ofpdf en 20 010 7285i 05 47053 0 · unlike the synoptophore, it offers not only diagnostic insights...
TRANSCRIPT
ZEISS Technology A Collection of Studies and white papers.
// DISPENSING TOOLS & INSTRUMENTS MADE BY ZEISS
1 29.08.2013 04:49:00
2
Technology by ZEISS 3
// 1 Exam and Refraction
Most Strongly Recommended:Polatest Visual Testing Instrument Designed by Haase
Friedrich Lorch, M.D. Article published in the Magazine for Practical Ophthalmology (Zeitschrift für praktische Augenheilkunde) 13:399-400 (1992) 4
The Evolution of Refraction: i.Profi ler® plus with i.Scription® technology.
White paper by Darryl Meister, ABOM Carl Zeiss Vision and Larry Thibos, PhD Indiana University (2010) 6
i.Scription® Clinical Study
Report on the clinical study performed by Carmel Mountain Vision Care, published in the Review of Optometry (2013). 14
// 2 Lens Fitting and Consultation
Standard Spectacle Lens Fitting
Article written by Darryl Meister, ABOM Carl Zeiss Vision (2010) 18
// 3 Technical data
i.Polatest® 22
i.Profi ler® plus 22
i.Terminal® 2 23
2 29.08.2013 04:49:00
3
The advances in technology, data transfer and communication channels characterizing
our world today have transformed our daily lives, and the field of eyecare is no
exception.
ZEISS has found a way to make these technological advances accessible to eye care
professionals and their patients through state-of-the-art instruments, such as the
i.Polatest®, i.Profiler® plus and i.Terminal® 2.
Not only do these instruments deliver precise results to optimize your patient´s visual
performance, but they also allow you to experience faster and more reliable data
management throughout your practice.
ZEISS dispensing tools have been the subject of many different tests and studies,
which have yielded very good results on their innovative qualities and reliability. This is
a collection of articles and white papers published in a wide range of media, aimed at
providing more in-depth technical information on ZEISS instruments.
Discover how instruments such as i.Polatest®, i.Profiler® plus with i.Scription technology
along with i.Terminal® 2 deliver on our promise of excellence and superior value for
you and your patients.
As one of the pioneers in optical technology, ZEISS offers innovative systems for measuring and fitting in today’s fields of eyecare. Besides providing solutions for your practice, ZEISS instruments support you in delivering a fully remarkable consultation experience for your patient.
Technology by ZEISS
i.Polatest®i.Profiler® plus i.Terminal® 2
3 29.08.2013 04:49:03
4
// 1 Exam and RefractionMost Strongly Recommended:Polatest Visual testing Instrument Designed by HaaseFriedrich Lorch, M.D. Article published in the Magazine for Practical Ophthalmology (Zeitschrift für praktische Augenheilkunde) 13:399-400 (1992)
In his letter to the editor “Normosensory
late onset strabismus” (ZPA 13: 145-
146/1992) Professor Lang attacks the
Polatest lobby and calls it “blatant
therapeutic wishful thinking”. This
accusation prompted me to write my
own letter to the editor reporting the two
years of experience that I gained from
an unbiased exposure to the Polatest
Visual Testing Instrument manufactured
by ZEISS according to the correction
recommendations of H.J. Haase.
Early in 1990 I received an invitation to
the Congress of the International Union
for Binocular Full Correction (IVBV) in
Stuttgart. The subject of the conference,
dyslexia, interested me. Because I was
unfamiliar with some of the vocabulary,
I was unable to understand much of the
material that was reported. But the results
of the heterophoria corrections were so
impressive that I most certainly wanted to
become more involved with the method.
After all, in my practice I saw at least
3-4 patients every day for whom I had
the feeling that I could not explain their
asthenopic symptoms, let alone resolve
them. It was with a sense of failure that I
had abandoned the training therapy that
had been possible in a few appropriate
cases which I had undertaken in the fi rst
years of my practice.
As I studied the literature on the correction
method according to H.J. Haase, it became
clear to me that I would be needing some
introductory seminars on binocular full
correction. As a doctor predominantly
oriented towards natural science, in the
fall of 1990 I started with binocular full
correction using the Polatest Visual Testing
Instrument without any therapeutic wishful
thinking and was thoroughly skeptical. My
assistants were my fi rst “test subjects”.
As one of them had complained of
daily headaches since childhood which
could not be explained by any medical
discipline as an experiment I prescribed
the prismatic correction I had established,.
Only 2 weeks after she received her glasses,
the headaches had disappeared, much to
our surprise. But they always came back
whenever, as an experiment, the glasses
were not used.
With this as motivation, I started to
prescribe full prismatic corrections for
patients with asthenopic symptoms with
increasing frequency. The numerous
successes were so encouraging that I gladly
accepted the considerable extra workload.
Today I can no longer imagine my daily
practice without the Polatest Visual Test
Instrument.
I have recorded the statistics of my full
prismatic correction “cases”. Here are the
most important results:
cm/m (n = 166)
0 - 0.75 36%
1 - 2.25 40%
2.5 - 4.75 21%
5 - 7.25 3%
Table 2: Distribution of vertical phorias
Of 198 patients:
73.3% were vertical esophorias
19.1% were vertical exophorias
3.5% were vertical phorias
3.5% were esophorias
0.6% were exophorias
Table 3: Distribution of vertical phorias
cm/m (n = 166)
0 - 1.75 24%
2 - 4.75 28%
5 - 4.75 21%
10 - 14.75 10%
15 - 19.75 4%
20 - 24.75 4%
25 - 29.75 3%
30 - 39.75 1%
gr. 40 5%
Table 1: Distribution of horizontal phorias
The large portion of phorias over 10 cm/m
are noticeable (Tables 1 and 2). Frequently,
these are patients that show only a slight
deviation or none at all, with the classic
methods such as an alternating cover test.
4 29.08.2013 04:49:04
5
Furthermore, work with the Polatest Visual
Testing Instrument shows that also small
heterophorias that cannot be detected
using the classical methods often cause
signifi cant problems. Very few exophorias
are to be found using the Polatest Visual
Testing Instrument (Table 3). In at least
21% of the heterophorias corrected in
the Polatest, the Maddox or alternating
color test would have lead to the wrong
direction. This does not even include those
cases in which these tests would not have
provided any conclusion.
Measured against many therapies in
medicine, the success rate of prismatic full
correction is very gratifying (Table 4).
Although so many positive things are
reported on the work with binocular full
correction according to H.J. Haase, it must
also be said that it is a very time-consuming
method, when applied correctly: A
binocular measurement and correction
phase, which lasts on the average at
least 45-60 minutes, follows the exact
monocular refraction. At present I perform
full prismatic corrections on 3-4 patients
every day; this cannot be integrated into
an already excessively long practice day
without considerable additional overtime.
In addition, the fi nancial losses are also
considerable. Nonetheless, I consider
H.J. Haase’s Polatest procedure to be
so revealing that it is worth the effort.
Unlike the synoptophore, it offers not
only diagnostic insights but also provides
valuable correction values at the same
time. For the optometrist that likes to get
involved with problems of binocular sight
himself and does not want to delegate
these cases to his orthopist, work on
the Polatest is always interesting. When
forced to participate in intensive case
history discussion, it becomes obvious
that many patients suffer considerably
from binocular problems. Thus the good
results of binocular full correction are all
the more satisfying and compensate for
the considerable workload and fi nancial
losses, at least in theory.
In the last 12 years of my practice I have
never had so much enjoyment from
my work as I have since I started using
“Polatesting”. To all my colleagues who
are self-critical and willing to take on a
new way of thinking, to jump beyond
their shadows and prescribe prisms in
high values as well, to perform refractions
– especially using a trial frame – and
are willing to make considerable time
and fi nancial sacrifi ces, I most strongly
recommend working with the Polatest
Visual Testing Instrument.
However, a basic, theoretical and practical
training course, as offered in seminars by
the IVBV and others, is indispensible. Such
training courses are also recommended
for later. The successes will be self-
evident despite all skeptical reservations.
“Therapeutic wishful thinking” is not
necessary. This is needed only by the
therapeutically unsuccessful.
Dr.med.Dipl.-Biochem. Friedrich Lorch:
“Wärmstens zu empfehlen:
Polatest-Sehprüfgerät nach Haase”
Zeitschrift für praktische Augenheilkunde
13:399-400 (1992)
Symptoms
n disappeared
% still seldom
% better
% unchanged
% worse
%
Visual problems up close 60% (81) 73 17 5 5 0
Forehead and periorbital headaches 48% (64) 73 16 6 5 0
Visual problems at a distance 22% (29) 72 17 4 7 0
Other headaches 18% (24) 54 33 8 5 0
Blinking, burning, foreign body sensation 10% (13) 62 23 8 8 0
Night driving problems 8% (10) 60 20 0 20 0
Near/far adjustment 5% (7) 28 57 0 15 0
Tears with visual strain 4% (6) 83 0 17 0 0
Neck pains 2% (3) 100 0 0 0 0
Sensitivity to glare during the day 2% (3) 33 67 0 0 0
Sensitivity to glare at night 0,7% (1) 100 0 0 0 0
Reduced amplitude of accommodation 0.7% (1) 100 0 0 0 0
Dizziness 0.7% (1) 100 0 0 0 0
Table 4: Symptoms of 134 patients undergoing full prismatic correction and changes in symptoms 8 weeks later
5 29.08.2013 04:49:05
6
Traditionally, the fi nal or manifest
refraction that serves as the basis for an
eyeglass prescription is the result of a two-
part process: The refractive errors of the
eye are fi rst estimated objectively, using
either retinoscopy or an autorefractor,
and then the prescription is subjectively
refi ned by comparing the vision of the
patient through trial lenses, using either
a refractor head or trial frame. Many
of the tools and techniques commonly
used during refraction procedures have
remained largely unchanged for over a
century, since the pioneering work of
Donders, Jackson, Copeland, and others.
Even today, with the widespread use
of sophisticated autorefractors and
photorefractors, manual subjective
refraction is still considered the “gold
standard” by clinicians. Nevertheless,
traditional subjective refraction techniques
suffer from certain inherent limitations. In
particular, subjective refraction procedures
often attempt to simulate ideal viewing
conditions by using a brightly-illuminated,
high-contrast visual acuity chart under
normal room lighting, which causes the
pupil of the eye to remain relatively small
in diameter. Because a small pupil size
restricts focusing to the central “paraxial”
region of the eye, the infl uence of ocular
aberrations upon vision is decreased,
while variation in subjective responses
from the patient due to the depth of
focus of the eye is increased.
// 1 Exam and RefractionThe Evolution of Refraction: i.Profi ler® plus with i.Scription® technology.White paper by Darryl Meister, ABOM Carl Zeiss Vision and Larry Thibos, PhD, Indiana University (2010)
Now, with the recent advent of
commercially-available aberrometers,
which measure the “wavefront”
aberrations of the eye, new enabling
technologies may fi nally elevate the
standard of refractive eyecare for
the fi rst time in decades. Evaluating
the wavefront aberrations of an
optical system has already become
commonplace in high-performance
optical fi elds such as astronomy.
There has been increasing interest in
the ophthalmic applications of this
technology, driven by advances in laser
refractive surgery.
Figure 1.
i.Profi ler® plus by ZEISS is
a “3-in-1” instrument
that incorporates the
functionality of an
aberrometer, corneal
topographer, and
autorefractor in a fast,
compact, and easy-to-use
system.
Refractive surgeons can now reduce
more of the wavefront aberrations of
the eye, in addition to the traditional
spherical and cylindrical refractive errors,
in an effort to achieve supernormal vision
with better than average visual acuity.
Although the “high-order” wavefront
aberrations of the eye cannot be
eliminated with a spectacle lens, without
introducing even greater aberrations
when the eye rotates behind the lens,
instruments such as i.Profi ler® plus by ZEISS
have made possible a new generation
of vision correction solutions that
should yield improved clinical outcomes
(Figure 1).
Aberrometry is elevating the standard of refractive eyecare by allowing clinicians to evaluate the optical characteristics of the eye more thoroughly, includi ng both low-order and high-order wavefront aberrations.
6 29.08.2013 04:49:06
7
OBJE
CT P
OINT
AT I
NFIN
ITY
IDEALWAVEFRONT
SINGLEPOINTFOCUS
ABERRATEDWAVEFRONT
ERROR
SPREADOF
FOCUS
OBJE
CT P
OINT
AT I
NFIN
ITY
B
A
Wavefront Aberrations
The propagation of “light” has been
described as a rapid movement of energy
particles – or photons – that travel in
a wave-like manner. The propagation
of light from an object point can be
represented conceptually using either
rays or waves emanating outward from
the light source. Just as rays of light
diverge from an object point, waves
of light spread out like ripples of water
traveling away from a stone that has
been dropped into a pond. At any given
distance from the original object point,
a wavefront exists that represents the
envelope bounding waves of light that
have traveled an equal distance from the
object.
As the distance from the object
increases, the curvature of these
wavefronts becomes progressively
fl atter, eventually appearing fl at beyond
“optical infi nity” (6 meters). The object
point serves as the common center
of curvature of these wavefronts.
Conversely, light converging to a point
focus can be described using spherical
wavefronts that become progressively
smaller, converging to the image point.
Further, a ray of light from the same
object point remains perpendicular to
the corresponding wavefront as both
propagate away from the object or
toward the image (Figure 2).
An aberration is essentially an error
in focus. There are several ways to
characterize the aberrations produced by
a lens or optical system. In geometrical
optics, ray tracing is often utilized to
calculate the path of a bundle of rays
from an object point as the rays are
refracted at the various surfaces of each
lens or optical element. Aberrations
are then determined by calculating the
distance of these refracted rays from
the intended focal point. Alternatively,
the deformation of the corresponding
wavefront of light as it passes through
the optical system may be also
determined.
In a perfect optical system, wavefronts
of light from an object point should
converge to a single point focus at
the desired image location, such as
the retina of the eye, after refraction
through the system. In the presence of
focusing errors or aberrations, however,
these wavefronts become either too
steep, too fl at, or distorted from their
ideal shape. Accordingly, the rays of light
corresponding to these wavefronts are
spread out at the plane of the desired
focus, instead of intersecting at a single,
sharp point. Consequently, optical
aberrations may be represented using
rays, wavefronts, or even the spread
of light intensity at the image plane
(Figure 3).
At any point across the aperture of
the optical system, such as the pupil
of the eye, the wavefront error is the
effective optical separation between
the actual wavefront and the ideal
spherical wavefront centered on the
desired focal point. Wavefront errors
are usually expressed in micrometers or
microns (mm), which are equal to one-
thousandth of a millimeter (0.001 mm).
For simple wavefront aberrations,
differences in curvature between the
aberrated and ideal wavefronts may
also be used to quantify the aberration.
The curvature of a wavefront, expressed
in diopters, is simply equal to the
reciprocal of its radius of curvature, in
meters. The sphere and cylinder powers
of a prescription actually indicate the
differences in curvature between the
aberrated and ideal wavefronts of
the eye. For more complex wavefront
aberrations, curvature, alone, does not
suffi ciently characterize the aberration.
Figure 2. Light diverging from an object point or
converging to an image point can be represented
using either rays or wavefronts representing the
envelope that bounds waves of light that have
traveled in unison or phase.
Figure 3. An optical system should produce
wavefronts of light that eventually converge to
the intended point focus (A), although wavefronts
of light do not converge to a sharp focus at the
desired point in the presence of aberrations (B).
WAVEFRONT
DISTANCE FROM OBJECT OR IMAGE POINTEQUALS RADIUS OF WAVEFRONT CURVATURE
WAVERAY
7 29.08.2013 04:49:07
8
Zenike Polynomial Series
Quantifying more complex wavefront
aberrations often relies on more
descriptive mathematics. Once the
wavefront errors have been measured
across the aperture (or pupil) of an
optical system, these measurements
can be “modeled” mathematically
for easier analysis and manipulation.
Most commonly, the shape of an
aberrated wavefront is modeled by
“fi tting”– or closely approximating –
the measurements with a series of
polynomial functions using mathematical
curve fi tting techniques. This allows
even complex wavefront aberrations
to be represented as a combination of
more basic shapes, often associated with
traditional optical aberrations.
The Zernike polynomial series is
commonly used to fi t wavefront
measurements. This is a set of functions
that each represent individual optical
aberrations, known as modes. Zernike
polynomials are typically grouped by
their radial order, which indicates how
rapidly the aberration increases with
pupil size (Figure 4):
• Low-order aberrations include
defocus and astigmatism, broken into
components at axis 45/135 (oblique)
and at axis 180/90 (WTR/ATR), which
are associated with the spherical and
cylindrical refractive errors of the eye,
respectively.
• High-order aberrations include
third-order aberrations, such as coma
and trefoil, fourth-order aberrations,
such as spherical aberration, and
aberrations of successively higher
radial orders, which increase more
rapidly as the pupil size increases.
Low-order aberrations are detrimental
to vision quality at both small and
large pupil sizes. These aberrations
are typically corrected by eliminating
the refractive errors of the eye using
sphero-cylindrical lenses. High-order
aberrations become more detrimental
to vision quality when the pupil size is
large. Although often less severe than
the low-order aberrations of an eye,
these aberrations can also degrade
visual acuity and reduce image contrast
(Figure 5).
Figure 5. Low-order aberrations degrade vision
quality even at smaller pupil sizes, whereas
high-order aberrations become more detrimental
at larger pupil sizes.
LARGE (6 MM) PUPIL SIZESMALL (3 MM) PUPIL SIZE
LOW
-ORD
ERAB
ERRA
TION
HIGH
-ORD
ERAB
ERRA
TION
Figure 4. Complex wavefront aberrations can be represented as a combination
of Zernike polynomial functions or “modes”:
• Each mode has a coeffi cient that specifi es the magnitude of the aberration in the wavefront.
• A wavefront is reconstructed by adding together quantities of the various modes.
• Modes are grouped by their radial order, which indicates the “power” of the function.
• Radial orders are whole numbers that begin with the zeroth order and continue indefi nitely.
• The total number of modes in each subsequent radial order increases by one mode.
• Modes in the second order are associated with spherical and cylindrical refractive errors.
• Modes in zeroth and fi rst radial orders are typically ignored in calculations of image quality.
NORMAL QUATREFOILOBLIQUE QUATREFOIL SPHERICAL ABERRATIONOBLIQUE SECONDARY ASTIG WTR/ATR SECONDARY ASTIG
VERTICAL COMA HORIZONTAL COMAOBLIQUE TREFOIL WTR/ATR TREFOIL
OBLIQUE ASTIGMATISM DEFOCUS WTR/ATR ASTIGMATISM
VERTICAL TILT OR PRISM HORIZONTAL TILT OR PRISM
RADI
AL O
RDER
Z 11Z -1
1
TRADITIONALEYEGLASS
PRESCRIPTION
1ST
2ND
3RD
4TH
Z -22 Z 2
2Z 02
Z 13Z -1
3Z -33 Z 3
3
Z -44 Z -2
4 Z 04 Z 2
4 Z 44
LOW
ORD
ERHI
GH O
RDER
IGNORED INIMAGE QUALITYCALCULATIONS
8 29.08.2013 04:49:11
9
Limits of Subjective Refraction
Subjective refraction is a clinical
procedure that determines the optimal
prescription for correcting the refractive
errors of an eye by having the patient
compare vision quality through different
spectacle lens powers, after establishing
an estimate of the refractive status of the
eye using objective refraction. Although
the subjective refraction is considered
the “gold standard” for judging the
accuracy and precision of objective
methods for performing refraction, the
outcome can vary between different
clinicians and between repeated
measurements by the same clinician.1
On average, refractions performed
by different clinicians agree to within
±0.12 D, but for individual patients the
discrepancies can be much larger (95%
limits of agreement = -0.90 to +0.65 D).2
This variability is due to a variety of
factors. The patient may use different
perceptual criteria when choosing
between lenses, such as sharpness,
contrast, or legibility. The depth of focus
of the eye may make it diffi cult for the
patient to discriminate between small
changes in image quality. The ideal
refraction will often vary with pupil size
and, therefore, luminance levels in the
presence of high-order aberrations. With
irregular corneas, sphere and cylinder
powers can vary by up to 1.00 D or
more between 3 mm and 7 mm pupil
sizes.3 Rounding errors due to the use of
trial lenses in 0.25-diopter steps limit the
precision of the refraction to ±0.12 D.
Moreover, high-order aberrations may
result in irregular astigmatism or multiple
combinations of cylinder power and axis
that yield relatively good vision quality.
Although the Jackson cross-cylinder
technique will isolate one of these local
maxima of vision quality, the technique
may not necessarily converge to the
combination of cylinder power and axis
that yields the best vision quality or global
maximum (Figure 6). Thus, the variability
of subjective refraction undermines any
attempt to validate objective refraction
techniques, because the “gold standard”
is, in effect, a “moving target.”
Although conventional vision corrections
are intended to correct only the low-
order aberrations of the eye, the optimal
sphere and cylinder powers for these
corrections are infl uenced by higher-
order aberrations. In the presence
of ocular high-order aberrations, the
vision correction that maximizes the
focus of paraxial rays of light passing
through the central region of the pupil
will differ from the vision correction
that maximizes the focus of marginal
rays passing through the periphery
of the pupil. Further, both vision
corrections will suffer from residual
blur (Figure 7). Maximum retinal image
quality will actually be achieved when
blur is minimized using a balanced vision
correction that represents a compromise
between the defocus of the paraxial rays
and the marginal rays.
In fact, experiments have shown that
patients typically judge the optimal focus
as lying somewhere between the paraxial
focus and the marginal focus.4 Because
subjective refraction often involves
viewing conditions that serve to restrict
the pupil size, however, this procedure
favors the paraxial focus. Additional
information regarding the high-order
aberrations of the eye should be applied
in order to compute a sphero-cylindrical
vision correction that optimizes image
quality for the entire pupil. Determining
the size of the patch of blur produced on
the retina is one such method, but there
are many ways to quantify image quality,
each emphasizing a different aspect of
the optical system and resulting image.
Because the patient ultimately decides
whether one retinal image is better than
another, the best metric of image quality
should refl ect the image processing
characteristics of the human visual
system.
Figure 6. In the presence of high-order aberrations,
a polar plot of retinal image quality as a function
of cylinder power may reveal multiple combinations
of cylinder power and axis that yield “good”
vision quality, although the Jackson cross-cylinder
technique may not necessarily converge to the
“best” vision correction.
Figure 7. Because ocular high-order aberrations,
such as spherical aberration, cause the best vision
correction for focusing the paraxial rays (A) to differ
from the best correction for focusing the marginal
rays (B), the ideal vision correction for the entire
pupil typically represents a compromise between
these two extreme cases.
PARAXIALFOCUS
OBJE
CT P
OINT
AT I
NFIN
ITY
MARGINALFOCUS
OBJE
CT P
OINT
AT I
NFIN
ITY
SPHERICALABERRATION
WEAKERMINUS LENS
STRONGERMINUS LENS B
A
045°
135°
180°090°
BESTVISION
WORSTVISION
2D
2D
060°
150°
075°
165°
030°
120°
015°
105°
1D
LOCALMAXIMUM
GLOBALMAXIMUM
9 29.08.2013 04:49:13
10
Wavefront-guided Objective Refraction
An aberrometer measures the wavefront
aberrations of the eye. Just as topographers
now provide more detail regarding the
surface characteristics of the cornea than
conventional keratometers, aberrometers
now capture more detail regarding the
refractive characteristics of the eye than
conventional autorefractors. Autorefractors
typically measure only the low-order
aberrations of the eye over a small region of
the pupil, roughly 3 mm in diameter, which
essentially restricts light to the paraxial
region of the eye. Aberrometers, on the
other hand, measure both the low-order
and high-order aberrations of the eye over
the entire pupil.
Many aberrometers utilize a Shack-
Hartmann wavefront sensor. A point source
of light – representing a distant object – is
projected onto the retina. The retinal
image of this point source then serves as
the object point for measurement. After
leaving the eye, light from this object point
passes through an array of small lenslets
that sample the optics of the eye over the
entire pupil. Each lenslet brings light to a
focus on a CCD sensor. Ideally, light from
the retina should produce a plane (fl at)
wavefront after passing back through
the optics of the eye. Any differences
between the plane wavefront and the
actual wavefront exiting the eye will cause
light to deviate as it passes through the
lenslets. The displacement of each focus
is then measured and used to model the
wavefront errors (Figure 8).
Once the wavefront errors have been
modeled, often using the Zernike
polynomial series, a vision correction can
be determined. Because the eye remains
in a constant state of movement, it is
not feasible to correct the high-order
aberrations of the eye with a spectacle
lens. A conventional sphero-cylindrical
vision correction eliminates the low-order
aberrations of the eye. Nevertheless, it is
possible to determine a wavefront-guided
vision correction using a combination of
conventional sphere power and cylinder
power that minimizes the blur that results
from the interaction between the low-
order and high-order aberrations.
For instance, approximating a map of the
vergence errors over the pupil of the eye
with the best-fi tting sphero-cylindrical
map is a simple way to determine an
ocular wavefront refraction.5 Alternatively,
diffraction and interference effects can
be taken into account by approximating
a map of the wavefront errors over the
pupil.6 Other methods are available
to determine the optimal wavefront
refraction as well. The refraction can
be determined by iteratively optimizing
the optical quality of either the retinal
image of a point source using the point
spread function or the image of sinusoidal
gratings using the modulation transfer
function.
These functions can be manipulated in a
variety of ways to develop novel measures
of optical quality. Some especially useful
metrics, such as the Strehl ratio, have
even been modifi ed to account for
the early stages of visual processing.
Furthermore, other criteria may also prove
benefi cial when optimizing a wavefront-
guided vision correction. For instance,
experiments have shown that observers
prefer a vision correction that maximizes
the area of the pupil for which the
vergence errors or the wavefront errors
are negligible.7
Although refractive errors are determined
clinically for distant objects, in everyday
life the eye must form the retinal image
for objects at a variety of distances,
simultaneously. Thus, providing a large
depth of focus is arguably more important
than optimizing the focus for any single
viewing distance. This is why the ZEISS
VoluMetric merit function was developed
to optimize the three-dimensional image
intensity produced in the vicinity of the
focus of the lens-and-eye combination.
Rather than limiting attention to a single
focal plane, this function integrates
the beam intensity along the depth
of the focus in addition to the cross-
sectional area in order to minimize
the volume of blur. The VoluMetric
function therefore provides a superior
wavefront refraction that is less sensitive
to the frequent fl uctuations in viewing
distance, accommodation, and pupil
size encountered throughout the day
(Figure 9).
POINT ONRETINA
ERRO
R
ABERRATIONS OF EYEDISTORT WAVEFRONT
CCDSENSOR
LENSLET FOCUSDISPLACED ON CCD
LENSLETARRAY
MINIMUMBLUR VOLUME
WAVEFRONTWITH SPHERICAL
ABERRATION
OPTIMUMFOCUS
WAVEFRONT-GUIDEDVISION CORRECTION
Figure 8. Many commercial aberrometers, like
i.Profi ler® plus by ZEISS, use a wavefront sensor that
is based on the Shack-Hartmann principle, which
measures the focusing errors of an array of small
lenslets caused by the distorted wavefront originating
from a point on the retina at the back of the eye (not
to scale).
Figure 9. The ZEISS VoluMetric merit function utilized
to calculate a wavefront-guided i.Scription® seeks
a combination of sphere and cylinder power that
minimizes the three-dimensional “volume” of a
focus distorted by high-order aberrations in order to
maximize vision quality as well as the depth of focus
of the eye.
10 29.08.2013 04:49:15
11
i.Scription® Technology by ZEISS
Traditional subjective refractions
are often performed under viewing
conditions involving luminance levels
that result in relatively small pupil sizes.
Ambient light levels vary considerably,
however, between photopic (or daytime),
mesopic (or twilight), and scotopic (or
nighttime) viewing conditions. Although
most light adaptation occurs within
the retina, the size of the pupil varies
inversely as a function of luminance,
becoming smaller as luminance increases
in order to control retinal illumination
and to maximize retinal image quality.
The pupil size varies from a minimum
of roughly 2 during photopic vision to
a maximum of roughly 8 mm during
scotopic vision (Figure 10).8 Because of
the increasing dependence on pupil size
of the aberration modes in higher orders,
the infl uence of high-order aberrations
upon the ideal vision correction increases
at low luminance levels when the pupil
size is large.
Although the low-order, sphero-
cylindrical refractive errors will remain
relatively constant in the absence of
high-order aberrations, regardless of
pupil size, all eyes suffer from at least
some high-order aberrations. Normal
eyes have on average a root-mean-
square error (RMS) of 0.33 mm for
high-order aberrations at a pupil size of
6 mm, which is roughly equivalent to
0.25 diopters of defocus.9 Determining
the endpoint of refraction by taking
into account the effects of high-order
aberrations on retinal image quality
may therefore result in superior sphero-
cylindrical vision corrections that deliver
optimal vision over a broader range of
luminance levels.
Using wavefront aberrometry data
captured by i.Profi ler® plus, the ZEISS
VoluMetric merit function calculates a
wavefront refraction with sphere and
cylinder powers that deliver the most
optimal vision quality over a range of
viewing conditions (Figure 11). The
clinician must then conduct a standard
subjective refraction as usual for
assessing binocular vision, performing
binocular balancing, and determining
the near addition. Lastly, the wavefront
refraction is reconciled against the
subjective refraction using a unique
subjective refi nement algorithm in order
to ensure that the spherical equivalent
of the fi nal refraction does not deviate
excessively from the subjective fi ndings.
The result of this patented process is
the patient’s i.Scription®: a precise,
wavefront-guided vision correction.
Because the VoluMetric merit function
maximizes depth of focus of the eye
based on the interaction between the
low-order and high-order aberrations,
i.Scription® should improve visual
performance under more demanding
viewing conditions. Wearers may
experience improvements in contrast
sensitivity, low-light vision, and
night driving due to a reduction in
“night myopia” caused by high-order
aberrations and the Purkinje shift in
ocular color sensitivity. Additionally,
wearers may experience a reduction in
the apparent effects of ocular chromatic
aberration.
Every i.Scription® spectacle refraction
comprises a sphere power, cylinder
power, and cylinder axis. Unlike traditional
eyeglass prescriptions, however, the
i.Scription® sphere and cylinder powers
are calculated to the nearest 0.01-diopter
step. Thus, i.Scription® is more precise
than a conventional prescription, which
can result in rounding errors of up to
±0.12 D due to the use of trial lenses
in 0.25-diopter steps. The use of more
precise prescription values further
enhances the optimized wavefront
refraction of i.Scription®.
LUMINANCE
PUPIL SIZE
10-6 10+6 10+810-4 10-2 100 10+410+2
7.9 2.0 2.07.5 6.1 3.9 2.12.5
STARLIGHT MOONLIGHT OFFICE LIGHT SUNLIGHT
MESOPIC PHOTOPICSCOTOPIC
CD/M2
MM
Figure 10. The pupil of
the human eye varies
from a minimum of
roughly 2 mm during
photopic vision to a
maximum of roughly
8 mm during scotopic
vision.
Figure 11. Conventional vision corrections
often provide optimal performance under ideal
viewing conditions (A), but may provide reduced
performance under more demanding conditions
due to high-order aberrations (B), whereas
i.Scription maximizes visual performance over a
broad range of viewing conditions (C).
OBJE
CT P
OINT
AT I
NFIN
ITYOB
JECT
POI
NT A
T INF
INITY
OBJE
CT P
OINT
AT I
NFIN
ITY
PUPIL BLOCKS HIGH-ORDER ABERRATIONS
REFRACTION AFFECTEDBY H-O ABERRATIONS
REFRACTION OPTIMIZEDFOR H-O ABERRATIONS
i.SCRIPTIONMODIFICATION
B
A
C
LOW-LIGHT VISION
EXAM ROOM VISION
11 29.08.2013 04:49:17
12
Precision Customized Lenses
Once an i.Scription® vision correction has
been determined, a spectacle lens must
be fabricated to the desired prescription
powers. Unfortunately, traditional
spectacle lenses often introduce
additional wavefront aberrations that
can compromise optical performance
and vision quality for the wearer
compared to the vision achieved during
the eye exam (Figure 12):
• Residual low-order aberrations occur
in traditional spectacle lenses because
of oblique astigmatism as a result of
either the tilt of the fitted lens (that is,
position of wear) or the angle that the
line of sight makes to the lens during
peripheral vision.
• Residual low-order aberrations also
occur in spectacle lenses due to the
power rounding errors inherent in
traditional lens surfacing, which
typically relies on hard “lap” tools
that are stocked in only 0.12-diopter
increments of surface power.
Traditional, semi-fi nished lens blanks
are typically available in only a handful
of unique base curves or lens designs,
which are factory-molded in mass
quantity. Changes to the basic design of
traditional spectacle lenses are limited to
subtle variations in optical design across
a small number of base curves that must
work suffi ciently well for relatively broad
prescription ranges. Traditional spectacle
lenses are therefore specifi cally designed
for a few “average” prescription powers,
using either “average” fi tting parameters
for progressive lenses or assuming a
fi tting condition free of lens tilt for
single vision lenses. The use of average
assumptions for each lens design results
in uncorrected low-order aberrations for
many wearers that can restrict, distort,
and blur the fi elds of clear vision.
Fortunately, the advent of free-form or
digital surfacing technology has freed
many lens designers from the constraints
of traditional, mass-produced spectacle
lenses. This modern manufacturing
platform relies on computer-controlled
generators that can precisely grind
surface curves of extremely high
complexity, including aspheric and
progressive designs, directly onto a semi-
fi nished lens blank.
Figure 12. Although vision through refractor-head
or trial-frame lenses may be clear (A), low-order
aberrations due to oblique astigmatism can occur
in fi tted spectacle lenses that will blur vision when
viewing through a lens that is tilted in front of the
eye (B) or when viewing through the periphery of
the lens (C).
Figure 13. Contour plots of ray-traced astigmatism demonstrate that the optics of traditional progressive
lens designs may be signifi cantly infl uenced by the position of wear, resulting in residual low-order
aberrations that can restrict, distort, and blur the zones of clear vision, whereas the fully-customized
lens designs utilized for i.Scription® lenses maintain the desired optical performance, regardless of the
prescription or position of wear.
LINE OF SIGHT AT ANGLETO LENS PERIPHERY
LENS AT ANGLETO LINE OF SIGHT B
C
A
+25+20+15+10+50–5–10–15–20–25
+25+20+15+10+50–5–10–15–20–25
+25+20+15+10+50–5–10–15–20–25
0.5
0.5
1.0
1.0
1.5
1.5
2.0
2.0
2.5
0.5 0.5
1.0 1.0
1.5 1.52.0
2.0
2.52.5
0.5 0.51.0
1.01.5
1.52.0
2.02.5
2.5
TRADITIONAL LENS DESIGNWITH SIGNIFICANT LENS TILT
+3.00 SPH RX: 15° PANTO & 15° WRAP +3.00 SPH RX: 15° PANTO & 15° WRAP
CUSTOMIZED LENS DESIGNWITH SIGNIFICANT LENS TILT
+3.00 SPH RX: NEGLIGIBLE LENS TILT
TRADITIONAL LENS DESIGNWITH “TRIAL FRAME” FITTING
EFFECTIVE RX:+3.00 SPH
EFFECTIVE RX:+3.51 –0.43 × 044
EFFECTIVE RX:+3.00 SPH
When combined with advanced optical
design software, free-form lens designs
can be calculated in “real time,” using
the wearer’s specifi c prescription and
fi tting parameters, immediately prior
to fabrication. Progressive and single
vision lens designs with i.Scription® by
ZEISS can therefore be fully customized
to the unique visual requirements of
the individual wearer. Customized
lenses preserve the intended optical
performance of the lens design by
minimizing residual low-order wavefront
aberrations, while ensuring that every
wearer enjoys the visual benefi ts of the
i.Scription® vision correction, regardless
of his or her prescription requirements
or position of wear (Figure 13).
Additionally, free-form surfacing is not
subject to the power rounding errors of
traditional lens surfacing.
12 29.08.2013 04:49:17
13
Clinical Proven Performance
Several clinical studies have demonstrated
the effi cacy of i.Scription® for improving
vision at night, color perception,
and contrast sensitivity, without
compromising visual performance
during normal viewing conditions.10 A
randomized, double-masked clinical study
comparing i.Scription® to conventional
spectacle refractions was conducted
at InSight Eyecare, an independent
optometric clinical research facility near
Kansas City, Missouri. This wearer trial
utilized a crossover design in which each
of 40 subjects wore a pair of ZEISS single
vision lenses with i.Scription® and a
pair of conventional single vision lenses
for one week each, in random order. A
variety of objective measures of visual
performance and subjective measures
of wearer preference demonstrated
improved results for vision quality under
demanding viewing conditions:
• Subjects rated ZEISS lenses with
i.Scription® higher for distance vision,
night vision, and color perception.
• ZEISS lenses with i.Scription® also
performed better on average than
conventional spectacle lenses in
measures of mesopic visual acuity and
contrast sensitivity.
A similar clinical study was also
conducted by investigators at the
prestigious Clinical Research Center of
the School of Optometry at the University
of California at Berkeley. Each of 30
subjects compared ZEISS single vision
lenses with i.Scription® to conventional
single vision lenses in a randomized,
double-masked wearer trial utilizing a
crossover design. Once again, a variety of
objective measures of visual performance
and subjective measures of wearer
preference demonstrated several positive
outcomes:
• Subjects with low- to moderate-
prescription powers preferred ZEISS
lenses with i.Scription® more often
for distance vision, active vision,
sharpness, changing focus, and overall
vision.
• Subjects preferred ZEISS lenses with
i.Scription® more often for night
vision, vividness of colors, and having
less glare.
• ZEISS lenses with i.Scription® also
performed better than conventional
spectacle lenses in measures of low-
contrast, mesopic visual acuity – by
approximately half a line of acuity.
Each i.Scription® vision correction
represents a synergy between a
thorough wavefront refraction, a skillful
subjective refraction, and a customized
lens design. The product of this patented
integration of optical science and clinical
art is a uniquely calculated wavefront
refraction, subjectively refi ned based
on the judgment of the clinician,
and precisely fabricated using digital
surfacing technology. ZEISS lenses with
i.Scription® will deliver excellent vision,
day or night, with improved contrast
sensitivity compared to conventional
spectacle lenses (Figure 14).
Automated refraction has not yet
replaced subjective refraction. Many
factors infl uence the fi nal manifest
refraction, including patient history,
binocular balancing, and cosmetic
appearance. Nevertheless, wavefront
refraction technology can determine a
starting point of refraction more quickly
and reliably by optimizing retinal image
quality more accurately. Subjective
refi nements can then be applied as
needed based on the professional
judgment of the clinician, in consultation
with the patient. Because aberrometers
characterize the optics of the eye so
thoroughly, the expectations are high
that this technology will yield improved
clinical outcomes.
Figure 14. Calculations
of the contrast
sensitivity of a
typical wearer for a
conventional spectacle
correction and an
i.Scription® wavefront-
guided vision
correction demonstrate
the improvement in
visual performance
with i.Scription®.
1000
100
10
1
0.11 10 100
IMPROVEDACUITY
IMPROVEDCONTRAST
CONT
RAST
SEN
SITI
VITY
SPATIAL FREQUENCY (CYCLES PER DEGREE)
CONTRAST SENSITIVITY COMPARISON FOR A TYPICAL EYE
i.SCRIPTIONSTANDARD
13 29.08.2013 04:49:19
14
// 1 Exam and Refractioni.Scription® Clinical StudyReport on the clinical study performed by Carmel Mountain Vision Care, published in the Review of Optometry (2013).
Our Study Design
We have served as clinical investigators
on scores of contact lens and
pharmaceutical studies and knew we
could execute a study to determine
if new technology was right for our
practice. We wished to avoid the pitfall
of very small experiences, and we
wanted to see if there was an indication
that we could do a better job. We used
a design similar to contact lens studies
we had conducted. A randomized,
single-masked, cross-over clinical study
comparing the effi cacy of customized
ZEISS Individual® Single Vision (SV)
lenses with i.Scription® technology to
customized ZEISS Individual® SV lenses
without i.Scription® technology was
selected for our practice, Carmel
Mountain Vision Care in San Diego. A
direct comparison of the visual quality
of 37 subjects was made between the
two single vision lenses, under mesopic
and photopic conditions. Each subject
was measured with the i.Profi ler® plus
and received a comprehensive eye
examination, including our customary
subjective manifest refraction. Objective
measurements were taken of the
wavefront aberrations of the eye with
the i.Profi ler® plus and combined with
our subjective manifest refractions
by use of the proprietary ZEISS Volu-
Metric merit function algorithm to
create customized ZEISS Individual®
SV lenses with i.Scription® technology
(Figure 3). A second pair of customized
ZEISS Individual® SV lenses without
i.Scription®, using only the conventional
subjective refraction, were manufactured
as the control lenses. The control and
test lenses were fi tted into two identical
frames, with the same position of
wear, measured with the i.Terminal®
by ZEISS. A refractive index of 1.6 was
used, as the lens material and lenses
were manufactured according to the
standard customization parameters of
ZEISS Individual® SV lenses. The study
spectacles were randomized for wearing
order and labeled to mask the type. The
subjects were compensated at their fi nal
visit with the study pair that they felt
gave them the best overall visual quality
and comfort.
Subject Selection
Our eligibility criteria for subjects
required normal, healthy eyes; patient
age between 18-40 years; best corrected
monocular distance visual acuity
(logMAR) of 0.10 (20/25) or better; and
subjects who were full-time eyeglass
wearers for all distances. Subjects were
classifi ed according to three different
prescription power categories. The
classifi cation was made according to the
eye with the highest Manifest Refraction
Spherical Equivalent (MRSE). Subjects
with an MRSE greater than or equal to
-4.00D were classifi ed as Subjects with
High Myopia; between -3.75D and
A group of Doctors of Optometry at Carmel Mountain Vision Care in San Diego conducted a randomized, singlemasked, cross-over clinical study comparing the effi cacy of customized ZEISS Individual® Single Vision (SV) lenses with i.Scription® technology to customized ZEISS
-0.25D inclusive, as Subjects with Mid-
Myopia and more plus or equal to plano,
as Subjects with Hyperopia. The study
concluded with 11 subjects with high
myopia; 21 subjects with mid-myopia
and fi ve subjects with hyperopia.
Clinical Measurements
Clinical measurements were conducted
under mesopic (30 lux) and photopic
(300 lux) conditions. The i.Profi ler® plus
measurement, Subjective Refraction and
Point-Spread-Function (PSF) test were
performed under mesopic conditions.
The PSF test was designed with a single
green LED light source, the same size as
a .0.1 LogMAR letter, that was mounted
against a matte black blackground
and placed at the same distance from
the patient as the visual acuity char.
Monocular und binocular high (100%)
and low (10%) contrast distance visual
acuities were measured in a straight-
ahead gaze position with the M&S Smart
System II 20/20™ 2010. Monocular
and binocular high-contrast near visual
acuities were measured in a straight
ahead gaze position, with ZEISS near
VA chart. Independent mesopic and
photopic pupil sizes were measured with
a Colvard Pupilometer.
Control and Test eyeglasses were worn
independently for a 10-day period
each, followed by a one-week direct
comparison period. Clinical measurement
14 29.08.2013 04:49:20
15
evaluations and patient-reported
outcomes were performed on days 10,
20 and 27, respectively. Results were
analysed using the following methods:
Percentage Analysis; Descriptive
Statistical Analysis; P-Test Analysis; Bland-
Altman Plots and ANOVA calculations. A
Visual Analog Scale response valuation
was used in all questionnaires (Figure 4).
Subjects were required to fi ll out a daily
study journal commenting on their
visual experiences under specifi c visual
conditions that were outlined in the
journal.
Results
Comparison to Habitual Rx. We asked
the subjects to compare the two pairs
of study eyewear to eyewear with
their habitual prescription prior to the
study. Subjects were asked to rate the
i.Scription® Rx and manifest refraction
Rx to their habitual Rx after they wore
each pair independently for a minimum
of 10 days. The overall preference for
the i.Scription® Rx vs. the Habitual Rx
was 95%. The overall preference for
the Manifest Rx vs. the Habitual Rx
was 92%. This outcome indicates that
both study prescriptions were preferred
over the habitual Rx and there was an
apparent need for prescription change.
Overall Preference Final Choice.
Subjects were asked which study
spectacles they preferred overall as their
fi nal choice for visual quality and visual
comfort, and to rate that choice. Answers
were recorded on a 1 to 6 numerical
scale and analyzed. A mean difference
of 0.514 higher for the Test lenses
resulted. Percentage analysis revealed
that 59.5% of subjects preferred the
Test lenses as their fi nal choice for visual
quality and visual comfort, compared to
the Control. This differenc was driven
mainly by the subjects with high myopia
and mid-myopia: 67% of subjects with
high myopia and 59% of subjects with
mid-myopia preferred the Test lenses and
subjects with Hyperopia had an equal
preference for the Test and Control lenses.
Mid Myopes High Myopes
The i.Scription® Rx was the prescription of choice overall for Visual Quality and Visual Comfort.
The overall preference for the i.Scription® Rx was mainly driven
by the High Myopes and Mid Myopes.
33%
67%
41%
59%
40,5%
59,5%
Final choice i.Scription Rx
Final choice Manifest RX
Final choice i.Scription Rx
Final choice Manifest RX
Final choice i.Scription Rx
Final choice Manifest RX
59.5% of subjects preferred the
i.Scription® Rx to the Manifest Rx.
15 29.08.2013 04:49:20
16
Visual Acuity: The Test lenses provided
better mean visual acuity in mesopic
conditions when compared to the Control
lenses. All visual acuity measurement
conditions resulted in no statistically
signifi cant mean logMAR acuity
differences with the Test and Control
lenses. In each case, the difference was
less than one line of vision improvement.
A difference of one line or more is
required to conclude that the diffrnce is
clinically signifi cant.5
Preference Under Lighting Conditions.
Our subjects wore each pair solely for
10 days, and then were allowed to make
a direct comparison with both study Test
and Control spectacles for one week
prior to their fi nal visit where clinical
measurements were also conducted with
direct comparison. The comparison is
reported in Figure 4. Overall, the Test
lenses were preferred for fi ve of the
seven visual conditions; for one visual
condition the Test lenses were preferred
equally to the Control and for the other
visual condition the Control lenses were
preferred over the Test. Adaptation Time.
Our subjects were asked how quickly they
adapted to the study spectacles after
having worn each pair independently for
a minimum of 10 days. Answers were
recorded on a 1 to 5 numerical scale
and analyzed. A mean difference of
0.027 resulted, indicating no signifi cant
difference and that the Test rated slightly
higher than the Control. There were no
cases of non-adaptation for either the
Test or Control.
Visual Conditions Questionnaire
Our subjects were asked which study
spectacles they preferred overall for
17 different visual conditions (Figure 5).
The Test lenses rated higher than the
Control lenses for all 17 different visual
conditions: distance vision; mid-range
vision; near vision; active vision;
brightness; brightness of environment;
colors more vivid; edges sharper;
less glare; peripheral vision; depth
perception; adaptation; visual comfort
for distance vision; visual comfort for
near vision; quicker to change focus;
night vision; natural vision. They were
asked which statement(s), out of eight,
best described their study spectacles.
They were allowed to select either or
both study spectacles for each statement
when applicable. The Test lenses rated
higher than the Control for all eight
statements as follows: Provides more
comfortable vision; provides fewer
headaches; provides good near vision;
provides good intermediate vision;
provides good distance vision; provides
a feeling of more relaxed vision; provides
less tiredness; provides less strain.
Likelihood to Recommend Lenses. Our
subjects were asked to rate on a scale
from 1 (very unlikely) to 10 (very likely)
how likely they would recommend the
study spectacles to family or friends.
A mean difference of 0.135 resulted,
which indicates that the Test lenses rated
higher than the Control. Percentage
analysis of the number subjects who
gave the highest rating (10) revealed
that 41% of subjects would “very likely”
recommend the Test lenses and 27% of
subjects would “very likely” recommend
the Control.
Our Conclusions
As clinicians and practice managers,
we try to make quantitative decisions
for constant product and service
improvements to meet our mission of
providing the highest quality of care in
our region. While we appreciate that our
in-offi ce studies may not have statistical
power for 95% confi dence, we endeavor
to execute our studies to allow us to
have more than anecdotal evidence.
In this study, we found that the blended
or optimized prescription of the lenses
with i.Scription® technology prevailed
in every category over the same ZEISS
Individual® lenses made according to
our subjective refractions only. The
overall preference rating was higher
with the Test spectacles than with the
Control spectacles. This difference was
driven mainly by the subjects with high
myopia and mid-myopia. While not
statistically our clinically signifi cant, the
preponderance of the evidence supports
a trend for enhanced performance and
patient satisfaction with the Test lenses.
The fi nding that the visual acuity was
better for the Test lenses under mesopic
lighting conditions was consistent with a
clear area where we wanted to improve
our patients(?) visual performance.
The preference for the Test lenses for
the Point Spread Function test also
supported our desire to improve visual
comfort in the presence of point sources
of light under dim light conditions.
Adaptation time was faster with the
Test spectacles than with the Control
spectacles. There were no cases of non-
adaptation for either the Test or Control
spectacles. We were impressed that the
Subjective ratings were higher with the
Test spectacles than with the Control
spectacles for all 17 visual conditions.
This complements the discovery that
the Test spectacles were more likely
to be recommended that the Control
spectacles. Under direct comparison, the
Test spectacles were preferred for visual
quality over the Control spectacles under
fi ve of seven lighting and contrast acuity
conditions. Further, the Test spectacles
16 29.08.2013 04:49:20
17
provided more comfortable vision;
fewer headaches; good near vision;
good intermediate vision; good distance
vision; a feeling of more relaxed vision;
less tiredness and less strain.
References1. Goss D, Grosvenor T. Rliability of refraction – a literature review. J Am Optom Assoc, 1996; Vol. 67, No. 10, pp619-630.2. Zadnik K, Mutti D, Adams A. The repeatability of measurement of the ocular components. Investigative Ophthalmology & Visual Science, June 1992;
Vol. 33, No. 7, p 23-29.3. Simms C, Durham D, The Jackson Cross Cylinder Disproved. T(?) Am. Ophth. Soc. 1986, vol. LXXXIV, pp 355-386.4. Bullimore M., Fusaro R., Adams C. The repeatability of automated and clinician refraction. Optom Vis Sci, 1998; Vol 75, No. 8, pp 617-622.5. ZEISS instrument guide.
Individual®, i.Scription® and i.Profi ler® plus are registered trademarks of Carl Zeiss Vision International GmbH. The authors wish to acknowledge the assistance of Jerome A. Legerton, OD, MS, MBA, and Barbara H. Bytomski, OD, in the preparation of this manuscript and the research from which it was derived.
Overall, our consideration of all the
investigational categories supports a
trend that i.Scription® by ZEISS provides
better visual quality and comfort for our
patients, and our patients are more likely
to recommend these lenses; thereby
adding to the growth of our practice.
We expect the use of this technology,
combined with our other study-based
decisions, will continue to support our
valuable fi nal product for the practice:
enthusiastic, satisfi ed patients.
27%
38% 35%24%
41%
35% 24%
38% 38%
51%
8%
41%
22%
19%59%
Mesopic High Contrast: i.Scription Rx Preference
Mesopic High Contrast: Manifest Rx Preference
Mesopic HighContrast: No Preference
Mesopic Neari.Scription Rx Preference
Mesopic NearManifest Rx Preference
Mesopic NearNo Preference
Mesopic PSFi.Scription Rx Preference
Mesopic PSFManifest Rx Preference
Mesopic PSFNo Preference
Mesopic Low Contrasti.Scription Rx Preference
Mesopic Low ContrastManifest Rx Preference
Mesopic Low ContrastNo Preference
Photopic Hight Contrasti.Scription Rx Preference
Photopic Hight ContrastManifest Rx Preference
Photopic Hight ContrastNo Preference
• Results: The i.Scription® prescription
was preferred by direct comparison
under Mesopic conditions.
i.Scription® was preferred under
Photopic High Contrast conditions
by direct comparison.
• Photopic Low Contrast: The
i.Scription® & Manifest Rx were
preferred equally, 57% of subjects saw
“No Difference”
• Photopic Near Vision: The Manifest
Rx was preferred by only 7% more
subjects. 59% of subjects saw
“No Difference”
KEY FACTS Study Spectacles: ZEISS Individual™ SV with i.Scription® vs. ZEISS Individual™ SV without i.Scription®
Study Subjects: 20 Males; 17 Females / 11 Subjects with High Myopia; 21 Subjects with Mid-Myopia; 5 Subjects with Hyperopia
Eligibility criteria: Normal healthy eyes 18 to 40 years / Full time spectacle lens wearers for all distances / Best corrected monocular distance VA 20/25 or better
Carmel Mountain Vision Care Optometry in San Diego, California, conducted a clinical study over the course
of eleven months. The aim of such study was to improve vision under mesopic conditions without compromising
vision under photopic conditions with i.Scription® by ZEISS.
17 29.08.2013 04:49:21
18
// 2 Lens Fitting and ConsultationStandard Spectacle Lens FittingArticle written by Darryl Meister, ABOM Carl Zeiss Vision (2010)
Traditional lens fi tting measurements are
limited in terms of both accuracy and
precision. Accuracy is the closeness of the
average measurement result to the actual
value. Random and systematic errors
introduced by parallax, or unwanted
head turn, and unwanted head tilt reduce
measurement accuracy. Errors introduced
by improper measurement techniques
or use of poorly maintained tools are
other causes for measurement errors.
Precision is the size of the smallest reliable
measurement or the overall spread of
the individual measurement results.
The standard “PD ruler” can be used to
measure reliably only to within ±½ mm
of the actual value, limiting measurement
precision. Random errors can also
reduce measurement precision, since
the repeatability of the measurement is
reduced. For any individual measurement,
limitations in either the accuracy or the
precision of a measurement technology
can result in a measurement error from
the true value (Figure 1).
Parallax error is the apparent change in the
location of the wearer’s pupil at the plane
of the lens when observed from different
positions. It is one common source of
High PrecisionHigh Accuracy
Low PrecisionLow Accuracy
Low PrecisionHigh Accuracy
High PrecisionLow Accuracy
Figure 1 Accuracy and precision measurement errors
Figure 3. Unwanted head tilt in vertical centrationFigure 2 Parallax errors in measurement
Eye Level Difference
Parallax Error
measurement error for both fi tting height
measurements and manual measurements
of interpupillary distance using a
pupilometer (Figure 2). For a head turn
of 12°, a 1.4 mm error in measurement
would occur for a binocular PD of
64 mm. Differences in eye level between
the dispenser and the eyeglass wearer
introduce signifi cant parallax errors.
Every 1 inch difference in eye level height
can result in an error in fi tting height
measurement of roughly 1.7 mm. In the
absence of a habitual head turn, the
wearer should be measured while facing
straight-ahead with his or her face parallel
to the dispenser’s face. Another signifcant
source of error derives from unwanted
head tilt. Every 1° of unwanted vertical
head tilt will result in an error in fi tting
height measurement of roughtly 0.5 mm.
In order to avoid erroneous vertical
centration, wearers should be measured
while maintaining a natural posture at the
same eye level as the dispenser (Figure 3).
Modern progressive lenses offer wearers better optical performance than ever, although the performance of these lenses is still limited by the accuracy and precision of the fi tting process.
18 29.08.2013 04:49:21
19
Lens centration errors reduce the width
of the binocular viewing zones, which
in turn degrades visual performance
and comfort. Every 1 mm of error in
monocular PD measurement will reduce
the width of the binocular viewing
zone by 2 mm (Figure 4). A 2 mm
error in centration from inaccurate
measurements can reduce the wearer’s
binocular fi eld of view by up to 25% or
more
This is especially critical for the
intermediate and near zones, which are
relatively narrow, or in lenses with high
add powers
Figure 4 Error in monocular PD
Pupilometers are not as accurate and
precise as eyecare professionals generally
assume. Repositioning of the dispenser’s
viewing eye and slight movements of the
device occur as the dispenser switches
fi xation from eye to eye, resulting in
random measurement errors. Inexact
calibration can result in systematic
measurement faults that are seldom
detected, while poor mechanical action
or friction can result in additional inexact
measurement.
Wesemann, Bartz, and Arnolds (1997)
showed that the standard deviations of
measurements taken with two popular
pupilometers by 30 observers were
0.52 and 0.55 mm. Obstfeld and Chou
(1998) showed that many pupilometers
suffer from systematic errors as high as
2 mm when the interpupillary distance
was signifi cantly larger or smaller than
60 mm.
Figure 5. Monocular interpupillary distances measured from center of the frame.
PD-R > PD-L
PD-R PD-L
Total PD
Example:PD-R = 31PD-L = 29Total = 60
Overlap = 18 mm
0 mm 2 mm
Overlap = 18 mm
Pupillometers do not always fi t the
wearer´s face like the chosen spectacle
frame. These may sit differently on the
bridge of the nose compared to the
positioning of pupillometers. Horizontal
centration for PD should be based
upon the center of the bridge of the
fi tted frame, not the bridge of the nose
(Figure 5).
This ensures that the fi tting point of each
lens remains aligned with the pupil center
of each eye in the mounted frame. This
may actually result in different monocular
interpupillary distances for the same
wearer in different frames.
19 29.08.2013 04:49:22
20
4.0
3.5
3.0
2.5
2.0
1.5
1.0
0.5
0.0
Maximum Fitting Errors with
ZEISS Centration Technology
Accurate and precise fitting
measurements are critical to the visual
performance and wearer satisfaction,
especially for progressive lenses.
ZEISS centration technology, implemented
in i.Terminal® 2, employs a precision
camera to eliminate potential
measurement erros due to parallax or
the location of the dispenser relative
to the wearer. A patented laser speckle
fixation target ensures that the wearer
maintains a fixation to optical infinity.
The device automatically compensates
for any unwanted head turn or head
tilt to avoid parallax errors (Figure 6).
Additionally, “Auto-X” functionality
compensates the monocular interpupillary
distance measurements for any unwanted
horizontal head turn.
Studies have confi rmed the accuracy,
precision and repeatability of i.Terminal®
(Figure 7). Winter showed an average
variance of 0.04 mm and a range of
0.4 mm with i.Terminal®, compared to 0.7
and 1.8 mm for pupilometers.
Wesemann (2010) also showed a
standard deviation of only 0.13 mm with
i.Terminal®, compared to 0.35 mm for
several pupilometers.
Figure 6. Patented laser speckle target
Collimating Optics
Laser Speckle
Beam Splitter
Figure 7. Comparison of measurement variance for i.Terminal®
Variance Range
Man
ual
PD
Ru
ler
Pu
pill
om
eter
No
.1
Pu
pill
om
eter
No
.2
Pu
pill
om
eter
No
.3
Pu
pill
om
eter
No
.4
i.Ter
min
al®
by
ZEIS
S
20 29.08.2013 04:49:22
21
Customized Spectacle Lens Fitting
Traditional progressive lenses are often
designed for a focimeter, not the patient.
These must assume a fi xed position of
wear, often replicating how a focimeter
measures optics.
The position of wear represents the
position of the fi tted spectacle lens on
the patient, including tilt, wrap, and
vertex (Figure 1). Conventional, semi-
fi nished progressive lens design must
assume a single, average position of wear
while ray tracing the lens-eye system
during the optical design process. In
practice, however, the actual position of
wear varies signifi cantly among eyeglass
wearers, which results in errors from the
intended optical performance of the lens
design.
The position of wear introduces greater
wavefront aberrations than the high-
order aberrations of either the eye or
spectacle lens. The average high-order
wavefront error of human eyes is less
than 0.35 mm, which is comparable
to roughly 0.29 D of spherical error
(Figure 2).
Customized lenses like ZEISS Individual®
and Individual® Single Vision can utilize
additional fi tting measurements to
maximize the optical performance of the
lens design for each wearer.
Prescription changes are introduced
by variations in the position of wearer
that will blur vision through the central
viewing zones. Deviations from the
“standard” position of wear utilized
during optical design result in lens
aberrations due to lens tilt. These optical
aberrations blur vision through the
central viewing zones of the lens and
restrict the fi elds of clear vision.
Customization for the position of
wear fi ne-tunes the optics of the lens
design using the patient’s exact fi tting
geometry. Customized lenses will deliver
the intended optics with crystal clear
viewing zones to patients, regardless of
fi tting geometry (Figure 3).
Figure 2. Wavefront aberrations
derived from the position of wear
Figure 3. Customized progressive lenses
Figure 4. Wearer satisfaction comparison
Clinical studies have demonstrated
improved optical performance with
customized lenses. A study conducted at
the prestigious Clinical Research Center
at UC Berkeley, fi tted ZEISS Individual®
lenses using i.Terminal®. Han, Graham,
and Lin (2011) showed greater wearer
satisfaction with progressive lenses
customized for the position of wear
(Figure 4). By compensating for how
lenses are actually worn, customized
lenses deliver the intended optical
powers across the entire lens.
Figure 1. Position of wear
Reading Distance
Vertex
Tilt
21 29.08.2013 04:49:22
22
// 3 Technical Data
i.Polatest®
Line voltage 100 V to 240 V AC ± 10 %
Frequency 50 to 60 Hz
Power consumption 70 VA
Ambient conditions for intended use Temperature: +10 °C to +35 °C
Relative humidity: 30% to 85 % (no condensation)
Transport and storage conditions (in original packing)
Temperature: -15 °C to +60 °C
Relative humidity: 10% to 85% (no condenation)
Protection class I
Test area size (width x height) 299.5 mm x 223.5 mm
Testing distance 1 m to 8 m
Polarization direction for analyzers Image for right eye: 45°
Image for left eye: 135°
Dimensions (H x W x D) 608 mm x 570 mm x 85 mm
Weight incl. wall mounting bracket 12.75 kg
IR remote control unit 3 V, 100 μA
Spare parts Line fuse: 2x T 2.0 AE/250V
Battery for IR remote control unit: 2x Micro AAA 1.5 V
i.Profiler® plus
Measuring range sphere: –20 D to +20 D
Axis 0° – 180°
Measuring surface 2.0 mm to 7.0 mm (3 zones)
No. of measuring points 2.0 mm to 7.0 mm (3 zones)
Method Hartmann-Shack
Reference wavelength 1 555 nm (ISO 24157)
No. of rings 22 (18 complete rings)
No. of measuring points 3,425
Detected corneal surface at 42.125 D Dia. 0.75 mm to 9.4 mm
Diopters Measurement range 25 to 65 D
Accuracy ± 0.05 D (± 0.01 mm)
Reproducibility ± 0.10 D (± 0.02 mm)
Type A Complies with 19980
General dimensions 420 mm x 600 mm
Weight 30 kilograms
Power connection 100 V~ to 240 V~
Supply frequency 50 Hz to 60 Hz
Power consumption ≤ 200 VA
22 29.08.2013 04:49:23
23
i.Terminal® 2Line voltage 100 V to 240 V AC
Supply frequency 50 Hz to 60 Hz
Line fuses F1, F2 T2.0A/E 250V 5x20 IEC 60127-2/3
Protection class I I
Overvoltage category II
Pollution degree II
Laser class I according to CFR and IEC 60825-1
Ambient conditions for intended use Temperature: +10°C to +40°C
Rel. humidity: 30%…85% (no condensation) condensation)
Max. height (MSL): 2000 m
Transport and storage conditions(in original packaging)
Temperature: 40°C to +70°C
Rel. humidity: 10%…85% (no condensation)
Max. height (MSL): 2000 m
Dimensions of base 60 cm x 60 cm
Instrument (height x width x depth) 1250–2100 mm x 600 mm x 600 mm
Weight approx. 47 kg
Min. ceiling height 2.1 m
Range of movement Minimum eye height: 110 cm (corresponds to approx. 120 cm body height)
Maximum eye height: 195 cm (corresponds to approx. 208 cm body height)
Lifting range: 85 cm
1. Goss D. and Grosvenor T. “Reliability of refraction – a literature review.” J Am Optom Assoc, 1996; Vol. 67, No. 10, pp 619-630.2. Bullimore M., Fusaro R., and Adams C. “The repeatability of automated and clinician refraction.” Optom Vis Sci, 1998; Vol 75, No. 8, pp 617-622.3. Collins M., Shaw A., Menkens E., Davis B., and Franklin R. “The Effect of Pupil Size on Subjective Refraction with Irregular Corneas.” Invest Opthalmol Vis Sci, 2002; Vol. 43: E-Abstract 20584. Cheng X., Bradley A., and Thibos L. “Predicting subjective judgment of best focus with objective image quality metrics.” J Vis, 2004; Vol. 4, No. 4, pp 310-321.5. Nam J., Thibos L., and Iskander D. “Describing ocular aberrations with wavefront vergence maps.” Clin Exp Optom, 2009; Vol. 92, No. 3, pp 194-205.6. Thibos L. “Principles of Hartmann-Shack aberrometry.” J Refract Surg, 2000; Vol. 16, No. 5, pp S563-565.7. Thibos L., Hong X., Bradley A., and Applegate R. “Accuracy and precision of objective refraction from wavefront aberrations.” J Vis, 2004; Vol. 4, No. 4, pp 329-351.8. Stockman A. and Sharpe L. “Into the twilight zone: the complexities of mesopic vision and luminous effi ciency.” Ophthal Physiol Opt, 2006; Vol. 26, No. 3, pp 225-239.9. Salmon S. and Van de Pol C. “Normal-eye Zernike coeffi cients and root-mean-square wavefront errors.” J Cataract Surg, 2006; Vol. 32, No. 12, pp 2064-2074.10. Clinical study data on fi le with Carl Zeiss Vision.
23 29.08.2013 04:49:26
EN
_20
_01
0_7
28
5I
Prin
ted
in G
erm
any
CZ-
VII/
2013
Subj
ect
to c
hang
e in
des
ign
and
scop
e of
del
iver
y an
d as
a r
esul
t of
ong
oing
tec
hnic
al d
evel
opm
ent.
Carl Zeiss Vision GmbHTurnstrasse 2773430 AalenGermany
Phone: +49 (0) 7361 598 5000 Fax: +49 (0) 7361 591 480 Email: [email protected]/vision
24 29.08.2013 04:49:26