operating instructions - laboshopdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · designed...
TRANSCRIPT
![Page 1: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/1.jpg)
Operating Instructions
High-Temperature Furnace (Tabletop Model)
LHT 01/17 D - LHT 03/17 D
-> 01.2016 Original instructions
www.nabertherm.com
Made
in
Germany
![Page 2: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/2.jpg)
2
Copyright
Copyright by Nabertherm GmbH Bahnhofstrasse 20 28865 Lilienthal Federal Republic of Germany Reg: M01.0068 ENGLISCH Rev: 2016-06 No responsibility is accepted for the correctness of this information. We reserve the right to make technical alterations.
![Page 3: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/3.jpg)
3
1 Introduction ..................................................................................................................................................... 5
1.1 Product Description .................................................................................................................................... 6
1.2 Overview of the Complete Furnace ............................................................................................................. 7
1.3 Safeguarding against Dangers Posed by Over-Temperature ........................................................................10
1.3.1 Key to the Model Names .......................................................................................................................11
1.4 Scope of Delivery ......................................................................................................................................11
2 Specifications ................................................................................................................................................. 13
2.1 Warranty and Liability ...............................................................................................................................14
3 Safety ............................................................................................................................................................. 14
3.1 Intended Use .............................................................................................................................................14
3.2 Requirements for the Furnace Operator ......................................................................................................16
3.3 Requirements for the Operating Personnel..................................................................................................17
3.4 Protective Clothing ....................................................................................................................................18
3.5 Basic Measures During Normal Operation .................................................................................................18
3.6 Basic Measures in Case of Emergency .......................................................................................................19
3.6.1 What to do in an Emergency ..................................................................................................................19
3.7 Basic Measures for Servicing and Maintenance ..........................................................................................20
3.8 Environmental Regulations ........................................................................................................................20
3.9 Explanation of the Symbols and Warnings .................................................................................................21
3.10 General Risks with the Furnace ..................................................................................................................24
4 Transportation, Installation, and Commissioning ........................................................................................ 26
4.1 Delivery ....................................................................................................................................................26
4.2 Unpacking .................................................................................................................................................28
4.3 Transportation Securing Equipment/Packaging ..........................................................................................30
4.4 Constructional and Connection Requirements ............................................................................................30
4.4.1 Installation (Furnace Location) ..............................................................................................................30
4.5 Assembly, Installation, and Connection ......................................................................................................32
4.5.1 Venting Exhaust Fumes .........................................................................................................................32
4.5.2 Connecting the Furnace to the Power Supply .........................................................................................33
4.6 Commissioning .........................................................................................................................................34
4.6.1 Molybdenum Disilicide (MoSi2) Heating Elements ................................................................................35
5 Operation ....................................................................................................................................................... 38
5.1 Controller ..................................................................................................................................................38
6 Operation, Display and Switch Elements (depending on design) ................................................................. 39
6.1 Turning on the Controller/Furnace .............................................................................................................39
6.2 Turning off the Controller/Furnace.............................................................................................................40
6.3 Air Inflow Lever........................................................................................................................................40
6.4 Loading/Charging ......................................................................................................................................41
6.4.1 Stackable Saggar ...................................................................................................................................41
7 Servicing, Cleaning, and Maintenance .......................................................................................................... 44
7.1 Furnace Insulation .....................................................................................................................................45
7.2 Shutting the System Down for Maintenance ...............................................................................................45
7.3 Regular Maintenance of the Furnace ..........................................................................................................46
7.4 Regular Maintenance Tasks – Documentation ............................................................................................48
![Page 4: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/4.jpg)
4
7.5 Regular Maintenance Tasks – Heating Elements/Furnace Chamber ............................................................48
7.6 Checking the Heating Element Screws .......................................................................................................49
7.7 Operating and Auxiliary Materials .............................................................................................................51
7.8 Cleaning Products......................................................................................................................................51
8 Malfunctions .................................................................................................................................................. 52
8.1 Error Messages of the Controller ................................................................................................................53
8.2 Warnings of the Controller.........................................................................................................................55
8.3 Malfunctions of the Switchgear ..................................................................................................................57
8.4 Controller Check List ................................................................................................................................58
9 Spare Parts/Wearing Parts ............................................................................................................................ 60
9.1 Electrical Schematics/Pneumatic Schematics..............................................................................................62
9.2 Replacing a Heating Element .....................................................................................................................63
9.3 Replacing a Thermocouple ........................................................................................................................67
9.4 Replacement/Readjustment of the Door Insulation Structure ......................................................................70
9.5 Replacing a Fuse .......................................................................................................................................71
9.6 Repairing the Insulation .............................................................................................................................73
10 Accessories (Options)..................................................................................................................................... 73
10.1 Gas Supply System (Accessories) ..............................................................................................................73
10.2 Operation of Compressed Gas Containers ..................................................................................................75
10.3 Separate the Snap-In Coupling (Plug) from the Furnace Housing ................................................................76
10.4 Over-temperature Limit Controller .............................................................................................................76
10.5 Electrical Schematics/Pneumatic Schematics..............................................................................................77
11 Nabertherm Service ....................................................................................................................................... 77
12 Shut-Down, Dismantling, and Storage .......................................................................................................... 78
12.1 Transportation/Return Transportation.........................................................................................................78
13 Declaration of Conformity ............................................................................................................................ 80
![Page 5: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/5.jpg)
5
Pos: 1 / TD/Einleitu ng/Ü bersc hrift - Einleit ung 1 @ 0\m od_ 116 782 321 2238 _51 .docx @ 5 139 @ 1 @ 1
1 Introduction Pos: 2 / TD/Einleitu ng/ Öfen @ 0\ mod _11 581 572 275 33_ 51.d ocx @ 208 4 @ @ 1
Dear Customer,
Thank you for choosing a quality product from Nabertherm GmbH.
You can be proud that you have chosen a furnace which has been especially tailored to suit
your manufacturing and production conditions.
This product is characterized by
professional workmanship
high performance due to its high efficiency
high-quality insulation
low power consumption
low noise level
simple installation
easy to maintain
high availability of spare parts
Your Nabertherm Team
Note
These documents are intended only for buyers of our products and may not be copied or
disclosed to third parties without our written consent. (Law governing copyright and
associated protective rights, German Copyright Law from Sept. 9, 1965)
Protective Rights
Nabertherm GmbH owns all rights to drawings, other documents and authorizations, also in
case of applications for protective rights.
Note
All the figures in the instructions have a descriptive character; in other words, they do not
represent the exact details of the furnace.
Pos: 3 / TD/Einleitu ng/Di e in d er Anl eitun g ge zeigt en Abbil dun gen k önn en a bhä ngig von Fu nktion, ... Ofen mod ell abw eiche n @ 24\m od_ 133 785 435 224 2_51 .docx @ 1 616 05 @ @ 1
Note
The pictures contained in the instruction manual may contain inaccuracies in terms of the
function, design and furnace model.
![Page 6: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/6.jpg)
6
Pos: 4 / TD/Einleitu ng/Pr oduk tbesc hrei bun g/Öf en/Üb ersc hrift - Pro dukt besch reib ung 1.1 @ 0\ mod _11 678 2194 380 7_5 1.do cx @ 5103 @ 2 @ 1
1.1 Product Description Pos: 5 / TD/Einleitu ng/Pr oduk tbesc hrei bun g/Öf en/Pro dukt besch reib ung -Bei di esen elekt risch b eheiz ten Öfen han delt es sich um ein Q ualitä tspr odukt ... @ 15\ mo d_1 3057 072 549 35_ 51.d ocx @ 117 576 @ @ 1
These electrically heated furnaces are a high-quality product which will give you many
years of reliable service if they are properly cared for and maintained. One basic
prerequisite is that the furnace is used the way it was designed to be used.
During development and production a high priority was placed on safety, functionality and
economy. Pos: 6 / TD/Einleitu ng/Pr oduk tbesc hrei bun g/Öf en/Pro dukt besch reib ung -LH T 01 -03 /17 D - Ho chte mpe rat uröf en (Mo dell 2 015 ) - Lab or @ 11 8\m od_1 461 855 247 901 _51. docx @ 46 785 0 @ @ 1
Designed as tabletop models, these compact high-temperature furnaces have a variety of
advantages. The first-class workmanship using high-quality materials, combined with ease
of operation, make these furnaces all-rounders in research and the laboratory. Pos: 7 / TD/Einleitu ng/Pr oduk tbesc hrei bun g/Öf en/Pro dukt besch reib ung -LH T 01 -03 /17 D - Ho chte mpe rat uröf en (Mo dell 2 015 ) - D ent al @ 1 18\ mod _146 219 478 141 0_5 1.doc x @ 4 690 33 @ @ 1
These high-temperature furnaces are perfectly suited for sintering of translucent zirconia units. The special molybdenum discilicide heating elements offer best possible protection
against chemical interaction between charge and heating elements. Pos: 8 / TD/Einleitu ng/Pr oduk tbesc hrei bun g/Öf en/Die Zirk onei nheit en we rde n in k era misch en Ch arg enb ehält ern positi onie rt. Bis zu dr ei Cha rge nbe hälte r @ 118\ mod _14 621 945 378 72_5 1.d ocx @ 469 006 @ @ 1
The zirconia units are positioned in ceramic saggars. Up to three saggars can be stacked into the high-temperature furnace.
Pos: 9 / TD/Einleitu ng/Pr oduk tbesc hrei bun g/Öf en/Bei Ve rwen dun g von Fä rbel ösun gen ist d ara uf zu acht en, dass die Wa re volls tändi g ge troc knet is t, ... @ 1 18\ mod _146 191 060 520 0_5 1.doc x @ 4 678 91 @ @ 1
When using liquid shading colors it must be ensured that the charge has been fully dried
before starting the sintering process. Pos: 10 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Ü bers chrift - Zusä tzlich zeic hne t sich dieses Prod ukt a us d urch: @ 1 5\m od_ 130 5713 428 306 _51. docx @ 11 772 4 @ @ 1
Other Characteristics of this Product are: Pos: 11 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Pr odu ktbesc hr eibun g-G run dauf bau -Hoc htem per atu röfe n LH T 0 1-0 3/17 D ( Mod ell 201 5). .. Den tal-Ka talog @ 1 18\ mod_ 146 191 209 016 4_5 1.docx @ 4 679 45 @ @ 1
Tmax 1650 °C (3002 °F)
Furnace chamber with a volume of 1 or 2 liters
Special heating elements made of molybdenum disilicide offer best possible
protection against chemical interaction between charge and heating elements
Furnace chamber lined with first-class, long-life fiber material, unclassified
Housing made of sheets of textured stainless steel
Dual shell housing with additional fan cooling for low surface temperature
Compact design with lift door, opening upwards
Adjustable air inlet
Exhaust air opening in the roof
Precise temperature control, also in the lower temperature range for drying
Controller P480 (LHT 01/17 D) or P470 (LHT 03/17 D)
NTLog for Nabertherm Controller: Recording of process data with USB-flash drive
Switchgear with adjustable control of the heating elements
Delivery incl. starter set to charge the zirconia works Pos: 12 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Ü bers chrift - Zusa tzauss tatt ung (Einzug Mitt e) @ 39\ mo d_13 649 840 854 02_ 51.d ocx @ 221 200 @ @ 1
Additional Equipment Pos: 13 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Pr odu ktbesc hr eibun g - Zus atza usstatt ung : Temp era turw ählbe gr enze r mit einst ellbar er . .. @ 116\ mo d_1 4593 325 707 94_ 51.d ocx @ 459 785 @ @ 1
Over-temperature limiter with adjustable cutout temperature for thermal protection
class 2 in accordance with EN 60519-2 as temperature limiter to protect the furnace
and load Pos: 14 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Pr odu ktbesc hr eibun g - Zu beh ör: Sch utzg asansc hluss für nicht bre nnb are Sc hutz - od er . .. LH T 0 3/17 D @ 1 18\ mod _14 6191 486 897 1_5 1.do cx @ 4 679 72 @ @ 1
Protective gas connection for non-flammable protective or reaction gases for LHT 03/17 D
Pos: 15 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Pr odu ktbesc hr eibun g - Zu beh ör: Man uelles ode r a utom atisch es Beg asun gssyste m. @ 11 6\m od_1 459 332 575 272 _51. docx @ 45 981 2 @ @ 1
Manual or automatic gas supply system Pos: 16 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Ü bers chrift - Zub ehö r (Einz ug Mitte ) @ 3 6\m od_ 135 997 1094 458 _51 .docx @ 2 0840 6 @ @ 1
Accessories Pos: 17 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Pr odu ktbesc hr eibun g - Zu beh ör: Eckig e Ch arg enb ehält er st apel bar für Be schicku ng in bis zu dr ei Eben e @ 118\ mo d_14 619 148 735 49_ 51.d ocx @ 467 999 @ @ 1
Square saggar for charging of up to three layers Pos: 18 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Pr odu ktbesc hr eibun g - Zu beh ör: Pr ozesss teue run g u nd - doku me ntatio n ü ber VCD -Softwa rep aket zur Ü berw a @ 116\ mod _14 593 327 106 04_ 51.d ocx @ 459 839 @ @ 1
Process control and documentation via VCD software package for monitoring, documentation and control
![Page 7: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/7.jpg)
7
Pos: 19 / TD/Einleit ung /Lief eru mfa ng/Öf en/Ü bersc hrift - G esa mtüb ersich t de r Anla ge @ 1\m od_ 117 430 263 6992 _51 .docx @ 1 133 2 @ 2 @ 1
1.2 Overview of the Complete Furnace Pos: 20 / TD/Einleit ung /Lief eru mfa ng/Öf en/ Gesa mtüb ersic ht Hoc hte mpe ratu röf en LH T 0 1-0 3/1 7 D ( Mod ell 20 15) @ 1 18\m od_ 146 191 746 798 7_51 .docx @ 4 680 27 @ @ 1
LHT 03/17 D (similar to picture)
LHT 01/17 D (similar to picture)
Fig. 1: High-temperature furnace (the figure shows a special design with over-temperature limiter)
![Page 8: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/8.jpg)
8
Pos: 21 / TD/Einleit ung /Lief eru mfa ng/Öf en/ Gesa mtüb ersic ht Rück ansich t Hoch tem per atu röfe n LH T 0 1-0 3/17 D ( Mod ell 201 5) @ 11 8\m od_ 146 2176 175 203 _51. docx @ 46 806 1 @ @ 1
Fig. 2: High-temperature furnace – rear view (similar to picture)
Pos: 22 / TD/Einleit ung /Lief eru mfa ng/Öf en/ Gesa mtüb ersic ht_ M01. 006 8_Tabell e Nr. und Bene nnu ng d er v erb aute n Teile f ür LH T 0 1-0 3/1 7 D @ 118\ mo d_14 618 532 290 53_ 51.d ocx @ 467 823 @ @ 1
No. Name
1 Lift door
2 Handle
3 Insulation from non-classified fiber material
4 Heating elements made from molybdenum disilicide (MoSi2)
5 Furnace chamber
6 Thermocouple
7 Floor plate LHT 01/17 D (accessory)
8 Fresh-air flap to control fresh air
9 Over-temperature limiter with adjustable cutout temperature for thermal protection class 2 in accordance
with EN 60519-2 as temperature limiter to protect the furnace and load (accessory)
10 Controller P480 (LHT 01/17 D) or P470 (LHT 03/17 D)
11 USB interface
12 Exhaust air tube
13 Power switch with integrated fuse (switching furnace on/off)
14 Additional power connection (for accessories)
15 Fuse for additional power connection (for accessories)
16 Ethernet interface (accessory)
17 Power plug with snap-in coupling
![Page 9: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/9.jpg)
9
Pos: 23 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Ü bers chrift - Zusa tzauss tatt ung (Einzug links) @ 3 6\mo d_1 360 061 559 626 _51. docx @ 20 848 5 @ @ 1
Additional Equipment Pos: 24 / TD/Einleit ung /Lief eru mfa ng/Öf en/ Gesa mtüb ersic ht Zusatz ausst attu ng_ Zub ehö r: TWB-TWW_B 410 _C450 _P48 0 ( Modell 201 5) @ 11 6\m od_ 145 8810 060 859 _51. docx @ 45 855 1 @ @ 1
Over-temperature limiter with adjustable cutout temperature for thermal protection class 2 in accordance
with EN 60519-2 as temperature limiter to protect the
furnace and load
Fig. 3: Example (similar to picture)
Pos: 25 / TD/Einleit ung /Lief eru mfa ng/Öf en/ Gesa mtüb ersic ht Zusatz ausst attu ng_ Zub ehö r: Be gasu ngsp aket f ür LHT 03/1 7 D ( Mo dell 2 015 ) @ 1 18\ mod _14 621 9289 124 0_5 1.d ocx @ 4684 82 @ @ 1
Protective gas connection for non-
flammable protective and reaction
gases for LHT 03/17 D
Gas supply system for non-flammable
protective and reaction gases with
shut-off valve and flow meter with control valve, piped and ready to
connect (similar to picture)
Fig. 4: Example (similar to picture)
Pos: 26 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Ü bers chrift - Zub ehö r (Einz ug links ) @ 53\ mod _13 898 6802 028 4_5 1.do cx @ 2 631 75 @ @ 1
Accessories Pos: 27 / TD/Einleit ung /Lief eru mfa ng/Öf en/ Gesa mtüb ersic ht Zusatz ausst attu ng_ Zub ehö r: Ru nde Cha rge nbeh älte r (LH T 0 3/1 7 D) @ 11 8\m od_ 1462 188 363 741 _51. docx @ 46 820 5 @ @ 1
Round saggar
Saggars are recommended when working with zirconium oxide. A saggar consists of the sintering bowl as the
bottom part and the spacer ring with ventilation
openings. The material is very resistant to temperature
changes and can also be used with short heating and
cooling times.
Fig. 5: Round saggar for LHT 03/17 D (similar to picture)
Pos: 28 / TD/Einleit ung /Lief eru mfa ng/Öf en/ Gesa mtüb ersic ht Zusatz ausst attu ng_ Zub ehö r: Eckig e Cha rge nbe hälte r mi t Deckel (LH T 0 1/17 D ) @ 11 8\m od_ 1462 180 194 169 _51. docx @ 46 808 8 @ @ 1
Rectangular saggars
For optimum utilization of the furnace chamber, the
material is placed in ceramic saggars. Up to three saggars
can be stacked in the furnace. The saggars have slits to
allow the air to circulate. The top saggar can be closed
with a ceramic lid.
Fig. 6: Rectangular saggar with lid for LHT 01/17 D (similar to picture)
![Page 10: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/10.jpg)
10
Pos: 29 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Ü bers chrift - Absic her ung von Gefa hre n b ei Übe rte mpe ratu r @ 37\ mod _136 248 973 088 4_5 1.doc x @ 2 139 87 @ 2 @ 1 Safety co ntr ol circui t
1.3 Safeguarding against Dangers Posed by Over-Temperature Pos: 30 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Sich erh eitsein richt ung - Üb erwac hun g d er Of enr au mte mpe ratu r ( TWB-TWW ) - B eschr eibu ng @ 37\ mod _13 624 9121 427 0_5 1.d ocx @ 2140 92 @ @ 1
Over-temperature limiters with manual reset/with automatic reset to protect against over-
temperature in the furnace chamber are available for Nabertherm GmbH furnaces either as
a standard feature (depending on the model series) or as additional equipment (customized
design).
The over-temperature limiter with manual reset/with automatic reset monitors the furnace
chamber temperature. The display shows the most recently set cut-off temperature. If the
furnace chamber temperature rises about the pre-set cut-off temperature the heating is shut
down to protect the furnace, the charge and/or the operating equipment.
Pos: 31 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Wa rnsy mbol _Gef ahr - Absic her ung v on G efa hre n bei Ü ber tem per atu r a m TWB- TWW @ 37\ mod _136 249 040 489 3_5 1.doc x @ 2 140 67 @ @ 1
DANGER
• Danger caused by incorrectly entered cut-off
temperature at the over-temperature limiter with
manual reset/over-temperature limiter with automatic
reset.
• Mortal danger
• If, as a result of over-temperature from the charge and/or the
operating equipment, a charge is likely to be damaged at this pre-
set cut-off temperature of the over-temperature limiter with
manual reset/over-temperature limiter with automatic reset, or if
the charge itself becomes a source of danger for the furnace or its
surroundings, the cut-off temperature must be reduced at the over-temperature limiter with manual reset/automatic reset to the
maximum permissible value.
Pos: 32 / TD/Bet rieb _Bedie nun g/Vor Inbe trie bna hme des Ofens ist di e Bedie nun gsanl eitun g... (TWB- TWW) - Text @ 37\ mod _13 624 9408 022 5_5 1.do cx @ 2141 42 @ @ 1
Read the operating instructions of the over-temperature limiter with manual reset/with automatic reset before starting the furnace. The safety sticker must be removed from the
over-temperature limiter with manual reset/with automatic reset. Any time a change is
made in the heat treatment program, the maximum permissible cut-off temperature (alarm
trigger temperature) at the over-temperature limiter with manual reset/with automatic reset
must be checked or re-entered. Pos: 33 / TD/Bet rieb _Bedie nun g/die maxi male S olltem per atu r des Wär mep rog ra mms C ontr oller zwische n 5 °C u nd 3 0 °C ( TWB-TWW ) @ 3 9\m od_ 136 378 461 0371 _51 .docx @ 2 185 00 @ @ 1
We recommend setting the maximum setpoint temperature of the heating program in the
limiter between 5 °C and 30 °C, depending on the physical characteristics of the furnace, below the trigger temperature of the over-temperature limiter with manual reset/with
automatic reset. This prevents an unwanted triggering of the over-temperature limiter with
manual reset/with automatic reset.
Pos: 34 / TD/Bet rieb _Bedie nun g/Aufkle ber vom TWB/TWW abzie hen - G rafik @ 37 \mo d_1 362 492 0200 88_ 51. docx @ 21 4117 @ @ 1
Description and function, see the Operating Instructions of the over-
temperature limit controller/guard
Fig. 7: Removing the sticker (similar to picture)
![Page 11: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/11.jpg)
11
Pos: 35 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/Ü bers chrift - Ent schlüss elun g de r M odellb ezeich nun g @ 2\m od_1 184 245 078 907 _51. docx @ 19 775 @ 3 @ 1
1.3.1 Key to the Model Names Pos: 36 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/En tschlüss ellun g de r M odell bezeic hnu ng LH T ... /17 D - Ho chte mpe rat uröf en (Tisc hmo delle 201 5) @ 11 9\mo d_1 462 261 697 997 _51. docx @ 46 924 1 @ @ 1
Example Explanation
LHT 03/17 D
LHT = High-Temperature Furnace
LHT 03/17 D
01 = 1 liter furnace chamber (volume in L)
03 = 2 liter furnace chamber (volume in L)
LHT 03/17 D
17 = Tmax 1650 °C (3002 °F)
LHT 03/17 D
D = Dental model
Pos: 37 / TD/Einleit ung /Prod uktb eschr eibu ng/Ö fen/En tschlüss ellun g de r M odell bezeic hnu ng LH T ... /17 D - Ho chte mpe rat uröf en (Mo dell 2 015 ) - Gra fik @ 1 19\ mod _14 622 6177 784 2_5 1.do cx @ 4692 68 @ @ 1
Fig. 8: Example model designation (type plate)
Pos: 38 / TD/Einleit ung /Lief eru mfa ng/Öf en/Ü bersc hrift - Li efe rumf ang @ 0\ mo d_1 1678 225 081 30_ 51.d ocx @ 511 2 @ 2 @ 1
1.4 Scope of Delivery Pos: 39 / TD/Einleit ung /Lief eru mfa ng/Öf en/ Liefe rum fan g - LHT . ../1 7 D - H ocht emp era tur öfen ( Tischm odell e 20 15) @ 1 19\m od_ 146 227 765 879 6_51 .docx @ 4 693 76 @ @ 1
The scope of supply includes:
System component Quantity Comment
High-Temperature Furnace LHT ../17 D 1 x Nabertherm GmbH
Allen key
1 x Nabertherm GmbH
Exhaust air pipe
1 x Nabertherm GmbH
Mains power cable
1 x Nabertherm GmbH
LHT 03/17 D SN 123456 2013
123456789 1650 °C -
- 220-240 V 1/N/PE~ -
50/60 Hz max.13,0 A 3,0 kW
![Page 12: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/12.jpg)
12
System component Quantity Comment
..
Charge saggars (starter-set)
1 x Nabertherm GmbH
.. Sintering dish/Spacer ring2)
4) Nabertherm GmbH
Gas supply system1) 1 x Nabertherm GmbH
Process documentation VCD software
package2)
1 x Nabertherm GmbH
Other components depending on variant - - - See delivery documents
Type of document Quantity Comment
Operating instructions, High-Temperature Chamber Furnace
1 x Nabertherm GmbH
Operating instructions, controller 1 x Nabertherm GmbH
Operating instructions, over-temperature limit controller1)
1 x Nabertherm GmbH
Operating instructions for gas injection pane2) 1 x Nabertherm GmbH
Operating Instructions VCD software package2)
1 x Nabertherm GmbH
Other documents depending on variant 1 x
1) = in scope of delivery depends on design/furnace model 2) = in scope of delivery depend on need, see shipping papers
3) = quantity depends on furnace model
4) = quantity depends on on need, see shipping papers
Note
Please carefully retain all documents. During fabrication and before delivery, all the
functions of this furnace system have been tested.
Pos: 40 / TD/Bet rieb _Bedie nun g/Die mitgeli efer ten U nte rlage n bei nhalt en nicht zw angsl äufig elekt rische Schaltpl äne bzw. Pn eum a ... @ 47 \mo d_1 380 028 074 616 _51. docx @ 24 766 8 @ @ 1
Note
The documents included do not always contain the electrical schematics and pneumatic
schematics.
If you need the respective schematics they can be ordered from Nabertherm Service.
Betriebsanleitung
Ma dein
Ger many
ProzessdokumentationControltherm MV
![Page 13: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/13.jpg)
13
Pos: 41 / TD/Einleit ung /Te chnisc he Da ten/ Öfen /Übe rschri ft - Tec hnisch e Dat en - mit Hi nweis @ 0\m od_ 116 782 284 073 7_51 .docx @ 5 121 @ 1 @ 1
2 Specifications
Electrical specifications are on the type plate located on the side of the furnace.
Pos: 42 / TD/Einleit ung /Te chnisc he Da ten/ Öfen /Mo dell- Tab elle fü r LH T .. ./17 D - H ochte mpe rat urö fen (Tisc hmo delle 201 5) - 1 @ 11 9\m od_1 462 263 874 337 _51. docx @ 46 929 5 @ @ 1
Model Tmax Inner dimensions in mm
Volume Maximum units
Outer dimensions in mm
Weight Minutes
°C w d h in l W D H+c2 in kg to Tmax1
LHT 01/17 D 1650 110 120 120 1 45 382 425 525+195 28 10
LHT 03/17 D 1650 135 155 200 2 75 470 620 770+260 75 60 1If connected at 230 V, 1/N/PE (including saggars) 2Including opened lift door
Pos: 43 / TD/Einleit ung /Te chnisc he Da ten/ Öfen /Mo dell- Tab elle fü r LH T .. ./17 D - H ochte mpe rat urö fen (Tisc hmo delle 201 5) - Gr afik - 1 @ 119 \mo d_1 462 263 9926 53_ 51. docx @ 469 322 @ @ 1
Fig. 9: Dimensions
Pos: 44 / TD/Einleit ung /Te chnisc he Da ten/ Öfen /Ta belle tech n. Dat en f ür LH T .. ./17 D - H ocht emp erat ur öfen (Tisc hm odelle 201 5) - 1 @ 11 9\m od_ 146 226 9305 985 _51 .docx @ 46 934 9 @ @ 1
Model Electrical
Connection
Voltage Electric
current
Connected
load
Frequency
in V in A in kW in Hz
LHT 01/17 D 1/N/PE or 2/PE
220 - 240 14,8 3,4 50 or 60
LHT 03/17 D 220 - 240 13,0 3,0 50 or 60
Thermal Protection Class Furnaces: According to DIN EN 60519-2
Without safety controller: class 0
With safety controller: class 2
Protection Class Furnaces: IP20
Ambient Conditions for
Electrical Conditions
Temperature:
Humidity:
+5 °C to +40 °C
max. 80 % noncondensing
H
D
B
h d b
C
![Page 14: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/14.jpg)
14
Model Electrical
Connection
Voltage Electric
current
Connected
load
Frequency
in V in A in kW in Hz
Weight Furnace with
accessories:
Depending on model (see delivery documents)
Pos: 45 / TD/Einleit ung /Te chnisc he Da ten/ Öfen /Ta belle D aue rsch alldruc kpeg el < 80 dB(A) - 2 @ 1\m od_ 117 075 098 5488 _51 .docx @ 8 913 @ @ 1
Emissions Continuous sound pressure level:
< 80 dB(A)
Pos: 46 / TD/Einleit ung /Gewä hrleis tun g_Haf tung /Übe rsch rift - Gewä hrleist ung un d Haft ung 1.1 @ 0\ mod _11 678 229 7949 2_5 1.d ocx @ 5130 @ 2 @ 1
2.1 Warranty and Liability Pos: 47 / TD/Einleit ung /Gewä hrleis tun g_Haf tung /Öfe n un d Schalt anla gen - G ewäh rleistu ng u nd H aftu ng @ 0\m od_ 115 7536 440 972 _51. docx @ 15 69 @ @ 1
§ As regards warranty and liability, the normal Nabertherm warranty terms apply,
unless individual terms and conditions have been agreed. However, the following
conditions also apply:
Warranty and liability claims for personal injury or damage to property shall be excluded if
they are attributable to one or more of the following causes:
Everyone involved in operation, installation, maintenance, or repair of the furnace
must have read and understood the operating instructions. No liability will be accepted
for damage or disruptions to operation resulting from non-compliance with the
operating instructions.
Not using the furnace as intended,
Improper installation, start-up, operation, or maintenance of the furnace,
Operation of the furnace with defective safety equipment or improperly installed or
non-functioning safety and protective equipment,
Not observing the references in the operating instructions to transportation, storage,
installation, start-up, operation, maintenance, or equipping the furnace,
Making unauthorized changes to the furnace,
Making unauthorized changes to the operating parameters,
Making unauthorized changes to the parameterization, the settings, or the program,
Original parts and accessories are designed especially for Nabertherm furnaces.
Replace parts only with original Nabertherm parts. Otherwise the warranty will be
void. Nabertherm accepts absolutely no liability for damage caused by using parts that
are not original Nabertherm parts.
Catastrophes due to third-party causes and force majeure. Pos: 48 / TD/Sich erh eit/Übe rsch rift - Sicher heit @ 0\ mod _11 588 4396 154 0_5 1.d ocx @ 3103 @ 1 @ 1
3 Safety Pos: 49 / TD/Sich erh eit/Übe rsch rift - Bestim mu ngsg emä ße Ve rwen dun g @ 0\m od_ 1167 823 503 921 _51. docx @ 51 48 @ 2 @ 1
3.1 Intended Use Pos: 50 / TD/Sich erh eit/Besti mmu ngsg emä ße Verwe ndu ng all gem eine r Ofe n-Anl agen _M ateri alien - Teil 1 @ 24\ mod _13 379 387 759 36_ 51.d ocx @ 161 920 @ @ 1 Safety
The Nabertherm furnace was designed and built in conformance with a careful selection of the applicable harmonized standards and other technical specifications. Hence, it
corresponds to the state of the art and assures the greatest degree of safety.
Only materials whose characteristics and melting temperatures are known may be heated.
Consult any available safety-related material data sheets.
Pos: 51 / TD/Sich erh eit/Besti mmu ngsg emä ße Verwe ndu ng Ka mme röf en -D ent albe reich LHT 02/1 7 LB... - LHT 03/1 7 D @ 46\ mod _13 7776 470 878 4_5 1.do cx @ 2 435 90 @ @ 1
Furnaces in this series are suitable for sintering technical ceramic pieces such as bridges and crowns made of zirconium oxide.
Pos: 52 / TD/Sich erh eit/De r Einsatz von Fä rbel ösun gen k ann zu ei nem üb erp rop ortio nale n Versc hleiß de r Heizel eme nte füh ren. .. @ 43\m od_ 137 111 529 958 8_51 .docx @ 2 335 30 @ @ 1 it
The use of paint solvents can lead to accelerated wear of the heating elements. If solvents are used, good pre-drying is necessary to minimize their impact on the heating elements.
![Page 15: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/15.jpg)
15
Pos: 53 / TD/Sich erh eit/Eine and ere ode r d arü ber hina usge hen de Be nutzu ng, wi e z.B. di e Ver arb eitun g ... @ 45 \mo d_1 377 242 071 687_ 51. docx @ 24 204 0 @ @ 1 Safety
Any other use, such as processing of products other than those for which the furnace was intended as well as handling hazardous materials or materials dangerous to health
is deemed IMPROPER and such uses must be approved in writing by Nabertherm. Pos: 54 / TD/Sich erh eit/Von den im Of en einge setzt en Mate rialien bzw. Aus gas unge n kö nne n sich unte r Um stän den Sc hadst offe .. @ 45\ mod _13 772 433 9221 7_5 1.d ocx @ 242 065 @ @ 1
Under certain circumstances gases or materials may be released from the materials in
the furnaces that settle on the insulation or the heating elements and destroy them. If
applicable, read the labels and instructions on the packaging of materials that
you use. Pos: 55 / TD/Sich erh eit/Bei Öf en mit Tem per aturw ählb egr enze r m uss di e Absch altte mpe ratu r so eing estellt wer den . .. @ 45\ mod _13 7724 441 712 2_5 1.do cx @ 2 423 90 @ @ 1
Furnaces with over-temperature limit controllers must have their shut-down temperatures set to prevent any overheating of the material.
Pos: 56 / TD/Sich erh eit/Ver änd eru nge n a m Ofe n, m üsse n mit N abe rth erm schri ftlich a bges tim mt we rden ... @ 45\ mo d_1 3772 433 986 94_ 51.d ocx @ 242 240 @ @ 1
The set-up instructions and safety regulations must be followed, otherwise the furnace
will be considered improperly used, effectively cancelling any claims against
Nabertherm GmbH. The EC Declaration of Conformity will cease to be valid if any
modifications are made to the machine without our approval. Pos: 57 / TD/Sich erh eit/Die Auf stellhinw eise und Sich er heitsb estim mun gen sind einzu halte n, a nde rnfalls gilt d er Of en als ... @ 45 \mo d_1 377 243 4105 07_ 51. docx @ 24 2265 @ @ 1
The set-up instructions and safety regulations must be followed, otherwise the furnace will be considered improperly used, effectively cancelling any claims against
Nabertherm GmbH. Pos: 58 / TD/Sich erh eit/Das Öffn en d es Of ens i m hei ße n Zustan d ü ber 200 °C (39 2 °F) kan n zu eine m e rhöh tem Versc hleiß ... @ 39\ mod _13 642 8991 163 5_5 1.do cx @ 2199 45 @ @ 1
Opening the furnace while it is still hot, over 200 °C (392 °F), can lead to increased
wear of the following components: insulation, door seal, heating elements and furnace
housing. No liability shall be accepted for any damage to the goods or the furnace
resulting from non-compliance with this warning. Pos: 59 / TD/Sich erh eit/De r Betri eb mit Kraf tqu ellen, Pr odu kten, Betri ebsmit teln, Hilfsstoff en usw., .. . @ 4 5\m od_ 137 724 108 3975 _51 .docx @ 2 419 90 @ @ 1
Operation with power sources, products, operating equipment, auxiliary materials, etc., which are listed as hazardous or which may in any way harm the health of the operator is
prohibited.
The furnace must not be filled with materials or substances that release explosive gases or vapors. Only materials and substances whose properties are known may be used.
Pos: 60 / TD/Allge mein e Hinwei se ( für alle Anleit ung en) /Hinweis - Da uer hafte r Bet rieb bei maxim aler Te mpe rat ur k ann z u e rhö hte m Versc hleiß an H eizele ment en @ 17\ mo d_13 100 506 109 42_ 51.d ocx @ 123 376 @ @ 1
Caution
Continuous operation at maximum temperature can lead to increased wear of the heating
elements and the insulation material. We recommend operating at approx. 50 °C below the
maximum temperature.
Pos: 61 / TD/Allge mein e Hinwei se ( für alle Anleit ung en) /Hinweis - Dies er Of en ist für die gewe rblich e Anwen dun g ... Der Ofen ist nic ht fü r die Erwär mu ng .. . @ 1 9\m od_ 131 5980 979 500 _51. docx @ 13 454 8 @ @ 1
- This furnace was designed for commercial use. The furnace is not designed for heating food, animals, wood, grain, etc.
- The furnace must not be used to heat the workplace.
- Do not use the furnace to melt ice or for similar purposes.
- Do not use the furnace as a clothes dryer.
Caution
Applicable safety instructions are contained in individual chapters.
![Page 16: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/16.jpg)
16
Pos: 62 / TD/Sich erh eit/Übe rsch rift - Für hi era us r esultie ren de Sch äde n ha ftet der Betrei ber @ 1 0\m od_1 256 731 411 117 _51. docx @ 68 223 @ @ 1
The operator is liable for any resulting damages. Pos: 63 / TD/Allge mein e Hinwei se ( für alle Anleit ung en) /Hinweis - Dies er Of en v erf ügt ü ber kein e Siche rheits tech nik fü r Proz esse, ..-nic ht b ren nba re G ase @ 8\ mod _12 350 3869 977 1_5 1.do cx @ 5148 3 @ @ 1
Caution
Operating the furnace with explosive gases or mixtures, including explosive gases or
mixtures created as a result of heating/drying, is prohibited.
This furnace features no safety technology for processes which produce combustible
mixtures, for example debinding.
If the furnace is still used for such processes despite this fact, the concentration of organic gas mixtures in the furnace must never exceed 3% of the lower explosion limit (LEL). This
pre-requisite applies not only to normal operation but, in particular, to exceptional
situations such as process disruptions (caused, for example, by the failure of a power unit).
You must ensure that the furnace is adequately ventilated and vented.
Nabertherm offers a broad range of furnaces which were especially developed for processes involving the use of combustible gas mixtures.
Pos: 64 / TD/Allge mein e Hinwei se ( für alle Anleit ung en) /Hinweis - Dies es Pro dukt entsp richt nicht der ATEX-Richtli nie u nd d arf nicht ..-nic ht b ren nba re G ase @ 2\ mod _11 8422 875 689 3_5 1.do cx @ 1 968 6 @ @ 1
Note
This product does not comply with the ATEX Directive and may not be used in ignitable
atmospheres. It must not be operated with explosive gases or mixtures or during processes
where explosive gases or mixtures are produced.
Pos: 65 / TD/Sich erh eit/Übe rsch rift - Anfo rde run gen an den Be treib er der Anlag e @ 0 \mo d_1 167 8237 755 31_ 51.d ocx @ 515 7 @ 2 @ 1
3.2 Requirements for the Furnace Operator Pos: 66 / TD/Sich erh eit/Anfo rde run gen an den B etrei ber de r Anlag e - RH TH, RH TC, LH T, R T - Roh röfe n @ 2\m od_1 184 570 280 797 _51. docx @ 19 913 @ @ 1
The set-up instructions and safety regulations must be followed, otherwise the furnace will be deemed to have been used improperly, effectively cancelling any claims against
Nabertherm GmbH.
This level of safety when operating the furnace can be achieved only if all the necessary
measures have been taken. It depends on the furnace operator's diligence in planning these
measures and controlling how they are carried out.
The operator must ensure that
all harmful gases are removed from the workplace, for example by an extraction
system,
the extraction system is switched on,
the workplace is properly ventilated,
the furnace is operated only in a perfect operating condition and, in particular, that the
functions of the safety components are checked regularly.
the required personal protective equipment is available for and used by the operating,
maintenance, and repair personnel.
these operating instructions, including the supplier documentation, are kept near the
furnace. These instructions must be available at all times for anyone working with or on the furnace;
all the safety and operating instruction signs on the furnace can be read properly.
Damaged or unreadable signs must be replaced immediately,
furnace personnel are informed regularly about all issues involving occupational
safety and environmental protection and are familiar with all the operating
instructions, especially those involving safety,
a risk assessment is carried out (in Germany, covered of the Occupational Safety Act)
to determine any other hazards that may result from the working conditions particular
to the furnace's location,
![Page 17: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/17.jpg)
17
all other instructions and safety guidelines that have been determined in a risk
assessment for the workplace are compiled in an operation manual (in Germany,
covered of the Ordinance Regulating the Use of Operating Equipment).
Only sufficiently qualified and authorized personnel may operate, maintain and repair
the system. This personnel must be trained in how to operate the furnace and must confirm their participation in the training with a personal signature. The training
program must be documented in detail. In case an operator is replaced, additional
training must also take place. The additional training may only be performed by
authorized, trained individuals familiar with the system. The additional training must
be painstakingly documented and participation must be evidenced by the names and
signatures of the participating employees. Pos: 67 / TD/Sich erh eit/De r Einsatz von Fä rbel ösun gen k ann zu ei nem üb erp rop ortio nale n Versc hleiß de r Heizel eme nte füh ren. .. @ 43\m od_ 137 111 529 958 8_51 .docx @ 2 335 30 @ @ 1 it
The use of paint solvents can lead to accelerated wear of the heating elements. If solvents
are used, good pre-drying is necessary to minimize their impact on the heating elements.
Pos: 68 / TD/Allge mein e Hinwei se ( für alle Anleit ung en) /Hinweis - In Deutsc hlan d ist di e allg emei ne Un fallve rhüt ungs vorsc hrift zu b each ten. @ 3\ mod _11 936 678 017 39_ 51.d ocx @ 261 22 @ @ 1 a
Note
In Germany, the general accident protection guidelines must be observed. The national
accident prevention regulations of the country of operation apply.
Pos: 69 / TD/Sich erh eit/Übe rsch rift - Anfo rde run gen an das Bedi enp erso nal @ 0\ mod_ 116 782 564 342 3_5 1.docx @ 5 166 @ 2 @ 1
3.3 Requirements for the Operating Personnel Pos: 70 / TD/Sich erh eit/Anfo rde run gen an das Be dien pers onal @ 0\ mod _11 582 1866 348 2_5 1.do cx @ 2155 @ @ 1
Everyone involved in operation, installation, maintenance, or repair of the furnace must have read and understood the operating instructions. No liability will be accepted for
damage or disruptions to operation resulting from non-compliance with the operating
instructions.
Only adequately qualified and authorized persons may operate, maintain, or repair the
furnace.
Operating personnel are instructed regularly in all aspects of occupational safety and
environmental protection and are familiar with all the operating instructions, in particular,
safety instructions.
Only trained personnel may operate the control and safety equipment.
The operator should complete these details:
Operator
_______________________________________________________________
The furnace may only be transported by
_______________________________________________________________
The furnace may only be installed by
_______________________________________________________________
The furnace may only be commissioned by
_______________________________________________________________
Initial instructions may only be given by _______________________________________________________________
Malfunctions may only be rectified by
_______________________________________________________________
The furnace may only be maintained by
_______________________________________________________________
The furnace may only be cleaned by
_______________________________________________________________
The furnace may only be serviced by
_______________________________________________________________
The furnace may only be repaired by
_______________________________________________________________
The furnace may only be shut down by
![Page 18: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/18.jpg)
18
_______________________________________________________________
Pos: 71 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Wa rnsy mbol _Gef ahr - Absic her ung v on G efa hre n bei Ü ber tem per atu r a m TWB- TWW @ 37\ mod _136 249 040 489 3_5 1.doc x @ 2 140 67 @ @ 1
DANGER
• Danger caused by incorrectly entered cut-off
temperature at the over-temperature limiter with
manual reset/over-temperature limiter with automatic
reset.
• Mortal danger
• If, as a result of over-temperature from the charge and/or the
operating equipment, a charge is likely to be damaged at this pre-
set cut-off temperature of the over-temperature limiter with
manual reset/over-temperature limiter with automatic reset, or if
the charge itself becomes a source of danger for the furnace or its
surroundings, the cut-off temperature must be reduced at the
over-temperature limiter with manual reset/automatic reset to the maximum permissible value.
Pos: 72 / TD/Sich erh eit/Übe rsch rift - Schutzkl eidu ng @ 0\m od_ 116 782 579 5750 _51 .docx @ 5 175 @ 2 @ 1
3.4 Protective Clothing Pos: 73 / TD/Sich erh eit/Schut zkleidu ng - Schut zkleidu ng t rag en @ 5\ mod_ 122 027 499 503 0_5 1.docx @ 4 210 8 @ @ 1
Wear protective clothing
Pos: 74 / TD/Sich erh eit/Schut zkleidu ng - Schut zbrille tra gen @ 5\ mod _12 202 739 548 30_5 1.d ocx @ 420 86 @ @ 1
Wear protective goggles.
Pos: 75 / TD/Sich erh eit/Schut zkleidu ng - Hitze bestä ndig e Han dsch uhe tra gen @ 9\ mod _12 465 4201 330 6_5 1.do cx @ 6247 3 @ @ 1
Wear heat-resistant gloves to protect your hands.
Pos: 76 / TD/Sich erh eit/Übe rsch rift - Gr undle gen de Ma ßna hme n bei Nor malb etrie b @ 0\mo d_1 167 825 919 827 _51. docx @ 51 84 @ 2 @ 1
3.5 Basic Measures During Normal Operation Pos: 77 / TD/Sich erh eit/Gr undl ege nde Ma ßna hme n b ei Nor malb etri eb Top -HO u nd F Öfe n @ 13\ mod _12 8939 573 565 2_5 1.doc x @ 1 075 62 @ @ 1
Risks during normal operation
Before switching the furnace on, check and ensure that only authorized persons are in the
working area of the furnace and that no one can be injured as a result of operating the
furnace.
Each time, before starting production check and ensure that all the safety equipment
functions as intended (for example, that the contact safety switch switches the heating off when the lid is opened).
Before starting production each time, check the furnace for obvious damage and ensure that
it is operated only in a perfect condition. Report any defects to Nabertherm Service
immediately.
![Page 19: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/19.jpg)
19
Before starting production each time, remove all materials and objects that are not needed
for production from the working area.
At least once every day (see also Servicing and Maintenance) check the following:
Check the furnace for obvious external damage (visual check), for example insulation,
heating elements, power cable, exhaust gas system, if applicable.
Check that all safety equipment is functioning (for example, that the contact safety
switch switches the heating off when the door is opened).
Pos: 78 / TD/Sich erh eit/Übe rsch rift - Gr undle gen de Ma ßna hme n im N otfall @ 1\ mod _11 7094 336 926 7_5 1.do cx @ 9 093 @ 2 @ 1
3.6 Basic Measures in Case of Emergency Pos: 79 / TD/Sich erh eit/Übe rsch rift - Verh alte n im N otfall @ 1\ mod_ 117 094 990 485 5_5 1.docx @ 9 123 @ 3 @ 1
3.6.1 What to do in an Emergency Pos: 80 / TD/Allge mein e Hinwei se ( für alle Anleit ung en) /Hinweis - Das Stills etzen im No tfall ist vor gese hen durc h Zi ehe n de s Netzs tecke rs. @ 6\m od_ 122 206 283 6860 _51 .docx @ 4 282 9 @ @ 1
Note
The power plug is to be pulled out to stop the furnace in case of an emergency.
Therefore, the power plug must be accessible at all times when the furnace is operating so
that it can be pulled out quickly in case of an emergency.
Pos: 81 / TD/Sich erh eit/Gr undl ege nde Ma ßna hme n im Notfall - Net zsteck er zie hen (alle Anleitu nge n) - Gr afik @ 36\ mod _13 6007 972 285 4_5 1.do cx @ 2 089 20 @ @ 1
Fig. 10: Pull the power plug (similar to picture)
Pos: 82 / TD/Allge mein e Hinwei se ( für alle Anleit ung en) /War nung - Bei une rwa rtet en Vor gän gen im Of en (z.B. st arke Rauc hentwickl ung ode r G eruc hsbel ästigu ng) @ 4\ mo d_12 053 065 797 37_ 51.d ocx @ 342 18 @ @ 1
Risks during Normal Operation!
Switch the furnace off immediately in case of unexpected occurrences in the furnace (e.g. a
lot of smoke or unusual smells). Wait until the furnace has cooled naturally to room
temperature.
![Page 20: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/20.jpg)
20
Pos: 83 / TD/Allge mein e Hinwei se ( für alle Anleit ung en) /War nung - Bei une rwar tete n Vor gän gen dar f de r Of en nic ht g eöff net w erd en - Gr afik LH T ../ 17 D - 2 015 @ 11 9\m od_ 146 2281 876 740 _51. docx @ 46 940 8 @ @ 1
In case of fire, keep
door and the air feed
lever closed. This will
allow you to prevent
the spread of smoke
and keep out a supply
of oxygen.
Pos: 84 / TD/Allge mein e Hinwei se ( für alle Anleit ung en) /War nung - Ar beite n an de r elek trisch en Aus rüst ung dür fen nur von ... @ 2\m od_ 118 422 802 101 9_51 .docx @ 1 967 3 @ @ 1
Warning - Danger of Electric Shock!
Work on the electrical equipment may be done only by qualified, authorized electricians.
Pos: 85 / TD/Sich erh eit/Übe rsch rift - Gr undle gen de Ma ßna hme n bei Wart ung und Inst andh altu ng @ 0\m od_ 116 782 6060 620 _51 .docx @ 51 93 @ 2 @ 1
3.7 Basic Measures for Servicing and Maintenance Pos: 86 / TD/Sich erh eit/Gr undl ege nde Ma ßna hme n b ei Wart ung un d Inst and haltu ng @ 0\m od_ 115 822 245 843 6_5 1.docx @ 2 188 @ @ 1
Maintenance work must be performed by authorized persons following the maintenance instructions and the accident prevention regulations. We recommend that the maintenance
and repair work be carried out by the service team of Nabertherm GmbH. Non-compliance
may cause injuries, death, or considerable damage to property.
Switch off the furnace and make sure it cannot be switched on again inadvertently (lock the
main switch and secure it with a padlock), or pull out the power plug.
Clear an adequate area around the furnace to facilitate the repair work.
Suspended loads are dangerous. Working beneath a suspended load is prohibited. There is a
risk of fatal injury.
Relieve the pressure on hydraulic or pneumatic equipment before carrying out maintenance or repair work (if applicable).
When cleaning furnaces, control cabinets, or electrical equipment housings, never spray
them with water.
When maintenance or repair work has been completed, before recommencing production
ensure the following:
Check that loosened screw connections have been re-tightened,
Reinstall protective equipment, screens, and filters,
Remove all material, tools, and other equipment used for the maintenance or repair
work from the working area of the furnace,
Remove any liquids that have leaked,
Check that all safety functions (e.g. emergency stop button) work properly,
Power cables may be replaced only with similar, approved cables. Pos: 87 / TD/Sich erh eit/Übe rsch rift - Umwel tschu tzvors chrif ten @ 0\ mod _11 6782 618 923 7_5 1.do cx @ 5 202 @ 2 @ 1
3.8 Environmental Regulations Pos: 88 / TD/Sich erh eit/Umw eltsch utzvo rschri ften Sc hmie rfet te-Hy dra uliköl-Kü hlmitt el @ 0 \mo d_1 158 2234 243 04_ 51.d ocx @ 219 9 @ @ 1
All statutory duties regarding waste avoidance, proper recycling, and disposal must be observed when work is carried out on and with the furnace.
![Page 21: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/21.jpg)
21
Problem materials that are no longer needed, such as lubricants or batteries, must not be
placed in normal waste disposal systems or allowed to enter the sewage system.
During installation, repair, and maintenance work, substances that are hazardous to water,
such as
lubricating grease and oils
hydraulic oils
refrigerants
solvent-based cleaning fluids must not be allowed to contaminate the soil or enter the
sewage system.
These substances must be stored, transported, collected, and disposed of in suitable containers.
Note
The operator must ensure that national environmental regulations are observed.
Pos: 89 / TD/Sich erh eit/Umw eltsch utzvo rschri ften Ele ktro nische Baut eile-Is olieru ng -Altmet all @ 4 \mo d_1 205 1433 148 53_ 51.d ocx @ 325 63 @ @ 1
When it is delivered, this furnace contains no substances that make a hazardous waste
classification necessary. However, residues of process materials may accumulate in the
furnace insulation during operation. These may be hazardous to health and/or the
environment.
Dismantle the electronic components and dispose of them as electric scrap.
Remove the insulation and dispose of it as hazardous waste (See Servicing, Cleaning,
and Maintenance with Ceramic Fiber Material)
Dispose of the housing as scrap metal.
Pos: 90 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Erl äute run g ANSI Z535. 6 @ 8\mo d_1 243 323 558 881 _51. docx @ 57 444 @ 2 @ 1
3.9 Explanation of the Symbols and Warnings
Note
In the following operating instructions, specific warnings are given to draw attention to
residual risks that cannot be avoided when the furnace is operating. These residual risks
include dangers for humans/products/ the furnace, and the environment.
The symbols used in the operating instructions are especially intended to draw attention to
safety information.
The symbols used cannot replace the text of the safety information. Therefore, always read the entire text.
Graphic symbols correspond to ISO 3864. In accordance with the American National
Standard Institute (ANSI) Z535.6 the following warning information and words are used in
this document:
The general hazard symbol, in combination with the words CAUTION, WARNING and DANGER warns about the risk of serious injury. Observe the following information to
prevent injury or death.
NOTICE Refers to a hazard that could damage or destroy the equipment.
CAUTION Refers to a hazard with a minor or medium risk of injury.
![Page 22: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/22.jpg)
22
WARNING Refers to a hazard that could cause death, serious or irreversible injury.
DANGER Refers to a hazard that could directly cause death, serious or irreversible injury.
Structure of the Warning: All Warnings are Structured as Follows
WARNING
• Type and source of the danger
• Consequences of non-compliance
• Action to prevent danger
or
DANGER
• Type and source of the danger
• Consequences of non-compliance
• Action to prevent danger
Graphical symbols (optional)
according to ISO 3864: Consequences, measures, and
prohibitions
Reference texts: • Type and source of the danger
• Possible consequences of non-
compliance
• Measures/prohibitions
Hazard symbol
Indicates the risk of injury
Signal word
Classifies the danger
Graphical symbols
(optional) according to ISO
3864: Instructions or prohibitions
Graphical symbols (optional)
according to ISO 3864:
Consequences, measures, and
prohibitions
Reference texts:
• Type and source of the danger
• Possible consequences of non-
compliance
• Measures/Prohibitions
Hazard symbol
Indicates the risk of injury
Signal word
Classifies the danger
![Page 23: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/23.jpg)
23
Pos: 91 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Üb ersc hrift - Hinw eissymb ole in de r Anleit ung @ 9\ mod _12 4705 342 962 6_5 1.do cx @ 6 275 1 @ @ 1
Information Symbols in the Instructions: Pos: 92 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Hinw eis - U nte r dies em Sy mbol erh alten Sie Anweis ungs hinweise un d ... @ 9\ mod _12 470 5393 231 1_5 1.do cx @ 6278 5 @ @ 1
Note
Below this symbol you will find instructions and particularly useful information.
Pos: 93 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Pikto gra mm i n d er Anleit ung - G ebot - G ebo tszeich en - Wichtig e Ge bote sind zu b efolg en @ 9\m od_ 124 705 617 598 2_51 .docx @ 6 285 3 @ @ 1
Rule - Rule Sign
This symbol draws attention to important rules that must be followed. Rule signs protect
people against injury and show what is to be done in certain situations.
Pos: 94 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Pikto gra mm i n d er Anleit ung - G ebot - Wich tige I nfo rma tion f ür den B edien er @ 9\ mod _12 4705 320 072 9_5 1.do cx @ 6 271 7 @ @ 1
Rule - Important Information for Operators
This symbol draws the operator's attention to important information and operating
instructions that must be followed.
Pos: 95 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Pikto gra mm i n d er Anleit ung - G ebot - Wich tige I nfo rma tion f ür das Wa rtu ngsp erso nal @ 9\ mod_ 124 705 320 604 2_5 1.docx @ 6 273 4 @ @ 1
Rule - Important Information for Maintenance Personnel
This symbol draws the maintenance personnel's attention to important operating and
maintenance instructions (service) that must be followed.
Pos: 96 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Pikto gra mm i n d er Anleit ung - G ebot - N etzsteck er zi ehe n @ 9 \mo d_1 247 055 471 526_ 51. docx @ 62 836 @ @ 1
Rule - Pull Out the Power Plug
This symbol tells the operator to pull out the power plug.
Pos: 97 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Pikto gra mm i n d er Anleit ung - G ebot - An heb en mit m ehr ere n Pers one n @ 9 \mo d_1 247 063 0580 02_ 51. docx @ 62 938 @ @ 1
Rule - Lift only with Several People
This symbol draws the personnel's attention to the fact that this device may only be lifted
and moved to its final destination by several people.
Pos: 98 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Pikto gra mm i n d er Anleit ung - Wa rnu ng - Hei ße O berfl äche - O ber fläche nicht be rüh ren @ 9\ mod _12 470 547 747 80_5 1.d ocx @ 628 02 @ @ 1
Warning - Hot Surface, Do Not Touch
This symbol warns the operator that the surface is hot and should not be touched.
Pos: 99 / TD/Sich erh eit/Siche rheits symb ole/Wa rnhi nweise -ISO-ANSI/Pikto gra mm i n d er Anleit ung - Wa rnu ng - elek trisch er Schl ag - zur Ver meidu ng Anw eisun g folg en @ 9\ mod _124 705 509 389 2_5 1.doc x @ 6 281 9 @ @ 1
Warning - Danger of Electric Shock
This symbol warns the operator that there is a risk of an electric shock if the following
warnings are not heeded.
Pos: 100 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m in der Anl eitun g - W arn ung - U mkipp en d es Ge rät es @ 9\mo d_1 247 059 198 372 _51. docx @ 62 870 @ @ 1
Warning – Risk of Device Toppling Over
This symbol tells the operator that there is a risk of the device toppling over if the following
warnings are not heeded.
Pos: 101 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m in der Anl eitun g - W arn ung - Schw ebe nde Last en @ 9\m od_ 125 145 050 030 9_51 .docx @ 6 545 4 @ @ 1
Warning – Suspended Load
This symbol warns the operator of potential dangers of suspended loads. Working below a
suspended load is strictly forbidden. Ignoring this can lead to fatal injury.
![Page 24: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/24.jpg)
24
Pos: 102 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m in der Anl eitun g - W arn ung - He ben schwe rer Last en @ 9\ mod _124 705 989 135 8_5 1.doc x @ 6 288 7 @ @ 1
Warning – Danger if Heavy Loads Are Lifted
This symbol warns the operator of the potential dangers of lifting heavy loads. Ignoring this can lead to injury.
Pos: 103 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m in der Anl eitun g - W arn ung - U mweltg efäh rdu ng @ 9\ mod_ 124 706 044 341 9_5 1.doc x @ 6 290 4 @ @ 1
Warning – Risk to the Environment
This symbol warns the operator of the risk to the environment if the following information is not heeded. The operator must ensure that national environmental regulations are
observed.
Pos: 104 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m in der Anl eitun g - W arn ung - Br and gefa hr @ 9\m od_ 125 144 527 282 2_5 1.docx @ 6 543 7 @ @ 1
Warning - Fire Danger
This symbol warns operators of the danger of fire if the following information is not
followed.
Pos: 105 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m in der Anl eitun g - W arn ung - Explo sions gefä hrlich e Stoff e @ 9 \mo d_1 247 061 148 859 _51. docx @ 62 921 @ @ 1
Warning – Risk of Explosive Substances or Explosive Atmosphere
These symbols warn the operator of explosive substances or an explosive atmosphere
Pos: 106 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m in der Anl eitun g - Gefa hr - nic ht mit Wasse r üb ersc hütt en @ 9\m od_ 124 782 641 0886 _51 .docx @ 6 381 9 @ @ 1
Prohibited - Important Information for Operators
This symbol warns the operator that water or cleaning products must NOT be poured over
the objects. A high-pressure cleaning device must also not be used.
Pos: 107 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Über schrif t - W arn hinweissy mbol e an de r Anlag e @ 9\m od_1 247 053 700 273 _51. docx @ 62 768 @ @ 1
Warning Signs on the Furnace: Pos: 108 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m an de r Anlag e - W arn ung - G efah r vo r h eiße r Ob erfl äche und Verb ren nun g @ 9 \mo d_1 247 052 957 145_ 51. docx @ 62 700 @ @ 1
Warning - Hot Surface, Danger of Burning – Do Not Touch
You may not always realize that surfaces, such as furnace components, furnace walls, doors
and materials, and even liquids are hot. Do not touch the surface.
Pos: 109 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m an de r Anlag e - W arn ung - G efah ren du rch el ektrisc hen Stro m @ 9\mo d_1 247 052 639 824 _51. docx @ 62 683 @ @ 1
Warning - Danger of Electric Shock!
Warning, dangerous electric voltage
Pos: 110 /TD/Sic her heit/Üb ersc hrift - Allge mein e Gef ahr en an d er Anl age @ 0\ mod _11 685 967 9628 8_5 1.d ocx @ 601 4 @ 2 @ 1
3.10 General Risks with the Furnace Pos: 111 /TD/Sic her heit/Allge mein e Ge fah ren (Ver bre nnu ng, Q uetsc hen , Stro m) - LH T-L abo röf en @ 46\ mod _13 777 739 9806 5_5 1.d ocx @ 243 640 @ @ 1
Warning! General Hazards!
Risk of burning on the furnace housing
The door handle/grip can become very hot during operation; wear gloves.
Risk of crushing on moving parts (door hinge)
The switchgear cabinet (if present) and the terminal boxes on the system contain dangerous
electrical voltages.
Do not insert any objects into the openings on the furnace housing, exhaust air holes, or
cooling slots on the switchgear or furnace (if present). This poses a risk of electric shock.
Fire hazard if an extension cable is used:
![Page 25: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/25.jpg)
25
With 230 V furnace models make sure that:
The distance between the circuit breaker and the power socket that the furnace is connected
to is as short as possible.
NO power board or extension cable is used between the power socket and the furnace.
Pos: 112 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Wa rnun g - Es dü rfen kein e Ge gens tänd e a uf d en/d er O fen/Sc halta nlage ab geleg t/ab gest ellt wer den ... @ 4\m od_ 120 392 4555 465 _51 .docx @ 3 1568 @ @ 1
Warning! General Hazards!
No objects may be placed or set down on the furnace or switchgear. Doing so creates a fire
or explosion hazard.
Pos: 113 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m in der Anl eitun g - W arn ung - vo r el ektr oma gnetis che Feld er (Im plant at-He rzsch ritt mach er) @ 1 16\m od_ 145 949 466 855 2_5 1.docx @ 4 606 73 @ @ 1
Warning – Influence on implants from electromagnetic fields
It can be assumed that there is no risk from electromagnetic fields. Deviating from this,
there may be a risk for people with active implants (such as a pacemaker, insulin pump) and
passive implants, as the function of the implant may be affected also with a field strength
below the permissible threshold. Proper function must be ensured on the basis of the
technical specifications of the implant through an individual assessment by a suitable expert
(for example, an occupational physician).
Pos: 114 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Warns ymb ol_Ge fah r - Gefa hr durc h fe hlend e o der nicht k or rekt ang eschloss ene Erdu ng a n ... @ 8 5\m od_1 423 745 935 634 _51. docx @ 35 946 1 @ @ 1
DANGER
• Danger from electrocution
• If there is no earth connection, or the
earth connection is poorly connected, the
result may be a deadly electrical shock.
• Do not insert any metallic objects such as thermocouples, sensors or tools into the
furnace chamber without having previously
ensured that the plant has been correctly
earthed. Entrust the job of making a earth
connection between the object and the
furnace housing to a qualified electrical
technician. Any objects inserted into the
furnace must be inserted only through those
openings intended for this purpose.
![Page 26: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/26.jpg)
26
Pos: 115 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Warns ymb ol_Ge fah r - Ab siche rung von Gef ahr en b ei Übe rte mpe rat ur a m TWB-TWW @ 37\ mo d_13 624 904 048 93_ 51.d ocx @ 214 067 @ @ 1
DANGER
• Danger caused by incorrectly entered cut-off
temperature at the over-temperature limiter with
manual reset/over-temperature limiter with automatic
reset.
• Mortal danger
• If, as a result of over-temperature from the charge and/or the
operating equipment, a charge is likely to be damaged at this pre-
set cut-off temperature of the over-temperature limiter with
manual reset/over-temperature limiter with automatic reset, or if
the charge itself becomes a source of danger for the furnace or its
surroundings, the cut-off temperature must be reduced at the
over-temperature limiter with manual reset/automatic reset to the
maximum permissible value.
Pos: 116 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Tr ansp ort, Mo ntag e u nd Ersti nbet rieb nah me @ 0\ mod _11 5884 422 741 6_5 1.do cx @ 3 112 @ 1 @ 1
4 Transportation, Installation, and Commissioning Pos: 117 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Anli efe rung @ 0\ mo d_1 1678 265 348 89_ 51.d ocx @ 522 0 @ 2 @ 1
4.1 Delivery Pos: 118 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Anliefe run g - (Hinweis e allg emei n) @ 0\m od_ 115 823 350 888 7_51 .docx @ 2 223 @ @ 1
Check that Everything is Complete
Compare the delivered items with the delivery note and the purchase order documents.
Immediately notify the carrier and Nabertherm GmbH of any missing or damaged parts, as
complaints at a later date cannot be acknowledged.
Danger of Injury
When the furnace is being lifted, parts of the furnace or the furnace itself could topple over,
slip, or fall. Before the furnace is lifted, make sure no one is in the working area. Wear
safety footwear and a hard hat.
Safety Instructions
Forklifts must be operated only by authorized personnel. The operator bears sole
responsibility for safe operation and the load.
When the furnace is being lifted, make sure that the ends of the forks or the load do
not catch on neighboring goods. Use a crane to move tall parts, such as control
cabinets.
Use only lifting equipment with sufficient load-bearing capacity.
Lifting gear must be attached only to positions that have been designated for this
purpose.
Attachments, piping, or cable conduits must never be used to affix lifting gear.
Unpackaged parts should only be lifted with ropes or straps.
Attach transportation equipment only to positions intended for this purpose.
Lifting and securing equipment must conform to the provisions contained in accident
prevention regulations.
Consider the weight of the furnace when choosing lifting and securing equipment. (see Specifications)
Stainless steel parts (including mounting elements) must always be kept separate from
unalloyed steel parts.
Do not remove corrosion protection until immediately prior to assembly.
![Page 27: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/27.jpg)
27
Risks during Normal Operation!
Suspended loads are dangerous. Working beneath a suspended load is prohibited. There is a
risk of fatal injury.
Note
Safety and accident prevention guidelines applicable for forklift trucks must be followed.
Pos: 119 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Anliefe run g - Öfen mit eine m Hubw age n tr ansp ortie ren @ 2 \mo d_1 184 6724 602 00_ 51.d ocx @ 200 00 @ @ 1
Transportation with a Pallet Truck
Observe the maximum permitted capacity of the pallet truck.
1. Our furnaces are delivered ex works on wooden frames to facilitate unloading. Transport the furnace in its original packaging and with suitable equipment to
prevent any damage. Remove the packaging only when the furnace is in its final
location. When transporting the furnace, make sure it is secured against sliding,
toppling over, and damage. The furnace should be transported and installed by at
least two persons. Do not store the furnace in damp rooms or outdoors.
2. Push the pallet truck underneath the transportation frame. Make sure that the pallet
truck is completely beneath the frame. Pay attention to neighboring goods.
Fig. 11: Pallet truck is pushed completely beneath the transportation frame
3. Lift the furnace carefully and pay attention to its center of gravity. When the furnace
is being lifted, make sure that the ends of the forks or the load do not catch on
neighboring goods.
4. Make sure that the furnace is balanced safely; if not, attach securing equipment.
Push the furnace carefully, slowly and with the pallet truck at its lowest position. Do not transport the furnace on inclines.
5. Carefully lower the furnace at its final position. Pay attention to neighboring goods.
Try not to set it down too abruptly.
Pos: 120 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Leg end e fü r Packst ücke z .B. ze rbr echlich @ 1\ mod _11 737 7671 330 7_5 1.do cx @ 1074 6 @ @ 1
Symbols:
The international standard symbols for handling packaging are defined in ISO R/780
(International Organization for Standardization) and in DIN 55 402 (German
Institute for Standardization).
Description Symbol Explanation
Fragile
This symbol is to be attached to fragile goods. Goods marked like this are to be handled carefully and must
not be thrown or tied up.
![Page 28: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/28.jpg)
28
Description Symbol Explanation
This side up
The freight must be transported, transshipped, and
stored in such a way that the arrows point upward. The
freight must not be rolled, folded, or stored on edge.
However, the package does not have to be packed on
top of other freight.
Keep dry
Products with this symbol must be protected against
high air moisture, hence, they must be stored under cover. If particularly heavy or bulky packages cannot
be stored in halls or sheds, they must be covered
carefully with a tarpaulin or similar.
Sling here
The symbol shows only where the sling should be
attached, not the method of slinging. If the symbols
are at an equal distance from the middle or center of
gravity of the package, the package hangs straight if the slings are the same length. If this is not the case,
the sling on one side has to be shortened.
Pos: 121 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Au spack en @ 2\m od_ 118 473 951 798 5_5 1.docx @ 2 001 3 @ 2 @ 1
4.2 Unpacking Pos: 122 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Allge mein er Hi nweis zu r Ve rpack ung (z.B. L abo röf en) @ 3\ mod _11 924 371 858 46_5 1.d ocx @ 235 09 @ @ 1
Note
The furnace packaging prevents damage during transportation. Make sure that you remove
all packaging material (also inside the furnace chamber). Keep the packaging and
transportation securing equipment in case it is needed for future transportation or storage.
At least two people are needed to carry/transport the furnace, more for larger furnaces.
Pos: 123 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Auspac ken kl eine Öfen mit St ülpka rton - Teil 1 @ 2\ mod _11 847 396 164 97_ 51.d ocx @ 200 26 @ @ 1
Wear protective
gloves
1. Check the transportation packaging for possible damage.
2. Remove tensioning straps from the transportation packaging.
3. Slacken screws and remove wooden casing from the covering box (if available 3a).
![Page 29: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/29.jpg)
29
Pos: 124 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Auspac ken Lab orof en - LH T- Tischm odell (M odell 2015 )- Teil 2 @ 1 19\ mod _14 622 860 3765 3_5 1.d ocx @ 469 457 @ @ 1
4. Carefully raise the covering box and remove from the pallet.
5. A flat box containing the accessories for your furnace (ceramic insert plate or
optionally a cable) is located on the rear wall of the furnace. Compare the scope of
delivery with the delivery note and the order documents, see chapter "Delivery".
Loosen screws or nails from the retaining strips and remove.
6. Remove retaining strips from the pallet.
Pos: 125 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Auspac ken -Tr age n L abo rofe n - LHT -Tisc hmo dell - Teil 3 @ 3\m od_1 192 436 920 381 _51. docx @ 23 483 @ @ 1
7. To carry, grip furnace from below at the sides and make sure you have a firm grip.
8. For furnaces weighing more than 25 kg, transport work must be carried out by at
least 2 people. If transport straps are used, they must be attached crosswise only.
Ensure that they are secure.
Note
In Germany, the general accident protection guidelines of VBG or BGZ must be observed.
The national accident prevention regulations of the country of operation apply.
![Page 30: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/30.jpg)
30
Note
Save the packaging for possible shipping or for storing the furnace.
Pos: 126 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Tr ansp ortsic her ung/V erp ackun g @ 0\mo d_1 167 826 775 847 _51. docx @ 52 29 @ 2 @ 1
4.3 Transportation Securing Equipment/Packaging Pos: 127 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Verp ackun g - Allg em eine Hinw eise z ur Ve rpac kung - s pezielle Tr ans portsic her ung für LHT -Öf en @ 2\m od_ 118 473 996 2051 _51 .docx @ 2 003 8 @ @ 1
To protect against transportation damage, the furnace is equipped with a transportation lock, which must be removed before commissioning the furnace.
First remove the adhesive strips and then pull out the transportation lock. Pulling the
furnace door slightly towards you makes it easier to remove the transportation lock from the furnace (see figure: Removing the transportation lock).
Fig. 12: Removing the transportation lock
Note
Save the transportation lock for possible shipping or for storing the furnace. To avoid
damage to the furnace door insulation in the event of reshipping, the furnace must be
locked again as shown in the above figure.
Pos: 128 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Ba uliche - u nd Ansc hlussvo rau ssetzu nge n @ 0\m od_1 167 826 890 943 _51. docx @ 52 38 @ 2 @ 1
4.4 Constructional and Connection Requirements Pos: 129 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Au fstellu ng (Stand ort des Ofens ) @ 2\m od_ 118 4848 718 972 _51. docx @ 20 166 @ 3 @ 1
4.4.1 Installation (Furnace Location) Pos: 130 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Beim Auf stellen des Ofe ns sind folg end e Siche rheits hinweis e zu beac hten : @ 4 9\m od_ 138 372 5883 608 _51 .docx @ 2 5263 5 @ @ 1
When setting up the furnace, these safety instructions must be followed: Pos: 131 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Aufstell ung - Kam me röfe n/La bo röfe n - B eispiel: Tisch mod elle @ 49\ mod _13 837 335 3212 6_5 1.d ocx @ 252 785 @ @ 1
The furnace must be installed in a dry room in accordance with the safety instructions.
The table/supporting surface must be flat to enable the furnace to be installed straight.
Place the furnace on a non flammable surface (fire safety class A DIN 4102 –
Example: concrete, tiles, glass, aluminum or steel) so that any hot material falling
from the furnace cannot ignite the surface.
The carrying capacity of the table must be designed to bear the weight of the furnace incl. accessories.
The floor covering must be made of nonflammable material so that hot material
falling out of the furnace will not cause the floor covering to ignite.
Remove adhesive strip
Remove adhesive strip
Remove transportation lock downwards
Remove transportation lock upwards
Furnace lift-door
Furnace housing
![Page 31: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/31.jpg)
31
Pos: 132 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Aufstell ort (für Kam mer öfen A rts & C rafts ode r Ro hröf en) @ 4 9\m od_1 384 335 338 473 _51. docx @ 25 289 5 @ @ 1
Installation Location
The operator is responsible for adequate ventilation by installing the appropriate systems to supply fresh air and to vent exhaust air. If a batch emits gases or vapors,
adequate ventilation of the installation site must be provided as well as a suitable
exhaust air venting system. A suitable vent for combustion exhaust must be supplied
by the customer
Make sure that the heat radiated by the furnace is vented (consult a ventilation expert, if necessary)
Although the furnace is well-insulated, the exterior surfaces of the furnace radiate
heat. If necessary, this heat must be dissipated (a ventilation technician may need to
be consulted). In addition, a minimum clearance (S) of 0.5 m on all sides and 1 m
above the furnace must be maintained to flammable materials. In individual cases,
more space must be chosen in order to match the local conditions. For nonflammable
materials, the minimum side clearance can be reduced to 0.2 m.
The furnace must be protected against weather and caustic atmospheres. Nabertherm
accepts no liability or warranty claim for any corrosion damage as a result of
installation in damp surroundings or for similar reasons. Pos: 133 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Min destsic her heitsa bsta nd zu br enn bar en Mate rialien -LH T 0 3/17 D ( Mod ell 20 15) - G rafik @ 11 9\mo d_1 462 347 835 812 _51. docx @ 46 949 6 @ @ 1
Fig. 13: Minimum safety distance to flammable materials (Table-top model) (similar to picture)
Pos: 134 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Warns ymb ol_Ge fah r - Br and gef ahr - Pikto gra mm B rand gef ahr @ 9\ mod _12 471 480 357 38_5 1.d ocx @ 630 10 @ @ 1
DANGER
• Fire- danger to health.
• Risk of fatal injury.
• Adequate ventilation must be ensured at the
installation location to conduct waste heat and any
exhaust gases away.
Pos: 135 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Vo r In betri ebn ahm e d es Of ens sollt e di eser 24 St unde n a m Aufst ellung sort akklim atisie rt .. @ 3\ mod _11 9556 801 433 6_5 1.do cx @ 2 787 1 @ @ 1
Note
Before starting the furnace for the first time, allow it to acclimatize at its installation
location for 24 hours.
![Page 32: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/32.jpg)
32
Pos: 136 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Warns ymb ol_Ge fah r - Gefa hre n du rch aut omatis che Lösch einric htu nge n (Wass er_ Lösc hgas ) @ 55\ mod_ 139 158 704 804 8_5 1.docx @ 2 697 40 @ @ 1
DANGER
• Danger associated with the use of an atomatic extinguishing
system
• Mortal danger posed by electrocution through wetness,
suffocation caused by extinguishing gas, etc.
• If automatic extinguishing systems are in place to fight fires and
protect the building, e.g. sprinkler systems, care must be taken
during their planning and installation that no additional hazards are
created, for example by extinguishing a pilot light, mixing hardening oil and extinguishing water, the disablement of
electrical equipment, etc.
Pos: 137 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Mon tag e, Ins tallati on u nd Ansc hluss @ 0\ mod _116 782 797 626 9_5 1.doc x @ 5 292 @ 2 @ 1
4.5 Assembly, Installation, and Connection Pos: 138 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - 1.1. 1 Abluft füh run g @ 0 \mo d_1 167 828 513 042_ 51. docx @ 53 10 @ 3 @ 1
4.5.1 Venting Exhaust Fumes Pos: 139 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Abgas füh rung allge mein er Text (NW 8 0 bis NW 120 ...) @ 66 \mo d_1 408 091 4898 59_ 51. docx @ 31 3526 @ @ 1
We recommend connecting an exhaust air pipe to the furnace to remove the exhaust gases.
For this purpose you can use a commercially available, metal exhaust gas pipe with NW80
to NW120. It must be installed continuously rising and fastened to the wall or ceiling.
Center the pipe over the furnace vent.
The exhaust gas pipe must not be installed with a tight fit to the furnace vent pipe since this
would prevent any bypass effect. This is necessary so that not too much fresh air is sucked
in by the furnace.
We recommend feeding the exhaust air through a flue.
Pos: 140 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Abluftf ühr ung LHT 03/ 17 D (Mo dell 2 015 ) - Gra fik @ 1 19\ mod _14 623 4794 674 1_5 1.do cx @ 4 695 23 @ @ 1
Fig. 14: Example: Assembling exhaust air piping (similar to picture)
![Page 33: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/33.jpg)
33
Pos: 141 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Es ist zu g ewäh rleiste n, d ass d urch die ausst röm end e h eiße Luft üb er di e Abluft ver roh run g @ 5 1\m od_ 138 685 684 5336 _51 .docx @ 2 582 60 @ @ 1
Caution
Steps must be taken to ensure that the flow of hot air emitted from the furnace's exhaust
outlet flap is not hazardous to people, property or the building.
Pos: 142 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Die Abgas e kö nne n nu r a bgeleit et we rde n, we nn der R aum üb er ei ne e ntsp rech end e Zuluft ... @ 4\m od_ 120 273 877 769 0_51 .docx @ 3 097 3 @ @ 1
Note
The exhaust gases can only be discharged if the room is aerated via an appropriate supply
air opening.
Pos: 143 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Ku nde nseitig e Dach - u nd o der Ma uer arb eiten sind dur ch di e Abga sfüh run g no twen dig ... @ 3\ mo d_1 1933 234 297 27_ 51.d ocx @ 259 93 @ @ 1
Note
Roof work and/or masonry by the customer is required for the exhaust gas discharge. The
size and design of the exhaust gas discharge must be determined by a ventilation technician.
The national regulations of the local country apply.
Pos: 144 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - 1.1. 1 Anschl uss a n das Elektr onetz @ 0\ mo d_1 1685 997 276 88_ 51.d ocx @ 605 0 @ 3 @ 1
4.5.2 Connecting the Furnace to the Power Supply Pos: 145 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Anschlus s an das Elek tro netz - fü r alle Stand ard öfe n @ 1 13\ mod _14 569 2407 839 0_5 1.d ocx @ 4514 32 @ @ 1
On the building side, the required services must be provided, i.e. the carrying capacity of the installation surface, provision of power (electricity), etc.
The furnace must be installed in accordance with its intended use. The power connection values must correspond to the values on the furnace type plate.
The power socket must be located close to the furnace and must be easily accessible.
The safety requirements are not met if the furnace is not connected to a socket with
grounding contact.
On use of an extension cable or a multipoint socket, the maximum electrical rating
must not be exceeded. Do not use the furnace with an extension cable if you are
uncertain whether grounding is guaranteed.
The power cable must not be damaged. Do not place any objects on the power cable.
Route the cable so that nobody can tread on or stumble over it.
A damaged power cable must be replaced immediately.
Ensure that the furnace's connection cable is routed so that it is protected. Pos: 146 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Hinweis - Vor de m Anschl uss d er Spa nnu ngsv erso rgu ng sic herst ellen, dass de r Netzsc halte r sich ... @ 51\ mo d_13 873 707 828 38_ 51.d ocx @ 258 640 @ @ 1
Note
Before connecting the power, make sure that the power switch is set to "Off" or "0".
Pos: 147 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Anschlus s an das Elek tro netz - Of en bis 36 00 Wat t (b eiliege ndes Netzk abel i m ... ) - Gra fik @ 5 1\m od_ 138 737 8147 753 _51 .docx @ 2 5884 0 @ @ 1
Fig. 15: Furnace up to 3600 kW (the power cable is supplied) (similar to picture)
![Page 34: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/34.jpg)
34
Pos: 148 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Anschlus s an das Elek tro netz - Das i m Li efer umf ang befi ndliche Netzz uleitun gska bel mi t „Sn ap-I n-Kup p @ 51\ mod _13 8737 080 308 7_5 1.do cx @ 2 586 90 @ @ 1
1. Plug the supplied connection cable with snap-in coupling into the rear wall of the furnace.
2. Then connect the power cable to the power supply. Use only a grounded socket.
Pos: 149 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Es gelte n die natio nale n Vorsc hrift en des jew eiligen Verwe ndu ngsla ndes @ 43 \mo d_1 371 459 790 861_ 51. docx @ 23 374 5 @ @ 1
Note
The national regulations of the respective country of operation apply.
Pos: 150 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Wa rnun g - Arbeit en a n d er el ektrisc hen Ausrü stun g dü rfe n nu r vo n ... @ 2\ mod _11 842 280 210 19_5 1.d ocx @ 196 73 @ @ 1
Warning - Danger of Electric Shock!
Work on the electrical equipment may be done only by qualified, authorized electricians.
Pos: 151 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Warns ymb ol_Acht ung - falsc he N etzsp annu ng - Piktog ra mm G ebo tszeich en @ 9\m od_ 124 747 523 138 1_51 .docx @ 6 313 1 @ @ 1
NOTICE
• Danger from incorrect voltage
• Damage to the furnace.
• Check voltage before connecting and
commissioning the furnace.
• Compare the voltage with the details
on the type plate.
Pos: 152 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Warns ymb ol_Ge fah r - Br and gef ahr - Pikto gra mm B rand gef ahr @ 9\ mod _12 471 480 357 38_5 1.d ocx @ 630 10 @ @ 1
DANGER
• Fire- danger to health.
• Risk of fatal injury.
• Adequate ventilation must be ensured at the
installation location to conduct waste heat and any
exhaust gases away.
Pos: 153 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - E rstinb etrie bna hm e @ 0 \mo d_1 167 8289 777 20_ 51.d ocx @ 533 7 @ 2 @ 1
4.6 Commissioning Pos: 154 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Erstin betri ebn ahm e LH T/N/NR/ TR/NAC/RHTH -RHTV - Öf en - 1 @ 2\m od_ 118 664 150 9092 _51 .docx @ 2 0833 @ @ 1
The furnace may be put into operation only by qualified persons and in compliance with the
safety instructions.
Read the section on "Safety". When the furnace is put into operation, the following safety
information must also be observed to prevent serious injury, damage to the furnace, and
damage to other property.
Make sure that the instructions and information in the controller instructions are observed
and followed.
The furnace may be used only for its intended purpose.
Ensure that only authorized persons remain in the working area of the furnace and that no
other persons are put at risk when the furnace is put into operation.
Before starting the furnace for the first time, make sure that all tools, foreign parts, and
transportation securing equipment have been removed.
Activate all safety equipment (power switch, emergency stop button if applicable) before
putting the furnace into operation.
Incorrectly wired connections may destroy electric/electronic components.
![Page 35: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/35.jpg)
35
Observe the special protective measures (e.g. grounding, …) for components that are at
risk.
Faulty connections can cause the furnace to start unexpectedly.
Before you switch on the furnace, make sure that you know what to do in case of faults or
emergencies.
Before starting the furnace for the first time, check the electrical connections and control
displays.
Before placing materials in the furnace, check whether they could harm or destroy the
insulation or the heating elements. Materials that could damage the insulation include:
alkalis, alkaline earths, metal vapors, metal oxides, chlorine compounds, phosphorous
compounds, and halogens.
Pos: 155 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - H eizele men te a us M olybd än-Disilizid (M oSi2) @ 2\ mod _11 866 4392 080 8_5 1.do cx @ 2084 6 @ 3 @ 1
4.6.1 Molybdenum Disilicide (MoSi2) Heating Elements Pos: 156 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Empf ehlun g fü r d as e rste Auf heize n d es Of ens in ca. 5 Stun den 100 °C unte rhal b... -MoSi2 H eizele men te @ 21\ mod _13 311 133 996 70_5 1.d ocx @ 151 240 @ @ 1
To generate a protective oxidated layer on the heating elements the furnace must be heated
to below the maximum temperature 100 °C (212 °F) for approx. 5 hours (Example: The
furnace is designed for a maximum temperature of approx. 1750 °C (3182 °F), hence a
temperature of approx. 1650 °C (3002 °F) should be entered on the controller. But this
temperature musst be reached over the full period of five hours).
The entered temperature (example: 1650 °C (3002 °F)) should be held for approx. 5 hours.
This process should be executed during the start-up, after replacing the heating elements or
to regenerate the oxide layer.
When the furnace is heated up for the first time, annoying odors may result, these are do to
the escape of binding agents from the insulation material. We recommend that the furnace
location be well ventilated during the first warm-up phase. Let the furnace cool down
naturally to room temperature.
Pos: 157 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - C hemis che Be stän digkeit @ 6 2\m od_1 403 074 144 601 _51. docx @ 30 044 2 @ @ 1
Chemical resistance: Pos: 158 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Che mische Bestä ndigkei t - H eizele ment e a us M olybdä n-Disilizid (M oSi2) - Besch reib ung @ 62 \mo d_1 403 073 7618 56_ 51. docx @ 30 0390 @ @ 1
Furnaces heated with molybdenum disilicide elements (MoSi2) are available for max.
furnace chamber temperatures of 1600, 1750 and 1800 °C. The max. furnace chamber
temperature refers to operation in air. At temperatures of over 800 °C, a layer of silicon
oxide occurs on the surface of the heating element, protecting it against further oxidation.
Such a protective layer is not formed at low temperatures. If the surface of the element is not protected, molybdenum and silicon oxidation can take place at temperatures of around
550 °C. This leads to the occurrence of a yellowish powder consisting primarily of
molybdenum oxide (MoO3). This chemical reaction has no disadvantageous effects on the
performance of the heating element.
Pos: 159 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Bei Nutz ung v on nicht bre nnb are n Schutz gase n wir d die max. zulässi ge O fen rau mte mpe ratu r u m 1 00 °C .. @ 62 \mo d_1 403 0851 839 34_ 51.d ocx @ 300 754 @ @ 1
If non-combustible protective gases are used, the max. permissible furnace chamber temperature is reduced to 100 °C (212 °F). When using forming gases (95/05 N2/H2 or
98/02 Ar/H2) oxidation burns at regular intervals are necessary and signs of increased wear
can appear on the heating elements and the insulation.
Pos: 160 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Wi derst and sfähi gkeit gege nüb er Met allen und Oxiden : @ 62\m od_ 140 307 415 364 8_51 .docx @ 3 004 68 @ @ 1
Resistance to metals and oxides: Pos: 161 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Gru nds ätzlich s ollten die Heiz elem ent e nie mals mit fes ten Mat erialie n in Be rüh ru ng ko mm en. @ 62\ mo d_1 4030 741 619 31_ 51.d ocx @ 300 494 @ @ 1
The heating elements should never come into contact with solid materials. Pos: 162 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Met allen: @ 62 \mo d_1 403 074 168 436_ 51. docx @ 30 052 0 @ @ 1
Metals: Pos: 163 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Met allen - In oxidi ere nde r Atmo sphä re z und ern (Aufg ru nd ei nes t her mische n Einflus ses in oxidie ren der @ 6 2\m od_1 403 074 176 407 _51. docx @ 30 054 6 @ @ 1
In an oxidizing atmosphere, the majority of metals form scale on the surface (oxide formation occurs due to a thermal influence in an oxidizing atmosphere). This metal oxide
may react with the silicon oxide of the heating element and shorten its service life. If the
![Page 36: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/36.jpg)
36
furnace is used to melt metals, it must be ensured that the furnace chamber is protected
against metal splashes, soiling and vapors which occur during the melting process
(particularly due to fluxing agents). Metal oxides also strongly corrode the insulation of the
furnace. Pos: 164 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Alkali: @ 6 2\m od_ 140 3074 183 083 _51. docx @ 30 057 2 @ @ 1
Alkalines: Pos: 165 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Alkali - Alkalih altige Verbi ndu nge n kön nen die H eizele ment e an gr eifen. Auße rd em f ühr en Alkali- ... @ 62\ mod _14 030 741 9064 9_5 1.d ocx @ 300 598 @ @ 1
Alkaline compounds may corrode the heating elements. Alkaline compounds also lead to
rapid destruction of the insulation. Pos: 166 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Ke ra mik: @ 62\ mod _14 0307 419 741 8_5 1.do cx @ 3 006 24 @ @ 1
Ceramic: Pos: 167 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Kera mik - Kera mik ka nn b ei dir ekte r Ber ühr ung das H eizele men t zer stör en. Au ße rde m kö nne n Salze ... @ 62\ mod _14 030 7420 403 2_5 1.do cx @ 3006 50 @ @ 1
Ceramic may destroy the heating element on direct contact. Salts and oxides from the ceramic may additionally react with the silicon oxide of the heating element and shorten its
service life. Pos: 168 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Übe rschri ft - Glas: @ 62\ mo d_14 030 742 140 16_ 51.d ocx @ 300 676 @ @ 1
Glass: Pos: 169 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Glas - Wird Glas gesch molze n, d ann ent hält die Of enat mos phä re Ve run reinig ung en, der en Zusa mm en... @ 6 2\m od_ 1403 074 262 824 _51. docx @ 30 070 2 @ @ 1
If glass is melted, the furnace atmosphere contains contaminants, the composition of which depends on the relevant glass product. These components usually act as a fluxing agent on
the silicon oxide layer. As a result, its viscosity is reduced and the silicon oxide flows
slowly down the heating element. However, new silicon oxide is immediately formed
again, with the result that the service life is not noticeably shortened. The insulation may
also be severely corroded depending on the type of glass.
Pos: 170 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Vo r In betri ebn ahm e d es Of ens sollt e di eser 24 St unde n a m Aufst ellung sort akklim atisie rt .. @ 3\ mod _11 9556 801 433 6_5 1.do cx @ 2 787 1 @ @ 1
Note
Before starting the furnace for the first time, allow it to acclimatize at its installation location for 24 hours.
Pos: 171 /TD/Sic her heit/De r Einsa tz vo n Färb elösu nge n kan n zu eine m ü ber pro porti onal en Ve rschlei ß d er Heiz elem ent e füh re n... @ 43\ mod _13 711 152 995 88_5 1.d ocx @ 233 530 @ @ 1 it
The use of paint solvents can lead to accelerated wear of the heating elements. If solvents are used, good pre-drying is necessary to minimize their impact on the heating elements.
Pos: 172 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - D aue rhaft er Be trie b bei maxi male r Temp era tur kan n zu erh öhte m Ver schlei ß an Heizel eme nten @ 1 7\m od_1 310 050 610 942 _51. docx @ 12 337 6 @ @ 1
Caution
Continuous operation at maximum temperature can lead to increased wear of the heating
elements and the insulation material. We recommend operating at approx. 50 °C below the
maximum temperature.
Pos: 173 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Sich erh eitshinw eise z u de n Bre nnhilfs mittel n (Sint ersc halen , Cha rge nbe hälte r usw. ) Kapi tel @ 119\ mo d_14 623 710 206 41_ 51.d ocx @ 469 822 @ @ 1
Notice
Observe the safety instructions for furnace furniture (sinter bowls, saggars, etc.) in the
“Stackable Saggars” section.
Pos: 174 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Wa rnun g - Zu m Schu tz d es Bedie ner s un d de s Ofe ns ist das Heiz pro gr amm gr unds ätzlich anzu halte n... @ 39 \mo d_1 365 0033 767 78_ 51.d ocx @ 221 385 @ @ 1
Warning - Danger of Electric Shock!
For the protection of the operator and the furnace, the heating program must be stopped
before the furnace is loaded. Ignoring this warning can result in electric shock.
Pos: 175 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Risse in de r Isolie run g - St and ard - Text @ 6\ mod _12 228 6141 243 6_5 1.do cx @ 4 354 8 @ @ 1
Cracks in the Insulation
The insulation of the furnace consists of a very high-quality refractory material. Heat
expansion may cause cracks in the insulation even after a few heating cycles. However,
these have no effect on the functioning or quality of the furnace. This is no reason for a
complaint.
![Page 37: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/37.jpg)
37
Pos: 176 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Risse in de r Isolie run g Hoc hte mpe ratu rof en H TCT 01/1 4-H TCT 01/1 6 - Gra fik @ 4 3\m od_ 137 120 922 758 9_51 .docx @ 2 336 35 @ @ 1
Fig. 16: Example: Cracks in the insulation after a few warm-up cycles (figure similar).
Pos: 177 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Bei de r Ersti nbet rieb nah me k ann es zu eine r Ver for mun g de r Heiz elem ente ko mme n... - M oSi2 @ 7 8\m od_ 141 648 962 2075 _51 .docx @ 3 419 40 @ @ 1
At the time of the first start-up and repeated heating up of the furnace one or more heating
elements may become deformed.
It is due to the nature of molybdenum-disilicide in the heating elements that they tend to
deform. But this appearance, however, has no influence on the function or quality of the
furnace and, hence, is no reason for a complaint.
Pos: 178 /TD/ Tr ansp ort _Mo nta ge_I nbet rieb nah me/Er stinb etrie bna hme /Verfo rm ung der Heizel eme nte (MoSi2 ) - Gra fik @ 78\m od_ 141 648 958 488 4_51 .docx @ 3 419 14 @ @ 1
Before the first start-up After the first start-up and/or after a few heating up phases.
Fig. 17: Example: Deformation of the heating elements made of molybdenum-disilicide
![Page 38: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/38.jpg)
38
Pos: 179 /TD/B etrie b_Bedi enu ng/Üb ersc hrift - Bedi enu ng 1 @ 3\ mod _11 867 406 850 02_ 51.d ocx @ 209 88 @ 1 @
5 Operation Pos: 180 /TD/B etrie b_Bedi enu ng/Co ntr oller/B1 50/B1 30/B17 0/C2 80/C2 90/C 295/P3 20/Ü bersc hrift - Co ntr oller @ 2\ mod _118 673 994 108 4_5 1.doc x @ 2 093 6 @ 2 @ 1
5.1 Controller Pos: 181 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Bedi enf eld Co ntroll er B4 00/C4 40/P4 70 - Gr afik/Text @ 67\ mod _14 084 547 2042 0_5 1.d ocx @ 314 641 @ @ 1
B400/C440/P470
4 1 2 3
Fig. 18: Control field B400/C440/P470 (similar to picture)
No. Description
1 Display
2 Control keys for "Start/Hold/Stop", "Menu" selection, "Back" function and information menu selection
3 Jog dial
4 USB interface for a USB stick
![Page 39: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/39.jpg)
39
Pos: 182 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Bedi enf eld Co ntroll er B4 10/C4 50/P4 80 - Gr afik/Text @ 67\ mod _14 084 535 9151 0_5 1.d ocx @ 314 615 @ @ 1
B410/C450/P480
4 1 2 3 5
Fig. 19: Control field B410/C450/P480 (similar to picture)
No. Description
1 Display
2 Control keys for "Start/Hold/Stop", "Menu" selection, "Back" function and information menu selection
3 Jog dial
4 USB interface for a USB stick
5 Over-temperature limiter with manual reset (optional)
Pos: 183 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Besc hrei bun g de r Eing abe von Te mpe ratu ren , Zeiten und „Sta rten “ d es Of ens sie he s epa rate A @ 3\ mo d_11 888 099 865 50_ 51.d ocx @ 210 15 @ @ 1
Note
See the separate operating instructions for a description of how to enter temperatures and
times and to "start" the furnace.
Pos: 184 /TD/B etrie b_Bedi enu ng/Sch altanl age/Ü bers chrift - Be dien -, Anzei ge- und Schaltel eme nte (je nach Au sfüh run g) @ 18\ mo d_13 141 919 615 17_ 51.d ocx @ 128 024 @ 1 @ 1
6 Operation, Display and Switch Elements (depending on design) Pos: 185 /TD/B etrie b_Bedi enu ng/Co ntr oller/B1 50/B1 30/B17 0/C2 80/C2 90/C 295/P3 20/Ü bersc hrift - Co ntr oller/O fen einsch alte n @ 0 \mo d_1 168 9470 513 96_ 51.d ocx @ 653 0 @ 2 @ 1
6.1 Turning on the Controller/Furnace Pos: 186 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Sym bole _Symb olda rstellu ngen /Cont rolle r ein schalt en_Abl auf-A nzeig e-Bem erk ung - Tabell e @ 6 7\m od_ 140 844 281 7311 _51 .docx @ 3 1430 2 @ @ 1
Switching on the Controller
Steps Display Comments
Turn on the power switch
/
Turn on the power switch by setting it to "I"
(power switch type differs depending on
features/furnace model)
The overview screen appears.
After a couple of
seconds, the temperature
is displayed.
If the temperature is displayed at the controller, the controller is ready to
operate.
![Page 40: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/40.jpg)
40
Pos: 187 /TD/B etrie b_Bedi enu ng/Co ntr oller/B1 50/B1 30/B17 0/C2 80/C2 90/C 295/P3 20/Ü bersc hrift - Co ntr oller/O fen aussc halte n @ 0\mo d_1 169 109 234 337 _51. docx @ 71 91 @ 2 @ 1
6.2 Turning off the Controller/Furnace Pos: 188 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Sym bole _Symb olda rstellu ngen /Cont rolle r a usschalt en_A blauf -Anzeig e-Be merk ung - Tab elle @ 67\ mod_ 140 845 045 782 9_5 1.doc x @ 3 144 07 @ @ 1
Turn off the controller
Steps Display Comments
Turn off the power
switch
/
Turn off the power switch by setting it to
"O"
(power switch type differs depending on
features/furnace model)
Pos: 189 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Alle n otwe ndige n Einst ellung en f ür ei ne ei nwan dfr eie Funkti on sin d b ereits im We rk e rfolg t. @ 67\ mod_ 140 845 010 186 5_5 1.doc x @ 3 143 55 @ @ 1
All the necessary settings for perfect functions have already been made at the factory. Pos: 190 /TD/B etrie b_Bedi enu ng/Üb ersc hrift - Zuluft heb el @ 3 \mo d_1 188 9820 213 89_ 51. docx @ 211 18 @ 2 @ 1
6.3 Air Inflow Lever Pos: 191 /TD/B etrie b_Bedi enu ng/ Zuluft heb el fü r Hub tür - u nd Klap ptü rmo dellen - 1 @ 3\ mo d_1 1889 775 247 41_ 51.d ocx @ 211 06 @ @ 1 D
The volume of air fed to the furnace can be adjusted with the fresh air lever. The fresh air lever is located at the side of the door at the bottom. The position is explained by the
symbols beside and on the lever.
FresFig. 1: Fresh air lever (figure similar)
Fresh air lever is Fresh air lever is max. open closed
Fig. 20: Regulating the feed of fresh air (symbols)
Pos: 192 /TD/B etrie b_Bedi enu ng/Hinw eis - Zulu fthe bel f ür Hu btü r- und Kla pptü rm odelle n - Te mpe rat urgl eichm äßig keit - 2 @ 12\ mod _12 833 342 9657 4_5 1.d ocx @ 977 05 @ @ 1
Note
If the fresh air lever is open, under certain circumstances this may affect temperature
uniformity in the furnace chamber.
Pos: 193 /TD/B etrie b_Bedi enu ng/Hinw eis - Bei geö ffne tem Zul ufth ebel wi rd Frischl uft in de n Ofe nra um z um Tr ockne n o der ... @ 62\ mod _14 031 6551 217 0_5 1.do cx @ 3008 21 @ @ 1
Note
When the fresh air lever is opened fresh air is guided in the furnace chamber for drying or
forced cooling. Please make sure that the fresh air lever is closed after drying in order to
achieve a good temperature uniformity during sintering.
The application notes of the zirconia manufacturer should be generally taken into
consideration in order to avoid damages at the charge e.g. due to fast cooling times.
![Page 41: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/41.jpg)
41
Pos: 194 /TD/B etrie b_Bedi enu ng/Üb ersc hrift - Beschic kun g/Cha rgie run g @ 0 \mo d_1 167 835 476 044_ 51. docx @ 54 18 @ 2 @ 1
6.4 Loading/Charging Pos: 195 /TD/B etrie b_Bedi enu ng/Üb ersc hrift - Of enb eschicku ng @ 46\ mo d_13 781 051 008 15_ 51.d ocx @ 243 960 @ @ 1
Furnace loading Pos: 196 /TD/B etrie b_Bedi enu ng/Die Ofe ntür ist vo rsichtig zu ö ffne n. - 1 @ 46\ mod _13 781 051 3685 5_5 1.d ocx @ 243 985 @ @ 1
Swivel the furnace door carefully away upwards. Pos: 197 /TD/B etrie b_Bedi enu ng/Es d ürf en n ur s olche Mat erialie n ei nges etzt we rde n, d ere n Eige nscha ften und Schm elztem pe ratu ren - 2 @ 4 6\mo d_1 378 105 152 975 _51. docx @ 24 401 0 @ @ 1
Only materials with known characteristics and melting temperatures may be used. Observe the material safety data sheets if necessary.
Pos: 198 /TD/B etrie b_Bedi enu ng/Bei m Beschick en d es Of ens ist da rauf zu a chte n, d ass d er Türk rag en s owie di e Heizel eme nte nicht - 3 @ 4 6\m od_1 378 105 175 083 _51. docx @ 24 403 5 @ @ 1
When loading the furnace, ensure that the door collar and the heating elements are not damaged. Under all circumstances, avoid coming into contact with the heating elements
when loading the furnace; this may lead to the immediate destruction of the heating
elements. Pos: 199 /TD/B etrie b_Bedi enu ng/We nn s ehr vi el Gut in d en O fen rau m ge stellt wi rd ka nn sic h die A ufheizz eit b eträ chtlich verlä nge - 5 @ 4 6\m od_ 1378 105 270 101 _51. docx @ 24 411 0 @ @ 1
If a large quantity of material is placed into the furnace chamber, the heating time may increase significantly.
Pos: 200 /TD/B etrie b_Bedi enu ng/Wir d ein Tie gel ve rwen det, so ist dar auf z u ac hten , das s das Schm elzgut vorsic htig i n de n Ti egel - 6 @ 46\ mo d_1 3781 052 432 76_ 51.d ocx @ 244 085 @ @ 1
If a crucible is used, make sure that the material to be melted is inserted carefully into the crucible. Crucibles are sensitive to shocks and impact. Metals expand more quickly and
strongly than the crucible when heated. Be sure that the recommendations of the crucible
manufacturer for maintenance and handling are carefully observed. Pos: 201 /TD/B etrie b_Bedi enu ng/Nac h d er Besc hickun g ist die Of entü r vo rsichti g zu sc hlie ßen. Allg em ein u nd M odell VH T- MO @ 46\ mod _13 781 0529 780 2_5 1.do cx @ 2441 35 @ @ 1
After loading, the furnace door must be closed carefully. The furnace door should be closed
gently in order not to damage the insulation. Make sure that the door is closed correctly. Pos: 202 /TD/B etrie b_Bedi enu ng/De r Of en ist nach Mö glichkeit nicht im h eißen Zusta nd z u öff nen . - 8 @ 4 6\m od_ 137 810 5380 970 _51 .docx @ 2 4416 0 @ @ 1
If possible, the furnace must not be opened when hot. If it has to be opened at a high temperature, this should take place as briefly as possible. Attention must be paid to
adequate protective clothing and room ventilation, see chapter "Safety". Pos: 203 /TD/B etrie b_Bedi enu ng/Ve rfär bun gen am E delsta hlblec h kön nen (b eson ders bei Ö ffnu ng i m hei ßen Zustan d) auft rete n - 9 @ 46\ mod_ 137 810 541 280 9_5 1.docx @ 2 441 85 @ @ 1
The stainless steel sheet may become discolored (particularly on opening when hot), but
this does not impair the function of the furnace. Pos: 204 /TD/B etrie b_Bedi enu ng/Weit ere Vorsich tsma ßn ahm en -Sorg en Sie daf ür, dass G ege nstä nde wi e Teeka nne n, Fl asch en vo m - 10 @ 46 \mo d_1 378 1054 434 24_ 51.d ocx @ 244 210 @ @ 1
Additional care measures
Ensure that objects like teapots, bottles, and similar items are kept away from the furnace.
Pos: 205 /TD/B etrie b_Bedi enu ng/Üb ersc hrift - Stap elba re C har gen behäl ter @ 42 \mo d_1 371 043 1592 42_ 51. docx @ 23 3330 @ 3 @ 1
6.4.1 Stackable Saggar Pos: 206 /TD/B etrie b_Bedi enu ng/ Für das C har gier en bi etet N abe rth erm spezi elle Ch arg enb ehält er a n. - Text-1 @ 4 7\m od_ 137 9429 321 242 _51. docx @ 24 727 5 @ @ 1
Nabertherm offers special-purpose saggars for charging the furnace. Pos: 207 /TD/B etrie b_Bedi enu ng/Üb ersc hrift - Ch arg enb ehält er f ür Sint eröf en LHT 03/1 7 D (Ein zug Mitte ) @ 119\ mo d_14 623 622 207 58_ 51.d ocx @ 469 741 @ @ 1
Saggar for sintering furnaces LHT 03/17 D Pos: 208 /TD/B etrie b_Bedi enu ng/Ein C har gen behäl ter setzt sic h g run dsätzlic h aus de r Sinte rschal e ... - Text- 1 @ 42\ mod _137 104 336 121 3_5 1.doc x @ 2 333 55 @ @ 1
A saggar basically consists of the sintering dish as base and the spacer ring with ventilation openings. The material is highly resistant to temperature fluctuations and can be used for
processes with short heat-up and cool-down times. Pos: 209 /TD/B etrie b_Bedi enu ng/Ch arg enb ehält er - Star ter -Set_Sint ersc hale _Distanz ring - LH T .. /.. LB_ LBR_LB Sp eed - Gr afik @ 43\ mod _13 710 433 840 41_5 1.d ocx @ 233 455 @ @ 1
Starter-Set Sintering Dish Spacer Ring with Ventilation Openings
Article No.: 699001066 Article No.: 699001054 Article No.: 699001055
Fig. 21: Charge Saggar
![Page 42: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/42.jpg)
42
Pos: 210 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Die ob en b esch riebe nen Bren nhilfmi ttel sin d fü r die Cha rgie run g un d Entn ahm e im kalte n ... @ 43 \mo d_1 371 467 2407 74_ 51. docx @ 23 3840 @ @ 1
Note
The accessories described above are designed for cold charging and discharging.
Removing the accessories in hot condition is not possible.
Pos: 211 /TD/B etrie b_Bedi enu ng/Bei m Cha rgie ren i m Of en ist da rauf zu ac hte n, d ass de r u nte re Ch arg enb ehält er - Text- 2 @ 4 2\m od_ 137 104 3365 375 _51 .docx @ 2 3338 0 @ @ 1
When charging the furnace it must be ensured that the lower charge carrier is generally resting on the spacer ring. This provides for air-circulation under this carrier and improves
the temperature uniformity. It is recommended to cover upper saggar with another sintering
dish as lid. Pos: 212 /TD/B etrie b_Bedi enu ng/ Zum Schutz des Ofen bod en_Dist anz ring - LH T 03 /17 D - Text -Gr afik @ 46\ mod_ 137 778 317 201 4_5 1.doc x @ 2 436 90 @ @ 1
For the protection of the sensitive furnace floor we
recommend that you make it a
general rule to leave the lowest
spacer ring in the furnace. The
spacer ring must be positioned
in the middle of the floor. This
ensures uniform heating of the
charge. When setting
down/lifting the sintering
dishes/spacer rings, they must
always be raised slightly to prevent any unwanted sideways movement of the sintering dishes/spacer rings.
Fig. 22: Spacer ring (similar to picture)
Pos: 213 /TD/B etrie b_Bedi enu ng/Das Start er -Set b esteh t aus ein em Ch arg enb ehält er, eine m Dista nzri ng... Tex t-3 @ 38 \mo d_1 363 105 335 161_ 51. docx @ 21 6100 @ @ 1
The starter set consists of a charge saggar, a spacer ring as a base and a second sintering dish as lid. The use of additional saggars (sintering dish and spacer ring) allows charging on
additional levels. Both furnace models are designed to get charged with up to three charge
saggars. Pos: 214 /TD/B etrie b_Bedi enu ng/Siche res C har gier en auf bis zu drei Eb ene n - LHT 03/1 7 D - Gr afik @ 46\ mod _137 778 093 061 2_5 1.doc x @ 2 436 65 @ @ 1
Fig. 23: Safe charging on up to three levels (similar to picture)
Pos: 215 /TD/B etrie b_Bedi enu ng/Üb ersc hrift - Anza hl be nötig ter Cha rgier ebe nen in d er Üb ersic ht: @ 43\ mod _13 710 4363 529 0_5 1.do cx @ 2334 80 @ @ 1
Number of required charge levels in overview: Pos: 216 /TD/B etrie b_Bedi enu ng/Sta pelb are C har gen beh älter 1_ 2_3 Eb ene - LH T .. . @ 4 2\m od_ 137 104 3370 653 _51 .docx @ 2 3340 5 @ @ 1
1 level: Starter set which includes 2 sintering dishes and 2 spacer rings = (A)
2 levels: Starter set (A) + 1 sintering dish + 1 spacer ring (B)
3 levels: Starter set (A) + 2 sintering dishes + 2 spacer rings (C)
180 °
180 °
180 °
2 level
1 level
3 level
![Page 43: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/43.jpg)
43
Pos: 217 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Wa rnun g - Zu m Schu tz d es Bedie ner s un d de s Ofe ns ist das Heiz pro gr amm gr unds ätzlich anzu halte n... @ 39 \mo d_1 365 0033 767 78_ 51.d ocx @ 221 385 @ @ 1
Warning - Danger of Electric Shock!
For the protection of the operator and the furnace, the heating program must be stopped
before the furnace is loaded. Ignoring this warning can result in electric shock.
Pos: 218 /TD/B etrie b_Bedi enu ng/Üb ersc hrift - Ch arg enb ehält er f ür Sint eröf en LHT 01/1 7 D (Ein zug Mitte ) @ 119\ mo d_14 623 623 001 85_ 51.d ocx @ 469 768 @ @ 1
Saggar for sintering furnaces LHT 01/17 D Pos: 219 /TD/B etrie b_Bedi enu ng/Bis zu dr ei Cha rge nbe hälte r lass en sich im O fen s tap eln - LH T/HTC T - Tex t-1 @ 43 \mo d_1 371 469 078 271 _51. docx @ 23 386 5 @ @ 1
Placing the zirconia product in charge saggars provides for optimum utilization of the furnace chamber. Up to three saggars can be stacked in the furnace. The integrated air slots
ensure a better air circulation of the charge. The upper saggar can be closed with a separate
ceramic lid.
Pos: 220 /TD/B etrie b_Bedi enu ng/Ch arg enb ehält er - Eckige C har gen beh älte r 11 0x75x 30 mit Deck el - LHT/H TCT - G rafik @ 43\ mo d_1 3714 712 026 37_ 51.d ocx @ 233 915 @ @ 1
Safe charging on up to three levels
Charge Saggar Lid for Charge Saggar
Article No.: 699000279 Article No.: 699000985
Fig. 24: Saggars with top lid
Pos: 221 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Die ob en b esch riebe nen Bren nhilfmi ttel sin d fü r die Cha rgie run g un d Entn ahm e im kalte n ... @ 43 \mo d_1 371 467 2407 74_ 51. docx @ 23 3840 @ @ 1
Note
The accessories described above are designed for cold charging and discharging.
Removing the accessories in hot condition is not possible.
Pos: 222 /TD/B etrie b_Bedi enu ng/Siche res C har gier en auf bis zu drei Eb ene n - LHT/H TCT - G rafik @ 43 \mo d_1 371 480 003 042 _51. docx @ 23 394 0 @ @ 1
Fig. 25: Safe charging on up to three levels (similar to picture)
Pos: 223 /TD/B etrie b_Bedi enu ng/Ch arg enb ehält er mittig ausric hte n... Hinw eise z um C har gen beh älter - Text -4 @ 43\ mod _13 714 8023 982 6_5 1.do cx @ 2 339 65 @ @ 1
The lowest charge saggar must be positioned in the middle (A/B) on the floor board
(ceramic insert board) to ensure uniforming heating of the charge.
2 level
1 level
3 level
![Page 44: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/44.jpg)
44
When charging make sure that neither the door collar nor the heating elements are
damaged. Make absolutely no contact with the heating elements. Any contact with the
heating elements can result in their destruction.
After charging, the furnace door must be closed carefully. The insulation of the furnace
door must not push the charge saggar into the furnace chamber.
Pos: 224 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Wa rnun g - Zu m Schu tz d es Bedie ner s un d de s Ofe ns ist das Heiz pro gr amm gr unds ätzlich anzu halte n... @ 39 \mo d_1 365 0033 767 78_ 51.d ocx @ 221 385 @ @ 1
Warning - Danger of Electric Shock!
For the protection of the operator and the furnace, the heating program must be stopped
before the furnace is loaded. Ignoring this warning can result in electric shock.
Pos: 225 /TD/W art ung _Reinig ung _Inst and haltu ng/Üb ersc hrift - Wa rtun g, Rei nigun g u nd I nstan dhalt ung @ 0\ mod _11 678 357 857 30_ 51.d ocx @ 543 6 @ 1 @ 1
7 Servicing, Cleaning, and Maintenance Pos: 226 /TD/W art ung _Reinig ung _Inst and haltu ng/Wa rtu ng, Rei nigu ng u nd I nsta ndhal tun g - Labo röf en @ 3\m od_ 118 890 843 796 9_51 .docx @ 2 107 1 @ @ 1
Warning! General Hazards!
Cleaning, lubrication, and maintenance tasks may only be performed by authorized experts
following the maintenance instructions and accident protection guidelines. We recommend
that maintenance and repair be performed by Nabertherm GmbH Service. Failure to comply
runs the risk of bodily injury, death, or significant property damage!
Warning - Danger due to Electrical Current!
Work on the electrical equipment may only be performed by qualified, authorized
electricians!
During maintenance work, the voltage supply to the furnace and/or switching system must
be switched off to prevent unintentional commissioning. Disconnect the mains power
connector due to reasons of safety.
Operators may only correct malfunctions which are obviously due to operational error!
Wait until the furnace chamber and attaching parts have cooled to room temperature.
The furnace must be visually inspected at regular intervals for damage. The interior of the
furnace must also be cleaned as required (e.g. vacuuming out) Attention: Do not bang
against the heating elements to avoid breaking them.
While work is being performed on the furnace, the furnace and work room must additionally be ventilated with fresh air.
Safety systems removed during maintenance tasks must be replaced after the work.
Warning of swinging loads in the workshop (e.g. crane systems). Work under a lifted load
(e.g. a lifted furnace or switching system) is not permitted.
Safety switches and any limit switches present must be checked for function periodically
(DGUV V3) or according to the national guidelines of the country of operation.
To ensure proper temperature regulation of the furnace, the thermocouple must be checked
for damage before every process.
If necessary, retighten the element holders (see chapter "Replacing the Heating Element").
Before carrying out this work, the voltage supply to the furnace and/or switching system
must be switched off (disconnect mains power connector). The regulations (DGUV V3) or
corresponding national regulations in the relevant country of operation must be observed.
There are one or more contactors in the control system. The contacts of these circuit
breakers are wearing parts and must therefore be serviced and/or replaced regularly
(DGUV V3) or according to the national guidelines of the country of operation.
The switching system cabinet (if available) contains vent grilles with integrated filter mats.
These must be cleaned and/or replaced at regular intervals in order to ensure sufficient
![Page 45: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/45.jpg)
45
intake and outflow of air from the switching system. During melting operation, the
switching cabinet door must always be firmly closed.
Pos: 227 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Üb ersch rift - Ofe nisolie run g @ 7 \mo d_1 233 821 365 549_ 51. docx @ 49 423 @ 2 @ 1
7.1 Furnace Insulation Pos: 228 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Isoli eru ng_R epa ratu ren an der Isolie run g od er der A ustau sch vo n Bau teilen i n d er Heizk am mer ... @ 91\ mo d_14 303 928 665 47_ 51.d ocx @ 379 618 @ @ 1
Repairs to the insulation or the replacement of components in the heating chamber may only be performed by persons who are trained about possible hazards and protective
measures and can apply this knowledge without supervision.
Pos: 229 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Üb ersch rift - Bei Arb eiten an der Isolie run g od er Aus tausc h vo n Baut eilen i m Of enr aum sind . .. @ 91\ mod _14 3039 319 296 8_5 1.do cx @ 3 796 44 @ @ 1
During the work on the insulation or the replacement of components in the furnace chamber, the following points must be observed:
Pos: 230 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Isoli eru ng (kF )_Bei Re par atu ren ode r ... könn en silik ogen e Stä ube f reig esetz t wed en .. . - Teil 1 @ 8 6\m od_ 142 5894 290 877 _51. docx @ 36 631 2 @ @ 1
When repairs are made or demolition work is performed, silicon dust may be released.
Depending heat-treated materials contained in the furnace, further contaminants may be contained in the insulation. To exclude possible health risks, dust concentrations must be
kept to a minimum during any work performed at or near the insulation. In many countries
there are specific occupational safety limits. You can acquire more relevant information by
investigating the relevant legal specifications in your country.
Pos: 231 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Isoli eru ng (kF )_Stau bkon zent ratio nen sollte n so nied rig wie möglic h g ehalt en we rde n ... - Teil 2 @ 86 \mo d_1 425 8993 832 39_ 51.d ocx @ 366 338 @ @ 1
Dust concentrations should be kept as low as possible. Dust must be removed using a suction device or a vacuum cleaner with a high-performance filter (HEPA – category H).
Strong air currents such as drafts, for example, must be prevented. Pressurized air or brush
must not be used for cleaning. Piles of dust must be sprinkled. Pos: 232 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Isoli eru ng (kF )_Bei Arb eite n an de r Isolie run g sollt e ein At emsc hutz mit FFP2 ... - Teil 3 @ 86\ mod _14 258 994 437 68_5 1.d ocx @ 366 364 @ @ 1
During work on the insulation a respirator mask with an FFP2 filter or an FFP3 filter must
be used. The work clothing must be loose and cover the body completely. Gloves and
goggles must be worn. Soiled clothing should be cleaned before it is removed with a
vacuum cleaner equipped with a HEPA filter. Pos: 233 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Isoli eru ng (kF )_De r Kont akt mi t Hau t un d Auge n sollt e ve rmie den w erd en. ... - Teil 4 @ 86\ mod _14 2589 944 840 1_5 1.do cx @ 3 663 90 @ @ 1
Contact with skin and eyes should be avoided. The impact of fibers on skin or eyes can lead to mechanical irritation which, in turn, causes reddening and itchiness. After completing the
work, or after direct contact, the skin must be washed with soap and water. If there is
contact with the eyes, they must be washed out carefully for several minutes. If necessary,
an eye doctor should be consulted. Pos: 234 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Isoli eru ng (kF )_Rau che n, Essen un d T rinke n a m Arbei tsplatz ist u nters agt. - Teil 5 @ 8 6\mo d_1 425 899 454 875 _51. docx @ 36 641 6 @ @ 1
Smoking, eating and drinking at the workplace is prohibited. Pos: 235 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Isoli eru ng (kF )_Bei Arb eite n an de r Isolie run g sin d in De utschl and die Tech nische n Re geln - Teil 6 @ 86\ mo d_1 4258 994 596 18_ 51.d ocx @ 366 442 @ @ 1
In Germany, during work involving insulation, the technical rules for hazardous substances must be applied. In particular: TRGS 500, TRGS 521, TRGS 558, TRGS 559, TRGS 900;
http://www.baua.de (German). Pos: 236 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Isoli eru ng (kF )_Weite rfü hre nde Info rm ation en z um U mga ng mit Faser mat erialie n e rhalt en .. . - Teil 7 @ 8 6\m od_ 142 5899 463 970 _51. docx @ 36 646 8 @ @ 1
Additional information regarding how to handle fibrous materials can be found at http://www.ecfia.eu (English).
Pos: 237 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Isoli eru ng (kF )_Bei d er Ent sor gun g de r M ateri alien si nd n atio nale und regi onale Richtlini en zu ... @ 91\ mod _14 303 928 610 87_ 51.d ocx @ 379 592 @ @ 1
When the materials are discarded, national and regional guidelines must be observed. The
possible presence of hazardous contaminants generated by the furnace process must be
taken into account.
Pos: 238 /TD/W art ung _Reinig ung _Inst and haltu ng/Üb ersc hrift - Stillsetz ung der Anlag e fü r War tung sar beite n @ 0 \mo d_1 167 835 907 715_ 51. docx @ 54 45 @ 2 @ 1
7.2 Shutting the System Down for Maintenance Pos: 239 /TD/W art ung _Reinig ung _Inst and haltu ng/Stillsetz ung de r Anlag e fü r Wa rtun gsar beite n @ 15\ mod _130 467 194 691 1_5 1.doc x @ 1 167 73 @ @ 1
Warning! General Hazards!
Cleaning, lubrication, and maintenance tasks may only be performed by authorized experts
following the maintenance instructions and accident protection guidelines. We recommend
that maintenance and repair be performed by Nabertherm GmbH Service. Failure to comply
runs the risk of bodily injury, death, or significant property damage!
![Page 46: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/46.jpg)
46
Wait until the furnace chamber and attached parts have cooled to room temperature.
The furnace must be completely emptied
Inform operating personnel and name supervisors
Switch off main switch and/or disconnect the power cord.
Lock the main switch (if available) and secure against restoration of power using a
padlock.
Attach a warning sign on the main switch
Clean up the maintenance area as far as possible.
Check for disconnection of power.
Ground and short-circuit the working area.
Cover any nearby parts still under power. Pos: 240 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Warn ung - Be rüh ren Sie keine n Ge gen stan d, o hne i hn zu vor auf s eine Te mpe rat ur übe rpr üft zu ha ben. @ 2 5\mo d_1 341 394 122 108 _51. docx @ 16 678 5 @ @ 1
Warning - Dangers During Normal Operation!
Do not touch any object without first having checked its temperature.
Pos: 241 /TD/E rsatz -/Versc hleiß teile/W arn ung - Ge fah ren dur ch el ektrisc hen St ro m - N etzstec ker zi ehe n - f ür alle Öf en @ 3\ mod_ 118 949 139 078 7_5 1.doc x @ 2 127 3 @ @ 1
Warning - Danger of Electric Shock!
Work on the electrical equipment may be done only by qualified, authorized electricians.
During work it must be ensured that the furnace and the switching equipment cannot be
activated by mistake (pull out the power plug) and that all moving parts in the furnace are
secured. Observe DGUV V3 or the corresponding national regulations in the country where
the furnace is installed. Wait until the furnace and the connected parts have cooled to room
temperature.
Pos: 242 /TD/W art ung _Reinig ung _Inst and haltu ng/Üb ersc hrift - Wa rtun gsar beit en a m Of en @ 0\m od_ 116 783 630 668 1_51 .docx @ 5 463 @ 2 @ 1
7.3 Regular Maintenance of the Furnace Pos: 243 /TD/W art ung _Reinig ung _Inst and haltu ng/Re gelm äßi ge Wa rtun gsa rbeit en (W art ungst abell e) a m LH T/LH T- LBR/RHT -H-V @ 119\ mod _14 623 528 285 55_ 51.d ocx @ 469 577 @ @ 1
A B C D E F
1 2 3 D W M Q J X
1
X
2
Safety inspection
compliant with DGUV V3 or the corresponding national regulations
Compliant with regulation
Compliant with regulation ●
EMERGENCY-OFF
system (if included)
Press the pushbutton
● ●
Safety and limit switch (if included)
Functional test ● ●
Furnace chamber, vent holes and vent pipes
Clean and check for damage, vacuum out carefully
● ●
Seal surfaces:
Door collar/furnace collar
Visual inspection ● ●
Seals (if included)
Clean/replace ● ●
![Page 47: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/47.jpg)
47
A B C D E F
1 2 3 D W M Q J X
1
X
2
Heating elements/Inner
process hood (if included)
Visual inspection
● ●
Screws of the element holders see
section entitled "Check Heating Element Screws" (if included)
Check and carefully tighten element holder screws.
●1
●2
●
1first time
2subsequent
Verify uniform power consumption of the heating system
Functional test ● ●
Thermocouple Visual inspection
(visual part of the thermocouple in the furnace chamber)
● ●
Setting lift bottom (if included)
Check whether the lift bottom completely closes
● ●
Settings on the
over-temperature limit controller (if included)
Every time the heat
treatment program is changed
Every time the heat treatment
program is changed check the
trigger temperature (alarm value)
on the over-temperature limit
controller.
●
Measurement and control precision
Calibration ● ●
Pos: 244 /TD/W art ung _Reinig ung _Inst and haltu ng/L ege nde: Reinig en,Pr üfe n,Ersatz teilstr ate gie,Wa rtun gsinte rvalle - f ür all e War tun gstab ellen @ 16 \mo d_1 306 221 4717 20_ 51. docx @ 118 223 @ @ 1
Symbols:
A = Component/Position/Function / B = Measure / C = Spare Part Stocks / D = Maintenance Interval / E = Performed by / F = Comment
Spare part stocks: 1 = Stocks urgently recommended
2 = Stocks recommended / 3 = As required, not relevant
Maintenance Interval Note: If environmental conditions are severe, the maintenance intervals must be shorter.
D = Daily, before each start of the furnace
W = Weekly
M = Monthly /
Q = Quarterly
Y = Yearly /
● = Check, replace
Implemented by: X1 = Operating personnel
X2 = Specialist personnel
Pos: 245 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Wa rnun g - Arbeit en a n d er el ektrisc hen Ausrü stun g dü rfe n nu r vo n ... @ 2\ mod _11 842 280 210 19_5 1.d ocx @ 196 73 @ @ 1
Warning - Danger of Electric Shock!
Work on the electrical equipment may be done only by qualified, authorized electricians.
Pos: 246 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Wa rtu ngsa rbeit en dürf en nur v on auto risier ten Fac hpe rson en u nte r Beac htun g vo n ... @ 1\ mod _11 739 7170 468 9_5 1.d ocx @ 1113 1 @ @ 1
Note
Maintenance work must be performed by authorized personnel following the maintenance instructions and the accident prevention regulations. We recommend that the maintenance
and repair work be carried out by the service team of Nabertherm GmbH.
![Page 48: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/48.jpg)
48
Pos: 247 /TD/W art ung _Reinig ung _Inst and haltu ng/Üb ersc hrift - Re gelm äßig e Wa rtun gsar beite n – D oku ment ation @ 1 7\m od_1 310 024 892 193 _51. docx @ 12 254 1 @ 2 @ 1
7.4 Regular Maintenance Tasks – Documentation Pos: 248 /TD/W art ung _Reinig ung _Inst and haltu ng/Re gelm äßi ge Re gelm äßig e Wa rtun gsa rbeit en – D oku ment atio n (He rdwa gen ofen ) @ 17\ mod _13 100 2496 832 2_5 1.do cx @ 1 225 64 @ @ 1
A B C D E F
1 2 3 T W M Q J X
1
X
2
Type plate Readable condition ● ●
Instruction Manual Ensure it is kept at
the furnace ● ● ●
Instructions for
components
Ensure they are kept
at the furnace ● ● ●
Pos: 249 /TD/W art ung _Reinig ung _Inst and haltu ng/L ege nde: Reinig en,Pr üfe n,Ersatz teilstr ate gie,Wa rtun gsinte rvalle - f ür all e War tun gstab ellen @ 16 \mo d_1 306 221 4717 20_ 51. docx @ 118 223 @ @ 1
Symbols:
A = Component/Position/Function / B = Measure / C = Spare Part Stocks / D = Maintenance Interval / E = Performed by / F = Comment
Spare part stocks: 1 = Stocks urgently recommended
2 = Stocks recommended /
3 = As required, not relevant
Maintenance Interval Note: If environmental conditions are severe, the maintenance intervals must be shorter.
D = Daily, before each start of the furnace
W = Weekly M = Monthly /
Q = Quarterly
Y = Yearly /
● = Check, replace
Implemented by: X1 = Operating personnel
X2 = Specialist personnel
Pos: 250 /TD/W art ung _Reinig ung _Inst and haltu ng/Üb ersc hrift - Wa rtun gsar beit en He rdwa gen ofen W...-H eizele men te/Of enka mm er @ 17\ mod _13 099 527 919 98_ 51.d ocx @ 121 646 @ 2 @ 1
7.5 Regular Maintenance Tasks – Heating Elements/Furnace Chamber Pos: 251 /TD/W art ung _Reinig ung _Inst and haltu ng/Re gelm äßi ge Wa rtun gsa rbeit en H T 04 /16 -HT 450/ 18-H eizele men te/Of enka mm er -1 @ 22\ mod _13 330 2152 565 0_5 1.do cx @ 1 536 25 @ @ 1
A B C D E F
1 2 3 T W M Q J X
1
X
2
Heating elements Visual check: Oxide layer formation, crack
formation
● ● ●
Heating elements Replace ● ● ●
Run through the heating elements
Clean
● ● ●
No later than when the heating
elements are
replaced
Run through the heating elements
Replacement ● ● ●
Connect the heating elements
Wiring to the connection ends,
corrosion tendency drill
ends traces of burning)
● ● ●
Support tubes Visual check, correct
position, sag, crack
formation
● ●
![Page 49: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/49.jpg)
49
A B C D E F
1 2 3 T W M Q J X
1
X
2
Support tubes Replace ● ● As necessary Pos: 252 /TD/W art ung _Reinig ung _Inst and haltu ng/L ege nde: Reinig en,Pr üfe n,Ersatz teilstr ate gie,Wa rtun gsinte rvalle - f ür all e War tun gstab ellen @ 16 \mo d_1 306 221 4717 20_ 51. docx @ 118 223 @ @ 1
Symbols:
A = Component/Position/Function / B = Measure / C = Spare Part Stocks / D = Maintenance Interval / E = Performed by / F = Comment
Spare part stocks: 1 = Stocks urgently recommended
2 = Stocks recommended /
3 = As required, not relevant
Maintenance Interval Note: If environmental conditions are severe, the maintenance intervals must be shorter.
D = Daily, before each start of the furnace
W = Weekly
M = Monthly /
Q = Quarterly Y = Yearly /
● = Check, replace
Implemented by: X1 = Operating personnel
X2 = Specialist personnel
Pos: 253 /TD/W art ung _Reinig ung _Inst and haltu ng/Üb ersc hrift - Sch raub en der H eizele men te ü ber prü fen @ 3\ mod _11 901 164 0510 2_5 1.d ocx @ 2187 8 @ 2 @ 1
7.6 Checking the Heating Element Screws Pos: 254 /TD/E rsatz -/Versc hleiß teile/W arn ung - Ge fah ren dur ch el ektrisc hen St ro m - N etzstec ker zi ehe n - f ür alle Öf en @ 3\ mod_ 118 949 139 078 7_5 1.doc x @ 2 127 3 @ @ 1
Warning - Danger of Electric Shock!
Work on the electrical equipment may be done only by qualified, authorized electricians.
During work it must be ensured that the furnace and the switching equipment cannot be
activated by mistake (pull out the power plug) and that all moving parts in the furnace are
secured. Observe DGUV V3 or the corresponding national regulations in the country where
the furnace is installed. Wait until the furnace and the connected parts have cooled to room
temperature.
Pos: 255 /TD/E rsatz -/Versc hleiß teile/Vo rsicht - Besc hädi gung von B auteil en! H eizele ment aust ausch en @ 3\ mod _11 8949 163 440 0_5 1.doc x @ 2 128 6 @ @ 1
Caution - damage to components!
Heating elements are extremely sensitive to breaking. Any strain on or rotation of the
heating elements must be avoided. Failure to observe this rule will lead to the immediate
destruction of the sensitive heating elements.
Pos: 256 /TD/E rsatz -/Versc hleiß teile/Abl uftr öhrc hen am LHT ../.. - Of en ( Tisch mod ell) -Dem onta ge - 2 @ 3\ mod _11 8949 359 365 3_5 1.do cx @ 2 132 5 @ @ 1
Exhaust air pipe
First slacken the screws (A) from the exhaust air pipe guard plate using the enclosed Allen
key. Lift the guard plate (B) from the exhaust air pipe and then carefully pull the exhaust air
pipe (C) out straight upwards. Store the exhaust air pipe in a safe place, as its material is
very sensitive.
![Page 50: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/50.jpg)
50
Fig. 26: Removing the exhaust air pipe
Pos: 257 /TD/E rsatz -/Versc hleiß teile/ Ofen deckel LHT ../.. - Of en ( Tisch mo dell) - De mont age - 3 @ 3\ mod _11 894 937 817 94_ 51.d ocx @ 213 38 @ @ 1
Removing the furnace lid
Slacken the screws (A) at the rear and front of the lid with the enclosed Allen key, pull the
lid forwards towards the furnace door (B) and remove upwards (C).
Fig. 27: Furnace lid
Pos: 258 /TD/W art ung _Reinig ung _Inst and haltu ng/Sch rau ben der Heizele me nte am LH T- Ofen üb erp rüfe n @ 3\m od_ 1190 122 975 915 _51. docx @ 21 903 @ @ 1
Checking the element holder screws and retightening if necessary
All screws must be carefully retightened after one week's operation and then once per year.
Any strain on or rotation of the heating element must be avoided. The element must be
secured with pliers to prevent rotation. Failure to observe this rule will lead to the
immediate destruction of the sensitive heating elements.
Fig. 28: Securing element holder screws.
A B C
Element holder
A
Hexagonal screw (WAF 7)
B C
![Page 51: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/51.jpg)
51
Pos: 259 /TD/E rsatz -/Versc hleiß teile/ Ofen deckel LHT ../.. - Of en ( Tisch mo dell) - M onta ge - 9 @ 3\ mod_ 118 949 508 589 2_5 1.docx @ 2 141 9 @ @ 1
Assembling the furnace lid
Position lid (A) and push to rear wall (B). Make sure that the lid holes align with the holes
on the housing. Secure the lid to the front and rear wall of the furnace housing using the
previously slackened screws (C).
Fig. 29: Furnace lid
Pos: 260 /TD/E rsatz -/Versc hleiß teile/Abl uftr öhrc hen am LHT ../.. - Of en ( Tisch mod ell) -Mo ntag e - 10 @ 3\ mod _11 8949 529 015 9_5 1.do cx @ 2 143 2 @ @ 1
Assembling the exhaust air pipe
Carefully push the exhaust air pipe (A) into the intended opening. The head of the exhaust
air pipe must lie on the furnace lid. Reassemble the exhaust air pipe guard plate (B) using the previously slackened screws (C).
Fig. 30: Assembling the exhaust air pipe
Pos: 261 /TD/E rsatz -/Versc hleiß teile/N etzste cker einst ecken (sie he Kapi tel „Ans chluss an das Elekt ron etz“ ) d anac h Netzsc halte r .. . @ 3 \mo d_1 190 0145 595 88_ 51.d ocx @ 218 47 @ @ 1
Commissioning
Insert the mains power connector (see chapter "Connection to the Mains Electricity"), then
switch on the power switch and check the function of the furnace (see chapter "Operation").
Pos: 262 /TD/W art ung _Reinig ung _Inst and haltu ng/Üb ersc hrift - Bet rieb - un d Hilfsmit tel @ 0\m od_ 116 783 7161 458 _51 .docx @ 54 99 @ 2 @ 1
7.7 Operating and Auxiliary Materials Pos: 263 /TD/W art ung _Reinig ung _Inst and haltu ng/Üb ersc hrift - Rei nigun gsmitt el 1. 1 @ 0\mo d_1 167 837 229 988 _51. docx @ 55 08 @ 2 @ 1
7.8 Cleaning Products Pos: 264 /TD/W art ung _Reinig ung _Inst and haltu ng/Rei nigu ngsmit tel ( für alle Ö fen ) @ 0 \mo d_1 158 6575 923 97_ 51. docx @ 284 3 @ @ 1
Follow the procedure for shutting down the furnace system (in the "Operation" section).
Then the power plug must be pulled out of the socket. Wait until the furnace cools down
naturally.
Use commercially available detergent which is either water-based or non-combustible and
free of any solvents to clean the housing of any deposits; use a vacuum cleaner for the
interior.
Follow the labeling and the instructions on the packaging of the detergent.
A B C
A B C
![Page 52: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/52.jpg)
52
Wipe the surface with a damp, lint-free cloth. The following detergents can also be used:
This list must be completed by the operator.
Component and
location
Detergent
Outer surfaces
(frames *)
Use commercially available detergent which is either water-or non-
combustible and free of any solvents for cleaning *)
Outer surface
(stainless steel)
Stainless still cleaner
Interior Carefully clean with a vacuum cleaner
(avoid the heating elements)
Insulation materials Carefully clean with a vacuum cleaner (avoid the heating elements)
Door seal (if included)
Use commercially available detergent which is either water-or non-combustible and free of any solvents for cleaning
Instrument panel Wipe the surface with a damp, lint-free cloth. (e.g. glass cleaner)
*) You must be sure that the cleaner does not damage the water-soluble and, hence, environmentally safe
paint (the clear should be tried first on an interior, normally unseen location).
Fig. 31: Detergent
Do the cleaning from beginning to end without breaks to protect the surfaces.
Remove the detergent completely from the surfaces by wiping them with a damp, lint-free
cloth.
After cleaning all the supply lines, check all the connections for leaks, loose connections,
abrasion and damage; report any shortcomings found immediately!
Please follow the section entitled "Environmental Protection Rules and Regulations"
Caution
The furnace, the furnace chamber and attached components must NOT be cleaned using a
high-pressure cleaner.
Pos: 265 /TD/St öru nge n/Fehle rsuch e/Öf en/Üb ersc hrift - Stö run gen @ 0\ mod _11 678 3772 478 1_5 1.do cx @ 5526 @ 1 @ 1
8 Malfunctions Pos: 266 /TD/St öru nge n/Fehle rsuch e/Öf en/Hinw eise z u Stör ung en a m Of en ode r Con trolle r ( allge mein e Hinweis e wie man sic h ve rhalt en s oll) @ 92\ mod _14 315 188 1254 4_5 1.d ocx @ 3810 80 @ @ 1
Work on the electrical system may be done only by qualified, authorized electricians. Operators may only rectify faults that are obviously due to operating errors.
Call your local electrician for faults that you cannot localize.
If you have any questions, problems, or requirements, contact Nabertherm GmbH. By mail,
phone, or e-mail -> See "Nabertherm Service".
Phone advice is free and non-binding for our customers – all you pay is the phone costs.
In case of mechanical damage, send an email containing the above information and a digital
photo of the damaged part and a photo of the complete furnace to the following address:
-> see "Nabertherm Service".
If a fault cannot be rectified with the described solutions, contact our service hotline
directly.
Have the following information at hand when you phone. This makes it easier for our
customer service to answer your questions.
![Page 53: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/53.jpg)
53
Pos: 267 /TD/St öru nge n/Fehle rsuch e/Üb ersch rift - Fehle rmel dun gen des C ontr ollers @ 7 2\mo d_1 409 671 241 104 _51. docx @ 32 554 2 @ 2 @ 1
8.1 Error Messages of the Controller Pos: 268 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/ Fehle rm eldu ngen des Cont rolle rs - Ta belle @ 12 4\mo d_1 465 893 737 464 _51. docx @ 54 795 0 @ @ 1
ID+
subID
Text Explanation Remedy
Communication errors
01-01 Bus zone The communication connection to a controller module is
malfunctioning
Check whether the controller module is securely seated.
Are the LEDs on the controller modules red?
Check the cable between the operating unit and
the controller module.
The plug of the connecting cable in the operating
unit is incorrectly plugged in.
01-02 Bus
communication module
The communication connection
to the communication module (Ethernet/USB) is
malfunctioning
Check whether the communication module is
securely seated
Check the cable between the operating unit and
the communication module
Sensor errors
02-01 TC open Check the thermocouple, thermocouple terminals and cable
Check the contact of the thermocouple cable in
connector X1 on the controller module (contacts
1+2)
02-02 Beyond TC range Check the set thermocouple type
Check the polarity of the thermocouple
connection
02-03 Reference junction error
The controller module is defective
02-04 Reference
junction too hot
The temperature in the switchgear is too high
(approx. 70 °C)
The controller module is defective
02-05 Reference junction too cold
The temperature in the switchgear is too low (approx. -10 °C)
02-06 Encoder disconnected
Fault on the 4-20 mA input of the controller (<2 mA)
Check the 4-20 mA sensor
Check the connection cable to the sensor
02-07 Sensor element defective
The PT100 or PT1000 sensor is defective
Check the PT sensor
Check the connection cable to the sensor (cable
break / short-circuit)
System errors
03-01 System memory Error after firmware updates1)
Defect in the operating unit1)
03-02 ADC error The communication between the
ADC converter and the
controller is malfunctioning
Replace the controller module1)
![Page 54: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/54.jpg)
54
ID+
subID
Text Explanation Remedy
03-03 System file faulty The communication between the display and the memory module
is malfunctioning
Replace the operating unit
03-04 System monitoring
Execution of the program on the operating unit is malfunctioning
(watchdog)
Replace the operating unit
The USB flash drive was pulled out prematurely
or is defective
Replace the controller and switch it on
03-05 Zone system monitoring
Execution of the program on a controller module is
malfunctioning (watchdog)
Replace the controller module1)
Replace the controller and switch it on1)
03-06 Self-test error Contact Nabertherm Service1)
Monitoring
04-01 No heat output No temperature increase in
ramps when the heating output
does not equal 100% for 12
minutes and if the temperature
setpoint is greater than the current furnace temperature
Acknowledge the error (deenergize if necessary)
and check the safety contactor, door switch,
heating control and controller.
04-02 Over-temperature The temperature of the guide
zone exceeds the max. program
setpoint or the maximum furnace temperature by 50°
Kelvin (200 °C or higher)
The equation for the shutdown
threshold is:
maximum program setpoint +
zone offset of the master zone +
batch control offset [max] (if
batch control is active) + over-
temperature shutdown threshold
(P0268, e.g. 50 K)
Check the solid state relay
Check the thermocouple
Check the controller
A program has started at a
furnace temperature that is
greater than the maximum
setpoint in the program
Wait until the temperature of the furnace has
dropped before starting the program.
If this is not possible, add a wait time as a start
segment, followed by a ramp with the desired
temperature (STEP = 0 minute duration for both
segments)
Example:
700 °C -> 700 °C, time: 00:00
700 °C -> 300 °C, time: 00:00
The normal program then starts here.
In version 1.14 and later, the actual temperature is
also taken into account at the start.
04-03 Power failure The set limit for restarting the furnace has been exceeded
If necessary, use an uninterruptible power supply
![Page 55: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/55.jpg)
55
ID+
subID
Text Explanation Remedy
04-04 Alarm A configured alarm has been triggered
04-05 Self-optimization failed
The measured values are not plausible
Do not perform self-optimization in the lower temperature range of the furnace work area
Battery weak A power failure may no longer be processed properly.
Export all the parameters to a USB flash drive
Replace the battery (see “Specifications”) Pos: 269 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/ Fehle rm eldu ngen kön nen dur ch zwei malig es Dr ücken des D reh rad es ( Jog Di al) zu rück geset zt we rde n... @ 72\ mod _14 098 220 273 29_ 51.d ocx @ 325 570 @ @ 1
Error messages can be reset by pressing twice the jog dial . If there is another error message, contact Nabertherm Service. Recirculation motors (if included) also remain
switched on in case of an error until the temperature falls below the set cut-off temperature.
Pos: 270 /TD/St öru nge n/Fehle rsuch e/Üb ersch rift - War nun gen des C ont rollers @ 8 8\m od_ 142 7208 346 050 _51. docx @ 37 174 6 @ 2 @ 1
8.2 Warnings of the Controller Pos: 271 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Wa rnu nge n wer den nicht im Fehl era rchiv a nge zeigt. Sie wer den nu r au f de r Anzei ge u nd in de r Dat ei... @ 88\ mod _14 272 084 595 25_5 1.d ocx @ 371 772 @ @ 1
Warnings are not displayed in the error archive. They are only displayed on the display and
in the file of the parameter export. Warning do not generally lead to a program crash.
Pos: 272 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Wa rnu nge n des Cont rolle rs - Ta belle @ 12 4\m od_ 1465 894 169 708 _51. docx @ 54 797 6 @ @ 1
No. Text Explanation Remedy
00 Gradient monitoring
The limit value of the configured gradient monitoring
has been exceeded
For the causes of this error, see “Gradient monitoring”
The gradient is set too low
01 No control
parameters
No “P” value has been entered
for the PID parameters
Enter at least one “P” value in the control
parameters. This may not be 0
02 Batch element
defective
No batch element has been
determined with the program
running and with batch control
activated
Insert a batch element
Disable the batch control in the program
Check the batch thermocouple and its cable for
damage
03 Cooling element
defective
The cooling thermocouple is not
plugged in or is defective
Plug in a cooling thermocouple
Check the cooling thermocouple and its cable for
damage
If a defect in the cooling thermocouple occurs
when controlled cooling is active, the system switches to the thermocouple of the master zone.
04 Documentation
element defective
The documentation
thermocouple is missing or
defective.
Plug in a documentation thermocouple
Check the documentation thermocouple and its
cable for damage
05 Power failure A power failure has been detected.
Program termination has not
occurred
None
06 Alarm 1 - belt The configured belt alarm 1 has
been triggered
Optimization of the control parameters
Alarm value too rigid
![Page 56: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/56.jpg)
56
No. Text Explanation Remedy
07 Alarm 1 - min The configured min alarm 1 has been triggered
Optimization of the control parameters
Alarm value too rigid
08 Alarm 1 - max The configured max alarm 1 has
been triggered
Optimization of the control parameters
Alarm value too rigid
09 Alarm 2 - belt The configured belt alarm 2 has been triggered
Optimization of the control parameters
Alarm value too rigid
10 Alarm 2 - min The configured min alarm 2 has been triggered
Optimization of the control parameters
Alarm value too rigid
11 Alarm 2 - max The configured max alarm 2 has
been triggered
Optimization of the control parameters
Alarm value too rigid
12 Alarm - external The configured alarm 1 on input
1 has been triggered
Check the source of the external alarm
13 Alarm - external The configured alarm 1 on input
2 has been triggered
Check the source of the external alarm
14 Alarm - external The configured alarm 2 on input 1 has been triggered
Check the source of the external alarm
15 Alarm - external The configured alarm 2 on input 2 has been triggered
Check the source of the external alarm
16 No USB flash drive inserted
When exporting data, insert a USB flash drive into the controller
17 Import / export of
data using the USB flash drive
was not
successful
The file has been edited using a
PC (text editor) and stored in the wrong format, or the USB
flash drive was not recognized.
You want to import data that are
not located in the Import folder
on the USB flash drive
Do not edit XML files with a text editor, but
always in the controller itself.
Format the USB flash drive (format: FAT32)
Use another USB flash drive (1-8 GB)
In a complete import, all data must be stored in
the Import folder on the USB flash drive.
When importing
programs,
programs are
rejected
The temperature, time or rate
are outside the limits
Import only programs that are suitable for the
furnace. The controllers differ in the number of
the program and number of segments and in the
maximum furnace temperature.
When importing
programs, “Error
occurred” appears
The complete set of parameters
(at least the configuration files)
has not been placed in the
“Import” folder
If you deliberately omitted files during the import,
the message can be ignored. Otherwise, please
check the completeness of the import files.
18 “Heating
disabled”
If a door switch is connected to
the controller, and the door is
open, this message appears
-Close the door
-Check the door switch
Pos: 273 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Üb ersc hrift - Stö rung en der Sc halta nlag e @ 7 4\m od_ 141 110 9885 873 _51 .docx @ 3 2952 0 @ 2 @ 1
![Page 57: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/57.jpg)
57
8.3 Malfunctions of the Switchgear Pos: 274 /TD/St öru nge n/Fehle rsuch e/Co ntroll er/B4 00/B41 0/C4 40/C4 50/P4 70/P48 0/Stö run gen der Sc halta nlag e- Fehle rsuc he_ 01 Co ntr oller B... /C.../P.. . - Tab elle @ 124 \mo d_1 465 8946 248 23_ 51. docx @ 548 002 @ @ 1
Error Cause Remedy
Controller does not light
up
Controller is switched off Switch the power switch to “I”
No power available Is the power cord plugged into the socket?
Check the building fuses.
Check the fuse of the controller (if
present) and replace it if necessary.
Check the fuse of the controller (if present) and replace it if necessary.
Switch the power switch on. If the error occurs again, contact Nabertherm
Service
Controller displays error See the separate instructions of the controller See the separate instructions of the controller
Furnace does not heat Door / cover is open Close the door / cover
The door contact switch is faulty (if present) Check the door contact switch
The “wait” or clock icon (series 400 controllers) lights up
The program is waiting for the programmed start time.
Set the wait time to 00:00 or disable it
Error in entering the program Check the heating program (see the
separate instructions of the controller)
Heating element defective Have this checked by Nabertherm Service or a qualified electrician.
Very slow heating of the
heating space
The fuse(s) of the connection is/are defective.
Check the fuse(s) of the connection and replace if necessary.
Notify Nabertherm service if the new
fuse fails again immediately.
The program does not
jump to the next segment
In one TIME segment in the program input,
the wait time is set to INFINITE (series 400
controllers).
If batch control is activated, the temperature
of the batch is higher than the zone
temperatures.
Do not set the wait time to INFINITE
If batch control is activated, the temperature of the batch is higher than the zone
temperatures.
The parameter [LOWER BLOCK] must be set to [NO].
The controller module
can not be registered on
the operating unit
Addressing error (series 400 controllers) Perform a bus reset
The controller is not
heating in the
optimization
No optimization temperature has been set The temperature to be optimized must be entered (see the separate instructions of
the controller)
![Page 58: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/58.jpg)
58
Pos: 275 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Stö run gen der Schalt anlag e- Fehl ersuc he_ 02_C ontr oller B.../C... /P... - Ta belle @ 88 \mo d_1 427 2110 033 31_ 51.d ocx @ 371 824 @ @ 1
The temperature rises
faster than the controller
setting allows
The switch element of the heating unit
(semiconductor relay, thyristor or switch contactor) is defective.
Individual defective components inside a
furnace cannot be completely ruled out in
advance. That is why the controllers and the
switchgear units must be equipped with
safety facilities. For example, the furnace shuts down the heating unit in response to
error message 04 - 02 via an independent
contact element.
Have the switch element tested by a
qualified electrician and replaced as necessary.
Pos: 276 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Üb ersc hrift - Ch eckliste C ontr oller @ 7 3\mo d_1 409 823 689 769 _51. docx @ 32 562 8 @ 2 @ 1
8.4 Controller Check List Pos: 277 /TD/B etrie b_Bedi enu ng/Co ntr oller/B4 00/B4 10/C4 40/C 450/P4 70/P4 80/Ch eckliste C ont roller Rekla matio n - Q0xx -B.../C. ../P... - Tabelle @ 7 3\m od_ 1409 825 205 309 _51. docx @ 32 564 8 @ @ 1
Customer:
Furnace model:
Controller model:
Controller version (see information menu ):
Controller serial number
Furnace serial number
Error code in the display:
The following errors are dependent on external
influences:
02-05 Ambient temperature too low: <-10 °C (-50 °F)
02-04 Ambient temperature too high: > 70 °C (158 °F)
Detailed error description:
Export of the service information: Please export all the data to a USB stick using the function [EXPORT COMPLETELY] Generate a zip file using the ZIP
function integrated in Windows (compression) of the exported
folder (see the section "Importing and Exporting Data and
Parameters") and send them to your contact at Nabertherm
Service.
i
![Page 59: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/59.jpg)
59
When does this error occur? At specific point in the program or at certain times of day:
At specific temperatures:
How long has the error existed? Error is new
Error has existed for a long time
Unknown
Error frequency Error occurs frequently
Error occurs regularly
Error occurs rarely
Unknown
Substitute controller:
Has a substitute controller already been used?
yes no
Did the error continue with the substitute
controller?
yes no
Checked according to the error search list
(see the furnace operating instructions)
yes no
Please enter the following test program so that the furnace heats up at full power:
Program point Value
Segment 01- Start Temperature 0 °C
Segment 01- Target Temperature 500 °C
Segment 01- Time 30 minutes
Segment 01- Target Temperature 500 °C
Close door/lid and start the example program
Please check the following items:
Does the furnace heat (temperature rise)?
Is the "Heating" symbol displayed?
Please call up the information menu in the heating up phase for further details.
Date: _______________ Name: ______________________ Signature: ___________________
![Page 60: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/60.jpg)
60
Pos: 278 /TD/E rsatz -/Versc hleiß teile/Ü bersc hrift - Ers atz-/V erschl eißt eile @ 0\m od_ 116 4125 252 056 _51. docx @ 42 43 @ 1 @ 1
9 Spare Parts/Wearing Parts Pos: 279 /TD/E rsatz -/Versc hleiß teile/Er satz- /Versc hleißt eile all gem eine r Text -E rsatzt eile b estelle n-Ve rfüg bar keit vo n Ers atz- und V ersch @ 1 3\mo d_1 290 428 577 708 _51. docx @ 10 872 0 @ @ 1
Ordering Spare Parts:
Our Nabertherm Service team is available to you all around the world. Due to our
considerable production depth we deliver most spare parts from the warehouse overnight or
can make them ready for delivery within short deadlines. You can order Nabertherm spare
parts easily and simply directly from the factory. If you fail to find the spare part you are
looking for in the spare part list or in the separate spare part list we would be happy to help
you. Spare parts can be ordered in writing, by phone or on the Internet -> see the section
entitled "Nabertherm Service".
Availability of Spare Parts and Wearing Parts:
Although Nabertherm has many spare parts and wearing parts on stock, we cannot guaranty the short-term availability of all of them. We recommend that certain parts be ordered in
advance. If you need any assistance when selecting spare parts and wearing parts, the staff
at Nabertherm will be glad to set aside time for you.
Pos: 280 /TD/E rsatz -/Versc hleiß teile/Hi nweis - Origi nalteil e un d Zube hör sind s peziell für N abe rth erm -Ofe nanl age n ... @ 8\ mod _12 427 198 129 94_5 1.d ocx @ 569 63 @ @ 1
Note
Original parts are designed especially for Nabertherm furnaces. Replace parts only with
original Nabertherm parts. Otherwise the warranty will be void. Nabertherm accepts
absolutely no liability for damage caused by using parts that are not original Nabertherm
parts.
Pos: 281 /TD/E rsatz -/Versc hleiß teile/ Fü r de n Aus- und Einba u von Ersatz -/Ver schlei ßteile n, we nde n Sie sich an unse ren N abe rth erm -Servic e. @ 4\m od_ 120 403 4379 624 _51 .docx @ 3 1716 @ @ 1
Note
Contact our Nabertherm Service for removing and installing replacement and wear parts.
See section on "Nabertherm Service". Work on the electrical equipment may only be
performed by qualified and authorized specialist electricians. This applies also to repairs
not described below.
![Page 61: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/61.jpg)
61
Pos: 282 /TD/E rsatz -/Versc hleiß teile/Er satzt eilzeichn ung eine s LH T 01 -03 /17 D - G esam tanl age (Mo dell 2 015 ) - Gra fik @ 1 19\ mod _14 623 721 3884 2_5 1.d ocx @ 4698 49 @ @ 1
LHT 03/17 D
LHT 01/17 D
Fig. 32: Replacement parts (example drawing)
1
1
1.5
1.5
1.5
1.2; 1.4 1.3
2
5
4
6
6.1
3
7
![Page 62: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/62.jpg)
62
Pos: 283 /TD/E rsatz -/Versc hleiß teile/Er satzt eilliste - Hocht em per atur ofe n LH T 01 -03 /17 D (M odell 201 5) - Liste @ 1 19\ mod _146 236 525 847 0_5 1.doc x @ 4 697 95 @ @ 1
Model LHT 01/17 D LHT 03/17 D
No. Name Part number Quantity Part number Quantity
1 Heating element 692253495 4 692253496 4
1.2 Clamp 691800042 8 691800033 8
1.3 Element holder 691800049 8 691800048 8
1.4 Connecting contact 691800047 3 691800046 2
691800144 2 6918000144 4
1.5 Thermocouple 540300384 12) 540300554 12)
2 Exhaust air tube 692040198 1 692040198 1
3 Floor plate 601604420 3) - -
4 Door insulation 601405357 1
5 Tension spring 691400598 2 691400599 2
6 P480 801000102 1 - -
6.1 P470 - - 801000095 1
7 Semi-conductor fuse - - 541500000 1
Power switch 541800259 1 541800259 1
Power cable (xx = state country
where it will be used) V0013xx 1 V0013xx 1
Tools
Heating element tongs 493000015
Allen key 493000021
1) Fiber paper/fiber wadding (delivered with the spare parts)
2) Furnaces with over-temperature limiters have 2 thermocouples 3) Quantity as required
Pos: 284 /TD/E rsatz -/Versc hleiß teile/ Leg end e fü r Ersat zteile - was v om K unde n o der eine r Fachk raft s elbst get ausch t wer den kan n @ 1 0\m od_ 125 2578 510 636 _51 .docx @ 65 905 @ @ 1
Symbols
Can be replaced by the customer with tools and instructions.
Can be replaced by trained personnel with tools and instructions.
NT Nabertherm Service required
Pos: 285 /TD/B etrie b_Bedi enu ng/Co ntr oller/B1 50/B1 30/B17 0/C2 80/C2 90/C 295/P3 20/Sch altplä ne/Üb ersc hrift - Elekt rische Sc haltpl äne/P neu matikpl äne @ 47 \mo d_1 380 029 813 550_ 51. docx @ 24 771 8 @ 2 @ 1
9.1 Electrical Schematics/Pneumatic Schematics Pos: 286 /TD/B etrie b_Bedi enu ng/Die mitg eliefe rten Unte rlag en b einh alten nicht zwan gsläufi g elek trisch e Schalt plän e bzw. P neu ma .. . @ 4 7\m od_ 138 002 807 461 6_51 .docx @ 2 476 68 @ @ 1
Note
The documents included do not always contain the electrical schematics and pneumatic
schematics.
If you need the respective schematics they can be ordered from Nabertherm Service.
![Page 63: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/63.jpg)
63
Pos: 287 /TD/E rsatz -/Versc hleiß teile/Ü bersc hrift - Heiz ele ment aust ausch en 1.1 @ 2\ mod_ 118 657 908 787 5_5 1.docx @ 2 075 6 @ 2 @ 1
9.2 Replacing a Heating Element Pos: 288 /TD/E rsatz -/Versc hleiß teile/W arn ung - Ge fah ren dur ch el ektrisc hen St ro m - N etzstec ker zi ehe n - f ür alle Öf en @ 3\ mod_ 118 949 139 078 7_5 1.doc x @ 2 127 3 @ @ 1
Warning - Danger of Electric Shock!
Work on the electrical equipment may be done only by qualified, authorized electricians.
During work it must be ensured that the furnace and the switching equipment cannot be
activated by mistake (pull out the power plug) and that all moving parts in the furnace are
secured. Observe DGUV V3 or the corresponding national regulations in the country where
the furnace is installed. Wait until the furnace and the connected parts have cooled to room
temperature.
Pos: 289 /TD/E rsatz -/Versc hleiß teile/Vo rsicht - Besc hädi gung von B auteil en! H eizele ment aust ausch en @ 3\ mod _11 8949 163 440 0_5 1.doc x @ 2 128 6 @ @ 1
Caution - damage to components!
Heating elements are extremely sensitive to breaking. Any strain on or rotation of the
heating elements must be avoided. Failure to observe this rule will lead to the immediate
destruction of the sensitive heating elements.
Pos: 290 /TD/E rsatz -/Versc hleiß teile/W erkze uge (wir d be nötig t fü r de n Austa usch der Heizel eme nte - LH T-Of en) - 1 @ 3\ mo d_1 189 4925 236 38_ 51.d ocx @ 212 99 @ @ 1
Tools
To change the heating elements, you will need the following tools:
1 Sharp knife, 2 Heating element pliers, 3 Needle-nose pliers, 4 Allen wrench (supplied
with furnace), 5 Open-end wrench (SW 7), 6 Slotted screwdriver
Fig. 33: Tools
Pos: 291 /TD/E rsatz -/Versc hleiß teile/Abl uftr öhrc hen am LHT ../.. - Of en ( Tisch mod ell) -Dem onta ge - 2 @ 3\ mod _11 8949 359 365 3_5 1.do cx @ 2 132 5 @ @ 1
Exhaust air pipe
First slacken the screws (A) from the exhaust air pipe guard plate using the enclosed Allen
key. Lift the guard plate (B) from the exhaust air pipe and then carefully pull the exhaust air
pipe (C) out straight upwards. Store the exhaust air pipe in a safe place, as its material is
very sensitive.
Fig. 34: Removing the exhaust air pipe
Pos: 292 /TD/E rsatz -/Versc hleiß teile/ Ofen deckel LHT ../.. - Of en ( Tisch mo dell) - De mont age - 3 @ 3\ mod _11 894 937 817 94_ 51.d ocx @ 213 38 @ @ 1
Removing the furnace lid
Slacken the screws (A) at the rear and front of the lid with the enclosed Allen key, pull the
lid forwards towards the furnace door (B) and remove upwards (C).
1 2 3 4 5 6
A B C
![Page 64: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/64.jpg)
64
Fig. 35: Furnace lid
Pos: 293 /TD/E rsatz -/Versc hleiß teile/ Fed erkl amm er und B reitb andk abel am LHT.. /.. - Ofe n ( Tisch mod ell) - 4 @ 3\m od_ 118 949 4139 200 _51 .docx @ 2 1351 @ @ 1
Clamps and connection contact/connection bridge
Carefully press the clamps open with the heating element pliers and pull upwards. Loosen
the connection contact or connection bridge upwards from the heating element.
Fig. 36: Clamps and connection contact
Pos: 294 /TD/E rsatz -/Versc hleiß teile/Ele men thalt er u nd I solier ung am LHT. ./.. - Of en ( Tisch mod ell) - 5 @ 3\m od_ 118 949 430 020 0_5 1.docx @ 2 136 4 @ @ 1
Element holder and insulation
Loosen the screws from the element holder with an open-ended wrench. Pull the element
holder from the heating element. Remove the underlying fiber paper and fiber wool from
the slot between the legs of the heating elements.
Fig. 37: Removing the element holder and insulation
Pos: 295 /TD/E rsatz -/Versc hleiß teile/H eizele ment e a ustau sche n am LH T../.. - Of en (Tisc hmo dell) - 6 @ 3\ mod _11 894 944 6679 4_5 1.d ocx @ 213 77 @ @ 1
Replacing heating elements
Carefully pull heating elements out upwards; do not press heating element legs together
whilst doing so. Clean heating chamber and passage holes (vacuum out). Carefully insert
new heating elements from above. Attention: Pay attention to the correct installation
height! See sketch. An assembly support (e.g. foam) between the heating elements and the base of the furnace is helpful.
Heating element pliers
Connection contact/connection bridge
Heating element
A B
Clamp
C
![Page 65: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/65.jpg)
65
Fig. 38: Heating elements
Pos: 296 /TD/E rsatz -/Versc hleiß teile/I solier ung an den H eizele men ten aust ausch en a m LH T../ .. - Ofen ( Tischm odell ) - 7 @ 3\mo d_1 189 494 818 362 _51. docx @ 21 393 @ @ 1
Insulation
Fill the slots between the ends of the element connections with fiber wool *). Place 2 mm
fiber paper *) on lightweight refractory bricks.
Fig. 39: Fiber wool Fiber paper
Pos: 297 /TD/E rsatz -/Versc hleiß teile/ Mon tag e de r Heizel em ente am LHT. ./.. - Ofe n ( Tisch mod ell) - 8 @ 3\m od_ 118 949 491 032 1_5 1.docx @ 2 140 6 @ @ 1
Assembling the heating elements
Place element holder onto ends of heating element. Carefully tighten the hexagonal screws
(WAF 7) of the element holders; the screw heads must point outwards. Any strain on or
rotation of the heating element must be avoided. The element must be secured with pliers to
prevent rotation. Position connection contact or connection bridge onto heating element
ends and secure with new clamps, making sure these are positioned correctly to each other.
Element holders, screws, spring clamps and cables must be properly checked.
Element holder
2 mm
Assembly support
(e.g. foam)
Sketch
Heating element pliers
Connection contact
Heating element end
Clamp
![Page 66: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/66.jpg)
66
Fig. 40: Assembly Additionally secure element holder with pliers
Pos: 298 /TD/E rsatz -/Versc hleiß teile/ Ofen deckel LHT ../.. - Of en ( Tisch mo dell) - M onta ge - 9 @ 3\ mod_ 118 949 508 589 2_5 1.docx @ 2 141 9 @ @ 1
Assembling the furnace lid
Position lid (A) and push to rear wall (B). Make sure that the lid holes align with the holes
on the housing. Secure the lid to the front and rear wall of the furnace housing using the
previously slackened screws (C).
Fig. 41: Furnace lid
Pos: 299 /TD/E rsatz -/Versc hleiß teile/Abl uftr öhrc hen am LHT ../.. - Of en ( Tisch mod ell) -Mo ntag e - 10 @ 3\ mod _11 8949 529 015 9_5 1.do cx @ 2 143 2 @ @ 1
Assembling the exhaust air pipe
Carefully push the exhaust air pipe (A) into the intended opening. The head of the exhaust
air pipe must lie on the furnace lid. Reassemble the exhaust air pipe guard plate (B) using
the previously slackened screws (C).
Fig. 42: Assembling the exhaust air pipe
Pos: 300 /TD/Einl eitun g/Tech nische D ate n/Fla nsch e/Dre hm ome nte f ür Sch rau ben /Mu tter n M 4_M 5_ M6_ M7_ M8_ M1 0 - Tab elle @ 97\ mod _14 355 7347 208 2_5 1.do cx @ 3936 04 @ @ 1
Screw tightening torque
Tighten power cable clamps and screws on the heating elements with a defined torque.
If this advice is not followed, the heating elements may be damaged.
Thread diameter
Metric thread (M)
Torque in Nm
M 4 2.0
M 5 6.0
M 6 8.0
M 7 14.0
A B C
Ø F
A B C
![Page 67: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/67.jpg)
67
Screw tightening torque
M 8 20.0
M 10 39.0
Pos: 301 /TD/E rsatz -/Versc hleiß teile/N etzste cker einst ecken (sie he Kapi tel „Ans chluss an das Elekt ron etz“ ) d anac h Netzsc halte r .. . @ 3 \mo d_1 190 0145 595 88_ 51.d ocx @ 218 47 @ @ 1
Commissioning
Insert the mains power connector (see chapter "Connection to the Mains Electricity"), then
switch on the power switch and check the function of the furnace (see chapter "Operation").
Pos: 302 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Ve rgess en Sie nicht die Mont age aufla ge/ Mont age hilfe a us d em Ofen zu e ntfe rne n. @ 3\m od_ 119 0010 794 387 _51. docx @ 21 835 @ @ 1
Note
Do not forget to remove the assembly support/assembly aid from the furnace chamber.
Pos: 303 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - * ) = lieg t de r Ersa tzteillief eru ng bei. @ 3\m od_ 119 001 053 7840 _51 .docx @ 2 182 3 @ @ 1
Note
*) = Enclosed with spare part delivery.
Pos: 304 /TD/E rsatz -/Versc hleiß teile/Ü bersc hrift - The rmo ele ment aust ausch en 1.1 @ 2\ mod_ 118 657 956 924 5_5 1.doc x @ 2 076 9 @ 2 @ 1
9.3 Replacing a Thermocouple Pos: 305 /TD/E rsatz -/Versc hleiß teile/W arn ung - Ge fah ren dur ch el ektrisc hen St ro m - N etzstec ker zi ehe n - f ür alle Öf en @ 3\ mod_ 118 949 139 078 7_5 1.doc x @ 2 127 3 @ @ 1
Warning - Danger of Electric Shock!
Work on the electrical equipment may be done only by qualified, authorized electricians.
During work it must be ensured that the furnace and the switching equipment cannot be
activated by mistake (pull out the power plug) and that all moving parts in the furnace are
secured. Observe DGUV V3 or the corresponding national regulations in the country where
the furnace is installed. Wait until the furnace and the connected parts have cooled to room
temperature.
Pos: 306 /TD/E rsatz -/Versc hleiß teile/Vo rsicht - Besc hädi gung von B auteil en! The rm oele ment aust ausch en @ 3\ mod _11 8949 292 193 5_5 1.do cx @ 2 131 2 @ @ 1
Caution - damage to components!
Thermocouples are extremely sensitive to breakage. Any strain on or rotation of the
thermocouples must be avoided. Failure to observe this rule will lead to the immediate
destruction of the sensitive thermocouples.
Pos: 307 /TD/E rsatz -/Versc hleiß teile/W erkze uge (wir d be nötig t fü r de n Austa usch der Th er moel eme nte - LH T-Of en) - 1 @ 3 \mo d_1 189 5008 513 50_ 51.d ocx @ 214 45 @ @ 1
Tools
To change the thermocouple, you will need the following tools:
1 Allen wrench (supplied with furnace), 2 Slotted screwdriver, 3 Phillips screwdriver
Fig. 43: Tools
Pos: 308 /TD/E rsatz -/Versc hleiß teile/Abl uftr öhrc hen am LHT ../.. - Of en ( Tisch mod ell) -Dem onta ge - 2 @ 3\ mod _11 8949 359 365 3_5 1.do cx @ 2 132 5 @ @ 1
Exhaust air pipe
First slacken the screws (A) from the exhaust air pipe guard plate using the enclosed Allen
key. Lift the guard plate (B) from the exhaust air pipe and then carefully pull the exhaust air
pipe (C) out straight upwards. Store the exhaust air pipe in a safe place, as its material is
very sensitive.
1 2 3
![Page 68: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/68.jpg)
68
Fig. 44: Removing the exhaust air pipe
Pos: 309 /TD/E rsatz -/Versc hleiß teile/ Ofen deckel LHT ../.. - Of en ( Tisch mo dell) - De mont age - 3 @ 3\ mod _11 894 937 817 94_ 51.d ocx @ 213 38 @ @ 1
Removing the furnace lid
Slacken the screws (A) at the rear and front of the lid with the enclosed Allen key, pull the
lid forwards towards the furnace door (B) and remove upwards (C).
Fig. 45: Furnace lid
Pos: 310 /TD/E rsatz -/Versc hleiß teile/ The rm oele ment aus tausc hen am LHT. ./.. - Ofe n ( Tisch mod ell) -D em onta ge - 4 @ 3\m od_ 118 950 100 393 7_5 1.docx @ 2 145 8 @ @ 1
Replacing the thermocouple
First slacken the two screws (A) from the thermocouple connection. Slacken screw (B)
from thermocouple retaining plate and pull thermocouple out upwards. Push a new
thermocouple carefully into the thermocouple channel (C) and assemble and connect in
reverse order. Attention must be paid to the correct polarity of the electrical connections (D) *).
Fig. 46: Replacing the thermocouple
A B C
A
B
A B
C
C
D
![Page 69: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/69.jpg)
69
Pos: 311 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Die Anschlüs se d er Ve rbin dun gsleitu nge n vo m Ther moel eme nt zu m Re gler sind mit + un d - @ 3\ mod _119 012 324 124 9_5 1.doc x @ 2 191 6 @ @ 1
Note
*) The connections of the connecting lines from the thermocouple to the controller are
labeled with and . It is absolutely essential to observe the correct polarity.
to to
Pos: 312 /TD/E rsatz -/Versc hleiß teile/ Ofen deckel LHT ../.. - Of en ( Tisch mo dell) - M onta ge - 9 @ 3\ mod_ 118 949 508 589 2_5 1.docx @ 2 141 9 @ @ 1
Assembling the furnace lid
Position lid (A) and push to rear wall (B). Make sure that the lid holes align with the holes
on the housing. Secure the lid to the front and rear wall of the furnace housing using the
previously slackened screws (C).
Fig. 47: Furnace lid
Pos: 313 /TD/E rsatz -/Versc hleiß teile/Abl uftr öhrc hen am LHT ../.. - Of en ( Tisch mod ell) -Mo ntag e - 10 @ 3\ mod _11 8949 529 015 9_5 1.do cx @ 2 143 2 @ @ 1
Assembling the exhaust air pipe
Carefully push the exhaust air pipe (A) into the intended opening. The head of the exhaust
air pipe must lie on the furnace lid. Reassemble the exhaust air pipe guard plate (B) using
the previously slackened screws (C).
Fig. 48: Assembling the exhaust air pipe
Pos: 314 /TD/E rsatz -/Versc hleiß teile/N etzste cker einst ecken (sie he Kapi tel „Ans chluss an das Elekt ron etz“ ) d anac h Netzsc halte r .. . @ 3 \mo d_1 190 0145 595 88_ 51.d ocx @ 218 47 @ @ 1
Commissioning
Insert the mains power connector (see chapter "Connection to the Mains Electricity"), then switch on the power switch and check the function of the furnace (see chapter "Operation").
+ -
+ + - -
A B C
A B C
![Page 70: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/70.jpg)
70
Pos: 315 /TD/W art ung _Reinig ung _Inst and haltu ng/Üb ersc hrift - Aust ausch /Nachj ustier ung des Tü r-Is oliera ufb aus @ 3\m od_ 119 494 712 644 5_51 .docx @ 2 748 1 @ 2 @ 1
9.4 Replacement/Readjustment of the Door Insulation Structure Pos: 316 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Wa rnun g - Arbeit en a n d er Aus rüst ung dür fen nur von qualifizi erte m u nd befu gte m Fachp erso nal @ 3\m od_ 119 520 836 487 6_51 .docx @ 2 775 7 @ @ 1
Warning! General hazards!
Work on the equipment may only be performed by qualified, authorized specialists. During
work, the voltage supply to the furnace/switching system must be switched off to prevent
unintentional commissioning (disconnect mains power connector) and all moving parts of
the furnace must be secured. Observe DGUV V3 or corresponding national guidelines of
the country of use. Wait until the furnace chamber and attaching parts have cooled to room
temperature.
Pos: 317 /TD/E rsatz -/Versc hleiß teile/W erkze uge (wir d be nötig t fü r de n Austa usch des Tü r-In nen aufb aus - LH T-O fen ) - 1 @ 3 \mo d_1 195 207 420 096_ 51. docx @ 27 743 @ @ 1
Tools
The following tools are required to replace/readjust the complete door insulation structure.
1 Open-ended wrench (WAF 8)
Fig. 49: Tools Pos: 318 /TD/W art ung _Reinig ung _Inst and haltu ng/Aust ausc h/Nachj ustie run g des Tü r-Is olier aufb aus am LHT . ../LH T... D - Of en (Tisc hmo dell) @ 46 \mo d_1 377 862 177 977_ 51. docx @ 24 3870 @ @ 1
*)Inclined surface of door insulation points upwards
Fig. 50: Replacement/readjustment of the door insulation structure (figure similar)
Item Quantity Name Comment
A 1 Furnace lift-door
B 1 Door spacer plate
C 1 Complete door insulation structure
See chapter "Spare/Wearing Parts"
F 1 Door collar insulation
B
C
H*)
I
A
F
![Page 71: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/71.jpg)
71
Item Quantity Name Comment
H 1 Door insulation
I 4 Hex screw
Fig. 51: Door insulation structure
1 Swivel the furnace lift-door carefully away upwards. Loosen all 4 hexagonal screws
(I) and complete door insulation structure (C) from the door spacer plate (B). Pull
complete door insulation structure towards the furnace and remove upwards (see
sketch on left). Carefully assemble door insulation structure in reverse order. Inclined surface of door insulation points upwards.
2. Secure new complete door insulation structure to the door spacer plate with the
enclosed hexagonal screws and align. The insulation is very sensitive; pay attention
to adjacent components. Door collar insulation must lie on the furnace collar
insulation all around. If the door insulation does not lie on the furnace collar all around, the complete door insulation structure must be readjusted using the
hexagonal screws (I). Pos: 319 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - I n Deu tschla nd ist die all gem eine U nfallv erh ütun gsvo rschrif t zu beac hten . @ 3\mo d_1 193 667 801 739 _51. docx @ 26 122 @ @ 1 a
Note
In Germany, the general accident protection guidelines must be observed. The national
accident prevention regulations of the country of operation apply.
Pos: 320 /TD/E rsatz -/Versc hleiß teile/Ü bersc hrift - Sich eru ng a usta usche n 1. 1 @ 9 \mo d_1 249 628 090 492 _51. docx @ 64 803 @ 2 @ 1
9.5 Replacing a Fuse Pos: 321 /TD/E rsatz -/Versc hleiß teile/Ein bau ort der Sic her ung - In ner halb des Sc halta nlage nge häus es @ 23\ mod _13 3552 535 370 8_5 1.do cx @ 1 563 80 @ @ 1
A Fuse is located inside the switchgear cabinet. The fuse is an important component of the power supply system and protects the furnace and its components against damage and fire.
When you insert a new fuse, make sure that the fuse rating is suitable for the voltage used
by your furnace.
Pos: 322 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Warns ymb ol_Acht ung - Verw end ung eine r Siche run g @ 2 3\m od_ 133 552 628 3205 _51 .docx @ 1 564 05 @ @ 1
NOTICE
• Damage to the furnace and its components
• Use of a fuse that is NOT suitable for the respective voltage
may damage the furnace and its components and is a fire
hazard.
• Use only a suitable fuse type. Check that the fuse type has the
correct nominal current value.
Pos: 323 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Befolg en Sie das Ve rfah ren zum Aussch alten der Ofe nanl age (sieh e ... @ 23 \mo d_1 335 940 466 853_ 51. docx @ 15 648 5 @ @ 1
Carry out the procedure to switch off the furnace (see "Operation"). Then pull the power plug out of the socket. Allow the furnace to cool naturally.
Pos: 324 /TD/E rsatz -/Versc hleiß teile/D emo nta ge d er Sch altanl age nab decku ng u nd Eins etze n de r Siche run g LH T 03 /17 D T - Text - Gr afik @ 46\ mod_ 137 786 507 268 0_5 1.doc x @ 2 438 95 @ @ 1
![Page 72: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/72.jpg)
72
Removing the switchgear cover
The number and position of the screws may differ from one furnace model to the next.
Fig. 52: Removing the switchgear cover (figure similar)
Remove the screws from the rear wall and keep them in a secure place for later use.
Pos: 325 /TD/E rsatz -/Versc hleiß teile/Sich eru ng austa usch en - Versio n C Siche run gshal ter ung - Besch reib ung @ 2 3\m od_1 335 946 646 277 _51. docx @ 15 651 0 @ @ 1
Remove the fuse from the fuse holder.
Replace the defective fuse with a similar fuse.
Before you replace the fuse, make sure that it has the correct nominal current.
Fig. 53: Fuse (fuse link)
Note
The nominal current is engraved into the metal cap of the fuse or can be found imprinted
directly on the fuse.
Insert the new fuse into the fuse holder. Make sure that the fuse is pushed fully into
the holder.
Check that the power cable is not damaged. The power cable must not be damaged.
Power cables may be replaced only with similar, approved cables.
Pos: 326 /TD/E rsatz -/Versc hleiß teile/ Mon tag e de r vo rhe r gel öste n Teile e rfolgt in u mge keh rter Reihe nfol ge @ 23\ mod _13 3594 964 752 9_5 1.do cx @ 1 565 35 @ @ 1
Note
Assemble the loosened parts in the reverse sequence.
T 4 A250V T
4 A
25
0V
F44
Nominal current (Example)
![Page 73: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/73.jpg)
73
Pos: 327 /TD/E rsatz -/Versc hleiß teile/Ü bersc hrift - Is olier ung rep arie ren @ 2\ mod _11 865 797 784 69_ 51.d ocx @ 207 82 @ 2 @ 1
9.6 Repairing the Insulation Pos: 328 /TD/E rsatz -/Versc hleiß teile/Bei Sc häd en a n d er Is olier ung - LH T- und Roh röfe n @ 3\m od_ 119 0127 300 968 _51. docx @ 21 928 @ @ 1
The insulation of the furnace consists of a very high-quality refractory material. Heat expansion may cause tears in the insulation even after a few heating cycles. However, these
have no affect on the function or quality of the furnace. However, if entire "sections" of the
insulation come loose, Nabertherm Service must be notified.
Pos: 329 /TD/ Zu beh ör/Üb ersc hrift - Zube hö r (O ption ) @ 2\m od_1 184 078 674 990 _51. docx @ 19 536 @ 1 @ 1
10 Accessories (Options) Pos: 330 /TD/B etrie b_Bedi enu ng/Üb ersc hrift - Beg asun gssyste m ( Zu beh ör) @ 1 16\ mod_ 145 943 223 865 8_5 1.doc x @ 4 606 34 @ 2 @ 1
10.1 Gas Supply System (Accessories) Pos: 331 /TD/B etrie b_Bedi enu ng/Ansc hluss des Beg asu ngssyst ems Beis piel LH T 0 3/1 7 D ( Mod ell 20 15) - G rafik @ 11 9\m od_ 146 2360 120 091 _51. docx @ 46 960 4 @ @ 1
Fig. 54: Connection to the gas supply system (similar to picture)
Pos: 332 /TD/B etrie b_Bedi enu ng/Ansc hluss des Beg asu ngssyst ems -Tabelle für das B egas ungssy ste m 1 @ 116 \mo d_1 459 331 830 822_ 51. docx @ 45 975 8 @ @ 1
1 Gas Supply System 1 for simple protective gas applications (no vacuum operation)This package represents a basic version suffcient for many
applications, for operation with nonf lammable protective gasses. The
standard working tube made of ceramic delivered with the furnace can still
be used.
2 The customer is responsible for providing the exhaust system
Pos: 333 /TD/B etrie b_Bedi enu ng/Üb ersc hrift (Mitt e) - Funkti onsb eschr eibu ng @ 26\ mod _13 418 258 014 78_5 1.d ocx @ 168 695 @ @ 1 Oper ation
Function Description Pos: 334 /TD/B etrie b_Bedi enu ng/ Mit de m Beg asun gssyst em ist es möglich , nich t br enn bar e Schu tz- u nd R eaktio nsga se (Beis piel: H elium @ 38 \mo d_1 362 652 5152 00_ 51. docx @ 21 4525 @ @ 1
The gas supply system permits you to feed non-combustible protective and reactive gases
(such as: helium (He), argon (Ar), forming gas, carbon dioxide (CO2) or nitrogen (N2)) into a furnace in a quantity for a defined period.
Pos: 335 /TD/Sic her heit/Üb ersc hrift - Siche rhei t (b ) @ 26\ mod _134 182 614 650 1_5 1.doc x @ 1 687 20 @ @ 1
Safety Pos: 336 /TD/Sic her heit/Sich erh eit a m Beg asun gssyste m @ 26\ mod _13 4182 623 441 6_5 1.do cx @ 1 687 45 @ @ 1
The gas supply system must be inspected to ensure it is in perfect condition before each use. If a defect is discovered the furnace must be shut down immediately.
During operation toxic gases and vapors can be released. They must be appropriately
vented out of the building. Ignoring this warning can be dangerous to your health.
Only use gases whose characteristics are known. If something unexpected happens in the
furnace (for example, build up of dense smoke or noxious smells) it must be shut down
immediately and you must wait until the furnace has cooled down naturally.
![Page 74: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/74.jpg)
74
The use of the gas supply system in connection with combustible gases is only permissible
with additional "safety devices".
You must ensure that the installation room is well ventilated and that protective gas
emissions do not present any danger.
It is the user's responsibility to ensure compatibility with local safety and installation
regulations.
The use of the furnace for the purposes for which is was designed also requires
observing the instructions contained in this operating manual for installation, start-up
and maintenance.
You must be aware of the real or potential combustibility and explosiveness of gases when they are when operating the furnace. Make sure in particular that no corrosive or
health endangering substances are generated and escape into the surrounding air.
Operating the system is prohibited with power sources, products, operating
equipment, auxiliary materials, etc., which are listed as hazardous or which may in
any way harm the health of the personnel operating the furnace.
Make sure the hose connection is perfectly tight before each use.
The gas supply system must be checked for leaks and deposits in the flow meter (use a
leak detection spray if necessary).
The functionality of the ball valve and the solenoid valve must be checked at regular
intervals.
Pos: 337 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Bei A rbeit en mit Schutz gas en ist i mme r f ür ei ne a usr eichen de Bel üftu ng .. . @ 1 1\m od_ 126 8226 124 218 _51. docx @ 72 783 @ @ 1
Note
When working with protective gases always make sure that the room is adequately
ventilated. Country-specific safety regulations must also be observed.
Pos: 338 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Besc hrei bun g un d Funkti on sie he s epa rate Bedie nung sanlei tung @ 4\ mo d_1 1968 445 843 14_ 51.d ocx @ 284 47 @ @ 1
Note
See additional operating instructions for description and function
Pos: 339 /TD/Sic her heit/Sich erh eitssym bole/W arn hinweis e-ISO -ANSI/Piktogr am m in der Anl eitun g - W arn ung - Erstic kun gsgef ahr _2 @ 16\ mod _13 088 357 619 84_ 51.d ocx @ 120 385 @ @ 1
Warning - Danger of Suffocation
There is a danger of suffocation when process/purge or exhaust gases escape, e.g. as a
results of leaks (e.g. at doors, pipes, valves, etc.).
Due to their specific weight gases can displace oxygen. This poses a danger of suffocation.
Measures: Switch on the extraction system.
![Page 75: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/75.jpg)
75
Pos: 340 /TD/Sic her heit/Üb ersc hrift - Bet reib en vo n Dr uckgas beh älte rn @ 3\m od_ 119 3760 404 960 _51. docx @ 26 461 @ 2 @ 1
10.2 Operation of Compressed Gas Containers Pos: 341 /TD/Sic her heit/Bet reib en v on Dr uckg asbe hälte rn @ 3\m od_ 119 376 049 132 6_51 .docx @ 2 647 5 @ @ 1
Compressed gas containers may be operated only by persons familiar with their handling. Before performing their duties, employees must be respectively instructed
in the operation of the compressed gas containers,
on the special hazards of handling compressed gas containers and
on the actions to be taken in case of accidents or malfunctions. This instruction must be
repeated at appropriate intervals.
Compressed gas bottles may be stood up in work rooms only in the absolutely necessary
quantities and in the smallest possible sizes, and only for immediate use.
Storage of compressed gas bottles in work rooms is impermissible.
Gas bottles should be kept in suctioned gas bottle cabinets if possible.
When no gas is being taken, always keep the main valve on the gas bottle closed. Gas
bottles without screwed-on pressure reducers must not be stood up without protective caps. Gas hoses must be regularly checked for fractured or porous areas and if necessary
immediately replaced.
Safety measures and appropriate conduct
Secure compressed gas containers against falling over, against jarring, shock and
heating (e.g. from heater or furnace system).
At the workplace only keep ready the number of compressed gas bottles required for
continuing the work.
Transport only with bottle transport vehicles and with protective caps firmly screwed
on.
Wear suitable gloves and if necessary goggles.
When changing bottles always check leaktightness of the valves of filled and empty
bottles.
Refilling bottles is prohibited.
Do not use force to open valves.
Sufficiently ventilate areas.
Smoking and open flames are prohibited.
Keep fire extinguisher ready.
The operator must produce operating instructions stating the hazards occurring in the
work room for people and the environment as well as stating the generally required
protective measures and rules of conduct. The operating instructions must be written
in an understandable form and kept available in the work room. The operating
instructions must also contain instructions on conduct in case of danger and on first-
aid measures.
Pos: 342 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Bei A rbeit en mit Schutz gas en ist i mme r f ür ei ne a usr eichen de Bel üftu ng .. . @ 1 1\m od_ 126 8226 124 218 _51. docx @ 72 783 @ @ 1
Note
When working with protective gases always make sure that the room is adequately ventilated. Country-specific safety regulations must also be observed.
Pos: 343 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Wa rnun g - Bei uns achg em äße r In stallati on ist die Funkti on und Sich erh eit d er Anl age nicht meh r ... @ 3\ mod _11 937 592 008 13_ 51.d ocx @ 264 38 @ @ 1
Warning! General hazards!
If installed improperly, functioning and safety of the system can no longer be guaranteed. The connection must be properly installed and put into operation by qualified personnel.
![Page 76: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/76.jpg)
76
Pos: 344 /TD/B etrie b_Bedi enu ng/Üb ersc hrift - Sna p-I n-Kup plung (Steck er ) vo m Ofe nge häus e tr enn en 1.1 @ 12\ mod _12 821 138 721 92_5 1.d ocx @ 969 23 @ 2 @ 1
10.3 Separate the Snap-In Coupling (Plug) from the Furnace Housing Pos: 345 /TD/B etrie b_Bedi enu ng/Bedi enu ng Sn ap-I n-Ku pplun g (St ecker ) vo m Of eng ehä use t ren nen @ 12 \mo d_1 282 114 156 861_ 51. docx @ 96 945 @ @ 1
With a small flat blade screwdriver carefully push the locking latch (2) upward while
pulling the plug (3) out of the coupling (4).
Fig. 55: Separate the snap-in coupling (plug) from the furnace housing (similar to picture)
Pos: 346 /TD/ Zu beh ör/Üb ersc hrift - Tem per atu rwählb egr enze r @ 3\m od_ 119 021 2914 589 _51 .docx @ 21 959 @ 2 @ 1
10.4 Over-temperature Limit Controller Pos: 347 /TD/Einl eitun g/Pro dukt besch reib ung/ Öfen /Sicher heits einric htun g - Ü berw achu ng der Ofen rau mte mpe rat ur - TWB @ 15\ mod _13 049 345 965 18_ 51.d ocx @ 116 826 @ @ 1
These Nabertherm GmbH furnaces are equipped as standard with a over-temperature limiter to protect against over-temperature in the furnace chamber.
The over-temperature limiter with manual reset monitors the temperature in the furnace
chamber. The display shows the last trigger temperature that was set. When the furnace
chamber temperature exceeds the set triggering temperature, the heating is switched off to
protect the furnace or charge.
Pos: 348 /TD/B etrie b_Bedi enu ng/Bedi enu ng Tem pe ratu rwähl beg renz er (TWB) Eu rot her m 2 132i - Teil 1- Gr afik @ 12\ mod_ 128 032 291 636 2_5 1.docx @ 9 669 5 @ @ 1
Fig. 56: Over-temperature limit controller with manual reset 2132i Pos: 349 /TD/Allg em eine Hinw eise (fü r alle Anl eitun gen )/Hinweis - Besc hrei bun g un d Funkti on sie he s epa rate Bedie nung sanlei tung @ 4\ mo d_1 1968 445 843 14_ 51.d ocx @ 284 47 @ @ 1
Note
See additional operating instructions for description and function
2132i
![Page 77: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/77.jpg)
77
Pos: 350 /TD/B etrie b_Bedi enu ng/Co ntr oller/B1 50/B1 30/B17 0/C2 80/C2 90/C 295/P3 20/Sch altplä ne/Üb ersc hrift - Elekt rische Sc haltpl äne/P neu matikpl äne @ 47 \mo d_1 380 029 813 550_ 51. docx @ 24 771 8 @ 2 @ 1
10.5 Electrical Schematics/Pneumatic Schematics Pos: 351 /TD/B etrie b_Bedi enu ng/Die mitg eliefe rten Unte rlag en b einh alten nicht zwan gsläufi g elek trisch e Schalt plän e bzw. P neu ma .. . @ 4 7\m od_ 138 002 807 461 6_51 .docx @ 2 476 68 @ @ 1
Note
The documents included do not always contain the electrical schematics and pneumatic
schematics.
If you need the respective schematics they can be ordered from Nabertherm Service.
Pos: 352 /TD/S ervice -Ty pensc hild/Na ber the rm -Service @ 0\ mo d_11 586 589 159 97_ 51.d ocx @ 286 1 @ 1 @ 1
11 Nabertherm Service
Contact Nabertherm Service at any time for maintenance and repair.
If you have any questions, problems, or requirements, contact Nabertherm GmbH. By mail,
phone or e-mail.
Nabertherm GmbH
Bahnhofstrasse 20
28865 Lilienthal/Germany
Phone or Fax
Phone: +49 (4298) 922-0
Fax: +49 (4298) 922-129
Web or E-mail
www.nabertherm.com
When you contact us, please have the type plate details of the furnace or controller at
hand.
Pos: 353 /TD/S ervice -Ty pensc hild/ Type nschild gekle bt_ 108 x 84 mit CE @ 43\ mod _13 711 189 6760 3_5 1.d ocx @ 233 580 @ @ 1
Provide the following details from the type plate:
Furnace model
Serial number
Article number
Year of construction
Fig. 57: Example (type plate)
1
2
3
4
![Page 78: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/78.jpg)
78
Pos: 354 /TD/A uße rbe trie bna hme _Dem ont age_ Lag eru ng_R ücktr ansp ort/Ü be rschrif t - Au ße rbet rieb nah me, D em onta ge u nd Lag eru ng @ 0\m od_ 116 783 8295 210 _51 .docx @ 5 580 @ 1 @ 1
12 Shut-Down, Dismantling, and Storage Pos: 355 /TD/A uße rbe trie bna hme _Dem ont age_ Lag eru ng_R ücktr ansp ort/A uße rbe trie bna hme ... vo m Bet reib er z u er gänz en - 1 @ 0\ mod _11 5867 794 776 6_5 1.do cx @ 2 921 @ @ 1
To be Completed by the Operator
When the furnace is shut down, the following safety information must be observed to
prevent serious injury, damage to property, and damage to the environment.
The furnace may only be shut down by authorized, trained personnel.
The following operating materials/parts are to be disposed of by:
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
Oils and other substances that are hazardous to water must be completely removed before the furnace is dismantled for reuse or scrap.
Ensure that operating materials, lubricants, and consumables are disposed of in an
environmentally compliant manner. Regulations relating to proper waste recycling and
disposal must be observed.
The furnace may be lifted only at the intended points.
Use only the stated lifting and securing equipment to lift the furnace/parts.
Consider a total weight of ___________kg when choosing suitable lifting equipment.
For transportation, consider a permitted floor weight of at least ___________kg/m².
Before transporting the furnace, attach the following securing equipment:
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
_________________________________________________________________
Note
Read the sections on "Safety" and "Transportation"
Pos: 356 /TD/A uße rbe trie bna hme _Dem ont age_ Lag eru ng_R ücktr ansp ort/Hi nweise für de n Rückt rans por t de r Of en-Anl age @ 4\ mod _12 024 590 538 63_ 51.d ocx @ 309 18 @ 2 @ 1
12.1 Transportation/Return Transportation
If you still have the original packaging, this is the safest way to send a furnace.
Otherwise:
Choose suitable, adequately sturdy packaging. During transportation, packages are often
stacked, bumped, or dropped; the packaging acts as external protection for your furnace.
+45 °
-5 °C
- Drain all piping and containers before transportation/return transportation
(e.g. cooling water). Pump off operating materials and dispose of properly.
- Do not subject the furnace to extreme cold or hot temperatures (direct sunlight).
Storage temperature -5 °C to 45 ° (-23 °F to 113 °F)
Humidity 5 % to 80 %, non condensing
- Place the furnace on a level floor to prevent distortion.
- Packaging and transportation may be carried out only by qualified and
authorized persons
If your furnace has transportation securing equipment (see "Transportation Securing
Equipment"), use this.
![Page 79: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/79.jpg)
79
Otherwise, in general:
"Fix" and "secure" (adhesive tape) all moving parts and cushion and protect any projecting
parts against breakage.
Protect your electronic equipment against moisture and make sure that no loose packaging
material can get inside it.
Fill gaps in your packaging with soft but adequately firm material (e.g. foam mats) and
make sure that the equipment cannot slide around in the packaging.
If the goods are damaged during return transportation due to inadequate packaging
or some other breach of duty, the costs will be borne by the customer.
As a rule:
The furnace is sent without accessories, unless the technician expressly requests them.
Enclose a detailed description of the malfunction along with the furnace – this saves the
technician time and costs.
Don't forget to enclose the name and phone number of a contact in case there are any
questions.
Note
Return transportation may only be carried out according to the information given on the
packaging or in the transportation documents.
Note
Transportation and return transportation not covered by a warranty claim are paid for by
the customer.
Pos: 357 /Steu er modul e/= == Seit enu mb ruch == = @ 0\m od_ 1158 819 844 943 _0.d ocx @ 298 3 @ @ 1
![Page 80: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/80.jpg)
80
Pos: 358 /TD/A nha ng/Üb ersc hrift - Konf or mitäts erklä run g @ 2 \mo d_1 186 581 095 030_ 51. docx @ 20 795 @ 1 @ 1
13 Declaration of Conformity Pos: 359 /EG-Erkl äru nge n/Kopf mod ul/Konf or mitäts erklä run g M aschin enric htlinie @ 10 \mo d_1 260 3621 997 08_ 51. docx @ 696 75 @ @ 1
EC Declaration of Conformity
Compliant with EC Directive 2006/42/EC on machinery, Annex II A
We,
Nabertherm GmbH
Bahnhofstr. 20, 28865 Lilienthal, Germany
Pos: 360 /EG-Erkl äru nge n/Pro dukta nga ben /dass folg ende s Pro dukt: @ 10 \mo d_1 260 343 5294 74_ 51. docx @ 693 24 @ @ 1
hereby declare that the following product: Pos: 361 /EG-Erkl äru nge n/Pro dukta nga ben /Hocht em per atur öfen LH T 01/ 17 D - LH T 0 3/17 D @ 1 18\ mod _14 618 466 2350 3_5 1.d ocx @ 467 769 @ @ 1
High-Temperature Furnaces
for the industrial use
Model LHT 01/17 D LHT 03/17 D
Pos: 362 /EG-Erkl äru nge n/Erklä run gen/ allen einschl ägig en Besti mm unge n d er Rich tlinie 2006 /42/EG ents prich t. @ 10\ mod _12 6035 738 425 7_5 1.doc x @ 6 952 7 @ @ 1
fulfills all the pertinent provisions contained in Directive 2006/42/EC. Pos: 363 /EG-Erkl äru nge n/Mit gelte nde Ric htlinie n/Das bezeic hne te Pr odukt ents prich t weit erhin allen Bestim mun gen de r Richtlini en: @ 10\ mo d_12 603 435 932 04_ 51.d ocx @ 693 40 @ @ 1
The product named is also compliant with all the provisions of the following directives: Pos: 364 /EG-Erkl äru nge n/Mit gelte nde Ric htlinie n/EG-Ni ede rspa nnu ngsric htlinie N r. 2 014/ 35/EU @ 11 6\m od_1 459 500 325 444 _51. docx @ 46 079 8 @ @ 1
Directive 2014/35/EU for electrical equipment designed for use within certain voltage limits Pos: 365 /EG-Erkl äru nge n/Mit gelte nde Ric htlinie n/EG-Ric htlinie Elek tro mag netisc he Ve rträ glichkeit Nr. 201 4/30 /EU @ 1 16\m od_ 145 950 036 987 3_51 .docx @ 4 608 25 @ @ 1
Directive 2014/30/EU on electromagnetic compatibility Pos: 366 /EG-Erkl äru nge n/Erklä run gen/D oku ment ation sver antwo rtlich e @ 1 0\m od_ 126 035 5006 988 _51 .docx @ 69 493 @ @ 1
The signatories are authorized to compile the relevant technical documents. The address is the
stated manufacturer's address. Pos: 367 /EG-Erkl äru nge n/Erklä run gen/Ä nde rung des Prod uktes @ 10 \mo d_1 260 355 011 435_ 51. docx @ 69 510 @ @ 1
Any change in the product not approved by the manufacturer invalidates this declaration. Pos: 368 .1 /EG -Erklä rung en/ Mitgel tend e No rme n/Folge nde har mo nisiert e No rme n wu rden an gewa ndt: @ 10 \mo d_1 260 3600 417 78_ 51.d ocx @ 695 44 @ @ 1
The following harmonized standards were applied: Pos: 368 .2 /EG -Erklä rung en/ Mitgel tend e No rme n/No rme n/DIN EN 6 101 0-1 @ 2 0\m od_ 1322 820 198 856 _0.d ocx @ 143 610 @ @ 1
DIN EN 61010-1 (07.2011) Pos: 368 .3 /EG -Erklä rung en/ Mitgel tend e No rme n/No rme n/DIN EN 6 100 0-6 -1, D IN EN 61 000 -6- 3 @ 32\ mod _135 159 567 282 5_0. docx @ 19 056 8 @ @ 1
DIN EN 61000-6-1 (10.2007), DIN EN 61000-6-3 (09.2011) Pos: 369 /EG-Erkl äru nge n/Or t un d Dat um @ 0\ mod _115 501 851 365 1_0. docx @ 10 93 @ @ 1
Lilienthal, 01.04.2016 Pos: 370 /EG-Erkl äru nge n/Unt ersch riftko mbi natio nen/U nte rschrif ten Obe rsch midt/Ad amek mit U nte rschrif t @ 65\m od_ 140 714 649 754 2_51 .docx @ 3 090 30 @ @ 1
________________________ _______________________
Michael Oberschmidt Thomas Adamek
Vice President R & D Quality Management === E nde der Liste für Tex tma rke I nhal t = ==
![Page 81: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/81.jpg)
81
![Page 82: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/82.jpg)
82
![Page 83: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/83.jpg)
83
![Page 84: Operating Instructions - LABOSHOPdata.laboshop.com/fileadmin/media/ga/43004_ga_enu.pdf · Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages](https://reader035.vdocuments.us/reader035/viewer/2022071107/5fe183d67a87b70501657e44/html5/thumbnails/84.jpg)
Headquarters: Reg: M01.0068 ENGLISCH
Nabertherm GmbH . Bahnhofstr. 20 . 28865 Lilienthal/Bremen, Germany . Tel +49 (4298) 922-0, Fax -129 . [email protected] . www.nabertherm.com