operating instruction manual model 2200c conductivity … · 4.4.1 two spdt (single pole double...

30
OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY ANALYZER REV. 7.0 AquaMetrix Inc. 22-121 Granton Drive Richmond Hill, ON Canada, L4B 3N4 Tel: (800) 742-1413 (905) 763-8432 Fax: (905) 763-9480 www.aquametrix.com

Upload: others

Post on 29-May-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

OPERATING INSTRUCTION MANUAL

MODEL 2200CCONDUCTIVITY ANALYZER

REV. 7.0

AquaMetrix Inc.22-121 Granton DriveRichmond Hill, ONCanada, L4B 3N4

Tel: (800) 742-1413(905) 763-8432

Fax: (905) 763-9480www.aquametrix.com

Page 2: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

TABLE OF CONTENTS

SECTION DESCRIPTION PAGE

1 GENERAL INFORMATION 1

2 SPECIFICATIONS 2

3 INSTALLATION 3

4 DESCRIPTION OF FUNCTIONS 6

5 START-UP AND OPERATION 11

6 OPERATING HINTS 16

7 UTILITY MENU 17

8 DIAGNOSTICS 21

9 TROUBLESHOOTING AND SERVICE 22

Page 3: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 1

MODEL 2200C CONDUCTIVITY CONTROLLERINSTRUCTION MANUAL

1.0 GENERAL INFORMATION

The AquaMetrix Model 2200C conductivity analyzer is a versatile industrial microprocessor basedinstrument. Setpoints and outputs are programmed through the menu with push buttons on the faceof the instrument. Calibration is achieved from the front panel menu. The instrument may be usedin conjunction with any AquaMetrix MS series cell.

The instrument provides non-isolated 0-5 Vdc, 0-1 mA and isolated 4-20 mA analog outputs. Theanalog output may be programmed to represent any segment of the measuring scale. One controlrelay, with fully adjustable deadband, and one high-low alarm relay are provided. The control relaycan be programmed to activate in response to rising or falling conductivity.

The integrity of the system is ensured with a watch-dog timer and system alarm. A passwordfeature protects the stored values.

The instrument is housed in a NEMA 4X enclosure. The standard unit is provided with mountinghardware for surface mount applications. Both panel and pipe mount kits can be orderedseparately.

Page 4: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 2

2.0 SPECIFICATIONS

DISPLAY:4 digit LED, 1/2" high digits

MEASURING RANGES:Eighteen ranges;The range is set at the factory as ordered, but maybechanged in the field to any other range through theUtility Menu.Temperature: 0°C to 100°C (32°F to 212°F)

POWER REQUIREMENTS:Standard: 98-132 Vac, 50/60 Hz (less than 5 VA)Optional: 196-264 Vac, 50/60 Hz (less than 5 VA)

23-26Vdc (nominal current 150mA)

AMBIENT CONDITIONS:-30 to 50°C (-22 to 122°F)0 to 90% R.H. non-condensing

CONTROL RELAY:5A 115/230 Vac, 5A 30 Vdc SPDTSelectable to activate in response to rising or fallingprocessFully adjustable setpoint and deadbandManual Override Auto / Off / OnCycling Feature: On time 0-600sec Off time 0-600secNormal or Fail Safe operation

ALARM RELAY:5A 115/230 Vac, 5A 30 Vdc SPDTProgrammable to activate at high and low alarmsetpointsDeadband fixed at 2% of Full Scale

Normal or Fail Safe Operation

RELAY INDICATORS:Three individual LED indicate status of the controlrelay and the alarm relay.

ANALOG OUTPUTS:Non Isolated 0-1mA, 100 ohms maximum loadNon Isolated 0-5Vdc, 1000 ohms minimum loadIsolated 4-20mA, 800 ohms maximum loadRange Expand: The 4-20 mA analog output can bemade to represent any segment of the measuringscale. Minimum segment is 10% of full scale.Output Hold: The analog outputs are automaticallyplaced on hold during calibration or other setupoperations.Temperature Output: The 0-5 Vdc output can beprogrammed to follow the process temperature orconductivity.

TEMPERATURE COMPENSATION:Automatic 0°C to 100°C (32°F to 212°F)

SYSTEM ERROR:A LED indicates a system error. The alarm relay canbe programmed to activate upon system memory loss.

DIAGNOSTICS:Invalid entries are identified by respective flashingLEDs. When a system error is indicated, use the CALLbutton to display STATUS, to determine the statuscode.

TEST:Display value and analog outputs can be set manuallyto any value for testing and diagnostic purposes. Thisfeature allows the operation of the control relay, alarmrelay, analog outputs to be tested independently of theprocess.

SAFETY AND SECURITY:Non-volatile memory (EPROM)Password protected if enabledWatch-dog timer monitors microprocessorInstrument automatically returns to on line operation ifaccidentally left in menu mode. (This feature may befield disabled if desired.)

SENSITIVITY: 0.1% of span

STABILITY: 0.1% of span per 24 hrs. non-cumulative.

NON-LINEARITY: 0.1% of span

REPEATABILITY: 0.1% of span or better

TEMPERATURE DRIFT:Zero: 0.01% of span per °CSpan: 0.01% of span per °C

RESPONSE TIME:0.5 second to 90% of value upon step change

ENCLOSURE:NEMA 4X molded fiberglass reinforced polyesterenclosure with four 1/2" conduit holes and mountingfeet for surface mount. A NEMA 4 plug is provided forone hole.

MOUNTING CONFIGURATIONS:Standard is surface mountOptional panel mount hardware Part No. C35-68Optional pipe mount hardware Part No. C35-69

NET WEIGHT: 3 1/2 lb. (1.6 kg)

Approvals: CSA

Page 5: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 3

3.0 INSTALLIATION

3.1 Location

3.1.1 Locate the instrument within 3000 feet of where the AquaMetrix MS or MCconductivity cell is installed.

3.1.2 Select an installation site which is:

• free of mechanical vibrations• reasonably clean and dry• protected from falling corrosive fluids within the ambient temperature and humidity

specifications• remote from high voltage relay and power switches

3.2 Type of Mounting

3.2.1 If the instrument is to be pipe or panel mounted a special hardware kit will berequired. For panel mount order part number C35-68. For pipe mount order partnumber C35-69. Instructions for both types of mounting will be included with thekits.

3.2.2 For surface mounting, four feet brackets, together with fastening screws, areprovided with the instrument. These should be fastened to the back of theinstrument and then it may be screwed or bolted in the selected location.

3.3 Conduit Connections

3.3.1 Four 1/2" conduit holes are provided in the bottom of the enclosure. One of these isfitted with an approved water tight plug. To maintain NEMA 4 integrity, approvedconduit hubs must be used to connect conduit. The hubs must be connected to theconduit before being connected to the enclosure. Any unused conduit holes mustbe closed with water tight plugs or connectors.

3.3.2 For convenience of internal connections the right conduit hole (viewed from thefront) should be used for power connection; the next hole to the left for relayoutputs; the next hole to the left for analog outputs and finally the fourth hole forsensor input.

Page 6: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 4

3.4 Electrical Connections

CAUTION: The instrument operates from line voltage. This constitutes a possible shock hazard.Ensure that line power is removed before attempting connections. Note: A separatesource of line power may have been connected to the floating relay contacts.

3.4.1 To access the terminal strips open the door of the instrument and then unscrew thecaptive retaining screw near the upper right hand corner of the panel. Now swingopen the panel to reveal the terminal strip on the power supply circuit board and thesmaller terminal strip on the back of the main circuit board.

3.4.2 The terminal strip on the power supply board at the back of the instrument islabeled for line power, relay outputs and analog outputs. Connect the wiring inaccordance with this labeling. (Refer to Fig.1)

Note: Control Relay B is not used in the 2200C.

Terminals 4 and 5 are wired directly to the input line voltage. These terminals canbe used to supply power to the relays. Use the common ground of Terminal 1as theground for the auxiliary devices. A separate source of power can also be used toprovide power to the relays. Ensure that the circuit breaker supplying the instrumentis able to supply enough current to the auxiliary devices, to ensure proper operationof the instrument.

Page 7: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 5

CAUTION: Connecting the line voltage to incorrect terminals may cause serious damage.

3.5 Cell Connections

3.5.1 Connect the 4 wires of the cell cable to the TB2 terminal strip on the main circuitboard being sure to match the colors as printed on the TB2 terminal strip. Whiteand black are the electrode wires; red and green are the temperature sensor wires.Place switches S41 and S42 in the down “on-line” position. Refer to Fig.2

Page 8: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 6

3.5.2 The test resistor R68, is used for the simulation feature. It has a resistance equal tothe equivalent resistance of the cell at mid scale. The resistor that is shipped withthe unit corresponds to the mid scale of the range ordered. If the range is changed,from the range configured at the factory. Refer to Sec. 9.1.5 for the correct resistorvalue.

4.0 Description of Functions

Page 9: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 7

4.1 Overview

4.1.1 The Model 2200C is a microprocessor based conductivity analyzer designed forindustrial applications. It is compatible with AquaMetrix’s MS and MC conductivitycells and offers several measurement ranges. The unit's software makesmaintenance and operation easy.

4.1.2 The outputs include voltage-free relay contacts and industry standard analogtransmission signals. Two programmable relays are provided for process controland alarm. The status of each relay is indicated on the front by a LED. The analogoutput signals transmit low power signals to peripherals such as data recorders orcontrol systems.

4.1.3 The software is designed for ease of operation. It uses a simple menu with all itemsindicated on the panel. The user interface consists of six buttons on the front panel.The buttons are scanned and responded continuously. In addition, an extensivesystem checking for values and parameters is performed by the software. All of theoperating parameters are stored in non-volatile memory, without the need of abattery.

4.1.4 The Model 2200C operates like a normal analog converter with a number ofadditional functions made possible by the microprocessor in the instrument.

Some of these are:• Recall and easy adjustment of relay and output parameters• Push button calibration• A HOLD function for outputs• Continuous sensor check during measurement• Continuous self check and watch-dog timer to ensure correct operation• Password protection of stored values• Temperature Output

Page 10: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 8

4.2 Calibration

4.2.1 All conductivity systems need to be calibrated when first placed in service andthereafter from time to time. The frequency of calibration can only be found by theoperator’s experience with each process. Calibration must always be performedwhen a new sensor is placed in service.

4.2.2 Calibration is accomplished by using buffer solutions, of known conductivity, andadjusting the instrument to show the known value. Buffers are available in 500 mLbottles and in 20 L packs from AquaMetrix. The procedure for performingcalibration is very simple and is given in Sections 5.2.

4.3 Temperature Compensation

4.3.1 Temperature compensation is an essential component of conductivitymeasuring instruments. Specific conductivity increases with temperature eventhough there has been no addition of dissolved solids. It is therefore the industrystandard to compensate for temperature change as if the temperature wereconstant at 25°C. The 2200C compensates for a 2% increase in specificconductivity per °C.

4.4 Relay Outputs

4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally opencontacts, NO, are open when no alarm or control is active.

4.4.2 The control relay can be programmed to close on either rising or falling conductivity.It is normally used to control a valve, pump or auxiliary device. It may be set toclose at any point on the scale. The deadband, sometimes termed hysteresis,defines the point at which the relays open.

4.4.3 Refer to Sections 5.4, to 5.6.

4.4.4 The second relay functions as a high-low alarm. It has two programmableSetpoints (high and low). The deadband is fixed at 2% of full span. The alarmcontact can be programmed to also signal memory loss in the controller. SeeSections 4.7, 5.7, and 5.8.

4.4.5 All relays can be programmed for "fail-safe" operation which reverses the normaloperation of the relay. In fail safe mode the Normally Closed contact functions asthe Normally Open contact. In the event of power loss to the unit, the relay will de-energize and the device connected to the Normally closed contact will be turned on.This option is selected with the DIP switches on the swing out board. See Section4.7.

Page 11: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 9

4.5 Analog Outputs

4.5.1 The analog output signals consist of a non-isolated 0-1 mA, 0-5 Vdc, and isolated 4-20 mA signals.

4.5.2 From the factory all of the analog outputs have a linear range corresponding to the

full range of the instrument. The analog outputs can be programmed to anotherlinear range by entering two values:

• Output High: This is the conductivity value at which you wish to have 100%output.

• Output Low: This is the conductivity value at which you wish to have 0% output.

4.5.3 When programming, you must ensure that the output range has a span of at least10% of the full range of the instrument. For applications that require the output todecrease as the process value increases, i.e. an inverted output, the Output Highvalue will be less than the Output Low. See Section 5.3.

4.5.4 The instrument may be used to measure the temperature of the process either in °Cor in °F. The 2200C does not control temperature but the 0-5 Vdc and 0-1 mAanalog outputs can be dedicated to follow the process temperature. See Section5.10.

4.6 Operation Menu

4.6.1 The operation menu allows the user to recall and to adjust the parameters, requiredby the analyzer functions.

4.6.2 When the analyzer is powered up, the program will display the conductivity reading.None of the LED’s in the operation's menu will be illuminated.

4.6.3 Five buttons on the panel are used to operate the menu. Use the CALL button tostep through the items in the menu. The red LED beside each item makes it veryeasy to follow the menu. The function of the RUN button is to return to the on-lineconductivity display from anywhere in the menu. The analyzer has a built-in timer,which returns the unit to the online mode if no button has been pressed for 10minutes. This time-out has the same effect as pressing the RUN button.

4.6.4 When in the menu mode, the display will shows the current value of the parameter,such as the Setpoint of the control relay. The two arrow buttons are used to adjustthe display value up or down. To accept the new value press ENTER twice. Whilein the operations menu, the relay outputs and the analog outputs remain on hold.

Page 12: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 10

The items that appear in the operations menu are:

• Password• Test• Calibration• Temperature• Status• Cell Constant• Range• Output High• Output Low• Cycle On / Off• Control Relay Setpoint• Control Relay Deadband• Alarm Relay High• Alarm Relay Low

4.7 DIP Switches

4.7.1 The controls which are frequently used in the normal operation of the instrument areall accessible on the control panel. Some switches, which are infrequently used,are located on the back of the main circuit board.

4.7.2 The DIP switches are scanned only on power-up and every time the instrument istaken out of the menu mode. Therefore, after changes to the DIP switch settings,you must bring up the menu by pressing CALL, and then return the instrument tothe online mode by pressing RUN, in order for the instrument to scan the new DIPvalues.

4.7.3 The following table describes the use of the 16 DIP switches:

Bank S1Dip Switch Description ON OFF1 Selects temperature unit °C °F2 Password Feature Disabled Enabled3 Fail-Safe mode for Relay A Enabled Disabled4 Direction of Control for Relay A Falling Rising5 Alarm Relay to be activated for memory loss Disabled Enabled6 Fail-Safe mode for Alarm Relay Enabled Disabled7 Auto return from menu Enabled Disabled8 0-5V / 0-1mA Select Conductivity Temperature

Bank S2Dip Switch Description ON OFF1 Cycling Control for Relay A Enabled Disabled2 Reserved3 Reserved4 Reserved5 Reserved

Page 13: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 11

6 Reserved7 Reserved8 Reserved

Note: Items in “BOLD” are the factory settings that the unit will be shipped with.

4.8 Output Hold

4.8.1 Output hold, is a function which freezes all output signals at the last value toprevent the occurrence of wild distortions during programming and maintenance.

4.8.2 When the Operations Menu is entered by pressing CALL, the relays and the analogoutputs are automatically placed on hold and remain on hold until the instrumentreturns to on line. The output hold will remain for a maximum of 10 minutes afterthe last button was pressed, if this feature has been enabled. See Section 4.7.

4.9 Parameter and Operation Checking

4.9.1 The instrument continuously checks all parameters in its memory, while measuringthe conductivity. When it detects an invalid value, it flashes the LED in theoperation menu to indicate the parameter that is at fault. You must then access theoperations menu to take corrective actions.

4.10 Simulated Input for Testing

4.10.1 TEST is function that can be used to check the setup of the 2200C. When in thetest mode the display will show a value between 0 and the max range the unit hasbeen configured for. The arrow keys are used to change the display. The relays andanalog outputs will react to the changes of the display as if it were monitoring theconductivity. This feature enables the user to verify that the controller has beenconfigured correctly, and that it will correctly react to the process. Of course, therelays and outputs are no longer on hold once the arrow buttons are used in TESTmode.

4.11 Control Relay AUTO/OFF/ON Switch

4.11.1 The AUTO/OFF/ON button is used to set the operating mode of Relay A. Thisuseful feature allows the operator to check the operation of the device controlled bythe relay. On power up the control relay always returns to the AUTO setting. Besure to return to AUTO after completing your test if you wish to work automatically.

4.12 Utility Menu

4.12.1 The Utility Menu is provided to enable authorized personnel to change the rangeand to fine tune the analog outputs. Refer to Section 8.0.

4.13 Watchdog Timer and Self Diagnostics

4.13.1 The 2200C continuously monitors the condition of all key components of themeasuring system to ensure that the measurements are reliable. Invalid entries andmemory loss are indicated on the panel. Refer to Section 8.0.

Page 14: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 12

5.0 START-UP AND OPERATION

5.1 Password

5.1.1 To enter the menu press CALL and the PASSWORD LED will illuminate. If thepassword feature has been disabled by the dip switch, TEST will be the first menuitem illuminated when the CALL button is pressed. With each press of CALL buttonyou will step through the menu. When the last item, ALARM LOW is reached themenu wraps around to TEST. If you have enabled PASSWORD by placing DIPSwitch No. 2 of Bank S1 in the off position you must enter the password “6” whenPASSWORD is indicated, if you wish to change any stored value. If the passwordis enabled and you do not enter the correct password. All the items in the menu willbe read only.

5.2 Calibration

5.2.1 To calibrate the instrument you will need:

• a buffer solutions of known conductivity• a beaker of clean water

The 2200C requires a one point calibration, using a conductivity buffer solution. It ispossible to calibrate the unit to the process conductivity. This can only be donewhen a sample is taken and measured with a certified lab unit. The process ofcalibrating is the same for both methods. If you are calibrating using the samplemethod, substitute the certified lab units reading for the know conductivity, andleave the cell installed in the process. The recommended calibration procedure iswith a certified lab meter.

To calibrate the 2200C in Manual Calibration Mode proceed as follows:

a) Enter the menu by pressing CALL. Proceed by pressing CALL until theCALIBRATION LED illuminates.

b) Remove the protective cap from the cell, rinse it in clean water, and place it inthe buffer solution. The display will change as the controller measures theconductivity and the temperature of the buffer solution. The display may flashwhich indicates that the instrument is receiving an unstable signal from the cell.The display will come to a rest and should display a value close to the buffersolution. If the value on the display is greatly different from the value of thebuffer solution, this indicates that the current calibration is not very close to theactual value, or the buffer solution is contaminated. If you are confident in yourbuffer solution proceed by using the arrow keys to correct the display to readthe actual value of the buffer. Press ENTER. The display will flash untilENTER is pressed again to confirm the value.

Page 15: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 13

c) The system is now calibrated. Press RUN to place the instrument on line oruse the CALL key to proceed to other menu items.

5.3 Analog Output Range Expand

5.3.1 To increase the resolution of the analog outputs, the range may spread over anysection of the scale. As long as it spans at least 10% of full scale. The best way todescribe this setup is by example. Suppose you wish the outputs to span from400uS to 600uS. Proceed as follows:

a) Enter the menu by pressing CALL until the OUTPUT HIGH LED is illuminated.Use the arrows keys to make the display read 600. Press ENTER. The displaywill flash until ENTER is pressed again to confirm the value.

b) Press CALL and the OUTPUT LOW LED will illuminate. Use the arrow keys tomake the display read 400. Press ENTER. The display will flash until ENTER ispressed again to confirm the value.

c) The analog output will now be at 0% when the conductivity of the solution is400 µS and will increase to 100% when the conductivity of the solution is 600µS.

d) Press RUN to place the instrument on line or press CALL for another menuselection.

NOTE: A range expand of less than 10% of full scale is an invalid entry which willbe indicated by the LED flashing when you return to RUN. To correct,return to the menu and correct the output settings.

5.4 Control Relay Setpoint

5.5.1 The Control Relays are configured by programming the Setpoint and Deadband.

•Setpoint: is the value at which the relay will energize•Deadband: is the value at which the relay will de-energize

As shipped from the factory, Control Relay, is configured to control risingconductivity. However, you may change the direction of the relay by changing theposition of the DIP switch. (Refer to Section 4.7)

5.5.2 The relay setpoint may be set at any point on the scale.To program the setpoint. Proceed as follows:

a) Enter the menu by pressing CALL, proceed by pressing CALL until the RELAYSETPOINT LED is illuminated. Now use the arrow keys to make the displayread the desired value. Press ENTER. The display will flash until ENTER ispressed again to confirm the value.

Page 16: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 14

b) Press RUN, to place the instrument on line, or press CALL for another menuselection.

5.5 Control Relay Deadband

5.5.1 If you have configured the relay to control decreasing conductivity, the deadbandwill be higher than the setpoint. i.e. the auxiliary device connected to the relay willincrease the conductivity. If you have configured the instrument to controlincreasing conductivity, the deadband will be lower than the setpoint. i.e. theauxiliary device connected to the relay will decrease the conductivity.

a) Enter the menu by pressing CALL, proceed by pressing CALL until the RELAY

DEADBAND LED is illuminated. Now use the arrow keys to make the displayread the desired value. Press ENTER. The display will flash until ENTER ispressed again to confirm the value.

b) Press RUN, to place the instrument on line, or press CALL for another menuselection.

5.6 Cycle Feature for Control Relay

5.6.1 The 2200C has a cycling feature for the control relay. This feature if enabled willcause the relays to cycle based on the on and off times selected. This usefulfeature can help eliminate overshooting, saving expensive chemicals.

a) If you wish to have cycling control for the Control Relay , you must put DIPswitch NO. 1 of Bank S2 in “ON” position. (See Section 4.7). If the dip switch isin the “OFF” position, the Relay will act in a simple ON-OFF manner.

b) The following example describes how to enter the ON and OFF times for theControl Relay cycling control.

i) Enter the menu by pressing Call, proceed by pressing CALL until theCYCLE LED illuminates.

ii) With the arrow buttons make the display read the number of seconds(maximum 600) you wish the relay to be activated. Example 30. PressENTER. The display will flash until ENTER is pressed again to confirmentry.

iii) Press CALL. The LED will flash. Now make the display read thenumber of seconds (maximum 600) you wish the relay to bedeactivated. Example 60. Press ENTER. Press ENTER again asabove.

Page 17: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 15

iv) With this configuration when the process reaches the Control RelaySetpoint the relay will cycle continuously 30 seconds activated, 60seconds deactivated until the Deadband Relay A setting is passed.

NOTE: As shipped from the factory the OFF time is set at 10 seconds and theON time is set at 5 seconds.

5.7 Alarm High

5.7.1 The instrument is fitted with a relay, which is set to activate on both high and lowalarm conditions. The deadband is factory set. To set the ALARM HIGH proceedas follows.

a) Press CALL to enter the menu, proceed by pressing CALL until the ALARMHIGH LED is illuminated. Use the arrow keys make the display read the desiredvalue. Press ENTER. The display will flash until ENTER is pressed again toconfirm the value.

b) Press RUN to place the instrument on line, or press CALL, for another menuselection.

5.8 Alarm Low

5.8.1 To set the ALARM LOW proceed as follows:

a) Press CALL to enter the menu, proceed by pressing CALL until the ALARMLOW LED is illuminated. With the arrows make the display read the desiredvalue. Press ENTER. The display will flash until ENTER is pressed again toconfirm entry.

b) Press RUN to place the instrument on line, or press CALL, for another menuselection.

NOTE: A low alarm point higher than the high alarm point is an invalid entry, whichis indicated by the LED flashing when you return to RUN. To correct, returnto the menu and reset the alarm points.

5.9 Test

5.9.1 To check the setup of the 2200C, the TEST feature can be used. The test featurewill allow the user to use the arrow keys to change the display. By changing thedisplay value the analog outputs and relay outputs will react, as if the process wasactually changing. This feature enables the user to simulate the process and test isthe auxiliary devices connected to the relays are functioning properly.

5.9.2 To use the TEST feature proceed as follows:

Page 18: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 16

a) To enter the menu press CALL, the TEST LED will be illuminated. Use thearrow keys to change the display. The outputs will react to the changes inthe display, as if the unit was online and the process was actually changing.

b) Press RUN to place the instrument on line, or press CALL, for anothermenu selection.

5.10 Temperature

5.10.1 The temperature of the process can be read at any time by entering the menu andscrolling through the menu until TEMPERATURE LED is illuminated. Either °C or°F will be indicated depending on the position of DIP switch No.1 of Bank S1. If thetemperature is suspected of being incorrect, place the S41 switch on the back of themain circuit board, in the up position, (this places a 3000 ohm across the RD andGN terminals) the temperature should read 25°C or 77°F.

5.10.2 The 0-5 Vdc and 0-1 mA analog outputs can be dedicated to follow the processtemperature by simply placing DIP switch No. 8 of Bank S1. in the OFF position.The temperature span of the output is set to the utility menu. Refer to Section 7.4.

5.11 Status

5.11.1 The 2200C continuously checks the integrity of all stored data and monitors thecondition of the measuring system. If a fault is detected, the FAIL LED above thedisplay will turn red. The STATUS in the operation menu will provide a numericalcode, giving a possible cause and a suggested remedy.

CODE POSSIBLE CAUSE SUGGESTED REMEDY0 Normal Condition No action required1 Reserved2 Temperature Sensor off Scale Check for open or short

connections of temperature sensorinput (Terminals RD & GN on TB2)

3 Conductivity reading off Scale Verify process. Check for open orshort connections. WH and BLwires

4* Memory Loss Call your AquaMetrix Rep. OrAquaMetrix directly

5 Calibration out of limit Check scale setting (See rangechange in Utility Menu)

NOTE: Code 4 could be a serious failure so the alarm relay will activate in addition to the redillumination of the FAIL LED if Dip Switch No. 5 of Bank S1 is off.

Page 19: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 17

5.12 Range

5.12.1 This menu item is read only, it displays the maximum range of the instrument ascurrently configured. You may change the range in the Utility Menu. Refer toSection 7.3.

5.13 Cell Constant

5.13.1 This menu item is read only, it displays the cell constant required for the range, as itis currently configured.

6.0 OPERATING HINTS

6.1 Cell Care

6.1.1 Keep the cell clean using the procedure recommended in the cell manual. Thefrequency of cleaning depends on the solution being monitored.

6.1.2 Be sure the cell cable is well protected. The cell cable should run in conduit butnever in the same conduit with line power. Excess cable should be cut off andnever coiled.

6.2 Calibration

6.2.1 The system will only be as accurate as your calibration technique. With a clean cellyou will need to calibrate only with a change in conditions or after installing a newcell.

6.2.2 Be sure of the value of your calibration solution. Conductivity solutions areavailable from AquaMetrix.

6.3 Reset

6.3.1 The instrument can be reset without losing calibration or any of the stored values bypressing and releasing the S9 reset button (located at the bottom of the board nearthe center) This action is equivalent to turning the power off and on.

6.4 Ground Loop Errors

6.4.1 If there is a conductive pathway from the sensor connections to earth, ground looperrors will occur. To avoid this condition keep all terminal connections in theinstrument or in a junction box dry and free of corrosion.

6.5 Output Hold

Page 20: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 18

6.5.1 It may be useful during some system maintenance procedures to place the relayand analog outputs on hold. To accomplish this simply press CALL. To return toon line operation press RUN. (If the instrument is in TEST the outputs will not be inHOLD if the arrow buttons are touched.)

NOTE: To safeguard against the operator forgetting to press RUN the instrumentwill automatically go back on line ten minutes after the last button waspressed provided this feature has been enabled by placing DIP switch No.7in the ON position.

7.0 UTILITY MENU

7.1 Utility Menu Functions

7.1.1 The Utility Menu enables authorized personnel to perform the following:

• Change the range• Adjust the temperature output range• Adjust the offset and span of the 0-1 mA/ 0-5 Vdc output• Adjust the offset and span of the 4-20 mA output

7.2 Access to Utility Menu

7.2.1 The Utility Menu is protected by password. To access the Utility Menu press andhold both RUN and ENTER for five seconds until the PASSWORD LED illuminates.The RUN LED will flash to warn that Utility Menu is in use. Now with the arrow keysmake the display read the password, "7". Press ENTER.

7.2.2 The Utility Menu is entirely separate from the Operations Menu but uses the sameLED display. Press CALL to step through the menu items.

Below is the cross-reference between menu items:

Operations Menu Utility MenuRange Range ChangeOutput High Temperature Output 100% (5Vdc)Output Low Temperature Output 0% (0Vdc)Control Relay Setpoint 0-1mA / 0-5Vdc Adjust lowControl Relay Deadband 0-1mA / 0-5Vdc Adjust highAlarm High 4-20 mA Output Adjust Low

Page 21: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 19

Alarm Low 4-20 mA Output Adjust High

7.3 Range Selection

7.3.1 The table below shows the ranges which are available for the 2200C together withthe relevant cell constant for each range.

Range uS Required Cell Constant0-2uS 0.01

0-5 uS 0.010-10 uS 0.050-20 uS 0.050-50 uS 0.05

0-100 uS 0.050-200 uS 0.50-500 uS 0.5

0-1000 uS 0.50-2000 uS 1.00-5000 uS 1.0

Range mS Required Cell Constant0-10 mS 100-20 mS 100-50 mS 10

0-100 mS 200-500 mS 50

0-1000 mS 50

Page 22: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 20

7.3.2 To change the range of your instrument to any range shown above proceed asdescribed below. (Note that changing the range may require a change ofconductivity cell.)

a) Enter the Utility Menu as described in 7.2.1.

b) Press CALL until the RANGE LED illuminates, which is “Range Change” in theUtility Menu. See table in 7.2.2.

c) The display will now show the range as currently configured. Use the arrowbuttons until the display shows the desired range. Press ENTER. The displaywill flash until ENTER is pressed again to confirm entry.

d) Return to the Operation Menu and press CALL until the CELL CONSTANT LEDilluminates. The display will show the cell constant required by the new range.You can also check the new range by calling RANGE. If the required cellconstant is other than the one used with the previous range it will be necessaryto purchase a cell with the correct constant.

e) Recalibrate. Refer to Section 5.2.

f) Press RUN to return to on line or press CALL to proceed to another item in theUtility Menu.

7.4 Temperature Output

7.4.1 The 0-5 Vdc and 0-1 mA outputs may be programmed to track the temperature ofthe process. The factory outputs have a linear range corresponding to 0°C to100°C, (or 32°F to 212°F.) Suppose you wish the output to span 10°C to 45°C.Proceed as follows:

a) Enter the Utility Menu as described in 7.2.1.

b) Press CALL until the OUTPUT HIGH LED illuminates, which is “Temperatureoutput, 100% point” in the Utility Menu (See table in 7.2.2). Use the arrow keysto make the display read 45.0. Press ENTER. The display will flash untilENTER is pressed again to confirm entry.

c) Press CALL and the OUTPUT LOW LED will illuminate, which is “Temperatureoutput, 0% point” in the Utility Menu (See table in 7.2.2). Use the arrow keys toshow 10.0 on the display. Press ENTER twice as above.

d) Press RUN to return to on line or press CALL to proceed to another item in theUtility Menu.

7.5 Adjust 0-1 mA / 0-5 Vdc Output

7.5.1 It may be desirable to fine tune the 0-1 mA / 0-5 Vdc output to take into account thecharacteristics of your particular loop. The following method involves a high andlow calibration and requires the use of a digital multi-meter. (DMM). Proceed asfollows ignoring the instrument display:

Page 23: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 21

a) Turn off the power to the instrument. Connect your DVM in parallel with the 0-5Vdc output terminals on the power supply board.

b) Turn on the power. Enter the Utility Menu as described in 7.2.1.

c) Press CALL until the CONTROL RELAY SETPOINT LED illuminates, which is"0-1 mA/ 0-5 Vdc output adjust, low" in the Utility Menu (See table in 7.2.2).Use the arrow keys to make your DMM read 1.25V. Press ENTER twice toconfirm.

d) Press CALL and the CONTROL RELAY DEADBAND LED will illuminate, whichis "0-1 mA / 0-5 Vdc Output adjust, high" in the Utility Menu (See table in 7.2.2.)Use the arrow keys to make your DMM read 3.75V. Press ENTER twice toconfirm.

e) Press RUN to return to on line or press CALL to proceed to another item in theUtility Menu.

7.6 Adjust 4-20 mA Output

7.6.1 It may be desirable to fine tune the 4-20 mA isolated output to take into account thecharacteristics of your particular loop. Before deciding that the adjustment isnecessary, be aware of the "Output High" and "Output Low" settings you may haveprogrammed as described in Section 5.4. The following method involves a high andlow calibration and requires the use of a Digital Multimeter (DMM). Proceed asfollows ignoring the instrument display.

a) Turn off the power to the instrument. Connect your DVM in series with the 4-20mA output terminals on the power supply board.

b) Turn on the power. Enter the Utility Menu as described in 7.2.1.

c) Press CALL until the ALARM HIGH LED illuminates, which is "4-20 mA Outputadjust, low" in the Utility Menu (See table in 7.2.2.) Now use the arrow keys tomake your DMM read 8 mA. Press ENTER twice to confirm.

d) Press CALL and the ALARM LOW LED illuminates, which is "4-20 mA Outputadjust, high" in the Utility Menu (See table in 7.2.2). Now use the arrow keys tomake your DMM read 16 mA. Press ENTER twice to confirm.

e) Press RUN to return to on line or press MENU to proceed to another item in theUtility Menu.

Page 24: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 22

8.0 DIAGNOSTICS

8.1 Description

8.1.1 The 2200C has diagnostic features which alerts the operator to invalid entries andmemory loss. Invalid entries are indicated by the flashing of the appropriate menuLED. The flashing will commence after RUN is pressed and will continue until theerrors are corrected. Memory loss is indicated by the flashing of TEST and by thealarm relay if enabled by DIP Switch No. 5 of Bank S1. See Section 4.7.3.

8.2 Invalid Calibration

8.2.1 Invalid calibration will be indicated when the input is out of range. This could occurif you had attempted to calibrate with a solution which was outside the range of theinstrument.

8.3 Invalid Output

8.3.1 Invalid output will be indicated if the expanded range is less than 10% of full scale.To correct, refer to Section 5.3.

8.4 Invalid Alarm Points

Page 25: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 23

8.4.1 Invalid Alarm will be indicated if the Low Alarm is set higher than the High Alarm.To correct, refer to Section 5.6 and 5.7.

8.5 Invalid Setpoint

8.5.1 The deadband setting must be lower than the Setpoint if you are controlling risingconductivity, and higher than the Setpoint if you are controlling falling conductivity.If you have not made these settings correctly, invalid entry will be indicated by aflashing LED. To correct, refer to Section 5.4 and 5.5.

9.0 TROUBLESHOOTING AND SERVICE

9.1 Isolate the cause

9.1.1 When a measurement problem occurs, the first step is to try to isolate the cause. Ifthe 2200C is powered, go through the menu and check your settings. A convenientway to do this is to call TEST. See Section 5.7

9.1.2 If your 2200C appears dead or intermittent, check the breaker, make sure that theinstrument is set up for the available line voltage and make sure the line voltage isactually available at the terminals. Now measure that sufficient voltage is availableat all times; it should be 98 Vac to 132 Vac or 187 Vac to 243 Vac respectively.Shut line power off, making sure it is off. CAUTION: Power to the relays may besupplied from a separate source, shut it off too. Check and if necessary replace theinternal 0.25A fuse. Push the connector of the ribbon cable firmly into its socket. Ifthese steps do not solve the problem it may be necessary to replace the powersupply board. See Section 9.3.1 below.

9.1.3 Should your process control element e.g. pump, valve or alarm not be activating,check that the correct power is supplied to the "floating” relay contacts, jumped over

Page 26: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 24

from the line voltage terminals or perhaps from a separate breaker or DC supply.Use the TEST menu feature to simulate the input values to observe the switchingpoints.

Check that your process control elements can function independently of the 2200C.Next check with a test light or an Ohmmeter (power off to the relay contacts, wiresto the control elements disconnected) that the contacts of the respective relay aremaking contact when actuated. If a relay is malfunctioning, it may be necessary tochange it. See Section below.

9.1.4 If the process value seems wrong, clean the cell as described in the cell manual.Inspect the cell, wire and interconnections. Calibrate and resume operation.

9.1.5 To find out whether the problem is in the sensor, or in the analyzer, use the self-testing features. Leave the instrument in RUN mode and proceed as follows:

a) Move the slide switches S41 and S42 on the back of the swing-out panel fromposition "ON LINE" to position "TEST."

b) Connect a test resistor R68 to the TB3 terminal strip on the swing-outboard according to the following table:

Range uS Required Cell Constant R68 Value0-2uS 0.01 10,000

0-5 uS 0.01 40000-10 uS 0.05 100000-20 uS 0.05 50000-50 uS 0.05 2000

0-100 uS 0.05 10000-200 uS 0.5 5000

Page 27: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 25

0-500 uS 0.5 20000-1000 uS 0.5 10000-2000 uS 1.0 10000-5000 uS 1.0 400

Range mS Required Cell Constant R68 Value0-10 mS 10 20000-20 mS 10 10000-50 mS 10 400

0-500 mS 50 2000-1000 mS 50 100

c) Set the temperature simulation DIP switch No. 1, S43 ON and Switch No. 2OFF. This is to simulate 25°C. Set the conductivity simulation DIP switch No. 1,S44 ON and switch No. 2 OFF. The display should indicate mid-range ±5%.(For example, for range 5000 µS, cell constant is 1 and R68 = 400 Ohmsdisplay should be 2500 µS ±5%.)

d) Now set the conductivity simulation DIP switch No. 1, S44 OFF and switch No.2 ON. The display should show full range ±5%.

NOTE: Do not use this mode for calibration if both c) and d) are satisfied theanalyzer is in order and the problem is with the probe.

9.2 Escape

9.2.1 If the instrument appears to be "DEAD", for example not responding to the buttons,or not performing on line measurement and control, always try the reset featurefirst, as described in 6.3.1.

9.2.2 a) The "ESCAPE" procedure is to be used normally at the factory only, when theunit is powered with a new processor. As a result, the internal non-volatilememory (EPROM) is "FORMATTED" and the factory values are loaded into it.

b) The "ESCAPE" procedure is to be used if a unit is serviced for a new processor,or in case of a memory loss problem, when so advised by AquaMetrix servicesupport.

9.2.3 Before performing the "ESCAPE" procedure it is important to know that thisprocedure provides the option to erase all programmed values and replace them

Page 28: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 26

with the factory set default values. This means that the range selection and theconductivity calibration must be performed after an "ESCAPE" procedure. Afterthat, all the control, alarm and scaling parameters mentioned above must be set tothe user values.

9.2.4 To perform the ""ESCAPE"" proceed as follows.

a) Turn off the power.

b) Press and hold the RUN button for about 3 seconds, while turning on t hepower. The FAIL LED will turn red and status code 7 will be obtained to showthat the "ESCAPE" procedure was just performed.

9.2.5 After the "ESCAPE" procedure it is necessary to do the following:

a) Tune the analog outputs and the temperature output span to suit your particularapplication and loop. See Sections 7.4, 7.5, and 7.6.

b) Calibrate the system, as described in Section 5.2.

c) Set up the user values for:

• Output High and Low - See Section 5.3.• Control Relay - See Sections 5.4 and 5.5.• Alarm Relay - See Sections 5.7 and 5.8.

d) After all the above operations are performed the FAIL LED will turn"green” and status "0" should be obtained.

e) The unit may be tested now, using the TEST menu item (See Section5.8.), or the built-in self-testing feature, (See Section 9.1.5.)

e) Bring the unit on line for measurement and control.

9.3 Printed Circuit Board Replacement

9.3.1 a) To replace printed circuit boards, shut off all power to the 2200C, including anyindependent power to the relay contacts. Make a record of the external wiring,then disconnect the wires. Unplug the ribbon cable connector.

b) The power supply circuit board is fastened to the back of the enclosure by fourscrews, remove the screws and the board is free. Reverse the procedure tomount a replacement board.

c) The microprocessor circuit board is located on the swing-out assembly behindthe door. Swing the assembly out, lift it up to unseat the lower hinge pin (theupper hinge pin is spring loaded.) The assembly is now free. The circuit board

Page 29: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 27

is fastened to the front panel by five screws. Remove the screws to release thecircuit board.

9.4 Customer Service

9.4.1 If a problem has not been resolved with the above procedures, a telephoneconsultation with your AquaMetrix representative, or directly with AquaMetrix willprovide the answer.

AquaMetrix Inc. Tel: (800) 742-141322-121 Granton Drive (905) 763-8432Richmond Hill, Ontario Fax: (905) 763-9480L4B 3N4 Canada Email: [email protected]

9.6 Parts and Accessories

9.6.1 The major parts are listed below. When ordering parts please use the completepart number.

Description Part #

Fuse, 0.25A Quantity of 5 A35-72Power Supply Circuit Board Assembly, 120 Vac C13-103APower Supply Circuit Board Assembly, 240 Vac C13-103BMicroprocessor Circuit Board Assembly C13-2200C

(Includes front plate and hinge.)500 ml Conductivity Solutions, Part No. varies with value

9.7 Instrument Return

9.7.1 If you are returning the instrument for service, please call an obtain an “ReturnMaterial Authorization” (RMA) number. The box must be clearly marked with theRMA number. Be sure to pack the instruments adequately because AquaMetrix willnot be responsible for shipping damage. For safety reasons, AquaMetrix cannotaccept instruments and sensors for repair that have not been thoroughly cleaned ofprocess materials.

Page 30: OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY … · 4.4.1 Two SPDT (Single Pole Double Throw) relays are provided. The normally open contacts, NO, are open when no alarm

Model 2200C Page 28

STATEMENTS OF CONFORMITY FROM THE MANUFACTURER

U.S.A.

WARNING: This equipment generates, uses, and canradiate radio frequency energy and if not installed andused in accordance with the instructions manual, maycause interference to radio communications. It has beentested and found to comply with the limits for a Class Acomputing device pursuant to Subpart J of Part 15 of FCCRules, which are designed to provide reasonableprotection against such interference when operated in acommercial environment.

Canada

This digital apparatus does not exceed the Class A limitsfor radio noise emissions from digital apparatus set out inthe radio interference regulations of the CanadianDepartment of Communications.

Le presént appareil numérique n’émet pas de bruitsradioélectriques dépassant les limites applicables auxappareils numériques de la classe A prescrites dans leRéglement sur le brouillage radioélectrique édicté par leministére des Communications du Canada.