octopus 3000 manual v 22
TRANSCRIPT
-
8/13/2019 Octopus 3000 Manual v 22
1/113
USER GUIDE
AquadyneVersion 2.2
OCTOPUS 3000
-
8/13/2019 Octopus 3000 Manual v 22
2/113
-
8/13/2019 Octopus 3000 Manual v 22
3/113
First EditionJULY 1998
Copyright AQUADYNE Computer Corp., 1994-1998. All rights reserved worldwide.No part of this publication may bereproduced, transmitted, transcribed,stored in a retrieval system or trans-lated into any language in any form byany means without the written permis-sion of AQUADYNE Computer Corp.
Licen ses and Tradem arksOctopus, AquaNode, AquaGuard,AquaWeb, H
2O/S are trademarks of
Aquadyne Computer Corp.
Credits
Designed and edited byKirkpatrick Price & Associates
Technical Review by Matt Kim,Patrick Johnston, Rick Kriss,Christophe Dugas, Rod Warlick,
and Dean McDaniel.
Part Number: 75-050705
AQUADYNEComputer Corporation7343 "P" Ronson RoadSan Diego, California 92111USA
Sales: (619) 569-2082Tech. Support: (619) 569-2082 x250Fax: (619) 495-3119E-Mail: [email protected]
[email protected]: http://www.aquadyne.com
Printed in USA
-
8/13/2019 Octopus 3000 Manual v 22
4/113
-
8/13/2019 Octopus 3000 Manual v 22
5/113
Table of Contents
i
1. INTRODUCTION
How the Octopus Works ............................................................... 1-3
User Guide Features ................................................................... 1-4
Quick Reference Menu Map......................................................... 1-4
Monitoring Capability .................................................................. 1-4
Flash Memory Architecture ......................................................... 1-5
Status Lights .............................................................................. 1-5
Infrared Printer Port .................................................................... 1-6
Powering the Octopus .................................................................. 1-6
Control Capability ....................................................................... 1-7
2. INSTALLING THE COMPONENTS
Step 1. Mount the Octopus Controller .......................................... 2-4
Step 2. Installing the Temperature, pH, and ORP Probes .............. 2-6
Step 3. Install the Digital Input Hardware .................................... 2-9
Step 4. Install Optional X-10 Control Module.............................. 2-11
Step 5. Install and Address Optional Satellite Module(s) ............. 2-11
Step 6. Install Communications Equipment ............................... 2-13
Step 7. Power Up the Octopus 3000 Controller ........................... 2-17
Step 8. Install the 9-Volt Backup Battery Pack (Optional) ........... 2-18
3. PROGRAMMING THE SOFTWARE
Overview of the Octopus Menus ................................................... 3-2
Using the Menu MapAn Example .............................................. 3-4
-
8/13/2019 Octopus 3000 Manual v 22
6/113
Table of Contents
ii
Initial Software Setup .................................................................. 3-4
Step 9. Set the System Clock ....................................................... 3-4
Step 10. Enter the Protective Password ........................................ 3-6
Step 11. Set the Noise Filter Control ............................................ 3-6
Step 12. Select the Conductivity Temp Compensation Settings ..... 3-7
Step 13. Select a Conductivity Range ........................................... 3-7
Step 14. Select a Temperature Scale. ........................................... 3-8
Step 15. Enter Setpoint and Range Values ................................... 3-8
Step 16. Entering Alarm Range Values ....................................... 3-12
Step 17. Set up Digital Input Alarm ........................................... 3-15
Step 18. Set up the Lighting System .......................................... 3-15
Step 19. Set up the WaveMaker System ..................................... 3-16
Step 20. Set Up System Hold Control ......................................... 3-16
Step 21. Assign System Control Addresses................................. 3-17
Step 22. Set Up the Communication Software ............................ 3-18
Step 23. Enable the Octopus System Control ............................. 3-21
Step 24. Test the System Configuration ..................................... 3-22
4. OPERATION AND MAINTENANCE
Daily check list ........................................................................... 4-2
Cleaning the Probes .................................................................... 4-3
Calibrating the Probes................................................................. 4-4
Replacing a Probe........................................................................ 4-8
Conducting a System Test ........................................................... 4-8
Changing a Control Address ........................................................ 4-9
Operating the Infrared Printer ..................................................... 4-9
-
8/13/2019 Octopus 3000 Manual v 22
7/113
Table of Contents
iii
5. SETTING UP AQUAWEB FOR WINDOWS 95
AquaWeb - Introduction ............................................................... 5-2
Step1-Configuring Windows Com port setting to support AquaWeb 5-4
Step 2 - Installing AquaWeb on your PC ....................................... 5-5
Step 3 - Starting and configuring AquaWeb .................................. 5-7
6. ADLOADER - Upgrading the H2O/S software
Upgrading the H2O/S software in your Aquadyne Computer ........ 6-2
7. TROUBLESHOOTING
Probe Troubleshooting................................................................. 7-2
Peripheral Equipment Response Troubleshooting ......................... 7-5
Modem/pager Troubleshooting .................................................... 7-6
AquaWeb Troubleshooting ........................................................... 7-7
APPENDICES
Appendix A: Menu Maps
Appendix B: Understanding Water Quality
Appendix C: Conversion Tables
Appendix D: Product Warranty
This manual is available in an Adobe pdf format from our web page athttp:/ / www.aquadyne.com . For the latest information, electronic up-dates to this documentation and other technical information, we recom-mend that you check with us regularly.
Aquadyne Computer Corp.7343 P Ronson Road
San Diego, California 92111USA
-
8/13/2019 Octopus 3000 Manual v 22
8/113
-
8/13/2019 Octopus 3000 Manual v 22
9/113
Chapter
1Introduction
-
8/13/2019 Octopus 3000 Manual v 22
10/113
CHAPTER1: INTRODUCTION1-2
Introduction
Figure 1.1 The Octopus 3000 Controller
-
8/13/2019 Octopus 3000 Manual v 22
11/113
CHAPTER1: INTRODUCTION 1-3
Introduction
AquaWebRemote AccessOur latestsoftware product that allows you tocommunicate with your Octopus 3000,perform advanced graphical analysis ofcurrent and historical datafrom anyWindows 95 PC. The standard versionof AquaWeb is available for no costthrough our WEB site at:
HTTP://www.aquadyne.com
Single User and Multi-User versions ofAquaWeb are available for additionalcost through your dealer.
Until now, performing even the simplestfunctions with a controller required technicalor electronic expertise. The Octopus 3000controller makes maintaining the optimumaquatic environment easy.
The Octopus 3000 is constructed with onlythe highest quality components. The heart ofthe Octopus 3000 is an INTEL microprocessorwhich manages all of the activities of thecontroller. All information gathered by thecontroller is processed digitally. The Octopus3000 Architecture includes Digital Signal
HOW THE OCTOPUS 3000 WORKS
Congratulations on your purchase of themost sophisticated, yet easy-to-use aquaticenvironment controller on the market today.Designed with a powerful INTELmicropro-cessor, your Octopus 3000 controllerincludes the following features:
Continuous monitoring of conductivity,water temperature, pH and oxidationreduction potential (ORP).
H2
O/S
software with menu-drivenprogramming environment.
Aux. Input supports power failure, waterlevel or pressure alarming.
FLASH memory that makes upgradingyour Octopus 3000 with new H2O/Ssoftware a simple task.
Local audible alarm or optional remotedial-up pager alarm capability.
Parameter control capability for man-agement of peripheral equipment suchas pumps, and lighting.
Powerful datalogging capability thatholds up to 7-days of probe data.
Operates on 9V AC or on an optional 9VDC battery backup.
optional X-10 support for up to 255control outputs.
An infrared Data Port that supports anoptional hand-held HP printer for datareporting and graphical analysis.
Single communication port supportingboth direct connection to a PC or remotemodem communication.
Note
The helpful Notes logo high-lights items which will saveyou time and effort, andincrease your understanding ofthe Octopus operation.
The CAUTION logo highlightsitems which are critical tosafety or which may cause theOctopus to malfunction orbecome damaged.
CAUTION
-
8/13/2019 Octopus 3000 Manual v 22
12/113
CHAPTER1: INTRODUCTION1-4
Introduction
Processing (DSP) hardware and software thatprovides accurate and repeatable readingsover the life of the Octopus 3000. In addition,the microprocessor design provides unsur-passed flexibility, as reflected by the variety ofsoftware options that are available for theOctopus 3000.
AQUADYNEis committed to buildingproducts that meet world class standards ofquality and workmanship. If you have any
questions regarding the use of this product,AQUADYNEprovides hotline service throughthe Technical Support Group. They can bereached between 9:00 AM and 5:00 PM PacificStandard Time at (619) 569-2082 x 250 or viathe Internet at: [email protected]
The primary source for technical and otheruseful information about the Octopus 3000,including the latest release of this User Guidein AdobePDF format, can be found on ourWeb Page:
h t tp :/ / w w w .aquadyne.com.
Our Web Page contains current softwarereleases and always reflects the latest infor-mation about our products, so we suggest thatyou visit it frequently.
USER GUIDE FEATURES
The Octopus 3000 User Guide is presented infive chapters: Introduction, Installation,Programming, Operation and Maintenance,and Troubleshooting.
In Chapter 1, we provide a basic overview of
the Octopus 3000 and its capabilities. Werecommend that all users review this infor-
mation before proceeding with Chapter 2,Installation and Software Programming.
This User Guide includes several quickreference features, including Notes, Cautionsand a Quick Reference Menu Map.
QUICK REFERENCE MENU MAP
The laminated Menu Map included with your
Octopus 3000 summarizes the H2O/S soft-ware menu structure. Post it in a convenientspot for quick reference. For more detailedmenu maps, refer to Appendix A.
MONITORING CAPABILITY
The Standard Octopus 3000 software providesfor monitoring, control, and alarm capabili-ties. The four water quality parametersmonitored are:
Conductivity
Temperature
pH
ORP
In addition to the four parameters theOctopus 3000 also provides an aux. inputwhich can be configured to provide a digitalinput for power failure detection, or waterlevel alarming.
Probes connected to the bottom of theOctopus 3000 gather information and relay itto the Octopus 3000 for processing and
display on the screen.
-
8/13/2019 Octopus 3000 Manual v 22
13/113
CHAPTER1: INTRODUCTION 1-5
Introduction
Figure 1.2 Parameter Status Lights
The Octopus 3000 is programmed using amembrane keypad on the face of the unit.Directional arrows on the keypad are used tomove horizontally across a given menu level.The Enter key is used to move down themenu map or to make a selection.
Once in the submenus, if there is no activityfor ten minutes in most menus, the OctopusLCD display window will return to theCurrent Status Display. Settings entered upto that time will be saved. When not inprogramming mode, current status informa-
tion gathered by the probes is shown on theOctopus LCD Display Window.
FLASH MEMORY ARCHITECTURE
A key feature of the Octopus 3000 is that it is
designed using FLASH memory and a special
application embedded in the Octopus 3000
called abootst rap loader. When the loader is
used in conjunction with other PC software
offered by Aquadyne, routine updates to the
Octopus 3000 H2O/S software can be easily
accomplished. Refer to Appendix D for
instructions on upgrading the H2O/S soft-ware in the Octopus 3000. Additional
Information about the current revision level
of the H2O/S software, the availability of
AquaWeb for Windows 95, and other techni-
cal details about the Octopus 3000 can be
found on the Aquadyne WEB page:
http://www.aquadyne.com.
STATUS LIGHTS
Parameter Status Lights
The Parameter Status Lights located on theface of the Octopus indicate whether asystem is operating within its preset limits ora control system has been activated. A status
Solid Green. (System is dosing to lowerconductivity value.)
Flashing Green. Dwell State. (Will readconductivity again after dwell time is complete.)
Solid Red. (System is Dosing to raise conductivityvalue.)
Flashing Red. Dwell State. (Will read conductivityagain after dwell time is complete.)
Solid Green. Chiller On. (Current temperature istoo high.)
Flashing Green. Chiller Dwell State. (Dwell time toprevent compressor burn out.)
Solid Red. Heater On. (Current temperature toolow.)
Flashing Green. Dwell State. (Will read pH againafter dwell time is complete.)
Flashing Red. Dwell State. (Will read pH againafter dwell time is complete.)
Solid Red. (Ozone is dosing to raise ORP.)
No Light. (Parameter is within setpoints and nodosing required, OR control system is not activated.)
Solid Red. (pH buffer is dosing to raise pH.)
Parameter Status Lights
Temperature
pH
ORP
Conductivity
Solid Green. (CO2is Dosing to lower pH.)
-
8/13/2019 Octopus 3000 Manual v 22
14/113
CHAPTER1: INTRODUCTION1-6
Introduction
light lit red indicates that the system de-tected a low condition, a green light indicatesa high condition. A summary of ParameterStatus Lights is given in Figure 1.2.
System Status Light
The System Status Light located at thebottom of the Octopus 3000 indicates thegeneral condition of the Octopus 3000controller operating system. Figure 1.3summarizes the System Status Lights andtheir meaning.
Infrared Printer Port
The infrared printer port located at the bot-
tom left of the face of the Octopus 3000 is
standard. This port provides access for the
optional infrared data-log reporting feature.
Using the Infrared Printer (optional) data
collected by the Octopus 3000 can be printed
out in a variety of formats, including the
current data, the high and low points of data
collected within the last 24 hours, 48 hours,
or 1 week. Temperature, pH, and ORP canbe graphed. Conductivity will be added in a
future release of H2O/S software
Powering the OCTOPUS 3000
Primary power for the Octopus is provided by
an A/C wall-mount transformer. A UL, CSA
or CE approved transformer rated for the
correct input voltage (110V or 220V) is used
to power the Octopus. The output voltage to
the Octopus must be 9V AC.
The Octopus design incorporates FLASH
Memory, RAM, and a battery backed-up real-
time clock. Should power be lost, none of the
parameter setpoints, or configuration datawill be lost. However, during a power failure,
data in the datalog will be erased.
Backup power is provided by either of two
methods:
An optional 9-Volt backup battery packspecially designed by Aquadyne.
An optional UPS Backup Power System(Uninterrupted Power Supply). (The A/Cwall-mount transformer is plugged intothis system.)
The backup battery-pack is an inexpensivemeans of providing about 45 minutes of
backup power during a power failure. While
operating from the battery pack, the Octopus
Solid Green. Operating properly and availablefor data entry.
Flashing Green. Busy processing data.
Flashing Green/Orange. Sending commandsthrough the power line to the Satellite Modules.
System Status Lights
Flashing Red. Controller hardware failure. Unplugthe Octopus controller and call Aquadynecustomer service.
Figure 1.3 System Status Lights
UPS power supplies have built innoise and surge suppression andWILL block X-10 control signals.Do not plug the X-10 power lineinterface module into a poweroutlet that is powered by theUPS, unless all the X-10 controlmodules are also powered by the
UPS as well.
CAUTION
-
8/13/2019 Octopus 3000 Manual v 22
15/113
CHAPTER1: INTRODUCTION 1-7
Introduction
Your Octopus is a precisionlaboratory instrument thatcould be damaged by unex-pected power surges or fluctua-tions.
The use of a quality surgeprotection device is highlyrecommended, and is requiredto preserve your warranty.However, DO NOT USE a surgesuppressor between an Octo-pus power line interface and acontrol module.
CAUTION
The proper and reliable function-ing of the Octopus and the X-10control system is dependent onthe proper installation of theOctopus, the X-10 Power LineInterface, and each ControlModule. Other electrical devices,such as motors, ballasts, anddimmer switches produce electri-cal interference that may affectX-10 signaling. Refer to Chapter
5, Peripheral Device Trouble-shooting, for further details.
CAUTION
will continue to operate normally until thebattery is exhausted.
If a UPS system is used as a backup powersource, it may be purchased from Aquadyne.A UPS backup is a good idea if you requiremore than 45 minutes of backup power, or ifyou wish to backup other critical peripheralor communications equipment such as apump, a bio-filter or a modem for example.
Note however, that the duration of backup ofthe UPS is reduced by the addition of suchequipment.
Control Capability
H2O/S software is used to program the
Octopus 3000 and give it the capability for
parameter control as well as monitoring.
With the purchase of the optional X-10
power line interface and one or more control
modules, a wide variety of monitoring and
control features are automatically activated.
The Octopus 3000 control system has six basic
components, as shown in Figure 1.4:
The Octopus 3000 Controller
Wall Mount Power Supply
Probes
The X-10 Power Line Interface
One or More Control Modules
A Modem
(Note: The modem, the X-10 power lineinterface, and the control modules areoptions.)
Control Using X-10 Management
The probes gather information and relay it tothe Octopus controller (Figure 1.4). Thecontroller then sends commands through the
X-10 power line interface to the controlmodules. The control modules receive signals
from the Octopus via the X-10 power line
-
8/13/2019 Octopus 3000 Manual v 22
16/113
CHAPTER1: INTRODUCTION1-8
Introduction
interface turning the devices on or off asneeded. Each device that the Octopuscontrols, such as a heater or chiller, has itsown control module. As many as 256 ad-dresses and a variety of devices can becontrolled by the Octopus, using X-10 controlmodules
The Octopus 3000 incorporates severalmechanisms to protect your aquatic environ-ment from controller failure. For example,the controller is engineered to overcomeproblems associated with power fluctuations
through the periodic re-sending of controlmodule commands to make sure that a X-10control is latched in the correct state.
In addition to controlling peripheral equip-ment to maintain water quality, the H
2O/S
software supports other optional featuresthat can be enabled on the control menus:
A Lighting System
System Hold Management
A WaveMaker System
An Alarm System
Each of these features is described in follow-ing sections.
Figure 1.4 Octopus Control System Components
Probe Cables
X-10
PowerSupply
SurgeSuppressor
Device(E.g. Pump)
OctopusControlModule
OctopusControlModule
OctopusControlModule
Octopus 3000
Device(E.g. Chiller)
Device(E.g. Heater)
ModemX-10
PowerLine
Interface
Octopus PowerTransformerPlugged Into
Surge Suppressor
-
8/13/2019 Octopus 3000 Manual v 22
17/113
CHAPTER1: INTRODUCTION 1-9
Introduction
Example Lighting Scheme24-hour period. The Octopus 3000 comesstandard with 4-channel lighting software. Aoptional 16 channel system is available as anoption.
With separate lighting timers, there is a highdegree of flexibility in setting up your lightingsystem. For example, in your application youmay wish to use a combination of metalhalide, fluorescent, and incandescent light-ing. And each lighting system may require adedicated ON and OFF time. The Octopus3000 lighting system can manage these and
other varied timing needs under softwarecontrol. (Figure 1.5).
System Hold
The H2O/S system hold software is a featuredesigned to aid in the feeding of fish andinvertebrates. When the Octopus controller isput into feed mode, all systems, except theLighting and WaveMaker systems, are shutdown for a period of time selected by the user.At the end of the specified time (up to 60minutes) all systems are turned back on andwill restart within 5 minutes. If desired, the
system hold option may be turned off bysetting the feed time to 0 minutes. TheOctopus controller will then no longerprompt you to enter the system hold mode.
WaveMaker System
The H2O/S Dual Channel WaveMakersoftware feature allows the Octopus 3000 tocontrol two independent wave or surgegeneration systems. Each system supportstwo pumps which are driven by ControlModules. The cycle time for each pump isselected by the user, from 1-999 minutes (over
16 hours).
Lighting System
(4-Channel or Optional 16-Channel)
The lighting control feature allows theOctopus 3000 to control your system lighting.The lighting system will support up to 16
independent lighting times per 24 hours.Each channel has one ON and OFF time per
Figure 1.5 Example Lighting Scheme
Timer 1 is for the fluorescent lamps.These lamps are the first lights to comeon in the morning and the last to go offlate at night.
Timer 2 is for the metal halide lamps.To approximate the sun rising, theselamps come on in the late morning andgo off in the late afternoon.
Timer 3 is for an additional halide
lamp. This lamp will come on for aboutone to two hours at noon to simulatethe midday sun.
Timer 4 is for a red bulb during thenight to allow observation of nocturnalcreatures that come out when the lightsare out, and or moonlight simulation.
If you have the 16-channel LightingSystem and are not using all 16 chan-nels, you can use the additional chan-nels to control other devices.
-
8/13/2019 Octopus 3000 Manual v 22
18/113
CHAPTER1: INTRODUCTION1-10
Introduction
Alarm System
The H20/Salarm system software provides
visual and audible notification in the eventthat a monitored parameter exceeds the userdefined range. The optional pager alarmsoftware upgrade used in conjunction with amodem provides remote alarm notification viaa digital pager.
-
8/13/2019 Octopus 3000 Manual v 22
19/113
Chapter
2Installing
the
Components
-
8/13/2019 Octopus 3000 Manual v 22
20/113
CHAPTER2: INSTALLINGTHECOMPONENTS2-2
Installing the Components
This Chapter will provide you with simplestep-by-step instructions to lead you throughthe installation process. But beforecontinuing...
STOPREAD THIS FIRST
DO NOT ALLOW CONTROLLERTO GET WET!
Your Octopus 3000 controller can belocated in any dry location. The controlleris not waterproof and cannot be sub-merged, permitted to get wet or exposedto high humidity. To avoid splashing theunit, install it properly as described inthis chapter.
AVOID DANGEROUSOPERATING CONDITIONS
Do not operate the Octopus controllerunder any of the following conditions:
With wet hands.
If you are standing in water.
If the AquaNode Controller is wet.
If there is salt creep on the box.
KEEP pH AND ORP PROBE TIPSWET AT ALL TIMES
To prevent loss of accuracy in the pH andORP probes, be sure to keep the tips wetat all times. NOTE: Do not completelysubmerge probes. Refer to probe installa-tion procedures (Step 2) in this chapterfor more information.
INSTALL ALL ELECTRICALCORDS CORRECTLY
Always take care when using electricityaround water. To prevent water from
running down the cord of an aquariumdevice, always use a drip loop as shownin Figure 2.1. To create a drip loop, makesure that the cord extends below theoutlet so that any runoff will not enterthe outlet or the device. Also, the Octopuscontroller and ALL other aquatic environ-ment devices should be plugged into aGFCI outlet.
If you are n ot famil iar with ins tal lation
of electrical equipment, i t is strongly
recom me nded that you hire a l i censed
electrician.
CAUTION
-
8/13/2019 Octopus 3000 Manual v 22
21/113
CHAPTER2: INSTALLINGTHECOMPONENTS 2-3
Installing the Components
Figure 2.1 Protective Drip Loop and GFCI Outlet
In addition to a surge suppressor, a Ground Fault Circuit Inter-rupter (GFCI) should be part of you installation. A GFCI is a sensi-tive device which immediately cuts off the electricity flow when anelectric current leak is detected. All devices t hat are part of t heOctopus 3 00 0 s ystem should be conn ecte d to a GFCI protect ed
circuit. If you do not have one installed already, you should pur-chase one. They are inexpensive and are available at any electronicsor hardware store. Dont wait, a GFCI can save your life!
CAUTION
-
8/13/2019 Octopus 3000 Manual v 22
22/113
CHAPTER2: INSTALLINGTHECOMPONENTS2-4
Installing the Components
INSTALLING THE COMPONENTS
There are two basic parts to installing theOctopus 3000 controller:
Installing the Octopus hardware andadditional components(Chapter 2, Steps 1 through 8).
Programming the Octopus (H2O/S)
software(Chapter 3, Steps 9 through 25).
Not all systems require the same installation
steps. The steps required for each individualsystem will depend on:
Whether the system is monitoring onlyor monitoring and control.
Which optional components are in-cluded in the system.
Table 2.1 indicates which componentsinstallation procedures your system requires.For example, for Monitoring Only systems,only installation steps 1, 2, and 7 would berequired.
STEP 1MOUNT THE OCTOPUSCONTROLLER
The Octopus controller will need to beinstalled within 5 to 10 feet of the probes. Ifyou wish to place the controller beyond thestandard 10-foot length of the probe cables,additional cable extension are available in avariety of lengths through special order.
The controller comes with two mountinghangers for easy installation. Use the screwsprovided to attach the controller to the wall.
The ORDER OF INSTALLA-TION is important to success-ful startup and operation ofthe system. The X-10 ControlInterface, Satellite Modules,and Modem must be installedbefore the Octopus controlleris plugged in to ensure properinitialization of thesecomponents.
CAUTION
Table 2.1 Installation of Components
Steps to Install Physical ComponentsMonitoring
System
Monitoring &Control System
1. Mount the Octopus Controller
2. Install the Probes
3. Install the Digital Input hardware N/A
4. Install the Optional X-10 Power Line Interface N/A
5. Install and Address Optional Control Module(s) N/A
6. Install the Communications Equipment Optional Optional
7. Power Up the Octopus Controller
8. Install 9-Volt Backup Battery Pack Optional Optional
-
8/13/2019 Octopus 3000 Manual v 22
23/113
CHAPTER2: INSTALLINGTHECOMPONENTS 2-5
Installing the Components
STEP 2INSTALL THE PROBES
Up to four probes can be installed as part ofthe Octopus 3000 system. Three of theprobes (temperature, pH, ORP) have the samebasic installation requirements. Theconductivity probe, requires specialinstallation procedures to ensure properoperation and accuracy.
In all cases, it is desirable to locate the probesaway from sources of electrical interferencesuch as UV sterilizers, florescent lighting,ballasts, pumps, etc.
PROBE UNPACKING
pH and ORP probe tipsmust be kept wet at alltimes. The probes should beinstalled immediately aftertheir tips are exposed to air.If probes are allowed to dry,they may be damaged andmay produce inaccuratereadings.
To prevent damage to the ORP probesensor, FIRST REMOVE THE TAPEOVER THE VENT SEAL on theprotective cap. Then remove theprotective cap.
Remove the protective caps from theremaining probes and install perStep 2 instructions.
OTHER CONSIDERATIONS
pH and ORP probes can require a
minimum of 48 hours to stabilizeafter installation. DO NOT ENABLEOCTOPUS CONTROL OF THESEPARAMETERS FOR 48 HOURS.
Be sure probes are plugged into thecorrect Octopus probe ports. If theyare plugged into the wrong ports, thesystem will not monitor properly.
Place temperature, pH, & ORP probeswithin 6 inches of each other.
The conductivity probe should belocated at least 18 away from thetemperature probe.
CAUTION
Note
The Octopus is a precision
instrument that has beendesigned to interface with high-quality laboratory grade probesmanufactured in the USA toAquadyne specifications. The useof other probes (especiallyinexpensive imported probes) notapproved by Aquadyne is notrecommended and will not besupported.
A genuine Aquadyne probe canbe identified easily by thepresence of date code stamped
on the top of the probe shaft.
-
8/13/2019 Octopus 3000 Manual v 22
24/113
CHAPTER2: INSTALLINGTHECOMPONENTS2-6
Installing the Components
4. Place the probes in the location selectedin Step 1. None of the probes can besubmersed. Mount the probes so thatthe top of the probe is above the waterline. The probes should be positioned sothat ONLY ABOUT HALF of the probeshaft is immersed in water.
Alternative Probe Locations
Select a probe installation that will allow thepH and ORP probes to be located within 6inches of the temperature probe.
Three possible probe installations are:
In-Filter Location Using Probe Mounts(Figure 2.2).
In-Filter Location Using a Float(Figure 2.3).
In-Line Location Using Tee-Fittings(Figure 2.4).
Figure 2.2 In-Filter Location Using Probe Mounts
Installing the Temperature, pH, andORP Probes
The four steps in probe installation are:
1. Select a location for the probes that willallow the pH and ORP probes to beplaced within 6-inches of the tempera-ture probe. (Alternatives are describedin this section). By placing the probesclose to each another, the Octopus isassured that the probes are not subjectto different electrical environments that
could affect their accuracy. Theconductivity probe is a special case andshould NOT be placed with the otherprobesif at all possible The diagrambelow shows the conductivity probeplaced in-line back to the pump.
2. Remove the vent seal on the ORP probeprotective cap, then remove all protec-tive caps.
3. Plug the probe connector cables intothe designated port at the bottom of theOctopus 3000 controller.
Temp.
pH
ORP
Octopus3000
ConductivityCell
ConductivityProbe
ToPump
Note: Conductivity Probe shown Mounted inline.
-
8/13/2019 Octopus 3000 Manual v 22
25/113
CHAPTER2: INSTALLINGTHECOMPONENTS 2-7
Installing the Components
plumbing, it has been found to be a veryaccurate and safe way to place your probes.In-line placement will give accurate readingsbecause water is moving across the probe tipsat all times, and because algae growth on theprobe tips is minimized.
The probe is placed into a tee-fitting, asshown in Figure 2.4. You will need to providethe tee-fitting for this installation. Thediameter of the in-line portion of the teeshould be 1/2 inch (3/4 for conductivityprobe) regardless of whether you are using
laboratory or standard grade probes as theyare supplied from Aquadyne. Aquadyne offerstwo compression fittings for installation ofthe probes. Flow rates across the pH andORP probes should be limited to < 10 GPM.
Be sure to install water shut-off valves, asshown in Figure 2.6, to permit probe replace-ment. All fittings must be secured so there is nowater leakage.
It is important that the in-line probe extendno more than about 50% into the flowstream. If the probe protrudes too far into theflow stream it will interrupt flow and interferewith the accuracy of the probe readings.
In-Filter Location Using Probe MountsThemost common place for probe location is inthe filter. Installing probes in a filter or sumprequires that the probes be secured usingprobe mounts that attach to the back of thefilter so the probes are only half-immersed inthe water, (See Figure 2.2.) When placingyour probes, be sure that there is adequatewater flow over the probe tips at all times.The pH and ORP probes must be placedwithin 6 inches of the Temperature Probe.
In-Filter Location Using a FloatThe float
installation shown in Figure 2.3 keeps probesfrom being submerged (or drying out), and iseconomical to construct.
To construct a probe float, punch a hole(slightly smaller than the probe diameter)through a small piece of styrofoam. Be surethe probes are no further than 6 inchesapart. Insert the probes into the float, asshown in Figure 2.3.
In-Line Locat ion Using a Tee -FittingAlthough in-line probe placement is a littlemore difficult and requires some extra
Figure 2.3 In-Filter Location Using Float
Tem
p
pH O
RP
Probe
ShutoffValve
ShutoffValve
CompressionFitting
Fig.2.4 In-Line Location Using Tee Fitting
-
8/13/2019 Octopus 3000 Manual v 22
26/113
CHAPTER2: INSTALLINGTHECOMPONENTS2-8
Installing the Components
Installing the Conductivity Probe
Installation of the conductivity probe issimilar to the installation of the pH, ORPand temperature probes discussed previously.
1. Select the proper probe location awayfrom the other probes.
2. Remove the probe protective cap.
3. Plug the probe connector cable into theconductivity port at the bottom of theOctopus 3000 controller.
4. Place the probe in the location selectedin Step 1.
Alternative Probe Locations
The conductivi ty probe must be located
away from the other three probes, and the
conductivi ty probe m ust be inst al led in a
control led volume environm ent cal led a
conductivi ty ce l l .(See Figure 2.5)
The conductivity cell provides the conductiv-ity probe with a controlled volume environ-ment. The conductivity cell is required to
achieve accurate conductivity readings. Itreduces the chances that the conductivityprobe will be affected by alternate ground(conductive) paths.
The conductivity probe can be installed ineither of two ways:
In-line, using a tee-fitting.
Off-line, using a Drip Cup.
In-Line Location
The in-line installation shown in Figure 2.6is the preferred installation method becausethe tee-fitting is a natural conductivity celland it provides the most accurate measure-ment environment.
The general procedure is the same as thatdescribed above for temperature, pH and ORPprobes. However, the conductivity cell, isutilized. Be sure that the installation providesthat about 10% of the probe protrudes intothe cell. Aquadyne provides a score-mark onall of its conductivity probes that guidesinstallation and assures proper insertiondepth.
If the conductivity probe is notinstalled in a conductivity cellor if the probe is exposed toelectrical interference frompumps, motors, or electricalballasts, unreliable readings
will result.
Note
Probe
ConductivityCell
Figure 2.5 In-Line Conductivity Cell
-
8/13/2019 Octopus 3000 Manual v 22
27/113
CHAPTER2: INSTALLINGTHECOMPONENTS 2-9
Installing the Components
Off-Line Location
The Drip Cup installation is a secondoption for the conductivity probe installation.As illustrated in Figure 2.6, a portion ofwater from the pump discharge line isdiverted to a small cup-type arrangement.
The water flows in and out of the cup at arate of about 1 cup per minute.
In a drip cup installation, there is no com-pression fitting to control insertion depthyou must carefully position the probe so thatthe end of the probe is approximately 1 inchfrom the bottom of the cup.
Stabilization Period for Probes
After the probes are installed, there is aperiod of time required for each probe tostabilize in its new environment. The conduc-tivity probe will be the first to stabilize in only5 minutes, followed by the temperature probewhich requires about 10 to 15 minutes before
accurate readings are possible. The pH andORP probes will take considerably longer tostabilize. Their readings are much moresensitive. The pH probe will stabilize in about24 hours while the ORP probe may require 48hours.
STEP 3INSTALL DIGITAL INPUT
HARDWARE
Various system conditions, such as high orlow water levels, high or low pressures, flowfailure, or power failure can be detected andalarmed using digital input hardware. Thesedevices use a switch and relay that detect andalarm when an electrical circuit changes fromopen-circuit to closed-circuit position, or viceversa. This section provides instructions forinstalling a float switch (water level detec-tion) and a power fail detector.
Figure 2.6 Drip Cup Installation
, , ,
, , ,
, , ,
, , ,
, , ,
, , ,
, , ,
, , ,
, , ,
, , ,
,
,
,
,
, , ,
, , ,
, , ,
, , ,
, , ,
Versel'Octopus
Pompe
Vase
Filtre ac
Valve
Sonde deConductivit
Installation "Vase d'Ecoulement"
The Octopus has been cali-brated at the factory to pre-setstandards. Since each probehas its own unique measuringcharacteristics, it is alwaysrecommended that you per-form a manual calibrationallowing sufficient time for theprobes to stabilize in thecalibration solution.
Note
-
8/13/2019 Octopus 3000 Manual v 22
28/113
CHAPTER2: INSTALLINGTHECOMPONENTS2-10
Installing the Components
Aquadyne offers digital input devices forwater levels, pressures, flow, and power faildetect.
Aquadyne offers an optional 8-pin connectorand wire that can be used to hook up amiscellaneous digital input.
Installing a Float Switch
A float switch consists of a stationary stemand a float that moves up and down with thewater level. When the float is in the raised
position, it is in the CLOSED position. Whenthe float is lowered, it is in the OPEN posi-tion.
Install the Float Switch in Tank or Filter
The float switch should be installed in thetank or filter at the desired level, according tothe manufacturers instructions. (Many floatswitches will require drilling into the tank atthe desired float location.) Be sure the floatis installed in the position shown in Figure2.7 (with the arrow on the side of the switch
pointing up).
Once the float switch has been installed inthe tank or filter, connect the 8-pin connec-tor to the bottom of the Octopus (in the 8-pin digital input port).
Installing the Power Fail Detector
A power fail detector is available from Aqua-dyne. This 9-volt wall mount device isconfigured with an 8-pin connector, as shownin Figure 2.8. Install the unit as follows:
1. Plug the 9-volt power detect switch into
a 110-V outlet.
2. Plug the 8-pin connector into thebottom of the Octopus. (At the 8-pinGeneral Input port.)
Figure 2.7 Float Switch with 8-pin Connector
Figure 2.8 Power Fail Detect Installation
Probe Cables
X-10
PowerSupply
SurgeSuppressor
Octopus 3000Modem X-10
ControlModule
Octopus PowerTransformerPlugged Into
Surge Suppressor
9-VoltPower FailDetector
-
8/13/2019 Octopus 3000 Manual v 22
29/113
CHAPTER2: INSTALLINGTHECOMPONENTS 2-11
Installing the Components
One Control module will be needed for eachperipheral device you want to control, asshown in Figure 2.9.
To install the Control module, simply plugthe module into a GFCI-protected electricaloutlet. Then plug the device to be controlled(i.e., pump, heater, chiller) into the bottom ofthe Control module.
Be sure you do not exceed the power specifi-cations of the Control module. The Controlmodules operate on power lines up to 110
volts and can be used with devices up to 15amps. Control modules which operate withhigher voltage and amperage or with differentinternational plug configurations, are avail-able from your dealer.
The Octopus identifies each Control moduleby its own code, which is called an address. Acontrol address must be set for each Controlmodule.
The address is selected using the two dials onthe face of the Control module, as shown inFigure 2.10. The alpha dial allows selectionsA-P and the numeric dial allows selection 1-16. The Control address shown in Figure 2.10is A-3.
STEP 4INSTALL THE OPTIONALX-10 POWER LINE INTERFACE
To install the X-10 Power Line Interface,simply plug it into a GFCI electrical outlet.Then use the 4-wire interconnect cableprovided with your unit to connect the X-10Power Line Interface (RJ-11 port at thebottom) to the Octopus 3000 controller. Forproper initialization of these components,remember to connect your X-10 Power LineInterface and Control modules before power-ing up your Octopus 3000 unit.
STEP 5INSTALL AND ADDRESSOPTIONAL CONTROL MODULE(S)
The Control modules can be located wherever
electrical power is available. It is essential,
however, that the Octopus 3000 controller,
the X-10 Power Line Interface, and all
Control modules be located on circuits on the
same phase in the electrical sub-panel and
have the distance of the A/C wiring between
the X-10 Power Line Interface and any
Control Module not exceed 50 feet. Other-
wise, these devices may not communicateproperly.
Figure 2.9 Control Module Installation
1
5
9
15 3
711
13
AC
E
GI
K
M
O
UNIT
ID
AQUADYNETM
TM
SATELLITEControl Module
GFCI
Heater1
5
9
15 3
711
13
AC
E
GI
K
M
O
UNIT
ID
AQUADYNE TM
TM
Control Module
1
5
9
15 3
711
13
AC
E
GI
K
M
O
UNIT
ID
Adresse: A-3
Figure 2.10 Addressing a Control Module
-
8/13/2019 Octopus 3000 Manual v 22
30/113
CHAPTER2: INSTALLINGTHECOMPONENTS2-12
Installing the Components
Before assigning any addresses, you shouldmake a list of all devices to be controlled andtheir proposed Control addresses. It isrecommended that you use the addressessupplied by Aquadyne. They will minimize therisk of control problems caused by interfer-ence. The default addresses are shown inTable 2.2. The Octopus 3000 comes pre-programmed with these addresses alreadyinstalled in the controller.
A sticker on each satellitemodule indicates the maximumload-rating of the device.
DO NOT OVERLOAD. Equip-ment such as compressors andpump motors can exceed theload rating of a Control module.Verify with the manufacturer ofyour peripheral device that thesurge current at startup doesnot exceed the load rating of thesatellite module. Damage to thesatellite unit could result!
In addition, be sure you areusing the correct plug configu-ration.
CAUTION
The X-10 Power Line Interfaceand all Control modules must beinstalled on circuits of the SAMEPHASE. If the X-10 Power LineInterface and the Controlmodules are not on the samephase, they will not be able tocommunicate with each other,except by means of a X-10Signal bridge.
Do not plug Control Modules or
the X-10 Power Line Interfaceinto power strips equipped withsurge suppression. Surgesuppression will filter out theOctopus X-10 control commandsignals.
Note
Table 2.2 Default Control ModuleAddresses
H-1 Heater C-1 Chiller O-1 Ozonizer D-1 Buffer System (pH)D-2 CO2 Doser F-1 Main Pump (System Hold)A-1 Alarm ModuleC-2 Low Cond. ControlC-3 High cond. Control
P-1 WaveMaker Pump 1AP-2 WaveMaker Pump 1BP-3 WaveMaker Pump 2AP-4 WaveMaker Pump 2B
L-1 Timer 1L-2 Timer 2L-3 Timer 3L-4 Timer 4L-5 Timer 5L-6 Timer 6 L-16 Timer 16
Default Control Addresses
-
8/13/2019 Octopus 3000 Manual v 22
31/113
CHAPTER2: INSTALLINGTHECOMPONENTS 2-13
Installing the Components
STEP 6INSTALL THECOMMUNICATIONS EQUIPMENT
There are two optional communicationfeatures available for the Octopus 3000system:
Alarm Paging
AquaWeb for Windows 95
The alarm paging feature allows your pagerto receive an alarm page when a water qualityparameter goes outside the acceptable rangeor when a digital input switch signals analarm condition. The pager message indicateswhich Octopus 3000 unit (via the last 3 digitsof the unit serial number) is affected (in theevent of multiple units), which parameter isinvolved, and whether the alarm represents ahigh or low condition. (Refer to Chapter 3,Step 23 for a discussion of pager messageinformation.)
As shown in Table 2.3, in order to have thealarm paging feature, you must install amodem to your system.
AquaWeb for Wind ows 9 5 provides remoteaccess to the Octopus 3000 internal data-logand permits network communication betweena PC and the Octopus. Depending on thesoftware version of AquaWeb, some of thefeatures include remote communications,trend analysis and networking with extensivegraphing, statistical and data export capabili-ties.
As indicated in Table 2.3, there are two waysto access AquaWeb:
Via modem Via direct cable connection
For a single PC to a single Octopus, a directcable connection can be used (refer to Figure2.11). Note that if a direct cable connection isused, alarm paging will not be available.(Both the direct cable connection and themodem connection require a serial portconnection. And PCs typically have only oneserial port available.)
For remote access to the Octopus 3000 data-log or for networking two or more Octopus
units, a modem connection is required.
Table 2.3 Communication Options
Communication FeatureConnection Options Available
ModemDirect CableConnection
Alarm Paging Only
AquaNet Only
Both Alarm Paging & AquaNet
N/A
N/A
-
8/13/2019 Octopus 3000 Manual v 22
32/113
CHAPTER2: INSTALLINGTHECOMPONENTS2-14
Installing the Components
Installing the Modem
Using the AquaWeb software requires aAquadyne modem. Non Aquadyne modems arenot supported by Aquadyne. The Aquadynemodem and 10-foot modem cable (RS-232)are preconfigured to work with the Octopus3000 unit and there are no adapters needed.If you did not purchase this hardware fromAquadyne you will need to follow additionalinstructions to connect your modem with theOctopus 3000 unit. (A Tech- Tip is availablefrom Aquadyne.)
1. Connec t the Modem and the OctopusUnitFirst, plug your modem cable intothe modem/serial port of the Octopus3000 unit. Then connect the other endof the modem cable into the 25-pinconnector on the back of the modem, asillustrated in Figure 2.12.
The connectors are keyed and can onlybe installed one way. Make sure that theplugs on the cable are fully inserted intothe jacks on the Octopus controller and
the modem.
Figure 2.11 AquaNet using Direct Cable or Modem Connection
Direct Cable Connection for Local Use Only
Modem Connection for Remote or Networking Access
-
8/13/2019 Octopus 3000 Manual v 22
33/113
CHAPTER2: INSTALLINGTHECOMPONENTS 2-15
Installing the Components
The Aquadyne modems default settingson the dip-switch (at the back of themodem) are DUDUUUDD for switchpositions 1-8 respectively. Verify that thedip-switch settings are in the correctpositions.
2. Power up the ModemAquadynemodems are powered by a small wall-mount transformer, as shown in Figure2.12. Plug the end of the cable from thepower transformer into the front of themodem. Then plug the modem walltransformer into a wall receptacle or,
preferably, into a surge suppressorpowerstrip.
If you have an Aquadyne modem, theserial port configurations are already setto work properly with your Octopus3000 unit. If you do not have anAquadyne modem, you will need toconfigure the serial port settings on themodem. Normally, this will require thatyour settings support the port settings ofyour Octopus 3000 (baud rate of 19,200and data format of 81N). If you are usingAquaWeb, you will also need to
configure the modem to auto-answer.
A small power switch is located on theback of the modem. Set the power switchon the modem to the ON position. Whenthe modem has power, several red lightsshould be illuminated on the front of thedevice.
3. Connec t the Modem and PhoneThemodem is supplied with a 6-foottelephone cable to connect the modem tothe telephone company (telco) phoneline. Plug the phone cable into the backof the modem in the jack labeled LINE.(DO NOT plug the cable into the jacklabeled PHONE.) Connect the other endof the phone line to the closest telco livetelephone jack.
If your installation does not use thistype of phone connection, adapters areavailable through your local computerhardware store. With the cableconnected to the phone line, you arenow ready to program the H
2O/S
software for modem use. (Refer toChapter 3, Step 23.)
Modem/ SerialPort
Modem
Wall Mount
Power Supply
TelephoneLine
Figure 2.12 Connecting a Modem
-
8/13/2019 Octopus 3000 Manual v 22
34/113
CHAPTER2: INSTALLINGTHECOMPONENTS2-16
Installing the Components
To hook up directly to the PC, one end of thedirect cable connects to the serial port on theOctopus 3000 unit and the other end con-nects to your PC com port. Aquadyne providesa optional 25 and 50-foot extension cablesfor longer runs.
Installing the Direct Cable Connection
Your Octopus 3000 Controller with its H2O/S
system software can be configured to connectdirectly to a PC by using Aquadynes directcable connection. Once connected, theOctopus 3000 can communicate withAquaWeb for Windows 95, Aquadynes remoteaccess application, and the H
2O/S Flash
downloader application.
CAUTION
There could be problems inmodem operation and pagerfunctionality if the telephoneline is shared. For properoperation, the modem willrequired a dedicated tele-phone line.
Use of any cable other thanthat supplied by Aquadynemay result in an incompleteconnection or, in extremecases, damage to the Octo-pus or computer.
The cable utilized for DirectCable connection is a non-standard cable. It can beordered from your dealer orfrom Aquadyne.
CAUTION
Figure 2.13 Bottom View of the Octopus Controller
Power
(9V-AC Power Jack)
ORP Probe (BNC)
Modem/Serial Port
(8-pin mini-DIN)Conductivity
(4-pin mini-DIN)
General Input(8-pin mini-DIN)
pH Probe(BNC)
Temp. Probe
(Telco)
X-10 Control
Module (Telco)
9V Battery B/U
(Stereo jack)
-
8/13/2019 Octopus 3000 Manual v 22
35/113
CHAPTER2: INSTALLINGTHECOMPONENTS 2-17
Installing the Components
Along with the cabling issue the Octopus3000 allows you to control a number ofdifferent serial data options such as baudrate, parity, and software flow control.
STEP 7POWER UP THEOCTOPUS 3000 CONTROLLER
If your system includes a modem, X-10 PowerLine Interface or Control modules, be surethat these components are plugged in BE-FORE powering up the Octopus 3000.
Plug the end of the cable from the powertransformer into the power jack located onthe bottom left of the Octopus 3000 (See
Figure 2.13). Be sure to use the protectivepower cord drip loop to prevent water fromrunning down the cord into the Octopus 3000controller. The cord should extend below theoutlet so that any runoff will not enter theelectrical outlet.
When the controller is first plugged in, it goesthrough a diagnostic self check. The LCDscreen will show a series of messages indicat-ing that a self diagnosis is in progress, asshown in Figure 2.14.
When powering up the Octopus for the firsttime, make note of your system serial numberand H2O/S software version. This informa-tion is required should you find it necessaryto obtain technical support. When thecurrent status screen begins to display youwill notice that the top line of the display willshow the current probe readings. The secondline will show system activity such as sys-tems being activated, and alarms occurring.When the Current Status Screen is shown,Figure 2.15, you are ready for programmingthe Octopus controller.
Figure 2.14 Startup Diagnostics
OCTOPUS 3000Aquadyne Inc.
InitialisationModem: +++
Auto Test ok
Ver. 2.0Ser #97-1234567
Version du Logicielet numro de serie
Installatin/Detectionde l'Interface et de Modules
Diagnosticsde dmarrage
Copyright (C)1992-1997
Auto Test
Initialisationsorties num.
Initialisation
Modem: Envoie
Initialisation duModem
InitialisationTermine
Figure 2.15 Current Status Window
Rdox= 350mV
08/15/97 11:45AMpH = 7.8
CD 45.4mS
Tous Systmes OK
-
8/13/2019 Octopus 3000 Manual v 22
36/113
CHAPTER2: INSTALLINGTHECOMPONENTS2-18
Installing the Components
CONGRATULATIONSYou have now com-pleted the first half of the Octopus installa-tioncomponents installation
In Steps 1-8, you mounted the Octopus 3000and installed the following components:
Probes (Temperature, ORP, pH andConductivity).
Digital Input Switch or Device (op-tional).
X-10 Power Line Interface (Optional).
Control Module(s) (Optional).
Communications Equipment (Optional).
You then powered up the Octopus andinstalled the Backup Battery Pack.
You are now ready for the second half of theinstallation process: programming theOctopus H
2O/S software (Steps 9-25),
presented in Chapter 3.
Dont worry, you dont have to be a computerprogrammer to successfully program theOctopus. We not only give you the roadmap
(the Menu Map), we guide you down the roadas well.
Please note that the 9-VoltBackup Battery incorporatesspecial hardware designedby Aquadyne specifically forthe Octopus. The use of anyother voltage source couldcause damage to yourOctopus.
CAUTION
STEP 8INSTALL THE 9-VOLTBACKUP BATTERY PACK(OPTIONAL COMPONENT)
To install the 9-Volt Backup Battery Pack,simply plug in the battery backup to thebottom of the Octopus controller.
Note: You need to install two 9V Alkalinebatteries into the battery backup pack. Donot use rechargeable batteries.
Figure 2.16 Battery Backup Installation
Cables de Sondes
Octopus
9V BatteriePack
TransformateurBranch dansSuppresseurde Tension
-
8/13/2019 Octopus 3000 Manual v 22
37/113
Chapter
3Programming
the
Software
-
8/13/2019 Octopus 3000 Manual v 22
38/113
CHAPTER3: PROGRAMMINGTHESOFTWARE3-2
Programming the Software
PROGRAMMING THE SOFTWARE
Your Octopus 3000 includes the latest versionof the H
2O/S software which was pre-loaded
into FLASH memory at the factory prior toshipment. The FLASH memory design of theOctopus 3000 allows future software upgradesto be installed by the user as they are madeavailable by Aquadyne.
For additional information pertaining to thelatest software release, visit our WEB page:h t t p : / / w w w . a q u a d y n e .c o m
For this part of the installation process, werecommend that you have your laminatedQuick Reference Menu Map card at hand tofollow along with the description. This willfamiliarize you with the Quick Reference cardas well as introduce you to the use of theOctopus 3000 software menus. Detailed MenuMaps are given in Appendix A.
Before you begin entering setpoint data, wethought it would be beneficial to review thegeneral structure of the Octopus 3000 menus.
Overview of the OCTOPUS MenusAll software menus and data entries areviewed through the LCD Screen Window onthe face of the Octopus 3000 (Figure 3.1). Thescreen typically shown in the window is theCurrent Status Screen. This screen givescurrent probe readings for each of the fourparameters: conductivity, temperature, pHand ORP
Top Level Display
The Current Status Screen is only one of
several screens that comprise the Top LevelDisplay. Side One of the Quick Reference
Menu Map shows all available Top LevelDisplays.
To scroll through the Top Level Display,simply push the right or left arrow on thekeypad. The screens will be shown in the orderindicated on the Menu Map. You can cyclethrough the screens as many times as youdesire in either direction.
The Current Status Screen provides bothcurrent parameter information (i.e. pH, ORP,Conductivity, and Temperature) on the topline and system status on the lower one. Assystems are activated or alarms are triggeredthe status line provides up to date status onthe system being managed. As events occurthey are entered into the lower display. Up to16 events are saved and displayed in se-quence. Figure 3.2 shows the current statuswindow.
Figure 3.2 Current Status Display
ORP= 350mV
08/15/97 11:45AMpH = 7.8
CD 45.4mS
All Systems Normal
-
8/13/2019 Octopus 3000 Manual v 22
39/113
CHAPTER3: PROGRAMMINGTHESOFTWARE 3-3
Programming the Software
Main Menu
The Main Menu is comprised of threeprimary sub-menus, as shown in Figure 3.3
Setpoints Menu
Operation Menu
Setup Menu
These menus are the first level down from theTop Level Display and comprise the Main
Menu for the Octopus 3000. This menu is thedoorway to all other sub-menus. To enter theMain Menu, simply push the Enter key fromany position on the Top Level Display.
Note that from the Top Level Display you willalways enter the Main Menu at the sameplace: the Setpoints Menu.
From there, you can scroll left or right in theMain Menu as desired, using the right or leftarrows on the keypad.
Setpoint Menu Overview
As the name implies, the Setpoint Menuallows you to enter the setpoint values for thefour water quality parameters:
Temperature Setpoint
ORP Setpoint
pH Setpoint
Conductivity Setpoint
In addition to setpoint values, you can inputparameter setpoint ranges, and parametercontrol information for use with the X-10control system and alarm generation.
Operation Menu Overview
The Operation Menu includes options thatmight be required in day-to-day operation ofthe Octopus 3000:
Control Menu
Probe Calibration
Lock/Unlock Password
Clean Probes
IR Printer Menu
Setup Menu Overview
The Setup Menu is used primarily duringinitial setup of the Octopus 3000 controller,including:
Test Menu Set Module ID
New Password
Temp Scale
Timers
Alarm
Data Logger
Wave-Maker
Filter
Pager (optional)
Conductivity
Digital Input
Figure 3.3 The Main Menu
ExitMainMenu
SetpointMenu
OperationMenu
SetupMenu
TOP LEVEL DISPLAY
-
8/13/2019 Octopus 3000 Manual v 22
40/113
CHAPTER3: PROGRAMMINGTHESOFTWARE3-4
Programming the Software
Using the Menu MapAn Example
Suppose, for example, you wish to enter thepH setpoint for the Octopus 3000 . To do this,you would follow these steps (See Figure 3.4):
From anywhere on the Top LevelDisplay, press Enter . After theSystem hold prompt you will be atthe Setpoints Menu.
Press Enter at the Setpoints Menu.This will take you to Temperature
Setpoint. Scroll to the right using the right
arrow key until you reach pHSetpoint. Then press Enter.
You will then be at pH Setpoint.Move the arrows left or right to findthe desired pH value. Then pressEnter to record the Setpoint valueinto memory.
TOP LEVEL DISPLAY
ExitMainMenu
OperationMenu
SetupMenu
ExitSetpoints
ORPSetpoint
SetpointMenu
pH Setpoint
Setpoint=x.xx
Cond.Setpoint
pHSetpoint
TempSetpoint
Figure 3.4 Entering pH Setpoint
Remember that the Main Menu selections arethe doorway through which all other sub-menus are accessed. From the Top LevelDisplay, press the Enter key to get to the MainMenu level. Then simply scroll to the left orright, as desired.
Initial Software Setup
You can now begin programming the Octopus3000 H
2O/S softwaredeveloped by Aqua-
dyne specifically for advanced water qualitymonitoring applications. The H
2O/S software
setup consists of Steps 9-23 in the installationprocess. A summary of these steps is given inTable 3.1. Step 15 (Entering Setpoint Values)is perhaps one of the most important of thesesteps.
STEP 9SET THE SYSTEM CLOCK(SETUP MENU)
The Octopus 3000 is equipped with aninternal clock that includes its own batterysource. So, you wont need to set the clock as
-
8/13/2019 Octopus 3000 Manual v 22
41/113
CHAPTER3: PROGRAMMINGTHESOFTWARE 3-5
Programming the Software
were May 16, 1997, then you would enter "05/16/97." The controller will then prompt you toverify the date before continuing to enter thetime. If you answer NO to the question"Correct?," then you have an opportunity tore-enter the date.
Next, enter the time in a 12-hour format.After the time is entered you are again giventhe opportunity to verify that what you
entered is correct. If you answer NO, then youwill be given another opportunity to enter thetime.
a result of power failure. However, becauseeach clock is set from the factory with PacificStandard Time, you may need to adjust theclock to your time zone. Control of the clock isaccessed under the Setup Menu under Timer.
Your first option in setup is to load thecurrent date. Data is loaded in the followingorder: First, the month (01=January, 02=Feb-ruary. . ., 12=December), then the day of the
month, and finally the last two digits of theyear. To enter the year 1997, you would onlyenter "97." For example, if the current date
Table 3.1 Initial Software Setup
puteSerawtfoSlaitinI uneMnO gnirotinoM
metsyS
&gnirotinoM
metsySlortnoC
kcolCmetsySteS.9 puteS X X
drowssaPretnE.01 puteS lanoitpO lanoitpO
lortnoCretliFesioNteS.11 puteS X X
.pmeTytivitcudnoCehttceleS.21
sgnitteSnoitasnepmoC puteS X X
egnaRytivitcudnoCtceleS.31 puteS X X
elacSerutarepmeTtceleS.41 puteS X X
seulaVegnaR&tniopteSretnE.51
)PRO,Hp,.pmeT,.dnoC( tniopteS X X
&seulaVegnaRmralAehtretnE.61
remiTffO-dloH puteS X X
mralAtupnIlatigiDputeS.71 puteS X X
metsySgnithgiLputeS.81 puteS AN X
metsySrekaMevaWputeS.91 puteS AN X
dloHmetsySputeS.02 puteS AN X
erawtfoSnosDIlortnoCngissA.12 puteS AN X
mralAregaPehtputeS.22 puteS X X
lortnoCmetsySsupotcOelbanE.32 snoitarepO AN X
noitarugifnoCmetsySehttseT.42 puteS AN X
-
8/13/2019 Octopus 3000 Manual v 22
42/113
CHAPTER3: PROGRAMMINGTHESOFTWARE3-6
Programming the Software
STEP 10ENTER THEPROTECTIVE PASSWORD
The use of a password will protect yourcontroller from unauthorized use. Anyoneattempting to use the controller will not beable to enter the Main Menu without enteringthe correct password. Your password mayconsist of up to five characters, letters and/ornumbers. You may choose not to utilize thepassword protection; if so, do not enter anypassword and move to the next setup proce-dure.
Many people find password numbers easier toremember; you may want to choose a numberyou already use somewhere else, such as anaddress. To enter your password, go to theNew Password selection in the Setup Menu.
When you press Enter, you will be promptedto enter your password of up to five charac-ters. To do so, push the arrow until the firstcharacter you desire appears, then pushenter. Follow this process for the remainingcharacters of your password. If your pass-word has less than five characters, push enter
a second time to signal that your password is
complete. You will be prompted to enter thesame password a second time to ensure thatthe password has been entered correctly.When your password has been entered andverified the LCD Screen returns to the "ExitPassword" Screen.
In order to activate your password selection goto the Operations Menu, press Enter, thenscroll using the arrows to the "Lock-Unlock"menu. Push Enter and the display will show"Lock Now." Push Enter and you have acti-vated the password.
STEP 11SET THE NOISE FILTERCONTROL
The Octopus 3000 comes with a softwarenoise filter which minimizes the effects ofelectrical noise on probe measurements. Thefilter may be set to one of four states: Off, Low,Medium, or High. Unless your Octopus 3000is having problems recording data into thedata-log, we recommend an initial setting ofOff. If your aquatic environment is stable, butyour HP hand-held printer output is showingunusual high-low readings, then some
adjustment to the filter setting is warranted. Itis always advisable to first identify andeliminate the source of the interference before attempting to use the filter to achievethe same goal. The filter menu is accessedunder the Setup Menu.
Note that the noise filter does not alleviate theX-10 control problems associated with electri-cal noise. Contact your nearest X-10 controlmodule dealer who can help troubleshootinstallation problems.
Entering AQDYN at the newpassword prompt willdisplay the system serialnumber and the version ofH
2O/S software installed in
the Octopus. This informa-tion will be required shouldyou need to contact technicalsupport for any reason.
Note
-
8/13/2019 Octopus 3000 Manual v 22
43/113
CHAPTER3: PROGRAMMINGTHESOFTWARE 3-7
Programming the Software
STEP 12SELECT THECONDUCTIVITY TEMPERATURECOMPENSATION SETTINGS
Temperature will affect conductivity readings.Therefore a temperature compensationadjustment is provided. This adjustmentallows you to artificially adjust the conduc-tance reading in percent per degree Celsiusfrom 0 to 5% per C. As a general rule, anadjustment of 2-3% is common for the waterenvironment.
To reach the temperature compensationmenu (Temp Comp Menu), press Enter onthe Setup Menu, then press the arrow keysuntil you scroll to the Conductivity menu(Cond. Menu). When you press Enter again,you will be at the Temp Comp Adjust Menu.
NOTE: You can test the required percentadjustment setting by using water tempera-tures of known conductivity value andobserving the conductivity readings. Thereadings should stay constant as the waterchanges temperature.
STEP 13SELECT ACONDUCTIVITY RANGE
The conductivity range menu allows you toset the appropriate conductivity scale. This isimportant since the conductivity range ofwater varies greatly depending on the applica-tion. Saltwater, for example, ranges fromabout 45,000 to 55,000 S (45mS - 55mS),whereas freshwater conductivity is typicallybetween 800 and 1,500 S, depending on thewater source. Hydroponic and other applica-tions can vary across a broad range as well,depending on the crop and the stage of
growth.
You will need to select the conductivity scalethat is appropriate for your system. Theconductivity scale choices are 0-1000 S (lowrange), or 0-100 mS (high range). The firstscale is a much more sensitive scale since itreflects changes in conductivity of only onemicroSiemen (1 S). The second scale is lesssensitive and reflects conductivity changes inincrements of 100 S.
From the Setup Menu, press Enter to reachthe submenus. Use the directional keys to
scroll through the Sub-menu options. Once atthe conductivity range, press Enter.
UNEMPUTESEHTNO
deriuqeReulaV gnirotinoM
metsyS&gnirotinoMmetsySlortnoC
ytivitcudnoC
noitasnepmoCerutarepmeT X X
egnaRytivitcudnoC X X
mralAytivitcudnoC X X
erutarepmeT
elacSerutarepmeT X X
mralAerutarepmeT X X
Hp
edoMgnisoD X
mralAHp X X
PRO
mralAPRO X X
Table 3.2 Parameter Settings Included in the
Setup Menu
-
8/13/2019 Octopus 3000 Manual v 22
44/113
CHAPTER3: PROGRAMMINGTHESOFTWARE3-8
Programming the Software
STEP 14SELECT ATEMPERATURE SCALE
The temperature scale menu allows you toselect the scale you wish to use for tempera-ture measurementCelsius or Fahrenheit.The Octopus 3000 controller is set to Fahren-heit at the factory.
Changing the temperature scale automaticallyrecomputes your setpoints. For example, ifyour setpoints were entered in Fahrenheit andyou change the temperature scale to Celsius,
all of your setpoints will be automaticallyconverted to their Celsius equivalents.
The Temperature Scale Menu is accessed bypressing Enter at the Setup Menu. Use thedirectional keys to scroll through the menuoptions until the temp scale menu is reached.Press Enter to access the item, and use thedirectional keys to choose a temp scale.
STEP 15ENTER SETPOINT ANDRANGE VALUES
Various setpoints and ranges must be estab-
lished for each of the four monitoring param-eters: conductivity, temperature, pH and ORP
There are three values that must be entered:
The optimal setpoint value (in Step 15).
The control range (in Step 15.
The alarm range (in Step 16).
When these three values are entered, thesystem will operate as illustrated in Figure3.5. As long as the water quality parametersremain within the acceptable operating range,no control or alarm will be activated. If,however, the parameter rises above or falls
below the acceptable operating range, theOctopus 3000 will activate the appropriatecontrol device. The control device remains inoperation until the parameter returns to theSetpoint value. (For ORP, the control deviceremains in operation until the ORP valuereturns to the HIGH control point.)
Entering Data on the Setpoints Menu
Generally speaking, you enter or change anysetpoint by first pressing Enter from anyposition on the Top Level Display. This will
bring you to the Setpoints Menu. Press Enterand you will be at the Temp Setpoint menu.Scroll left or right using the arrows until youarrive at the desired parameter. Press Enter.When the current setpoint is displayed, pushthe right arrow to increase the value or theleft arrow to decrease the value. Once thesetpoint value you want is displayed, pushEnter once to register your selection. Table 33shows which parameters need to be set formonitoring and alarming functions and theadditional settings needed for adding control.
As you choose a setpoint value, check the
System Status light. The light should be solidgreen. If the light turns yellow as you areselecting your value, the setpoint is out of thenormal value range for that parameter. Adjustthe setpoint accordingly.
Conductivity Setpoints
To enter the conductivity optimal setpointvalue, press Enter from any place on the TopLevel Display. You will be at the SetpointsMenu. Press Enter and you will be at theTemperature Setpoints Menu.
Using the left arrow key, scroll to the conduc-tivity menu and press Enter. At the flashing
-
8/13/2019 Octopus 3000 Manual v 22
45/113
CHAPTER3: PROGRAMMINGTHESOFTWARE 3-9
Programming the Software
need to enter other values in the SetpointsMenu including dosing times for low and highconductivity, and dwell times.
A dwell time is the length of time that theOctopus 3000 will wait until dosing can occuragain. This feature prevents overdosing asystem with a dosing solution and protectsagainst short cycling of equipment.
Conductivity Example
Suppose that you desire an optimal setpoint
value of 45 mS, an operating range of 5 mSand an alarm range of 3mS. Figure 3.5illustrates the relationship between theoptimal setpoint value, and the control range.
For these systems, if the conductivity valueremains within the target operating range, theparameter status light will not be illuminated.If the conductivity value enters the controlrange, the parameter status light will turneither red if the low conductivity dosingsystem is enabled, or green if the high con-ductivity dosing system is active. If theparameter status light is flashing red, the low
conductivity dosing system is in a dwell state.A high conductivity dosing system dwell stateis indicated by a flashing green parameterstatus light (See Figure 1.2).
Acceptable
OperatingRange
Low
Control
ON
Low
AlarmON
HighAlarm
ON
HighControl
ON
OptimalSetpoint
Value
Lower
Parameter
ValuesHigh
Alarm
Setpoint
Low
Alarm
Setpoint
High
Control
Setpoint
LowControl
Setpoint
Higher
ParameterValues
Figure 3.5 Control Systems Setpoint and Ranges
screen, scroll to the left or right until thedesired conductivity value appears and press
Enter. This registers the optimal conductivitysetpoint value.
In a control application, in addition tooptimal setpoint value you will need to entera range value as well, in addition you will
Table 3.3 Parameter Setpoints
ON SETPOINT MENU
Value Required
Conduct iv ity
Conductivity Setpoint X X
Conductivity Range X
X
X
X
High Conductivity Dwell Time X
Temperature
Temperature Setpoint X X
Temperatrue Range X
Hp
X X
X
X
Dwell Time X
OC 2 Doser Time X
OC 2 Dwell Time X
PRO
ORP Setpoint X X
X
X
X
Dose Time
pH Range
pH Setpoint
High Conductivity Dose Time
Low Conductivity Dwell Time
Low Conductivity Dose Time
Monitoring
System
Monitoring &ControlSystem
-
8/13/2019 Octopus 3000 Manual v 22
46/113
CHAPTER3: PROGRAMMINGTHESOFTWARE3-10
Programming the Software
Temperature Setpoints
The temperature setpoint defines the opti-mum water temperature. In a control applica-tion you will also need to enter the controlrange on the Setpoint Menu.
To enter the temperature setpoint, go to theSetpoints Menu. Press Enter and you will beat the Temperature Setpoint Menu.
When you press Enter again the currenttemperature setpoint will begin flashing onthe LCD display window. Pushing the rightarrow will increase the setpoint value andpushing the left arrow will decrease thesetpoint value. When the desired value is
displayed, push the Enter key to select it.
If the system is configured for monitoring andcontrol, you will need to enter values for thetemperature range in addition to the tem-perature setpoint. This determines thecontrol range for heating and chilling devices.
Since most heaters and chillers have theirown thermostat, these thermostats can beused as a backup temperature control system.The heater thermostat should be set severaldegrees above your desired temperature andthe chiller several degrees below. If your
desired temperature is 78F then set yourheater thermostat at 81F and your chillerthermostat at 75F.
Temperature Example
Suppose you have a setpoint value is 78.2 F,your target operating range is 0.1 F, and thealarm range is 0.1 F. (Figure 3.5 illustratesthe relationship between the setpoint value,and the control and alarm ranges.)
If the temperature stays within the targetoperating range (78.178.3 F), the parameterstatus light will remain off.
If, however, the temperature enters the controlrange (between 78.1F on the low side and78.3F on the high side), the parameter statuslight will either turn green if the chillingsystem is active, or red if the heating systemis active.
At the beginning of the control range, theOctopus will turn on the chiller or heater, asneeded. These peripheral devices will remainactivated to drive the temperature value back
to the optimal setpoint, and once achieved,the peripheral device will be turned off.
Configuration data such assetpoint values, are stored inpermanent (nonvolatile)memory. Should primary orbackup power to the Octopusbe lost, the setpoint values willbe restored to their previousvalues when power is restoredto the Octopus.
Note
Note that the chiller system hasa protective feature that pre-vents the chiller from turningon/off more often than every 10minutes.
Note
-
8/13/2019 Octopus 3000 Manual v 22
47/113
CHAPTER3: PROGRAMMINGTHESOFTWARE 3-11
Programming the Software
pH Setpoints
The pH setpoint defines the optimum pH levelthat you wish the Octopus 3000 controller tomaintain in the aquatic environment. Toenter a pH optimal setpoint value, go to theSetpoint Menu. Press Enter and use the rightarrow to reach the pH Setpoints Menu. Whenyou press Enter, the current pH setpoint willflash on the LCD display. Push the rightarrow to increase the setpoint value or the leftarrow to decrease the setpoint value. Whenthe desired value is displayed, push Enter toregister your selection. Other setpoint valuesare entered similarly. For controlling pHadditional values must be entered on the Set-points Menu, including the dose and dwell values.
The Octopus can control the pH system by dosinga buffer solution or injecting CO
2gas or do both.
The pH control range is selectable from 0.05pH units to 1.00 pH units, and determinesthe ON points for the pH and CO
2dosing
systems.
The Current Dose in minutes (time per dose of
buffer solution) will flash on the screen. Thisis the number of minutes that the Octopuscontroller will dose buffer to bring the pHlevel back up into the Setpoint . The accept-able values for buffer solution dosing are from1 to 99 minutes.
Dose times will range from 0-99 minutes. Youare also required to enter a dwell time TheDwell Time is the length of time that theOctopus 3000 will wait until dosing can occuragain. This feature prevents overdosing asystem with buffer.
The next option is the setting of the CO2Dosing, in seconds. This controls how longthe CO2gas valve will remain open. Thepossible values range from 0 to 99 seconds.For example, if you enter 10 seconds, whenthe upper end of the target operating range for
pH is reached, the Octopus will turn on theCO2gas for 10 seconds to reduce the pH to alower value. A CO
2Dwell Time will force the
Octopus 3000 to wait to allow the gas to beabsorbed into the water. This will prevent thepH value from dropping to fast.
If you enter a value of 0 seconds, then theOctopus 3000 will turn the gas valve on whenthe pH level reaches the high end of the targetoperating range and will keep the valve openuntil the pH drops below the optimal pH level.The 0 seconds option is for systems withpressurized CO2injection.
The last pH control selection is the CO2dwell
time. This is the time that the Octopus waitsbefore allowing another CO2gas injection tooccur. This time is selectable from 1-255minutes.
If the Octopus 3000 senses that the pH isoutside of the target operating range, it will
Dose Tim e CO2 Sys te m
0 Sec. Pressurized Systems1-99 Sec. Non Pressurized Systems
Table 3.4 CO2Dose Time
Note that if the dose time is setto 0, the dwell time entered willbe ignored by the Octopus andthe unit will continue to dosebuffer until the pH returns to
the Optimal Setpoint Value.
Note
-
8/13/2019 Octopus 3000 Manual v 22
48/113
CHAPTER3: PROGRAMMINGTHESOFTWARE3-12
Programming the Software
turn on the pH doser or CO2injector untilthe pH returns to the optimal setpoint range(See Figure 3.5).
An red parameter status light indicates theCO
2dosing system is active, while an green
status light indicates the buffer dosingsystem is active
A CO2 dwell time is indicated by a flashing
red status light while a flashing green lightindicates a buffer dosing system dwell time.
ORP Setpoints
The ORP setpoint defines the acceptable levelof oxidation reduction potential in theaquatic environment. The ORP level is anindicator of ozone. Acceptable levels arebetween 200 and 450 mV. The ORP setpointcan be maintained through the use of anozonizer in Monitoring and Control System.However, unlike the proceeding parameters,the ORP setpoint has a FIXED range of 5mV.
Verify All Setpoint Settings
You can easily view the parameter setpoints.From the Top Level Display, simply pusheither the right or left arrow keys to scrollthrough the screens. Each successive push ofan arrow will display another setpoint untilit returns to the current status screen. PushEnter at any time to enter the other menus.
STEP 16ENTERING ALARMRANGE
General Alarm System
The alarm is located in the Setup Menu underAlarms. All alarm ranges are entered in thismenu. The alarm range can only be entered ifthe alarm system itself is also enabled first. Inaddition to an alarm range for each systemthere is a common alarm hold-off time whichcan be entered. This will prevent any systemfor generating an alarm until the hold off time
has expired. Table 3.5 lists all system param-eters that are located in the Setup menurelating to alarming or system configuration.
When a alarm occurs the status line on theLCD screen will show which system(s) had thealarm and whether the alarm was a high orlow alarm condition. When the hold-off timehas expired the controller will generate anaudible alarm and initiate a alarm page if thepager system is enabled on the controller. Thealarm condition is reset by pressing the Enterbutton on the front of the controller. Pressingthe Enter key during an alarm will stop the
audible alarm and cancel any pending pagesbut the alarm status messages will NOT becleared until the alarm condition is no longersensed by the controller.
Alarm will be triggered when the value mea-sured falls outside of the control range. (See
Figure 3.5).
The high or low alarm triggerpoints are computed by addingthe control value and the alarm
value to the setpoint value.
Note
-
8/13/2019 Octopus 3000 Manual v 22
49/113
CHAPTER3: PROGRAMMINGTHESOFTWARE 3-13
Programming the Software
Conductivity Alarm
The conductivity alarm range is entered in theSetup Menu under the Alarms sub-menu.From the Setup Menu, press the Enter key.Then press the right or left arrow until youreach the alarm menu. Press the Enter keyand you will go to the pH menu. Scroll leftusing the arrows until you get to CondAlarm. Press enter and then select Enabledand press Enter again. You will automaticallybe directed to the conductivity alarm rangemenu.
As long as the conductivity value of the waterstays outside the alarm range, (i.e. within theacceptable operating range or control range),the alarm will not be triggered.
If the conductivity value goes outside thecontrol range, an alarm will be triggered. Thealarm condition will be confirmed by an alarmstatus message appearing in the LCD screen,the sounding of an audible alarm, and theissuance of a pager alarm (if installed), oncethe hold-off period has expired. (See Figure3.5)
In addition to the general conductivity alarmthere is also two separate conductivity alarmoutputs which can be configured in the SetCtrl Address menu. One for a low conductivityalarm and the other for a high conductivityalarm. Theses outputs will go active in a lowor high conductivity alarm condition.
Temperature Alarm
The temperature alarm range can be set from0.10 degrees to 2.00 degrees. The tempera-ture alarm range can only be set if the tem-
perature alarm is enabled.
An alarm will be triggered when the valuemeasured falls outside of the control range.
For example, suppose you have a optimaltemperature setpoint value of 78.0 F and anacceptable operating range of 0.1. Figure 3.6illustrates the relationship between thesetpoint value and the acceptable operatingrange. When the parameter value stays withinthe acceptable operating range (77.978.1 F),the system monitoring function will notdetect an alarm and the parameter statuslight will be off.
If, however, the temperature drifts past theacceptable operating range (below 77.9 F or
above 78.1 F), an audible alarm will betriggered, the monitored parameter will begin
Table 3.5 Parameter Settings Included in the Setup Menu
ON THE SETUP MENU
Value Required Monitoring
System
Monitoring &Control System
Conductivity
Temperature Compensation X X
Conductivity Alarm Range X X
Temperature
Temperature Scale X X
Temperature Alarm Range X X
pH
ph Alarm Range X X
ORP
ORP Alarm Range X X
-
8/13/2019 Octopus 3000 Manual v 22
50/113
CHAPTER3: PROGRAMMINGTHESOFTWARE3-14
Programming the Software
flashing on the current status screen, and apager alarm will be issued (if installed), uponthe expiration of the hold-off period.
As a second example , suppose you have asystem with the same values as above butwith a temperature range of 0.2F. If thetemperature value drifts away from the targetoperating range and comes to the controlrange (lower than 77.8F or higher than78.2F), then the peripheral device (heater orchiller for example), will be activated. If thetemperature continues to drift away from the
target operating range and comes to the end ofthe control range (i.e. 78.3F),then the heater or chiller will remain activebut an audible alarm will be triggered, themonitored parameter will begin flashing onthe LCD display window, and a pager alarmwill be issued (if installed), upon expiration ofthe hold-off period.
pH Alarm
The pH alarm range is e