Occlusion alarms during administration of lipid emulsion with vitamins in neonates: in vitro evaluation M. Mattiuzzo 1 , C. Fonzo-Christe 1 , L. Bouchoud 1,3 , F. Sadeghipour 1 , R. Pfister 2 , P. Rimensberger 2 , P. Bonnabry 1,3 1 Pharmacy, 2 Neonatology and PediatricIntensive Care Unit, Geneva University Hospitals (HUG) and 3 School of pharmaceutical sciences, University of Geneva, University of Lausanne, Switzerland Introduction and objectives One week-stable lipid emulsion syringes with vitamins are used in the neonatology unit of our hospital 1 . An increased incidence of occlusion alarms has been reported by nurses compared to lipid emulsion syringes without vitamins prepared daily. The objective was to compare experimentally the occurrence of occlusion alarms when infusing lipid emulsion with or without vitamins, stored for 2 hours or 1 week at +2-8°C and if differences were observed when stored in syringes or in glass vials. Material and methods Smart pump (Module DPS, Fresenius Kabi, pressure alarm at 300 mmHg), 10 ml BD-Plastipak syringes, syringe double- tubing 3-way connector ramp 3-way connector filter catheter alarm at 300 mmHg), 10 ml BD-Plastipak syringes, amber type II glass vials, translucent 150 cm double- tubing (BD Extension Tube) and 3-way connectors (BD Connecta Luer-Lok and BD Multiflo), in-line filter (1.2 μm Lipipor NLF, PALL Medical), catheter 27 G (Deltec, Smith Medical). Lipid emulsion (Lipofundin ® MCT/LCT) tested with or without vitamins (1 amp. Cernevit ® ), and stored for 2 hours or 7 days at +2-8°Cin syringes or glass vials. Tests were conducted in duplicate. Flow rate started at 1 mL/h with 1 mL/h increments every 25 min, pressure determined on the pump. Kinematic viscosity (Ubbelohde viscometer, Schott-Geräte, capillary 0.63 mm) and apparent pH measured at 25°C. Initial flow rate 1 ml/h with increment of 1 ml/h every 25 min Results (B) viscosity in syringes 1. Vitamins addition increases the risk of occlusion alarm 2. Storage for 7 daysincreases the risk of occlusion alarm 1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h 1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h 1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h 1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h 1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h Lipids without vitamins Lipids with vitamins Lipids without vitamins Lipids with vitamins storedfor 2 hours storedfor 7 days 2.37 2.58 2.33 2.57 mm 2 /s 4. Vitamins addition slightly increases viscosity (A) (A) Internal pressure of the system, determined on the pump. (B) Kinematic viscosity measured at 25°C. Emulsions were stored i n syringes at +2-8°C. (C) Apparent pH measured at 25°C. Emulsions were stored in syringes at +2-8°C. Conclusions Dramatic increase of alarm occlusion risk occurs when lipid emulsion mixed with vitamins is stored for 7 days in syringes at 2-8°C. Modification in emulsion’s properties an d increased viscosity when adding vitamins to the lipid emulsion could be partially responsible for this observation. However, syringe stiction due to extraction of the silicone oil 2 by the emulsion mixed with vitamins is probably the main cause, as results are not reproduced when stored in a glass vial. Due to lack of data about silicone oil administration in neonates, production of syringes of lipids mixed with vitamins has been reconsidered. (C) References 1. BouchoudL et al. Clin Nutr2010. 29(6):808-12. - 2. Capes DF et al. PDA J PharmSciTechnol1996. 50(1):40-50. Acknowledgments Weare gratefulto the Group of Pharmaceutical Technology(Universityof Geneva) for providingthe viscometer. apparent pH in syringes Vienna http://pharmacie.hug-ge.ch/rd/posters.html Authorsreport no conflictof interest internal pressure 3. Storage in a syringe increases the risk of occlusion alarm 1 ml/h 2 ml/h 3 ml/h 4 ml/h 1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h 5 ml/h storedfor 2 hours storedfor 7 days 7.23 6.15 8.01 6.10 5. Vitamins addition decreases apparent pH 2011

Upload: others

Post on 17-Aug-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Occlusion alarms during administration of lipid emulsion ... · Occlusion alarms during administration of lipid emulsion with vitamins in neonates: in vitro evaluation M. Mattiuzzo1,

Occlusion alarms during administration of lipid emulsion with

vitamins in neonates: in vitro evaluationM. Mattiuzzo1, C. Fonzo-Christe1, L. Bouchoud1,3, F. Sadeghipour1, R. Pfister2, P. Rimensberger2, P. Bonnabry1,3

1Pharmacy, 2Neonatology and Pediatric Intensive Care Unit, Geneva University Hospitals (HUG) and3School of pharmaceutical sciences, University of Geneva, University of Lausanne, Switzerland

Introduction and objectivesOne week-stable lipid emulsion syringes with vitamins are used in the neonatology unit of our hospital1. An increasedincidence of occlusion alarms has been reported by nurses compared to lipid emulsion syringes without vitaminsprepared daily. The objective was to compare experimentally the occurrence of occlusion alarms when infusing lipidemulsion with or without vitamins, stored for 2 hours or 1 week at +2-8°C and if differences were observed when storedin syringes or in glass vials.

Material and methodsSmart pump (Module DPS, Fresenius Kabi, pressurealarm at 300 mmHg), 10 ml BD-Plastipak syringes,

syringedouble-tubing

3-wayconnectorramp3-way

connectorfiltercatheter

alarm at 300 mmHg), 10 ml BD-Plastipak syringes,amber type II glass vials, translucent 150 cm double-tubing (BD Extension Tube) and 3-way connectors (BDConnecta Luer-Lok and BD Multiflo), in-line filter (1.2 µmLipipor NLF, PALL Medical), catheter 27 G (Deltec, SmithMedical). Lipid emulsion (Lipofundin® MCT/LCT) testedwith or without vitamins (1 amp. Cernevit®), and storedfor 2 hours or 7 days at +2-8°C in syringes or glass vials.Tests were conducted in duplicate. Flow rate started at 1mL/h with 1 mL/h increments every 25 min, pressuredetermined on the pump. Kinematic viscosity (Ubbelohdeviscometer, Schott-Geräte, capillary 0.63 mm) andapparent pH measured at 25°C.

Initial flow rate 1 ml/h with increment of 1 ml/h every 25 min

Results

(B)viscosity in syringes

1. Vitamins

addition

increases the

risk of occlusion

alarm

2. Storage for 7

days increases

the risk of

occlusion alarm

1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h 1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h

1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h 1 ml/h 2 ml/h 3 ml/h 4 ml/h 5 ml/h

Lipids without

vitamins

Lipids with

vitamins

Lipids without

vitamins

Lipids with

vitamins

stored for 2 hours

stored for 7 days

2.37 2.58

2.33 2.57

mm2/s

4. Vitamins addition slightly increases viscosity

(A)(A) Internal pressure of the system, determined on the pump.(B) Kinematic viscosity measured at 25°C. Emulsions were stored i n syringes at +2-8°C.(C) Apparent pH measured at 25°C. Emulsions were stored in syringes at +2-8°C.Conclusions

Dramatic increase of alarm occlusion risk occurs when lipid emulsion mixed with vitamins is stored for 7 daysin syringes at 2-8°C. Modification in emulsion’s properties an d increased viscosity when adding vitamins tothe lipid emulsion could be partially responsible for this observation. However, syringe stiction due toextraction of the silicone oil2 by the emulsion mixed with vitamins is probably the main cause, as results arenot reproduced when stored in a glass vial. Due to lack of data about silicone oil administration in neonates,production of syringes of lipids mixed with vitamins has been reconsidered.

(C)

References 1. Bouchoud L et al. Clin Nutr 2010. 29(6):808-12. - 2. Capes DF et al. PDA J Pharm Sci Technol 1996. 50(1):40-50.

Acknowledgments We are grateful to the Group of Pharmaceutical Technology (University of Geneva) for providing the viscometer.

apparent pH in syringes

Vienna

http://pharmacie.hug-ge.ch/rd/posters.htmlAuthors report no conflict of interest

internal pressure

3. Storage in a

syringe

increases the

risk of occlusion

alarm

1 ml/h 2 ml/h 3 ml/h 4 ml/h 1 ml/h 2 ml/h 3 ml/h 4 ml/h5 ml/h 5 ml/h

vitamins vitamins

stored for 2 hours

stored for 7 days

7.23 6.15

8.01 6.10

5. Vitamins addition decreases apparent pH

2011