OPERATOR, ENVIRONMENT,
PRODUCT PROTECTION
www.lamsys.com
The system is patented and available only with LAMSYSTEMS cabinets
– Physical isolation (containment and controlled removal from the work zone) of pathogenic biological agents (PBA) and microorganisms to prevent airborne infection of the staff and contamination of the air in the work room and laboratory environment.
– Minimization of risk of contamination and cross-contamination of the product.
– Possibility to work with small amount of toxic chemicals and radionuclides as well as to remove work agent odors in case of mandatory connection to an active exhaust system using an exhaust hood supplied upon customer's additional request.
– Equipment of individual work places in medical, pharmaceutical and other institutions working with pathogenic biological agents and microorganisms.
2
Operating mode monitoring system with audible and visible alarm indicates any disturbances of the airflow balance in the work chamber.
A proximity electromagnetic key reliably protects the control system from unauthorized access.
Front sash and UV unit position is controlled with optical sensors.
A removable ergonomic armrest prevents accidental blocking of the front air grille and ensures comfortable support of operator's arms.
Simple filter replacement.
A tabletop is divided in sections allowing their simple disinfection (within the work zone) and autoclaving.
PULLOUT UV UNIT:
– does not disturb the laminar flow– does not require any additional space for storage– reliable and simple in operation– well-adapted for disinfection- completely closes the front opening when equipment is idle– controlled with optical position sensor– passed TÜV NORD certification
900 mm 1200 mm
BMB-II-«Laminar-S»
Conforms to EN 12469:2000-09
The cabinet is featured with an audible and visible alarm system REFLEX LS activated in case the front air grille is blocked to indicate the reduction of the level of safety provided by the cabinet.
NE
OT
ER
IC –
th
e ne
wes
t, a
dvan
ced,
mod
ern
Article.................................................................................................................................................... .....................
Maximum dimensions of the cabinet with stand* (WхDхH), mm....................................................... ................
Dimensions of the work chamber (WхDхH), mm................................................................................... ....................
Rated input power** (without built-in electric sockets), W................................................................................... ...................................
Maximum allowed load on the built-in electric sockets, W................................................................................... ................................... 3Downflow, m /h............................................................................................................................................... ..............................
3Inflow, m /h..................................................................................................................................................... .............................
Illumination of the work zone, lx........................................................................................................................... ....................................
Maximum net weight of the cabinet with stand, kg............................................................................................. .....................................
Maximum noise level at 1 m distance from the cabinet, dB(A)........................................................................... ...................................
21
50
1490
1373
660
1000/1200 770
3
AIRFLOW LAYOUT
room airclean (filtered) aircontaminated air
MAIN CHARACTERISTICS
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
– for particles of 0.5µm and more.................................................................................................................................................................ISO 5
– for particles of 5.0µm and more..............................................................................................................................ISO M (20; ≥5µm); LSAPC
Cabinet class according to ЕN 12469:2000, NSF/ANSI 49 ................................................................................................................................
as per NSF/ANSI 49..........................................................................................................................................................
Class of HEPA filters as per EN 1822-1..........................................................................................................................................................
Average velocity of the inflow though the work opening, m/s................................................................................................................
Average downflow velocity in the work chamber, m/s............................................................................................................................
Air recirculation, %..........................................................................................................................................................................................
:2015
Airflow recirculation
I
А2
Н14
0,47±0,03
0,35+0,01
70
I
MAIN PARAMETERS AND DIMENSIONS
1Е-B.001-12.0
1200х770х2150*
1105х610х750
110**
1000
795-817
333-378
1000
230
55***
1Е-B.001-09.0
1000х770х2150*
905х610х750
100**
1000
656-674
273-309
1000
195.5
59***
BSC Class II Type A2
BMB-II-«Laminar-S» NEOTERIC
* Dimensions do not account for outstanding supports(The supports stand out of the cabinet's dimensions by 10 mm on each side).
** Power when HEPA filters are new (not clogged).*** Level of noise measured as per DIN EN ISO 11201 in free sound field
over the sound-reflecting surface (noise level in real operating environment depends on the dimensions of the operating site as well as on the total background noise and may vary by 3-4 dB(A)).
DIMENSIONAL DRAWING
www.lamsys.com
BMB-II-”Laminar-S”
Code: 222.120 222.150 222.180
– controlled removal from the work zone) of pathogenic biological agents (PBA) and microorganisms to prevent airborne infection of the staff and contamination of the air in the work room and laboratory environment.
– Minimization of risk of contamination and cross-contamination of the product.
– Possibility to work with small amount of toxic chemicals and radionuclides as well as to remove work agent odors in case of mandatory connection to an active exhaust system using an exhaust hood supplied upon customer's additional request.
– Equipment of individual work places in medical, pharmaceutical and other institutions working with pathogenic biological agents and microorganisms.
Physical isolation (containment and
SIMPLE ADJUSTMENT OF AIRFLOW VELOCITYThe system individually controls the inflow and downflow velocities as well as automatically maintains airflow balance. Since there is no need to adjust the airflow balance manually, the time of maintenance at qualification, filter replacement and periodic checks is significantly reduced.
HIGHLY PRECISE MAINTENANCE OF PRESET VELOCITY of the airflows at any level of filter clogging and in case of environment change (humidity, temperature, pressure).
A TOUCHSCREEN CONTROL PANEL conveniently visualizes operating modes, simplifies the control process and provides extensive service and maintenance information.
SIMPLE AND SAFE FILTER REPLACEMENT PROCEDURE minimizes the risk of filter damage in the process of their replacement thanks to configuraiton of the cabinet and new spring fixation system.
4
LOW NOISE LEVEL
47 dBA in operating modeat testing laboratory
900 mm 1200 mm 1500 mm 1800 mm
BMB-II-«Laminar-S»
SAV
VY
– s
ophi
stic
ated
, pr
ofic
ient
, w
ise
Article................................................................................................ ........ ....... ........
Maximum dimensions of the cabinet with stand* (WхDхH), mm... .... ..... ....
Dimensions of the work chamber (WхDхH), mm................................ ........ ....... .......
Rated input power** (without built-in electric sockets), W................................ ....................... ...................... ......................
Maximum allowed load on the built-in electric sockets, W................................ ....................... ...................... .......................3Downflow, m /h........................................................................................... ................. ............... ..............
3Inflow, m /h................................................................................................. ................. ................. .................
Illumination of the work zone, lx....................................................................... ...................... ....................... ......................
Maximum net weight of the cabinet with stand, kg.......................................... ........................ ........................ ........................
Maximum noise level at 1 m distance from the cabinet, dB(A)........................ ...................... ........................ ......................
5
20
95
1425
670
1373
7701000/1200/1500/1800
BSC Class II Type A2
BMB-II-«Laminar-S» SAVVY
MAIN CHARACTERISTICS
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
– for particles of 0.5µm and more.................................................................................................................................................................ISO 5
– for particles of 5.0µm and more..............................................................................................................................ISO M (20; ≥5µm); LSAPC
Cabinet class according to ЕN 12469:2000, NSF/ANSI 49 ................................................................................................................................
as per NSF/ANSI 49..........................................................................................................................................................
Class of HEPA filters as per EN 1822-1..........................................................................................................................................................
Average velocity of the inflow though the work opening, m/s................................................................................................................
Average downflow velocity in the work chamber, m/s............................................................................................................................
Air recirculation, %..........................................................................................................................................................................................
:2015
Airflow recirculation
I
А2
Н14
0,47±0,03
0,35+0,01
70
I
AIRFLOW LAYOUT
room airclean (filtered) aircontaminated air
MAIN PARAMETERS AND DIMENSIONS
* Dimensions do not account for outstanding supports(The supports stand out of the cabinet's dimensions by 10 mm on each side).
** Power when HEPA filters are new (not clogged).*** Level of noise measured as per DIN EN ISO 11201 in free sound field
over the sound-reflecting surface (noise level in real operating environment depends on the dimensions of the operating site as well as on the total background noise and may vary by 3-4 dB(A)).
DIMENSIONAL DRAWING
1Е-B.002-15.0
1500х770х2095*
1405х610х700
142**
1000
1008-1036
426-484
2000
260
53***
1Е-B.002-18.0
1800х770х2095*
1705х610х700
2000
295
55***
300**
1000
1210-1245
510-580
1Е-B.002-12.0
1200х770х2095*
1105х610х700
110**
1000
2000
230
47***
795-817
333-378
1Е-B.002-09.0
1000х770х2095*
905х610х700
110**
1000
656-674
273-309
2000
195.5
47***
www.lamsys.com
HAS ALL THE ADVANTAGES OF SAVVY MODEL (see page 4)
DISTINCTION FROM SAVVY MODEL:POWER-DRIVEN FRONT SASH
Five preset positions of the front sash:1. OPERATING POSITION in Main Operation mode.2. UP POSITION for work chamber loading/unloading.3. STOP BEFORE CLOSING for safety (at ~40 mm height from the work surface to prevent any trauma of operator's hands).4. CLOSED POSITION in Standby mode or for workchamber UV irradiation.5. DOWN POSITION for disinfection of the upper part of the front sash.
For safety reasons, the moving front sash automatically stops ahead of obstacle (in case of contact with an object or a hand of the operator).
900 mm 1200 mm 1500 mm 1800 mm
6
BMB-II-«Laminar-S»
– controlled removal from the work zone) of pathogenic biological agents (PBA) and microorganisms to prevent airborne infection of the staff and contamination of the air in the work room and laboratory environment.
– Minimization of risk of contamination and cross-contamination of the product.
– Possibility to work with small amount of toxic chemicals and radionuclides as well as to remove work agent odors in case of mandatory connection to an active exhaust system using an exhaust hood supplied upon customer's additional request.
– Equipment of individual work places in medical, pharmaceutical and other institutions working with pathogenic biological agents and microorganisms.
Physical isolation (containment and
LOW NOISE LEVEL
47 dBA in operating modeat testing laboratory
7
~40 мм
Dimensions of the cabinet assembled with stand** (WxDxH), mm....... ... ... ....
Dimensions of the cabinet with stand and completely
opened front sash /WхDхH/, mm............................................................. ..... ..... .....
Dimensions of the work chamber (WxDxH), mm......................................... ........ ....... ........
Rated input power without built-in electric sockets, W, not more than......................... ......................... .......................... .........................
Maximum allowed load on the built-in electric sockets, W......................................... ........................ ......................... .........................3Downflow, m /h..................................................................................................... .................. ............... ...............
3Inflow, m /h........................................................................................................... .................. ................... ...................
Illumination of the work zone, lx, not less than ......................................................... ........................ ....................... ........................
Weight of the cabinet assembled with the stand (net), kg, not more than................... .......................... ......................... .........................
Maximum noise level at 1 m distance from the cabinet, dBA ................................... ........................ ......................... .......................
Article.................................................................................................. ........ ...... ........ 1Е-B.008-15.0*
1500х810х2095**
1500х810х2180
1405х610х700
2000
-
-
570
1000
1008-1036
451-512
1Е-B.008-18.0
1800х810х2095**
1800х810х2180
1705х610х700
2000
320
55***
610
1000
1210-1245
548-623
* Articles 1Е-B.008-09.0 and 1Е-B.008-15.0 are in the process of development. Some of the parameters may change.**
*** Level of noise in free sound field over the sound-reflecting surface (noise level in real operating environment
depends on the dimensions of the operating site as well as on the total background noise and may vary by 3-4 dB(A)).
Dimensions do not account for outstanding supports(The supports stand out of the cabinet's dimensions by 10 mm on each side).
1Е-B.008-12.0
1200х810х2095**
1200х810х2180
1105х610х700
2000
255
47***
570
1000
795-817
354-402
1Е-B.008-09.0*
1000х810х2095**
1000х810х2180
905х610х700
2000
215
47***
570
1000
656-674
289-328
20
95
1425
670
1200/1800810
BSC Class II Type A2
BMB-II-«Laminar-S» SAVVY SL
MAIN CHARACTERISTICS
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
– for particles of 0.5µm and more.................................................................................................................................................................ISO 5
– for particles of 5.0µm and more..............................................................................................................................ISO M (20; ≥5µm); LSAPC
Cabinet class according to ЕN 12469:2000, NSF/ANSI 49 ................................................................................................................................
as per NSF/ANSI 49..........................................................................................................................................................
Class of HEPA filters as per EN 1822-1..........................................................................................................................................................
Average velocity of the inflow though the work opening, m/s................................................................................................................
Average downflow velocity in the work chamber, m/s............................................................................................................................
Air recirculation, %..........................................................................................................................................................................................
:2015
Airflow recirculation
I
А2
Н14
0,47±0,03
0,35+0,01
70
I
AIRFLOW LAYOUT
room airclean (filtered) aircontaminated air
MAIN PARAMETERS AND DIMENSIONS
DIMENSIONAL DRAWING
www.lamsys.com
8
Equipped with a mode monitoring system, component (sash, UV unit and work opening screen) position sensors, control panel with LCD screen that visualizes operating information. In case of safety disturbances, an audible and visual alarm is generated.
A laminar air downflow in the work zone prevents any cross-contamination of the materials.
A preliminary filter (to retain animal hair) is installed under the tabletop.
A pullout UV unit is part of the standard configuration.Please refer to page 2.
The cabinet is designed for simultaneous work OF TWO OPERATORS OPPOSITE ONE ANOTHER
VIS-A-VIS
BMB-II-«Laminar-S»
– controlled removal from the work zone) of pathogenic biological agents (PBA) and microorganisms to prevent airborne infection of the staff and contamination of the air in the work room and laboratory environment.
– Minimization of risk of contamination and cross-contamination of the product.
– Possibility to work with small amount of toxic chemicals and radionuclides as well as to remove work agent odors in case of mandatory connection to an active exhaust system using an exhaust hood supplied upon customer's additional request.
– Equipment of individual work places in medical, pharmaceutical and other institutions working with pathogenic biological agents and microorganisms.
Physical isolation (containment and
1200 mm
Neoteric
9
* The power consumption with newly installed uncontaminated HEPA filters.
Article................................................................................................................................................................................................
Maximum dimensions of the cabinet with stand* (WхDхH), mm...................................................................................................
Dimensions of the work chamber (WхDхH), mm..............................................................................................................................
Rated input power (without built-in electric sockets), W....................................................................................................................................
Maximum allowed load on the built-in electric sockets, W...............................................................................................................................3Downflow, m /h...........................................................................................................................................................................................
3Inflow with two front windows opened, m /h...............................................................................................................................................
Illumination of the work zone, lx.......................................................................................................................................................................
Maximum net weight of the cabinet with stand, kg............................................................................................................................................
Maximum noise level at 1 m distance from the cabinet, dB(A).........................................................................................................................
* Dimensions do not account for outstanding supports(The supports stand out of the cabinet's dimensions by 10 mm on each side).
** Level of noise measured as per DIN EN ISO 11201 in free sound field.over the sound-reflecting surface (noise level in real operating environment depends on the dimensions
of the operating site as well as on the total background noise and may vary by 3-4 dB(A)).
1Е-B.004-12.0
1470х730х2050*
1110х700х670
620
1000
690-710
665-755
1500
330
60**
730 1470
600
1450
2050
BSC Class II Type A2
AIRFLOW LAYOUT
room airclean (filtered) aircontaminated air
DIMENSIONAL DRAWING
MAIN CHARACTERISTICS
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
– for particles of 0.5µm and more.................................................................................................................................................................ISO 5
– for particles of 5.0µm and more..............................................................................................................................ISO M (20; ≥5µm); LSAPC
Cabinet class according to ЕN 12469:2000, NSF/ANSI 49 ................................................................................................................................
as per NSF/ANSI 49..........................................................................................................................................................
Class of HEPA filters as per EN 1822-1..........................................................................................................................................................
Average velocity of the inflow though the work opening, m/s................................................................................................................
Average downflow velocity in the work chamber, m/s............................................................................................................................
Air recirculation, %..........................................................................................................................................................................................
:2015
Airflow recirculation
I
А2
Н14
0,47±0,03
0,35+0,01
50
I
MAIN PARAMETERS AND DIMENSIONS
BMB-II-«Laminar-S» NEOTERIC VIS-A-VIS
www.lamsys.com
Not for sale in AT CH DE DK NL
10
– Protection of the personnel from the impact of cytostatic and cytotoxic agents that are manipulated in the work chamber.
– Aseptic production of antineoplastic agents.
– Physical isolation (containment and controlled removal from the work zone) of pathogenic biological agents (PBA) and microorganisms to prevent airborne infection of the staff and contamination of the air in the work room and laboratory environment.
– Minimization of risk of contamination and cross-contamination of the product.
– In case of work with cytostatics, psychotropic substances, methyl methacrylates, phenols and formaldehydes, organic solvents, aniline colorants etc. as well as to remove work agent odors, the cabinet shall be connected to an active exhaust system using an exhaust hood that is part of the standard configuration of the cabinet.
– Equipment of individual work places in medical, pharmaceutical and other institutions working with pathogenic biological agents, microorganisms and toxic powder substances.
WORK CHAMBER:– tabletop made of stainless steel, power-driven sliding front sash made of laminated safety glass, side windows made of tempered glass;– tray under the tabletop is made of stainless steel, has easy-to-clean configuration and prevents any leakages inside the cabinet or on the preliminary HEPA filters;– work chamber lighting is LED.
SUPPLY AND EXHAUST AIR FILTRATION SYSTEM:– one supply HEPA filter H14 located above the work chamber;– one exhaust HEPA filter H14 located in the upper ventilation chamber;– two preliminary panel HEPA filters H14 located under the tabletop. THE FILTERS ARE REPLACED within the work chamber WHEN THA FANS ARE ON and do not require any disassembly of the cabinet, thus, minimizing the risk of contamination of the personnel or the environment.
CONTROL SYSTEM COMPONENTS:– automatic airflow maintenance independent of the level of the HEPA filter clogging;– automatic switching-off of the UV light in case of opening the front sash or UV unit screen.
1200 mm
The model is in the process of development. Some of the parameters may change
BMB-II-«Laminar-S»
TOXIC CYTOSTATIC
11
Article................................................................................................................................................................................................
Maximum dimensions of the cabinet with stand* (WхDхH), mm...................................................................................................
Dimensions of the work chamber (WхDхH), mm..............................................................................................................................
Rated input power** (without built-in electric sockets), W...............................................................................................................................
Maximum allowed load on the built-in electric sockets, W...............................................................................................................................3Downflow, m /h...........................................................................................................................................................................................
3Inflow, m /h.................................................................................................................................................................................................
Illumination of the work zone, lx......................................................................................................................................................................
Maximum net weight of the cabinet with stand, kg............................................................................................................................................
Maximum noise level at 1 m distance from the cabinet, dB(A)........................................................................................................................
1Е-B.012-12.0
1200х800х2355*
1105х610х700
350**
1000
795-817
354-402
2000
270
58***
* Dimensions do not account for outstanding supports(The supports stand out of the cabinet's dimensions by 10 mm on each side).
** Power when HEPA filters are new (not clogged).*** Level of noise measured as per DIN EN ISO 11201 in free sound field
over the sound-reflecting surface (noise level in real operating environment depends on the dimensions of the operating site as well as on the total background noise and may vary by 3-4 dB(A)).
1200
800
537
310
1200
253
2050 6
10
2180
2355
BSC Class II Type A2
BMB-II-«Laminar-S» SAVVY SL CYTOS
MAIN CHARACTERISTICS
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1:2015
– for particles of 0.5µm and more.................................................................................................................................................................ISO 5
– for particles of 5.0µm and more..............................................................................................................................ISO M (20; ≥5µm); LSAPC
Cabinet class according to ЕN 12469:2000, NSF/ANSI 49 ................................................................................................................................
Airflow as per NSF/ANSI 49..........................................................................................................................................................
Class of HEPA filters as per EN 1822-1..........................................................................................................................................................
Average velocity of the inflow though the work opening, m/s................................................................................................................
Average downflow velocity in the work chamber, m/s............................................................................................................................
Air recirculation, %..........................................................................................................................................................................................
recirculation
I
А2
Н14
0,47±0,03
0,35+0,01
70
I
MAIN PARAMETERS AND DIMENSIONS
AIRFLOW LAYOUT DIMENSIONAL DRAWING
room airclean (filtered) aircontaminated air
www.lamsys.com
12
– Physical isolation (containment and controlled removal from the work zone) of pathogenic biological agents (PBA) and microorganisms to prevent airborne infection of the staff and contamination of the air in the work room and laboratory environment.
– Minimization of risk of contamination and cross-contamination of the product.
– Equipment of individual work places in medical, pharmaceutical and other institutions working with pathogenic biological agents and microorganisms.
THE CABINET DOES NOT PROTECT FROM TOXIC CHEMICALS AND RADIONUCLIDES!
In case of work with small amount of toxic chemicals and radionuclides as well as to remove work agent odors, the cabinet shall be connected to an
3active exhaust system of 1135-1205 m /h capacity.
Thereat. the room of installation shall be equipped with a supply ventilation 3system of at least 1000m /h capacity.
In all other cases, it is up to the organization operating the cabinet to decide whether it is necessary to connect the cabinet to the exhaust system based on the risk assessment and analysis.
No recirculation in the cabinet
BMB-II-«Laminar-S»
1200 mm 1500 mm 1800 mm
13
Article........................................................................................................................... ......... ..........
Maximum dimensions of the cabinet with exhaust hood* (WхDхH), mm................ ....... .......
Dimensions of the work chamber (WхDхH), mm......................................................... ......... .........
Rated input power (without built-in electric sockets), W...................................................... ........................... ..........................
Maximum allowed load on the built-in electric sockets, W.......................................................... ........................ ........................3Downflow, m /h...................................................................................................................... ................ .................
3Inflow, m /h......................................................................................................................... ................ ...............
Illumination of the work zone, lx.................................................................................................. ......................... ........................
Maximum net weight of the cabinet with exhaust hood, kg.......................................................... ........................... ..........................
Maximum noise level at 1 m distance from the cabinet, dB(A).................................................. .......................... .........................
1E-B.003-12.0
1200х810х2325*
1105х610х705
720/174**
1000
795-820
1135-1205
1000
265
59***
1E-B.003-15.0
1500х810х2220*
1405х610х705
730
1000
1008-1036
1434-1520
2000
330
59***
1E-B.003-18.0
1800х810х2200*
1705х610х705
960
2000
1070-1100
1590-1690
2000
370
59***
* Dimensions do not account for outstanding supports(The supports stand out of the cabinet's dimensions by 10 mm on each side).
** Power when HEPA filters are new (not clogged).*** Level of noise measured as per DIN EN ISO 11201 in free sound field
over the sound-reflecting surface (noise level in real operating environment depends on the dimensions of the operating site as well as on the total background noise and may vary by 3-4 dB(A)).
2013
2325/2
220
312
1241
772
1200/1500/1800
BSC Class II Type В2
BMB-II-«Laminar-S» NEOTERIC В2
MAIN CHARACTERISTICS
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
– for particles of 0.5µm and more.................................................................................................................................................................ISO 5
– for particles of 5.0µm and more..............................................................................................................................ISO M (20; ≥5µm); LSAPC
Cabinet class according to ЕN 12469:2000, NSF/ANSI 49 ................................................................................................................................
as per NSF/ANSI 49..........................................................................................................................................................
Class of HEPA filters as per EN 1822-1..........................................................................................................................................................
Average velocity of the inflow though the work opening, m/s................................................................................................................
Average downflow velocity in the work chamber, m/s............................................................................................................................
Air recirculation, %.........................................................................................................................................................................
:2015
Airflow recirculation
I
В2
Н14
0,47±0,03
0,35+0,01
I
no recirculation
MAIN PARAMETERS AND DIMENSIONS
AIRFLOW LAYOUT DIMENSIONAL DRAWING
room airclean (filtered) aircontaminated air
www.lamsys.com
14
– Protection of an operator and the environment from aerosols generated during the work with pathogenic biological agents (PBA) and microorganisms including the highest (especially hazardous) risk groups.
– Protection of the work agents in the work chamber from external contamination.
– Possibility to work with small amount of toxic chemicals and radionuclides as well as to remove work agent odors in case of mandatory connection to an active exhaust system.
– Equipment of individual work places in medical, pharmaceutical and other institutions working with I-II-III-IV risk group pathogens and parasitic disease causative microorganisms.
The model of 1800 mm width has 4 glove ports allowing simultaneous work of two operators
THE CABINET IS CONTROLLED via a touchscreen control panel.
FRONT SASH is featured with oval-shaped glove ports; the material of the window is laminated safety glass resistant to UV and disinfectants.
Each glove port is equipped with a CHAMBER GLOVE highly resistant to UV, ozone, natural aging, acids and alkalies.
In the bottom part of the work chamber, there is a 10-liter LIQUID-COLLECTING TRAY (without drain).
LED LIGHTING of the work chamber.
REMOVABLE SET OF SOCKETS.
UV LIGHT in the work chamber.
LEAK-TIGHT PASS BOX.
BMB-II-«Laminar-S»
1200 мм 1800 мм
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
– for particles of 0.5µm and more.................................................................................................................................................................ISO 5
– for particles of 5.0µm and more..............................................................................................................................ISO M (20; ≥5µm); LSAPC
Cabinet class according to ЕN 12469:2000, NSF/ANSI 49 ...............................................................................................................................
Class of the installed НЕРА filters according to EN 1822-1............................................................................................................................
Prefilter class according to EN779:2002...........................................................................................................................................................
Minimal negative pressure in the working chamber, Pa, not less than ...........................................................................................................
Permanently provided working negative pressure in the working chamber, Pa, not less than........................................................................
Air inflow filtering...................................................................................................................................................................
Air outflow filtering...............................................................................................................................................................
Air recirculation, %........................................................................................................................................................................
:2015
15
342.120
1200 1800
Article....................................................................................................................................................... ...............
Maximum dimensions of the cabinet with pass box* (WхDхH), mm.................................................... ............
Dimensions of the work chamber (WхDхH), mm..................................................................................... ..............
Rated input power (without built-in electric sockets), W........................................................................................... ................................
Maximum allowed load on the built-in electric sockets, W...................................................................................... ...............................3Downflow, m /h.................................................................................................................................................. .........................
Illumination of the work zone, lx.............................................................................................................................. ...............................
Maximum net weight of the cabinet with stand, kg................................................................................................... ................................
1Е-C.001-18.0
2320х750х1940*
1740х640х675
460
1000
400-450
2000
400
1Е-C.001-12.0
1720х750х1940*
1140х640х675
420
1000
270-320
1500
350
1720/2320
7501140/1740
1270
670
1940
BSC Class II I
BMB-III-«Laminar-S» PROTECT
MAIN CHARACTERISTICS
I
H14
G4
200
250
I
two-stage (G4, H14)
two-stage (Н14, H14)
no recirculation
I
MAIN PARAMETERS AND DIMENSIONS
AIRFLOW LAYOUT DIMENSIONAL DRAWING
room airclean (filtered) aircontaminated air
* Dimensions do not account for outstanding supports(The supports stand out of the cabinet's dimensions by 10 mm on each side).
www.lamsys.com
16
Code: 332.120 332.180*
CONTROL SYSTEM COMPONENTS:– control system boards are located under the front panel outside of the contamination area to ensure access for electronic component repair or replacemnt without prior decontamination required;– cabinet control is carried out via a touchscreen control panel;– pass box control panel is pushbutton with door and lock position indication.
WORK CHAMBER:– work chamber is made of stainless steel;– hinged front sashes for equipment loading; each front window has oval-shaped glove ports. The windows are made of laminated safety glass;– 10-liter liquid-collecting tray is located in the bottom part of the work chamber;– LED lighting in the work chamber;– removable set of electric sockets.
SUPPLY AND EXHAUST AIR FILTRATION SYSTEM:– filtration of the air coming into the work chamber is carried out via successive filters: prefilter G4 and supply HEPA filter H14;– filtration of the exhaust air coming out of the cabinet is carried out via two successive exhaust HEPA filters H14;– each HEPA filter has an access for scanning.
LEAK-TIGHT PASS BOX.
The cabinet is designed for simultaneous work of two operators opposite one another.The model of 1800 mm width has 8 glove ports
1200 mm 1800 mm
BMB-II-«Laminar-S»
– Protection of an operator and the environment from aerosols generated during the work with pathogenic biological agents (PBA) and microorganisms including the highest (especially hazardous) risk groups.
– Protection of the work agents in the work chamber from external contamination.
– Possibility to work with small amount of toxic chemicals and radionuclides as well as to remove work agent odors in case of mandatory connection to an active exhaust system.
– Equipment of individual work places in medical, pharmaceutical and other institutions working with I-II-III-IV risk group pathogens and parasitic disease causative microorganisms.
17
Article....................................................................................................................................................... ...............
Maximum dimensions of the cabinet with pass box* (WхDхH), mm.................................................... ............
Dimensions of the work chamber (WхDхH), mm.................................................................................... ...............
Rated input power (without built-in electric sockets), W........................................................................................... ...............................
Maximum allowed load on the built-in electric sockets, W....................................................................................... ..............................3Downflow, m /h.................................................................................................................................................. ........................
Illumination of the work zone, lx.............................................................................................................................. ..............................
Maximum net weight of the cabinet with stand, kg................................................................................................... ...............................
1Е-C.002-18.0
2320х880х1940*
1640х655х685
460
1000
400-450
2000
400
1Е-C.002-12.0
1720х880х1940*
1040х655х685
450
1000
270-320
2000
350
1720/2320
520
1040/1640
1940
1270
555
BSC Class III
BMB-III-«Laminar-S» PROTECT VIS-A-VIS
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1
– for particles of 0.5µm and more.................................................................................................................................................................ISO 5
– for particles of 5.0µm and more..............................................................................................................................ISO M (20; ≥5µm); LSAPC
Cabinet class according to ЕN 12469:2000, NSF/ANSI 49 ...............................................................................................................................
Class of the installed НЕРА filters according to EN 1822-1............................................................................................................................
Prefilter class according to EN779:2002...........................................................................................................................................................
Minimal negative pressure in the working chamber, Pa, not less than ...........................................................................................................
Permanently provided working negative pressure in the working chamber, Pa, not less than........................................................................
Air inflow filtering...................................................................................................................................................................
Air outflow filtering...............................................................................................................................................................
Air recirculation, %........................................................................................................................................................................
:2015
MAIN CHARACTERISTICS
I
H14
G4
200
250
I
two-stage (G4, H14)
two-stage (Н14, H14)
no recirculation
I
MAIN PARAMETERS AND DIMENSIONS
AIRFLOW LAYOUT DIMENSIONAL DRAWING
room airclean (filtered) aircontaminated air
* Dimensions do not account for outstanding supports(The supports stand out of the cabinet's dimensions by 10 mm on each side).
www.lamsys.com
18
DESIGN, MANUFACTURE, SALE AND MAINTENANCE OF HIGH-TECHNOLOGY LABORATORY EQUIPMENT, CLEAN ZONES AND CLEAN ROOMS
MANUFACTURE
SERVICEStrict requirements to service quality are significant part of our corporate culture. Well-experienced service engineers are able to perform any type of works related to maintenance and repair of LAMSYSTEMS equipment.
We continuously test materials, structures and technologies using up-to-date instruments and equipment in our own testing laboratory. As a result, we are able to design the equipment that corresponds to international quality and safety requirements.
About 7000 square meters of production facilities are fitted with up-to-date machines and equipment of well-known international brands, i.e. from German, Japanese and Swiss manufacturers. Annual output amounts to about 2500 units. More than 30 models are serial. LAMSYSTEMS products are supplied to numerous countries of CIS, Europe and Asia. LAMSYSTEMS is a top supplier in the market of the Russian Federation – 80%.
ACCEPTANCE TESTINGAll of the products manufactured by LAMSYSTEMS are certified.The acceptance testing site is designed as a Class 7 clean room and is fitted with qualified equipment for testing each item released.
LABORATORY
LAMSYSTEMS CC
www.lamsys.com
19
WE ARE ABLE TO CREATE CONTINUOUS TECHNOLOGICAL PROCESS LINES BASED ON OUR EQUIPMENT.
The lines can consist of biosafety cabinets, laminar flow cabinets, fume hoods, positive and negative-pressure isolators etc.
LAMINAR FLOW CABINETS
Class III
IVF WORKSTATIONS
PCR-CABINETS FUME HOODS STERILE STORAGE CABINETS
LABORATORY EQUIPMENT
ISOLATORS(GLOVE BOXES)
FOR FACTORIES AND LABORATORIES WITH STRICT REQUIREMENTS TO AIR CLEANLINESS
Class I Class II Type A2 Class II Type B2
Opera
tion w
ith c
ytost
atic
and c
ytoto
xic
pro
duct
s
Opera
tion
with
path
ogenic
bio
logic
al
agents
(P
BA
)
CLEAN ZONESOur protfolio comprises more than 500 clean zones ensuring a required cleanliness class as per ISO 14644-1
Opera
tion
with
P
BA
, sm
all
am
ounts
of
toxi
c and
chem
ical
subst
ance
s
CUSTOMIZED ASSEMBLY
MICROBIOLOGICAL SAFETY CABINETS
www.lamsys.com
Class IIIA SELF-CONTAINED MODULE FOR GLOVE INTEGRITY TESTING
SAFE MATERIAL, OBJECT AND WASTE REMOVAL SYSTEM
A self-contained module for glove integrity testing ensures detection of any defects including the smallest ones like needle punctures.
It consists of a glove port plug with a built-in pressure gauge and a valve for compressed air line connection.
The table top of the Class III cabinets can be featured with an opening for installation of a removable sealed container with two lids: an inner and an outer one.
Published in 2020.Manufacturer reserves its right to change technical specification and configuration of equipment in the course of its further development.
Turgoyak Road, 2/4Miass, Chelyabinsk region456300 RussiaTel./fax: [email protected]
Representative in EU: LAMSYSTEMS GmbH Campus Berlin-Buch, Robert-Rössle-Str. 1013125 Berlin, GermanyTel.: +49 (0)30 9489 2080Fax.: +49 (0)30 9489 [email protected]
MICROBIOLOGICAL SAFETY CABINETS
Class II
ADDITIONAL OPTIONS
– Technical gas tap with electromagnetic valve
– Inflammable gas tap with electromagnetic valve
– Vacuum tap with electromagnetic valve
– Exhaust hood for connecting the cabinet
to the exhaust system
– Additional sockets
– ULPA filters
– Airflow and HEPA-filter integrity testing kit
– Formaldehyde (HCHO) disinfection kit
– Cabinet base stand