noninvasive monitoring in the intensive care unit iskander al-githmi, md,frcsc, fccp assistant...

31
Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Post on 22-Dec-2015

216 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Noninvasive Monitoring in The Intensive Care Unit

Iskander Al-Githmi, MD,FRCSC, FCCP

Assistant Professor of Surgery

King Abdulaziz University

Page 2: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Learning Objectives

• Know the different noninvasive monitoring techniques commonly use in ICU.

• Know the advantages and limitation of the different noninvasive monitoring methods.

• Know the different technologies used in noninvasive monitoring.

• Correlate findings observed during noninvasive monitoring with the patient’s changing physiology.

Page 3: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Goals of Monitoring

• Assessment of vital organs function• Detection of early life-threatening complications• Determine the needs for intervention.

Page 4: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Respiratory Monitoring

• Pulse Oximetry

• Capnography

Page 5: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Pulse Oximetry

• Measures four types of hemoglobin: deoxyHb, oxyHb, carboxyHb and metHb.

• Estimate functional hemoglobin saturation.

Page 6: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Principles of Pulse Oximetry

Page 7: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Principles of Pulse Oximetry con’t

• Spectorphotometry• Discriminate between oxyHb and deoxyHb by

the difference in light absorption at 660nm and 940nm

• Estimate heart rate by measuring cyclic changes in light transmission.

Page 8: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Absorption Spectra

Page 9: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Pitfalls and Limitations

• Margin of error is +/- 4% at Sao2 95%• Margin of error is upto 15% at SaO2 <70%• Does not measure arterial oxygen (PaO2)• Wide Pao2 level 60-160mmHg at O2-Dissociation curve• Is not a substitute for arterial blood gas• SPO2= O2Hb+COHb• SPO2>90%even with COHb 70%• Methylene blue underestimate the saturation• Low perfusion e.g. low cardiac output• Extreme anemia

Page 10: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Oxygen - dissociation Curve

Page 11: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Capnometry

• Clinical practice:• Confirmation of endotracheal tube placement• Estimation of arterial CO2 with End-Tidal CO2• Monitoring the integrity of patient-ventilator

system• Provide a noninvasive means of facilitating

weaning from mechanical ventilation

Page 12: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

The principles of capnometry

• Main stream and side stream analyzer• Infrared spectrometry• CO2 absorption takes place at 4.3 micm

Page 13: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Normal capnogram

• Phase I: inspiratory baseline• Phase II: expiratory upstroke• Phase III: expiratory plateau

Page 14: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

PaCo2 – PETCo2 gradient

• PETCo2 is usually 1-3mmHg lower than PaCo2

• The difference between PETCo2 and PaCo2 is caused by V/Q mismatch

• PETCo2 does not reflect PaCo2 in the presence of V/Q mismatch

• PaCo2 – PETCo2 gradient is usually < 5mmHg

• The gradient increased when cardiac output decreased

Page 15: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Limitations

• Alteration of dead space ventilation• Breathing patterns• Patient stability• Tidal volume• V/Q ratio

Page 16: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Capnogram abnormalitiesSudden loss of ETCO2

Page 17: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Sudden loss ETCO2

• Airway disconnection• Dislodgment of ET tube• Total obstruction of ET tube

Possible causes:

Page 18: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Gradual decrease in ETCO2 to non zero level

Page 19: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Gradual decrease in ETCO2 to non zero level

• Leak in the circuit• Partial disconnection from ventilator

Possible causes:

Page 20: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Gradual decrease in ETCO2 level

Page 21: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Gradual decrease in ET CO2

• Hypothermia• Pulmonary embolism• Cardiopulmonary bypass

Possible causes:

Page 22: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Rise in baseline ETCO2

Page 23: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Rise in baseline ETCO2

• Defective exhalation valve • Rebreathing of previously exhaled CO2

Possible causes:

Page 24: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Quiz

Page 25: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

The capnograph also has a use in correctly and exactly identifying end expiration during the analysis of haemodynamic wave forms. This would classically be seen when measuring the Pulmonary Artery Wedge Pressure at end expiration.The point of end expiration is seen at:

a. Just after the peak on the capnography waveform.

b. Just before the peak on the capnograph waveform.

c. In the exact middle of the capnography waveform.

d. In the middle of the capnography "trough".

Question 1

Page 26: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Capnography has a place in confirming endotracheal tube placement but should be used in conjunction with other, simpler techniques. What is the most correct sequence of events in confirming endotracheal tube placement?

a. Visualise the tube with laryngoscopy / observe and

auscultate the chest and epigastrium / check the capnogram.

b. Visualise the tube with laryngoscopy / check the capnogram / observe and auscultate the chest and epigastrium

Question 2

Page 27: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

The Capnograph may also be used in the ICU to avoid constant arterial sampling for ABGs and also to monitor patients with brains injuries who need to be kept with a normal PaCO2.This is possible because of a correlation between the maximum partial pressure of CO2 (PetCO2) at end expiration and the arterial CO2 levels. With normal ventilation and perfusion the gradient between PetCO2 and PaCO2 should be between 1-5 mm Hg.The following statement is most true:

a. Capnography is unreliable in the case of massive pulmonary embolism.

b. Capnography is unreliable in the case of ARDS.

c. Once a PetCO2 to PaCO2 gradient is established, capnography completely obviates the need for ABGs.

d. Once a PetCO2 to PaCO2 gradient is established capnography can be reliable used.

Question 3

Page 28: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Cardiac arrest can produce a flat capnograph trace, due to low pulmonary blood flow, even if the endotracheal tube is correctly placed above the carina.

• True

• False

Question 4

Page 29: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Severe bronchospasm can produce a flat capnograph trace even if the endotracheal tube is correctly placed.

a. True

b. False

Question 5

Page 30: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Question 6

12 year-old was rescued by a firefighter form a smoke field room of a burning 12-story apartment building. The following procedures are appropriate, except which of the following:

a. a careful inspection of the upper airway passage to detect signs of inhalation injury

b. A chest x-ray on admission to look for inhalation injury

c. A 24 hour hospital observation

d. A normal oxygen hemoglobin saturation by pulse oximetry to exclude the presence of CO poisoning

Page 31: Noninvasive Monitoring in The Intensive Care Unit Iskander Al-Githmi, MD,FRCSC, FCCP Assistant Professor of Surgery King Abdulaziz University

Question 7

The following statement about pulse oximetry are true except which of the following:

a. Modern pulse oximetry uses two wavelength of light, red and infrared to discriminate between oxygenated and deoxygenated blood

b. Pulse oximetry is not affected by low cardiac output state

c. Different form of dyshemoglobinemia can affect the accuracy of oxyhemoglobin measurement by pulse oximetry

d. The pulse oximetry degrades with oxygen saturation <65%