operating instructions for temperature sensor model: … · operating instructions for temperature...
TRANSCRIPT
Operating Instructions
for
Temperature Sensor
Model: TWL
Explosion Proof Protection Ex d
TWL
page 2 DT0454
1. Contents
1. Contents ........................................................................................................ 2 2. Note .............................................................................................................. 3 3. Instrument Inspection .................................................................................... 3 4. Regulation Use .............................................................................................. 3 5. Operating Principle ........................................................................................ 4 6. Electrical connection ..................................................................................... 4 7. Use in hazardous Areas ................................................................................ 5
7.1 Area of validity ..................................................................................... 5 7.2 Guidelines ............................................................................................ 5 7.3 General ................................................................................................ 6 7.4 Protection against E.S.D. (Electro Static Discharge) ........................... 6 6.5 Maintenance and repairs ......................................................................... 6 7.5 Storage ................................................................................................ 6
8. Description of the Factory Label.................................................................... 7 9. Installation in the classified area ................................................................... 8
9.1 Examples of installation in explosion-proof areas conform to the protection type “Ex d” ........................................................................... 8
10. Technical Details ........................................................................................... 9 10.1 Sensor wiring ....................................................................................... 9 10.2 General details ..................................................................................... 9 10.3 Materials .............................................................................................. 9 10.4 Process connection ............................................................................. 9 10.5 ATEX-approval .................................................................................. 10 10.6 Head transmitter ................................................................................ 10 10.7 Display ............................................................................................... 10
11. Dimensions ................................................................................................. 11 11.1 Sensors .............................................................................................. 11 11.2 Thermowells ...................................................................................... 12 11.3 Lagging extension at thermowell “T”. ................................................ 17
12. Order details ............................................................................................... 18 13. CE-Certificates ............................................................................................ 20 14. ATEX-Certificates ....................................................................................... 21 15. Notes ........................................................................................................... 25
Manufactured by:
Kobold Mesura S.L.U
C/Guifré 655 08918 Badalona
Tel.: +34 93 460 38 83 Fax: +34 93 460 38 76
E-Mail: [email protected] www.kobold.com
Edition August 2012
TWL
DT0454 page 3
2. Note
Please read these operating instructions before unpacking and putting the unit in operation. Follow the instructions precisely as described herein. The devices are only to be used, maintained and serviced by persons familiar with these operating instructions and in accordance with local regulations applying to Health & Safety and prevention of accidents. When used in machines, the measuring unit should be used only when the machines fulfil the EWG-machine guidelines.
3. Instrument Inspection
Instruments are inspected before shipping and sent out in perfect condition. Should damage to a device be visible, we recommend a thorough inspection of the delivery packaging. In case of damage, please inform your parcel service / forwarding agent immediately, since they are responsible for damages during transit. Scope of delivery: The standard delivery includes: • Device model: TWL • Operating Instructions
4. Regulation Use
The temperature sensors of series TWL are suitable for all applications where processes involving fluids, solid bodies or materials, or gases, require temperature recording and measurement. Our temperature sensors are suitable for use in the following industrial areas: chemicals, petrochemicals, water, feed, food, sanitary, etc. Any use of the Temperature Sensor, model: TWL, which exceeds the manufacturer’s specification may invalidate its warranty. Therefore, any resulting damage is not the responsibility of the manufacturer. The user assumes all risk for such usage.
TWL
page 4 DT0454
5. Operating Principle
Resistance thermometers work by using the continuous change of resistance of metals subject to temperature rising and decreasing. The most common resistance material used is platinum, as it is very stable and has very good repeatability. The temperature coefficient of platinum is positive, so its resistance increases as the temperature rises. This property is defined in the IEC751 standard, which defines measurements deviations categories A and B.
6. Electrical connection
1 x Pt100 2 wires 1 x Pt100 3 wires 1 x Pt100 4 wires 2 x Pt100 2 wires 2 x Pt100 3 wires
Note: For electrical connection of transmitter please refer to their separate manual
TWL
DT0454 page 5
7. Use in hazardous Areas
7.1 Area of validity These security instructions apply to TWL Series temperature sensors and their accessories for use in explosion-proof atmospheres conform to CE certificate LOM 08ATEX2016 X 7.2 Guidelines The TWL Series temperature measuring instruments work according to the heat resistance measuring principle. This measuring principle acts to check and measure the temperature also in explosion-proof areas. TWL devices can be with a standard 4 to 20 mA signal transmitter, Hart protocol, Profibus/Fieldbus or with direct access to the sensor. They are appropriate for use in Group IIC and Categories 2G/D explosion-proof atmospheres. The sensor element can be installed in Category 2 explosion-proof areas in accordance with Certification 2G Ex d. The mechanical connecting element to the process can be installed in Category 2 explosion-proof areas, conform to Certification 2G Ex d. Devices labelled “Ex d” which have a mechanical separating element, such as a thermowell, are approved for Zone 0/20 to Zone 1/21, when using a sensor certified as 2GD. The temperature class and/or the surface temperature relates solely to a device operated at ambient temperature. On installation the actual temperature class for process operation has to be determined. The inlet bushings used must conform to the certification for their type in accordance with the directive.
The requirements of Regulation 94/9/CE, and the applicable national regulations for the use of measuring instruments in explosion-proof areas, such as EN 60079-0:2011, EN 60079-1:2004, EN 61241-0:2006, EN 61241-1:2004 and other regulations relating to this certification type, must be fulfilled. Only qualified specialist personnel may install devices in explosion-proof areas.
TWL
page 6 DT0454
7.3 General • When installing the sensor it is necessary to follow all the instructions and
regulations for explosion-proof areas and the safety instructions included in these instructions.
• Make sure that the details on the sensor’s type label correspond to the working conditions for the application.
• When installing the device, make sure you do not create any mechanical deformation as a result of solder spots or the application of mechanical force.
• Important: Make sure there is an electrical connection between the device’s earth and the earth of the system.
• Make sure the lid are closed before putting the device into operation.
• Before re-opening the lid, remove the plug from the mains or de-energise the device and make sure there is no danger of explosion.
• Only use cable glands according norms for Ex d protection.
• Important: Due to the use of a flame path and its fit tolerance it is not
permissible to use standard measuring inserts as spare parts 7.4 Protection against E.S.D. (Electro Static Discharge) Temperature sensors with plastic parts that can become electrically charged bear a warning label. Electrical charging must be avoided at all costs. Pay attention to the following: - Avoid rubbing the device - Never clean the device dry - Do not install the device near material airflows or near steam outlets 6.5 Maintenance and repairs The instrument does not require maintenance or servicing. Repairs must be only carried out by Kobold Mesura ( manufacturer ) 7.5 Storage Measuring instruments should be protected against humidity and dust. Storage temperature: -40ºC….+85ºC for sensors without transmitter. Storage temperature: See manual of the corresponding transmitter and display model.
TWL
DT0454 page 7
8. Description of the Factory Label
TWL
page 8 DT0454
9. Installation in the classified area
9.1 Examples of installation in explosion-proof areas conform to
the protection type “Ex d”
Hazardous area Safe area
TWL
DT0454 page 9
10. Technical Details
10.1 Sensor wiring
- 2 wires Error due to the lead resistance of the sensor - 3 wires With connecting wires up to 25 m, lead resistance is negligible. - 4 wires The lead resistance of the connecting wires is negligible
10.2 General details
Ambient temperature: -40…+150ºC with ceramic terminal base -40… +85ºC with transmitter -20…..+70ºC with LCD display -20…..+80ºC with led display Meas. Range: -30…..+550ºC
-80…..+600ºC ( Other on request )
Operating pressure: 250 bar ( depending on thermowell ) Connection head: form XD ( IP54÷IP68 depending on cable gland and
sealing “not included” M20 x 1,5mm )
10.3 Materials Sensor: Stainless steel 1.4404 ( others on request ) Thermowell: Stainless steel 1.4404 ( others on request ) Neckpipe: Stainless steel 1.4404 ( others on request ) Connection head: Aluminium, painted Terminal base: Ceramic ( without transmitter )
10.4 Process connection Thread: G1/2” , G3/4, G1”, ½” NPT, ¾” NPT, 1” NPT Din-flange: DN 15, 20, 25, 32, 40, 50 Ansi flange: ½”, ¾”, 1”, 1 ½”, 2“ Weld-in: ¾”, 1”, 1 ¼”
TWL
page 10 DT0454
10.5 ATEX-approval
II 2 GD Exd IIC T6 / Ex tD A21 IP65 T85 ºC
10.6 Head transmitter - Output: analogue output 4…20 mA - Communication: HART®-protocol
PROFIBUS®/Fieldbus - Minimum meas. span: standard transmitter 25ºK transmitter with HART® 10ºK transmitter with PROFIBUS®/Fieldbus 5ºK
- Supply voltage: 8…35 Vdc for standard transmitter and transmitter with Hart
9…32 Vdc for transmitter with PROFIBUS®/Fieldbus
10.7 Display - Only for 4..20 mA or Hart transmitters. - Type: 4 digit LCD or LED - Supply: loop powered - Voltage drop out: LCD max. 2,5 Vdc LED 3,3 V at 4 mA 3,7 V at 20 mA Note For programming of transmitter and display please refer to their separate programming manuals.
TWL
DT0454 page 11
11. Dimensions
11.1 Sensors Dimensions Temperature Sensor TWL-1 Dimensions Temperature Sensor TWL-2
HL: Neckpipe length 130 mm TWL-1 ( other on request ) U: Immersion length thermowell ( see drawing thermowell ) 150 mm TWL-2 ( other on request ) H: Length thermowell ( see drawing thermowell )
LN: Screw-in-length by hand ( approx. 8,1 mm at ½”NPT ) EL: Immersion length EL : U+H-7 mm for thermowell type B EL : U+H-8 mm for thermowell type G and D Measuring Insert TWL-3 Dimensions Connection Head with Display
IMPORTANT!! * Due to the use of a flame path and its fit tolerance it is no permissible to use standard measuring inserts as spare parts.
TWL
page 12 DT0454
11.2 Thermowells Dimensions Thermowell Model TWL-…B… Nominal pressure: max. PN25 at 400ºC Cylindrical thermowell, welded, with process connection G-thread
Nominal pressure: max. PN25 at 400ºC Cylindrical thermowell, welded, with process connection NPT- thread
Process Connection
Max. Total length AF F i d H L Q
G-thread 5000 mm
27 G ½ B 10 12
36 14 26 12 14
32 G ¾ B 10 12
38 16 31.7 12 14
41 G 1 B 10 12
40 18 39 12 14
Process Connection
Max. Total length AF F i d H L
NPT-thread 5000 mm
24 ½ NPT
10 12 42 17
12 14
27 ¾ NPT
10 12 43 18
12 14
36 1 NPT
10 12 46 21
12 14
TWL
DT0454 page 13
Nominal pressure: max. PN6..40 at 400ºC Cylindrical thermowell, welded, with process connection flange acc. DIN or ANSI
Dimensions Thermowell Model TWL-…G…
Nominal pressure: max.PN100 at 400ºC Cylindrical thermowell out of solid material with stepped shaft and process connection G-thread
Process connection
Max. Total Length F i d H e
With flange
ANSI ½”
5000mm
½” NPT
10/12 12/14 40 30 ANSI ¾” ½ NPT
ANSI 1” ¾ NPT ANSI 1” ½ 1 NPT 35 ANSI 2” 1 NPT DIN DN 15 ½ NPT
10/12 12/14 40
30 DIN DN 20 ½ NPT DIN DN 25 ¾ NPT DIN DN 32 1 NPT
35 DIN DN 40 1 NPT DIN DN 50 1 NPT
Process Connection
Max. Total Length AF F i d D H L Q
G-thread 1000mm
27 G ½ B 7-8-9 15 17.5
46 20
26 10-12 17,5 17,5
36 G ¾ B 7-8-9 15 18
31.7 10-12 18 21
41 G 1 B 7-8-9 15 21
51 39 10-12 18 25
TWL
page 14 DT0454
Nominal pressure: max. PN100 at 400ºC Cylindrical thermowell out of solid material withstepped shaft and process connection NPT-thread
Nominal pressure: as flange rating ( max. PN100 at 400ºC ) Cylindrical thermowell out of solid material with stepped shaft and process connection flanges acc. DIN or ANSI
Process Connection
Max. Total Length AF F i d D H L
NPT-thread 1000mm
24 ½ NPT 7-8-9 15 17.5
46 20 10-12 17,5 17,5
27 ¾ NPT 7-8-9 15 18
10-12 18 21
36 1 NPT 7-8-9 15 21
51 25 10-12 18 25
Process Connection
Max.total Length F i d D H e
Flange
ANSI ¾”
1000mm
½ NPT 7-8-9 15 17.5
60
30 10-12 17,5 17,5
ANSI 1” ¾ NPT 7-8-9 15 18
10-12 18 21
ANSI 1 ½” 1 NPT 7-8-9 15 21
35 10-12 18 25
ANSI 2” 1 NPT 7-8-9 15 21
10-12 18 25
DIN DN 20 ½ NPT 7-8-9 15 17.5
60
30 10-12 17,5 17,5
DIN DN 25 ¾ NPT 7-8-9 15 18
10-12 18 21
DIN DN 32 1 NPT 7-8-9 15 21
35
10-12 18 25
DIN DN 40 1 NPT 7-8-9 15 21
10-12 18 25
DIN DN 50 1 NPT 7-8-9 15 21
10-12 18 25
TWL
DT0454 page 15
Nominal pressure: PN100 at 400ºC Cylindrical thermowell out of solid material with stepped shaft and process connection for weld-on
Dimensions Thermowell Model TWL-…D…
Nominal pressure: PN250 at 400ºC Conical thermowell out of solid material with process connection G-thread
Nominal pressure: PN250 at 400ºC
Process Connection
Max. Total Length F i d D H
For weld-on
DN ¾”
1000mm
26,9 7-8-9 15
19 46 10-12 18
DN 1” 33.4 7-8-9 15
22 51 10-12 18
Process Connection
Max. Total Length AF F i d D H L Q
G-thread 1000mm
36 G ¾ B 7-8-9 18
23 46 20 31.7 10-12 21
41 G 1 B 7-8-9 18
29 51 25 39 10-12 21
TWL
page 16 DT0454
Conical thermowell out of solid material with process connection NPT-thread
Process Connection
Max. Total Length AF F i d D H L
NPT-thread 1000mm
27 ¾ NPT 7-8-9 18
23 46 20 10-12 21
36 1 NPT 7-8-9 18
29 51 25 10-12 21
Nominal pressure: as flange rating ( max. PN250 at 400ºC ) Conical thermowell out of solid material with process connection flange acc, DIN or ANSI
Nominal pressure: PN250 at 400ºC
Process Connection
Max. Total Length F i d D H e
Flange
ANSI 1”
1000mm
¾ NPT 7-8-9 18 23
60
30 10-12 21
ANSI 1 ½” 1 NPT 7-8-9 18
29 35 10-12 21
ANSI 2” 1 NPT 7-8-9 18 10-12 21
DIN DN 25 ¾ NPT 7-8-9 18 23
60
30 10-12 21
DIN DN 32 1 NPT 7-8-9 18
29 35
10-12 21
DIN DN 40 1 NPT 7-8-9 18 10-12 21
DIN DN 50 1 NPT 7-8-9 18 10-12 21
TWL
DT0454 page 17
Conical thermowell out of solid material with process connection flange acc. DIN or ANSI
11.3 Lagging extension at thermowell “T”.
In case of ordering with lagging extension at thermowell please see drawing and specify length “T” ( total length is “T”+”H” ). “H” is fix.
Process Connection
Max. Total Length i d D H
For weld-on
DN 1”
1000mm
7-8-9 18 33.4 51
10-12 21
DN 1 ¼” 7-8-9 18
38.1 51 10-12 21
TWL
page 18 DT0454
12. Order details
Order Details (example: TWL-113LGBG4AA0)
Model Sensor specification
Sensor type Sensor type/cat. Sensor wiring Connection head/transmitter Sensor connection
TWL-
0 = without 0 = without 0 = without 0 = without
N2)=1/2” NPT male G= G1/2” male X= special
1= standard 2= with nipple and union
1= 1 x Pt100 cat B -30…+550ºC 2= 1 x Pt100 cat B -30…+550ºC 3= 1 x Pt100 cat B -80…+660ºC 4= 1 x Pt100 cat B -80…+660ºC 5= 1 x Pt100 cat A -30…+550ºC 6= 1 x Pt100 cat A -30…+550ºC 7= 1 x Pt100 cat A -80…+660ºC 8= 1 x Pt100 cat A -80…+660ºC X= special
2= 2-wire 3= 3-wire 41)= 4-wire
L= Atex Exd without transmitter A= Atex Exd with programmable 2-wire Transmitter B= Atex Exd with 2-wire transmitter HART® protocol C= Atex Exd with transnmitter Profibus®/ Fieldbus
3= measuring insert
0= without
Order Details continued:
Thermowell specification Length ( meas. Insert, sensor, thermowell )4)
Special option Thermowell type Process
connection Process
connection size Nominal pressure6)
( process connection ) 0 = without 0 = without 0 = without 0 = without
Only for TWL-0 ( only thermowell )
0 = without lagging extension “T” 1 = with lagging extension “ T “
Sensor with thermowell ( only for TWL-1/TWL-2 )
A = with standard neckpipe “HL” and without lagging extension “T” B = with standard neckpipe “HL” and with lagging extension “T“ C5) = without neckpipe “HL” and with lagging extension “T“ D5) = without neckpipe “HL” and without lagging extension “T“ E = with special neckpipe “HL“ length and with lagging extension “T“ F = with special neckpipe length “HL” and without lagging extension “T”
Sensor without thermowell ( only TWL-1/TWL-2 )
G = with standard neckpipe “HL” H = with special neckpipe “HL” J = without neckpipe length “HL” X = special option ( specify in clear text )
Only TWL-3 M = measuring insert ( only for TWL-3, specify length “ML” )
0 = without 16) = with LCD display 26) = with led display Y = special option ( specify in clear text )
B = cylindrical, multipart, welded G = cylindrical, bar stock/drilled with stepped shank D = tapered shank, bar stock/drilled X = special option
G = G-thread
4 = 1/2” ( not for thermowell type D ) 5 = 3/4" 6 = 1 “ Y= special
A = PN25 ( only for thermowell type B ) B = PN100 ( only for thermowell type G ) C = PN250 ( only for thermowell type D )
N = NPT-thread
4 = 1/2" ( not for thermowell type D ) 5 = 3/4” 6= 1” Y = special
A = PN25 ( only for thermowell type B ) B = PN100 ( only for thermowell type G ) C = PN250 ( only for thermowell type D )
S3) = welded
5 = 3/4" ( only for thermowell type G ) 6 = 1” 7 = 1 1/4" ( only for thermowell type D ) Y = special
B = PN100 ( only for thermowell type G ) C = PN250 ( only for thermowell type D )
F = DIN flange
4 = DN15 5 = DN20 6 = DN25 7 = DN32 8 = DN40 9 = DN50 Y = special
1 = PN6 2 = PN16 3 = PN40 4 = PN100 Y = special
A = ANSI flange
4 = 1/2" 5 = 3/4" 6 = 1” 8 = 1 1/2" 9 = 2” Y = special
5 = 150 lbs 6 = 300 lbs 7 = 600 lbs 8 = 900 lbs ( not for 1/2" ) 9 = 1500 lbs ( not for 1/2" ) Y = special
TWL
DT0454 page 19
1) 4-wire only for 1 sensor 2) choose N for TWL-2 3) not for thermowell type B 4) immersion length “U“ and hole dia. “i“ (when ordering a/with thermowell) or “EL“ (when ordering without thermowell), neckpipe length “HL“ (when different from stdd. i.e. for TWL-1 stdd. Is 130 mm, for TWL-2 standard is150 mm), lagging extension “T“ (when ordering) and measuring insert length “ML“ (when ordering TWL-3) must be specified in clear text when ordering. Pls. check lengths very precisely in order to ensure a perfect match between sensor and thermowell. 5) not for TWL-2 6) display only available for 4...20 mA or Hart transmitters. NOTE: Nominal pressure for TWL-3 and sensors without thermowell is atmospheric pressure. Order Details for ordering only the thermowell (example: TWL-0000NBG4A00)
Model Sensor specification
Sensor type Sensor type/cat. Sensor wiring Connection head/transmitter Sensor connection
TWL-
0 = without 0 = without 0 = without 0 = without
N1)=1/2” NPT male G= G1/2” male X= special
1) choose N for TWL-2 Order Details for ordering only the thermowell continued:
Thermowell specification Length ( meas. Insert, sensor, thermowell )1)
Special option Thermowell type Process
connection Process
connection size Nominal pressure (
process connection ) Please use the specification codes according to order table shown on page 16
0 = without lagging extension “T” 1 = with lagging extension “T”
0 = without Y = special option ( specify in clear text )
1) immersion length “U“ , hole dia. “i“ and lagging extension “T“ must be specified in clear text when ordering. Pls. check lengths very precisely in order to ensure a perfect match between sensor and thermowell.
TWL
page 20 DT0454
13. CE-Certificates
DECLARACIÓN DE CONFORMIDAD CE EC DECLARATION OF CONFORMITY
EG-KONFOMITÄTSERKLÄRUNG DÉCLARATION DE CONFORMITÉ
DICHIARAZIONE DI CONFORMITÀ CE
KOBOLD MESURA SLU Guifré, 655 08918 Badalona (España)
Declara, bajo la propia responsabilidad, que el producto Declares under our sole responsibility, that the product Erklärt in alleiniger Verantwortung, daß das produkt Déclare sous sa seule responsabilité, que le produit Dichiara sotto la propia responsabilità, che il prodotto TWL-1……. TWL-2……. A los cuales se refiere esta declaración, son conformes a las siguiente Directivas Europeas: To which this declaration relates is in conformity with the following European Directives: An auf das diese Erklärung verweist, sie mit den Europäischen Richtlinien im Einklang stehen folgend: À auxquels se réfère cette déclaration, ils sont conformes aux Directives Européennes suivant : A ai quali si riferisce questa dichiarazione, sono conformi alle direttive europee seguente:
EMC89/336/CE Directiva Ex 94/9/EC Normas armonizadas y documentos de la normativa aplicados: Applied harmonised standards and normative documents: Angewandte harmonisierte Normen oder normativer Dokumente: Normes harmonisées et documents normatifs appliqués Norme armonizzate e documenti normativi applicati: EN 61010-1 :2001 EN 60079-1:2004 (acc.EN 60079-1:2005) EN 61241-0:2006 EN 61000-6-2 :2006 EN 61241-1:2004 (acc. EN 61241-1:2005) EN 60079-0:2006 Certificado de examen CE de tipo Marcado EC-type examination certificate Marking EG-baumusterprübescheinigung Markierung Attestation d´examen CE de type Inscription Certificazione per esame di tipo CE Marcatura
LOM 08ATEX2016 X II 2 GD Exd IIC T6 / Ex tD A21 IP65 T85 ºC Fabricado en: KOBOLD MESURA SLU C/ Guifré, 655 08918 BADALONA (Spain) Made in: Hergesteltlt in: Fabriqué dans: Fabbricato in: Organismo notificado : LOM 0163 Número notificación : LOM 05ATEX9070 Notified organism Number notification Mitgeteilter Organismus Zahlmitteilung Organization annoncée Nombre notification Organismo informato Notifica di numero Badalona Junio 2012 Gerente Antonio Sánchez Tomás
TWL
DT0454 page 21
14. ATEX-Certificates
TWL
page 22 DT0454
TWL
DT0454 page 23
TWL
page 24 DT0454
TWL
DT0390 page 25
15. Note
TWL
page 26 DT0390
TWL
DT0390 page 27
KOBOLD MESURA S.L.U C/Guifré 655 08918 Badalona Tel.: +34 93 460 38 83 Fax: +34 93 460 38 76 E-Mail: [email protected] www.kobold.com
Technical data Subject to change without prior notice