to determine which auto-standard was being used and therefore determine the standards temperature corrected value for calibration. The instrument Auto-recognizes only three Orion conductivity standards (1413 μS/cm ,100 μS/cm, and 12.9 mS/cm).

Conductivity Temperature Reference: By definition, temperature compensated conductivity of a solution is the conductivity which that solution exhibits at the reference temperature. The temperature is chosen to be either 25 ˚C, 20 ˚C, or 15 ˚C. A measurement made at the reference temperature needs no compensation. The closer the sample is to the reference temperature, the smaller the error will be if the meter temperature coefficient is incorrect. The meter automatically compensates for temperature changes based on the temperature coefficient and the reference temperature when a temperature measurement is simultaneously made.

Conductivity Cell Type & Manual Ranging: This allows you to select a Standard cell where the conductivity plate are parallel to each other or Planar where the conductivity plate are on the same plane. Almost all conductivity cells are of the Standard arrangement. This also allows the user to fix the meter in one range. This can be useful with extreamly long cables where cable capacitance could cause non-linearities due to the drive frequency change between ranges.

Page 54: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVII-4 Orion Star Series Meter

Conductivity

Conductivity Setup Option Menu Screen.Press the UP & DOWN arrow to select the Conductivity setup Options as listed below:

tC = Temperature Compensation (Linear or nLF) COEF = Temperature coefficient for Linear temp compensation

- Screen only shown if Linear Temp Comp Selected

tdSF = TDS Factor for Total Dissolved Solids measurement CELL = Conductivity Cell Constant rEFt = Reference temperature for temperature compensation tYPE = Conductivity cell type (Planar or Conventional)

• Press the RIGHT ARROW key to select the option editing line (Bottom Line). Press the MEASURE key to return to measurement mode.

• Temperature Compensation Selection Menu Screen.

Press the UP & DOWN arrow to select the Conductivity Temperture compensation selection as listed below:

OFF = No temperature compensation LIn = Linear temperature compensation nLF = Natural Water Non-Linear Temperature Compensation

• Press the RIGHT ARROW key to accept the selection and select the Conductivity Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Temperature Coefficient Setting Menu Screen.

Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

0.0 to 10.0 %/˚C

• Press the RIGHT ARROW key to accept the selection and select the Conductivity Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 55: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

VII-5Orion Star Series MeterThermo Electron Corporation

Conductivity

• TDS Factor Setting Menu Screen.

Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

0.00 to 10.00

• Press the RIGHT ARROW key to accept the selection and select the Conductivity Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Conductivity Cell Constant Setting Menu Screen.

Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled and the place of the decimal poit.

0.001 to 199.0

• Press the RIGHT ARROW key to accept the selection and select the Conductivity Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Temperature Compensation Selection Menu Screen.

Press the UP & DOWN arrow to select the Conductivity Temperture compensation selection as listed below:

15 = 15 degrees C 20 = 20 degrees C 25 = 25 degrees C

• Press the RIGHT ARROW key to accept the selection and select the Conductivity Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Cell Type & Manual Ranging Selection.

Page 56: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVII-6 Orion Star Series Meter

Conductivity

Press the UP & DOWN arrow to select the Conductivity Temperture compensation selection as listed below:

PLnr = Planar Conductivity Cell Std = Standard Conductivity Cell 1 = Manual Range 1 2 = Manual Range 2 3 = Manual Range 3 4 = Manual Range 4 5 = Manual Range 5 6 = Manual Range 6 7 = Manual Range 7

• Press the RIGHT ARROW key to accept the selection and select the Conductivity Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

How to Change the Conductivity Setup OptionsTo access the Conductivity Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

• Press the UP & DOWN arrow to scroll through the Setup Menu’s until COND is displayed on the Top line.

• Press the RIGHT ARROW key to select the Conductivity setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

Page 57: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

VII-7Orion Star Series MeterThermo Electron Corporation

Conductivity

Conductivity Calibration for 1-5 Points1. Select a conductivity probe with an appropriate cell constant for the expected

samples to be measured. A flow cell is recommended for greater low level measurement accuracy.

2. Prepare the conductivity probe for use according to the probe instructions.

3. Verify the correct reference temperature to be used for the measurement has been selected in the conductivity setup menu.

4. All conductivity calibrations are performed in mS/cm or μS/cm. The other measurement modes of conductivity are calculated using the cell constant determined from the calibration and any other necessary parameters selected in the setup menu.

5. Place the probe into the least concentrated standard for the first calibration point. When the mS/cm or μS/cm icon stops flashing in the measurement mode with the arrow pointed at the conductivity line press the Calibrate key.

6. The screen will display the cell constant on the Top line and the conductivity on the middle line. To perform a manual calibration use the UP & DOWN arrow keys to change the value of the flashing digit on the cell constant. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Edit the cell constant until the conductivity measurement matches the conductivity of the standard at the actual measurement temperature.

a. Note: If you do not start the scrolling within 5 seconds the screen will automatically proceed to the AUTOCal/DirectCal screen.

b. In the AutoCal/DirectCal screen when the electrode is stable the mS/cm or μS/cm icon will stop flashing and the temperature corrected value for the standard is displayed. Press CALIBRATE to accept the AutoCal value and proceed to the next calibration point or if performing a DirectCal calibration use the UP & DOWN arrow and DECIMAL/DIGIT keys to change the conductivity value to the actual conductivity of the standard at the measurement temperature then press CALIBRATE key.

Page 58: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVII-8 Orion Star Series Meter

Conductivity

c. The standard value is stored and the meter automatically switches to standard two as indicated by the “CAL2” on the display in the lower field.

7. Rinse well with deionized water and gently blot the electrode with a lint-free tissue to remove water drops before moving to the next beaker or field sample.

8. Continue from the lowest to the highest concentration standard to reduce the potential carryover effects.

9. If doing less than five calibration points and if this is your last calibration point press the MEASURE key instead pf the CALIBRATE key to accept the value and return to the measurement mode.

10. Repeat steps 5 and 6 for each standard.

11. After the standard value for the last standard has been entered the cell constant will be displayed automatically CELL appears in the lower field while the actual CELL constant appears in the main field. Then the meter automatically advances to the measure mode. The probe icon is displayed above the main field.

To Take a Conductivity MeasurementRinse electrode and place into sample. Record conductivity directly from the main meter display when the mS/cm or μS/cm icon stops flashing indicating a stable value. Temperature is displayed in the top left corner of the display.

Page 59: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter VIII Dissolved OxygenSetup MenuThe following table describes the DO Setup Options:

Dissolved Oxygen Setup

Description Display Range of Values Default Value

Method Specific

Dissolved Oxygen%Saturation Resolution

DOrES1

1, 0.1DO % SaturationResolution

0.1 Yes

Dissolved OxygenConcentration Resolution

DOrES0.1

0.1, 0.01DO Concentration “mg/L” resolution

0.01 Yes

Dissolved OxygenBarometric PressureCompensation Selection

DObArAUtO

AUtO, mAnSelects Internal Barometeror Manual Pressure

AUtO Yes

Dissolved Oxygen ManualBarometric PressureSetting

DOPrES760.0

450.0 - 850.2Manual PressureCompensation Value

760.0 Yes

Dissolved Oxygen SalinityCorrection Selection

DOSALAUtO

AUtO, mAnSelects Method of SalinityCorrection; Meters w/Cond Only

AUtO Yes

Dissolved Oxygen Manual Salinity Correction Factor

DOSALF0

0 & 2 - 45Manual Salinity CorrectionFactor

0 Yes

Dissolved OxygenCalibration Type Selection

DOCALtAir

Air, H2O, mAn, SEt0Air = Water Sat Air, H2O = Air Sat Water, mAn = Manual, SEt0 = Zero Cal

Air Yes

VIII-1Orion Star Series MeterThermo Electron Corporation

Page 60: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVIII-2 Orion Star Series Meter

Dissolved Oxygen

DO % Saturation Resolution: This selects the displayed resolution in measurement mode. In addition the resolution selection is used by Thermo Orion’s SMART STABILITY & SMART AVERAGING algorithms to optimize the accuracy, precision, and response of the DO measurement.

DO (mg/L) Concentration Resolution: This selects the displayed resolution in measurement mode. In addition the resolution selection is used by Thermo Orion’s SMART STABILITY & SMART AVERAGING algorithms to optimize the accuracy, precision, and response of the DO measurement.

DO Barometric Pressure Compensation Selection: This selects whether to use the instruments internal barometer for pressure compensation or use a manually set pressure. In general we recommend using the internal barometer, but you may wish to use manual pressure compensation if you are measuring DO with a submerged probe or in a pressurized vessel where the probe is exposed to higher pressures then the meter.

DO Manual Pressure Setting: This screen will only appear if Manual Barometric Pressure compensation has been selected. This is where you enter the pressure to use for pressure compensating the DO measurement. The pressure in entered in mm of Hg (Mercury). See the following table to convert from other units of pressure.

hPa (mBar) mm Hg (torr) Inches Hg lbf/in2 (PSI)

1 hPa (mBar) 1 0.75006 0.02953 0.014504

1 mm Hg (torr) 1.3332 1 0.039370 0.019337

1 Inches Hg 33.864 25.400 1 0.491154

1 lbf/in2 (PSI) 68.9475 51.7149 2.03602 1

Example: To convert 14.1 PSI to mm Hg, multiple 14.7 by 51.7149 to get 750.21 mm Hg.

DO Salinity Correction Selection: This screen will only be shown on meters that have conductivity as well as DO. The Automatic setting takes the conductivity reading to calculate the salinity correction factor and automatically applies that to the DO measurement. A conductivity probe must be used in addition to the DO probe for automatic salinity correction. The manual setting allow the user to enter a manual salinity correction factor. Select the Manual setting and set it to zero to disable salinity correction.

Page 61: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

VIII-3Orion Star Series MeterThermo Electron Corporation

Dissolved Oxygen

DO Manual Salinity Correction Factor Setting: This factor adjusts the dissolved oxygen concentration in mg/L based on the samples salinity. Use the following table to select an appropriate salinity factor. A typical value for Sea Water is 35, but does vary considerably based on location.

Conductivity Salinity Conductivity Salinity Conductivity Salinity in mS/cm Value in mS/cm Value in mS/cm Value

5 3 20 13 35 25

6 4 21 14 36 25

7 4 22 15 37 26

8 5 23 15 38 27

9 6 24 16 39 28

10 6 25 17 40 29

11 7 26 18 42 30

12 8 27 18 44 32

13 8 28 19 46 33

14 9 29 20 48 35

15 10 30 21 50 37

16 10 31 22 52 38

17 11 32 22 54 40

18 12 33 23 56 42

19 13 34 24 Salinity depending on the conductivity at 20 ˚C

ppt = Parts per Thousand Salinity

This table was calculated from the International Oceanographic Tables.

International Oceanographic Tables, Vol. 1, National Institute of Oceanography of Great Britian, Womley, Godaming, Surrey, England and Unesco, Paris 1971

DO Calibration Type: This allows you to select the type of DO calibration that will be performed when the CALIBRATE key is pressed. AIR is a water saturated air calibration. This is generally the simpliest and most accurate calibration method. H2O is air saturated water. MAN is manual calibration which can be used to calibrate the sensor to the value achieved by a Winkler titration. Set0 is a zero calibration, which is used for very low level DO measurements. Not generally required unless you are making measurements below 5% Saturation or 0.5 mg/L.

Page 62: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVIII-4 Orion Star Series Meter

Dissolved Oxygen

How to Change the DO Setup OptionsTo access the pH Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

• Press the UP & DOWN arrow to scroll through the Setup Menu’s until DO is displayed on the Top line.

• Press the LINE SELECT key to select the DO setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

DO Setup Option Menu Screen.Press the UP & DOWN arrow to select the DO setup Options as listed below:

rES = % Saturation Resolution (%Sat icon is lit) rES = Concentration Resolution (mg/L icon is lit) bAr = Automatic/Manual Pressure Compensation PrES = Manual Pressure Pressure (This screen will only appear if Manual Pressure compensation has been selected) SAL = Automatic/Manual Salinity Compensation SALF = Manual Salinity Factor (This screen will only appear if Manual Salinity Compensation has been selected) CALt = Calibration Type

Page 63: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

VIII-5Orion Star Series MeterThermo Electron Corporation

Dissolved Oxygen

• Press the LINE SELECT key to select the option editing line (Bottom Line). Press the MEASURE key to return to measurement mode.

• % Saturation Resolution Selection Menu Screen.

• Press the UP & DOWN arrow to select the DO resolution options as listed below:

1 or 0.1

• Press the LINE SELECT key to accept the selection and select the DO Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Concentration Resolution Selection Menu Screen.

• Press the UP & DOWN arrow to select the DO resolution options as listed below:

0.1 or 0.01

• Press the LINE SELECT key to accept the selection and select the DO Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Pressure Compensation Selection Menu Screen.

• Press the UP & DOWN arrow to select the DO resolution options as listed below:

AOtO = Automatic Press. Comp. using the Internal Barometer mAn = Manual Press. Comp. Using entered value.

• Press the RIGHT ARROW key to accept the selection and select the DO Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Manual Pressure Setting Menu Screen.

Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

450.0 to 850.0

Page 64: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVIII-6 Orion Star Series Meter

Dissolved Oxygen

• Press the LINE SELECT key to accept the selection and select the DO Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Salinity Correction Selection Menu Screen.

Press the UP & DOWN arrow to select the DO resolution options as listed below:

AOtO = Automatic Salinity Correction using conductivity mAn = Manual Salinity Correction using entered value.

• Press the RIGHT ARROW key to accept the selection and select the DO Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Manual Salinity Correction Setting Menu Screen.

Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

0 & 2 to 45

• Press the RIGHT ARROW key to accept the selection and select the DO Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Calibration Type Selection Menu Screen.

Press the UP & DOWN arrow to select the DO resolution options as listed below:

AIr = Water Saturated Air Calibration H2O = Air Saturated Water Calibration mAn = Manual/Winkler Cal (Cal to Known Concentration) SEt0 = Zero calibrate the sensor in a Zero Oxygen Solution

• Press the RIGHT ARROW key to accept the selection and select the DO Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 65: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

VIII-7Orion Star Series MeterThermo Electron Corporation

Dissolved Oxygen

Dissolved Oxygen Calibration Prior to calibration the probe must be prepared and polarized. The desired options have to be chosen from the DO Setup Menu, and the active line has to correspond to the DO mode (bottom line).

Assemble and prepare the electrode for use according to the instructions received with the electrode. Since DO is dependent upon temperature, barometric pressure and salinity ensure the proper settings have been selected in the DO setup. The Star series DO probes have built-in themistors to accurately measure temperature. The Star series meters with conductivity measurement capability can use the real-time measured salinity of the sample for the salinity correction.

The simplest and most common calibration is an air calibration. Moisten the sponge or absorbent cloth in the calibration sleeve with water and insert the probe into the sleeve above, but not into, the sponge. This water-saturated air has 100% humidity and is a standard for calibration. For BOD measurements, this calibration can be performed in a BOD bottle. An air saturated water calibrated probe can be used for measurement of dissolved oxygen in water. For low oxygen levels a second zero calibration point is often required, although some probes generate zero current at zero oxygen, making a second calibration point unnecessary.

A Winkler calibration involves performing a manual Winkler titration and then using that sample as a standard. The titration oxygen level result is entered in a Winkler calibration as the DO standard value. This correlates the meter input to the Winkler titration. Although the inherently less accurate titration method adds error when the calibration is set to these test results, some customers are required to correlate their result to Winkler. This does allow them to take save time and have greater reproducibility by taking most reading using an electrochemical method.

Water calibration requires a standard of water saturated with oxygen and is the least commonly used calibration method.

Page 66: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVIII-8 Orion Star Series Meter

Dissolved Oxygen

There are 4 ways to calibrate the probe.

Air Calibration: The calibration is performed in water saturated air using the air calibration sleeve. This is the easiest and most accurate way to calibrate the probe.

H2O “Water” Calibration: The calibration is performed in a water sample that is 100% saturated with air.

Manual Calibration: The calibration is performed using a water sample with a known concentration of dissolved oxygen.

In addition for measurements below 1 mg/L a zero calibration should be performed.

Probe Polarization: The dissolved oxygen probe are continuously polarized when they are connected to the meter. But when first connected to the meter, or has been unplugged from the meter for more then an hour it needs to be given 30 to 60 minutes to polarize the probe. Interrupted connections of less then an hour will take 5 – 25 minutes for the readings to stabilize.

Zeroing the Probe

A DO probe can generate a 0.02 to 0.05 mg/L positive error in an oxygen free (anoxic) solution. If this error is unacceptable then zero the probe with the following procedure when:

• Using a new sensing-membrane

• Using fresh internal filling solution

• Measuring dissolved oxygen levels below 1 mg/L or 10% saturation

Page 67: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

VIII-9Orion Star Series MeterThermo Electron Corporation

Dissolved Oxygen

Calibrating with Water Saturated Air:Calibration of the Oxygen Probe

It is recommended to calibrate the oxygen probe before performing measurements. The Meter has four types of calibration. The meter may be calibrated via Air Calibration (using water-saturated air as the calibration standard), Water Calibration (using air-saturated water as the calibration standard), or Winkler Calibration (using the Winkler titration as the basis for the calibration). The meter may also be calibrated by performing a Probe Zero calibration. The Probe Zero calibration is used in rare applications and should be performed if the probe is suspected of having residual current. This “zeroing out” of the residual current is particularly helpful when samples are at a low level. The mode that is used in calibration (% saturation vs. concentration) is governed by the measurement mode prior to starting the calibration.

Note: The probe and calibration standard (water-saturated air or air-saturated water) should be allowed to reach equilibrium before the system is calibrated. If a polarographic probe is used, it is also important to allow sufficient time (at least 30 to 50 minutes) for the system to polarize. ▲

Air Calibration (AutoCal)(using water-saturated air as the calibration standard)

The default calibration mode is Air calibration. If the calibration selection in SETUP has not been changed, then pressing the CALIBRATE key will automatically perform an Air calibration. Due to the inherent differences between water-saturated air and air-saturated water, upon stability, the air standard is set to 101.7 % saturation or the corresponding concentration value.

The highest possible accuracy is reached when calibration temperature is the same as the measuring temperature.

1. Prepare the calibration sleeve* by removing the reservoir cap from the bottom of the sleeve. Remove the sponge from the cap and saturate it with water. Squeeze all excess water out of the sponge so that it is merely damp. Replace the sponge into the cap and replace the cap onto the calibration sleeve.

Page 68: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVIII-10 Orion Star Series Meter

Dissolved Oxygen

2. Place the dissolved oxygen probe into the calibration sleeve. Ensure that the probe membrane does not come into contact with the sponge in the calibration sleeve. Also ensure that there are no drops of water on the membrane.

3. Make sure Setup has AIR calibration set.

4. To initiate the Air calibration, press the CALIBRATE key. The “cal” indicator will light up and the current slope (SLP) will be displayed briefly with the calibration type.

5. While the electrode signal is stabilizing, the meter will flash a dissolved oxygen reading (based upon the current calibration data) in the main field. Once the input has stabilized, 101.7 % saturation will be displayed for three seconds. The meter will return to measure mode.

* A standard 300 mL BOD bottle may also be use for AIR calibration. Fill the bottom of the bottle with 50 mL of distilled water. Be sure to use the BOD funnel accessory (Orion Cat. Nos. 080044 or 080160, depending on the probe in use). Place the probe into the funnel so that it is just above the stirring paddle. Ensure that there are no drops of water on the surface of the membrane. Ensure that the water level is 1/2 inch below the tip of the probe. Allow at least 30 minutes for the bottle to equilibrate.

Water Calibration (using air-saturated water as the calibration standard)

The Water calibration function must be enabled in SETUP before the calibration is started. When stability is reached, the water standard is set to 100 % or the corresponding concentration value.

The highest possible accuracy is reached when calibration temperature is the same as the measuring temperature.

Page 69: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

VIII-11Orion Star Series MeterThermo Electron Corporation

Dissolved Oxygen

1. Prepare the calibration standard by bubbling air through a beaker of stirred deionized water for at least one hour. Ensure that the stirring is vigorous enough to provide adequate mixing yet not forceful enough to create a vortex in the beaker. Make sure that the water level is high enough to fully submerge the probe thermistors, which are located inside the metal band above the membrane cap. Ensure that no air bubbles accumulate on the surface of the membrane.

Note: Some magnetic stirrers generate enough heat to change solution temperature. To avoid this, place a piece of cardboard, foam rubber or other insulating material between the stir plate and sample beaker. ▲

2. Place the dissolved oxygen probe into the aerated water so that the sensors are immersed. Ensure that there are no air bubbles accumulating on the membrane.

3. To initiate the Water calibration, press the CALIBRATE key. The “cal” indicator will light up and the current slope (SLP) will be displayed briefly with the calibration type.

4. While the electrode signal is stabilizing, the meter will flash a dissolved oxygen reading (based upon the current calibration data) in the main field. Once the input has stabilized, 100.0 % saturation will be displayed for three seconds.

5. The slope value (SLP) will then be displayed. The meter will return to measure mode.

Calibration to the Winkler Titration Method(using the Winkler titration as the basis for the calibration)

Many regulatory agencies require the determination of dissolved oxygen in water to be referenced to the Winkler Titration Method. The Meter allows the user to scroll the dissolved oxygen measurement to correlate with the value obtained from the Winkler titrated sample. The man calibration function must be enabled in SETUP before the calibration is started.

Page 70: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVIII-12 Orion Star Series Meter

Dissolved Oxygen

1. Place the dissolved oxygen probe into the sample. Make sure that the water level is high enough to fully submerge the probe thermistors, which are located inside the metal band above the membrane cap. Ensure that the stirring is vigorous enough to provide adequate mixing yet not forceful enough to create a vortex in the beaker. Ensure that there are no air bubbles accumulating on the membrane.

Note: Some magnetic stirrers generate enough heat to change solution temperature. To avoid this, place a piece of cardboard, foam rubber or other insulating material between the stir plate and sample beaker. ▲

2. To initiate the Winkler titration calibration, press the CALIBRATE key. The “cal” indicator will light up and the current slope (SLP) will be displayed briefly with the calibration type.

3. While the electrode signal is stabilizing, the meter will flash a dissolved oxygen reading (based upon the current calibration data) in the main field. Once the input has stabilized, the concentration value, which corresponds to saturation at the current temperature, pressure and salinity, will be displayed.

4. The concentration value should then be adjusted (using the UP/DOWN arrow & DECIMAL/DIGIT keys) to the value that was obtained via the Winkler titration. The user must press the MEASURE key to accept the concentration value.

5. The slope value (SLP) will then be displayed for three seconds. The meter will then return to measure mode.

Probe Zero Calibration If the expected sample range is at the lower end of the dissolved oxygen scale, you may want to perform a zeroing of your probe. If the probe exhibits any residual current, this calibration will compensate for it. A Probe Zero calibration should always follow as the second point determination of a two-point calibration. To remove the effects of a Probe Zero calibration, perform an Air, Water or Winkler calibration as a one-point calibration. To perform a Probe Zero calibration, follow the steps outlined below.

Page 71: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

VIII-13Orion Star Series MeterThermo Electron Corporation

Dissolved Oxygen

1. Perform either an Air, Water or Winkler calibration before continuing with the Probe Zero calibration. See the appropriate calibration procedure above.

2. Enable the Set0 calibration function in SETUP before the calibration is started.

3. Prepare an oxygen scavenging solution, such as sodium sulfite. Add 15 grams of sodium sulfite to 250 mL of deionized water. Stir until most of the salt has dissolved. Transfer the solution into a BOD bottle. Place the overflow funnel into the bottle so that it is snugly sealed.

4. Place the probe into the bottle ensuring that the thermistor is fully submerged in the solution. Place the bottle on a magnetic stirrer & stir gently.

5. Allow the solution to eliminate all oxygen from the membrane module. This should take at least 5 minutes.

6. To initiate the Probe Zero calibration, press the CALIBRATE key. The “cal” indicator will light up and the current slope (SLP) will be displayed briefly with the calibration type.

7. While the electrode signal is stabilizing, the meter will flash a dissolved oxygen reading (based upon the current calibration data) in the main field. Once the input has stabilized, 0.0 will be displayed.

8. The user must press the MEASURE key to zero the probe.

9. The slope value (SLP) will then be displayed for three seconds. The meter will then return to measure mode.

10. After performing the probe zeroing, thoroughly rinse the probe with deionized water.

Note: Prior to using the BOD funnel and bottle again, be sure to remove all traces of the oxygen scavenging solution for accurate measurements. ▲

Page 72: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationVIII-14 Orion Star Series Meter

Page 73: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter IX Key Feature OverviewKey Feature OverviewAudible Signals

The meter will beep whenever a key is pressed providing immediate verification that the user’s input was received.

Automatic Shut-off

When Automatic shutoff is enabled (this is the instrument default) the instrument will shutdown after 20 minutes without a keypress. This maximizes battery life of portable meters or benchtop meters being run on battery power. Benchtop meters using line power are switched into standby mode which reduces power consumption.(verify occurs as stated on BT).

Display Backlight

A quick press of the power key will toggle the backlight on & off. On the portable meters (and the benchtop meters when on battery power) the backlight will automatically turn off after 2 minutes to conserve battery power.

When line power is attached to the benchtop meter the the backlight will remain on until turned off.

AUTO-READ™

If you are in AUTO-READ mode then press the MEASURE key to start the measurement. Once the ready is stable the meter will automatically Log & Print the results and freeze the display. If the stirrer is enabled then the stirrer will turn on when the MEASURE key is pressed and turn off once the reading has stabilized.

• With the AutoStir BOD Probe pressing the button on the probe will start the AUTO-READ measurement.

IX-1Orion Star Series MeterThermo Electron Corporation

Page 74: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-2 Orion Star Series Meter

Key Feature Overview

Stirrer Control

The Benchtop meters have stirrer control for the Thermo Orion 096019 Stirrer and the AutoStir BOD probe. The 096019 is an overhead stirrer that fits into Thermo Orion’s Electrode stand and is much more convient to use then magnetic stirrers. There are no stirrer bars to handle and the stirrer blade can be rinsed with a rinse bottle with the electrode.

• The stirrer button will toggle the stirrer on & off.

• If AUTO-READ and the stirrer are enabled then the stirrer will turn on when the MEASURE key is pressed and turn off once the reading has stabilized.

• With the AutoStir BOD Probe pressing the MEASURE key will start the AUTO-READ™ measurement and stirring.

Methods

If methods have been turned on, then this screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the MEASURE button to implement the selected method and return to measurement mode. Press the SETUP or RIGHT arrow key to edit the selected method.

• Up to 10 methods can be created, edited, & selected. Initially all 10 methods are set to the instrument defaults.

Passwords

If a password has been setup, then this screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

Page 75: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-3Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

Calibration Alarm Setup MenuThe following table describes the DO Setup Options:

Calibration Alarm Setup

Description Display Range of Values Default Value

Method Specific

pH Calibration AlarmSetting

dUEPH0

0000 - 9999Time in Hours; 0000 = CalAlarm Off

0000 Yes

ORP Calibration AlarmSetting

dUEOrP0

0000 - 9999Time in Hours; 0000 = CalAlarm Off

0000 Yes

ISE Calibration AlarmSetting

dUEISE0

0000 - 9999Time in Hours; 0000 = CalAlarm Off

0000 Yes

Conductivity CalibrationAlarm Setting

dUECOnd0

0000 - 9999Time in Hours; 0000 = CalAlarm Off

0000 Yes

Dissolved Oxygen Calibration Alarm Setting

dUECOnd0

0000 - 9999Time in Hours; 0000 = CalAlarm Off

0000 Yes

pH Calibration Alarm Setting: This selects the time in hours after a calibration before the meter will remind you that the pH needs to be recalibrated. It will provide this warning by lighting the Caution icon and flashing the Probe icon. In addition, the units icon for pH will flash about once every 5 seconds. A setting of 0000, turns off the calibration alarm.

ORP Calibration Alarm Setting: This selects the time in hours after a calibration before the meter will remind you that the ORP needs to be recalibrated. It will provide this warning by lighting the Caution icon and flashing the Probe icon. In addition, the units icon for Rel mV will flash about once every 5 seconds. A setting of 0000, turns off the calibration alarm.

Page 76: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-4 Orion Star Series Meter

Key Feature Overview

ISE Calibration Alarm Setting: This selects the time in hours after a calibration before the meter will remind you that the ISE needs to be recalibrated. It will provide this warning by lighting the Caution icon and flashing the Probe icon. In addition, the units icon for ISE will flash about once every 5 seconds. A setting of 0000, turns off the calibration alarm.

Cond Calibration Alarm Setting: This selects the time in hours after a calibration before the meter will remind you that the Conductivity needs to be recalibrated. It will provide this warning by lighting the Caution icon and flashing the Probe icon. In addition, the µS/cm, mS/cm, mg/L, ppt, or MW-cm icon will flash about once every 5 seconds. A setting of 0000, turns off the calibration alarm.

DO Calibration Alarm Setting: This selects the time in hours after a calibration before the meter will remind you that the DO needs to be recalibrated. It will provide this warning by lighting the Caution icon and flashing the Probe icon. In addition, the %Sat or mg/L icon will flash about once every 5 seconds. A setting of 0000, turns off the calibration alarm.

How to Change the Cal Alarm Setup OptionsTo access the Cal Alarm Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

• Press the UP & DOWN arrow to scroll through the Setup Menu’s until dUE is displayed on the Top line.

• Press the RIGHT ARROW key to select the Cal Alrm setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

Page 77: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-5Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

Press the UP & DOWN arrow to select the Cal Alarm setup Options as listed below:

PH = pH Calibration Alarm OrP = ORP “Rel mV” Calibration Alarm ISE = ISE Calibration Alarm COnd = Conductivity Calibration Alarm DO = DO Calibration Alarm

• Press the RIGHT ARROW key to select the option editing line (Bottom Line). Press the MEASURE key to return to measurement mode.

• pH Calibration Alarm Setting Menu Screen.

Scroll in the number of hours between calibrations. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

0000 to 9999, where 0000 turns the calibration alarm off.

• Press the RIGHT ARROW key to accept the selection and select the Cal Alarm Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• ORP Calibration Alarm Setting Menu Screen.

Scroll in the number of hours between calibrations. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

0000 to 9999, where 0000 turns the calibration alarm off.

• Press the RIGHT ARROW key to accept the selection and select the Cal Alarm Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 78: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-6 Orion Star Series Meter

Key Feature Overview

• ISE Calibration Alarm Setting Menu Screen.

Scroll in the number of hours between calibrations. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

0000 to 9999, where 0000 turns the calibration alarm off.

• Press the RIGHT ARROW key to accept the selection and select the Cal Alarm Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Conductivity Calibration Alarm Setting Menu Screen.

Scroll in the number of hours between calibrations. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

0000 to 9999, where 0000 turns the calibration alarm off.

• Press the RIGHT ARROW key to accept the selection and select the Cal Alarm Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Dissolved Oxygen Calibration Alarm Setting Menu Screen.

Scroll in the number of hours between calibrations. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

0000 to 9999, where 0000 turns the calibration alarm off.

• Press the RIGHT ARROW key to accept the selection and select the Cal Alarm Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 79: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-7Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

Datalog Setup MenuThe following table describes the Measurement Setup Option:

Data Logging Setup

Description Display Range of Values Default Value

Method Specific

Data Log Roll Over orDelete on Download OptionSelection

LOgdELyES

nO, yESSelects whether to DeleteLog after downloading to aComputer or printer

yES Yes

Datalog Delete Option: This allows you to select whether the Datalog will automatically delete “clear” the log after it is downloaded to a printer or computer or if instead it will roll over and start overwriting the oldest data. If “YES” is selected it will stop datalogging and report a datalog full error when all 200 datalog points are filled. To continue datalogging you must download the data to a printer or computer. If “No” is selected when all 200 datalog points are filled it will start overwriting the datalog starting at point 1.

How to Change the Datalog Setup OptionTo access the Datalog Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

Page 80: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-8 Orion Star Series Meter

Key Feature Overview

• Press the UP & DOWN arrow to scroll through the Setup Menu’s until LOg is displayed on the Top line.

• Press the RIGHT ARROW key to select the dEL setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

Datalog Setup Option Menu Screen.• Press the RIGHT ARROW key to select the option editing line (Bottom Line).

Press the MEASURE key to return to measurement mode.

• Datalog Auto-Delete Selection Menu Screen.

Press the UP & DOWN arrow to select the Measurement Type options as listed below:

yES = Automatic log delete after download nO = Roll over log when full

• Press the RIGHT ARROW key to accept the selection and select the Log Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 81: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-9Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

General Instrument Setup MenuThe following table describes the Measurement Setup Options:

General Instrument Setup

Description Display Range of Values Default Value

Method Specific

Manual TemperatureCompensation Setting

gEndEgC25.0

-5.0 - 105 25.0 Yes

Stirrer Speed Setting gEnStIr4

OFF, 1, 2, 3, 4, 5, 6, 7 4 Yes

Auto-Shut Off Selection dUEISE0

gEnAUtOOn

off, on Yes

• Manual Temperature Setting: The is the temperature that will be used for temperature compensation when no temperature sensors are attached to the instrument.

• Stirrer Speed Setting: This screen is only present on the Benchtop meters. It is used to set the stirrer speed. The speeds are 1 through 7 with speed 1 being the slowest qand speed 7 being the fastest. We generally recommend a speed of 3 or 4 for most applications. When a stirrer speed is selected:

• In continuous or timed reading mode the stirrer is toggled on and off with the stirrer button.

• In AUTO-READ mode pressing the MEASURE key will start the stirrer and the stirrer automatically turn off once the measurement is stable.

• When the AUTO-STIR BOD probe and AUTO-READ mode is selected, pressing the button on the AUTO-STIR BOD probe will act just like pressing the MEASURE key. It will turn on the stirrer, then when the reading stabilizes it will automatically turn off the stirrer, freeze the display and print & log the results.

Page 82: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-10 Orion Star Series Meter

Key Feature Overview

• Auto-Shut Off Selection: When Auto-Shut off is “On” the meter will automatically shut-off after 20 mintues without a key press. This is to save battery power. The meter will restart with any key press from auto-shut off. If Auto-shut off is turn “OFF” the instrument will not go into Auto-Shut off low power mode.

How to Change the General Instrument Setup OptionsTo access the General Instrument Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

• Press the UP & DOWN arrow to scroll through the Setup Menu’s until gEn is displayed on the Top line.

• Press the RIGHT ARROW key to select the rEAd setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

General Instrument Setup Option Menu Screen.Press the UP & DOWN arrow to select the rEAd setup Options as listed below:

dEgC = Manual Temperture Setting StIr = Enable & Set Stirrer Speed AUtO = Auto-Shut Off Enable/Disable

Page 83: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-11Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

• Press the RIGHT ARROW key to select the option editing line (Bottom Line). Press the MEASURE key to return to measurement mode.

• Manual Temperature Setting Menu Screen.

Enter the manual temperature setting by Pressing the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

• Press the RIGHT ARROW key to accept the selection and select the General Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Stirrer Speed Selection Setting Menu Screen.

Press the UP & DOWN arrow to select the Stirrer options as listed below:

OFF = Disables the stirrer 1 = Slowest Speed 2 = 2nd Speed 3 = 3rd Speed 4 = 4th Speed 5 = 4th Speed 6 = 6th Speed 7 = Fastest Speed

• Press the RIGHT ARROW key to accept the selection and select the General Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Auto-Shut Off Selection Menu Screen.

Press the UP & DOWN arrow to select the Stirrer options as listed below:

OFF = Disables Auto-Shut Off On = Enable Auto-Shut Off

• Press the RIGHT ARROW key to accept the selection and select the General Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 84: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-12 Orion Star Series Meter

Key Feature Overview

GLP Instrument Setup MenuThe following table describes the Measurement Setup Options:

GLP Instrument Setup

Description Display Range of Values Default Value

Method Specific

Enable or Disable Methods “Setting”

gIPSEtOn

OFF, OnWhen on will allow storingAnd using 10 stored Methods

OFF No

Enable & Set InstrumentPassword

gIPPASS0000

0000 - 9999Sets the Setup/MethodPassword; 0000 disablespasswords

0000 No

Methods Setting: This allows selecting between a single setup or a methods based setup. When this is turned “On” the user will be able to enter, edit, & select from 10 stored methods. Methods are actual always in use. This sets whether different methods can be selected, & created or edits or just the current method. The default current method is method 1.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

Password Setup: This allows the user to Enable and setup a password to prevent unauthorized individuals from changing the Setup options or editing/creating methods. Setting the Password to 0000, disables passwords.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

Page 85: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-13Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

How to Change the GLP Instrument Setup OptionsTo access the GLP Instrument Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

• Press the UP & DOWN arrow to scroll through the Setup Menu’s until gEn is displayed on the Top line.

• Press the RIGHT ARROW key to select the rEAd setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

Page 86: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-14 Orion Star Series Meter

Key Feature Overview

GLP Instrument Setup Option Menu Screen.Press the UP & DOWN arrow to select the GLP setup Options as listed below:

SET = Method “Set” Enable/Disable PASS = Enable & Set Password

• Press the RIGHT ARROW key to select the option editing line (Bottom Line). Press the MEASURE key to return to measurement mode.

• Methods On/Off Selection Menu Screen.

Press the UP & DOWN arrow to select the Stirrer options as listed below:

OFF = Disables Methods On = Enables Methods

• Press the RIGHT ARROW key to accept the selection and select the GLP Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Password Setting Menu Screen.

Set the Password by Pressing the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

0000 to 9999, where 0000 is password disabled

• Press the RIGHT ARROW key to accept the selection and select the GLP Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 87: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-15Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

Time & Date Setup MenuThe following table describes the Time & Date Setup Options:

Time & Date Setup

Description Display Range of Values Default Value

Method Specific

Time: Hour Setting dAtEHOUrHH12

00 - 23Sets the hour of the currenttime in 24Hr format

12 Yes

Time: Minute Setting dAtEtImEmm12

00 - 59Sets the minutes for the current time

00 Yes

Date Format dAtEtyPEdAy

mdy, dmyMonth, Day, Year or Day,Month, Year Selection

mdy Yes

Date: Year Setting dAtEyEAr2004

00 - 99Sets the year for thecurrent date from 2000 to2099

04 Yes

Date: Month Setting dAtEdAtEmm01

01 - 12Month Setting 01 “Jan”Thru. 12 “Dec”

01 Yes

Date: Day of the Month Setting

dAtEdAydd01

01 - 31Day of the Month

01 Yes

Time: Hour Setting: This is to set the Hour of the current time in 24 hour “Military Time” format (for example, 4 pm equals 16).

Timed: Minute Setting: This is to set the minutes of the hour of the current time (for example, 4:15 pm would be 15).

Date Format: This sets the printer and computer output format. Either mdy (Month, Day, Year) or dmy (Day, Month, Year).

Page 88: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-16 Orion Star Series Meter

Key Feature Overview

Date: Year Setting: Sets the year of the current date from year 2000 to year 2099.

Date: Month Setting: Sets the Month of the current Year from 01 “Jan” to 12 “Dec”.

Date: Day of the Month: Sets the day of the current month from 01 to 31.

How to Change the Time & Date Setup OptionsTo access the Measurement Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

• Press the UP & DOWN arrow to scroll through the Setup Menu’s until rEAd is displayed on the Top line.

• Press the RIGHT ARROW key to select the dAtE setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

Measurement Setup Option Menu Screen.Press the UP & DOWN arrow to select the dAtE setup Options as listed below:

HOUr = Current Time Hour Setting tIE = Current Time Minute Setting tYPE = Date output format mdy or dmy YEAr = Year of current date dAtE = Month of current date dAY = Day of current month

Page 89: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-17Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

• Press the RIGHT ARROW key to select the option editing line (Bottom Line). Press the MEASURE key to return to measurement mode.

• Hour Setting Menu Screen.

Enter the Hour of the current time by Pressing the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

00 to 23

• Press the RIGHT ARROW key to accept the selection and select the Time & Date Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Minute Setting Menu Screen.

Enter the Minutes of the current time by Pressing the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

00 to 59

• Press the RIGHT ARROW key to accept the selection and select the Time & Date Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Date Format Selection Menu Screen.

Press the UP & DOWN arrow to select the Date Format used for printer and computer output as listed below:

mdy = Month, Day, Year dmy = Day, Month, Year

• Press the RIGHT ARROW key to accept the selection and select the Time & Date Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 90: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-18 Orion Star Series Meter

Key Feature Overview

• Year Setting Menu Screen.

Enter the year of the current date by Pressing the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

2000 to 2099

• Press the RIGHT ARROW key to accept the selection and select the Time & Date Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Month Setting Menu Screen.

Enter the Month of the current date by Pressing the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

00 to 12

• Press the RIGHT ARROW key to accept the selection and select the Time & Date Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Day Setting Menu Screen.

Enter the Day of the current month by Pressing the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled.

01 to 31

• Press the RIGHT ARROW key to accept the selection and select the Time & Date Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 91: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-19Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

RS-232 Instrument Setup MenuThe following table describes the Measurement Setup Options:

RS232 Instrument Setup

Description Display Range of Values Default Value

Method Specific

RS-232 Baud Rate Selection R232bAUd1200

1200,2400,4800,9600Baud Rate

9600 No

Printout Format R232OUtFPrnt

Print,COmPPrinter or Computer“Comma Delimited”

Print No

• RS-232 Baud Rate Selection: This allows the user to select between 4 baud rates for printer and computer output.

• Printout Format Selection: This allows the user to select between an output which is formatted for a printer and a comma delimited output for direct interfacing to LIM’s systems, Excel, or data analysis software.

How to Change the General Instrument Setup OptionsTo access the RS-232 Instrument Setup Menu:

• From the Measurement Screen press the SETUP key.

a. If methods have been turned on then the Method selection screen will appear when setup is pressed. Press the UP & DOWN arrows to select the Method to be used or edited. Press the SETUP or RIGHT arrow key to edit the selected method.

b. If a password has been setup, then the password entry screen will appear and a password will need to be entered to enter setup and edit the method. Press the UP & DOWN arrow keys to change the value of the flashing digit. Press the DECIMAL/DIGIT key to change the digit to be scrolled. Press the RIGHT Arrow key to move to enter setup. Press the MEASURE key to exit to measurement mode.

Page 92: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-20 Orion Star Series Meter

Key Feature Overview

• Press the UP & DOWN arrow to scroll through the Setup Menu’s until R232 is displayed on the Top line.

• Press the RIGHT ARROW key to select the R232 setup option line (Middle Line). Press the MEASURE key to return to measurement mode.

RS-232 Instrument Setup Option Menu ScreenPress the UP & DOWN arrow to select the R232 setup Options as listed below:

bAUd = Baud Rate “1200, 2400, 4800, 9600” OUtF = Printer or Computer output format

• Press the RIGHT ARROW key to select the option editing line (Bottom Line). Press the MEASURE key to return to measurement mode.

• Baud Rate Setting Menu Screen.

Enter serial ports baud rate by Pressing the UP & DOWN arrow keys to change the selection.

• Press the RIGHT ARROW key to accept the selection and select the RS-232 Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

• Output Format Selection Menu Screen.

Press the UP & DOWN arrow to select the Output options as listed below:

Prnt = Printer format COmP = Computer Format “Comma Delimited Output”

• Press the RIGHT ARROW key to accept the selection and select the RS-232 Setup Menu line (Top Line). Press the MEASURE key to return to measurement mode.

Page 93: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-21Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

Example of a multi-parameter printoutA multi-parameter printout is the same as the single parameter printouts, with the exception of a few minor changes.

1. the header and footer is only printed once.

2. Each section is separated by a blank line. The print out is determined by what is on the display.

3. If a particular line is turned off, then that section will not be printed out.

4. If all lines are turned off, nothing other than the header will get printed.

PH 10.45 pH MV -183.3 mV Temperature Temperature = 25.0 C Blank line Conductivity Conductivity = 1413 uS/cm Conductance Conductance = 30.2 mS Temperature Temperature = 25.0 C Temp. Coefficient Temp. Coefficient = 10.0 %/C (or OFF, or NLF) Temp. Reference Temp. Reference = 25.0 C Cell Constant Cell Constant = 1.00 /cm Blank line Concentration Concentration = 100.0 mg/L % Saturation Saturation = 60.0 %sat Current Current = 100.0 nA Temperature Temperature = 25.0 C Barometric Pressure Barometric Pressure = 760 mmHg Salinity Correction Salinity Correction = 20 Calibration number Calibration # 3 Log number Log # 2 Sample ID Sample # ____________ Operator Operator ____________

Page 94: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-22 Orion Star Series Meter

Key Feature Overview

Log View:1. In measurement mode press the Log View Button.

2. The top line will say “Log”

3. The bottom line will allow you to use the scroll key (up & down arrows) to select between vIEw, Send, & CALO

4. vIEw will display the log point on the Temperature line, the time of that log point on the top line, the date of the log point on the middle line, and the year of that log point on the bottom line.\

a. Use the Up & Down arrow key to scroll to different time and dates or hit the Log View Button to view the log point.

b. In viewing the log point use the Up & Down scroll & view key to different log points or press the Log View key to return to the Time & Date selection screen.

c. Press the MEAS key to escape to measurement mode.

5. Send will output all measurement log points to a printer or computer.

6. CALO will output all calibration logs to a computer or printer.

Page 95: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

IX-23Orion Star Series MeterThermo Electron Corporation

Key Feature Overview

Error Codes & Debugging

If an error code is displayed on the screen, the user must press the MEAS key to clear.

Error/Event codes will be in the format “E###”. Some of these are errors, some are warnings, and some are purely informational. The first digits identifies the type of error or event as identified in the table below:

System Events/Errors

Description Error Code Description Error Code

System Event or Error E0## Calibration Event or Error E7##

pH Event or Error E1## Setup Event or Error E8##

ISE Event or Error E2## SW Event or Error E9##

ORP Event or Error E3## HW Event or Error EA##

Conductivity Event or Error E4## Memory Event or Error EB##

DO Event or Error E5## ISE Event or Error EC##

Measurement Event or Error E6##

System Events/Errors

Error Code

Description Press MEAS to Clear

Cause Trouble Shooting

E001 No Model Number Yes Memory Corrupted Call Customer Service

E002 No HW Calibration Yes Memory Corrupted Call Customer Service

E003 Bad HW Calibration Yes Memory Corrupted Call Customer Service

E004 RTC Failure Yes HW Transeint or Failure

Reset Meter, if error reoccurs contact Customer Service

E005 ADC Failure Yes HW Transeint or Failure

Reset Meter, if error reoccurs contact Customer Service

Page 96: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationIX-24 Orion Star Series Meter

Key Feature Overview

E033 Keypad Failure Yes Hardware problem or user missed hitting one of the keys during the test

Re-run the self test to insure you have a hardware problem, then contact Customer Service

E035 Password Wrong Yes User entered incorrect password

Re-enter the passwordIf you have forgotten your pass-word contact Customer Service

E036 User Entered Wrong Value

Yes User entered value is out of range

Re-enter value. Check specifica-tions for range of correct values

pH Events/Errors

Error Code

Description Press MEAS to Clear

Cause Trouble Shooting

E105 Auto-Buffer Failure Yes Auto-buffer not recognized, will only happen in remote control

Page 97: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter X Star Series Specifications3-Star Meter Specifications (Benchtop and Portable)

3-Star pH

pH Range -2.000 to 19.999

Resolution 0.1/0.01/0.001

Relative accuracy +/-0.002

ORP mV/RmV/Eh Range +/-1999.9

Resolution +/-0.1

Relative accuracy +/-0.2 mV or 0.5%, whichever is greater

3-Star ISE

ISE Concentration Range 0.0 to 19,900.0

Units M, mg/L, %, ppb, none

Resolution 1, 2, or 3 sig. Digits

Relative Accuracy +/-0.2 mV or 0.05%, whichever is greater

mV/RmV +/-1999.9

3-Star DO

DO Concentration 0.000 to 90.00 mg/L

Relative Accuracy 0.2 mg/L

% Saturation 0.0 to 600

Salinity factor 0 to 45 ppt

Barometric pressure 450-850 mm Hg

X-1Orion Star Series MeterThermo Electron Corporation

Page 98: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationX-2 Orion Star Series Meter

Star Series Specifications

3-Star Conductivity

Conductivity Range 0-3000 mS/cm cell constant dependent

Resistivity 0.0001 to 100 Megohms

Salinity 0.01 to 80.0 ppt NaCl

TDS 0.0 to 19.999.0 mg/L

Cell Constant 0.001 to 199.9

Reference temperature 15 °C, 20 °C or 25 °C (default), autoranging

4-Star Meter Specifications (Benchtop and Portable)

4-Star pH/ISE

pH Range -2.000 to 19.999

Resolution 0.1/0.01/0.001

Relative accuracy +/-0.002

ISE Concentration Range 0.0 to 19,900.0

Units M, mg/L, %, ppb, none

Resolution 1, 2, or 3 sig. Digits

Relative Accuracy +/-0.2 mV or 0.05%, whichever is greater

mV/RmV +/-1999.9

ORP mV/RmV/Eh Range +/-1999.9

Resolution +/-0.1

Relative accuracy +/-0.2 mV or 0.5%, whichever is greater

Page 99: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

X-3Orion Star Series MeterThermo Electron Corporation

Star Series Specifications

4-Star pH/DO

pH Range -2.000 to 19.999

Resolution 0.1/0.01/0.001

Relative accuracy +/-0.002

DO Concentration 0.000 to 90.00 mg/L

Relative Accuracy 0.2 mg/L

% Saturation 0.0 to 600

Salinity factor 0 to 45 ppt

Barometric pressure 450-850 mm Hg

ORP mV/RmV/Eh Range +/-1999.9

Resolution +/-0.1

Relative accuracy +/-0.2 mV or 0.5%, whichever is greater

4-Star pH/Conductivity

pH Range -2.000 to 19.999

Resolution 0.1/0.01/0.001

Relative accuracy +/-0.002

Conductivity Range 0-3000 mS/cm cell constant dependent

Resistivity 0.0001 to 100 Megohms

Salinity 0.01 to 80.0 ppt NaCl

TDS 0.0 to 19.999.0 mg/L

Cell Constant 0.001 to 199.9

Reference temperature 15 °C, 20 °C or 25 °C (default), autoranging

ORP mV/RmV/Eh Range +/-1999.9

Resolution +/-0.1

Relative accuracy +/-0.2 mV or 0.5%, whichever is greater

Page 100: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationX-4 Orion Star Series Meter

Star Series Specifications

5-Star Meter Specifications (Benchtop and Portable)

5 Star pH/ISE/DO/Cond

pH Range -2.000 to 19.999

Resolution 0.1/0.01/0.001

Relative accuracy +/-0.002

ISE Concentration Range 0.0 to 19,900.0

Units M, mg/L, %, ppb, none

Resolution 1, 2, or 3 sig. Digits

Relative Accuracy +/-0.2 mV or 0.05% whichever is greater

mV/RmV +/-1999.9

DO Concentration 0.000 to 90.00 mg/L

Relative Accuracy 0.2 mg/L

% Saturation 0.0 to 600

Salinity factor 0 to 45 ppt

Barometric pressure 450-850 mm Hg

Conductivity Range 0-3000 mS/cm cell constant dependent

Resistivity 0.0001 to 100 Megohms

Salinity 0.01 to 80.0 ppt NaCl

TDS 0.0 to 19.999.0 mg/L

Cell Constant 0.001 to 199.9

Reference temperature 15 °C, 20 °C or 25 °C (default), autoranging

ORP mV/RmV/Eh Range +/-1999.9

Resolution +/-0.1

Relative accuracy +/-0.2 mV or 0.5%, whichever is greater

Page 101: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

X-5Orion Star Series MeterThermo Electron Corporation

Star Series Specifications

Description SpecificationsRegulatory and Safety CE,CSA, TÜV, UL, FCC Class A limits

Environmental Operating Conditions Ambient Operating Temperature 5 to 45 °C Relative Humidity 5 to 85% Non-Condensing Storage Temperature -20 to 60 °C Storage Humidity 5% to 85%, Non-Condensing

Case Material ABS

Shock and Vibration Vibration, Shipping/Handling Per ISTA #1A Shock, Drop Test in Packaging Per ISTA #1A

Enclosure IP67 (Portable Meter) IP54 (Benchtop Meter)

Page 102: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationX-6 Orion Star Series Meter

Page 103: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter XI Terms and Conditions1. General Seller hereby offers for sale to the buyer (“Buyer”) the products herein (the “Products”) on the express condition that Buyer agrees to accept and be bound by the terms and conditions set forth herein. Any provisions contained in any document issued by Buyer are expressly rejected and if the terms and conditions in this Agreement differ from the terms of Buyer’s offer, this document shall be construed as a counter offer and shall not be effective as an acceptance of Buyer’s document. Buyer’s receipt of Products or Seller’s commencement of the services provided hereunder will constitute Buyer’s acceptance of this Agreement. This is the complete and exclusive statement of the contract between Seller and Buyer with respect to Buyer’s purchase of the Products. No waiver, consent, modification, amendment or change of the terms contained herein shall be binding unless in writing and signed by Seller and Buyer. Seller’s failure to object to terms contained in any subsequent communication from Buyer will not be a waiver or modification of the terms set forth herein. All orders are subject to acceptance in writing by an authorized representative of Seller.

2. WarrantyThermo Electron warranty for Orion products covers failures due to manufacturer’s workmanship or material defects from the date of purchase by the user. User should return the warranty card to Thermo Electron and retain proof of purchase. Warranty is void if product has been abused, misused, or repairs attempted by unauthorized persons.

Warranties herein are for product sold/installed by Thermo Electron or its authorized dealers.

Any product sold by a U.S. or Canadian distributor must be returned to Thermo Electron for any warranty work. Please contact our Technical Service department for further information. A Return Authorization Number must be obtained from Thermo Electron Technical Service before returning any product for in-warranty repair or replacement.

XI-1Orion Star Series MeterThermo Electron Corporation

Page 104: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationXI-2 Orion Star Series Meter

Terms and Conditions

In the event of failure within the warranty period, Thermo Electron will at Thermo Electron’s option, repair or replace product not conforming to this warranty. There may be additional charges, including freight, for warranty service performed in some countries. For service, call Thermo Electron (or its authorized dealer outside the United States and Canada). Thermo Electron reserves the right to ask for proof of purchase, such as the original invoice or packing slip.

Field Service is available on Orion BOD AutoEZ™, Orion EZ Flash® GC Accessory and Orion TEA Analyzer®. Contact our Field Service department for details on quotations, service, and other field service-related activities.

The following products are warranted to be free from defects in material and workmanship in the period listed below from the date of purchase from the user or from the date of shipment from Thermo Electron, whichever is earlier, provided use is in accordance with the operating limitations and maintenance procedures in the instruction manual and when not having been subjected to accident, alteration, misuse, abuse or breakage of electrodes:

Thirty-six months from date of purchase by the user (or forty-two months from date of shipment from Thermo Electron)

All Orion Star Series pH, ISE, DO and Conductivity Meters, Orion Waterproof Meters (Orion 630, 635, 830A, 835A, 260A, 261S, 265A, 266S, 130A, 131S, 135A, 136S), Conductivity Meters (Orion 105Aplus™, 115Aplus™, 125Aplus™, 145Aplus™, 150Aplus™ and 162A), PerpHect® pH/ISE Meters (Orion 310, 320, 330, 350, 370) pH/ISE Meters (Orion 210Aplus™, 230Aplus™, 250Aplus™, 290Aplus™, 410Aplus™, 420Aplus™, 520Aplus™, 525Aplus™, 710Aplus™, 720Aplus™ and 920Aplus™), pHuture MMS™ Meters (Orion 535A and 555A), pH/Conductivity Meter (Orion 550A), Dissolved Oxygen Meters (Orion 805Aplus™, 810Aplus™, 850Aplus™ and 862A).

Twenty-four months from date of purchase by the user (or thirty-six months from date of shipment from Thermo Electron)

Orion Ross Ultra™ Electrodes, AQUAfast® IV Colorimeters, AQUAfast® IV Turbidimeter, 925 Flash Titrator™, Series 100 DuraProbe™ Conductivity Cells and Series 800 Dissolved Oxygen Probes.

Page 105: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

XI-3Orion Star Series MeterThermo Electron Corporation

Terms and Conditions

Twelve months from date of purchase by the user (or eighteen months from date of shipment from Thermo Electron)

Orion Laboratory pH Meters, (Orion 301, 611 and 940), SensorLink®, pHuture™ pH Meters (Orion 610 and 620), Smart Chek™ meters, Sage® Pumps, Cahn® Balances, 930 Ionalyzer®, 950 ROSS™ FAST QC™ Titrator, 960 Titrator PLUS®, Karl Fischer Titrators, Autosamplers, pHuture® Conversion Box, Wine Master®, 607 Switchbox, rf link™, AQUAfast® II Colorimeters, Vacuum Degasser and Flowmeter.

Orion EZ Flash® GC Accessory, TEA Analyzer® Orion 610 and 510 excluding consumable items carry twelve months warranty from date of installation or sixteen months from shipment.

Orion Ion Selective Electrodes, ionplus® Electrodes, ROSS™ Electrodes, Sure-Flow® Electrodes, PerpHecT® Electrodes, AquaPro Professional Electrodes, No Cal™ pH electrodes, Standard Line pH Electrodes, Tris pH Electrodes, KNIpHE® electrode, ORP Triode™ (Orion 9180BN), pHuture™ pH Probes (Orion 616500) and pHuture MMS™ Quatrode™ and Triode™ (Orion 616600 and 617900), Orion 97-08 DO Probe, Series 100 Conventional Conductivity Cells, temperature probes and compensators (except those models noted).

Orion 93 and 97 ionplus Series sensing modules are warranted to give six months of operation if placed in service before the date indicated on the package, except 93-07 and 97-07 Nitrate modules are warranted to give ninety days of operation if placed in service before the date indicated on the package.

Six months from date of purchase by the user (or twelve months from date of shipment from Thermo Electron)

Orion Flash Titration™ Probe (Orion 092518), pHuture™ Electrode (Orion 615700), pHuture MMS™ Pentrode™ (Orion 617500), Quatrode™ (Orion 617800) and Triode™ (Orion 615800), Low Maintenance Triode™ (Orion 9107BN), ORP Low Maintenance Triode™ (Orion 9179BN), and PerpHecT® Low Maintenance Triode™ (Orion 9207BN), Waterproof Triode™ (Orion 9107WP, 9107WL, 9109WL and 9109WP), QuiKcheK® Meters and Micro Electrodes.

Page 106: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationXI-4 Orion Star Series Meter

Terms and Conditions

Three months from date of purchase by the user (or six months from date of shipment from Thermo Electron)

Economy Line Electrodes, Orion 91-05, 91-06, 91-15, 91-16, 91-25, 91-26, 91-35, 91-36, 92-06. Warranty also includes failure for any reason (excluding breakage), except abuse, provided the electrode is not used in solutions containing silver, sulfide, perchlorate, or hydrofluoric acid; or in solutions more than one (1) Molar in strong acid or base at temperatures above 50 °C.

“Out-of-Box” Warranty - Should any of the following products fail to work when first used, contact Thermo Electron immediately for replacement.

Orion Solutions, Standards, Reagents, Cables, Line adapters, Printers, Software, Cases, Stands, Probe Membranes, AQUAfast® Test Strips, EZ Flash® columns and general accessories.

For products in the catalog not listed in this warranty statement, please visit our website at: www.thermo.com

THE WARRANTIES DESCRIBED ABOVE ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER STATUTORY, EXPRESS OR IMPLIED INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE AND ALL WARRANTIES ARISING FROM THE COURSE OF DEALING OR USAGE OF TRADE. THE BUYER’S SOLE AND EXCLUSIVE REMEDY IS FOR REPAIR OR REPLACEMENT OF THE NON-CONFORMING PRODUCT OR PART THEREOF, OR REFUND OF THE PURCHASE PRICE, BUT IN NO EVENT SHALL THERMO ELECTRON (ITS CONTRACTORS AND SUPPLIERS OF ANY TIER) BE LIABLE TO THE BUYER OR ANY PERSON FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES WHETHER THE CLAIMS ARE BASED IN CONTRACT, IN TORT (INCLUDING NEGLIGENCE), OR OTHERWISE WITH RESPECT TO OR ARISING OUT OF THE PRODUCT FURNISHED HEREUNDER. REPRESENTATION AND WARRANTIES MADE BY ANY PERSON, INCLUDING ITS AUTHORIZED DEALERS, REPRESENTATIVES AND EMPLOYEES OF THERMO ELECTRON WHICH ALTER OR ARE IN ADDITION TO THE TERMS OF THIS WARRANTY SHALL NOT BE BINDING UPON THERMO ELECTRON UNLESS IN WRITING AND SIGNED BY ONE OF ITS OFFICERS.

Page 107: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

XI-5Orion Star Series MeterThermo Electron Corporation

Terms and Conditions

3. Limitation of LiabilityNOTWITHSTANDING ANYTHING TO THE CONTRARY CONTAINED HEREIN, THE LIABILITY OF SELLER UNDER THESE TERMS AND CONDITIONS (WHETHER BY REASON OF BREACH OF CONTRACT, TORT, INDEMNIFICATION, OR OTHERWISE, BUT EXCLUDING LIABILITY OF SELLER FOR BREACH OF WARRANTY (THE SOLE REMEDY FOR WHICH SHALL BE AS PROVIDED UNDER SECTION 2 ABOVE)) SHALL NOT EXCEED AN AMOUNT EQUAL TO THE LESSER OF (A) THE TOTAL PURCHASE PRICE THERETOFORE PAID BY BUYER TO SELLER WITH RESPECT TO THE PRODUCT(s) GIVING RISE TO SUCH LIABILITY OR (B) ONE MILLION DOLLARS ($1,000,000). NOTWITHSTANDING ANYTHING TO THE CONTRARY CONTAINED HEREIN, IN NO EVENT SHALL SELLER BE LIABLE FOR ANY INDIRECT, SPECIAL, CONSEQUENTIAL OR INCIDENTAL DAMAGES (INCLUDING WITHOUT LIMITATION DAMAGES FOR LOSS OF USE OF FACILITIES OR EQUIPMENT, LOSS OF REVENUE, LOSS OF DATA, LOSS OF PROFITS OR LOSS OF GOODWILL), REGARDLESS OF WHETHER SELLER (a) HAS BEEN INFORMED OF THE POSSIBILITY OF SUCH DAMAGES OR (b) IS NEGLIGENT.

4. Miscellaneous(a) The rights and obligations of the parties hereunder shall be governed by and construed in accordance with the laws of the Commonwealth of Massachusetts, without reference to its choice of law provisions. Each party hereby irrevocably consents to the exclusive jurisdiction of the state and federal courts located in Suffolk County, Massachusetts, in any action arising out of or relating to this Agreement and waives any other venue to which it may be entitled by domicile or otherwise. (b) In the event of any legal proceeding between the Seller and Buyer relating to this Agreement, neither party may claim the right to a trial by jury, and both parties waive any right they may have under applicable law or otherwise to a right to a trial by jury. Any action arising under this Agreement must be brought within one (1) year from the date that the cause of action arose.

Page 108: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationXI-6 Orion Star Series Meter

Page 109: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Chapter XII Declaration of Conformity

Manufacturer: Thermo Electron

Address: 166 Cummings Center Beverly, Ma 01915 USA

The above named Manufacturer hereby declares that the product(s) described below conforms to the Standards and Directives listed below:

5-Star pH/ISE/Cond/DO benchtop 5-Star pH/Cond/DO portable 4-Star pH/ Cond benchtop 4-Star pH/ Cond portable 4-Star pH/ DO benchtop 4-Star pH/ DO portable 4-Star pH/ ISE benchtop 4-Star pH/ ISE portable 3-Star Cond benchtop 3-Star Cond portable 3-Star DO benchtop 3-Star DO portable 3-Star pH Benchtop 3-Star pH portable

Directive / Standard:89/336/EEC Electromagnetic Compatibility (EMC) Directive

EN 61326:1997 + A1:1998 + A2:2001, Electrical equipment for measurement, control and laboratory use: EMC-requirements for measurement device per Annex D.

XII-1Orion Star Series MeterThermo Electron Corporation

Page 110: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron CorporationXII-2 Orion Star Series Meter

Declaration of Conformity

73/23/EEC Low voltage directive

EN 61010-1:2001 Safety requirements for electrical equipment for measurement, control and laboratory use - general requirements

These products have been manufactured in compliance with the provisions of the relevant Thermo Electron manufacturing and test documents and processes. Further, these documents and processes are recognized as complying with ISO9001: 2000 by QMI, listed as File #001911.

Place and date of issues:

Beverly, MA, February 22, 2005

______________________________

Robert Manning

Manager of QA & Mfg Engineering

Page 111: Orion Star Series Meter - Cole-Parmer...ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results

Thermo Electron Corporation

Environmental InstrumentsWater Analysis Instruments

166 Cummings CenterBeverly, MA 01915 USATel: 978-232-6000Toll Free: 800-225-1480Dom. Fax: 978-232-6015Int’l. Fax: 978-232-6031

www.thermo.com/water

248590-001 RevA