PN_R16-2_Ocean.docx 2016-12-09 Product Note Ocean 2014 – News in Release R16-2 Release information Release date 2016-12-09 Release name R16-2 Installer Ocean 2014 Installer 2016.12A Ocean 2014 version Ocean 2014 Setup 2016.12.7.242 EXE file version Ocean 2014.exe 1.0.2.18832 DLL version RTICMLib.dll 2.6.1.9111 This is a new release of Ocean 2014 for Cobia and Black Piranha. This release of Ocean 2014 has been mainly dedicated to maintenance. Reported bugs have been fixed, and some new features have been added. New Customer Values Below the most important news are described. New mammo radiation quality selection in QuickCheck Now the radiation quality can be selected based on Manufacturer and model. This makes it easier and quicker to select the proper calibration for each machine type. The example below shows the selection of Rh/Ag target-filter combination for the GE Senographe Pristina, by first selecting Manufacturer, then model and last the radiation quality amongst those available for that model.

Upload: others

Post on 30-Apr-2020

8 views

Category:

Documents


0 download

TRANSCRIPT

PN_R16-2_Ocean.docx 2016-12-09

Product Note

Ocean 2014 – News in Release R16-2

Release information

Release date 2016-12-09 Release name R16-2 Installer Ocean 2014 Installer 2016.12A Ocean 2014 version Ocean 2014 Setup 2016.12.7.242 EXE file version Ocean 2014.exe 1.0.2.18832 DLL version RTICMLib.dll 2.6.1.9111

This is a new release of Ocean 2014 for Cobia and Black Piranha. This release of Ocean 2014 has been mainly dedicated to maintenance. Reported bugs have been fixed, and some new features have been added.

New Customer Values Below the most important news are described. New mammo radiation quality selection in QuickCheck Now the radiation quality can be selected based on Manufacturer and model. This makes it easier and quicker to select the proper calibration for each machine type. The example below shows the selection of Rh/Ag target-filter combination for the GE Senographe Pristina, by first selecting Manufacturer, then model and last the radiation quality amongst those available for that model.

PN_R16-2_Ocean.docx 2016-12-09

Product Note

Lock cursors and waveform zoom state

Now if moving cursors and/or zooming the waveform in Ocean and Ocean QuickCheck, the zoom and cursors will stay in that state for next exposure. This allows quick analysis of repeated exposures, when a specific analysis is done. The zoom state and cursor positions are also stored for when saving data as well as for reports. New Excel connect mode A new connection mode to MS Excel have been implemented. This allows exporting data to Excel in real-time in a pre-defined format. The format is the same as used by the Raysafe Xi and X2, which allows to use same Excel template either using an RTI meter or a Raysafe meter. To enable this function – From Quick Check and Options, check the function Fixed format (Excel connect).

Also when working in Excel connect the minimized Ocean window has changed. The most important functions are available, and made touch friendly. See below.

PN_R16-2_Ocean.docx 2016-12-09

Product Note

New k-factors for CTDI Following changes of k-factors for CTDI calculations have been done.

Added GE revolution (issue #8463) Changed "Optima CT660" to "Optima 660" The following models are equal:

Optima 580 = Light Speed RT GE Revolution EVO = GE Optima 660 GE Revolution GSI = GE Optima 520 = Light speed 16 GE Optima 540 = Light speed Pro 16

Improved how a lost communication is handled in Quick check In Quick Check a lost communication caused Ocean to hang. That is now resolved. Improved GUI when changing order of displays in Real-Time Displays In the Real-Time Display the order of displays can be customized. This has been available in previous versions too. Now it is better visualized in the drag and drop function. HVL limit rules - Tungsten targets in mammography HVL limits have been added for: "ACR Mammo W/Rh", ACR Mammo W/Ag" and "ACR Mammo W/Al". Example for W/Al shown below.

Improved template protection Some parts in Ocean that have not been affected by protection before, are now covered by the different protection levels that is avilable. There are no additional levels. Only a more stringent coverage. Allow Quick check to be used without a meter

PN_R16-2_Ocean.docx 2016-12-09

Product Note

Now when opening a saved Quick Check, the measurement will be shown “as a Quick Check”. In previous versions, a saved Quick Check was opened as a Real-Time Display. Now possible to save a Quick check favorite from within Quick check Under the Options menu the selection Save as Favourite… is now available.

Now possible to protect and hide Quick check favorites in the library For groups that uses Quick Check favourites and need a strict control of the use of those, a possibility to protect those is implemented. This function is available only for larger groups and requires support from RTI to access. Allow Quick Check favourites for the Connect license level Quick Check favourites is now available also for the Connect license level. In previous versions, the Professional level has been required. Calibration date in reports Ocean reports can now include date of last calibration of the meter that was used during test.

Database format This version of Ocean has the same database structure as last version of Ocean. Last change of database structure came with the Ocean 2014 release R15-2, 2015.12A. Any old database from previous versions is restructured when the new version of Ocean 2014 is started. This is an irreversible process and it is not possible to go back to an earlier version once the database has been restructured. For security reasons a backup file is created automatically and saved in the Ocean 2014 data folder. The file name is “DataBackupBeforeNewVersion.accdb”.

Cobia and Piranha firmware Piranha There is a new firmware for Piranha in this release, and consequently the lowest firmware that shall be used with this release of Ocean 2014 is 4.2B. Cobia To run the new version of Ocean 2014 existing Cobia shall be upgraded with a new firmware. Lowest Cobia system version that shall be used with this version of Ocean 2014 is 2016.12A.

PN_R16-2_Ocean.docx 2016-12-09

Product Note

Bluetooth Pairing Note that it may take some time to pair the Piranha or Cobia with a new device the first time.

Resolved Problems Below is a list of the most important resolved problems, for a complete list please read the “Version History” document distributed with Ocean 2014 and RTI internal bug handling system.

Issue # Description

8516 At short exposure times or very low signals Ocean could sometimes show a false Peak kV value as kV(PPV). This is now resolved.

8480 Improved how a lost communication is handled in Quick check

8465 New Excel mode, compatible with the format used with Xi and X2 View from RaySafe

8481 Now possible to select calibration based on manufacturer and model

8449 Improved GUI when changing order of displays in Real-Time Displays

8430 HVL limit rules for: "ACR Mammo W/Rh", ACR Mammo W/Ag" and "ACR Mammo W/Al"

8463 Extended with more k-factors

8486 Improved template protection

8424 Allow Quick check to be used without a meter

8417 Now possible to save a Quick check favorite from within Quick check

8073 Now possible to protect and hide Quick check favorites in the library

8418 Allow Quick check favorites for the Connect license level

7515 Now possible to lock the waveform zoom state

8422 Analysis exposure selection doesn't work if "set" value is in general settings

8355 User X-Y graph analysis not showing failure in grid

8405 Ocean can't always update "Time Diff %" when adding new rows

8374 A possible “range check” error when opening a document

8479 Dental model 255 never measure above 105 kVp

8183 An error in an HVL calculation block Ocean from going to the next row

8450 In Quick check wasn’t the template’s stored TF reflected in meter settings

8404 When checking version in Ocean, you're told that the meter firmware is a betaversion

8471 Possible crash when measuring while the cell editor in Excel is active

8468 Set values from Excel doesn't update Ocean columns when moving to next row

8365 Corrected k-factor for Siemens Somatom Emotion

8482 If the CT manufacturer is unknown, there is an error message when listing k-factors

8488 Adjusted the post delay in Combo mode template

8412 AGD is shown in "R" when selecting "R" as default unit in Quick Check

*** END ***

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

Ocean 2014 – News in Release R16-1.1 & R16-1

Release information Release name R16-1.1 Release date 2016-06-27 Installer 2016.6A Ocean 2014 version 2016.06.14.214 EXE file version 1.0.2.18801 DLL version 2.6.1.9096 Release name R16-1 Release date 2016-05-23 Installer 2016.5A Ocean 2014 version 2016.05.19.211 EXE file version 1.0.2.18794 DLL version 2.6.1.9093 This is a summary of the two releases R16-1 and R16-1.1 released in May and June 2016, of Ocean 2014 for Cobia and Black Piranha.

Ocean 2014 – News in Release R16-1.1 This release of Ocean 2014 has been dedicated to the new radiation qualities M27 and M28 for Hologic Affirm Prone. Reported bugs have been fixed.

New Customer Values Below the most important news are described. Mammography Improvements

To handle the mammography spectra for the Hologic Affirm Prone, new calibrations have been added to the Black Piranha.

Affirm Prone W / 50 µm Ag, [M27] Affirm Prone W / 0.70 mm Al, [M28]

See the Piranha Product Note, PN_R16-1_Piranha, for more information. USB service – Bug fix A reported bug in USB communication have been fixed.

Database format This version of Ocean has the same database structure as last version of Ocean. Last change of database structure came with the Ocean 2014 release R15-2, 2015.12A. Any old database from previous versions is restructured when the new version of Ocean 2014 is started. This is an irreversible process and it is not possible to go back to an earlier version once the database has been

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

restructured. For security reasons a backup file is created automatically and saved in the Ocean 2014 data folder. The file name is “DataBackupBeforeNewVersion.accdb”.

Cobia and Piranha firmware Piranha There is no new firmware for Piranha in this release, and consequently the lowest firmware that can be used with this release of Ocean 2014 is 4.1B. Cobia There is no new firmware for Cobia in this release, and consequently the lowest firmware that can be used with this release of Ocean 2014 is 2016.5A.

Resolved Problems Below is a list of the most important resolved problems, for a complete list please read the “Version History” document distributed with Ocean 2014 and RTI internal bug handling system. Issue # Description

8266 Position check can at some circumstances fail

8233 RTI USB Service isn’t always started when installing for the first time

2660 "Select detector/calibrations" are blank (requiring "Show All") for mA, mAs, and light

8270 Import may notify about database upgrade for import file

Ocean 2014 – News in Release R16-1 This release of Ocean 2014 has been mainly dedicated to maintenance. Reported bugs have been fixed, and some new features have been added.

New Customer Values Below the most important news are described. New mammography calibrations Support for the new GE Senographe PristinaTM See the Piranha Product Note, PN_R16-1_Piranha, for more information. New zoom function for waveforms The zoom function for waveforms has been extended to work also for the y-axis.

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

Zoom the waveform to see details. It is possible to zoom in both horizontal and vertical direction. Hold down the mouse pointer and move it to mark the zoom, release and the waveform part marked is enlarged.

Horizontal Cursor for waveforms A horizontal cursor for the waveform data is added. The value for the horizontal cursor is shown at the cursor.

You can move the waveform cursors by use of the mouse or the keyboard. Cursor labels for waveforms The cursors in the waveform screen is now labeled with their values. kV waveform extension The kV waveform is now visible from a lower kV value than before. This is mainly seen in beginning of the exposure or a pulse where the kV waveform is now visible from a much lower level. Bluetooth communication improvements Increased security. Ocean now warns directly if it happens that communication with the meter is lost. Back function in Quick Check Dialogues A back button has been added to the dialogues in Quick Check. Allows to easily correct a wrong selection. Toggle button in Quick Check In case there is both displays, grid and analysis present, the toggle button will show up in Quick Check. Click

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

the button to toggle between displays and grid.

Calibration reminder Ocean 2014 now has a reminder function that is used to give a message when it is time for recalibration of the meter. – There is a message system included where different information is displayed. The message inbox is shown to the right on the lower status bar:

When a new message arrives the icon changes:

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

You can click on the icon to open the message list:

Here all your messages are shown. Double-click in the list to read a message. Unwanted messages can be deleted. Calibration reminders can be turned off in the Program Options. Displays in Quick Check Application To increase readability displays have been added to the Quick Check Applications. Faster transfer of waveforms for Cobia Cobia waveforms are now displayed much faster in Ocean 2014.

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

Inplace editors Dropdown fields are now visible without the need to select the cell first.

MS Excel features added

Add columns available for - Read settings from Office (#8121)

Added CT dose profiler columns

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

Add Detector, serial number, date- and time stamp to Excel export (#8122)

Database format This version of Ocean has the same database structure as last version of Ocean. Last change of database structure came with the Ocean 2014 release R15-2, 2015.12A. Any old database from previous versions is restructured when the new version of Ocean 2014 is started. This is an irreversible process and it is not possible to go back to an earlier version once the database has been restructured. For security reasons a backup file is created automatically and saved in the Ocean 2014 data folder. The file name is “DataBackupBeforeNewVersion.accdb”.

Cobia and Piranha firmware Piranha There a new firmware for Piranha in this release, and consequently the lowest firmware that can be used with this release of Ocean 2014 is 4.1B. Cobia To run the new version of Ocean 2014 existing Cobia units must be upgraded with a new firmware. Lowest Cobia system version that can be used with this version of Ocean 2014 is 2016.5A.

Bluetooth Pairing Bluetooth pairing is faster and more stable. Issues with Bluetooth pairing with Windows 10 have been further resolved. The effect is that the Ocean 2014 starts up faster after first time the Piranha and Cobia have been connected with a particular Tablet or laptop. The new installer can also be used with Windows 10. Note that it may still take some time to pair the Piranha or Cobia with a new device the first time.

Resolved Problems Below is a list of the most important resolved problems, for a complete list please read the “Version History” document distributed with Ocean 2014 and RTI internal bug handling system. Issue # Description

8211 When starting a Quick Check while a QC is already starting, there can be a crash

8181 Improvements in Quick Check startup

8177 Read settings from Office doesn't work in sessions

8159 Possible crash when changing analysis' exposure selection

8153 Crash can occur when clicking in the waveform graph

8139 Enable Mammography beam qualities for Cobia Flex external probes

8132 Starting a Quick Check from within a CTDP session having manufacturer and model selected can give an undefined k-factor

8131 Refresh calibration list is enabled even when no document is open

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

8126 Ocean connect license has issues exporting examples with analyses and then importing these

8112 Add a toggle button for Quick Check having analysis and displays

8110 Add detector name and serial number to Quick Check meter settings

8085 Ocean waveform, sometimes it has an extra piece of measurement that doesn't exist.

8041 Very high sensitivity is not supported in Quick Check

8026 Some waveform graph points are not joined together

8023 Possible crash when editing report templates

8015 Show approximately 50% of the kV waveform in the beginning of the exposure

8009 Session Template requires some equipment in this room

7999 Some customer’s computer can't find Bluetooth devices

7991 Indicate battery level when using USB

7987 Add a "Back" button when using the manufacturer/model selection

7985 Missing first sample in graphical waveform

7979 A template with Normalized dose doesn't work as application in Quick Check

7946 Misspelled word on the Ribbon bar

7936 Possible crash when using the Adjust button from the Excel control window

7934 The Excel control window cannot be removed

7931 Calibrations are never allowed to change when saved

7927 Problem using Ocean with Cobia and an external probe having one calibration

7926 Last Modified date not being updated

7925 Trending on a cell value, current test’s cell value is not shown in the trend graph

7919 Possible crash in "Export for user analysis"

7855 Possible crash when using gestures

7805 Space between waveforms in the Excel "dump" depends on selected columns

7760 Quick Check applications for dental pan should have a long recording time

7629 Al filter thickness cannot be properly entered in Quick Check Application for HVL

7627 Update the hint in the Scatter & Leakage templates we provide

7601 Possible crash after changing detector within a test

7570 Measurement reset nullifies values but not waveforms in Ocean.

7532 Cobia Smart now gives a proper message that it isn’t supported by Ocean

7254 Update the Trial button

5700 Possible crash when activating a test after deleted other tests in a session

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

Ocean 2014 – News in Release R15-2

Release information Release date 2015-12-22 Release name R15-2 Installer 2015.12C Ocean 2014 version 2015.12.22.198 EXE file version 1.0.2.18348 DLL version 2.6.0.9078

This is a new release of Ocean 2014 for Cobia and Black Piranha. This release of Ocean 2014 has been mainly dedicated to maintenance. Reported bugs have been fixed, and some new features have been added.

New Customer Values Below the most important news are described. Extended Meter settings in Quick Check The following settings have been added to the Meter Adjust page: Filtration, Beam correction factor, and Window. Quick Check now has all settings that are available in the QABrowser. Normalized Dose in Quick Check It is now possible to present Normalized dose in Quick Check. New fields on the Meter adjust page to specify SSD and SDD to get a normalized dose and dose rate value. Improved readability on tablets with high screen resolution Tablets have higher and higher resolution. Ocean 2014 objects appear small on these devices. The Font feature that allows user to increase font size has been improved and now increases size of more object than just text. Also one more size, Extra Large, has been added. Read more in the section Minimum screen resolution later in this document. New mammography calibrations See the Piranha Product Note, PN_R15-2_Piranha, for more information. New macros New macros have for Used detector, Detector s/n and Detector Calibration date. This information can now be included into the analysis text. Log scale for user-defined graphs User-defined graphs can now have x- and y-axis with log scale.

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

Improved Bluetooth communication The Bluetooth communication has been further improved. A new protocol is used that eliminates the problems with Windows 10 that exists with earlier versions. The pairing process is also improved that gives more stability and speed. Effective time A measured parameter Effective time is added. Effective time is the time radiation is present during an exposure and all “dead-time” is excluded. “Dead-time” is the time when no radiation is present. Limit the amount of mammography calibrations shown The extensive support for mammography we have in Piranha gives a long list of mammography calibrations. Not everybody uses all of them. It is now possible, in Program Options, to hide mammography calibrations you don’t want to see when you are asked for a mammography calibration. You can also give your own name to the calibrations. This name is shown instead of the default name. In all places where this feature is used an optional “Show all” choice is available. Improved report print out Different sections of the report are now better “kept together”. “Not so good” page breaks are avoided.

Database format This version of Ocean has a new database structure. Any old database from previous versions is restructured when the new version of Ocean 2014 is started. This is an irreversible process and it is not possible to go back to an earlier version once the database has been restructured. For security reasons a backup file is created automatically and saved in the Ocean 2014 data folder. The file name is “DataBackupBeforeNewVersion.accdb”.

Cobia and Piranha firmware Piranha There is no new firmware for Piranha in this release, and consequently the lowest firmware is 4.0A. Cobia To run the new version of Ocean 2014 existing Cobia units must be upgraded with a new firmware. Lowest Cobia system version that can be used with this version of Ocean 2014 is 2015.12A.

Bluetooth Pairing Bluetooth pairing is faster and more stable. Issues with Bluetooth pairing with Windows 10 have been resolved. The effect is that the Ocean 2014 starts up faster after first time the Piranha and Cobia have been connected with a particular Tablet or laptop. The new installer can also be used with Windows 10. Note that it may still take some time to pair the Piranha or Cobia with a new device the first time.

Minimum screen resolution Ocean 2014 is designed for a minimum screen resolution of 1280 x 800. If screen resolution is higher, objects in Ocean can look very small. There are different ways to increase visibility on the screen:

Increase font size in Ocean 2014. This will only affect texts in Ocean, not buttons and other objects.

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

Change settings in Windows (increase size of text, apps and other items). This feature will increase size of all objects.

Increase font size You can change the Font size to increase the size of texts and ribbon bar in Ocean 2014. There are four different sizes to choose from; Small (default), Medium, Large and Extra Large. If you use Extra large font size be aware that some texts may be truncated. Increase object size in Windows You can increase everything that is shown on the screen by changing the Display settings in Windows (Windows 8 and 10). You can set the "size of texts and apps" between 100 % and 200 %. It is not recommended that you increase more than what gives an "effective" screen resolution lower than 1280 x 800. For example, if you have a screen resolution of 1980 x 1200 and you think that Ocean 2014 "looks small". It is then not recommended to go higher than 150% (1980/1.5=1320 and 1200/1.5=800). You can combine these two possibilities to maximize size of Ocean 2014 texts and buttons.

Resolved Problems Below is a list of the most important resolved problems, for a complete list please read the “Version History” document distributed with Ocean 2014 and RTI internal bug handling system.

Issue # Description

6630 Calibration list in Quick Check doesn’t scroll when it is too long to fit on the screen

7050 Long analysis may be truncated in the report

7200 Ocean can’t use pictures with transparent background

7482 When you add a test to a session Ocean wants to open Excel

7503 Possible crash when remove a test from a session

7508 “Link to waveform data” column can lose its value when editing a related analysis

7578 When using on-screen keyboard with a tablet in the Setup wizard, Ocean may shut down and show an error

7623 Crash after import and right-click in the database tree

7664 Bad switching between Ocean and Excel when using embedded Workbook

7759 Wrong document may be deleted from the tree

7764 Set Time and Meas time doesn’t work with the unit “Minutes”

7774 Report comment texts may be truncated

7806 “Auto-send” doesn’t work with embedded Workbook and without cell-to-cell links

7808 Possible crash when deleting an exposure with an analysis

7811 Sometimes time out when retrieving data from the Cobia

7813 Incorrect dose and dose rate measurement in Quick Check with the external dose probe if it is detached and attached again. See separate Product Note “PN_Ocean_005_External_Dose_Probe_Failure.pdf”

7848 Column light unit is not retained

7856 No mAs in Quick Check when using Cobia

PN_R16-1.1+_Ocean.docx 2016-06-29

Product Note

Known Limitations The table below describes limitations that have been observed and not been fixed in the release.

Description

Extra Large font size: in some places texts doesn’t fit into available space when “Extra Large” font size is used.

Slow data transfer of waveforms with Cobia.

It takes long time to recognize that an external detector is connected, especially if it has many calibrations.

If you Cancel “Save” when using Quick Check, Ocean returns to Quick Check but the wrong menu is shown. Leave Quick Check and go back to solve the problem.

Not possible to use longer time than 60 seconds when using Timed mode.

When using “Normalized dose” in Quick Check, the waveform graph is labeled “(norm)” but scale is showing non normalized dose.

When activating a row in a checklist, the analysis tab is shown instead of the question tab.

A test template may get the wrong title when used in a session (measurement) if you leave the “Title” field empty when you design tests.

Quick Check doesn’t support that you measure only with the internal mAs when you use Cobia.

Quick Check sometimes doesn’t react when the external dose probe is connected. If this happens; click on the “New Quick Check” button.