nuclear decay and counting statistics: measure the half-life of a … · 2013-10-08 · nuclear...
TRANSCRIPT
Nuclear Decay and Counting Statistics:
Measure the Half-life of a Radioactive Nuclide with a Geiger-Müller Counter
Joshua Webster
Partners: Billy Day & Josh Kendrick
PHY 3802L
10/4/2013
Webster Lab 1: Geiger-Müller Counter
1
Abstract:
The purpose of this lab was to observe the nuclear decay of particles from known sources,
perform statistical calculations, and to finally observe the nuclear decay of an unknown
radioactive isotope in order to determine its decay constant, mean-life, and half-life within a
range of uncertainty. The results of this experiment conclude that the unknown radioisotope has
a decay constant of 4.77 ± 0.78 milliseconds-1
, a mean-life of 209.64 ± 34.19 seconds, and a half-
life of 145.31 ± 34.19 seconds.
Webster Lab 1: Geiger-Müller Counter
2
Table of Contents Abstract: ........................................................................................................................................................ 1
Introduction ................................................................................................................................................... 4
Background ................................................................................................................................................... 5
Part 1: ........................................................................................................................................................ 5
Part 2: ........................................................................................................................................................ 5
Part 3: ........................................................................................................................................................ 8
Part 4: ........................................................................................................................................................ 9
Part 5: ........................................................................................................................................................ 9
Experimental Techniques ............................................................................................................................ 12
Diagrams and Images .............................................................................................................................. 12
Part 1: ...................................................................................................................................................... 13
Part 2: ...................................................................................................................................................... 13
Part 3: ...................................................................................................................................................... 13
Part 4: ...................................................................................................................................................... 13
Part 5: ...................................................................................................................................................... 14
Data ............................................................................................................................................................. 15
Part 1: ...................................................................................................................................................... 15
Part 2: ...................................................................................................................................................... 19
Part 3: ...................................................................................................................................................... 19
Part 4: ...................................................................................................................................................... 20
Part 5: ...................................................................................................................................................... 20
Analysis ...................................................................................................................................................... 25
Part 1: ...................................................................................................................................................... 25
Part 2: ...................................................................................................................................................... 25
Part 3: ...................................................................................................................................................... 27
Part 4: ...................................................................................................................................................... 28
Part 5: ...................................................................................................................................................... 28
Discussion ................................................................................................................................................... 32
Part 1: ...................................................................................................................................................... 32
Part 2: ...................................................................................................................................................... 32
Part 3: ...................................................................................................................................................... 32
Part 4: ...................................................................................................................................................... 32
Webster Lab 1: Geiger-Müller Counter
3
Part 5: ...................................................................................................................................................... 32
Questions and Answers: .......................................................................................................................... 33
Conclusion .................................................................................................................................................. 35
Part 1: ...................................................................................................................................................... 35
Part 2: ...................................................................................................................................................... 35
Part 3: ...................................................................................................................................................... 35
Part 4: ...................................................................................................................................................... 35
Part 5: ...................................................................................................................................................... 35
Appendix ..................................................................................................................................................... 36
References ................................................................................................................................................... 38
Webster Lab 1: Geiger-Müller Counter
4
Introduction
Radioactive is the word used when describing nuclear isotope that emits radiation in the
form of particles (alpha and beta), and/or electromagnetic waves (gamma rays). The reason for
the emission of these particles (and radioactivity in general) is due to the instability of an isotope.
Radioactive sources can be naturally occurring or man-made, with half-lives varying from less
than a second to billions of years. The disintegration of an atom or its radioactive decay can be
measured using a device popularly known as a Geiger counter. In the case of the experiments
performed in this report a Geiger-Müller counter was utilized.
The Geiger-Müller (GM) counter consists of a tube (cathode) with a wire running through
it (anode), and is filled with easily ionized gas (in our case halogen). The GM counter was
connected to a counting device that also supplied the voltage and to an oscilloscope. The
oscilloscope was used to visually observe the pulses during detection. The shape of the pulse
corresponds to the cascade of electrons during an “avalanche”. This “avalanche” occurs because
electrons traveling to the anode of the Geiger-Müller tube ionize other atoms, which in turn,
causes a voltage drop that the counter recognizes. This voltage drop is then represented on the
oscilloscope as a pulse, signifying the presence of a detected particle. The avalanche is stopped
by the halogen gas inside the Geiger-Müller tube that acts as a barrier, forcing positive ions to
transfer energy to the gas molecules. Some of the halogen gas molecules are broken down in
place of ionization. Any of these molecules that are accelerated toward the cathode quickly
disassociate and can no longer produce a signal. Put simply, the Geiger-Müller counter acts as a
detector, which sends a signal to the “counting device” which records the number of counts
during a set time interval. An oscilloscope can also be connected to the setup to display a graph
of voltage versus time, and indicate the detection of particles by a “blip” in the graph.
The purposes of the experiments conducted in this report were to achieve a general
knowledge of the procedures involved in detecting radioactive particles and to understand the
statistical nature of radioactive decay. The final experiment conducted was to determine the
decay constant of an unknown radioisotope. The ability to determine a radioisotope’s decay rate
is an important scientific concept that allows us to have a greater understanding for the world we
live in and the universe that surrounds us. The sections of the report that follow will be: The
Background, consisting of theoretical expectations and mathematical derivations; Experimental
Techniques, providing insight into the equipment and procedures used during experimentation
and related diagrams; Data, listing all the data gathered from experiments; Analysis, where
calculations of values will be shown in detail, as well as discussion of uncertainties or other
errors involved; Discussion, in which results will be compared to expected values; Conclusion,
summarizing the results.
Webster Lab 1: Geiger-Müller Counter
5
Background
Part 1:
The first part of the experiment consisted of constructing a Geiger-Müller plateau graph
for our specific device. This will allow us to determine the optimum operating voltage. The
graph should begin from a zero value in counts to an abrupt jump in counts that begins to level
off, or plateau, before it eventually rises exponentially (the discharge region). After taking the
first measurements for the plateau, the slope of the plateau can be calculated. The value that
results from this calculation determines the Geiger-Müller devices reliability, and represents
percent per 100 V. If the value for the slope is greater than 10%, the accuracy of the GM tube is
compromised, and it should no longer be used. The equation for the slope is as follows,
( ) ( )
( )( ) ( )
For the equation above: N1 and N2 are the low end and high end counts respectively, and V1 and
V2 are the lowest and highest voltages.
To understand this equation better it breaks down to,
( )
( )
For the equation above: ∆counts represents the change in counts (N2-N1), ∆voltage represents the
change in voltage (V2-V1), and total counts represents the quantity (N2+N1). This equation is
similar to the standard slope formula, however, since the graph that we are finding the slope for
is not a straight line, we must use the change in counts divided by the change in voltage (instead
of ∆y/∆x), and then divide that by the sum of the high end and low end counts (to give you a
fractional percentage). The 2 multiplier term coupled with a 100 multiplier term normalizes the
equation to (% per) 100V, because in the experiment the voltage was reduced to about 200V
from the starting voltage. There is also a 100 multiplier to convert the fractional percentage to a
percent.
Part 2:
The next section of the experiment deals with finding the dead-time of the Geiger-Muller
counter. This is the amount of time that must pass after one decay is counted until a subsequent
decay can be counted. This is calculated using a split source. The first count is recorded, then
another split source is used and counts are recorded. Finally the sources are combined and the
combined counts recorded. The equation for dead-time, τ:
( )
Webster Lab 1: Geiger-Müller Counter
6
In the above equation, N1 is the count rate of the first half source, N2 is the count rate of the
second half source, Nc is the combined source count rate, and T is the time during counting.
Adapting equation 2 for the dead-time uncertainty, we first take the natural log of both sides,
then take the derivative to obtain dτ, and using the fact that σN = √N (or dN = √N) we square both
sides.
( ) ( ) ( ) ( ) ( )
(
) (
) (
)
(
)
(
)
(
)
(
)
√ (
)
(
)
(
)
( )
Now calculating the dead-time uncertainty, we use Poisson uncertainty for the number of
counts. Equation 2.2 is for calculating the mean. Equation 2.3 is for calculating the standard
deviation. The formulae are as follows:
∑
( )
For the equation above: xi, the counts, are summed over the number of counts, N.
√ √ ( )
The standard deviation, σ, for a Poisson distribution is simply the square root of the mean value
found before.
The dead-time can be used to make corrections to the observed count rate. Finding and equation
to account for the true counting rate:
For a dead-time, τ, we need to define the following variables:
real-time or count time
true time or time detector can record counts
Webster Lab 1: Geiger-Müller Counter
7
total counts
measured count rate, true count rate,
Taking the ratio of counts, we find:
⁄
⁄
We need an equation relating the dead-time. We can approximate the true time:
From the above equation: Cτ is the total time detector dead-time. Continuing, we divide the
equation for above by :
Finally, we relate the L.H.S. of the above equation to the R.H.S. of the equation for the ratio of
counts, and solve for R.
( )
For the equation above: is the true count rate, is the observed count rate, and τ is the dead-
time.
For the uncertainty in the true count rate we need to differentiate the above equation 3 with
respect to r, but since ⁄ is very small:
( ) ( )
Once the count rates are corrected, they should satisfy the following equation (within a margin of
error), where the c’s denote the corrected values:
( ) ( ) ( ) ( )
Webster Lab 1: Geiger-Müller Counter
8
Part 3:
During the next section we conducted another counting experiment, and determined the
mean, standard deviation for a Poisson distribution, individual deviations, and the standard
deviation (square root of the sample variance). Rewriting eq. 2.2 from part 2 algebraically to find
the mean value:
( )
( )
For the equation above: is the mean, x1-xN are the counts, and N is the number of counts. The
true mean can only be determined by averaging an infinite number of measurements, but for our
case (a finite set of measurements) the sample mean is the best approximation.
For a Poisson distribution, the standard deviation is the square root of the mean (also
known as the R.M.S. or Root-Mean-Square). This R.M.S. value represents the magnitude of the
deviations of individual measurements, as shown in equation 2.3 above.
Also in the case of this section of the experiment, the best approximation for the standard
deviation is the unbiased sample variance. The equation for the standard deviation is:
√∑ ( )
( )
Which can expressed algebraically as,
√( ) ( ) ( )
( )
For the equation above: σx represents the standard deviation, is the arithmetic mean (as
calculated before in part 2), xi is the count, and N is the number of counts. This equation shows
that the standard deviation is the sums of the arithmetic mean minus the count squared and
divided by the number of counts minus one, all square rooted.
When only a single observation can be made, the mean can be approximated to be the
single measurement, x, in place of the true mean (m or as we have in this report). The
uncertainty then, is just σx = √x.
For a gross counting rate standard deviation is:
√
√
√
( )
Webster Lab 1: Geiger-Müller Counter
9
For the equation above: σRg is the standard deviation of the gross counting rate, x is the counts,
Rg is the gross counting rate, and t is the time of the counting.
Part 4:
In this section, we conducted measurements of the background radiation over a period of
5 minutes with no source in the GM counter. This background radiation measurement plays an
integral part in the accuracy of all counts taken by a specific GM counter in a specific location.
The background readings are to be subtracted from all counts.
Part 5:
This part of the experiment deals with finding a decay constant of a rapidly decaying
unknown source. The decay constant will be determined by using the equations outlined in this
section.
The activity of a radioisotope, or rather the number of disintegrations per unit time, is determined
by the equation below.
( ) ( ) ( ) ( )
For the equation above: A(t) is the activity at time t, t2 and t1 are the beginning and end times (in
seconds) of the two activities taken into consideration, and λ is the decay constant (specific to a
certain radioisotope, representing the probability per unit time for a radioactive decay to occur).
Taking the natural logarithm of both sides of equation 8 we find:
[ ( )
( )] ( )
which can also be represented as,
[ ( )] ( ) [ ( )]
This is the equation that will be plotted later in this report, and from which the decay constant
can be easily identified. It shows that the decay constant, –λ, represents the slope of the equation.
Solving eq. 8 for λ, we find:
[ ( ) ( )
]
( )
Uncertainty in the decay constant can be found using equation A.3 from the appendix. Note: The
uncertainty in the time is small so it can be left out of the error propagation.
Webster Lab 1: Geiger-Müller Counter
10
√(
)
(
)
(
)
( )
√(
( ))
( ) (
( ))
( )
√(
( )( ))
( ) (
( )( ))
( ) ( )
To find the weighted average decay constant we use the formula:
∑
∑
( )
The variables in this equation are defined for our purposes as (the decay constant) and
(σi is the uncertainty in the decay constant).
The mean life of the isotope can be found using the decay constant in the following equation,
where τ is the mean life:
( )
We can find the uncertainty in the mean-life by taking the ln of equation 9 and differentiating:
( )
The half-life of a radioisotope is the interval of time in which the activity decreases to one-half
its value in the beginning of the time interval.
( )
( )
( )
To calculate the uncertainty in the half-life we must find the derivative of equation 10. It is easy
to do if we take the natural log of both sides, and we can also negate the ln(2) term, because it is
small:
Webster Lab 1: Geiger-Müller Counter
11
( ) (
)
( )
The relationship below determines the error bars for Graph 2 in Part 5 of the Data section.
σR is the uncertainty in the corrected counts, and R is the corrected count.
( ) ( )
Webster Lab 1: Geiger-Müller Counter
12
Experimental Techniques
Diagrams and Images
Diagram 1: This diagram depicts the device setup for the series of experiments performed in this report.
The only difference between this setup and our setup is that our scaler, timer, high voltage supply (HV),
and voltmeter were all in one device called the counter. Our setup also included an oscilloscope and GM
tube.
Image 1: The picture below was taken in the lab, and shows our exact laboratory setup. The
digital oscilloscope (Tektronix TDS 2002B) is on the left, SPECTECH ST360 counter (high
voltage supply, timer, and scaler) is in the middle, and GM tube is on the right. What might not
be easily visible is the makeshift cardboard shelf that we had to construct for our GM tube, as it
was missing its originally included glass shelf.
Webster Lab 1: Geiger-Müller Counter
13
Part 1:
A radioactive source was placed into the GM tube. The radioactive source used was
Cesium-137. The plateau curve for the specific GM tube used was found during this part of the
experiment. The voltage was raised on the counter/HV until counts were detected. The minimum
voltage required to detect counts was 760 V (starting voltage). The counter was set to a counting
time of 30 seconds, and the voltage was set to the starting voltage. The second shelf in the GM
tube was used, as it provided at least 1000 counts once the voltage was raised 50 V above the
starting voltage. Every 30 seconds of counting, the voltage and counts were recorded, and the
timer and counts were reset. The voltage was raised after each 30 second count time by 40 V. As
the voltage was increased, the count rate rose. The pulse height in the oscilloscope was around
3.6 V as depicted in Image 2 in the data section of this report. The highest voltage reached was
1200 V, and no exponential increase in counts was detected at this voltage. The data was then
plotted, and the slope of the plateau was determined. The optimum operating voltage was
determined by this experiment to be 1000 V.
Part 2:
A radioactive split source was used to determine the dead-time of the GM tube. The split
source used was Titanium-204. One half of the source was placed on the GM tube tray, and the
tray was placed on the second shelf under the GM tube. The HV was set to the determined
operating voltage of 1000 V. As a trial, the counter-timer was set to 60 seconds, and a count rate
higher than 20,000 was detected (which is desirable). An order of magnitude estimate on the
dead-time was made by observing the shortest time interval between two pulses on the
oscilloscope. The counter-timer was then set to 200 seconds, and the counts were recorded as N1.
Then, very carefully, the other half of the source was placed directly next to the piece already
counted. The timer was reset to 200 seconds, the count number was reset to 0, and the counting
for the combined pieces began. The count was then recorded as Nc. Finally, the first half counted
was then removed, leaving the final half to be counted. The timer was reset to 200 seconds, and
the count number was reset to 0. The count of the remaining half was recorded as N2.
Part 3:
Using Titanium-204, we adjusted the GM tube shelf height to produce about 2000 counts
per minute. The count-timer was set to 30 seconds, and the HV was set to the operating voltage
of 1000 V. A total of 10 separate counting intervals were made, and deemed “Set A”. Then,
another set of 10 separate counting intervals were made and deemed “Set B”. Statistical
calculations were then made using the equations found in the background section of this report.
Part 4:
Measurements were conducted in the section of the experiment to determine the
background radiation count. All radioactive sources were removed from the GM tube tray, the
tray was put back in place under the GM tube, the count-timer was set to 300 seconds (5
minutes), the HV was set to 1000 V, and the background radiation count was recorded.
Webster Lab 1: Geiger-Müller Counter
14
Part 5:
An unknown radioactive sample was obtained from the instructor. The count-timer was
set to 30 seconds, HV supply was set to the operating voltage of 1000 V, the source was placed
under the GM tube on the second shelf, and a total of 20 separate counting intervals were
recorded with as little time between each counting interval as possible. Once the 30 seconds for a
specific counting interval was over, the counter-timer was immediately reset to begin counting
the next 30 second interval. The decay constant was then determined using the equations found
in the background section of this report, as well as by the slope of the logarithmic graph found in
the data section.
Webster Lab 1: Geiger-Müller Counter
15
Data
Part 1:
Image 2: The picture below was taken in the lab, and shows the pulse shape as seen through the
oscilloscope for the Cesium-137 source used in part 1. This graph plots the voltage (2.00 V
increments) on the vertical axis, and time (500 μs increments) on the horizontal axis. Using the
graph, the voltage of the pulse is estimated to be at 3.6 V, and the pulse duration around 750
microseconds. Note: Channel 1 (CH1) is the channel that was being used. This image was taken
with a camera and digitally enhanced to improve contrast.
Webster Lab 1: Geiger-Müller Counter
16
Table 1: The voltage and counts represented in this table are the data collected from this part of
the experiment using Cesium-137. They comprise the points on the Geiger plateau graphs
(graphs 1 & 2 below). The voltage uncertainty is an estimated uncertainty based on the accuracy
of the device. The uncertainty for the counts was calculated using standard deviation for the
individual counts.
Plateau Points
Time (s) Voltage (V) Counts Count Uncertainty Voltage Uncertainty
30 800 11839 108.81 0.5
30 840 27434 165.63 0.5
30 880 32567 180.46 0.5
30 920 36325 190.59 0.5
30 960 40821 202.04 0.5
30 1000 43263 208.00 0.5
30 1040 46373 215.34 0.5
30 1080 47155 217.15 0.5
30 1120 49152 221.70 0.5
30 1160 49601 222.71 0.5
30 1200 50955 225.73 0.5
Webster Lab 1: Geiger-Müller Counter
17
Graph 1: This graph plots the plateau points given in Table 1 with counts as the vertical axis and
voltage in volts as the horizontal axis. The slope of the plateau part of this graph needs to be less
than 10% per 100 V for the GM tube to be considered reliable. This graph shows the fast jump in
counts after the first detection interval, and the plateau area which is consistent with theoretical
expectations. The vertical error bars were calculated using the uncertainty for a single
measurement, the horizontal error bars shown are ± 0.5 volts due to uncertainty in the accuracy
of the device. The manual for the HV supply (ST360) does not state the accuracy of the internal
voltmeter.
0
10000
20000
30000
40000
50000
60000
700 800 900 1000 1100 1200
Co
un
ts
Voltage (V)
Geiger Plateau Graph
Counts vs. Voltage
Webster Lab 1: Geiger-Müller Counter
18
Graph 2: This is another graph for the Geiger plateau which was created with the program
LineFit using the same data, but fitting a line to the points considered to be the “plateau”
(starting at 960 V). Error bars are included. The equation of the fit line shown is y = 41.5x +
1.90.
Webster Lab 1: Geiger-Müller Counter
19
Part 2:
The data shown in the section will be used to calculate the dead-time. The dead-time
uncertainty will also be calculated, as well as corrections to the observed count rates.
Table 2: This table lists the counts for the split source of Titanium-204. The first count, N1, is
for the first half. The second count, Nc, is for the two combined split sources. The third count,
N2, is for the second half. All counts were taken for a time of 200 seconds at the operating
voltage of 1000 V. The count rate is simply the counts for the trial divided by the time, and the
units are counts per second.
Split Source Titanium-204
Trial Time (s) Counts Count Rate
N1 200 70570 352.85
Nc 200 126998 634.99
N2 200 70903 354.52
Part 3:
In this section there were ten counts recorded for two sets (Set A and Set B), giving 20
separate data points as shown in Table 3 below. The source used was Titanium-204. The data
from the table is later used for statistical analysis.
Table 3: This table shows the data recorded after counting for 30 seconds a total of 10 times for
two separate sets (A and B). It also shows the individual count deviations from the mean, and
then shows that they sum to a value near zero.
Trial
Time
(s)
Set A
Counts
Set B
Counts
Set A Individual
Devs.
Set B Individual
Devs.
1 30 1117 1077 29 1
2 30 1061 1058 -27 -18
3 30 1080 1061 -8 -15
4 30 1053 1045 -35 -31
5 30 1044 1096 -44 20
6 30 1082 1038 -6 -38
7 30 1079 1063 -9 -13
8 30 1120 1140 32 64
9 30 1120 1101 32 25
10 30 1132 1077 44 1
Sums
8 -4
Webster Lab 1: Geiger-Müller Counter
20
Part 4:
A background radiation count was determined during this section of the experiment. The
voltage used was the operating voltage of 1000 V. The background count data in the table below
is used to adjust the count numbers in other parts of the experiment.
Table 4: This table shows that after a 300 second counting interval and background count of 707
was detected.
Background Count Time (s)
707 300
Part 5:
A rapidly decaying unknown source was obtained. The counts in Table 5 were recorded
at 30 second time intervals at the operating voltage of 1000 V. The data in Table 5 is used to
determine the decay constant and associated uncertainty. Image 3 displays the pulses on the
oscilloscope during the counting.
Image 3: This image shows the pulses on the oscilloscope during this part of the lab. The large
amount of pulses displayed in short time intervals represents a high count rate. The operating
channel of the oscilloscope is channel one (CH1). Voltage (in 1 V increments) is along the
vertical axis, and time (1.00 ms increments) is along the horizontal axis. This image was taken
with a camera and digitally enhanced to improve contrast.
Webster Lab 1: Geiger-Müller Counter
21
Table 5: Shown in this table are the counts recorded during 30 second counting time intervals.
Also shown are the corrected values accounting for the background count, and the dead-time.
The Corrected Counts column uses the dead-time equation and multiplies the result by the 30
second time interval. The Ln(Corrected Count Rate) column is for use in plotting the decay as a
linear function. The uncertainty for the count rate, σ(corrected count rate), was calculated using
equation 3.1, and the error for the log function was calculated using equation 11 in the
Background section. The time uncertainty was due to human error and was determined to be a
maximum of 2 seconds.
Unknown Source Data
Trial
Elapsed
Time (s) Counts
Count
Rate
Corrected
Count Rate
Corrected
Counts
Ln(Corrected
Count Rate)
σ(corrected
count rate) σln
σ(time)
(s)
1 30 11955 398.50 448.069 15442 6.10 25.50 0.06 2
2 60 10359 345.30 381.264 12854 5.94 22.94 0.06 2
3 90 9221 307.37 335.007 11128 5.81 21.12 0.06 2
4 120 8085 269.50 289.931 9493 5.67 19.31 0.07 2
5 150 7013 233.77 248.365 8027 5.51 17.59 0.07 2
6 180 6226 207.53 218.432 6994 5.39 16.31 0.07 2
7 210 5431 181.03 188.681 5987 5.24 14.98 0.08 2
8 240 4813 160.43 165.885 5227 5.11 13.93 0.08 2
9 270 4152 138.40 141.815 4436 4.95 12.77 0.09 2
10 300 3545 118.17 119.992 3729 4.79 11.65 0.10 2
11 330 3138 104.60 105.507 3265 4.66 10.88 0.10 2
12 360 2757 91.90 92.053 2837 4.52 10.12 0.11 2
13 390 2451 81.70 81.321 2498 4.40 9.48 0.12 2
14 420 2097 69.90 68.986 2112 4.23 8.71 0.13 2
15 450 1895 63.17 61.985 1893 4.13 8.25 0.13 2
16 480 1713 57.10 55.702 1698 4.02 7.81 0.14 2
17 510 1411 47.03 45.325 1378 3.81 7.05 0.16 2
18 540 1268 42.27 40.433 1227 3.70 6.66 0.16 2
19 570 1113 37.10 35.146 1065 3.56 6.22 0.18 2
20 600 980 32.67 30.621 927 3.42 5.83 0.19 2
Webster Lab 1: Geiger-Müller Counter
22
Graph 3: The graph below plots corrected counts vs. time for the unknown radioactive isotope.
For the vertical axis the error bars were calculated using equation 3.1 modified as described in
the appendix (A.5), and for the horizontal axis the error bars are due to human error and were
estimated to be ±2 seconds.
0.000
50.000
100.000
150.000
200.000
250.000
300.000
350.000
400.000
450.000
500.000
0 60 120 180 240 300 360 420 480 540 600 660
Co
un
ts
Time (s)
Unknown Source Counts vs. Time Graph
Counts vs. Time Plot
Webster Lab 1: Geiger-Müller Counter
23
Graph 4: This graph depicts the logarithmic of the counts of the unknown radioactive source vs.
time. Plotting logarithmically allows us to see the decay as a linear function. Using the natural
log of the count rates makes the slope of the fit line a good estimate of the decay constant (the
decay constant equals negative slope of the natural logarithmic function). There are vertical and
horizontal error bars, but the horizontal are very small (due to human error). The vertical error
bars are calculated using equation A.6 in the Appendix, and the values are shown in Table 5.
This linear fit accounts for x and y error, and the equation of the line is ( )
.
Webster Lab 1: Geiger-Müller Counter
24
Table 6: This table represents the data analysis of the individual decay constants and
uncertainties associated. It also shows the values for the average decay constant and the average,
standard deviation, and shows the values used to calculate the weighted average. This weighted
average of the decay constant is another means of finding the decay constant value.
Trial
Elapsed Time
(s)
Decay
Const. λ(s-1
)
Uncertainty
σi (s-1
)
Dev. From
Mean λi/σi2
Weight
(1/σi2)
1 30 n/a n/a n/a n/a n/a
2 60 5.38E-03 2.76E-03 4.55E-07 7.06E+02 1.31E+05
3 90 4.31E-03 2.90E-03 1.57E-07 5.11E+02 1.19E+05
4 120 4.82E-03 3.06E-03 1.20E-08 5.15E+02 1.07E+05
5 150 5.16E-03 3.24E-03 2.03E-07 4.91E+02 9.52E+04
6 180 4.28E-03 3.43E-03 1.82E-07 3.64E+02 8.50E+04
7 210 4.88E-03 3.63E-03 2.99E-08 3.70E+02 7.58E+04
8 240 4.29E-03 3.85E-03 1.72E-07 2.89E+02 6.74E+04
9 270 5.23E-03 4.10E-03 2.68E-07 3.10E+02 5.94E+04
10 300 5.57E-03 4.41E-03 7.44E-07 2.86E+02 5.13E+04
11 330 4.29E-03 4.72E-03 1.76E-07 1.92E+02 4.49E+04
12 360 4.55E-03 5.02E-03 2.57E-08 1.80E+02 3.96E+04
13 390 4.13E-03 5.34E-03 3.31E-07 1.45E+02 3.51E+04
14 420 5.48E-03 5.73E-03 6.02E-07 1.67E+02 3.05E+04
15 450 3.57E-03 6.11E-03 1.30E-06 9.54E+01 2.68E+04
16 480 3.56E-03 6.44E-03 1.31E-06 8.58E+01 2.41E+04
17 510 6.87E-03 6.98E-03 4.68E-06 1.41E+02 2.05E+04
18 540 3.81E-03 7.55E-03 8.10E-07 6.67E+01 1.75E+04
19 570 4.67E-03 8.06E-03 1.29E-09 7.18E+01 1.54E+04
20 600 4.59E-03 8.66E-03 1.30E-08 6.12E+01 1.33E+04
AVG Decay
Const. (λ)
Std. Dev. Sum
Weighted
AVG λ
4.71E-03 7.77E-04 5.05E+03 4.77E-03
Webster Lab 1: Geiger-Müller Counter
25
Analysis
Part 1:
Equation 1 is used to find the slope of the Geiger plateau found in Graph 1 using the data in Table
1. A slope of less than 10% signifies a GM tubes reliability. A GM tube with a slope greater than 10%
should not be used. A sample calculation is shown below.
( ) ( )
( )( ) ( )
( ) ( )
( )( )
Possible errors in this value can be attributed to the counts and dead-time, although dead-time
isn’t necessary to account for in this part of the experiment. Also, another error could also possibly be
associated with the HV. Since there was no separate voltmeter attached to the device, it could not be
verified that the HV supply was delivering every volt it stated it was. Furthermore, if the power outlet that
the HV supply was connected to had a drop in voltage due to environmental effects (improper wiring of
electrical outlets, or a bad circuit breaker) this could possibly affect the voltage delivered to the GM tube
resulting in systematic errors. It is not believed that this occurred during the experiment, but it also could
not be verified due to not having an available voltmeter to test/observe the voltage delivered by the outlet.
Doing this experiment in Part 1 and the other parts of this lab, we are working on the assumption that the
HV supply is able to maintain a constant voltage, and there are no internal faults in the device itself.
While these environmental errors are systematic (affecting all parts of this experiment) I will not re-iterate
those aforementioned.
Part 2:
This section of the analysis uses values from Table 2. Equation 2 is used to find the dead-time, τ,
of the GM tube. Equation 2.1 is also used to find the uncertainty associated with the dead-time, στ.
( )
( )( ) ( ) (
)
Now calculating the dead-time uncertainty using Poisson uncertainty for the numbers of counts,
we will use equation 2.1.
Webster Lab 1: Geiger-Müller Counter
26
√ (
)
(
)
(
)
( )
To make things a bit easier, calculate the common fractional term ahead of time:
( )√ (
)
(
)
(
)
( )√( ) ( ) ( )
Therefore,
To make a dead-time correction to count N1 using equation 3, the count rates need to be used
(shown in Table 2). The value then obtained will be the corrected count rate. This value then needs to be
multiplied by the count time, 200 seconds, to obtain the value for the corrected count.
( )
( )( )
For the equation below: N1c represents the corrected count value for N1.
(
) ( )
To calculate the error propagation we can use equation 3.1, noting that σr is just the √r:
( ) ( )
Webster Lab 1: Geiger-Müller Counter
27
( ( )( )
Using the same steps as before we obtain the values, and . These
values can now be checked by ensuring that they satisfy equation 4.
( ) ( ) ( ) ( )
This result is acceptable because it is within the uncertainty of counts. Although it is
insignificant, it should also be stated that the counter (ST360) itself has a resolving time (dead-
time). From the ST360 manual, “The resolving time of the ST360 RADIATION COUNTER is
very short and is not a significant factor compared to that of the GM tube.”1
Part 3:
Using the data from Table 3, equation 5 is used to determine the arithmetic average, and
equation 2.3 is used to determine the standard deviation of the mean. A sample set of
calculations for Set A of the average and standard deviation is shown below.
( )
( )
For Set A:
(
)
√ √ ( )
√
Now using equation 6.1, we find the standard deviation by taking the square root of the sample
variance.
1 (Spectrum Techniques, 2002)
Webster Lab 1: Geiger-Müller Counter
28
√( ) ( ) ( )
( )
For Set A:
√( ) ( ) ( ) ( )
7/10 counts in Set A fall within
For Set B: , , and .
7/10 counts in Set B fall within
The two mean values are fairly close with a difference of 12 between the two sets. The
values for sigma from set to set are very close. The sums of the individual deviations sum to
nearly zero as shown in Table 3.
Part 4:
The background radiation was measured in this part of the experiment. The background
radiation value recorded (707 counts in 300 seconds) seems a bit high. Our GM tube was located
near the box containing other radioactive sources, so this may have been the cause of the high
background count. Also, the GM tube was missing its glass shelf, so a shelf had to be constructed
out of cardboard. This cardboard could have added a minor effect to the background radiation.
Calculating the uncertainty associated with this single measurement:
√
Part 5:
Table 5 shows the corrected count values after doing the dead-time calculation in Part 3
and accounting for the background radiation. Accounting for the background radiation and dead-
time is an important step when accuracy is required; such is the case when determining the decay
constant of an unknown radioactive isotope. Though it is basic algebra, for thoroughness, sample
calculations to obtain one of the corrected counts is shown below.
Using Trial 10, which has 3545 counts, the count rate will be calculated, the background
radiation will be taken into account, and the dead-time correction will be applied. To calculate
the count rate, simply divide the counts by the time interval they were detected in.
Webster Lab 1: Geiger-Müller Counter
29
Accounting for background radiation we need to use the value obtained in Part 4, which
was 707 counts in 300 seconds. Since the counts were detected over a 300 second time interval,
we divide 707 counts by 300 to obtain the count rate.
The background count rate must then be subtracted from the count rate of the sample to obtain
the background corrected count rate for Trial 10.
Now the count rate needs to account for dead-time, so we apply equation 3.
( )
( )( )
To find the corrected count we simply multiply this (background and dead-time) corrected count
rate by the 30 second time interval.
(
) ( )
Using equation 3.1 we can account for the propagation of errors:
( ) ( )
( ( )( )
To determine the decay constant we use eq. 8.1. We can then find the mean-life using eq.
9. With the mean-life determined, we simply plug that value into equation 10 and solve for the
half-life. Finding the decay constant for the first two activity values in Table 5:
[ ( ) ( )
]
( )
Webster Lab 1: Geiger-Müller Counter
30
[ ]
Uncertainty in the decay constant can now be found using equation 8.2, noting that the
uncertainties in the activities at time 2 and time 1 were previously calculated and shown in Table
5:
√(
( )( ))
( ) (
( )( ))
( ) ( )
√(
( ))
( ) (
( ))
( )
Using the weighted average for the decay constant we found using the values tabulated in
Table 6, we can find the mean-life by inserting this value into equation 9. The mean life can also
be calculated using each individual decay constant or the average decay constant, but in theory
the more reliable value would be the weighted average decay constant, so we shall use it here.
( )
The uncertainty in the mean-life can be determined by using equation 9.1:
( )
( )( )
We can now solve the final equation (10) to find the half-life of the unknown radioactive isotope:
( )
( )
( )
Webster Lab 1: Geiger-Müller Counter
31
( ) ( )
( )
The uncertainty in the half-life is found using equation 10.1. The value used for σλ is the standard
deviation of the individual values, as calculated in Table 6.
( )( )
The decay constant can also be estimated by taking the negative value of the slope of the
logarithmic graph, which was determined to be 4.71 ± 0.14 ms-1
. The error on this value is given
by the program LineFit as part of the equation of the line. This value is fairly close to the value
arrived at by using the weighted average decay constant.
Human errors also come in to effect with Part 5. The errors can be attributed to the fact
that a human cannot press the button on the count-timer (in order to reset the counts and begin
counting again) at the exact moment the prior 30 second interval has finished. This error shows
up in the counts, which in turn affects the count rate. These effects are more pronounced when
viewing the graph, resulting in a slightly imperfect linear graph (when in theory it should be
perfectly linear).
Webster Lab 1: Geiger-Müller Counter
32
Discussion
Part 1:
The results for the Geiger plateau are considered reliable. The slope was determined to be
9.20% and the accepted value for reliability is any value under 10%.
Part 2:
The value obtained for the dead-time of the Geiger-Müller tube, 292.29 ± 9.29 μs, seems
to be an acceptable value. A reference to known or acceptable dead-times was not found, so a
comparison could not be made. After comparing this result with colleagues, it was found to be an
accepted value.
Part 3:
The statistical calculations performed in Part 3 produce anticipated results. The two mean
values are fairly close with a difference of 12 between the two sets. The values for sigma from
set to set are very close. The sums of the individual deviations sum to nearly zero as shown in
Table 3.
Part 4:
At first glance 707 counts seems like an above average background radiation count. Upon
further inspection, it was realized that our GM tube was located near the box that contained all of
the radioactive sources. If repeated, it would be preferable to keep the GM away from any
coincidental radioactive sources. A smaller background count could result in fewer errors due to
a fluctuating background count. Furthermore, repeated background counts allowing for the
statistical treatment of the background would result in an even greater accuracy.
Part 5:
The values obtained for the decay constant, mean-life, half-life, and associated
uncertainties are listed in a table in the conclusion for easy comparisons. The half-life seems to
be fairly short. It is definitely within the broadly acceptable values of half-lives.
Webster Lab 1: Geiger-Müller Counter
33
Questions and Answers:
What are physical phenomena that are initiated by the passage of an ionizing particle in the GM
tube?
The Townsend avalanche (or gas multiplication) is a physical phenomenon initiated by
the passage of an ionizing particle in the GM tube. This “avalanche” occurs because electrons
traveling to the anode of the Geiger-Müller counter ionize other atoms, which in turn, causes a
voltage drop that the counter recognizes. This voltage drop is then represented on the
oscilloscope as a pulse, signifying the presence of a detected particle. The avalanche is stopped
by the halogen gas inside the Geiger-Müller counter that acts as a barrier forcing positive ions to
transfer energy to the gas molecules. Some of the halogen gas molecules are broken down in
place of ionization. Any of these molecules that are accelerated toward the cathode quickly
disassociate and can no longer produce a signal.
Coincidence is another physical phenomenon. A decaying particle or a ray can produce
an ion pair by ionization inside the GM tube. Electrons arrive at the anode faster than the positive
ions arrive at the cathode. While the positive ions are still traveling to the cathode there is a short
time interval in which no other radiation can be detected. The other “coincidental” radiation
produces another Townsend avalanche that effectively adds to the one already present. The GM
tube then sees this as one large avalanche.
What makes it possible for the GM tube to also count X-rays and γ-rays (photons)?
An absorber can be used to shield the tube from incident X-rays. Taking into account
counts recorded for a source with no absorber present, and counts with an absorber present the
X-rays can be differentiated from the γ-rays.
What is the size (in V) of the smallest signal from the GM tube that is big enough to trigger the
counter circuit?
From the oscilloscope it seemed that 1 V was large enough to trigger the counter. I am
not entirely certain; it could be that less than 1 V could trigger the counter. The starting voltage
on the counter however, was around 760 Volts to produce any counts.
What is the smallest time between two successive GM tube output signals that you can see on the
oscilloscope?
From close examination of the pulses on the oscilloscope, the smallest time interval
between two pulses was recorded to be around 50 μs. Unfortunately an image was not taken of
this. It can be seen in Image 3 that each block has a time interval of 1.00 ms. Some of the pulses
in the image seem to overlap. From this we can determine that the time between two successive
pulses must be much less than 1.00 ms.
Webster Lab 1: Geiger-Müller Counter
34
Given the dead-time determined by you, what is the counting rate at which the error due to the
finite dead-time is 1% of the rate?
as calculated in the Appendix section (A.3)
When taking the counts for the decay constant measurement, why is it important that the time
between two successive measurement periods be minimized? If you had appreciable delays
between successive counting periods (without correcting for this), would that cause you to over-
or underestimate the half-life? (Explain)
When trying to determine the decay constant of a radioactive isotope it is important to
minimize the time between counts. This is because the isotope is rapidly decaying, and any
added time between two counts produces an error if it is not accounted for and corrected. This
added time would result in a lower count, because the count is continually decreasing. Without
correction, the original total time (t2-t1) would still be used. Solving the equations outlined in part
5 using a lower count rate along with the same time interval would result in the decay constant
being underestimated.
Webster Lab 1: Geiger-Müller Counter
35
Conclusion
Part 1:
The slope was determined to be 9.20% per 100V, which is an acceptable value. It isn’t a
great value, but it is under the 10% per 100V which is necessary for accuracy.
Part 2:
The dead-time of the device was determined to be 289 ± 9.29 μs. Possibly more advanced
or specialized equipment would improve this dead-time.
Part 3:
For Set A, the mean was determined to be 1088, the standard deviation was 32.98, the
root-mean-square deviation was 31.44. For Set B, the mean was determined to be 1076, the
standard deviation was 32.80, and the root-mean-square deviation was 30.27.
Part 4:
Background radiation was determined to be 707 ± 26.6 counts over a period of 300
seconds. A longer background count time and/or more individual background counts would yield
more accurate results for use in Part 5.
Part 5:
The values obtained for the decay constant, mean-life, half-life, and associated
uncertainties are listed in the table below.
Decay Constant
λ (s-1
)
Mean-life
τ (s)
Half-life
T1/2 (s) σλ (s-1
) στ (s) σT1/2 (s)
Average 4.71E-03 212.31 147.17 7.78E-04 35.07 16.85
Weighted Average 4.77E-03 209.64 145.31 7.78E-04 34.19 16.43
From Graph 4.72E-03 211.86 146.85 1.32E-04 5.93 2.85
Equipment that could record counts at set intervals and reset automatically would greatly
improve the results in this part of the experiment.
Webster Lab 1: Geiger-Müller Counter
36
Appendix
A.2 Graph: This is another graph of the natural logarithm of count rates. Made in Excel, it
shows a slightly different perspective. Horizontal and vertical error bars are shown. Vertical
errors calculated by the standard deviation for each individual count. The plot points are fit with
a line. This linear fit does not take the error bars into account like the Graph 4 does. The equation
of the line is also shown, with the coefficient of determination. We can see that without taking
the error bars into account the value of the slope changes slightly.
A.3 Propagation of errors equation:
For an equation, f, that is dependent on the variables x, y, …, n the equation below is used for
finding the uncertainty in f that has propagated from the other variables uncertainties (σ).
√(
)
(
)
(
)
( )
y = -4.691E-03x + 6.227E+00 R² = 9.996E-01
2.50
3.00
3.50
4.00
4.50
5.00
5.50
6.00
6.50
0 100 200 300 400 500 600 700
Log(
cou
nt)
Time (s)
Logarithm of Counts Graph
Series1
Linear (Series1)
Webster Lab 1: Geiger-Müller Counter
37
A.4 Rate at which the error due to dead time effects (i.e. the difference between raw and
true counting rate) is 1%.
Using equation 3:
( )
We want to satisfy the following relation:
Solving for :
( )
Now plug this value back into equation 3:
( )
( ) ( )
We know τ, so we can solve for R:
A.5 Calculating the error bars for Graph 3:
In order to calculate the error bars for Graph 3, we must use a modified version of equation 3.1
that accounts for the uncertainty in the count being the Poisson distribution for an individual
count. That is, σr is just √r.
( ) ( )
A.6 Calculating error bars for Graph 4 (logarithmic decay graph):
To calculate the errors associated with the logarithmic values the following equation is used:
( )
Using the first trial values from Table 5:
( )
Webster Lab 1: Geiger-Müller Counter
38
References
Lindberg, V. (2000, July 1). Uncertainties and Error Propagation. Retrieved September 19,
2013, from rit.edu: http://www.rit.edu/~w-uphysi/uncertainties/Uncertaintiespart2.html
Spectrum Techniques. (2002, June). Instrument Manuals. Retrieved September 19, 2013, from
SPECTECH: http://spectrumtechniques.com/manuals/ST360manual.pdf
Spectrum Techniques. (n.d.). Lab Manuals. Retrieved September 19, 2013, from SPECTECH:
http://spectrumtechniques.com/Lab_Manuals/Studentmanual.pdf
Stuve, E. M. (2004, December). Estimating and Plotting Logarithmic Error Bars. Retrieved
September 19, 2013, from University of Washington:
http://faculty.washington.edu/stuve/uwess/log_error.pdf