optimizing gas chromatography - gulf coast …...optimizing gas chromatography brian rohrback,...

19
Infometrix Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018

Upload: others

Post on 30-Aug-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Optimizing Gas Chromatography

Brian Rohrback, Infometrix, Inc.

Gulf Coast Conference October 2018

Page 2: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

The Overarching Infometrix Goal

• The process world has a disconnect as we reduce staffing and add more analyzers to the monitoring and control mix.

• More data does not get us into a better position to achieve our quality objectives; we must extract the information content.

• As much as possible, we need to make the interpretation more automated, objective and do so in the highest quality manner.

• We also need to cut the lifecycle costs, which means lower costs of installation, maintenance and training of personnel.

3

Page 3: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Delivering information

Just having the measurements does not translate into control

• Remember, there are not enough skilled technicians to handle even

the current workload.

• Chemometrics aids the processing problem with 2 technologies:

– Alignment enables us to sell instruments that have vastly-lower

calibration requirements.

– Interpretation algorithms automate the generation and the

qualification of the information derived from the raw data.

And if we can make all of our instruments

look as much alike as possible.

Interchangeability

Common interpretive base

4

Page 4: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Chromatograms – RT drift

5Crude Standard: 3 years of repeat injections

Page 5: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Aligned profiles

6

Well aligned 2014

10.000027 11.666751 13.333475 15.0002 16.666925

Wave Number

100

150

200

Re

sp

on

se

Page 6: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

20 40 60

Time (seconds)

Raw data3 instruments

Automated alignment across

instruments

20 40 60

Time (seconds)

alignedRaw data3 instruments

7

Page 7: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Taking this a step beyond

• Think about how alignment relates to simulated

distillation…

– We run an n-paraffin standard to correlate

temperature to retention time.

– We use this new axis to map the cumulative percent

of total area as we progress along this set of

temperatures.

8

Page 8: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Location 1 – aligned to the Location 1 standard

Page 9: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Location 2 – aligned to the Location 2 standard

Page 10: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Location 2 – aligned to the Location 1 standard

Page 11: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Impact of alignment

0.0

1.0

2.0

3.0

4.0

5.0

IBP 5% 10% 20% 30% 40% 50% 60% 70% 80% 90% 95% FBP

Unaligned

AlignedDif

fere

nce

bet

wee

n s

ample

and R

T st

andar

d (

°F)

12ASTM 7798 ILS

Page 12: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

GC that is (almost) calibration-free

We can correct retention times to match an application-

specific relevant sample

You can use this to make all instruments performing a

similar task to look identical (Plug and Play)

This raises the possibility of having a universal calibration

At the least, the frequency with which we really need to

run calibration standards is significantly lower that what

is currently being done

AND, we get the ability to apply continuous (and reliable)

data interpretation

13

Page 13: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

SimDist D7798: lab versus on-line

Laboratory

Run a Blank Run the RT

Standard

Run the

RGO

Run Samples

Pass PassPass

4. Run Samples

3. Run the Reference

to check results with a known

1. Run a Blank

to ensure a clean system

2. Run the n-Paraffin Standard

relates retention time to temperature

14

Page 14: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

SimDist D7798: lab versus on-line

On-Line Chromatographic files from the laboratory

Run Samples

4. Run Samples

Run a Blank

Pass

1. Run a Blank

to ensure a clean system

Use the Lab’s

RT Standard

2. Use the same n-Paraffin Standard

from the run in the laboratory

Use the RGO as

Alignment

Target

3. Use the RGO or Process Sample

as an alignment target

15

Page 15: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

SimDist test samples

16

Page 16: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ixReorienting our thinking about

chromatography

Kowalski, 1980

17

Page 17: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Comparison of PCA scores

18

85% of all of the variation in the raw

data is due to the misaligned peaks.

Correcting for this shows us that there are three different production regimes in these data.

Before alignment After alignment

Page 18: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Identify cause for outlier groupings

19

3600 process chromatograms

Bad

Good

Page 19: Optimizing Gas Chromatography - Gulf Coast …...Optimizing Gas Chromatography Brian Rohrback, Infometrix, Inc. Gulf Coast Conference October 2018 o rix The Overarching Infometrix

Info

metr

ix

Acknowledgements

• Robert Lorenz and Brian Cudby, Chevron

• Rajko Puzic, ExxonMobil

• Joe Perron, Falcon

• ASTM 7798 ILS Laboratories

20