Non-randomized Medical Device Clinical Studies:

A Regulatory Perspective

Sep. 16, 2005

Lilly Yue, Ph.D.*CDRH, FDA, Rockville MD 20850

*No official support or endorsement by the Food and Drug Administration of this presentation is intended or should be inferred.

Page 2: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

2

Acknowledgements

Thanks to my colleagues in the Cardiovascular and Ophthalmic Devices Branch, CDRH, FDA, for their help with this presentation.

Page 3: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

3

Outline

1. Types of non-randomized studies in medical device

2. Why non-randomized?

3. Major concerns with non-randomized studies

4. Conclusions

Page 4: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

4

Type of Non-randomized Study

• Concurrent 2-arm non-randomized study

• One-arm study:• With comparison to a historical control, where

patient-level data of historical control is available and used in treatment comparison;

Pseudo 2-arm comparative study

• With comparison to a fixed target value obtained from multiple historical trials;

OPC: Objective performance criterion

Page 5: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

5

Why Non-Randomized?

• RCT is sometimes not ethical or practical.

• Sample size determination based on one-sample hypothesis test smaller sample size?

• May save time and money

• “Least Burdensome”?

• To keep the rapid pace of new technology development.

Page 6: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

6

Example 1 --- Coronary artery bare mental stent

• In 1994, two superiority RCTs: compared novel Palmaz-Schatz (P-S) stent with standard balloon angioplasty

• Subsequently, randomized non-inferiority trials: compared several new stents with the P-S stent

• Over 30 different coronary stents developed for use in USA or Europe over the past 10 years

• Design changes: minor modification resulting in small or local effect on patient outcome.

• Approx. 1.5 millions patients per year world-wide undergoing the catheter-based coronary treatment

• Stent life cycle < 2 yrs, required for randomized trial

• Some non-randomized studies were conducted to keep the rapid pace of new stent development.

Page 7: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

7

Major Concerns with Non–randomized Studies

• In 2-arm comparative study:– Selection bias and comparability of treatmeent groups

• In a pseudo 2-arm comparative study– Is historical active control good? – Bio-creep– Are test and historical control comparable in pat.

population?

• In one-arm study with OPC– Share all the problems associated with historical

controls – Problems with the validity of its determination– Problems with the appropriateness of its use

Page 8: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

8

Comparability of Treatment Groups

• In RCT, expect that all patient covariates, measured or unmeasured, are balanced between the two treatment groups.

• So, the two treatment groups are comparable and observed treatment difference is an unbiased estimate of true treatment difference.

• None of advantages provided by randomized trials is available in non-randomized studies.

• A potential problem: Two treatment groups were not comparable before the start of

treatment, i.e., not comparable due to imbalanced covariates between two treatment groups.

• So, direct treatment comparisons are invalid.

Page 9: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

9

Traditional Adjustments for Covariates

• Three common methods of adjusting for confounding covariates:

– Matching

– Subclassification (stratification)

– Regression (Covariate) adjustment

Page 10: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

10

Propensity Score Methodology

• Replace the collection of confounding covariates with one scalar function of these covariates: the propensity score.

Age Gender

Duration…….

1 composite covariate:

Propensity Score

Balancing score

Page 11: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

11

Propensity Score Methodology (cont.)

• Propensity score (PS): conditional prob. of receiving treatment A rather than treatment B, given a collection of observed covariates.

• Purpose: simultaneously balance many covariates in the two treatment groups and thus reduce the bias.

• PS construction: multiple logistic regression model based on patient data of all measured covariates and actual treatment received.

Page 12: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

12

• Properties of propensity scores

– A group of patients with the same propensity score are equally likely to have been assigned to trt A.

– Within a group of patients with the same propensity score, e.g., 0.7, some patients actually got trt A and some got trt B, just as they had been randomly allocated to whichever trt they actually received.

“Randomized After the Fact”PS=0.7

Trt A Trt B

Page 13: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

13

– When the propensity scores are balanced across two treatment groups, the distribution of all the covariates are balanced in expectation across the two groups.

– Use the propensity scores as a diagnostic tool to measure treatment group comparability.

– If the two treatment groups overlap well enough in terms of the propensity scores, we compare the two treatment groups adjusting for the PS.

• Compare treatments adjusting for propensity score

– Matching – Subclassification (stratification)– Regression (Covariate) adjustment

Page 14: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

14

•Stratification

– All patients are sorted by propensity scores.

– Divide into equal-sized subclasses.

– Compare two trts within each subclass, as in a randomized trial; then estimate overall trt effect as weighted average.

– It is intended to use all patients.

– But, if trial size is small, some subclass may contain patients from only one treatment group.

PS 1 2 ……. 5

Page 15: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

15

Example 2

• New vs. control in a non-randomized study

• Primary endpoint: MACE incidence rate at 6-month after treatment

• Non-inferiority margin: 7%, in this study

• Sample size: new: 290, control: 560

• 14 covariates were considered.

Page 16: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

16

Table. Distribution of patients at five strata

Subclass Control New Total 1 142 28 170

2 127 43 170

3 122 48 170

4 119 51 170

5 50 120 170

Total 560 290 850

Page 17: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

17

Estimated Propensity Scores

N(new)=560, N(control)=290

Control New

0.0

0.2

0.4

0.6

0.8

1.0

Est

imat

ed P

rop

ensi

ty S

core

Page 18: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

18

Baseline covariate balance checking before and after PS stratification adjustment

Mean p-value New Control Before After --------------------------------------------------------------------------------------

Mi 0.25 0.40 <.0001 0.4645

Diab 0.28 0.21 0.0421 0.8608

CCS 2.41 2.75 0.0003 0.3096

Lesleng 11.02 12.16 <.0001 0.5008

Preref 3.00 3.08 0.0202 0.2556

Presten 62.75 66.81 <.0001 0.4053

Page 19: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

19

Diagnostic check for covariate balance: Percentage of patients with prior Mi

0

10

20

30

40

50

60

70

80

1st 2nd 3rdsubcla

4th 5th BeforeAdj

NewCtl

Page 20: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

20

Example 3

• Non-concurrent, two-arm, multi-center study

• Control: Medical treatment without device,

N=65, hospital record collection

• Treatment: Device A, N = 130• Primary effectiveness endpoint: Treatment success

• Hypothesis testing: superiority in success rate• 20 imbalanced clinically important baseline

covariates, e.g., prior cardiac surgery• 22% patients with missing baseline covariate values

Page 21: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

21

Enrollment Time

0

5

10

15

20

25

1992 1993 1994 1995 1996 1997 1998 1999 2000 2001 2002

Ctl Trt

Page 22: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

22

• Two treatment groups are not comparable

– Imbalance in multiple baseline covariates– Imbalance in the time of enrollment

• So, any direct treatment comparisons on the effectiveness endpoint are inappropriate.

• And, p-values from direct treatment comparisons are un-interpretable.

• What about treatment comparisons adjusting for the imbalanced covariates?

– Traditional covariate analysis – Propensity score analysis

Page 23: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

23

• Performed propensity score (PS) analysis

• Handed missing values – MI: generate multiple data sets for PS analysis – Generate one data set: generalized PS analysis– Others

• Included all statistically significant and/or clinically important baseline covariates in PS modeling.

• Checked comparability of two treatment groups through estimated propensity score distributions.

• Found that the two treatment groups did not overlap well.

Page 24: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

24

Estimated Propensity Scores (with time)

Ctl Trt

0.0

0.2

0.4

0.6

0.8

1.0Es

timat

ed P

rope

nsity

Sco

re

Page 25: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

25

Estimated Propensity Scores (w/o time)

Ctl Trt

0.0

0.2

0.4

0.6

0.8

1.0E

stim

ated

Pro

pens

ity S

core

Page 26: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

26

Patients in Propensity Score Quintiles

1 2 3 4 5 Total

Ctl 38 18 8 1 0 65

(w/time) 58% 28% 12% 2% 0%

Trt 1 21 31 38 39 130

1% 16% 24% 29% 30%

Page 27: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

27

Patients in Propensity Score Quintiles

1 2 3 4 5 Total

Ctl 29 24 8 4 0 65

(w/o time) 45% 37% 12% 6% 0%

Trt 10 14 32 35 39 130

8% 11% 24% 27% 30%

Page 28: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

28

• Conclusion: – The two treatment groups did not overlap enough to

allow a sensible treatment comparison.

– So, any treatment comparisons adjusting for imbalanced covariates are problematic.

• Question: Given that the two treatment groups are not

comparable, what can we do NOW?

Page 29: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

29

Risks and Dangers of Non-randomized Studies

• A study with an historical control may result in much riskier and potentially more burdensome than a RCT.

• It may be impossible to predict in advance whether the patient population with the new treatment is comparable to the population for the historical control.

• The sponsor must have legal access to the historical data at the patient level and all the right baseline covariates need to have been measured in both groups.

Page 30: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

30

One-arm study with OPC

• OPC: Objective Performance Criterion• Introduced to the FDA approx. 10 yrs ago, in the

evaluation of prosthetic heart valves• Compared a new heart valve against a fixed number,

e.g., a complication rate, obtained from multiple approved heart valve trials by outside experts

• Data and guidance for the OPC in public domain• Now, used for some coronary artery stents, e.g.,

Ho: 6 mo. MACE rate ≥ point estimate + delta

Ha: 6 mo. MACE rate < point estimate + delta

Page 31: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

31

• Delta: often a clinical call, no FDA guidance

• Point estimate: often estimated mean of outcome

but currently no universally accepted way

• OPC: point estimate, by some people,

point estimate + delta, by some others

• “One-sample OPC equivalence”, stated by some

Question: equivalent to what? To a fixed number!?

• “One-sample OPC equivalence” -- inappropriate claim!

Page 32: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

32

Problems with OPC:• Limited good historical data available for the development of

OPC

• Disregarded variability associated with the estimate in historical studies

• Bio-creep problem

• Time sensitive

• Patient population sensitive

• Who is responsible for developing the OPC for a particular device?

• Who is responsible for checking if the OPC developed is appropriate?

• Who is responsible for updating an existing OPC?

Page 33: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

33

Example 4

• Primary effectiveness endpoint: acute procedure success

• Evaluated for the entire SVT patient populationSVT = (AVNRT AVRT AF)

• OPC = 85%

• Hypotheses:

• Study results: N = 200, # of successes =164 Observed success rate = 82% (< 85%) C.I.: (76%, 87%)

• OPC was not met!

%85 : Ha. %85 :Ho pvsp

Page 34: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

Post-hoc Subgroup Analysis and Claim

Acute Success (OPC:85%)AVNRT AVRT AF Total

N 128 60 12 200

# Success 118 40 6 164

Success Rate

92% 67% 50% 82%

95% C.I. (86%,

96%)

Page 35: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

35

• One of major problems with the post-hoc analysis and claim:

– The OPC, 85%, was developed for the entire SVT population, not for a particular patient subpopulation

– In fact, an OPC for AVNRT, if exists, should be much higher than 85%

Page 36: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

36

Example 5. Weighted OPC

• Primary endpoint: one-year adverse event rate

• Patient population: Co-morbid group & Anatomic group

• Expected event rate: Co-morbid: 14%, Anatomic: 11%

• A common delta, 3%

• Individual OPC: Co-morbid: 17%, Anatomic: 14%

• Weighted OPC:

• n1 and n2 were # of patients actually enrolled

• Hypotheses:

1 1 1 2/( )w n n n 2 2 1 2/( )w n n n

1 2OPC 17% 14%w w w

o aH : OPC vs. H : OPCw wp p

Page 37: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

37

Problems:

• N = n1 + n2 was fixed in protocol, but n1 and n2 and hence w1 and w2 were not.

• So, the weighted OPC is a random variable, e.g.,– If w1= w2=50%, then OPCW = 15.5%

– IF w1= =70%, w2= 30%, then OPCW = 16%

• The setting of hypotheses is inappropriate.

• The weighted OPC leads to the study subject to questionable manipulation.

a 1 2 a 1 a 2H : 17% 14% H : 17% AND H : 14%p w w p p

Page 38: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

38

Study Result:

• Enrolled patients:

Co-morbid Anatomic Total 264 36 300 (88%) (12%) (100%)

• Post-hoc determined weighted OPC = 88% * 17% +12% * 14% =16.6%. • Observed event rate: Co-morbid: 8% < Anatomic:16%

( Expected event rate: Co-morbid: 14% > Anatomic:11% )

• Overall observed event rate:

88%*8% +12%*16% = 9% Reject Ho

Page 39: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

39

• What if enrolled: Co-morbid: 12% , Anatomic: 88%

• Then, post-hoc determined weighted OPC =

12% *17% + 88% *14% =14.4%

• Overall observed event rate:

12% * 8% + 88% * 16% = 15% (>14.4%)

Can’t reject Ho!

• The weighted OPC leads to the study subject to questionable manipulation.

Page 40: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

40

• Test statistic:

• If treat n1 and n2 as fixed, then the calculated C.I. would be narrower than it should be.

• What if pre-specify w1 and w2 in protocol?

• Should comply with protocol

• If the w1 and w2 are not achieved in the actual enrollment, then a protocol deviation has been committed.

1 1 2 2ˆ ˆT w p w p

Page 41: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

41

Conclusions

• Select comparable control prospectively!• Bio-creep problem should be avoided.• OPC should be determined by sufficient solid scientific evidence.• Variability associated with the point estimate from historical

studies should be incorporated in the determination of OPC.• OPC should be appropriately adjusted for different patient

populations, and different indications for use.• OPC would need to be updated constantly.• RCT is still the gold standard for clinical studies. • RCT should be preserved for new technology!

Page 42: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

42

References

• Rubin, DB, Estimating casual effects from large data sets using propensity scores. Ann Intern Med 1997; 127:757-763

• Rosenbaum, PR, Rubin DB, Reducing bias in observational studies using subclassification on the propensity score. JASA 1984; 79:516-524

• D’agostino, RB, Jr., Propensity score methods for bias reduction in the comparison of a treatment to a non-randomized control group, Statistics in medicine, 1998,17:2265-2281

Page 43: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

43

References

• Blackstone, EH, Comparing apples and oranges, J. Thoracic and Cardiovascular Surgery, January 2002; 1:8-15

• Grunkemeier, GL and et al, Propensity score analysis of stroke after off-pump coronary artery bypass grafting, Ann Thorac Surg 2002; 74:301-305

• Wolfgang, C. and et al, Comparing mortality of elder patients on hemodialysis versus peritoneal dialysis: A propensity score approach, J. Am Soc Nephrol 2002; 13:2353-2362

Page 44: Non-randomized Medical Device Clinical Studies: A Regulatory Perspective Sep. 16, 2005 Lilly Yue, Ph.D.* CDRH, FDA, Rockville MD 20850 * No official support

44

Thanks!