nova phox - service manual

110
Service Manual

Upload: armando-aguilar

Post on 09-Dec-2015

296 views

Category:

Documents


35 download

DESCRIPTION

Manual de Usuario, Gasometro Nova Biomedical

TRANSCRIPT

Service Manual

i

Preface pHOx Service Manual

Stat Profile® pHOx® Service Manual

Part Number and Ordering Information

The Stat Profile® pHOx® Service Manual (PN 24301) can be ordered from Nova Biomedical Order Services. Write or call:

Nova Biomedical200 Prospect StreetWaltham, MA 02454-9141

U.S.A.Tel: 1-800-458-5813 FAX: (781) 893-6998 (in the U.S.A.) or (781) 899-0417 (outside the U.S.A.)

Technical Assistance (USA)For technical assistance, call Nova Biomedical Technical Services at:

Tel: 1-800-545-NOVA or FAX: (781) 894-0585 (781) 894-0800

TrademarksStat Profile® is a registered trademark of Nova Biomedical.pHOx® is a registered trademark of Nova Biomedical.

CopyrightPrinted in the U.S.A. Copyright 2005, Nova Biomedical, Waltham, MA 02454.

ii

Preface pHOx Service Manual

PN 24301 Rev. B 2005/11 TOC-1

Stat Profile pHOx Service Manual

Contents1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

1.1 The Stat Profile pHOx Analyzer ............................................................................1-11.1.1 Environmental ............................................................................................1-11.1.2 Energy consumption ..................................................................................1-11.1.3 Physical Dimensions ..................................................................................1-11.1.4 Regulatory Compliance .............................................................................1-11.1.5 Location - Stable stationary surface. .........................................................1-21.1.6 Operational .................................................................................................1-21.1.7 General .......................................................................................................1-2

1.2 Cautions and Hazards ............................................................................................1-31.3 Required tools .......................................................................................................1-4

2.0 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . 2-12.1 Mechanical Assemblies ..........................................................................................2-1

2.1.1 Pump ..........................................................................................................2-22.1.2 Sampler Assembly ....................................................................................2-22.1.3 Rotary Valve Assembly .............................................................................2-22.1.4 Sensor Module Assembly .........................................................................2-3

2.2 Electronic Assemblies. ...........................................................................................2-32.2.1 Power Entry Module .................................................................................2-32.2.2 Power Supply ............................................................................................2-32.2.3 Digital Control Board ...............................................................................2-42.2.4 Analog Board .............................................................................................2-62.2.5 SO2 Control Board .....................................................................................2-62.2.6 Electrode Interface Boards .........................................................................2-72.2.7 Door Assembly Electronics ........................................................................2-7

Table of Contents

TOC-2 PN 24301 Rev. B 2005/11

3.0 Replacement Procedures . . . . . . . . . . . . . . . . . . . . 3-13.1 Cover Removal .......................................................................................................3-13.2 Tubing Harness .....................................................................................................3-23.3 Power Supply Removal ..........................................................................................3-43.4 Sampler Assembly .................................................................................................3-53.5 Rotary Valve Assembly ........................................................................................3-103.6 Valve Assembly ....................................................................................................3-113.7 Rotary Valve Motor ..............................................................................................3-123.8 Pump Assembly ...................................................................................................3-133.9 Digital Control Board ..........................................................................................3-143.10 Analog Board ......................................................................................................3-153.11 Printer Assembly .................................................................................................3-163.12 Sensor Module Light Source ...............................................................................3-173.13 Door Removal /Reinstallation Procedure ...........................................................3-183.14 Display Assembly ................................................................................................3-203.15 Distribution board ...............................................................................................3-213.16 Keypad Assembly ...............................................................................................3-213.17 Sensor Module Assembly ....................................................................................3-223.18 Electrode Interface Boards ...................................................................................3-233.19 Pinch Valve Assemblies .......................................................................................3-24

4.0 Troubleshooting Information . . . . . . . . . . . . . . . . . . 1-14.1 Service Screens ......................................................................................................1-3

4.1.1 System Test Screen ....................................................................................1-44.1.1.1 Operator Flow Test .........................................................................1-44.1.1.2 Service Flow test ............................................................................1-5

4.1.2 Analog Input Screen ..................................................................................1-64.1.3 Sensor Subsystem .....................................................................................1-74.1.4 Printer Menu ..............................................................................................1-84.1.5 Communication Tests .................................................................................1-9

4.2 Flow Problems .....................................................................................................1-104.2.1 Isolating a Plug in the W/R to Waste Tubing ...........................................1-114.2.2 Isolating a flow problem in the Sampler/Sensor module .........................1-114.2.3 Isolating a flow problem between the Fluid Pack and the Fluid Fountain 1-12

4.3 Status Codes .......................................................................................................4-134.3.1 Analyte Troubleshooting ( Series code 1-6 and Digit codes 1-4) ............4-184.3.2 Single Channel Failures ......................................................................4-204.3.3 Temperature Problems .............................................................................4-214.3.4 Air Detector (ADT ) Problems ................................................................4-21

Operator Flow Test ...............................................................................................4-26 Sensor Module Back Flush Procedure .................................................................4-27

PN 24301 Rev. B 2005/11 TOC-3

Stat Profile pHOx Service Manual

A Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1A.1 Quick Reference Guide .........................................................................................A-1A.2 General Maintenance/QC ......................................................................................A-8

A.2.1 Other Analyzer Maintenance ..................................................................A-12A.3 pHOx Electrode/Standards ..................................................................................A-13A.4 Electronics and Mechanicals Overview ..............................................................A-13A.5 Shutdown Procedure ...........................................................................................A-14A.6 Start Up Procedure ..............................................................................................A-14 Installation Check List ........................................................................................A-15 Preventive Maintenance Check List ....................................................................A-17 Explosion Parts List ............................................................................................A-19

S Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-1#10-1087A Wiring Doucument ...................................................................................... S-1S-2 Wireing Charts ............................................................................................. S-2#10-1071B 24292-01 pHOx Digital Board Processor and Memory ............................. S-3#10-1072B 24292-02 pHOx Digital Board Comm. Printer, Display Touch Memory ... S-4#10-1073B 24292-03 pHOx Digital Board Motor & Valve Drivers .............................. S-5#10-1074B 24292 pHOx Digital Board Assembly Drawing .................................... S-6#10-1075A 21642 pHOx SO2 Board ........................................................................ S-7#10-1076B 24727 pHOx Distribition Board ............................................................ S-8#10-1077B 24291 Analog Board - Cortex Assembly Drawing .............................. S-9#10-1078B 24292-01 New Gen. Sensor Board ........................................................... S-10#10-1079B 24292-02 New Gen. Sensor Board ........................................................... S-11#10-1080B 24292-03 New Gen. Sensor Board ........................................................... S-12#10-1081B 24292-04 New Gen. Sensor Board ........................................................... S-13#10-1082B 23897 Pump Assembly ........................................................................ S-14#10-1083B 23899 Sampler Assembly .................................................................... S-15#10-1084B 23898 Rotary Valve Drive Assembly ................................................... S-16#10-1085B 24614 Printer Interface IF3001 Seiko ................................................. S-17#10-1086A 24720 Power Supply Universal input .................................................. S-18

Table of Contents

TOC-4 PN 24301 Rev. B 2005/11

1-1

1 Introduction1. Intro.

1.0 Introduction

This manual contains information and procedures used in servicing the Stat Profile pHOx Analyzer. The intent is to supplement the information provided in the Stat Profile pHOx Instructions for Use Manual (PN 37865) and the Stat Profile pHOx Advanced User Applications Manual (PN 35235). This manual does not supersede any current performance specification claims, expendable maintenance proce-dures, or warranty criteria as outlined in the reference manual.

1.1 The Stat Profile pHOx Analyzer

1.1.1 Environmental

Ambient Temperature: 15° - 32° C ( 59° - 86° F)Ambient Humidity: 0% - 95% Non-condensing

1.1.2 Energy consumption

Air Conditioning Load: Peak 44 BTU/HourPower Consumption: ~130 watts peakPower requirements: 100-120; 230-240 VAC, 50/60 Hz

• 2 Amp Time Delay (SB 2A or T2A) at 100-120VAC line• 1 Amp Time Delay (T1A) at 220-240 VAC line

1.1.3 Physical Dimensions

Weight: 8.1 Kg (18 lbs)Dimensions: 38 cm high (15 in) x 31 cm (12 in) wide x 38 cm (15 in) depth

1.1.4 Regulatory Compliance

The analyzer is tested and acceptable to attach the CSA, GS (Europe), JIS (Japan), and CE (self declaration) safety marks, and compliant with EN1010-1, EN55011, and IEC 801-2.

1-2

pHOx Service Manual

1.1.5 Location - Stable stationary surface.

• 8 cm ( 3 in) clearance to a wall or adjacent instrumentation• Not intended for use on vibrating (helicopter, etc.) surface• Placement on a cart is acceptable if the cart is stationary during operation.

1.1.6 Operational

The analyzer is designed to be on at all times. It may be turned off without special procedure or consideration for up to 1 hour with no detrimental effect on the sen-sors or fluidics.The tubing should be flushed and the pump tubing relaxed prior to extended shut-down periods.Any shutdown period will require a 2-point calibration and verification by control material upon restarting operation.

1.1.7 General

The analyzer utilizes an Intel 386EX processor, PCMCIA ROM programming.There are three RS-232 connectors, utilizing ASTM protocol, and one Bar Code reader connection.The display is a backlit liquid crystal (LCD) with type that is CCFT backlit with 320 x 240 pixel resolution.The internal optional printer is a 20 column thermal international font /166 dot graphics printer with ~5 cm (2 in) wide paper.

NOTE: The circuit boards use fine pitch surface mount technol-ogy. Field repair of these circuit boards is not possible. DO NOT attempt to repair any components on the board.

1-3

1 Introduction1. Intro.

1.2 Cautions and Hazards

There are NO user serviceable assemblies inside the analyzer. Only a trained, authorized service representative should remove the cover of this analyzer.

WARNING: Removal of the top cover allows access to power supply voltages. Care should be taken to avoid electrical shock. Remove the power cord prior to accessing any internal assemblies.

WARNING: Blood samples and blood products are potential sources of hepatitis and other infectious agents. Handle all blood products and flow path components (waste-line, capillary adapter, probe, sensor module, etc.) with care. Gloves and protective clothing are recommended.

NOTE: This International Caution Label appears on the rear of the pHOx Analyzer and means refer to the manual.

It is the responsibility of the service representative to decontaminate any assembly or analyzer being returned to Nova for repair or warranty claim.Decontamination of external surface of the flow path may be accomplished by a wipe down with cleaning agent or 10% bleach solution.The Internal surfaces of the flow path and tubing should always be considered contaminated. Prior to removal from the analyzer, ensure that the flow path and tubing are emptied.

WARNING: Internal surfaces may have sharp edges. Care should be taken to avoid cuts and scrapes when accessing internal assemblies.

Used tubing, biosensors, electrodes, reagents, controls, etc. may be disposed by normal laboratory waste procedures.

1-4

pHOx Service Manual

1.3 Required tools

The analyzer is designed to require no unique or special tools. Flow path cleaning wires/kits are provided in the accessory kit. Avoid metal wires or other material that may scratch the internal surfaces of the flow path.

• Digital Volt Meter (2 decimal accuracy)• Phillips® screwdriver (crosshead) preferably one with 15 cm (6 in) shank• Spring clip expander wrench or comparable means of releasing tension of

retention clips• Pliers - needle nose or small snub nose

2-1

2 Product Description2. Descrip.

2.0 Product Description

The pHOx analyzer accepts syringe or capillary samples through a sample probe. The operator selects the sample mode via the keypad. The probe moves from its home position in the analyzer to an extended 45° position for syringe sampling. If the capillary sample is selected, the probe moves to a horizontal position that is back inside the capillary adapter.

2.1 Mechanical Assemblies

The analyzer is designed to calibrate using calibration stan-dards and reference solution contained in a reagent pack. The tubing automatically con-nects with the pHOx when the pack is placed into the analyz-er. Likewise, a space is pro-vided for a plug-in control pack. The content and se-rial number of the pack is coded in a “Dallas Semicon-ductor Add-Only Memory,” which is read by a One-Wire MicroLAN® system. The ana-lyzer has 4 major mechanical assemblies to move calibrator or sample in front of the biosen-sors.

Figure 2-1 Flow Diagram

2-2

pHOx Service Manual

2.1.1 Pump

The analyzer uses one pump (PN 21492) for 2 functions. The pump aspirates either sample or calibration standards from the sample probe, through the sensor module/reference electrode, through a solenoid actuated pinch valve, and delivers to the waste bottle of the reagent pack.The second function of the pump is to aspirate the reference solution from the reagent pack and deliver it to the reference electrode through a solenoid actuated pinch valve.

2.1.2 Sampler Assembly

The sampler assembly (PN 21504) utilizes a probe that rotates to 1 of 3 possible positions. The HOME position is vertical with the tip of the probe seated against a "Fluid Fountain” assembly that is the common port of the rotary valve.The syringe position is the probe tip extended out in front of the analyzer at ap-proximately a 45° angle. The capillary position is horizontal, but the tip is totally inside the capillary adapter. This allows the capillary to be used without a separate adapter. (This position is also used to change the capillary adapter.) The capillary adapter is pushed onto the end of the sampler. Extending the probe outward from the capillary position ensures that the probe will be seated inside of the adapter. The sampler will exert enough force to push the adapter off of the sampler if it is not aligned correctly.

2.1.3 Rotary Valve Assembly

The rotary valve assembly (PN 22215) has one common port connected to the sampler’s fluid fountain. Each calibration standard and quality control standard is connected to a port on the rotary valve. The rotor positions a central port con-necting one of the fluids to the outlet port. The outlet port is connected to the Fluid Fountain by the S-Line. Starting the pump aspirates the selected fluid into the flow path.The rotary valve's manifold also has positions that allow air to enter the S-line. Re-fer to the replacement procedures for actual positions.

2-3

2 Product Description2. Descrip.

2.1.4 Sensor Module Assembly

The Sensor Module Assembly contains the measuring sensors, flow cell, and preheater. The measuring sensors clip into place upon insertion. The sensors have contacts instead of cabled connectors. When mounted, the sensor module forces the sensor contacts to align with the matching contacts on the left/right intercon-nect boards. The module also contains the light sources and detector for the SO2% channel and 2 air detectors. Please refer to the drawing section and Appendix A for individual part numbers.

2.2 Electronic Assemblies.

2.2.1 Power Entry Module

The power entry module (PN 22818) provides the interconnect to wall power. The fuses are found in the accessory kit for the operating voltage (120 VAC PN 22814, 236 VAC PN 22817). The power entry module connects to the power supply through TB1.

2.2.2 Power Supply

The power supply (PN 21602) provides all of the power requirements for the ana-lyzer. The analog and logic voltage used in the pHOx are + 5VDC. The stepper motors are driven by +24 VDC. This module has no repairable parts inside. The power supply connects to the digital control board by TB2.

2-4

pHOx Service Manual

2.2.3 Digital Control Board

There are 2 Digital Control Boards: all units with monochromatic display use PN 24292, and all units with a color display use PN 40797. The main controlling board of the analyzer uses an Intel 386X processor with the following integrated peripheral functions:

• Clock and Power Management unit• Watchdog Timer Circuit• Asynchronous Serial I/O unit• Synchronous Serial I/O unit• Parallel I/O unit• Refresh control unit• JTAG- Complaint Test Logic Unit

The digital control board provides one 72 pin socket for a 4 MB SIMM module of RAM. This volatile memory stores the calibration, slope, and millivolt data.Major circuit identification:

U25 is the main clock generator; it is crystal controlled 14.318 MHz and a phase-locked-loop frequency generator. This generates a 50.11 MHz CPU clock, 14.318 MHz for VGA video timing, and 1.84 MHz for Baud rate generation.

U35 is the real time clock containing a battery which will retain the date/time, setup options, daily QC (up to 144 points), Error log (last 96 errors FIFO), and sample counter data. The battery has an expected 10 year life.

U37 PEROM stores the monthly (256 data points) and YTD QC data, QC setup information, and recognition of the QC fluid pack presence.

PCMCIA contains the operational program and boot program. This card must be in place for the unit to function.

U43 is the video control chip. U26 controls the printer output and the I/O ports Comm 1 and Comm 2. U36 is the DUART that controls the Comm 3 I/O port.

U16 provides the serial interface for the (RMS) interface and fluid pack memories.

The mechanical devices are controlled by U18. The Keypad and “home” sensors for the rotary valve and sampler are ported through U18. The stepper motors are 4 phase stepper motors. U18 controls the speed and direction each phase is pow-ered. U3 and U4 provide the power for each winding of the pump motor. U5 and U6 power each phase winding of the sampler motor. U7 and U8 power each phase of the rotary valve motor.U18 also provides the logic signal to operate the other mechanical devices. Typi-cally, the devices are connected to either 12 VDC or 24 VDC and the logic provides a ground path through a transistor. The waste valve solenoid (12 VDC) grounds through Q3, the reference solenoid (12 VDC) through Q4, the sensor module lamp (24 VDC) through Q6, and the preheater (24 VDC) through Q2. The 24 VDC valve boost circuit is on the board but not used on the pHOx analyzer.

2-5

2 Product Description2. Descrip.

Power is received from the power supply through the J7 connector on the digital control board.Power Supply Connector TB2

1 2 3 4 5 6 7 8 +12 VDC. -12 VDC +24 VDC. Ground Ground Ground +5 VDC +5 VDC

Digital Board Connector J7

Pin 1 2 3 4 5 6 7 8 Ground +24 VDC. +5 VDC +5 VDC Ground +12 VDC. Ground -12 VDC

Sensing Wire Color Black Violet White Red Black Orange Green Yellow

Individual circuits are protected from an over current condition by the use of Posi-tive Temperature Coefficient Resistors (PTC). These resistors act as a direct short if the current is below a set current. If the current increases, the resistors heat up and open, similar to a fuse. When the device cools off, the connection is remade, allowing the circuit to function. The following lists these devices:

R22 - Solenoid common boost circuit (not used on pHOx)R24 - Heater power circuitR29 - Printer motor powerR43 - Print Head heater powerR44 - +24 VDCR45 - +5 VDCR71 - Bar Code Reader (+5 VDC out)R249 - Distribution board power (This board is mounted in the door.)

The LED indicators on this board are mostly unused. All will light at power up and then go off. The exception is D8 - 100 Hz Clock. D9 - 386 Strobe will appear to be always on. D7 is your program fail indicator and will come on with a software reset.Connectors:

J1 Bar Code ReaderJ2 COMM 1J3 COMM 2J4 COMM 3J5 JTAGJ6 PCMCIAJ7 Power SupplyJ8 Mechanical AssembliesJ9 Printer ConnectorJ10 Analog Board J11 - Display/Keypad (Motors/heater/lamp/solenoids and door)

2-6

pHOx Service Manual

2.2.4 Analog Board

There are 2 Analog Boards: pHOx, pHOx Basic, BiopHOx use PN 24291 and pHOx Plus, pHOx Plus L, and pHOx Plus C use PN 33390. U34 is the micropro-cessor that controls the signal acquisition and digital conversion. The signals are connected through several connectors.

• Left Sensor Interface board J6 - PCO2, pH, Reference, Air Detector (ADT) 4• Right Sensor Interface board J7 - PO2, Na+

• Sensor module J4 -SO2% detector, Preheater temp, ADT2, ADT3/Hct• Mechanical assemblies J3 - ADT1

Barometric pressure sensor U9 is on the analog board. The Analog to Digital Conver-sion (ADC) is performed by U31. The signals are multiplexed through U26 and U32.Preheater temperature control circuits are on the analog board. +24 VDC are ported through J3 to the SO2%/sensor module board (J4). The heater power wir-ing is protected on the digital control board by a PTC Resistor; the Heater has an in line thermostat (50°C) in the sensor module. The thermistor has 18.25 Kohms resistance at 37°C. The signal will change at a rate of 22.5 millivolts/0.1°C.There are 4 ADTs in the system. ADT1 holds the sample probe (40 uL setting). ADT2 at the inlet to the sensor module (70 uL). ADT3 is at the input to the sensors and is also used for the Hct. ADT4 is in the reference electrode housing.Logic and analog power is supplied to the analog board from the digital control board. Logic voltage (+5 VDC) enters J1 pin 14. The analog +5 VDC enter J1 pin 3 and -5 VDC J1 pin 4. The analog board provides power to the SO2 board through +5 VDC J4 pin 5 and -5 VDC J4 pin 6.

2.2.5 SO2 Control Board

SO2 Control Board is an integral part of the Sensor module. It controls the 2 LED light sources that glow at 880 nM and 660 nM, respectively.

2-7

2 Product Description2. Descrip.

2.2.6 Electrode Interface Boards

NOTE: The electrode interface boards are found only on pHOx and pHOx Basic.

The Electrode Interface Boards have 2 part numbers: PN 20641 (Left) and PN 20674 (Right). The electrodes are wireless. The electrodes have point contacts which engage the Interface board contacts when inserted into the sensor module. The left interface board, connecting to the analog board through J6, has 3 sets of contacts: one set for the reference electrode/ADT4, one set for Na+, and one set for PCO2 sensors. The reference electrode/ADT4 has 3 individual point contacts while the Na+ and PCO2 have 2 individual point contacts.The right interface board has 2 sets of contacts: one for the pH and one for the PO2 sensors; each consisting of 2 individual point contacts. This board is connect-ed through J7 to the analog board.

2.2.7 Door Assembly Electronics

The door assembly electronics include the Display Assembly, the Distribution Board, and the Keypad Assembly. The digital control board sends/receives signals to the door via the flat cable J104 on the distribution board.

NOTE: See Section 3.13 for part numbers that are model specific.

The keypad, which is glued onto the door, connects to the distribution board via J102 and has a ground cable clamped to the conductive surface on the inside of the door. The Analyze key also connects to the distribution board through this con-nector.The distribution board also provides the signal path to the display assembly and power to the display back lighting lamp. The display is a liquid crystal, graphics dis-play. It has a 320 x 240 dot matrix (1/4 VGA) yielding a 122 x 92 mm (4.99 x 1.33 inch) viewing area. This translates to a 40 x 20 character display using a standard character set. The display is backlit by cold cathode fluorescent tube (CCFT).An 8 ohm voice coil loudspeaker is mounted on the distribution board, producing all associated sounds.

2-8

pHOx Service Manual

3-1

3 Replacement Procedures3. Replace.

3.0 Replacement Procedures

CAUTION: Assemblies may have loose hardware mounting. En-sure no loose hardware is left in the unit before restoring power to the analyzer.

WARNING: Removal of the analyzer cover will allow access to line voltage wiring. Power should be turned off and the power cord disconnected before any assembly replacement procedure is performed.

3.1 Cover Removal

1. Turn the unit OFF, disconnect the line cord, and have the back of the unit facing you.

2. Remove the 4 rear-panel-corner-mounting screws. Lift the printer cover and remove the 2 cover mounting screws.

3. Slightly spread the bottom sides and slide the cover to-ward the back. This will disengage the internal clips at the front of the cover.

4. Lift the cover off the unit. Take care not to damage any cables or wiring during re-moval.

Figure 3-1 Cover Removal

DW

G #

10-1

029A

Screws (2)

MountScrews (4)

3-2

pHOx Service Manual

3.2 Tubing Harness

1. Remove the analyzer cover per Section 3.1.2. Remove the digital control shield. There are 4 mounting screws: One at the top front One at the bottom front Two at the bottom rear (above and below the PCMCIA card)

NOTE: Take care not to lose the spacers on the screws.

NOTE: It is optional to remove the digital control board and mount plate when accessing the reference and waste line. To remove the digital control board and mount plate, follow these steps:

a. Disconnect the electrical connectors (J7, J8, J9, and J10) from the top of the board.

b. Disconnect J11 connector at the bottom front of the board.

c. The digital control board and mount plate come off as one assembly. There are 3 mounting screws: two on the rear panel and one at the top front on the printer box. (Take care not to dam-age the board.)

d. Slightly lift the board and mount plate upward and out from the side of the analyzer.

e. Remove the ground strap on the back of the board.

Figure 3-2 Removing the Digital Control Board/Mount Plate

3. Remove the R and W-lines from their front panel fitting. (Looking at the front panel, R is on the left.) If you have not removed the digital control board, it may be easier to remove the panel fitting nut on the front panel and move the fitting to the new tubing.

Digital Control BoardMount Plate

MountScrews (4)

DW

G #

10-1

030A

Digital Control Board

(3)

3-3

3 Replacement Procedures3. Replace.

4. Carefully remove the tubing from the rotary valve including the center port to fluid fountain.

NOTE: Take care not to damage the valve ports. Push the tubing end off the port to reduce the risk of damag-ing the valve.

Figure 3-3 Removing Tubing from Rotary Valve

5. The tubing bracket is held in place on each end by a screw. Two screws hold the bracket in alignment; access them from the bottom of the analyzer. Remove both screws.

6. Lift the bracket and tubing set, removing it from the sampler side of the analyzer.7. Slide the new assembly into the

analyzer and reconnect the W and R-tubing to the front panel ports. Ensure that they go to the correct ports and brackets.

8. Reinstall the 2 mount screws at the bottom of the analyzer and the 2 end mount screws.

9. Place the reagent tubing and S-line onto their ports on the rotary valve. Take care that the metal ports are not damaged.

NOTE: The A line connects at the lower left (7 O’clock position).

Figure 3-4 Reconnect W and R-lines

10. If the control board and mount plate were removed, replace them.11. Reinstall the cover and restore the unit to operation.

Front Panel W and R Fittings

W

W R

R

DW

G #

10-1

039A

DW

G #

10-1

040A

3-4

pHOx Service Manual

3.3 Power Supply Removal

The Power Supply (PN 24720) is removed as follows:

WARNING: Unplug the Analyzer before proceeding.

1. Remove the analyzer cover per Section 3.1.2. Disconnect the power harness connectors: P1 and P3 (top of the power

supply assembly).3. Remove the 4 power supply mounting screws from the rear analyzer panel.

NOTE: Support the power supply before removing the mounting screw.

4. Carefully lower the supply. Rotate it to access the line input module wiring (TB1) on the bottom of the power supply.

Figure 3-5 Power Supply Removal

5. Disconnect TB1 and remove the power supply.6. Position the new power supply with the lower side toward the analyzer. Plug

in TB1.

Power Supply

DW

G #

10-1

038B

RearPanel

On/Off Switch

Power Plug

Use Center Holesto Mount

Power SupplyMount Screws(4)

P3P1

3-5

3 Replacement Procedures3. Replace.

NOTE: TB1 male end has 3 wires. The female end has 5 sockets. The Green/Yellow wire is closest to you.

7. Position the new power supply to align with the mounting holes in the back panel.

NOTE: The holes closest to the middle of the supply bracket are the ones used to mount the supply.

8. Connect P1 and P3.

NOTE: The wires leave the top of P1. The white wire of P3 goes to the digital display board P7.

9. Reinstall the unit cover and power cord.

Figure 3-6 Power Supply Position

3.4 Sampler Assembly

The Sampler Assembly (PN 23899) is removed as follows:1. Remove the probe and the air detector.

a. From the Menu screen, select Flowpath/Probe Maintenance and press Enter.

b. Press Move Probe (soft key). Open the door.c. Remove the capillary adapter from the front of the probe by gently pull-

ing it off.d. Disconnect the air detector's sample line from the sensor module.e. Disconnect the 2-prong cable.f. Push the air detector down and pull the air detector with the probe out

of the sampler assembly.

WARNING: Unplug the Analyzer before proceeding.

2. Remove the analyzer cover per Section 3.13. Cut the tie wraps on the rear of this assembly.

NOTE: Care should be taken to avoid cutting wires.

Brown Black Green &Yellow

Guide PowerSupply

DW

G #

10-1

037A

TB1

3-6

pHOx Service Manual

4. Disconnect the following connectors:J5 Motor cable connectorJ10 Home detectorP6 door switch cable- Sensor module light cable- Ground wire

5. Remove the S-Line from the fluid fountain assembly.

STOP: There are washers and nuts that are NOT captive. Do not lose any hardware into the unit.

Figure 3-7 Sampler Assembly

MountScrew

Ground

S-Line

MountScrews

Flow Cell Light Source

PowerSupply

J5, J10

CN3

DW

G #

10-1

032B

3-7

3 Replacement Procedures3. Replace.

6. Remove the 4 mounting screws.• One at the sensor mod-

ule light bracket to front panel

• One at the lower front of bracket

• Two at the rear of the sampler bracket, they also mount the rotary valve.

7. If NOT replacing the entire assembly, skip to Step 8, otherwise:a. Install the sensor mod-

ule light bracket onto the new sampler as-sembly.

b. Position the new as-sembly onto the unit. Install the mounting screws and tie wraps.

Figure 3-8 Removal of Sampler Assembly

8. Remove sampler subassemblies.a. Motor assembly1. From the Gear set side

of the assembly, remove the flywheel (solid metal disk) by loosening the set screws on its side.

2. On the motor side remove the 2 mount screws.

3. Remove the motor (some tie wraps may need cutting). Install the new motor by reversing these steps. Use new tie wraps as required.

b. Home detector (PN 21504) 1. Remove the sensor mod-

ule light and bracket.

NOTE: Touching the light glass will shorten use life.

Figure 3-9 Removal of Motor Assembly

Mount Screws

DW

G #

10-1

035A

DW

G #

10-1

034B

Mount Screws

Mount Screw

MountScrew

3-8

pHOx Service Manual

2. Remove the 2 mounting screws from the home detector.3. Install the new detector by reversing these steps. Use new tie wraps as

required to prevent fouling of the wires.4. After restoring power, ensure the capillary adapter seals against the

Fluid fountain in the Home position. If not, adjust the Home flag on the top gear.

NOTE: Ensure the home flag enters the detector without obstruc-tion.

C. Gear Set (PN 22634)1. The gears are held in position by compression springs. These can be

removed by spreading the open end and lifting.2. Remove the old gears

noting the position and orientation of each gear. The noise damper is behind the middle gear.

3. After installing the new gears, ensure that the gears run freely.

Figure 3-10 Removal of the Compression Spring

DW

G #

10-1

045A

Gear

Post

Spread CompressionSpring to loosen

Compression Spring

3-9

3 Replacement Procedures3. Replace.

D. Fluid Fountain Assembly (PN 24620)1. Rotate the gears so the capillary adapter lifts off the fluid fountain as-

sembly. 2 . Remove the S-Line

from the bottom of the spike assembly.

3. Loosen the mount-ing nut (under the sampler bracket).

4. Slide the fluid foun-tain assembly off the sampler.

5. Move the washer and mount nut to the new assembly.

6. Slide the new assem-bly onto the sampler.

7. Rotate the sampler gears until the capil-lary presses down on the fluid fountain assembly.

NOTE: The fluid fountain should be centered and should move downward between 0.06 to 0.08 inches (1.5 to 2.0 cm).

Figure 3-11 Fluid Fountain AssemblyDWG #10-1041A

CapillaryAdapter

FluidFountain

WasherNut

S-Line

Sampler

3-10

pHOx Service Manual

3.5 Rotary Valve Assembly

The Rotary Valve Assembly (PN 23898) is removed as follows.

1. Remove the cover per Sec-tion 3.1.

2. Optional: Perform this step to enhance accessibility. Remove the sampler as-sembly per Section 3.4.

3. Remove the 7 tubes and S-Line from the rotary valve.

CAUTION: Care should be taken not to scratch the metal tubes of the valve.

4. Disconnect the motor connector J2 and the optical detector connector J9.

5. Loosen the 2 mounting screws at the bottom of the assembly.

6. Slide the assembly toward the front and lift.

NOTE: The front of the valve bracket is hook shaped.

7. Slide the assembly toward the back of the unit to clear the sampler bracket.

Figure 3-12 Removing Rotary Valve Assembly

8. If replacing the entire assembly, install the new rotary valve assembly by following Steps 1 thru 7 in reverse order.

NOTE: Ensure the red, nonmetallic washers for the sampler mount screws stay in place during the remounting pro-cess. If only repairing the rotary valve assembly, refer to Section 3.6. If repairing the rotary valve motor, refer to Section 3.7.

DW

G #

10-1

040A

MountScrews

(2)

Flow Cell Light Source

DW

G #

10-1

036B

3-11

3 Replacement Procedures3. Replace.

3.6 Valve Assembly

The Valve Assembly (Valve Repair Kit PN 24626) is removed as follows:NOTE: The rotary valve assembly must be removed first before proceeding. (If not already removed, see Section 3.5.)

1. Remove “C” clamp on the valve shaft and remove the drive gear.

CAUTION: Do not lose the shaft pin when removing the drive gear.

2. With the drive gear removed, access and remove the 2 optical-detector mount screws.

3. Remove the optical detector.4. Remove the 4 screws from the compression spring washer.5. Remove the washer and compression spring.6. Slide the rotor and tube manifold out of the valve body.

NOTE: Before proceeding, note the alignment pin in the valve body that aligns with the slot in the manifold.

NOTE: Before installing the new parts, lubricate with a small drop of sealing oil on the mating surfaces.

7. Slide the new rotor into the valve body. The large flange side goes in first. Align with the drive pins.

DW

G #

10-1

019ADWG #10-1033A

A

B

CD

E

F

G

S J9

Figure 3-13 Valve Assembly Removal

3-12

pHOx Service Manual

8. Align the manifold with the pin and slide it into place.9. Reinstall the compression spring and washer

NOTE: Ensure the drive gear goes all the way over the shaft pin.

10. Install the “C” clamp on the drive gear shaft.11. Install the valve by reversing Steps 1 - 10 above.12. Install the rotary valve assembly. See Section 3.5 and reverse Steps 1 thru 7.

3.7 Rotary Valve Motor

WARNING: Unplug the power from the analyzer.

The Rotary Valve Motor (PN 24627) is removed as follows:

NOTE: The rotary valve assembly must be removed first before proceeding. (If not already removed, see Section 3.5.)

1. Remove the drive gear/opto detector as in Steps 1 and 2 in Section 3.6.2. Remove the 4 mounting screws from the motor.3. Remove the fly wheel from the motor shaft by loosening the Allen (hex)

screw.4. Mount the new motor and opto detector.5. Reinstall the rotary valve assembly. See Section 3.5 and reverse Steps 1 thru 7.

OpticalDetectoron Reverse Side

Access Holesin Gear

DW

G #

10-1

027A

Figure 3-14 Optical Detector

OpticalDetector

Mounting Screws

Flow Cell Light Source

DW

G #

10-1

026B

3-13

3 Replacement Procedures3. Replace.

3.8 Pump Assembly

The Pump Assembly (PN 23897) is removed as follows:1. Remove the analyzer cover per Section 3.1.2. Remove the digital control board shield. There are 4 mounting screws: One at the top front One at the bottom front Two at the bottom rear (above and below the PCMCIA card)

NOTE: Take care not to lose the spacers on the screws.

3. Remove the digital control board and mount plate.a. Disconnect the electric connectors (J7, J8, J9, and J10) from the top

of the board. b. Disconnect J11 connector at the bottom front of the board.c. The digital control board and mount plate come off as one assembly. There

are 4 mounting plate screws: two on the rear panel and two under the printer box.

NOTE: Take care not to damage the board.

d. Slightly lift the mount plate and the board upward and out the side of the analyzer.

Figure 3-15 Pump Assembly

4. Disconnect the motor connector J1.5. Remove the pump cap (on the front side of the pHOx) by pulling it off.6. Remove the 4 corner mounting screws then remove the pump.7. Install the new pump by reversing the above steps.

PumpAssembly

W

W R

R

DW

G #

10-1

048A

3-14

pHOx Service Manual

3.9 Digital Control Board

WARNING: Unplug the power from the analyzer.

The Digital Control Board is removed as follows:

NOTE: Units manufactured after June 1, 2005, use PN 40797; Units manufactured through June 1, 2005, use PN 24292.

1. Remove the analyzer cover per Section 3.1.

2. Remove the digital control board shield. There are 4 mounting screws:

-One at the top front, -One at the bottom front

-Two at the bottom rear (above and below the PCMCIA card)

NOTE: The shield is not pres-ent on units manu-factured after June 1, 2005.

Figure 3-16 Digital Board Removal

NOTE: Take care not to lose the spacers on the screws.

3. At the top of the board, disconnect the electri-cal connectors:

• J 7 - Power Supply • J 8 - Mechanical de-

vices harness • J 9 - Printer (Pull side

tabs outward to release the cable.)

• J10 - Analog Board 4. At the front lower corner

disconnect J11 – Display/keyboard (pull side tabs outward to release cable).

Figure 3-17 Install PCPMIC Card

Shield

DW

G #

10-1

047A

MountScrews

(4)

PCMCIARelease

Button

ShieldStandOffs

DW

G #

10-1

046B

Digital ControlBoard

Mount Screws

3-15

3 Replacement Procedures3. Replace.

5. Remove the PCMCIA card (PN 22805) Push the button below the PCMCIA card (bottom rear corner) to release the card.

6. Remove the 2 mounting screws and 2 shield standoffs: One screw is at the top rear corner

One screw is between J8 and J9 connectors One shield standoff is at the top front and the other is at the bottom front

7. Remove the 8 COM connectors screw standoffs at the rear of the pHOx by unscrewing 2 standoffs from each COM port.

8. Remove the board. 9. Position the new board on the mount plate.10. Install the 2 mounting screws and the 2 shield standoffs.11. Install the 8 COM connector screw standoffs.12. Reinstall the PCMCIA card into its slot.13. Reconnect all the cable connectors.14. Reinstall the digital board shield (4 screws).15. Reinstall the cover and restore the analyzer to operation.

3.10 Analog Board

The Analog Board is removed as follows:

NOTE: Basic/BiopHOx, pHOx/pHOx use PN 24291; pHOx Plus, Plus L,Plus C use PN 33390; all jumpers will be in the 3 to 4 position. To access the top/rear mounting screw, use a long shank (15 cm or 6 inch) Phillips® head (cross head) screwdriver.

1. Remove the analyzer cover per Section 3.1.

2. Disconnect the sampler ground strap.

3. Disconnect the cable connectors from the Analog board Sensor Interface board connectors:

J6 - 3 sensor board J7 - 2 sensor board J4 - Sensor module/SO2 cable J3 - RMS/Fluid Deck (Pull side

tabs outward to release cable.) J1 - Digital control board

Figure 3-18 Analog Board Removal

AnalogBoard

MountScrews

DW

G #

10-1

049B

J7

J6

J4J3

J1

3-16

pHOx Service Manual

4. Remove the 3 mounting screws: one in each corner of the board. You may need to place a screw driver through the holes in the power supply to access the screw in the upper back corner of the board.

5. Install the analog board by reversing the above steps.

3.11 Printer Assembly

WARNING: Unplug the power from the analyzer.

The Printer Assembly (PN 24615) is removed as follows:1. Remove the analyzer cover per Section 3.1.2. Remove the digital control board per Section 3.7.3. Open the front door; remove the 2 screws on top of the printer cover plate;

remove the printer cover plate by pushing the plate from the inside outward; then remove the paper and roller bar.

4. Remove the 3 print h e a d m o u n t i n g screws: two at the top of the print head and one at the lower right.

5. The flat ribbon cable from the print head sits in a connector. Slide the top of the connector upward to release the cable. The connector stays on the board. Tilt the print head from the assembly. Remove the paper advance motor cable that is connected into the printer control board.

Figure 3-19 Printer Cover Removal

6. The Printer Control Board (PN 24614) is mounted to the paper box by 2 screws that fit into the slots. Loosen the screws and slide out the board.

NOTE: Save cable that goes from the printer control board to J9 of the digital control board.

DW

G #

10-1

050A

Printer Cover

2 Screws

3-17

3 Replacement Procedures3. Replace.

7. When installing the new assembly, install the printer control board first. Do not tighten the 2 slotted screws. This allows access when connecting the print head cable. Attach the paper advance motor cable to the printer control board.

8. Install the print head. Connect the flat ribbon cable. The easiest way to plug in the print head cable is using one finger on each side. Tighten the 3 print head screws.

NOTE: The print head cable is clamped in place. Lift up on the white connector to release the flat cable for easy removal or replacement.

9. After connecting the print head, tighten the printer control board mounting screws inside the paper box.

10. Attach the cable from the printer control board (key faces up, blue or black wire faces towards the front of unit) to the digital control board.

11. Reinstall the digital control board, shield, printer cover, analyzer cover, and paper. Restore unit to opera-tion.

Figure 3-20 Attaching the Printer Cable

3.12 Sensor Module Light Source

The Sensor Module Light Source (PN 24717) is removed as follows:

1. Remove the analyzer cover per Section 3.1.

2. Remove the 2 screws of the light source mounting bracket.

3. The light bulb is removed by push-ing it into the socket, rotating it counterclockwise a 1/4 turn, then releasing it.

Figure 3-21 Removing Mounting Screws

CAUTION: Do not touch the glass surface of the new light.

4. Place the glass end of the new light bulb into the tubing provided.5. Insert the light bulb into the socket. Push it in and rotate it clockwise a 1/4

turn. Then pull off the tubing.6. Reinstall the bracket.7. Reinstall the covers and restore the unit to operation.

DW

G #

10-1

051A

Use 2 Fingers toConnect Printer Cableto Printer Board

DW

G #

10-1

052A

Flowcell Light Source Mount Screws

3-18

pHOx Service Manual

Figure 3-22 Tubing for Light

3.13 Door Removal /Reinstallation Procedure

1. Turn the unit OFF and disconnect the line cord.

STOP: Use a pencil to trace the position of the hinge onto the door mounting plate.

CAUTION: Until free of the hinge and cables, support the door to prevent stress to other cables/assemblies.

Figure 3-23 Door Removal

Push and rotate 1/4 turn counterclockwise to unscrew light.

Push plastic tubeover lamp . Do not touch glass of new light bulb.

Push and rotate 1/4 turn to clockwise to install new bulb into socket

DW

G #

10-1

053B

DW

G #

10-1

054B

3 MountingScrews

3-19

3 Replacement Procedures3. Replace.

2. Remove the 3 hinge screws from the door plate.3. Remove the 5 plate-mounting screws: one in each corner, plus one at right

side.

NOTE: Care should be taken not to strip the threads.

4. Remove the plate and disconnect the flat cable (J101). This frees the door. Place it face down on a protec-tive surface to prevent scratching.

Figure 3-24 Door Removal Continued

5. Reinstall the Door by reversing the above steps. Use the pencil marks made on the door plate to realign the door to the hinge. Ensure the door opens and closes easily and aligns with the chassis.

Analyzer PN of Door pHOx 21493 pHOx Basic/ BiopHOx 33102 pHOx Plus 30817 pHOx Plus L 34082 pHOx Plus C 35350

DW

G #

10-1

055B

Keypad Display

DistributionBoard

Door

Shield

5 PlateMountingScrews

3-20

pHOx Service Manual

3.14 Display Assembly

The Display Assembly is removed as follows: Prior to June 1, 2005 June 1, 2005 forwardDisplay Assy. 24716 40910Door Distribution Board 24727 40795Digital Control Board 24292 40797Door Interface Cable 21983 40495

1. Remove the analyzer door assembly per Section 3.13.2. Disconnect the cable from the display board (CN2).3. Disconnect the light source, J103, on the distribution board.4. Remove the 4 corner mounting screws.

CAUTION: When reinstalling the mount screws, DO NOT overtight-en. Assure the door plastic is not stripped. The screws will tighten in about one turn clockwise.

5. Position the new display and install the 4 mounting screws. Reconnect the cables CN2 and J103.

6. Reinstall the door per Section 3.13.

Figure 3-25 Display Removal

Keypad

KeypadGround

Strip

J102

Clip

CN2

J104

J101

Light SourceJ103

J102DW

G #

10-1

065A

3-21

3 Replacement Procedures3. Replace.

3.15 Distribution board

The Distribution Board (PN 24727) is removed as follows:1. Remove the analyzer door per section 3.13.2. Disconnect the cable connectors J104 and J103 coming from the display

board, J101 coming from the digital control board (should already be discon-nected), and J102 coming from the Keypad. (Refer to Figure 3-25.)

3. Remove the 4 corner mounting screws.4. Position the new distribution board (J101 will be on the hinge side of the

door) and remount.

CAUTION: When reinstalling the mount screws, DO NOT overtight-en. Assure the door plastic is not stripped. The screws will tighten in about one turn clockwise.

5. Reconnect cables J104, J103, J102, and J101.6. Reinstall the door per Section 3.13.

3.16 Keypad Assembly

The Keypad Assembly (PN 24625) is removed as follows:1. Remove the analyzer door per Section 3.13.2. Disconnect the cable connector J102 from the distribution board.3. Remove the clip holding the keypad

ground strip to the door. Save this for reuse.

4. The Keypad is held by adhesive to the front panel. Peel the old assembly from the front of the door.

5. Clean the surface of any debris.6. Remove the protective backing from

the new keypad.7. Slide the cable/ground strap through

the hole in the door. Align and press new keypad in place.

8. Fold the ground strap over the ledge and reinstall the clip.

9. Plug in J102 and reinstall the door per Section 3.13.

Figure 3-26 Keypad Removal

KeypadGround

Strip

Clip

DW

G #

10-1

056A

3-22

pHOx Service Manual

3.17 Sensor Module Assembly

WARNING: Unplug the power from the analyzer.

The Sensor Module Assembly is re-moved as follows:

1. Remove the sensors (electrodes) and sample probe/ADT1 assem-blies.

2. Remove the reference electrode assembly.

3. Rotate the retaining clamp on the top and the bottom of the assembly.

4. Pull the assembly toward you.5. Remount by reversing these

steps.

NOTE: The preheater and SO2 component are all part of this assembly.

Figure 3-27 Removing the Sensor Module

Sensor Module Part NumbersAnalyzer Part NumberpHOx 21494 pHOx Basic/BiopHOx 34653pHOx Plus 27807 pHOx Plus L 34738pHOx Plus C 37343

W

R

DW

G #

10-1

059A

3-23

3 Replacement Procedures3. Replace.

3.18 Electrode Interface Boards

This section pertains to the pHOx, pHOx Basic, and BiopHOx Analyzers only.

WARNING: Unplug the power from the analyzer.

The Electrode Interface Boards (PN 24719 {left} and PN 24718 {right}) are re-moved as follows:

1. Remove the analyzer's cover per Section 3.13.2. Remove the sampler assembly per Section 3.4.3. Remove the sensor module assembly per Section 3.17.4. Remove the spring loaded sensor module retaining clips (front of pHOx).5. Remove the 3 remaining mount screws.

NOTE: The top left corner screw was removed with sampler re-moval.

Figure 3-28 Removing the 3 Mount Screws

6. Pull the electrode interface board out thru the inside of the pHOx.

DW

G #

10-1

057A

3 Mount Screwsto Remove Electrode Interface BoardBracket

3-24

pHOx Service Manual

7. Each interface board is held to the plate by 2 screws. Ensure that the 3 connector board is on left, and the 2 connector board is on the right when looking at it from the front.

Figure 3-29 Correct position of Interface Board

8. Disconnect the affected board from the analog board: 3 connector board to J6, 2 connector board to J7.

9. Install new assembly by reversing the above steps.

3.19 Pinch Valve Assemblies

The Pinch Valve Assemblies (PN 16928) are interchangeable. The assembly plugs into the front panel and is held in by one screw.

Figure 3-30 Pinch Valve Assemblies

3 ConnectorBoard

2 ConnectorBoard

DW

G #

10-1

058A

DW

G #

10-1

059A

#10 Screw

#10 Screw

4 Troubleshooting4. Trouble.

4-1

4.0 Troubleshooting Information

This section contains suggestions for individual error codes, cycle information to help identify the cause of an error condition, and hints on basic troubleshooting.

1. Troubleshooting is recognizing the variance from normal operation. The calibration cycle can be used as the sequence of events where the cycle response and results are known.

Fluid Function Start cycle Std. A all 4 ADTs, Hct, PCO2, Na+, K+, iCa, Cl-, Gluc, Lact Std. B PCO2 Std. C pH, Na+, K+, Cl-, Gluc, Lact, (iCa on pHOx +L) Std. D all ADTs, pH, Hct, (iCa on pHOx+C)

PO2 uses Room Air (A-air position of the rotary valve). SO2/Hb utilizes external calibrators. (These are model specific.)

2. Sensor Methodology

ISE sensors - Na+, K+, iCa, Cl-, pH, and PCO2 ISE requiring a reference signal - Na+, K+, iCa, Cl-, pH ISE with its own internal reference - PCO2 Reflected Impedance - Air Detectors 1,2,4, 3/Hct Amperometric - PO2, Glucose, Lactate Optical Reflectance - SO2%

3. Calibrator Functions other than calibration.

PO2 Std. B is used to flush the system. PO2 is monitored on all fluids in the Calibrator pack. If the PO2 millivolt read-

ing varies from that of room air, a 72 Delta mV error is generated indicating a contaminated calibrator pack.

Std. D has surfactant and is used to clean the flowpath between pump cycles.

4-2

pHOx Service Manual

4. Dependencies (Refer to the quick reference guide in the appendix.)

Hct Calibrated Na+ required

NOTE: The pHOx Na+ is a 30 day expendable electrode.

SO2% Calibrated Hct and PO2 required Hb Calibrated Hct and SO2% required PCO2 Calibrated pH required

Calculated values will not be reported if a needed test is suppressed. P50 is reported only if PO2 is between 30 mmHg (4 kPa) and 75 mmHg (10 kPa).

5. Password protection - One password for Set Up, QC, and Service menus. Should the password be lost, enter the following number when asked for the password.

Date Code“Back-door” Password = (((month x 31) + day) x 137) + (year x 16)) + 5987 2 digits 2 digits last two digits

Example: If today is 02 January 2003 the “back-door” password would be:

Password = (((01 x 31) + 02) x 137) + (03 x 16)) + 5987 = (33 x 137) + 48 + 5987 = 4521 + 48 + 5987 = 10556

Change the password to a known value and write it down.

4 Troubleshooting4. Trouble.

4-3

Service Menu

System Test

Analog Input - Real Time mV

Sensor Subsystem

Printer Menu

Error Log

Comm unications Tests

Exit

Use to select. Press to change / enter.

Home

Manual Device Control

Manual Device Control + Flow test

Calibration data

Error log(Last 100 FIFO)

Active analog input signal (1 - 25)

Printer enable, Error log,System configuration print

I/O Signals / Loop Back Test

4.1 Service Screens

Pressing the “MENU” soft key from the “Ready Screen” displays an “Operation Menu.” Note that the soft key labels change. The “QC” label changes to “Service” (second key from the left). Pressing this key will display the Service Menu. If the home key is pressed, the system returns to the “Ready Screen.”

4-4

pHOx Service Manual

System Test Rev: 123.456 11-23-96 10:54 am

pH: +047.43

HCT: +007.43

PCO2: -082.12

PO2: -051.20 AD2: +314.3

Na: +147.4LED1: +314.3

LED2: +314.3 AD1: +314.3

Sampler:

Ref. valve:

Open

Rot. Valve:

Pump:

Waste valve:

Closed

Home STD A

CTRL 1

Exit

Use to select. Press to select / enter.

AD3: +314.3

AD4: +314.3

SO2% LEDs: On

Flow Test

Temp: 37.0

Air Osc: On

X level: Hct

Home

4.1.1 System Test Screen

Measured Parameters/ Current signal (millivolt)

Device Names (column 1&3) /Current state of the device (column 2&4)

Manual Control of a mechanical device:• Move the cursor over the device /Press “ENTER.”• The Option selections will appear. (If there are only 2 options, the device will toggle its state when the “ENTER”

key is pressed, i.e., solenoid valves.)• Move the cursor over the option wanted /Press “ENTER.”

NOTE: To successfully move fluids, a knowledge of the complete flow path is required.

4.1.1.1 Operator Flow Test

Pressing the soft key under the Operator Flow Test turns on the pump and raises the sample probe. Use the guide in this Section titled Flow Problems in Conjunc-tion with the Operator Flow Test.

ExitOperatorFlow Test

ServiceFlow Test

4 Troubleshooting4. Trouble.

4-5

4.1.1.2 Service Flow test

Pressing the soft key under the Service Flow Test brings up a screen that displays the results of this test. On this screen, pressing the soft key under Run Test causes the unit to pump each fluid twice: once with the Reference valve shut and once with the Reference valve Open.The expectation is that the flow time will increase with the decrease in vacuum caused by pumping reference solution into the waste line junction in the reference electrode. Ideally the 2 pump windings are equal and the flow time will double with the reference solenoid valve open. The pumping times will yield 2 pieces of information:

1. The individual flow times appear at the end of pumping that fluid.2. All the fluids will pump at the same rate. Usually Standard D is the one stan-

dard that varies in flow rate from the other fluids. A good rule of thumb for the flow rates is

pHOx Basic/BiopHOx and pHOx is in the 1000 – 1200 range.pHOx Plus/C is in the 1200 – 1300 range.pHOx Plus L pumps in the 1300 – 1400 range.

4-6

pHOx Service Manual

At the end of the Test, the Average flow time for Reference Valve Closed and Reference Valve Open is displayed. A ratio of the average times is presented. The ratio should fall between 0.5 – 0.65.

• Ratios approaching 1 indicate the loss of reference flow.• Ratios approaching 0.0 indicate an obstruction or air leak in the system

before the flowpath reaches the Reference electrode.

4.1.2 Analog Input Screen

Selecting this option, displays the 25 possible input channels. The display is active.Chnl Signal Unit Chnl Signal Unit1 Analog GND ALL 15 PO2 pHOx+, pHOx+ C2 K+ pHOx+,+L,+C 15 Lac pHOx+L3 Cl-/Ca++ pHOx+,+L 16 Glu pHOx+,+L,+C3 Cl- pHOx+C 17 PO2 pHOx+L4 Ca++ pHOx+C 18 PO2 pHOx+,+C5 pH pHOx 18 Lac pHOX+ L5 Na+ pHOx+,+L,+C 20 Temp 0.00=37C6 Na+ pHOx 22 LED1 ALL6 pH pHOx+,+L,+C 23 LED2 ALL7 PCO2 All 24 ISE Ref ALL8 Barr.Press. All 25 PCO2,Ref ALL9 +5V Ref All13 HCT pHOx, pHOx+C 14 PO2 pHOx, pHOx+C10 Voltage Monitor Sums +24 and +5

CAL D 1170 RATIO 0.58

CAL A 1190 AVG FLOW 1 1180

CAL C 1190

QC 1 1195

QC 2 1193

QC 3 1195

FLOW TEST RESULTS

CAL B 1195 AVG FLOW 2 2010

EXIT Run Test Print

Average Flow time Reference Valve Shut

Flow 1/Flow 2 (nominal 0.5 – 0.65)

Average Flow time Reference Valve Open

4 Troubleshooting4. Trouble.

4-7

4.1.3 Sensor Subsystem

This set of screens shows the calibration data for each analyte measured by the pHOx. Use the “NEXT” soft key to page to the analyte of interest.

Last 2-point calibration data Millivolts for each standard Slope calculated

Last 1 point data (Millivolts should be similar to the same Standard in 2 point cal)

Sample millivolts and concentration - Unknown and therefore not relevant to troubleshooting Useful to the clinician for any possible dilution

SO2 sensors - Each LED has its own screen. Displays similar data to the other Measured parameters, plus

The “dark” (off) and “Light” (full) Signals are shown

Air Detector Subsystems

Displays the millivolts for Air and Liquid. (Air column is actually Standard D.)

Shows the AIR/Liquid threshold (set by the analyzer) millivolt level

NOTE: Numbers shown on this screen are not real values.

SLOPE

SAMPLE

CONC

Exit

LIGHT

STD1

STD2

DARK

Callibration & Analysis Data

SO2 LED 1

PrintHomeNextPage

+11.4

mV mV

+ 379.1

+ 379.1

+ 379.1

+ 379.1

+ 379.1

Air Detectors

THRESHOLDAIR FLUID

AD3

Callibration Data

AD1

Exit

AD2

AD4

NextPage

PrintHome

+ 117.0

+ 117.0

+ 117.0

+ 117.0 + 54.0

+ 54.0

+ 54.0

+ 54.0

+ 99.0

+ 99.0

+ 99.0

+ 99.0

pH Sensor

Callibration & Analysis Data

Exit

SLOPE

STDA

STDB SAMPLE

STDA

CONCSTDC

mV mV

+ 43.1+ 43.1

+ 43.1

+ 43.1

+ 43.1

+ 43.1

97.4

- 7.1

+ 43.1

Home

STDD

PrintNextPage

4-8

pHOx Service Manual

4.1.4 Printer Menu

This menu is straight forward. Move the cursor to the desired action and press “ENTER”

Printer Menu

Printer: Enabled

Print System Error Log

Print System Dump

Character Set Test

Use to select. Press to change / enter.

Exit CancelPrintout

Home

Enable/Disable internal printer

Prints last errors

Prints setup configuration and millivolt data

Test Print

4 Troubleshooting4. Trouble.

4-9

4.1.5 Communication Tests

This screen allows you to transmit out a particular port for testing. Likewise using a test plug you can perform a Loopback test for each I/O function.

Pressing the “HOME” key returns to the READY screen, or EXIT to go back one screen.

Communications Tests

Loop Back Tests

Send test characters.

Exit

Use to select. Press to change / enter.

Home

Loopback Tests

COOX

ASTM

Remote

Bar Wand

Exit

1. Select Port and install loopback plug.

2. Press to begin testing.

Tests Completed: 0

RXD DTR RTS

TXD DSR DCD CTS RI

Errors 0 0 0 0 0

Use to select. Press to change / enter.

Home

Loopback Tests

COOX

ASTM

Remote

Bar Wand

Exit

1. Select Port and install loopback plug.

2. Press to begin testing.

Tests Completed: 0

RXD DTR RTS

TXD DSR DCD CTS RI

Errors 0 0 0 0 0

Use to select. Press to change / enter.

Home

4-10

pHOx Service Manual

4.2 Flow Problems

NOTE: Each of the following troubleshooting procedures has an associated flow diagram that may be found at the end of this chapter.

General procedures when isolating flow problems:

NOTE: Refer to the FLOW TEST decision tree. Do not be con-fused with the Flow Test Cycle that only shows the flow times.

Unit calibrates, but the sample will not aspirate. This usually means you have an air leak into the system. The sample, being more viscous than the aqueous standards, causes more vacuum to be created by the roller pump. If a membrane is leaking, this increase in vacuum will pull the aqueous fill solution from the electrode, easier than pulling sample from the probe. Likewise, a poor connection at the intercon-nect tubing or between the electrode sealing washer and the sensor module will allow air to enter easier than pulling sample up the flow path.

• Check the connections between probe and sensor module, sensor module, and reference electrode.

• Check the electrodes for the loss of internal fill solution.• Check the flow cell window during a pump cycle, looking for bubbles entering

the flow path or pump and waste tubing.

A No flow or obstructed flow cause can be isolated by performing the FLOW TEST. The fluidic system can be isolated into three large sections. This will limit the area of your search. The 3 sections are Fluid Pack to Fluid Fountain, Sampler/Sensor module, and W/R to Waste tubing. Use the SYSTEM TEST screens to control the mechanical devices.

1. Extend the sampler to the syringe position.2. Open the waste line solenoid.3. Turn on the pump and attempt to aspirate water then air from a sample

cup.4. If water is aspirated from the cup, the sample probe, sensor module, external

tubing, and waste line are functioning. The problem exists between the Fluid Fountain and Fluid Pack.

5. If water is NOT aspirated from the cup, remove the tubing from the outlet port of the Reference electrode. Place it in a cup of water.

NOTE: This assumes the waste solenoid has functioned and is not pinching the tubing.

4 Troubleshooting4. Trouble.

4-11

6. If step 5 fails to aspirate from the cup, the problem lies between the W/R tubing and the waste bottle of the Fluid Pack.

7. If Step 5 does aspirate water, the problem is in the Sampler/ Sensor mod-ule.

4.2.1 Isolating a Plug in the W/R to Waste Tubing

NOTE: If the reference flow is obstructed, the sample may not stop in the sensor module due to increased vacuum. Flow time decreases due to increased flow rate.

1. Disconnect the waste tubing from the reference electrode and place in a cup of water. If water is NOT aspirated, there is a plug.

2. Lift the waste tubing out of the waste solenoid. If flow starts, the valve is dirty. Clean the dirt from the waste solenoid or replace it.

3. If flow does NOT start, remove the waste line from the front panel connec-tion. If water flows from this tube there is a plug in the internal waste tubing or waste bag of the fluid pack.

4. Turn off the pump. Connect a syringe of water to the waste port on the front panel. Attempt to inject water through the line to clear the plug.

5. If there is no flow, remove the Fluid Pack to ensure the problem is not a poorly seated or damaged waste connection. Should the problem continue, replace the internal tubing set.

4.2.2 Isolating a flow problem in the Sampler/Sensor module

1. You noted flow when aspirating from the waste tubing in step 2 of Section 4.2.3.

2. Raise the Sampler to the Syringe position. The Pump should be OFF for this test.3. Connect a syringe of water to the reference electrode outlet port. Attempt to

inject the water.4. If water flows freely, air is entering the system preventing aspiration.

a. Inspect each electrode to assure the internal fill solution level is not dropping.

b. If yes, replace the electrode or membrane as required.c. Inspect the seal between the reference electrode and sensor module

(interconnect tubing) .

NOTE: If the flow path starts to empty while the syringe is still attached, only the water below the leak will leave the flow path out of the tip of the probe.

4-12

pHOx Service Manual

d. Inspect the sensor sealing washers for distortion or misalignment. Re-place as necessary.

e. The seal between the air detector and the sample probe may be tested by attaching the syringe to the sample probe, blocking the air detector outlet tube. Apply pressure to the syringe and look for water leaking out.

5. If water does not flow freely, there is a blockage.a. Loosen the reference electrode. Water should pass through or the

electrode is plugged. Flushing with cleaning agent or hot water may free the blockage.

b. If the reference is clear, reinstall it and remove the bottom electrode. c. Apply a moderate pressure with the syringe.d. If water does not enter the sensor module of the bottom electrode the

blockage is above it.e. Removing one electrode at a time. Look for water entering the sensor

module. If no water enters the module as the electrode is removed, the plug is above it.

f. If water did enter the bottom sensor module in step C above, the plug is in the Sample Probe/Air detector. Remove these assemblies and flush through them or replace them.

4.2.3 Isolating a flow problem between the Fluid Pack and the Fluid Fountain

If you have successfully aspirated a sample, but the analyzer will not calibrate or aspirate on-board controls, proceed as follows.

1. Position the sampler in the syringe position and inspect the capillary adapter for signs of wear. Replace if necessary.

2. Lower the sampler to the Home position; connect a syringe to the reference outlet port.

3. Remove the analyzer cover.4. Disconnect the S-Line (center tube) from the rotary valve.5. Inject the water and look for flow from the S-Line.6. Look for a leak between the fluid fountain and the capillary adapter. Realign

if necessary.7. If there is no flow from the S-Line, remove the fluid fountain/S-Line and

flush with hot water or cleaning agent, or replace the assembly.8. If there is free flow through the S-Line, the problem is in the rotary valve,

tubing harness, or connection to the fluid pack.9. Reconnect the S-Line and remove the fluid pack.10. Move the rotary valve to all the possible positions, observing which ports do

not allow flow.11. If there is no flow in ALL positions, remove the valve and clean or replace it.

4 Troubleshooting4. Trouble.

4-13

12. Remove the tubing for each position that does not flow freely.13. If water can be flushed through the individual lines, remove and clean or

replace the valve.14. If water cannot be flushed through the affected lines, replace the tubing har-

ness.

4.3 Status Codes

Error codes come in 2-digit groups. The tens (first) number, called the “series code,” identifies the affected component. Series 1 through 6 are sensors,0 = pH 1 = PCO2 2 = PO2 3 = SO2 % 4 = Hb 5 = Hct 6 = Na+

Series 7 = flow, 8 = Q.C. 9 = Electronic/Mechanical/Hardware 10 = CO-Oximeter codes.The ones (second) number, called the “digit code,” provides more information on the failure.

Series - 1 - 6 7 8 9 10

Digit code -1 Slope No Std.A No QC1 Barometer Not assigned -2 Instability No Std.B No QC2 Printer -3 Overload No Std.C No QC3 Flow Cell Temp -4 Drift No Std.D Communication -5 Dependency Digital Board -6 Sample flow Analog Board (ADX) -7 Air flow Software

pH Status CodesCode Status Screen Text Error Log Text01 pH Slope pH Slope02 pH Instability pH Instability03 pH Overload pH Overload04 pH Drift pH Drift96 Hardware pH ADC97 Software pH Math

4-14

pHOx Service Manual

PCO2 Status CodesCode Status Screen Text Error Log Text11 PCO2 Slope PCO2 Slope12 PCO2 Instability PCO2 Instability13 PCO2 Overload PCO2 Overload14 PCO2 Drift PCO2 Drift15 PCO2 Dependency PCO2 Dependency96 Hardware PCO2 ADC97 Software PCO2 Math

PO2 Status CodesCode Status Screen Text Error Log Text21 PO2 Slope PO2 Slope22 PO2 Instability PO2 Instability23 PO2 Overload PO2 Overload 24 PO2 Drift PO2 Drift96 Hardware PO2 ADC97 Software PO2 Math

SO2 LED 1 Status CodesCode Status Screen Text Error Log Text31 SO2 % Slope SO2 % LED132 SO2 % Instability SO2 % LED1 33 SO2 % Overload SO2 % LED134 SO2 % Drift SO2 % LED135 SO2 % Dependency SO2 % Dependency 96 Hardware SO2 % LED1 ADC 97 Software SO2 % LED1 Math

SO2 LED 2 Status CodesCode Status Screen Text Error Log Text31 SO2 % Slope SO2 % LED232 SO2 % Instability SO2 % LED2 33 SO2 % Overload SO2 % LED234 SO2 % Drift SO2 % LED296 Hardware SO2 % LED2 ADC 97 Software SO2 % LED2 Math

Hb Status CodesCode Status Screen Text Error Log Text41 Hb Slope Hb Slope45 Hb Dependency Hb Dependency 97 Software Hb Math

4 Troubleshooting4. Trouble.

4-15

Hct Status CodesCode Status Screen Text Error Log Text51 Hct/Air Det Slope Hct52 Hct/Air Det Instability Hct53 Hct/Air Det Overload Hct54 Hct/Air Det Drift Hct55 Hct/Air Det Dependency Hct96 Hardware Hct ADC97 Software Hct Math

Na+ Status CodesCode Status Screen Text Error Log Text61 Na Slope Na Slope62 Na Instability Na Instability63 Na Overload Na Overload64 Na Drift Na Drift96 Hardware Na ADC97 Software Na Math

AD1 Status CodesCode Status Screen Text Error Log Text51 Hct/Air Det Slope AD152 Hct/Air Det Instability AD153 Hct/Air Det Overload AD196 Hardware AD1 ADC97 Software AD1 Math

AD2 Status CodesCode Status Screen Text Error Log Text51 Hct/Air Det Slope AD252 Hct/Air Det Instability AD253 Hct/Air Det Overload AD296 Hardware AD2 ADC97 Software AD2 Math

AD3 Status CodesCode Status Screen Text Error Log Text51 Hct/Air Det Slope AD352 Hct/Air Det Instability AD353 Hct/Air Det Overload AD396 Hardware AD3 ADC97 Software AD3 Math

4-16

pHOx Service Manual

AD4 Status CodesCode Status Screen Text Error Log Text51 Hct/Air Det Slope AD452 Hct/Air Det Instability AD453 Hct/Air Det Overload AD496 Hardware AD4 ADC97 Software AD4 Math

Flow Status CodesCode Status Screen Text Error Log Text71 StdA Flow StdA Flow72 StdB Flow StdB Flow73 StdC Flow StdC Flow74 StdD Flow StdD Flow75 Flowtime Flowtime76 Sample Flow Sample Flow77 Air Flow Air Flow78 Back Flow Back Flow

Control Status CodesCode Status Screen Text Error Log Text81 Ctrl 1 Flow Ctrl 1 Flow82 Ctrl 2 Flow Ctrl 2 Flow83 Ctrl 3 Flow Ctrl 3 Flow

System Status CodesCode Status Screen Text Error Log Text91 Barometer Barometer92 Printer Printer93 Temperature Temperature94 Communication Communication95 Hardware PEROM Write95 Hardware PEROM Read95 Hardware PEROM CRC96 Hardware Hardware97 Software Software99 Scheduled QC Not Run Scheduled QC Not Run

RMS Status CodesCode Status Screen Text Error Log Text688 N/A Software (Pack Not Present)690 N/A Software (Pack Empty)691 N/A Software (Pack Lot Date Expired)693 N/A Software (Pack Use Life Expired)694 N/A Communication Error

4 Troubleshooting4. Trouble.

4-17

CMS Status CodesCode Status Screen Text Error Log Text696 N/A Software (Pack Not Present)698 N/A Software (Pack Empty)699 N/A Software (Pack Lot Date Expired)701 N/A Software (Pack Use Life Expired)702 N/A Communication Error

Standards Information:

Sensor Standards Used Delta mV Calibration To calibrate (between standards)_______________________________________________________pH C & D 33Na+ (pHOx) A & D 17(pHOx+/C/L) A & C 16PO2 Electronic 0.00 / Room Air, Std A used for 1 pt Ca++

PCO2 A & B 17Insert sheet shows HCO3- value. HCO3 x 1.5=CO2 in mmHg

Hct& ADET’s A & D (all ADET’s) > 500 Hct ideal > 800(Hb) Based on SO2 & Hct (mV not shown)SO2 External Stds 1 & 2 n/aK+ A & C 24iCa A & D 12Cl- A & C 15Gluc A & C 38-55Lact A & C 35-50_______________________________________________________

4-18

pHOx Service Manual

4.3.1 Analyte Troubleshooting ( Series code 1-6 and Digit codes 1-4)

Multiple Channel failures:Slope

• Ensure the Reagent Pack is fully plugged in. • The pack box edge closest to you should

lie inside the small lip of the chassis. • Check the flow times displayed on the

service screens.• Inspect the Reference tubing to assure

flow.• Inspect the waste line

for bubbles.• Prime the system,

check each fluid for air.• pH and Na+ require

a working reference electrode.

Figure 4-1 Reagent Pack Placement

• Note if the channels that fail use the same standard from the fluid pack.• Loss of the +5 Vdc on the analog board• Loss of the Analog board A/D/ converter• Loss of +5 Vdc to the Analog Board• Analog Board over heated (PTC resister open) - If unit is allowed to cool off

by shutting it off for 5 minutes. Investigate why the unit or Analog board is overheating.

Dependency Rules

• Hct requires calibrated Na+ for calibration and results.• PCO2 requires calibrated pH for results (pH checked on standards to assure

bicarbonate is correct).• SO2 requires calibrated Hct and PO2 for results. Calculated SO2% is re-

ported if measured SO2 is not available.• Hb requires Hct and SO2 for reporting. Priority:

1. If linked with COOX, the COOX HGB is reported.2. If no COOX link, the measured result is reported, if available.3. If no COOX or measured result, the calculated result from the Hct reading

(Hct/3) is reported.4. If no COOX, no measured, and no Hct, the default value (Hb) is re-

ported.

4 Troubleshooting4. Trouble.

4-19

• P50 is measured and is reported only when PO2 is between 30-75 mmHg. P50c is reported when PO2 is outside those limits. P50 also requires cali-brated SO2% and Hb/Hct for reporting.

• Qsp/Qt requires 2 separate blood samples for determination: mixed venous and arterial.

• RI utilizes input or default FIO2% for calculation.• Calculated Results require calibrated and proper results by measured tests

used in the calculation.• Suppression rules - If measured test is suppressed in the Setup, the calcu-

lated results based on test also will not be shown or printed.• Password (One) is used for Setup, Service Menu, and QC. (No user pass-

words are available.)• Interfaces are NOVA COOX, PDM, and ASTM.• Standby Mode: There is no time dependency. Analyzer will prime and calibrate after coming

out of standby. Idle pumping is active, but no fluidic or gas calibrations are performed. Features Auto-Wake-Up, Auto-Cal, and Auto-QC• Micro Sample Analysis: Micro-Mode Tests: pH, PCO2, and PO2 Stepped Analysis: Stage 1 = Tip of probe to ADT1 (40 µL) Stage 2 = Up to ADT3 Stage 3 = Between ADT3 (up to ADT4): Other Tests Reading (including contact with reference electrode)

Instability

• Inspect sensor module for bubbles during the measurement cycle• Ensure that the waste valve stops the fluid flow during analysis. Fill the flow

path with any standard, raise the probe to the syringe position Look for droplets forming at the sampler tip.• Ensure that the waste and Reference solenoid valve are functioning.• Fill the flow path with any standard and view the system test screen. The signals should be steady and there is no flow of fluid there are three Possible conditions, a loose sensor module, an analog board failure, or a

poor connection between a set of electrodes and an interconnect board.• Inspect the electrode washers to assure proper alignment. Nonalignment

may cause a flow problem.• In rare cases the tubing harness may have become contaminated. Using the clean-

ing adapter, flush the system with cleaning agent followed by distilled water.

4-20

pHOx Service Manual

4.3.2 Single Channel Failures

SlopeView the Sensor Subsystem screens. If the change in millivolts between standards is 0 or less than that mentioned in Section 4.3.1, there are several possibilities:

• Membrane or sensor has reached the end of its use life.• Bubble of Air is covering the sensing material at the electrode tip.• The electrode is not making good contact with the Interconnect board con-

tacts.• Loss of the Analog board multiplexer will cause a zero input to the A/D con-

verter.• Ensure calibration on required channels: Sodium is required for Hct, Hct is

required for SO2%.• Viewing the Sensor subsystem screens you note an overload condition (999.9 mv).• Check for NO connection to the Interconnect board.• Check the PO2 /PCO2 for loss of the Internal Fill solution (membrane rup-

ture).• Loss of the multiplexer or individual channel circuit on the Analog board.• The pH electrode may need conditioning. • Na+ sensor over 30 days of use. The Na+ sensor will be destroyed by con-

ditioning.

InstabilitySingle channel instability is seldom a flow problem unless a standard is unique to a particular channel.

• De-bubble the electrodes.• Ensure a good connection between the electrode and the Interconnect

Board.• Ensure a good connection between the sensor module and Chassis.• Ensure the analyzer's door is closed and the ground strap is making good

connection.

4 Troubleshooting4. Trouble.

4-21

4.3.3 Temperature Problems

• Ensure that the room ambient temperature is within specification ( 18o - 32o C).• Ensure that the sensor module is plugged in correctly.• View the Analog Input screen channel 20. This channel will be a 0 if the

module is at 37o C and change at a rate of 22.5 mv/1oC, + if above and - if below 37o C.

• The power for the preheater comes from the digital control board.• The heater control circuitry is on the analog board.

4.3.4 Air Detector (ADT) Problems

• Air Detectors are only calibrated during a 2-point calibration.• Typically the Fluid /Air millivolt difference will be about 700.• Check the sample probe for air being introduced below the sensor mod-

ule.• Check the sensor module for air being introduced at the electrode/sensor

module seal.• Ensure that the electrode sealing washers are centered and seated correctly.• The Air detector signals are processed by the analog board. They are only

displayed on the System Test screen. Alternately aspirate air and then fluid to see the change in the signal.

NOTE: Remember to turn on the air oscillator to obtain an ADT reading.

• Flow Test- During a flow test, the rotary valve rotates to each fluid position.- The pump turns on to fill the lines from the rotary valve S-line port to

the ADT1.- The rotary valve rotates to an air position.- The pump turns on and the time in milliseconds until the ADT senses

air is recorded.- The time is normally 1200 msec. The acceptable range is 100 to 1700

milliseconds.• Run a Flow Test; press the soft key on the Service screen.

- Typical result for Standard A, B, C, D is 1300.- If all are high (3000-5000), there is a flow path obstruction.- If only one is high,

1. Check for correct fluid pack installation.2. Check that the valve is in the correct position.3. Connect a syringe to the S-line port of the valve. Remove the

reagent pack and move the valve to each position. Verify that the link is clear by purging air through the S-port.

4-22

pHOx Service Manual

4 Troubleshooting4. Trouble.

4-23

4-24

pHOx Service Manual

Replace or backflush the air detector.

Replace or use plas-tic cleaning line to clear the plug.

Clear the plug by flush-ing plastic line or re-place the module.

Inspect for air leak. Loss of IFS, bad washer, ruptured membrane, or interconnect tubing seal.

4 Troubleshooting4. Trouble.

4-25

4-26

pHOx Service Manual

������������������

���� ��������������������������������������������� ���������������������������������������������������������������� �������������������������������������������������������������������������������������������� ��� ����������������������������������������������� ���� ��������������������

��� ������������������������������������������������������������ ��������������������������������������������������������������������������������������������������� ��� ���������������������������������������������������� � ���������������������������������������������������������� � �������������������� ���� ���������������������

��� �������������������������������������������������������������� ��������������������������������������������������������������� ����������������������������������������������������������������������������������������������������������������������������� ��� ���������������������� ���� ��������������������������������������������������

��� ������������������������������������������������� ��������������������������������������������������� ������������������������������� �������������������������������������������������������� �������������������������������������������������� ��������������������������������������������������������� � �������������������������������������������� ��������������������������������������������������� ������������������������������������������������������������� �������������������������������������������������������������������������������������� ������������������������������������������������������������������ �������������������������������������������������������������� ���������������������������������������������������������

�������������������������������������������������

4 Troubleshooting4. Trouble.

4-27

����������������������������������

��

��

��

��

��� ������������������������������������������������� �����������������������������������������������

��� �������������������������������������������� ����������������������������������������������������������������� ���������������������������������

��� ���������������������������������������������������������� ������������������������

���� ������������������������������������������������������������������������������������������������������������������������������������������

���� �����������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������

���� ���������������������������������������� �������������������������������������������������������� ������������������������������������������������������������������������������������������������������������������������������������������������������

��� ���������������������������������������������������������� ��������

���� ��������������������������������������������������������� ������������������������������������������� ��� ������������������������ ����������������������������

���� ����������������������������������������������� ������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������

����������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������

�������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������

��������������������������������������������������������������������������������������������

��

��

��

��

��

��

��

4-28

pHOx Service Manual

A-1

AppendixAppendix

A Appendix

A.1 Quick Reference Guide

Measured Tests: pH, PO2, PCO2, SO2%, Hb/Hct, (Na+ is not reported)

Calculated Tests:• A Alveolar Oxygen Concentration• a/A Arterial Alveolar Oxygen Tension Ratio • A-aDO2 Arterial Alveolar Oxygen• BEecf Base Excess, Extracellular Fluid• BEblood Base Excess, Blood• HCO3- Actual Bicarbonate Concentration• O2Ctc Oxygen Content• P50c P50 Calculated (when PO2 is <30 or >75)• Qsp/Qt Physiologic Shunt• RI Respiratory Index• SBC Standard Bicarbonate• SO2%c Oxygen Saturation % (Calculated)• TCO2 Total Carbon Dioxide• Temp pH/PCO2/PO2 (Temperature Patient Corrected, if selected)

Slope Limits: pH 9.1 to 11.6 (1,7) PCO2 7.9 to 12.6 (1,2,4,8) PO2 -15.0 to -1.6 (1,2,4,8) SO2% 6.9 to 18.2 (1,5,8) Na+ 8.8 to 11.5 (1,7,8,9) Hct: 12.0 to 50. (1,6,7) Electrode Notes:1 = Long Life Sensor 6 = Impedance 2 = Int’l Ref Electrode 7 = No Washer 3 = Membrane Cap; Push-On 8 = Internal Washer 4 = Membrane Cap; Screw-On 9 = Non-Reported Test5 = Reflectance 10= Obtained from SO2% and Hct

A-2

pHOx Service Manual

Flow Limits: Flow Test (Service Menu): Typical/Ideal Value = 1300 (milliseconds) ** High Values (ie: > 3000) or “ERROR” note on Std Channel: Indicates major flow blockage.

Offset Limits (Slope/Intercept): Test Y-Intercept Range Slope Range pH +/- 100 mpH NONE H+ NONE 0.8 to 1.2 PCO2 +/- 10 mm Hg 0.8 to 1.2 PO2 +/- 20 mm Hg 0.9 to 1.1 SO2 % NONE 0.9 to 1.1 Hct +/- 10.0 % 0.8 to 1.2 Hb +/- 2.0 g/dl 0.9 to 1.1 Flow Cell Positions:

Top Reference Biosensor Adet 1 Located above Sample Probe Used to detect 40 uL MicroSampleAdet 2 In Sensor Mod/Flow Cell Used to detect 70 uL sampleAdet 3 Reads Hct, then blood pulled up and air is seen telling stop for electrode sample reading; sample positionAdet 4 Positions top of sample for electrode readings

Calibration Details (Analysis Time/Throughput):Two Point Calibration: User Selects (not auto control), 2,4,6 hour intervals (turned off in Standby Mode).One Point Cal: User selects for 30 or 45 minutes (same as STP/UL), One point cal for ALL (gases plus pH/Hct), One point SO2 done with each full 2-point calibration.Standard One Point Cal: Done with blood samples: Single Mode (each sample); NOT available with product release; will be for low volume users only.

Reference

Na+

pCO2

sO2

pH

pO2

ADET 4

ADET 3

ADET 2

A-3

AppendixAppendix

Throughput Details: Throughput specs subject to change without notice due to software sequence changes. pHOx/B: Analysis Cycle: Mode A = 44 sec.; Mode B = 64 sec (time to view results) Run Time: Mode A = 62 sec; Mode B = 74 sec (full cycle) Throughput: Approx 50/Hour (varies by mode) Micro Mode: Analysis (view) = 43 sec; Run Time = 65 sec.

pHOx+C: Analysis Cycle: Mode A = 47 sec.; Mode B = 100 sec (time to view results) Run Time: Mode A = 65 sec; Mode B= 109 sec (full cycle) Throughput: Approx 33/Hour (varies by mode) Micro Mode: Analysis (view) = 45 sec; Run Time = 65 sec.

pHOx+-L: Analysis Cycle: Mode A = 60 sec.; Mode B = 126 sec (time to view results) Run Time: Mode A = 89 sec; Mode B= 134 sec (full cycle) Throughput: Goal: 25-30/Hour Micro Mode: Analysis (view) = 71 sec; Run Time = 95 sec.

Default Values:Hb 14.5 g/dL (Range 10.0 to 20.0 g/dL)Temperature 37 ° C (Range 10.0 to 40.0°C)FIO2 % 20.9 % (room air) (Range 20 to 100.0%)

Sample Volume: Regular Mode 70 uL pHOx/B (all tests) 115 uL pHOx-Plus (all tests) 120 uL pHOx-Plus L (all tests) Micro Mode 40 uL pHOx/B (gases) 55 uL pHOx-Plus- Plus L (gases)

Software Version: 4.xx

Sample Type: Arterial, venous, capillary, mixed venous, Quality Controls, and Proficiency

A-4

pHOx Service Manual

One-Point Cal Modes:Mode A One point fluid cal performed with each Gas Cal I (30-45 min intervals). (Best for medium-high volume customers)Mode B One point fluid and Gas Cal performed with EACH analysis. (Best for LOW volume customers only - No separate Gas Cal I)

Rotary Valve Positions:

Figure A-2 Valve Manifold Assembly (Tubing Connections and Rotation Positions)

Control/Position Valve Manually and View Positions: Service Menu, System Test

AIR Positions: Air Std A, Air Std C, Air Cntrl 1, Air Cntrl 3Closed Positions: HOME, Air Std B, Air Std D, Air Cntrl 2, Dummy Fluid Positions: Std A Std B Std C Std DConnections on Valve: (A) (B) (C) (D) Control (Fluid) Positions: Cntrl 1 Cntrl 2 Cntrl 3Connections on Valve: (E) (F) (G) Not Used Position: Dummy

Home: ClosedAIR ("Air Std A")

Std A

Closed("Air Std B")

Std B

AIR

Std C

Closed("Air Std D")

Std D

AIR("Air Cntrl 1")

Closed("Air Cntrl 2")

Cntrl 1

Cntrl 2

AIR("Air Cntrl 3")

Cntrl 3

"Dummy"/Closed(not used)

("Air Std C")

Sample

A-5

AppendixAppendix

Linear (Measurement) Ranges:pH 6.50 to 8.00 pH unitsPCO2 3.0 to 200 mmHg (0.4 to 26.7 kPa)PO2 0 to 800 mmHg (0.0 to 106.7 kPa)SO2% 0.0 to 100%Hct 12% to 70%Hb 4 - 23.3 g/dL - based on HctNa+ not reported - 1.0 to 200.0 mmol/L)

Memory Storage Capability:Patient Storage 18 Samples (FIFO)QC Storage 6 QC Files (FIFO) Stores up to 10 daily points/level - condensed to 1 monthly point Monthly Data - 32 points for each levelError Codes 96 total codes stored (FIFO) 6 shown per screen (16 potential screens)

QC Files Details:6 QC Files plus “File 0” (proficiency)

3 On-Board QC files (combined gases/Hct/SO2 controls; 3 levels) 3 External QC files (3 levels for BG; 2 levels for Hct/SO2)Features Auto QC (on-board QC) w/ preset times

Control reminder/lockout; Setting daily QC analysis times.

Tracking Daily, Monthly, Cumulative, Levey-Jennings.

Fluids Details:Std A - It is used to calibrate PCO2, Na+, Hct, and PO2 (1-point check) and sits

in flow cell during idle.Std B - It is used to calibrate PCO2, contains surfactant, and is used to clean flow

path during cycles.Std C - It is used to calibrate pH.Std D - It is used to calibrate pH, Na+, and Hct.External SO2 Calibrators - There are 2 levels used to calibrate SO2 (approx 50

and 95 %; final values TBD). Calibrate monthly (or as needed).

A-6

pHOx Service Manual

Dependency Rule and Setup Information:HCT Requires Na+ calibrated for calibration/results.PCO2 Requires pH calibrated for results. (pH checked to assure bicarbonate

is correct.)SO2% Requires Hct calibrated, uses SO2%c if measured SO2 is not available.Hb Requires Hct and SO2 for reporting. Hb method priority: (1) If linked with COOx, then COOx Hb will be reported. (2) If no COOx link, measured results are reported (if available). (3) If no COOx or measured result, calculated result from Hct reading

(Hct/3) is reported. (4) If no Coox, no measured, and no Hct: Reports ‘Default’ Value (Hb).P50 Measured Reported only with PO2 between 30 and 75 mmHg. Also requires calibrated SO2% and Hb/Hct for reporting.Qsp/Qt Requires 2 separate blood samples for determination: Mixed Venous

and Arterial.RI Utilizes input or default FIO2 % for calculation.nCa pHOx-Plus Only. Requires properly installed/calibrated iCa and pH.Anion Gap pHOx-Plus Only. Requires properly calibrated/installed Na+,K+,Cl-, PCO2.Calculated Results require calibrated and ‘proper results’ by measured tests used

in calculation.Suppression (Hb): Only Hb can be suppressed: Option to turn Hb from measured

to calculated (one or the other will always appear).

Passwords:• One password (System Password) for Set Up• 200 user unique passwords available - 3 levels of priviledge:

1. Privilege Level 1: Operators have access to all areas of the analyzer except those protected by the existing System Password. Level 1 opera-tors do not require PDM review and may override this feature. Level 1 operators can override QC lockout.

2. Privilege Level 2: Operators have access to all areas of the analyzer except those protected by the existing System Password. Level 2 op-erators DO require PDM review, but may override this feature. Level 2 operators cannot override QC lockout.

3. Privilege Level 3: Operators have access to analysis only. Level 3 op-erators require PDM review and cannot override this. Level 3 operators cannot override.

Test Suppression: If measured test is suppressed, any calculated/associated tests will also be

suppressed. Tests still will be calibrated w/errors displayed on suppressed channels.

A-7

AppendixAppendix

Interfaces:NOVA COOX, PDM, ASTM interfaces.

Standby Mode:No time dependency will prime and calibrate after coming out of Standby,In Standby only does idle pumping, no fluidic/gas calibration,Auto-Wake-Up, Auto-Cal, and Auto-QC

Micro-Sample Analysis:Micro-Mode Tests: pH, PCO2, PO2Stepped Analysis: Stage 1 = Tip of Probe to Adet 1 (45/55 uLStage 2 = Up to Adet 3.Stage 3 = Between Adet 3 (up to Adet 4): Other Tests Reading (including contact with Reference Electrode)

QC Lockout:Mode A Lockout:

If lockout enabled, and tests fail, unit can be ‘unlocked’ by running any control level and passing all tests.

Mode B Lockout: If lockout enabled, and tests fail, unit can only be ‘unlocked’ by running

specific control level(s) that failed and passing all tests.QC Storage Options:

Mode A QC Storage: If all QC results ‘pass’ range limits, QC results are automatically stored

in memory. If any results ‘fail’ range limits, all results will remain on unit display until either user accepts/rejects results or another QC or sample result is run/displayed (at which time past [non-released] QC results) will be deleted.

Mode B QC Storage: All QC results, whether passing or failing range limits are automatically

stored in memory at end of control analysis.Glucose (Overrange Testing) pHOx-Plus/C/L only:

Gluc Overrange Solution included in Reagent Pack: Done Automatically Daily; Elimination of manual - ampule running of Performance Check Solution

Automatic Glucose Performance done at operator-programmed time daily. (Nova recommends to set time at 5:00 AM for optimum performance.) PASS - FAIL is reporting. Change Gluc Membrane if test lists as FAILED and

repeat.Glucose Auto Polishing Cycle: Automatic polishing of Gluc sensor occurs each day at midnight, followed by 2-point cal, and additional 2-pt calibrations at 2 and 4 AM.

A-8

pHOx Service Manual

Error Code Overview: Code Groups Summary Group: 1 - 6 pH Group: 7 - 13 PCO2 Group: 14 - 19 PO2 Group: 20 - 32 SO2% Group: 33 - 35 Hb Group: 36 - 42 Hct Group: 43 - 48 Na+

Group: 49 - 53 ADET 1 Group: 54 - 58 ADET 2 Group: 59 - 63 ADET 3 Group: 64 - 68 ADET 4 Group: 69 - 75 Flow-Related Group: 76 - 78 QC-Related Group: 79 - 104 Hardware-Software Related Group: 602 - 684 QC Related Group: 686 - 703 RMS Reagent Pack Related

(Refer to the reference manual for specific code details.)

A.2 General Maintenance/QC

1. Flowpath Conditioning (AS NEEDED) (Use serum/plasma/whole blood - Automated Cycle.)

(Sales/Customer Typical Information: Customers as below; Sales- Do once at start of demo/evaluation.)a. Done daily for low-mid volume users (as needed).b. Not required for higher volume users.

2. Control(s) Analysis (DAILY) (Sales Information: SetUp QC Files fully before demo/evaluation;

Customers: Run QC as they wish.)

NOTE: Internal & External Controls are unique for all pHOx models. Controls for pHOx cannot be used on pHOx-Plus or “L”, and visa-versa.

a. Internal Controls - 3 levels (All Tests including gases and SO2)b. External Controls - 3 levels in Multi-Pack Only (pHOx/B =All Tests inc.

SO2 & Gases; pHOx+/L = All Tests, inc. electrolytes/Glu/Lac)

A-9

AppendixAppendix

3. Glucose Overrange Solution (Analysis) pHOx+/L/C Only (DAILY) (Sales Information: Verify Performance Chk passes- AutoRun at start of

demo/evaluation & at each followup) Software Versions 2.0.x or Higher (Elimination of manual-ampule running of Performance Check Solution.) Automatic Glucose Performance done at operator-programmed time daily. PASS - FAIL reporting: Change Gluc Membrane if test lists as FAILED then

repeat test. For Failed Test: Gluc is BACKLIT on Ready screen indicating failure; tests

are unavailable for use until resolved.

4. Flowpath Cleaning (MONTHLY) a. For product release: Do monthly.b. Automated aspiration cycle, use Deproteinizing Solution.c. For pHOx only, remove Na+ electrode (non-glass sensor) and use a

blank electrode.

5. Calibrate SO2 Channel pHOx/+/L/C (MONTHLY) a. Calibrate using 2 external Calibrators (latex suspension).b. Do monthly or more frequently as needed.

6. Clean SO2 Sensor (Optics) Surface: (pHOx/+/L/C) (QUARTERLY)a. Wipe optics surface of SO2 Sensor with lint-free tissue soaked with

Cleaning Solution or 10% Bleach. Rinse with H2O.

7. Replace Na+ Electrode pHOx (AS NEEDED)a. Required to be done based on slope or results performance.b. Correct Hct QC is NOT reliable indicator of Na+ channel perfor-

mance. Condition Na+ Electrode pHOx+/L/C (AS NEEDED)

a. Required to be done based on slope or results performance.b. Use Na+ Conditioning Solution (PN 36856). Remove Na+ and soak in

Conditioning Solution for 2 minutes.

8. Electrode Maintenance Details

pH Electrode• Condition (manually) troubleshooting w/ pH Cond Soln (PN 06856); 10

min soak.• NO need to condition at start of use (Cond Soln in electrode pouch:

pre-soaked)• Long-Life Sensor (similar to Stat Profile M)• Expected Lifetime: 1-2 Years; Warranty = 6 months from installation

A-10

pHOx Service Manual

PO2 Electrode• Screw-On Membrane Cap (w/ Washer attached)• Polish electrode (manually) (10 sec. circular on paper) w/ each mem-

brane change.• Electrode = Long Life; Membrane = Short Life• Electrode estimated life = 1-3 Years; Membrane estimated life = 1-5

months• Warranty: Electrode: 6 months; Membrane: Free of defects at installation• Blood Condition after new electrode is installed (membrane changed

as needed).• Prefilled Membrane Caps (no IFS addition required or to be added)

PCO2 Electrode• Screw-on Membrane Cap (NEW Version; Different than STP/UL);

washer attached.• Membrane Life (new membranes/sensors = 2-12 weeks• NO Conditioning (PCO2 Condition Solution) at StartUp (Caps pre-

soaked in package)• Electrode = Long Life; Membrane = Short Life• Electrode estimated life = 1-3 Years; Membrane estimated life = 1-5

months• Warranty: Electrode: 6 months; Membrane: Free of defects at installation• Blood Condition after new electrode install, membrane change and

as needed.• Pre-filled Membrane Caps (no IFS addition required or to be added)

Na+ ElectrodepHOx:• Short- Life Disposable Electrode. NO maintenance.• Expected Lifetime: 2-4 months. Warranty = 1 month (from install).• Do NOT condition with pH Cond Solution (will damage electrode).• Must remove Na+ when doing Flowpath Cleaning (use ‘blank’).pHOx-Plus/L/C:• Long-Life Sensor (similar to SP-M)• Expected Lifetime: 1-3 Years. Warranty = 6 months (from install).• Condition (as needed): Requires manual soak in Na+ Conditioning

Solution.• NO need to remove Na+ (from flowpath) when doing Flowpath Clean-

ing.

A-11

AppendixAppendix

SO2 Channel pHOx/+/L/C Only• Clean flow-channel with cotton-tipped swab.• Wipe optics surface of SO2 Sensor with lint-free tissue soaked w/

Cleaning Solution or 10% Bleach. Rinse with H2O. (Quarterly)• Expected Lifetime: 6-12 Months Warranty = 3 months from installation

K+ Electrode pHOx-Plus/L/C Only• Mid-life (age) Sensor (similar to SP-M); Max Use Life = 5 months• Blood/protein conditioning for low-volume customers (as needed) iCa Electrode pHOx-Plus/L/C Only• Short-life (age) Sensor (similar to STP-M)• Expected Lifetime: 1-3 Months; Warranty = 7 days or 300 samples from

installation• Blood conditioning (optional) can improve sensor performance.

Cl- Electrode pHOx-Plus/L/C Only• Short-life (age) Sensor (similar to STP-M)• Expected Lifetime: 1-3 Months; Warranty = 7 days or 300 samples from

installation• Blood conditioning (optional) can improve sensor performance• No membrane cap; pre-filled/sealed sensor (similar to STP-M)

Glucose Electrode pHOx+/L/C Only• Electrode Short-life (age) Sensor (similar to STP-M)• Electrode = Long Life; Membrane = Short Life• Electrode estimated life = 1-3 Years; Membrane estimated life = 1-4

weeks (replace membrane if Glucose Performance Check fails or slope drops out)

• Warranty: Electrode: 6 months; Membrane: 3 days from installation• Polish using new YELLOW (1 u) Polishing Paper only.• NO CONDITIONING (soaking) of Sensor required routinely.• Polish sensor ONLY FOR TROUBLESHOOTING (not with each mem-

brane change). (Polish AS NEEDED.)• Polish 10 seconds. (Polish DRY on paper: no water added.)• Rinse with water; blot dry.• Replace new membrane directly.• Condition sensor in Soaking Solution (PN 13409) for 5 minutes with

polishing.

A-12

pHOx Service Manual

Lactate pHOx-Plus-L Only• Electrode = Long Life; Membrane = Short Life• Electrode estimated life = 1-3 Years; Membrane estimated life = 1-3

weeks (Replace membrane if Lact Std D Hi or Lo errors occur or slope drops out.)

• Warranty: Electrode: 6 months; Membrane: 3 days from installation• Polish using new YELLOW (1 u) Polishing Paper only.• NO CONDITIONING (soaking) of Sensor required routinely.• Polish sensor ONLY FOR TROUBLESHOOTING (not with each mem-

brane change). (Polish AS NEEDED.)• Polish 10 seconds. (Polish DRY on paper: no water added.)• Rinse with water; blot dry.• Condition sensor in Soaking Solution (PN 13409) for 5 minutes with

polishing.

A.2.1 Other Analyzer Maintenance

1. Tubings Detailsa. Pump Tubings QUARTERLY May need more frequent changes due to any flow-related problems.b. External Tubings: Waste/Reference Lines (through Solenoids) YEARLYc. Other External Tubings YEARLYd. Internal Tubings (Manifold) YEARLY (done only at time of PM by field engineer)

2. Analyzer Shutdown AS NEEDEDa. Short-Term (<2 Hours)

• Power-off the analyzer• No need to rinse/clean unit for this period of time

b. Long Term (> 48 Hours)• Requires water and air purge of flowpath and all reagent lines• Utilize special Purge Manifold to attach to reagent needle ports and

place• Put all lines (except waste) in distilled water. Process 2-3 fluid primes

with water.• Then remove lines from water and leave open to air; process 2 fluid

primes with air.• After air prime, unit may be shut-off and remove electrodes as needed.• Take pump tubing off pump roller to relieve tubing tension.

A-13

AppendixAppendix

A.3 pHOx Electrode/Standards

Electrode Standards Used in Approximate Delta mV Calibration/Slope (between standards)_________________________________________________________________

pH Stds C & D 33 Na+ Stds A & D 17

PO2 Stds Room Air, Std A and 0 Not Shown (Air to 0) (Ratio Air to Fluid (A) 1 (typical ratio Air/A) for 2 pt Cal; (should be almost same) Std A used for 1 pt Cal) PCO2 Stds A and B 17 (HCO3- for dissolved gas>>monitor pH)

Hct and ADET’s Stds A and D (all ADET’s) > 500 (Adet 3 = Hct) (ideal > 800)

Hb Based on SO2 and Hct TBD

SO2% External Stds 1 and 2 n/a _________________________________________________________________

A.4 Electronics and Mechanicals Overview

ElectronicsCircuit Boards and Main Assemblies

NOTE: Almost all circuit boards on the pHOx utilize surface-mount technology, therefore board component repair is not recommended except for socketed chips.

• Analog Board/Sensor Board• Digital/Controller Board• Display Assembly

A-14

pHOx Service Manual

• Distribution Board• Electrode (Left and Right) Interconnect Boards• Keypad• Power Entry Module• Power Supply Assembly• Printer Printhead and Printer Interface Board• SO2 Sensor Board

A.5 Shutdown Procedure

If the analyzer is to be turned off for more than 24 hours, flush the tubing harness first with distilled water then with air. Use the Flush Fixture (PN 24327) to perform this procedure.

1. Remove the Reagent Pack from the analyzer.2. Install the Flush Fixture into the analyzer in the same way as the Reagent

Pack.3. Place the W-line of the fixture into an empty container.4. Place the other tubing ends into a beaker of distilled water.5. From the Ready/Not Ready screen, press Menu (soft key).6. Select Flowpath Maintenance and press Enter.7. Press Purge (soft key).8. When the cycle ends, take all the tubings out of the distilled water. Leave

the W-line in its container.9. Repeat the purge with air.10. When this cycle is completed, remove the flush fixture.11. The pHOx Analyzer is now ready to be powered off for extended time.

A.6 Start Up Procedure

1. Insert the Reagent Pack and the Quality Control Pack.2. Reinstall the pump tubing.3. Power up the analyzer.4. Perform normal fluid priming and biosensor maintenance.5. Calibrate the analyzer.6. Run QC per your normal procedures.

A-15

Stat Profile pHOxInstallation Check List

Model _____________________ Serial # ____________________ SMR(s) ____________________

Set Up Check Out

Inspect Unit for Shipping Damage Remove Cover, Check all Cable Connec-

tions Fuses Correct Line Voltage Barometric Pressure Sensor Module Temperature

Operator Training Subjects System Identificaion Running an Analysis Operational Status Calibrating the Analyzer

Using and Handling Controls Replacing Reagents and Paper Troubleshooting Flow Test and Flushing Maintenance Set Up

DO

NE

Instrument Check Out

W/R Pump Tubing Harness Reagent Cartridge Control Cartridge (when available) Tubing Harness Electrode / Membranes SO2 Sensor Module Software Card and Version Probe/Microsample Adapter W and R Lines

Troubleshooting Training Subjects Solenoid Valve Tubing Incorrectly Seated Water Test Flow Blockage Other __________________________ Other __________________________ Other __________________________

Fill out and review PACE forms Review customer warranty or service

financing coverage

DO

NE

YE

S

N/A

DO

NE

Nova Support Representative Date Customer Signature Date

The above checks have been performed on your NOVA analyzer as part of your INSTALLATION. We at NOVA BIOMEDICAL believe that you, our valued cus-tomer, are entitled to the highest quality support available. As part of this service, we are leaving a copy of this check list for your records. If any questions arise, feel free to contact us TOLL FREE at 1-800-545-NOVA.

YE

S

N/A

A-16

A-17

Nova Biomedical pHOxPreventive Maintenance Check List

Model _____________________ Serial # _____________________ SMR _____________________

Mechanical Checkout

Pump Assy Sampler Assy W & R Valve Rotary Valve Printer Assy

Fluidic Checkout

Flowcell SO2 Sensor Sensors (replace membrane,

if applicable) External Tubing Internal Tubing Fluids Bay Fluid Fountain Sample Probe Capillary Adapter

INS

PE

CT

RE

PLA

CE

AD

JUS

T

CLE

AN

Electronic Checkout

LCD Screen Electronics Bay Electrical Grounds RS232 Connections Software Version ______

Nova Support Representative Date Customer Signature Date

INS

PE

CT

RE

PLA

CE

AD

JUS

T

CLE

AN

INS

PE

CT

RE

PLA

CE

AD

JUS

T

CLE

AN

The above checks have been performed on your NOVA analyzer as part of your scheduled PREVENTIVE MAINTENANCE service call. We at NOVA BIOMEDI-CAL believe that you, our valued customer, are entitled to the highest quality support available. As part of this service, we are leaving a copy of this check list for your records. If any questions arise, feel free to contact us TOLL FREE at 1-800-545-NOVA. The PM sticker placed on the upper, front corner of the left side cover indicates today’s date, the month your next scheduled PM is due, the initials of the Nova Support Representative and the service report number.

DO

NE

A-18

A-19

�������������

���

����

���

����

��

�����

���

����

���

����

����

����

����

���

����

����

�����

����

���

����

����

����

����

����

����

����

���

����

����

����

����

����

����

����

���

����

����

���

����

����

����

���

����

����

���

����

����

���

����

����

����

����

���

����

����

����

���

����

����

����

����

���

����

����

����

���

����

����

����

����

���

���

����

����

���

����

����

���

����

���

����

����

���

�����

����

����

���

����

����

����

�����

���

����

����

����

����

����

����

���

����

����

���

����

����

���

���

���

�����

����

���

����

���

����

����

���

����

����

���

���

����

�����

���

����

����

���

����

����

����

����

���

����

����

���

���

���

�����

���

����

���

����

����

����

���

����

����

���

���

����

�����

���

����

����

����

�����

����

����

����

�����

����

����

����

���

���

��

����

���

���

����

����

���

����

���

����

����

����

�����

����

����

���

����

���

����

����

����

���

���

��

����

���

���

����

����

���

����

���

����

�����

�����

����

����

�����

���

����

�����

���

������

����

����

����

����

����

����

���

���

����

����

����

����

���

����

����

���

����

����

����

���

����

��

����

���

����

�����

����

���

����

����

����

����

���

����

���

����

���

����

����

����

����

���

���

����

���

���

����

���

����

���

����

����

����

����

���

���

����

�����

���

���

����

����

����

����

����

���

����

���

����

����

����

����

���

����

����

���

����

�����

���

���

�����

���

���

����

����

����

����

����

����

����

���

����

����

����

����

����

����

����

���

����

����

����

����

����

����

����

���

����

����

���

����

���

�����

���

����

����

���

����

���

����

����

����

����

���

����

����

��

���

����

���

����

����

����

��

��

���

����

��

� �

��

��

��

��

��

�� ��

��

��

��

����

��

��

��

��

��

��

���� �� �� ��

��

��

��

��

�����

���

����

����

����

����

����

����

����

��

���

����

���

����

����

����

����

����

����

��

���

����

���

����

���

�����

���

����

���

����

��

���

���

����

����

���

���

����

�����

�����

��

���

����

���

���

����

����

���

���

���

���

����

����

����

�����

����

���

����

����

����

����

���

����

���

����

����

���

����

����

����

����

���

����

���

����

����

���

����

����

����

����

��

��

����

����

���

���

���

����

����

���

����

���

���

����

���

A-20

CN3

CN1

TB1TB2

P2

BARCODE

11

COM1COM3 COM2PCMCIA

CC

FL

MODULER

IGH

T E

LEC

TR

OD

E C

LIP

ACINPUT

LCD PANELSENSOR

J11

KEYPADOPTREX

FP

WIRING DOCUMENT

DSize Drawing Number

Do not scale drawing

Title:

Sheet of

Rev

P1

SEIKOPRINTER

J104

LEF

T E

LEC

TR

OD

E C

LIP

J3

J1

J6 J7

J2

J103

DIGITAL CONTROL BOARD

J8

J7

J9

J10

POWER SUPPLY

ROTARYVALVE

FLUIDPACK

CONTROLPACK

FRONT

PANEL

SAMPLERPUMP

TEST

J102

J101 DISTRIBUTION BOARD

FOR PRODUCT: NOVA PHOX

ANALOG BOARD

P5

P4

CPCN2

DWG #10-1087A

Page S

-1

Main Wire Harness J8 to Pump P1PDM1 Red 1

PDM3 Yellow 2

PDM0 Black 3

PDM2 Grey 4

Digital Control Board J10 to Analog Board J1PGND --- 1ANA V+ --- 2

+5VANA --- 3-5VANA --- 4

ANAGND --- 5ANAGND --- 6ANAGND --- 10

ANARXD# --- 12ANATXD# --- 13

VCCPROT --- 14GND --- 15GND --- 16

INT7 --- 18HTRCTL# --- 19

FAIL# --- 20$1N3313 --- 22PGND --- 24

Analog Board J3 to Sensor Module CN1ADT1-E Yellow 1

ADT1-M Blue 2

THERMISTOR Red 3

FPACK White 4

AGND Black 5

N.C. --- 6

HEATER+ Red 7

HEATER- Brown 8

N.C. --- 9

N.C. --- 10

Digital Control Board J11 to Distribution Board J101 GND --- 1D LP --- 2GND --- 3D M --- 4GND --- 5D FLM --- 6GND --- 7D CP --- 8GND --- 9D DD0 --- 10GND --- 11D DD1 --- 12GND --- 13D DD2 --- 14GND --- 15D DD3 --- 16GND --- 17D SPARE --- 18GND --- 19D ENAVDD --- 20D VEEN --- 21D ENABKL --- 22D YELLOW# --- 23D GREEN# --- 24D VCC --- 25D VCC --- 26D VCC --- 27D VCC --- 28KBY5# --- 29KBY4# --- 30KBY3# --- 31KBY2# --- 32

KBY1# --- 33KBY0# --- 34

KBX3# --- 35KBX1# --- 36KBX2# --- 37

KBX0# --- 38SPCLK --- 39

DACLK --- 41DADAT --- 42SPKR- --- 43

SPKR+ --- 44

Distribution Board J102 to Membrane Keypad KBY5# --- 1

KBY4# --- 2

KBY3# --- 3

KBY2# --- 4

KBY1# --- 5

KBY0# --- 6

KBX3# --- 7

KBX2# --- 8

KBX1# --- 9

KBX0# --- 10

G AN --- 11

G KA --- 12

Y AN --- 13

Y KA --- 14

Power Supply TB2 to Digital Control Board J7PGND Black 1

+24VDC Violet 2

POWER-FAIL White 3

+5VDC Red 4

GND Black 5

+12VDC Orange 6

GND Green 7

-12VDC Yellow 8

Distribution Board J104 to Optrex LCD Panel CN2D FLM --- 1D LP --- 2

D CP --- 3D M --- 4

VADJ --- 5VCCS --- 6GND --- 7VEE --- 8D DD3 --- 9D DD2 --- 10D DD1 --- 11D DD0 --- 12DISPOFF# --- 13

AC Mains Entry Module To Power Supply TB1AC Line Brown 1

AC Neutral Blue 3

Earth Ground Green/Yellow 5

Digital Control Board J9 to Seiko Printer CN3PV1 --- 1

GND --- 2

PV2 --- 3

PV2 --- 4

PGND --- 5

PGND --- 6

P STB# --- 11

P DR1 --- 12

P DR2 --- 13

P DR3 --- 14

P DR4 --- 15

P DR5 --- 16

P DR6 --- 17

P DR7 --- 18

P DR8 --- 19

P BUSY --- 20

P ACK# --- 21

P ERR# --- 22

P PE --- 23

GND --- 24

P RESET# --- 25

P ONLINO --- 26

P ONLINI --- 27

P FEED# --- 28

P BFEED# --- 29

P NMI# --- 30

PV2 --- 31

PV2 --- 32

GND --- 33

PV1 --- 34Main Wire Harness J8 to Control Pack CPTOUCH Orange 1

N.C. --- 2

N.C. --- 3

GND Black 4

Main Wire Harness J8 to Fluid Deck P4VALVE1# White 1

VALVE0# Violet 2

VALVECOM Black 3

N.C. --- 4

ADT1-E Yellow 5

ADT1-M Blue 6

Main Wire Harness J8 to Sampler P5SMD1 Yellow 1

SMD3 Red 2

SMD0 Black 3

SMD2 Grey 4

N.C. --- 5

N.C. --- 6

Digital Control Board J8 to Main Wire HarnessPGND P9 Black 1

PGND P6 Black 3

PMD1 P1 Red 5

PMD3 P1 Yellow 6

PMD0 P1 Black 7

PMD2 P1 Grey 8

VALVECOM P4 Black 9

TOUCH FP Orange 10

SMD3 P5 Red 11

DOORSW# P6 White 12

SMD2 P5 Grey 13

PGND P10 Black 14

SMD1 P5 Yellow 15

GND P10 Black 16

SMD10 P5 Black 17

PGND FP Black 18

RVD3 P2 Red 19

SAMPLED P10 Red 20

RVD2 P2 Grey 21

ROVLED P9 Red 22

HEATER+ P3 Red 23

HOME# P9 White 24

DOORLED --- N.C. 25

SAMHOME# P10 White 26

HEATER- P3 Brown 27

LAMP- P7 Brown 28

GND --- N.C. 29

LAMP+ P7 Red 30

VALVE1# P4 White 31

RVD0 P2 Black 32

VALVE0# P4 Violet 33

RVD1 P2 Yellow 34

PGND P9 Black 2

PGND --- N.C. 4

Main Wire Harness J8 to Rotary Valve P2RVD1 Yellow 1

RVD3 Red 2

RVD0 Black 3

RVD2 Grey 4

N.C. --- 5

N.C. --- 6

Main Wire Harness J8 to Fluid Pack FPTOUCH Orange 1

N.C. --- 2

N.C. --- 3

GND Black 4

Distribution Board J103 to Optrex Backlight CN1COM White 1

N.C. --- 2

N.C. --- 3

HOT Black 4

Main Wire Harness J8 to Door Switch P6DOORSW# White 1N.C. --- 2

PGND Black 3

Main Wire Harness J8 to Lamp P7LAMP- Red 1

LAMP+ Brown 2

Main Wire Harness J8 to Rot Val P9N.C. --- 1

N.C. --- 2

ROVHOME# White 3

ROVLED Red 4

GND Black 5

GND Black 6

Main Wire Harness J8 to Sampler P10N.C. --- 1

N.C. --- 2

SAMHOME# White 3

SAMPLED Red 4

GND Black 5

GND Black 6

10-5-1998_13:54

-12VDC

TCLKTCLK

TCLK

TCLK

TMS

TMS

TMS

TMS

DRIVENB#

RAMADD10

RAMADD9

RAMADD8

RAMADD7

RAMADD6

RAMADD5

RAMADD4

RAMADD3

RAMADD2

WDTOUT

ADS#

RD#

W/RD#

D/C#

BRST#

CLK2

CLKOUT

MECHENB#

BBLE#

ADB1

ADB2

ADB3

ADB4

ADB5

ADB6

ADB7

ADB8

ADB9

ADB10

ADB11

ADB12

ADB13

ADB14

ADB15

ADB16

ADB17

ADB18

ADB19

ADB20

ADB21

ADB22

CHOKEWR#

CS5B#

CS5A#

CS4B#

CS4A#

CS3B#

CS3A#

BBHE#

VCC

MEMSIZ0#

DB2

DB1

DB0

DB3

DB4

DB5

DB6

DB7

DB8

DB9

DB10

DB11

DB12

DB13

DB14

DB15

DB0

DB1

DB2

DB3

DB4

DB5

DB6

DB7

DB8

DB9

DB10

DB11

DB12

DB13

DB14

DB15

D13

D15

D14

D12

D11

D10

D9

D8

D7

D6

D5

D4

D3

D2

D1

D0

BS8#

ADB1

DB7

ADB18

ADB17

CHOKEWR#

BRD#

CS1#

DB0

DB1

DB2

DB3

DB4

DB5

DB6

BBLE#

ADB2

ADB3

ADB4

ADB5

ADB6

ADB7

ADB8

ADB9

ADB10

ADB11

ADB12

ADB13

ADB14

ADB15

ADB16

UCS#

ADB1

ADB2

ADB3

ADB4

ADB5

ADB6

ADB7

ADB8

ADB9

BBLE#

ADB20

ADB19

ADB18

ADB17

ADB16

ADB15

ADB14

ADB13

ADB12

ADB11

ADB10

ADB25

ADB24

ADB23

ADB22

ADB21

DB14

DB0

DB1

DB2

DB3

DB4

DB5

DB6

DB7

DB8

DB9

DB10

DB11

DB12

DB13

BRSTSYNC

BWR#

BRD#

DB15

VCC

ADB17

RD#

DB14

ADB16

DB0

ADB13

ADB14

ADB15

VCC

CHOKEWR#

BRD#

CS2#

LAMPON#

A16

BLE#

BHE#

A1

A2

A3

A4

A5

A6

A7

A8

WR#

RD#

M/IO#

D/C#

WR#

RD#

M/IO#

D/C#

W/RD#

ADS#

A25

A24

A23

A22

A21

A20

A19

A18

A17

ADS#

W/RD#

A25A15

A24A14

A23A13

A22A12

A21A11

A20A10

A19A9

A18A8

A17A7

A16A6

A15A5

A14A4

A13A3

A12A2

A11A1

A10BHE#

A9BLE#

BADS#

DACK0#

DREQ0

BUSY

CLK100

CLK100

VCC

INT7BIST#

ANARXD#

BRSTSYNC

BADS#

BCODE

BCTXD

BCRXD

VCC

BRST#RST# RST#

BRST#

DREQ0

DOOR

TDO

BS8#

BW/RD#

VCC

VCC

DACK0#

BADS#

ANATXD#

TMRGATE0

TMROUT1

BCVCC

TDI

VCC

AUDIOENB

TMROUT0

VCC

BCLK2

BWR#

BRD#

1.8MCLK

VIDCLK

VCC VCC

CS3#

CS2#

CS1#

CS0#

1.8MCLK

INT3

INT2

INT1

INT4

INT6

PMPCLK

XCDONE

XCRLD#

XCDATA

CLKOUT

WDTOUT

CS6#

UCS#

READY#

CLK2

CLK2

CLK2

D15

D14

D13

D12

D11

D10

D9

D8

D7

D6

D5

D4

D3

D2

D1

D0

NMIENB#

NMI

RSTSYNCRSTSYNC

DRIVENB#

BD/C#

BM/IO#

MECHENB#

VCC

DRIVENB#

DOORSW#

VCC

VCC

VCC

XINIT

DB0

DB1

DB2

DB3

DB4

DB5

DB6

DB7

DB8

DB9

DB10

DB11

DB12

DB13

DB14

DB15

GND

ADB7

ADB6

ADB5

ADB4

ADB3

ADB2

ADB1

ADB8

ADB9

ADB10

ADB11

ADB12

ADB13

ADB14

ADB15ADB23

ADB22

ADB21

ADB20

ADB19

ADB18

ADB17

ADB16

GND

ADB25

ADB24

BBHE#

BBLE#

BW/RD#

BD/C#

BM/IO#

RAMADD0

RAMADD1

RAMADD2

RAMADD3

RAMADD4

RAMADD5

RAMADD6

RAMADD7

BCLK2

CLKOUT

1.8MCLKRAS0#

GND

RAMADD10

RAMADD9

RAMADD8

CASL#

CASH#

RAS1#

CS3

VCC

GND

BRSTSYNC

PMPCLK

CHOKEWR#

LATERD#

BWR#

BRD#

READY#

VCC

DB1

DB2

DB3

DB4

DB5

DB7

DB6

ADB2

BBLE#

ADB1

ADB3

ADB4

ADB5

ADB6

ADB7

ADB8

ADB9

ADB10

ADB11

ADB12

DB15

DB13

DB12

DB11

DB10

DB9

DB8

DB7

DB6

DB5

DB4

DB3

DB2

DB1

DB0

BBLE#

BBHE#

ADB1

ADB2

ADB3

ADB4

ADB5

ADB6

ADB7

ADB8

ADB9

ADB10

ADB11

ADB12

ADB13

ADB14

ADB15

ADB16

ADB25

ADB24

ADB23

ADB22

ADB21

ADB20

ADB19

ADB18

VCC

TP1

1

TP2

1

D9

R254

4.75K

R164

4.75K

ANARXD#

U21

11 10

U21

9 8

4.75KR145

1.0 OHM 1W

R16

1.0 OHM 1W

R1

1.0 OHM 1W

R18

R71

SMD050

SMD050

R44

SMD050

R45

U10

3 1

4

1 OHM 1W

R149

16V47

C75 C64

0.1

C63

0.1

10 OHM 1WR70

10 OHM 1W

R48-5VANA

ANAGND

+5VANA

40V22

C27

C3322

40V

100R98

16V

47

C119100PF

C8

C28

100PF 100PF

C7

C65

100PF

C96

100PF

C29

2240V

R51

1.00K

R123221

R138221

R137221

4.75K

R148

221

R134

Y1

14.318MHZ

1

4

R199 33

4.75KR165

4.75KR166

4.75KR162

1.00KR143

221

R188

R169 4.75K

R156 4.75K

R140 4.75K

R168 4.75K

R167 4.75K

R141 4.75K

R124 4.75K

R139 4.75K

R157 4.75K

R127 4.75K

R142 4.75K

R125 4.75K

R129 4.75K

R135

4.75K

4.75K

R170

R158

4.75K

221

R189

R176

221

R155 4.75KR173 4.75K

R218

4.75K

U244

56

4.75K R2094.75K

R213

U33

333436373840414243444547484951525455

689

1012131415

19202123

3027292831

56

126

17

2

24

16

53

U34

333436373840414243444547484951525455

689

1012131415

19202123

3027292831

56

126

17

2

24

16

53

10R136

R150 33

R133 33

J5

42

97531

1086

PGND

PWRFAIL#

VCC

16V47

C41

16V47

C34

+24VDC

4.75K

R15

PGND

+5VANA

-5VANA

ANAGND

VCCPROT

ANA_V+

FAIL#

ANATXD#

J10

24

2

2322212019181716151413121110987

43

6

1

5

D10

U32

75

8

4

1

U27

333436373840414243444547484951525455

689

1012131415

19202123

30272928 31

56

126

17

2

24

16

53

4.75KR144

4.75KR212

4.75KR211

R182

4.75K

R183

4.75K

U29

107108111

106

132104

131102129105101

90

1103341

73120

126125124123122

8786858482807574

93949596

928991

118112117128113

98777978

76252426

119103

11421

5678

101112131416181920212223

373942434445484950515253545556575859616263656667687072

3534

43227293040

115

99

R197 33R195 33

R187 33

R198 33R196 33

R154 33

35V

1000

C2

+24VDCIN

4.75KR175

P11

2

3

4

5

6

7

8 9

10

11

12

13

14

15

16

P21

2

3

4

5

6

7

8 9

10

11

12

13

14

15

16 P31

2

3

4

5

6

7

8 9

10

11

12

13

14

15

16 P61

2

3

4

5

6

7

8 9

10

11

12

13

14

15

16

P41

2

3

4

5

6

7

8 9

10

11

12

13

14

15

16

R17

1.0 OHM 1W

+12VDC

221R49

221R46

U93 4

1

J7

8

7

6

5

4

3

2

1

U21

3 4

4.75K

R200

U14

12 13

U37

312422

121110

98765

262325

42829

32

1314151718192021

27

4.75K

R146

U19

65

R147

4.75K

U11

43

INT7

4.75K

R47

R50

4.75K

J1162738495

R126 4.75K

R12833 4.75K

221

R174

R215

221

U25

711

614

9101516

2

3

1

U28

1932312818171615141312

47

42414340

3444

64626058565452502725232197536563615755535149262422208642

4533

70696867

D7

D8

J6

30313223456646566373839

4140

5655545350

4847461920141321108

111222232425262728297915

163336424344455758596061626367185217511343568

49

74HCT14

U23

21

74HCT14

U23

8 9

74HCT14

U2343

74HCT14

U23

1011

74HCT14

U23

12 13

74HCT14 U23

6 5

U44

12 13

HTRCTL#

ANARST

DRIVENB#

U35

5

3

2

18

19

20

21

22

23

24

25

26

27

28

29

30

31

32

33

34

17

16

15

14

13

12

11

10

9

8

7

6

4

1

E6

R1634.75K

+24VDC

ADB23

MEMSIZ1#

MEMSIZ0#

BWR#

R2084.75K

BIST#

VCC

M/IO#

XTDO

CS0#

RAS1#

RAS0#

CASL#

CASH#

BWR#

LATERD#

RSTSYNC

POEH#

POEL#

BUSY

PSTB#

READY#

ACK#

PDEN

DREQ0

DACK0#

RAMADD0

RAMADD1

DB0

DB1

DB2

DB3

SUICI

U38

50

55

70

71

72

69

66

65

80

81

79

78

82

86

85

83

54

67

74

73

57

62

60

58

59

77

56

12

15

30

16

14

9

7

25

24

34

37

18

23

1

3

4

99

98

100

33

22

39

2

29

31

32

8

92

89

91

95

26

27

10

11

1764756

94

96

IPCMTX

RXD3TXD3

IPCMRX

PWRFAIL#

4.75KR114

E54.75KR177

SUICIDE

E7

E501SUICIDE SUICI

GREEN

7406DOC

7406DOC

T

T

T

MC78M05CDT

GND

INOUT

74HC02D

74ABT18245ADL

2A9

2A8

2A7

2A6

2A5

2A4

2A3

2A2

2A1

1A9

1A8

1A7

1A6

1A5

1A4

1A3

1A2

1A1

1B4

1B5

1B6

1B7

1B8

1B9

2B1

2B2

2B5

2B6

2B7

2B8

TDI

TDO

TCK

TMS2OE#

1OE#

1DIR

2DIR

2B4

1B1

2B9

2B3

1B3

1B2

74ABT18245ADL

2A9

2A8

2A7

2A6

2A5

2A4

2A3

2A2

2A1

1A9

1A8

1A7

1A6

1A5

1A4

1A3

1A2

1A1

1B4

1B5

1B6

1B7

1B8

1B9

2B1

2B2

2B5

2B6

2B7

2B8

TDI

TDO

TCK

TMS2OE#

1OE#

1DIR

2DIR

2B4

1B1

2B9

2B3

1B3

1B2

MB

RS

140T

MAX707CSA

RST

RST#

PFO#PFI

MR#

74ABT18245ADL

2A9

2A8

2A7

2A6

2A5

2A4

2A3

2A2

2A1

1A9

1A8

1A7

1A6

1A5

1A4

1A3

1A2

1A1

1B4

1B5

1B6

1B7

1B8

1B9

2B1

2B2

2B5

2B6

2B7

2B8

TDI

TDO

TCK

TMS 2OE#

1OE#

1DIR

2DIR

2B4

1B1

2B9

2B3

1B3

1B2

KU386EXTC

P1.5/LOCK#

P1.6/HOLD

P1.7/HLDA

P1.4/RI0#

P2.7/CTS0#

P1.2/DTR0#

P2.6/TXD0

P1.1/RTS0#

P2.5/RXD0

P1.3/DSR0#

P1.0/DCD0#

NMI

RESET

BS8#

NA#

SMI#

SMIACT#

P2.4/CS4#

P2.3/CS3#

P2.2/CS2#

P2.1/CS1#

P2.0/CS0#

P3.7/COMCLK

P3.6/PWRDWN

P3.5/INT3

P3.4/INT2

P3.3/INT1

P3.2/INT0

P3.1/TMROUT1

P3.0/TMROUT0

TMRCLK0/INT4

TMRGATE0/INT5

TMRCLK1/INT6

TMRGATE1/INT7

BUSY#/TMRGATE2

PEREQ/TMRCLK2

ERROR#/TMROUT2

DRQ1/RXD1

DACK1#/TXD1

DRQ0/DCD1#

DACK0#/CS5#

EOP#/CTS1#

STXCLK/DSR1#

SRXCLK/DTR1#

SSIOTX/RTS1#

SSIORX/RI1#

TCK

TDI

TDO

TMS

TRST#

CLKOUT

WDTOUT

CS6#/REFRESH#

UCS#

D0

D1

D2

D3

D4

D5

D6

D7

D8

D9

D10

D11

D12

D13

D14

D15

BLE#

BHE#

A1

A2

A3

A4

A5

A6

A7

A8

A9

A10

A11

A12

A13

A14

A15

A16

A17

A18

A19

A20

A21

A22

A23

A24

A25

WR#

RD#

LBA#

READY#

M/IO#

D/C#

W/R#

ADS#

CLK2

FLT#

MC79M05CDT

GND

INOUT

7406DOC

ACT05

OD

AT29C010A-90JC

A8

D7

D6

D5

D4

D3

D2

D1

D0

A16

A15

A14

A13

A12

A11

A10

A9

A6

A4

A3

A2

A1

A0

CE#

RD#

WR#

A7

A51A18

30A17

74LS14D ACT05

OD

SGND

RI

DTR

CTS

TXD#

RTS

RXD#

DSR

DCD

SHELL

AV9154-27CS16

1.8MHZ

12MHZ

14.3MHZ

24MHZ

50MHZ

F0

F1

F2

OE

4VCC1

X1

X2

13VCC2

MT8D132M-6

AT0

AT1

AT2

AT3

NC0

NC1

DQ0

DQ1

DQ2

DQ3

DQ4

DQ5

DQ6

DQ7

DQ8

DQ9

DQ10

DQ11

DQ12

DQ13

DQ14

DQ15

DQ16

DQ17

DQ18

DQ19

DQ20

DQ21

DQ22

DQ23

DQ24

DQ25

DQ26

DQ27

DQ28

DQ29

DQ30

DQ31

RAS0#

RAS1#

CAS0#

CAS1#

CAS2#

CAS3#

WE#

A0

A1

A2

A3

A4

A5

A6

A7

A8

A9

A10

GREENGREEN

FUJITSU FCN565PO68G0231E4

D0

D1

D2

D3

D4

D5

D6

D7

D8

D9

D10

D11

D12

D13

D15

D14

A25

A24

A23

A22

A21

A19

A18

A17

A16

A15

A14

A13

A12

A11

A10

A9

A8

A7

A6

A5

A4

A3

A2

A1

A0

CE1#

OE#

WE#

BSY#

WP

CD1

CE2#

RFSH

RFU1

RFU2

RFU3

RST

WT#

RFU4

REG#

BVD1

BVD2

CD2

VPP1

VPP2

VCC1

VCC2

GND1

GND2

GND3

GND4

A20

74LS14D

DS1644P

NC1

PFO

WE

OE

CE

DQ7

DQ6

DQ5

DQ4

DQ3

DQ2

DQ1

DQ0

GND

NC3

NC2

A14

A13

A12

A11

A10

A9

A8

A7

A6

A5

A4

A3

A2

A1

A0

A15

A16

VCC

2

1

NOVA Biomedical 200 Prospect Street P.O. Box 9141Waltham, MA 02254-9141

SIZE DRAWING NUMBER

DRAWING TITLE

DO NOT SCALE DRAWING LAST MODIFIED ON: SHEET OF

REV.

E

8 7 6 5 4 3 2 1

8 7 6 5 4 3 2 1

B

C

D

F

G

A

E

H

G

F

E

D

C

B

A

H

/RESET

A1

A10

A11

A12

A13

A14

A15

A16

A17

A18

A19

A2

A20

A21

A22

A23

A3

A4

A5

A6

A7

A8

A9

44ACK

ADS#

BHEN#

BLEN#

42BUSY

CASH#

CASL#

CHWR#

CLK2

CLKOUT

CS0# CS3A#

CS3B#

CS4A#

CS4B#

CS5A#

CS5B#

D/C#

48DACK0

47DREQ0

DRVEN#

IO#

LTRD#

MECHENB#

46PDEN

43PSTB#

RA0

RA1

RA10

RA2

RA3

RA4

RA5

RA6

RA7

RA8

RA9

RAS0#

RAS1#

RD#

51RDYIN

RSTSYN

SIZ0

SIZ1

21DB00

TCK

TDITDO

TMS

W/R#

WDTOUT

87 386BIST

19 BWR

POEH

52POEL

35PP35

38PP38

49PP49

63PP63

Processor & Memory

3

24292-01

Epm7128SQC100-7

CONNECTS TO CHASSIS GND

FRAME OF PCMCIA

NET MUST CARRY 5A CURRENT

RIGHT NEXT TO J7

LOCATE THESE PARTS

CLOCK GENERATOR

POWER ON RESET

DRAM SELECTION

1MEG = No jumper

4MEG = No jumper

16MEG = Yes jumper

JTAG PORT

ESR =< 0.1 ohms

Ripple I => 1.5AS12

105c / 3000 hr

IPCM PORT

14.318 MHz

50.113 MHz

1.8474 MHz

MANUAL RESET

16-BIT

LOGIC ANALYZER TEST POINTS

DRAM CONTROLLER

SYSTEM DRAM

JUMPER SELECTS SIZE

OR 4M X 32 SIMM DRAM

256K X 32, 1M X 32,

DIAGNOSTIC LEDS

"ADS" "DRV""RST"

BARCODE

STORAGE

PATIENT DATA

SYSTEM BUS BUFFERS

RTC / NVRAM

128K x 8

8-BIT

ACCESSCARD

install BARCODE pullup OR pulldown only

so software sees the right levels when nothing is connected

POWER SUPPLY CONNECTOR

PC

MC

IA C

AR

D C

ON

NE

CT

OR

ANALOG BOARD CONNECTOR

ALTERA

Schematic, Cortex Digital Board

DWG #10-1071B

Page S

-3

10-5-1998_16:11

TXD3

RXD3

VCC

ADB2

DB7

PE_INT

PERR_INT

POEL#

POEH#

IPCMRX

IPCMTX

P_DR6P_DR5P_DR4P_DR3P_DR2P_DR1

PDEN

VCC

DB8

VCC

VCC

DCD1#DSR1#RXD1RTS1#TXD1CTS1#DTR1#RI1#

ACK#

PV2

PV1

RI1

CTS1

RTS1

DSR1

DTR1

TXD1#

RXD1#

DCD1

DCD2DSR2RXD2#RTS2TXD2#CTS2DTR2RI2

LATERD#CHOKEWR#

INT3INT2

BRST#1.8MCLK

CS5A#CS3B#CS3A#

DCD2#DSR2#RXD2RTS2#TXD2CTS2#DTR2#RI2#

DB0DB1DB2DB3

DB4DB5DB6DB7

DB9DB10DB11

DB12DB13DB14DB15

FSELAFSELBFSELCFSELD

DB0DB1DB2DB3DB4DB5DB6

BBLE#ADB1

BUSY

P_RESET#

P_NMI#VCC

P_BFEED#P_FEED#P_ONLINIP_ONLINO

P_PEP_ERR#P_ACK#P_BUSY

P_DR8P_DR7

P_STB#

BPSTB#

VCC

CTS3

DSR3

PROGPULSETXD4

PGMEN#

PERR_INTPE_INT

RXD4

VCC

DCD3

RI3DTR3

RXD3#RTS3TXD3#

VCC

YELLOW#GREEN#

XCCLK

DB7DB6DB5DB4DB3DB2DB1DB0

ADB2ADB1BBLE#

CS4A#CS4B#CS5B#

CHOKEWR#LATERD#

1.8MCLKBRST#

INT4INT6

CTS3#

RTS3#

DSR3#DCD3#

OPTION

R504

6.81K

R400

DB9

DB15DB14DB13DB12DB11DB10

DB8DB7DB6DB5DB4DB3DB2

DB0

ADB1ADB2ADB3ADB4ADB5ADB6ADB7ADB8ADB9ADB10ADB11ADB12ADB13ADB14ADB15ADB16ADB17ADB18ADB19

ADB22

ADB20

BBLE#

DB1

TP4

1

TP3

1

PROGPULSE

D_SPARE

D_ENAVDD

U2

16

12

11 1

32569

1314171819

15

874

D_YELLOW#

TXD4

AUDIOENB

CLICK

U24

1089

KBX3#

KBY1#

1.00kR244

D_LP

D_VEEN

D_M

D_FLM

D_CP

D_DD0

D_DD1

D_DD2

D_DD3

D_ENABKL

4.75kR203

4.75kR153

4.75kR38

4.75kR35

DD0DD1DD2

RXD4

C540.1

C43

0.1

C70 0.1

C67 0.1

C86 0.1

C970.1

C710.1

C1210.1

C900.1

C1200.1

C87

0.1

C55 0.1

C42 0.1

C680.1

C66

0.1

221R33221R8

R192

221

U26

3025282624413129

5352515049484746

2120191817161514

333435

32338

3637

439444345605963656667681

8562121013116

57565558

2

U22

1089

16V

C11647

R221 10

16V

C115

47

R22010

READY#

R201

4.75k

R204 4.75k

R205 4.75k

R2194.75k

R230

4.75k

R224 33R222 33R223 33

33R225

R2334.75K

R2314.75K

1.00K

R119

221R181

221R184221R185221R186

221R190

1.00KR160

1.00KR100

1.00K

R73

BRSTSYNCVIDCLK

ADS#M/IO#

W/RD#

BCLK2

VMA1

VMA2

VMA3

VMA4

VMA5

VMA6

VMA7

VMA8

VMD0

VMD1

VMD2

VMD3

VMD4

VMD5

VMD6

VMD7

VMD8

VMD9

VMD10

VMD11

VMD12

VMD13

VMD14

VMD15

U20

21

25

17

13

19687

20225

26

24

2818

3921

234

27

16

1514

12

U15

12

1415

16

274

231293

1828

24

265

2220

786

19

13

17

25

21

FLMLPCPM

DD3

U39 19

8

1

53

1517

642

1179

13

12141618

221R31

221R6221R32221R7

221R34R9 221

221R12221R41

221R40

221R5

221R39

221R37

221R36

221R10

221R11

C44

0.1

R118

3.32K

R43

SMD100

D_VEEN

D_DD3D_DD2D_DD1

D_YELLOW#

D_FLMD_LPD_CPD_M

D_ENAVDD

D_DD0

J11

40383634323028262422201816141210864

39373533312927252321

17151311

97531 2

19

U30

12

1415

16

274

231293

1828

24

265

2220

786

19

13

17

25

21

U43

158157

48

55

53

711

134

145146147148149150151152153

1

135136137138140141143144

133

3130

181716151413

2827262524232120

5654

76

112113

66656463

47

100101102103104105106107108

114116

126127128129130131

117119120121122123124125

110

61605857

111

109

33

3736

2

45

3

159155

8

403938

3435

6

VCC

CS3

221R202

KBX1#

KBY4#KBY2#

KBY0#

KBX0#

KBY3#

KBY5#

R217221

D_ENABKL

BPSTB#

R210

221

221R214

U40

262524232219181716

3837363433323110

9875432

R2164.75K

R2324.75K

D_VCC

D_GREEN#

D_VCC

D_VCC

1.00K

R72

1.00K

R117

U31

16

12

11 1

32569

1314171819

15

874

R13 4.75k

221R42

0.1

C85

VCC

DADAT

SPKR+

U21

5 6

SMD050

R29

E8

R151

1.0 OHM 1W

R171

1.0 OHM 1W

R99

1.0 OHM 1W

R161 4.75k

SPKR-

0.1

C142C141

20.5K

R28

20.5K

R4

560

R401

R103 4.75K

R120

10K

TOUCH

LM4861M

U45

5678

4321

C143

J9

302928272625

23222120

19181716151413121110987

3324

2

323143

34

6

1

5

J2162738495

J3162738495

J4162738495

R101

1.0 OHM 1W

D_GREEN#YELLOW#

GREEN#

U41

19

8

1

53

1517

642

1179

13

12141618

R179

220

D_SPARE

SMD100

R249

VCC

221R255

U21

1 2

20KR121

C5

VMA0

221R172

D13 D14D12D11

U24

123

TMROUT0

U46

1

23

U501

5

3

2

1

4

6

R1521.00K

221R193R194 221

PSTB#

U36

3025282624413129

5352515049484746

2120191817161514

333435

32338

3637

439444345605963656667681

8562121013116

57565558

2

BBHE#

ADB21

ADB23

1.00KR159

R102

49.9

Q5

3

1

2

D5

D6

D500

VCC

+15VDC

10PFC56

R104

4.75K

C400

1000PFU16

15

24

3

+15VDC

U500

15

24

3+2.5VREF

24.9K

R501

324K

R502

R122 4.75K

11.8VDC

C58

100PF

C500 .1UF

+15VDC

R30

2 31

R3

4.75k

OPTION

R503

DACLK

KBX2#

SPCLK

P_INIT#

2

221

R206221

R207

R191

221

DTR3#

IPCMRX

IPCMTX

N74F373D

OEEN

74HC02D

TL16C552FN

TRI

SLIN#

STB#

AFD#

INIT#

RI1#

DTR1#

CTS1#

TXD1

RTS1#

RXD1

DSR1#

DCD1#

BIDEN

ACK#

PE

BUSY

SLCT

ERR#

INTP

INT1

INT0

INTSEL#

BDO

RESET#

CLK

IORD#

IOWR#

CSP#

CS1#

CS0#

A0

A1

A2

D0

D1

D2

D3

D4

D5

D6

D7

PD7

PD6

PD5

PD4

PD3

PD2

PD1

PD0

DCD0#

DSR0#

RXD0

RTS0#

TXD0

CTS0#

DTR0#

RI0#

74HC02D

MAX211ECAI

T4D

SHDN

V-

V+

R5Q

T2D

R1Q

T1D

T3D

R4Q

R2Q

R3Q

EN#

T4Q

R5D

T2Q

R1D

T1Q

T3Q

R4D

R2D

R3D

C2-

C2+

C1-

C1+

MAX211ECAI

T4D

SHDN

V-

V+

R5Q

T2D

R1Q

T1D

T3D

R4Q

R2Q

R3Q

EN#

T4Q

R5D

T2Q

R1D

T1Q

T3Q

R4D

R2D

R3D

C2-

C2+

C1-

C1+

N74F244D

T

4143

4244

MAX211ECAI

T4D

SHDN

V-

V+

R5Q

T2D

R1Q

T1D

T3D

R4Q

R2Q

R3Q

EN#

T4Q

R5D

T2Q

R1D

T1Q

T3Q

R4D

R2D

R3D

C2-

C2+

C1-

C1+

CVCC1

CVCC2

ENAVDD

M

FLM

STANDBY#

BE3#

ADDHI

CLKIN

CPURST

BCFG0

BCFG1

BCFG2

BHE#

14.3MCLK

RESET

A21

A23

A22

A20

LDEV#

LRDY#

RD#

ADS#

M/IO#

WE#

PNL0

PNL1

PNL2

PNL3

RAS#

MD9

MD8

MD7

MD6

MD5

MD4

MD3

MD2

MD15

MD14

MD13

MD12

MD11

MD10

MD1

MD0

MA8

MA7

MA6

MA5

MA4

MA3

MA2

MA1

MA0

ENAVEE

PNL4

PNL5

PNL6

PNL7

LCAS#

HCAS#

ENABKL

LP

SHFCLK

D9

D8

D7

D6

D5

D4

D3

D2

D15

D14

D13

D12

D11

D10

D1

D0

BLE#

A9

A8

A7

A6

A5

A4

A3

A2

A19

A18

A17

A16

A15

A14

A13

A12

A11

A10

A1

F65535A

MT4C16257-70

27OE#

14RAS#

13WE#

28HCAS#

29LCAS#

A8

A7

A6

A5

A4

A3

A2

A1

A0

39D15

D14

D13

D12

D11

D10

D9

D8

D7

D6

D5

D4

D3

D2

D1

D0

N74F373D

OEEN

7406DOC

T

2

1.5uF

+IN

-IN

BYPASS GND

SHUTDOWN

VDD

VO1

VO2

.22uF

SGND

RI

DTR

CTS

TXD#

RTS

RXD#

DSR

DCD

SHELL

SGND

RI

DTR

CTS

TXD#

RTS

RXD#

DSR

DCD

SHELL

SGND

RI

DTR

CTS

TXD#

RTS

RXD#

DSR

DCD

SHELL

N74F244D

T

7406DOC

1.5uF

GREENGREEN

GREENGREEN

74HC02D

DS2505

NC

DATA

GND

DS2505P

NC NC

DATA

GND NC

NC

TL16C552FN

TRI

SLIN#

STB#

AFD#

INIT#

RI1#

DTR1#

CTS1#

TXD1

RTS1#

RXD1

DSR1#

DCD1#

BIDEN

ACK#

PE

BUSY

SLCT

ERR#

INTP

INT1

INT0

INTSEL#

BDO

RESET#

CLK

IORD#

IOWR#

CSP#

CS1#

CS0#

A0

A1

A2

D0

D1

D2

D3

D4

D5

D6

D7

PD7

PD6

PD5

PD4

PD3

PD2

PD1

PD0

DCD0#

DSR0#

RXD0

RTS0#

TXD0

CTS0#

DTR0#

RI0#

22657

C

E

B

MB

RS

140T

3

MBRS140T

MBRS140T

MIC6211

V-

V+

MIC6211

V-

V+

10K

CW

NOVA Biomedical 200 Prospect Street P.O. Box 9141Waltham, MA 02254-9141

SIZE DRAWING NUMBER

DRAWING TITLE

DO NOT SCALE DRAWING LAST MODIFIED ON: SHEET OF

REV.

E

8 7 6 5 4 3 2 1

8 7 6 5 4 3 2 1

B

C

D

F

G

A

E

H

G

F

E

D

C

B

A

H

DUART

PRINTER CONNECTOR

stb used for b_feed output

DUART/PRINTER PORT

VGA VIDEO CONTROLLER

COM1

COM2

option for either package

install only one ds2505

SYSTEM STATUS LEDS

COM3

TOUCH MEMORY INTERFACE

IPCM DEBUG PORT

RS232 DRIVER/RECEIVERS

24292-02

Schematic, Cortex Digital BoardComm, Printer, Display

Touch Memory

3

DOORBOARD FUSE

DWG #10-1072B

Page S

-4

ROVHOM2#

XXSP8

XXSP7

ROV2IN

VCC

SPCLK

VCC

ROVLED2

ROVHOM2#

ROVLED

BRSTSYNC

ANARST

XINIT

PUMPHOLD

SAMPHOLD

SMPCMP12

+2.5VREF

+2.5VREF

+2.5VREF

ROVGOOSE#

ROVHOLD

PUMP1

XOUT3

XOUT0

VCC

VALBOOST

DENNA

ROV2

FSELA

VCC

FSELC

ADB2

FSELB

FSELD

DENNA

VCC

+24VDC

VCC

VCC

VCC

PGND

+24VDC

VCC

DACLK

DADAT

VCC

KBX0#

KBX1#

KBX2#

KBX3#

KBY0#

KBY1#

KBY2#

KBY3#

KBY4#

KBY5#

+24VDC

VALEN0#

DRIVENB#

+24VDC

VALEN1#

DRIVENB#

+12VDC

VALVECOM

+24VDC

RVD3

RVD2

RVD1

RVD0

PMD0

PMD1

PMD2

PMD3

SMD3

SMD2

SMD1

SMD0

VALVE0#

VCC

SAMLED

SAMHOME#

ROVHOME#

VCC

VALVE1#

VCC

CLICK

VALBOOST

VALEN1#

ADB3

CHOKEWR#

VALEN0#

TMS

XTDOTDI

TCLK

BRD#

PROGPULSE

XCRLD#

PMPCLK

PGMEN#

INT1

XCDONE

XCDATA

DB7DB6DB5DB4DB3DB2DB1DB0

CLK100

CS6#

XCCLK

ADB1BBLE#

PUMP2

PUMP3

PUMP4

PMPCMP12

PMPCMP34

SAMP1

SAMP2

SAMP3

SAMP4

SMPCMP34

ROV1

ROV3

ROV4

ROVCMP12

XOUT1

XOUT2

YIN0

YIN1

YIN3

YIN4

YIN5

1.8MCLK

VCC

VCC

+24VDC

+24VDC

+24VDC

+24VDC

DENNA

DENNA

DENNA

DENNA

ADB4

YIN2

ROVCMP34

PMD1PMD3PMD0PMD2VALVECOMTOUCH

SMD3DOORSW#SMD2SMD1GNDSMD0RVD3SAMLEDRVD2ROVLEDHEATER+ROVHOME#

DOORLEDSAMHOME#HEATER-LAMP-GNDLAMP+

ROVLED2

RVD0 VALVE1#VALVE0#RVD1

R506 221

221R505

VCC

HTRCTL#

TP5

1

Q6

3

4

1

U21

13 12

220PF

C49

220PFC47

C32

220PF

220PF

C48

U12

11

1013

D2

U14

89U14

21

U14 43

U11

1011

U11

1213

U11

89

U11

65

R20

2.49K

Q2

3

4

1

U24

131112

Q3

3

4

1

Q4

3

4

1

C57

0.10.1

C51C35

0.10.1

C139C39

0.10.1

C107C74

0.10.1

C109C113

0.10.1

C94C133

0.10.1

C61

C89

0.10.1

C122C126

0.10.1

C91C37

0.10.1

C77C69

0.10.1

C80C81

0.10.1

C100C99

0.10.1

C98C92

0.1

0.1

C84C78

0.10.1

C79C60

0.10.1

C50C62

0.10.1

C38C40

0.10.1

C88C128

0.10.1

C53C59

0.10.1

C52

C36

0.10.1

C73C76

0.10.1

C72C82

0.10.1

C83C102

0.10.1

C101C93

0.10.1

C95C104

0.10.1

C103C117

0.1

C112

0.1

C118

0.10.1

C105C106

0.10.1

C114C110

0.10.1

C111C123

0.10.1

C134C108

0.10.1

C127

221R237

221R236

C22

0.01

0.01

C23

C25

0.01

C26

0.01

0.1

C21

0.1

C24

0.1C18

0.1C15

C20

0.01

C19

0.01

0.01

C17

C16

0.01

0.1C12

0.1C9

C13

0.01

0.01

C11

C10

0.01

C460.001

U42

21

R85 21.5K

R95 21.5K

R2 221

R90 100K

4.75KR64

R88 100K

4.75KR59

4.75KR60

R89 100K

U4

6

4

9

57

2

11

8

1

3

10

4.75KR55

R248 100

R246 100

R247 100

R245 100

221R235

221R234

221R227

221R226

U42

1213

U42

1011

U42

89

U42

65

R81 100K

4.75K

R107

4.75K

R109

E1

E2

E3

E4

C45

0.1

U221

23

Q1

1

2

3

C6

22

40VR21324K

D3

D4

0.1

C4

HEATER-

+24VDC

0.1

C3C122

40V

U1

4

2

67

1 3

0.1

C140

R54 4.75K

0.01

C14

4.75K

R61

4.75K

R105

R664.75K

U6

6

4

9

57

2

11

8

1

3

10

U7

6

4

9

57

2

11

8

1

3

10

R22

D1

U3

6

4

9

57

2

11

8

1

3

10

U17

2

4 356

8

4.75K

R22

8

R22

9

4.75K R23

8

4.75K

R23

9

4.75K

4.75K

R24

0

4.75K

R24

1

R116

20.5K

R113

20.5K

R112 221

R115 221

U224

56

R27221

221

R26

DOORLED

U191213

U1989

1000

R106

U12

9

814

R53

4.75K

R56

4.75K

U5

6

4

9

57

2

11

8

1

3

10

220PF

C30

C31

220PF

R131 4.75K

R91

100K

R19

220

1.0 OHM 1W

R52

R57

1.0 OHM 1W

1.0 OHM 1W

R62

1.0 OHM 1W

R65

1.0 OHM 1W

R69

R67

1.0 OHM 1WLAMPON#

LAMP-

U22

131112

+24VDC LAMP+

DRIVENB#

U8

6

4

9

57

2

11

8

1

3

10

U42

43

U44

10 11

U44

8 9

U44

6 5

U44

4 3

U4421

4.75K

R108

R128

221

R250

221

R58

1000

R251

1000

U11

21

VCC

DRIVENB#DENNA

4.75K

R1104.75KR132

4.75KR130

C144

+15VDC

R243

1.0 OHM 1W

4.75KR252

4.75KR253

C13

810

0PF

100P

FC

136

C13

710

0PF

100P

FC

135

C13

210

0PF

100P

FC

131

C13

010

0PF

100P

FC

129

C12

510

0PF

C12

410

0PF

U18

100

82

73

74

99

72512423

30

5

604948375946

45474342

10

6

9

7978

34951218351932

16

71

6517

6869

89708890

382062586463

56

55

6186

50

57

292527

33

2144

3931

81

8384

76

53

75

185221336119387

2

98

77

969794

R86

10K

R92

10K

R68

4.75KU14 1011

R84 21.5K

R94 21.5K

18.2KR96

18.2KR82

U135

42

U1311

1013

R93 20K

R87 20K

4.75KR63

U137

61

24.9KR83

24.9KR97

U139

814

R80 100K

R76 20K

R79 20K

R78 21.5K

21.5KR77U12

7

61

24.9KR75

24.9KR74

U125

42

E500

U14

65

4.75K

R111

R25 1000

R24

SMD100

HEATER++24VDC

DRIVENB#

R14

4.75K

R500

4.75K

221R23

U19

21

U19

1011

20.5K

R242

U19

43

J8

40 393738

98 76 54

36 3534 3332 3130

3

2928 2726 2524 2322 2120

2

1918 1716 1514 1312 1110

1

10-5-1998_13:57 3

MTD3055ELG

D

S

7406DOC

LM339AD

MBRS140T

ACT05

OD

ACT05

OD

ACT05

OD

ACT05

OD

ACT05

OD

ACT05

OD

ACT05

OD

MTD3055ELG

D

S74HC02D

MTD3055ELG

D

S

MTD3055ELG

D

S

74LS14D

L6203

BOOT1

GND

IN2

IN1

VREF

ENABLE

VS

SENSE

OUT1

OUT2

BOOT2

74LS14D

74LS14D

74LS14D

74LS14D

74HC02D

VP

0610TDS

G

MB

RS

140T

MB

RS

140T

GNDOUT4

OUT3

OUT1

OUT2IN

LM317LM

L6203

BOOT1

GND

IN2

IN1

VREF

ENABLE

VS

SENSE

OUT1

OUT2

BOOT2

L6203

BOOT1

GND

IN2

IN1

VREF

ENABLE

VS

SENSE

OUT1

OUT2

BOOT2

SMD050 T

MBRS140T

L6203

BOOT1

GND

IN2

IN1

VREF

ENABLE

VS

SENSE

OUT1

OUT2

BOOT2

MAX872CSA

COMP

GNDTEMP

TR

VINVO

74HC02D

74LS14D

74LS14D

LM339AD

L6203

BOOT1

GND

IN2

IN1

VREF

ENABLE

VS

SENSE

OUT1

OUT2

BOOT2

74HC02DL6203

BOOT1

GND

IN2

IN1

VREF

ENABLE

VS

SENSE

OUT1

OUT2

BOOT2

74LS14D

74LS14D

74LS14D

74LS14D

74LS14D

74LS14D

ACT05

OD

1.5uF

XC5202-6PQ100C

AB1

AB2

AB3

AB4

CCLK

CE#

CLICK

CLK100

COMCLK

DB0

DB1

DB2

DB3

DB4

DB5

DB6

DB7

DIN

DONE

DOUT

GCK2

HDC

INIT#

KYBDINT

LDC#

M0

M1

M2

PGMEN#

PMPCLK

PMPCMP12

PMPCMP34

PROG#

PROGPLS

PUMP1

PUMP2

PUMP3

PUMP4

PUMPHOLDRD#

ROV1

ROV2

ROV3

ROV4

ROVCMP12

ROVCMP34

ROVGOOSE#

ROVHOLD

ROVHOME

SAMHOME

SAMP1

SAMP2

SAMP3

SAMP4

SAMPHOLD

SMPCMP12

SMPCMP34

SP1

SP2

SP3

SP4

ANARST

BRSTSYNC

SP7

SP8

TCK

TDI

TDO

TMS

VALVE0#

VALVE1#

VBOOST

WR#

XOUT0

XOUT1

XOUT2

XOUT3

YIN0

YIN1

YIN2

YIN3

YIN4

YIN5

BBLE

ACT05

OD

LM339AD

LM339AD

LM339AD

LM339AD

LM339AD

LM339AD

2

ACT05

OD

T

74LS14D

74LS14D

74LS14D

NOVA Biomedical 200 Prospect Street P.O. Box 9141Waltham, MA 02254-9141

SIZE DRAWING NUMBER

DRAWING TITLE

DO NOT SCALE DRAWING LAST MODIFIED ON: SHEET OF

REV.

E

8 7 6 5 4 3 2 1

8 7 6 5 4 3 2 1

B

C

D

F

G

A

E

H

G

F

E

D

C

B

A

H

HEATER CONTROL POLARITY

VALVE BOOST

BC

Pump Motor

ROV Motor

Sampler Motor

+2.5V REFERENCE

VALVE DRIVERS

INSTALL ONE DIODE ONLY

Local +15VDC

eepot direction

eepot increment

KEYBOARD CONNECTOR

KEYBOARD SCANNER

MOTOR CONTROLLER

SYSTEM CLOCK DIVIDER

POWER GROUND

LAMP DRIVER

FUNCTION SELECT

D

A

UNUSED GATES

RUN: 1700 MA

RUN: 1300 MAPUMP and SAMPLER MOTORS:

ROTARY VALVE MOTOR:

LAMP IS T3/4 BULB 28V .17A

Jumper 1-2: normal sense (original state)

HEATER DRIVER

HOME DETECTORS

Jumper 2-3: inverted sense (requires new sw)

24292-03

Motor & Valve DriversSchematic, Cortex Digital Board

3

DWG #10-1073B

Page S

-5

PN 24292

DWG #10-1074B

Page S

-6

PN 21642

P10

DWG #10-1075A

Page S

-7

D_LP

D_VEEND_YELLOW

VCCUKBY5#KBY3#KBY1#KBX3#KBX2#

D_MD_FLM

D_CPD_DD0D_DD1D_DD2D_DD3

D_ENAVDDD_ENABKL

D_GREEN

KBY4#KBY2#KBY0#KBX1#KBX0#

DACLK DADAT

.15PHL2

J101

SPAREPIN

SPKR-

SPCLK SPDAT

1KR37

VCCU

VCCU VCC

TP1SPKR+

C316V

47C4

Y2.2W

19

213579

11131517

21232527293133353739

46810121416182022242628303234363840

4143

4244

0.1

C7.1UF

C10.1UF

C12.1UF

VCC

C11.1UF .1UF

C8

VCC

DACLK

VEE

DADAT

R181K

C9.1UF

8.45KR16U6

3

5

6

1

7

2

4

8

R1720.5K

R21182K

VADJ

Q2

3

1

2

X9C103

U/D

CS

INC

VSS

VCC

VH

VW

VL

MMBT4403LT1

E

C

B

7-15-1998_12:16 1

6 5 3 24

D

NOVA Biomedical 200 Prospect Street P.O. Box 9141Waltham, MA 02254-9141

SIZE

7 1

2 134567

D

C

B

A

D

C

B

DRAWING NUMBER

DRAWING TITLE

A

8

8DO NOT SCALE DRAWING LAST MODIFIED ON: SHEET OF

REV.

BYPASSING NOTES:

Bypass each chip with one 0.1uF from Vcc to Gnd unless bypass capacitors are explicitly shown.

DISPOFF#

VCC

U8

1213

D_ENAVDD

D_M

D_LPD_CP

D_FLM

VADJVCCS

D_DD0D_DD1D_DD2D_DD3

VEE

Q3VP0610T1

2 3

R10 R9

U5

2 8

1

7

65

3

4

D_VEEN

R110.47

R21K1K

R4

U8

1011

U8

89

1KR5

SPCLK

VCC

0.1C1

4716V

C16

C20.1

MTD2955EQ6

1 3

4

1KR8

1KR33

35V22

C17

MBRS140T3D2

47UHL1

J104

1KR6

74ACT05

OD

DSG

20.5K 324K

MAX775

GNDFB

REF

SHDN

CS

EXT

OUT

VCC

74ACT05

OD

74ACT05

OD

D

SG

14139101112

87654321

note:

pixels swizzeled here

-23.5V @ 10mA

N.C.

LCD DISPLAY

D_YELLOW

KBX0#

KBX1#

KBX2#

KBX3#

KBY0#

G_AN

Y_AN

G_KA

VCC

Y_KA

D_GREEN

KBY5#

KBY4#

KBY3#

KBY2#

KBY1#

R12

R13

U8

43

4.75KR14

U8

65

4.75KR15

1KR36

1KR35

J102

9

8

7

6

5

4

3

2

14

13

12

11

10

1

221

221

74ACT05

OD

74ACT05

OD

KEYPAD

D_ENABKL

R19U8

21 16V47C15

U4

VCC

1KR1

1KR3

1K

R29

MMBT4401LT1Q1

3

1

2

HOT

COM

J1031K74ACT05

OD

E1636C

5 ADJ

4HOT

3COM

1VIN

2

GND

C

B

E

4321

BACKLIGHT

Place power supply filter as close to J101 as possible. CORTEX DISTRIBITION BOARDSCHEMATIC,

247272

DWG #10-1076B

Page S

-8

DWG #10-1077B

PN 24291

Page S

-9

4

6

4

37

10

R86

2.5V

2.5V

~CLEAN0,~CLEAN1,PEAK0,PEAK1

REF1

ISE3

ISE2

ISE0

ISE1

-

+

-

-

G

24292-01

SHEET 1 OF 4

HIGH RES ANALOG BUS

REV.SIZE

NOVA BIOMEDICAL

3

2

1

12

312.4K

R65

U4

24.9K

R51

100PF

C27

24.9K

R69

U9

24.9K

R500

24.9K

R197

100PF

C26

100PF

C34

BAR

A

A

A

LM324

MPX4115AP

654

NCNCNC

5V REF.

BAROMETER

SEE SHEET 3

CLEAN/PEAK CTRL BUS

APC CTRL BUS

NEW GEN. SENSOR BD.

12

3U21

R80

10.0KC130

0.01

R178

24.9MEG

LMC6484

89

10

R81

21.5K

1000PF

C54

R82

24.9K

12

13

U21

C129

0.01

A

LMC6484

81

5

63

2

U23

R17724.9MEG

20K

R47

R79

10.0K

R97R59

10.0K

R97A

10.0K

I/V1-

DG444DY

RG

AD622

+REF

RG

A

8

4

6

86

231

76

5

14

U21

U21

U16

51.1K

.01C55

R83

44.2K

10.0K

R180

D19

LT1004

D21

LT1004

.1R83A

A

LMC6484

LMC6484

DG444DY

DS

cw

LT1004CD-1

LT1004CD-1

1.235V

1.235V

1413

12

R196

51.1K

.22C57

U17

R84100K

A

LMC6484

1

23

89

10

119

U5

U17R89100K

D13 R90

10.0K

100K4700pf

C71

Q4

2N7002PEAK1

PK1

DS+5V

LMC6484

MM

BD

914L

T1

S D

G

2

3

21

C127

R88

100KR85

U17

100KR92

4700PFC72

A

A

.22UFCTHRU2

LMC6484

MM

BD

914L

T1

S

1

3

768

5

67

100K

R91

10.0K

U17

U5

C123

D14

Q5

2N7002PEAK0

PK0

+5V

LMC6484

DG444DY

DS

.22UFCTHRU2

D

R172

.22C58

R87

51.1K

CLEAN0

I/V0+

R74

51.1K

A

cw

+5V

81

5

63

2

10.0K

51.1K

R75

10.0KR14

R4

10.0KU15

I/V0-

A

A

RG

AD622

+

-

REF

RG

86

8

4

6

151416

76

5U13

U16

D11

LT1004

D9

LT1004

R42

44.2K

.01

C23

R52

C20

.1

1KR42A

LMC6484

DG444DY

DS

LT1004CD-1

LT1004CD-1

1.235V

1.235V

2

1

3

81

5 63

2

2.49K

U14

10.0K

R18

D10

A

RG

AD622

+REF

RG

BAV99LT1

10

8

11

9

7

1

40

39

38

J8

J8

J8

J8

J8

J8

J8

1K

R48

R60

100K

GNDCHS

35

33

36

34

32

5

3

1

24

23

22J7

J7

J7

J7

J8

J8

J8

J8

J8

J8

J8

J8

J8

J8

1KR50

A

GNDCHS

6

4

2

24

23

22

1J6

J6

J6

J6

J8

J8

J8

R62

100K

A/D+,A/D-,F+,F-,E+,E-,D+,D-,C-,ISE[0:5],BAR,2.27V,HX,CAIR,HCTD,I/V[0:2]+,I/V[0:2]-,PK[0:2],TMPR,REF[1:2],VMN,HTR-,BTMP,MH

GNDCHS

ISE5

ISE4

REF23

21

1413

12

53

6

2

7

11

9

4J6

J6

J6

J6

J6

J6J6

U1R167

200MEG

1KR3

C12

1000PF

R166

200MEGR2

A

LMC6484

LMC6484

2

1

3

23 1

231

23 1

231

D1

CLAMP2

CLAMP1

D2

D4

D5

1000PF

C2

1000PF

C3

R11

4.75K

R10

4.75K

D3

A

SSTPAD5

-5V

+5V

A

SSTPAD5

SSTPAD5

SSTPAD5

BAV99LT1

COMMON MODE VOLTAGE CLAMP

ISE 5 = PCO2+

10

98

15

8

10

16

13

12

14

J6

J6

J6

J6

J6

J6

J6

J6

CLAMP2

1K

R164

200MEG

1K

R1

R165

U1

C8

1000PF

U1

A

A

LMC6484

LMC6484ISE 4 = PH

REF2

REF1

810

12

1314

5

67

20

17

19

5

18

21

3

6

4

2J7

J7

J7

J7

J6

J6

J7

J7

J7

J6

CLAMP1

200MEG

U1

U2

U2

1000PF

C14

1K

R9

R171

200MEG

R169

200MEG

A

A

LMC6484

LMC6484

ISE 3 = NA

ISE 2

REF1

76

5

9

25

23

26

24

22

20

18

21J7

J7

J7

J8

J8

J8

J8

J8

1K

R6

1K

R7

1K

R55

R174

200MEG

1K

R54

U2

1000PF

C13

1000PF

C24

A

A

LMC6484ISE 1

12

3

30

28

31

29

27

1TP1

J8

J8

J8

J8

J8

1K

R57

R175200MEG

1K

R56

U2

C25

1000PF

R170

24.9MEG

C124 100K

LMC6484

A

12

3

10

8

11

7

9J7

J7

J7

J7

J7

R19100K

1K

R8

2200PFR17

10.0K

R16

1000PF

C18

R33

24.9K R34

21.5K

U13

A

LMC6484

10

98

1413

12

13

15

12

16

14J7

J7

J7

J7

J7

1K

R5

R15

100K

U13

U13

R20

100K

R168

24.9MEG

C118

2200PF

A

LMC6484

LMC6484

PO2-

PO2+

I/V 0

ISE 0

GUARD

I/V 1

81

53

2

31D20

R98

51.1K

U22

10.0K

R61 20K

R49

C56

2.49K

R94

1K

R98A

10.0K

CLEAN1

I/V1+

+5V

BAS21LT1

RG

AD622REF

RG

A

A

Not used in CORTEX

HIGH RES ANALOG BUS

LOW RES ANALOG BUS

E

DWG #10-1078B Page S-10

1

9

21

3

U5

A+

U10

100KR160

2.5V

2.5V,AIR[1:4],PHTR,ACH6,MHTRTEMP

HSO.3

PWM0BUF

PWM0BUF,HSO.3HX

A

DG444DY

DS

A

A

DS

G

24292-02

GEN CTRL BUS

SHEET 2 OF 4

23

~CLEAN2,PEAK2

VMN

D40

+5V

A-100KR159

4.99K

R162

6.81KR146

C95

.22

R125

51.1K

R163

10.0K

10.0KR96A

A

A

V+

-5V

A

MM

BD

914L

T1

1413

12

1011

U16Q6

2N7002

U30

C944700pf

100K

R123

+5V

DG444DY

LMC6484

100KR161

2.5V

A/D+,A/D-,F+,F-,E+,E-,D+,D-,C-,ISE[0:5],BAR,2.27V,HX,CAIR,HCTD,I/V[0:2]+,I/V[0:2]-,PK[0:2],TMPR,REF[1:2],VMN,BTMP,MH

23

1

12

1314

89

10 U30

U33

R126

100K

R127

10.0K

100K

R124

D37

C126

100K

R158

PEAK2

ACH6

PK2

LMC6484S D

LMC6484

MM

BD

914L

T1

.22UFCTHRU2

SEE SHEET 3

CLEAN/PEAK CTRL BUS

HI RES ANALOG BUS

REV.SIZE

NOVA BIOMEDICAL

NEW GEN. SENSOR BD.

LOW RES ANALOG BUS

HCT/AIR DETECTOR #3

38J2

R100

10.0

200

R101

R130

1K

AIR1

A

39J2

U6

10.0

R115

AIR2

13

7

10.0K

R102

1.25V

U10

AIR3

AIR4

A

LMC6484

49.9

R1041.0UFC64A

1.0UF

C64B

1.0UFC64C

C64D1.0UF

C64E1.0UF

A

2

13

17

1

HCT-E

J4 15

J3

J6

J4 17

C37

.22

D8

C624700pf

2.49K

R103

4700pf

C67

R11310.0K

C39

.22

AD2E

AD4E

AD1EC38

.22

-5V

+5V

BAV99LT1

A

AIR EXCITATION MONITOR

AIR EXCITATION

HCT EXCITATION

6

5

108

1

5

67

19

2.5V

J6

J418 HCT-M

.22

C70

1K

R117

10PF

C66

R112

475K

U10

C17

10PF

475K

R31

200

R110

10.0K

R114

AD4M

.15UH

L1

1K

R32

A

LMC6484

BAV99LT1 LMC6484

LMC6484

IND11

HCT/AIR DETECTOR

AIR DETECTOR #4

12

1314

92

3

2

J4 16

J3

R116

1K.22

C69

.15UH

L2

D7

C65

10PF

475K

R111

10PF

C63R107

475K

U10

R108

1K.22

C61AD1M

AD2M

.15UH

L7

IND11

LMC6484

IND11

AIR DETECTOR #1

AIR DETECTOR #2

2

J4 11

J4 13

J4 14

J4 12

C59

4700pf

R109

10.0K

HCT1

HCTD

FLOW CELL TEMP

5

4

3J3

J3

J3

PHTR

MHTRTEMP

2.5V

MH

TMPR

BTMP

CAIR

JUMPER

T

3

21

2

13

1 2 PAKPRESENTU25

R133

30.1K

D26

475K

R143

U33

1000PFC104

9.53MEG

R144

100

R105 4.75K

R106 A

+5V

+5V

74HC14

BAV99LT1

LMC6484

FLUID_TEMP.

5

67

R136

18.2K

R137

18.2K

R138

18.2K

R139

18.2K

475K

R142

R141

475K

1000PFC105

9.53MEG

R145

U33

1000PF

C101R140

9.53MEG

+5V

LMC6484

2

13

2

13

6

9

8

7

10J3

LAMPHI

J4 23

J4 25

HEATERHIJ3

J3

J4 24

HEATERLO

J3

J4 26

LAMPLO

J3

D39

D38

A

A

BAV99LT1

BAV99LT1

[PASS THRU SIGNALS ONLY]

2.5V

HEATER TEMP

THERMISTOR RETURN

500mV/C

500mV/C

E

DWG #10-1079B Page S-11

6

8

4

5

2

5

123

645

15141312111097

13

52

4

1516

14

76

5

1110

98

1312

56

34

5 6

3 4

2

132

13

31

23

1

2

2

13

2

13

2

13

32

1

2

13

2

13

1 2

1918171615141312

1112

56789

43

10

98

5

67

3

21

116152

141339

1011127654

8

22

14

15

16

5

7

11

20

4

3

21

19

23

17

9

25

18

6

10

12

13

22

21

24

23

20

19

10

8

6

4

9

7

5

3

1

RESET

.01UF

C22

J5A

J5A

J5A

J5A

J5A

J5A

J5A

J5A

J5A

2.27V

D-C-BTMPMHVMNREF2REF1I/V2-I/V1-I/V0-HCTDCAIRHX

LEDDRV

TXD

PREHTR

PREHTR

SYNC

SYNC

DRDY

WRL

RFS

LEDSEL4

BLKHTR

A-

J2

J2

J2

J2

J1

J1

J1

J1

J1

SIGOUTJ1

A+

Q3

J4 7

J4 9

LEDSEL7

LEDSEL3LEDSEL2LEDSEL1LEDSEL0

LEDSEL6

LEDSEL7

LEDSEL3

LEDSEL6LEDSEL5

RXD

J4 10

J4 8

LED1-A

LED2-A

FAILIN

INTIN

RESETIN

SRXD

STXD

SYNCIN

HTRCTL

PAKPRESENT

J2

J1

J1

J1

J1

J1

TFS

J1

J1

J1

J1

J1

J1

J1

J1

J1

E-

SMDXTAL

FPX-20MHZ

X1

J2

MOD0MOD1

MOD2

MOD2MOD3

MOD1MOD0

MODON

MODCTRL

MOD2

MODON

MOD3MOD2MOD1MOD0

U26 U32

U28

U30

U30

U33

U31

C99.1

C98

.1

C109

22PF

C108

22PFU27

.1C97

R132

100K

C93.1

C96.1

C106

.1

C103

.1

C85

.1

C84

.110.0K

R118

C91

.01

R121

4.75K

R155

4.75K 4.75K

R156

4.75K

R157

4.75K

R151

4.75K

R152

4.75K

R153

4.75K

R154

R148

4.75K 4.75K

R147

4.75K

R149

4.75K

R150

U24

D33

D28

47UF

C134

4.75K

R129

4.75K

R128

4.75K

R120

4.75K

R122

D2747UF

C135 47UF

C133

C137

.1

C136

.1

R193

4.75K

D31

4.75K

R189

4.75K

R182

4.75K

R184

4.75K

R191

D34D35

D29D32

D30D36

U24

U25

U25

U24

U24

U24

U24

100R192

U4

U5

49.9K

R73

9.09K

R67

1KR68

100K

R70

Q2 Q1

R195

475

R35

1K

475

R27

D6

24.9R28

1KR36

150

R37

U3

C33

.1

10.0

R6310.0

R3810.0

R64

U29

10.0

R24

10.0

R23C36

470PF

R72

200K

MCLKOUT

0.0R

135

R131121K

C100.1

100K

R199

R198 2.49k

10.0R22

10.0K

R119

LEDCOM

AGND

AIR4

HX

SO2DET2

AIR3

HCTD

SO2DET1

.1

C138

U34

+5VF

CLEAN0

CLEAN2CLEAN1

PEAK0PEAK1PEAK2

PEAK[0:2],~CLEAN[0:2]

PMUXENA0NMUXENA0

PMUX0PMUX1PMUX2PMUX3NMUX0NMUX1NMUX2NMUX3

PMUXENA1PMUXENA2PMUXENA3NMUXENA1

PMUXENA0

PMUX2PMUX1

NMUX1 NMUX2 NMUX3

PMUX1PMUX0

PMUX1

PMUX0PMUX1PMUX2PMUX3NMUX0NMUX1NMUX2NMUX3

PMUX0

PMUX0

PMUX3NMUXENA0NMUX3NMUX2NMUX1

PMUXENA1

PMUXENA2NMUXENA1

PMUXENA3

NMUX0

NMUX0

PMUX3PMUX2PMUXENA1PMUX2

2.5VACH6

MHTRTEMP

PHTR

2.5V

2.5V

AIR3AIR4

AIR1AIR2

PHTR,ACH[0:2],ACH6,2.5V,AIR[1:4],MHTRTEMP

SO2DET2

ISE0

D+E+

TMPRPK2PK1PK0I/V2+I/V1+I/V0+HCTDCAIRHX

BARISE5ISE4ISE3ISE2ISE1

A/D-A/D+

SO2DET1C90

.01

100

R188

.15UH

L3

.15UH

L6

L5

.15UH

J5A

1K

R21

A/DSYNC

D38A

100R194

100R185

100R183

100R187

R186100

FAIL

FAIL

C29

4700pf

SPARE2

HSO.3

P2.3

SPARE1

PWM0

MIX/VAL

HSO.5

PWM0BUF

HSO.5

P2.3

PWM0

EXTINT

R134

0.0

R190100

R200681

A

A

A

V+

A

VDD

VDD

GND

VDD

GND

GND

VDDVDD

VDD

A

A

VDD

A

A

VDD

GND

GND

GND

A

A

GND

A

GND

MMBT4403LT1

-5V

+5V

-5V

+5V

-5V

+5V

+5V

A

VDD

GND

GND

VDD

A

+5V

VD

D

A

-5V

DG406DN

8910112625242322212019

76

1716151418

12

2754 1

28

2

3

NC

2

01

3456789ABCDEF

A0A1A2A3EN

NC

13GNDNC-V

+V

OUT

DG406DN

8910112625242322212019

76

1716151418

12

2754 1

28

2

3

NC

2

01

3456789ABCDEF

A0A1A2A3EN

NC

13GNDNC-V

+V

OUT

DG508ADY

2

01

34567

A0A1A2EN

V+GNDOUT

V-

LMC6484

LMC6484

LMC64841112 13

1415161718

222324

212019

21

54

78910

3

6

DVdd

/DRDY

IoutREF OUT

REFIN(-)Vbias

AD7710SCLKMCLK IN

A0

MODE

AIN1(-)AIN2(+)AIN2(-)VssAVdd

REFIN(+)

AGND/TFS/RFS

SDATA

AIN1(+)

DGND

/SYNC

MCLK OUT

74HC574

1Q2Q3Q4Q5Q6Q7Q8Q8D

7D6D5D4D3D2D1D

CLKOC

74HC14

BAV99LT1

MMBZ5V6A

MMBZ5V6A

BAV99LT1

BAV99LT1BAV99LT1

BAV99LT1BAV99LT1

BAV99LT1BAV99LT1

74HC14

74HC14

74HC14

74HC14

74HC14

74HC14

74HC14

LM324

DG444DY

DS

MMBT4403LT1 MMBT4403LT1

LED1

MIC6211B

VP-

VP+

ABC

G1G2AG2B

74HC138

Y0Y1Y2Y3Y4Y5Y6Y7

A

+5V

A

A

GND

-5V

+5V

2 /EA

16 /RESET

6 ACH05 ACH1

7 ACH2

4 ACH3

11 ACH410 ACH5

8 ACH69 ACH7

62ADV

12 ANGND

64BUS

65

CLKOUT

15 EXTINT

24 HSI.025 HSI.1

34HSO.2

35HSO.326 HSO.427 HSO.5

63INST

3 NMI

19 P1.020 P1.121 P1.2

44P2.3

42P2.4

33

P2.6

38P2.7

60P3.059P3.158P3.257P3.356P3.455P3.554P3.653P3.752P4.051P4.150P4.249P4.348P4.447P4.546P4.645P4.7

23 PWM2 39PWM022 PWM1

61RD

43READY

17 RXD18 TXD

37VPP

13 VREF

41WRH40WRL

67XTAL1

66XTAL2

29

HSO.1

32P1.7

28HSO.0

31

P1.6

30P1.5

8XC196KC

IND11

IND11

IND11

26

28

18

17

16

15J2

J2

J2

J2

J2

LEDSEL2

LEDSEL1

LEDSEL0

J2

J2

ALE

PWM2

.1C109A

.01C109B

L8

.15UH

+5V

37

35

36

33

34J2

J2

J2

J2

J2

D+

E+

D-

E-

C-

13

11

914

12

10

32

31J2

J2

J2

J2

J2J2

J2

J2

PMUXENA3

PMUXENA2

A/D-

A/D+GND

1

8

7

6

5

4

3

2J2

J2

J2

J2

J2

J2

J2

PMUX0

PMUX1

PMUX2

PMUX3

NMUX0

NMUX1

NMUX2

NMUX3

NMUXENA1

PMUX[0:3],NMUX[0:3],NMUXENA[0:1],PMUXENA[0:3]

J2

1

10

8

6

4

2

9

7

3

28

32

34

30

26

24

22

20

18

16

14

1211

13

15

17

19

21

33

31

29

27

25

23

40

27

29

30

1TP2

BTMP

PK1

VMNTMPR

MH

I/V2+

J2

I/V1-

I/V0+I/V0-

I/V1+

PHTR

BAR

HX

CAIR

PK2

I/V2-

J2

J2

J2

MHTRTEMP

ACH6

PGND

2.5V

J5J5J5J5J5J5

J5J5J5J5J5J5

J5

J5

J5

J5

J5

J5

J5

J5

J5

J5

J5

100

R181

A+A-

AGND

AIR2

AIR1

J5

J5J5J5J5

J5

J5

J5

J5

J5PK0

ISE5

ISE3

ISE1

REF1AGND

ISE0

ISE2

ISE4

REF2

J5

GND

GND

IND11

LT10

09S

8

A

A

TO SHEETS 1 & 2

GEN CTRL BUS

MOD CTRL BUS

TO SHEETS 1 & 2

ANALOG MUX BUS

HI RES ANALOG BUS

REV.SIZE

NOVA BIOMEDICAL

NEW GEN. SENSOR BD.

ANALOG SUPPLIES

/LEDSEL

LED2

LED1

SHEET 3 OF 4

LOGIC SUPPLY

+5V MONITOR

24292-03

U34-14,36,68

U34-1

LOW RES ANALOG BUS

CLEAN/PEAK CTRL BUS

NOT INSTALLED IN CORTEX

1 2

E

DWG #10-1080B Page S-12

3

21

89 8

98

10

12

1314

1516

14

109

11

1516

14

109

11

78

6

78

6

21

3

21

3

1311

12

108

9

56

4

23

1

123

645

15141312111097

1413

128

9

10

13 1211 10

2

SO2DET-C

MCLKOUT

J4 4

J4 1

J4 5

J4 6

MEAS

SO2-2

J4

J4 3

MEAS

MEAS

IN1

BKGND_IN

SAM2

SAM1

MODCTRL

MOD3MOD2MOD0

B2B1

MEASSAM2SAM1

SAM1B1

SAM2B2

IN2

IN1 B1B2

BKGND_IN

IN2

SO2DET2

.1

C28

.1

C68

.1

C92 C44

.1 .1

C87

.1

C86 C83

.1

U25 U25

.1

C5A C45

.1 .1

C40 C19

.1 .1

C52 C89

.1 .1

C16 C48

.1 .1

C102 C107

.1.1

C73

.1

C5

.1

C77 C81

.1 .1

C31

.1

C42 C75

.1

C79

.1

C7

.1

U4 U4

U12 U11

U11

U11

U11

.1

C35 C46

.1.1

C41C21

.1 .1

C53 C88

.1

C49

.1 .1

C60

.1

C74

.1

C10

.1

C78 C82

.1 .1

C32 C43

.1

C76

.1

C80

.1

C1

.1

10.0K

R41

U7

U8

U8

U7

U7

CTHRU

C119

.033UF

C120

.033UFCTHRU

CTHRU.033UF

C122

C121

.033UFCTHRU

U7

U8

C114

.22UFCTHRU2

U8

C117

.22UF

CTHRU2

C116

.22UF

CTHRU2

475K

R66

C115

.22UF

CTHRU2

10.0K

R29

R26

10.0K

R25

36.5K

U6

U6

U25

SO2-1R40

475K

J1

SCOPETRIGGER

CLKOUTR300

100

.1

C30

200K

R30

C113

SO2DET1

U6

CO2INPUT

A

-5V

+5V

A

A

A

A

A

A

GND

VDD

V+

P

A

+5V

-5V

74HC14 74HC14

LM324 LM324

ABC

G1G2AG2B74HC138

Y0Y1Y2Y3Y4Y5Y6Y7

74HC08

74HC08

74HC08

74HC08

DG444DY

D S

DG444DY

DS

DG444DY

DS

DG444DY

DS

DG444DY

DS

DG444DY

DS

DG444DY

DS

DG444DY

DS

LMC6484

LMC6484

74HC14

A

.22UF

CT

HR

U2

LMC6484

MAKE CONNECTION TO ALL SPARES WITH ETCH

REV.SIZE

NOVA BIOMEDICAL

NEW GEN. SENSOR BD.

SO THAT ECCH CUTS AND JUMPERS CAN USE THESE DEVICES FOR ECO'S

SHEET 4 OF 4

24292-04E

DWG #10-1081B Page S-13

PN 23897

DWG #10-1082B

Page S

-14

PN 23899

ITEM P/N DESCRIPTION65 22745 Flowcell Lamp5, 40, 58, 38 24628 Kit Sampler Gears31 24622 Optosensor8 24621 Motor, Sampler52 20861 Fluid Fountain

DWG #10-1083B

Page S

-15

PN 23898

ITEM P/N DESCRIPTION2 24627 Motor Rotary Valve3 24626 Kit Rotary Valve Gear4 24613 Rep. Opto Sensor Rotary Valve5 24624 Rep. Kit Ceramic Rotary Valve

DWG #10-1084B

Page S

-16

SPECIFICATIONSHEET

TITLE SIZEA

DWG.NO. REV

REV ECO# APPROVEDDRP/CO-DRP/DATE

EFFECTIVEECO MGR/DATE REV ECO# APPROVED

DRP/CO-DRP/DATEEFFECTIVE

ECO MGR/DATE

P/N SHT 1 OF 1

For

Doc

.Cnt

rl U

seO

nly

Dis

k N

o.

PRINTER INTERFACE IF3001 SEIKO

24614

CN01: Head flexible printed circuit connector

CN02: Motor control connector

CN03: Input/Output control connector

SW01: Function switch

a, b: Position selection notches for mounting to the LTP3245

c: Holes for mounting to the LTP3245

d: Two holes for mounting to the outer case (or frame)

NOTE: CN01: Lift white plastic part to release print head flat cable

PT300P01

CN02

To PrinterMotor

To Print Head

CN01

SW01

PT300GA1

CN03

d

dc

a b

To ControlBoard J9

DWG #10-1085B

Page S

-17

TB1

V2ADJ.

F1T3.5A/250V~

3

3

1

1

V1ADJ.

1

1

TB2

J2

NC+S-SPF

8.

.

000" (203.2mm)

8.

.

50" (215.9mm)

SPECIFICATIONSHEET

TITLE SIZEA

DWG.NO. REV

REV ECO# APPROVEDDRP/CO-DRP/DATE

EFFECTIVEECO MGR/DATE REV ECO# APPROVED

DRP/CO-DRP/DATEEFFECTIVE

ECO MGR/DATE

P/N SHT 1 OF 1

For

Doc

.Cnt

rl U

seO

nly

Dis

k N

o.

POWER SUPPLY UNIVERSAL INPUT5+/-12 24Vo

24720

INPUT: J1 AMP P.C.B. HEADER P/N 640445-5 PIN 1) AC LINE TB1 PIN 2) N/C PIN 1) AC LINE PIN 3) AC NEUTRAL PIN 2) AC NEUTRAL PIN 4) N/C PIN 3) AC GROUND PIN 5) AC GROUND

SIGNALS: J2 AMP P.C.B. HEADER P/N 640456-4 PIN 1) N/C PIN 3) -SENSE PIN 2) +SENSE PIN 4) POWER FAIL

OUTPUT: J3 AMP P.C.B. HEADER P/N 1-640445-6 PIN 1) OUTPUT #4 PIN 14) OUTPUT #1 PIN 2) OUTPUT #4 PIN 15) OUTPUT #1 PIN 3) OUTPUT #3 PIN 16) OUTPUT #1 PIN 4) OUTPUT #3 PIN 5) OUTPUT #2 TB2 PIN 6) OUTPUT #2 PIN 1) OUTPUT #4 PIN 7) COMMON PIN 2) OUTPUT #3 PIN 8) COMMON PIN 3) OUTPUT #2 PIN 9) COMMON PIN 4) COMMON PIN 10) COMMON PIN 5) COMMON PIN 11) COMMON PIN 6) COMMON PIN 12) COMMON PIN 7) OUTPUT #1 PIN 13) COMMON PIN 8) OUTPUT #1

NOTE: 5A MAXIMUM RECOMMENDED CURRENT PER CONNECTOR PIN

WEIGHT: 2.3 LBS. MAX.

Output Current Total Model Output Output Minimum Maximum Maximum Peak Noise Regulation (Note A) Voltage (Note C) (Note D) P-P (Note B)

GPx130D 1 +5V 3A 16A 20A 25A 50mV 2% 2 +24V 0A 3.5A 3.5A 5A 240mV 2% 3 -12V 0A 1.2A 1.2A 1.5A 120mV 3% 4 +12V 0A 1.2A 1.2A 1.5A 120mV 3%

DWG #10-1086A

Page S

-18