not only a peripheral stent available in renal sizes

24
W e m a k e i d e a s c o m e a l i v e 001IHP050501 For internal use only Not only a peripheral stent Not only a peripheral stent available in Renal sizes... available in Renal sizes...

Upload: lucky

Post on 11-Jan-2016

24 views

Category:

Documents


2 download

DESCRIPTION

Not only a peripheral stent available in Renal sizes. Specification SDS. Catheter design Guide wire compatibility Usable shaft length Length of RX section Balloon material Prox./distal Shaft diameter Crossing profile Introducer compatibility Guiding Catheter comp. Nominal pressure - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Not only a peripheral stentNot only a peripheral stentavailable in Renal sizes... available in Renal sizes...

Page 2: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Specification SDS

Catheter designCatheter design

Guide wire Guide wire compatibilitycompatibility

Usable shaft lengthUsable shaft length

Length of RX section Length of RX section

Balloon material Balloon material

Prox./distal Shaft Prox./distal Shaft diameter diameter

Crossing profile Crossing profile

Introducer compatibilityIntroducer compatibility

Guiding Catheter comp.Guiding Catheter comp.

Nominal pressureNominal pressure

RBPRBP

Rapid Exchange with NiTi-Rapid Exchange with NiTi-wire reinforced RX-section, wire reinforced RX-section, hypotube shafthypotube shaft

0.014“0.014“

80, 145 cm 80, 145 cm

15 cm15 cm

FLEXITEC™ LPFLEXITEC™ LP

2,3F / 3,5F 2,3F / 3,5F

< 0.065”< 0.065”

5F5F

6F6F

8 bar8 bar

14/15 bar14/15 bar

Page 3: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Specification Stent

Stent designStent design

Range of diameterRange of diameter

Maximum Maximum expansionexpansion

Range of lengthRange of length

Stent material Stent material

Strut thicknessStrut thickness

Strut widthStrut width

Metal to Artery Metal to Artery RatioRatio

Closed cell, slotted tubeClosed cell, slotted tube

Ø 4.0, 5.0, 5.5, 6.0, 6.5, 7.0 Ø 4.0, 5.0, 5.5, 6.0, 6.5, 7.0 mmmm

7.5 mm7.5 mm

10, 15, 20, 24 mm10, 15, 20, 24 mm

Stainless steelStainless steel

165 µ165 µmm

110 110 µµmm

16% @ Ø 6 mm 16% @ Ø 6 mm

Page 4: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Hippocampus in DetailHippocampus in Detail

Page 5: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

How does it work? 1/6

Insert 0.014” guidewire through a guiding catheter or long introducer sheath into the renal artery.

Page 6: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

How does it work? 2/6

Advance slowly the Hippocampus stent system through GC or IS into the renal ostium. Since the system shows a

progressive flexibility, coming from the long tip, followed by the long balloon-cone, the guidewire will not be straightened and possibly flipped out of the ostium!

Non-Flip-TipNon-Flip-Tip

Page 7: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

How does it work? 3/6

Advance slowly the Hippocampus stent system through GC/IS into the renal ostium. As soon as the balloon

segment with the crimped stent is advanced through the curve of the GC/IS, the long tip is already inserted in the ostium so the position cannot be lost again.

Progressive FlexibilityProgressive FlexibilitySuperior Vessel Accessibility Superior Vessel Accessibility

Page 8: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

How does it work? 4/6

Position the stent inside the lesion, so that the proximal 1-2 mm of the stent will overlap into the Aorta. Since it will not shorten, this position can be easily maintained during inflation.

Zero Stent-Shortening Zero Stent-Shortening for Precise Positioningfor Precise Positioning

Page 9: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

How does it work? 5/6

Inflate the stent carefully, with slow increasing pressure until a proper dilation of the lesion is achieved. High Pressure BalloonHigh Pressure Balloonfor Optimal Resultsfor Optimal Results

Page 10: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

How does it work? 6/6

Flaring the proximal stent section into the ostium is optional, because this might facilitate any following re-intervention of this renal artery.Ostial Scaffolding CapabilityOstial Scaffolding Capability

Facilitating Re-interventionsFacilitating Re-interventions

Page 11: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Non-Flip-Tip

Long Non-Flip-TipNon-Flip-Tip with progressive flexibility and minimal entry profile Tip length 7mm

6 folded high pressure balloon

Tight FIX secure crimping

Page 12: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Progressive Flexibility

Progressive Flexibility on distal balloon segment starting from the long Non-Flip-Tip followed by the Long Cones provides a superior vessel accessibility.

NO balloon overh

ang!

NO balloon overh

ang!

It is t

he long sm

all angle

cone th

at looks l

ike a balloon

overhang!

30°30°

Page 13: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Stent System Dimensions in Detail

Tip length

7 mm

Stent Ø(mm)

Cone length(mm)

4 5,35 7,2

5,5 8,16 9,0

6,5 10,07 10,9

Stent Ø(mm)

Cone length(mm)

4 5,35 7,2

5,5 8,16 9,0

6,5 10,07 10,9

Rapid Exchange section

15 cm

Stent length

10, 15, 20, 24 mm

Distal shaft Ø

3,5F

Ste

nt I.D

. Ø

Page 14: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Crossing Profiles

Ø 4.0, 5.5, 6.0 mm 1.45 - 1.48mm (0.0571 – 0.0583”) 0.0577”

Ø 5.0, 6.5, 7.0 mm 1.54 - 1.57mm (0.0606 – 0.0618”) 0.0612”

Page 15: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Zero Stent-Shortening

Zero Stent-Shortening and precise alignment of the Stent with the proximal marker band enables the operator to precisely position the stent inside the ostium. Proximal Stent

Edge/ Balloon Marker

Page 16: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Enhanced Pushability

For the normal RX- sections, there will be an increased friction, when passing through the curve of GC

The Ni-Ti wire is straightening the RX- section in order to transfer the maximum of push through the curve.

Push Loss

Push Loss

Tapered NiTi-wire inside the 15 cm long rapid exchange section reinforces the distal catheter section and clearly enhances the pushability.

Page 17: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Ostial Scaffolding Capability

Ostial section Distal section

Flaring the proximal stent section into the ostium is optional, because this might facilitate any following re-intervention of this renal artery.

Page 18: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Everything is Possible

Shaft length of 80 and 145 cm enable you to perform radial approaches, which have significant benefits for the patient.

Page 19: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

SDS Compliance

4,00 5,00 5,50 6,00 6,50 7,00

6 3,76 4,77 5,22 5,64 6,13 6,70

7 3,88 4,88 5,36 5,82 6,29 6,85

8 4,00 5,00 5,50 6,00 6,50 7,00

9 4,12 5,11 5,65 6,19 6,68 7,26

10 4,20 5,19 5,73 6,30 6,77 7,38

11 4,26 5,28 5,81 6,40 6,91 7,48

12 4,32 5,35 5,89 6,47 6,99 7,59

13 4,37 5,39 5,95 6,54 7,05 7,68

14 4,43 5,45 6,01 6,63 7,12 7,75

15 4,48 5,51 6,10

Stent Inner Diameter (mm)

Compliance Chart

Pressure (bar)

Nominal, in vitro results; ±5% of balloon diameter. Results do not take into account lesion resistance. Confirm balloon sizing angiographically. Do not exceed RBP

Nominal Pressure

Rated Burst Pressure

Controlled compliance of the FLEXITEC™ HS with high strength balloon material, which ensures a proper Stent deployment including a large working range at high pressures of 14-15 bar.

Page 20: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Size Mix

Stent Ø - L 10 15 20 24

4.0 5.0 5.5 6.0 6.5 7.0

Usable shaft length: 80 and 145 cm

Page 21: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Frequently Asked Questions

Why such a long balloon overhang? It’s no balloon overhang, it is a long cones with a small opening

angle for a progressive flexibility from the long Non-Flip-Tip over the Long Cones till the crimped stent segment.

In the common ostial stenting procedures it is NOT acceptable to engage the GC in the ostium, or even advance it through the ostium. NO – TOUCH - PROCEDURE The GC has to remain in front of the renal artery inside the aorta.

Why is the rapid exchange section only 15 cm long? The SDS is dedicated to renal application, where you only have to

reach the renal ostia. That shorter the RX-section is, that higher is the pushability.

What for a NiTi-wire inside the RX-section? To reduce the friction inside the GC/IS and consequentially improve

the pushability.

Page 22: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Frequently Asked Questions (cont.)

Are there any clinical data available? Not yet for the Hippocampus. But we are about to start a

multicenter (4) registry “PRECISION” (Renal artery angioplasty in Patients with hypertension and renal insufficiency using Hippocapmus renal stent) with the PI Dr. Thomas Zeller from Germany enrolling 50 patients.

The ostial scaffolding capability is not needed! Maybe that this operator does not care about any potential re-

intervention. But without flaring the ostial section of the stent, the engagement of the ostium with a guidewire can become a very tricky procedure.

Why is the guiding catheter compatibility 6F and the introducer sheath comp. 5F? A guiding catheter has always the outer diameter indicated. In

case a long introducer sheath will be used instead of a GC, the inner diameter of the IS is sufficient with 5F.

Page 23: Not only a peripheral stent available in Renal sizes

W e

m

a k

e i d

e a

s c

o m

e a l i v

e

00

1IH

P0

50

50

1

For

inte

rnal u

se o

nly

Positioning

The Hippocampus is a Renal Stent System of the latest generation.

The design and architecture of the stent AND the delivery system is absolutely dedicated to the renal artery approach or similar anatomies (e.g. mesenteric artery). It is NOT a normal peripheral stent, with a size mix suitable for renal

arteries!

Due to the ease of insertion of the stent system into the renal artery, the product has the potential to reduce the procedural risks

decrease the duration of the procedure

improve the technical success

Hippocampus should be sold on premium price!Hippocampus should be sold on premium price!

Page 24: Not only a peripheral stent available in Renal sizes

Select Properly Where to Go, and Seize the Best Opportunity!