non pharmachological treatment of hf

Upload: cristiano-spadaccio

Post on 06-Apr-2018

221 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/3/2019 Non Pharmachological Treatment of HF

    1/31

    Non pharmachological treatment of

    heart failure

  • 8/3/2019 Non Pharmachological Treatment of HF

    2/31

    Heart Failure

    Venezia, 5-6 November 2009

    Jessup M, Brozena S. N Engl J Med 2003;348:2007-18.

    The clinical syndrome of heart failure is the final pathway for

    myriad diseases that affect the heart.

  • 8/3/2019 Non Pharmachological Treatment of HF

    3/31

    Heart Failure

    Venezia, 5-6 November 2009

    Therapeutical targets

    Jessup M, Brozena S. N Engl J Med 2003;348:2007-18

  • 8/3/2019 Non Pharmachological Treatment of HF

    4/31

    Non pharmachological approaches

    Venezia, 5-6 November 2009

  • 8/3/2019 Non Pharmachological Treatment of HF

    5/31

    Cardiac Resynchronization Therapy

    Venezia, 5-6 November 2009

    Elevated ventricular end-diastolic pressure

    can lead to wall stretch causing atrialfibrillation or delays in ventricular conduction

    and bundlebranch

    block.

    Left bundle-branch block is a significant

    predictor of sudden death and a common

    finding in patients with myocardial failure.Left ventricular contraction is

    dyssynchronous, with the interventricular

    septum contracting before the left ventricular

    free wall.

    Delayed opening and closure of the mitral

    and aortic valves, and abnormal diastolicfunction

    Dyssynchronous contraction is mechanically

    inefficient, leading to decreases in the left

    ventricular ejection fraction (LVEF) and

    cardiac output.

  • 8/3/2019 Non Pharmachological Treatment of HF

    6/31

    Cardiac Resynchronization Therapy

    Venezia, 5-6 November 2009

    INDICATIONS

    Dilated cardiomyopathy (ischemic or nonischemic)

    LVEF of 35 percent or less

    QRS interval of at least 120 msec

    NYHA class III or IV heart failure despite

    optimal medical therapy

    most candidates for biventricular pacing are also

    candidates for a defibrillator.

    GREY ZONE AND UNCERTAINITIES

    Mild heart failure (NYHA class II)

    QRS interval as best criterion for benefit from CRT.

    Possible role of echocardiographic evidence of

    ventricular dyssynchrony Atrial fibrillation

    Ablation of the atrioventricular node performed to

    ensure complete control of ventricular activation.

    REFERENCES

    Gregoratos G, Abrams J, Epstein AE, et al. ACC/AHA/NASPE2002 guideline update for implantation of cardiac pacemakers

    and antiarrhythmia devices: summary article: a report of the

    American College of Cardiology/American Heart Association

    Task Force on Practice Guidelines (ACC/AHA/NASPE

    Committee to Update the 1998 Pacemaker Guidelines).

    Circulation 2002;106:2145-61.

    7DQJ$65RVV+6LPSVRQ&6HWDO&DQDGLDQ

    &DUGLRYDVFXODU6RFLHW\&DQDGLDQ+HDUW5K\WKP6RFLHW\

    SRVLWLRQSDSHURQLPSODQWDEOHFDUGLRYHUWHUGHILEULOODWRU

    XVHLQ&DQDGD&DQ-&DUGLRO6XSSO$

  • 8/3/2019 Non Pharmachological Treatment of HF

    7/31

    Cardiac Resynchronization Therapy

    Venezia, 5-6 November 2009

    COMPLICATIONS

    Coronary sinus dissection or perforation

    Hemopericardium and tamponade

    Pneumothorax

    Complete heart block

    Asystole

    Dislodgment of the left ventricular pacing lead

    PITFALLS

    Lead failure

    Device infection in PMK pocket

    Atrial dysrhythmias, with device tracking of the atrial

    rate

    inability to implant the left ventricular because ofunfavorable coronary venous anatomy.

    Proximity to the left phrenic nerve, and the resulting

    uncomfortable diaphragmatic stimulation during pacing

    External electromagnetic fields

    REFERENCES

    McAlister FA, Ezekowitz JA, Wiebe N,et al. Systematic

    review: cardiac resynchronization in patients with

    symptomatic heart failure. Ann Intern Med 2004;

    141:381-90.

  • 8/3/2019 Non Pharmachological Treatment of HF

    8/31

    Cardiac Resynchronization Therapy

    Venezia, 5-6 November 2009

    Most candidates for biventricular pacing are alsocandidates for a defibrillator.

  • 8/3/2019 Non Pharmachological Treatment of HF

    9/31

    Non pharmachological approaches

    Venezia, 5-6 November 2009

  • 8/3/2019 Non Pharmachological Treatment of HF

    10/31

    Surgical Strategies

    Venezia, 5-6 November 2009

    Heart failure subsequent to ischemic or non ischaemic

    disease often leads to functional mitral regurgitation dueto LV dilation worsening prognosis

    Mitral valve repair or replacement (MVR)could have a

    dramatic effect on cardiac function unloding the heart

    and improving ejection fraction and optimizing oxygen

    consumption

    MVR procedures can be combined with surgical

    revascularization with improved resultsTechniques for minimally invasive MVR have been

    developed

  • 8/3/2019 Non Pharmachological Treatment of HF

    11/31

    Surgical Strategies

    Venezia, 5-6 November 2009

    Surgical ventricular reconstruction

    DOR procedure allows ventricular reshaping

    in case of ventricular dilation

    Several techniques have been developed for

    ventricular reconstruction

    The procedure can be associated to MVR in

    order to improve valve function as well

    Surgical ventricular reconstruction is

    performed in selected patients,usually in

    conjunction with CABG

  • 8/3/2019 Non Pharmachological Treatment of HF

    12/31

    Surgical Strategies

    Venezia, 5-6 November 2009

    UP-TO-DATESTICH 6XUJLFDO7UHDWPHQWIRU,VFKHPLF+HDUW)DLOXUH67,&+WULDO&OLQLFDO7ULDOVJRYQXPEHU1&7LQWKH+\SRWKHVLV

    VXEVWXG\FRPSDUHG&$%*DORQHZLWKWKHFRPELQHGSURFHGXUHRI

    &$%*ZLWKVXUJLFDOYHQWULFXODUUHFRQVWUXFWLRQ

    %RWK&$%*DORQHDQGWKHFRPELQHGSURFHGXUHZHUHHTXDOO\

    VXFFHVVIXOLQLPSURYLQJWKHSRVWRSHUDWLYH&&6DQJLQDFODVVDQG

    1

  • 8/3/2019 Non Pharmachological Treatment of HF

    13/31

    Non pharmachological approaches

    Venezia, 5-6 November 2009

  • 8/3/2019 Non Pharmachological Treatment of HF

    14/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    Ventricular assist devices are mechanical pumps that take over the function of the

    damaged ventricle and restore normal hemodynamics and end-organ blood flow.

    These devices are useful in two groups of patients.

    1. Patients who require ventricular assistance to allow the heart to rest and recover its

    function.

    obtain complete drainage of the ventricle

    unload the ventricle

    Diminish myocardial work

    maximize subendocardial perfusion.

    2. Patients with myocardial infarction,

    acute myocarditis,or end-stage heart diseasewho are not expected to recover adequate

    cardiac function and who require

    mechanical support as a bridge to transplantation

  • 8/3/2019 Non Pharmachological Treatment of HF

    15/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    Available ventricular assist device

  • 8/3/2019 Non Pharmachological Treatment of HF

    16/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    INDICATIONS FLOW CHART

  • 8/3/2019 Non Pharmachological Treatment of HF

    17/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    ValvularHeartDisease

    Patients with preexisting mitral stenosis or aortic

    regurgitation may require correction of the

    valvulopathy before implantation of the device

    CoronaryArteryDiseasePatients with inoperable coronary artery disease

    sometimes continue to have angina without adverse

    hemodynamic effects while being supported by a left ventricular assist device. However, right

    ventricular ischemia and myocardial injury soon after implantation of the device can cause

    right-sided heart failure, resulting in decreased flow to the left ventricular assist device.

    GENERAL PRINCIPLES FORVAD APPLICATION

    CARDIAC FACTORS

    Arrhythmias

    Atrial and ventricular arrhythmias continue to occur after implantation of the device.Atrial fibrillation hinders right ventricular filling but is reasonably well tolerated in recipients of left

    ventricular assist devices.CongenitalDefects

    Intracardiac septal defects should be repaired at the time of implantation of the device to avoid the right-

    to-left shunt and subsequent oxygen desaturation that would be created by the sudden reduction in left

    filling pressures

  • 8/3/2019 Non Pharmachological Treatment of HF

    18/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    GENERAL PRINCIPLES FORVAD APPLICATION

    EXTRACARDIAC FACTORS

    Functional compromise of end-organs

    Irreversible major neurologic deficits

    Severe obstructive or restrictive pulmonary disease is a contraindication, because

    these patients often have oxygen desaturation in the perioperative period that can

    result in hypoxic pulmonary vasoconstriction and right-sided heart failure.

    Hepatic insufficiency and coagulopathy needs to be corrected aggressively

    Dependence on hemodialysis is a contraindication but moderate renal insufficiency

    does not preclude VAD implantation

    Arterial and aortoiliac disease of the lower extremities complicates transfemoral

    institution of cardiopulmonary bypass, which is sometimes necessary at the time of

    removal of the device.

    Patients with type O blood and for patients with a positive panel-reactiveantibody titer

    for whom prospective crossmatching is required may be candidate to BRIDGING

    Venezia, 5-6 November 2009

  • 8/3/2019 Non Pharmachological Treatment of HF

    19/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    Available ventricular assist device

  • 8/3/2019 Non Pharmachological Treatment of HF

    20/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    Extracorporeal Pulsatile

    Abiomed BVS 5000This is a short-term mechanical system that can provide left, right, or biventricular support

    for patients whose hearts have failed but have the potential for recovery.

    The device can also be used as a bridge to definitive therapy.

    The pump houses an Angioflex membrane

    and two proprietary tri-leaflet valves.

    The AB5000 can provide flow rates of up to 6 lit

    per min

    The console recognizes which blood pump

    is connected and automatically ensures

    proper system function.

    Although the left and right sides of the pump

    work independently of each other, the console

    automatically adjusts heart rates and

    flows accordingly.

  • 8/3/2019 Non Pharmachological Treatment of HF

    21/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    Extracorporeal Pulsatile

    ThoratecOne or two Thoratec VADs can be used to provide left, right, or biventricular support.

    The blood pump is positioned outside the body (extracorporeally) and connected to tubes

    (cannulas) inserted into the heart.

    The pump has a rigid plastic case that contains a flexible pumping sac.

    Blood is ejected from the pump when the pumping sac is

    compressed by air from the external control console.

    Within the inflow and outflow conduits, mechanicalvalves control the direction of blood flow.

    The Thoratec VAD has a stroke volume of 65 milliliters

    It can be operated at up to 100 beats per minute

    (blood flow rates of up to 7 liters per minute)

    3 different modes of operation:

    asynchronous mode (pumping occurs at a preset rate)

    synchronous mode (pumping is synchronizedwith the patient's heart rate)

    volume mode (pumping is adjusted according

    to the left ventricular filling volume).

  • 8/3/2019 Non Pharmachological Treatment of HF

    22/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    TandemHeartTandemHeart Percutaneous Ventricular Assist Device (pVAD)

    can be inserted either by cardiovascular surgeons in the operating room or by cardiologists in the

    cardiac catheterization laboratory.

    The TandemHeart has been used in postcardiotomy cardiogenic shock patients (those who have

    developed heart failure as a result of heart surgery or a heart attack) and as a bridge to a definitive

    therapy.

    The TandemHeart pVAD provides short-term support from a few hours up to 14 days, giving the heart

    time to strengthen and potentially regain native function.

    The TandemHeart pVAD is a continuous-flow centrifugal assist device placed

    outside the body (extracorporeally).

    Cannulas are inserted percutaneously through the femoral vein and advancedacross the intraatrial septum into the left atrium.

    The pump withdraws oxygenated blood from the left atrium,

    propels it by a magnetically driven, six-bladed impeller through

    the outflow port, and returns it to one or both femoral arteriesvia arterial cannulas.

    The pump weighs 8 ounces

    The pump is capable of delivering blood flow up to 5.0 liters/minute.

    Extracorporeal Non Pulsatile

  • 8/3/2019 Non Pharmachological Treatment of HF

    23/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    Implantable Pulsatile

    Thoratec HeartMate IP LVAS (implantable pneumatic)

    The pneumatic (air-driven) LVAS is a titanium alloy pump that weighs 570 grams and

    consists of a blood chamber, an air chamber, a drive-line, and inflow and outflow

    conduits.

    Each conduit is a titanium cage

    that contains a 25-mm porcine (pig)valve within a woven Dacrongraft.

    A flexible polyurethane diaphragm

    separates the blood chamber

    and the air chamber.

    Textured surfaces within the

    blood chamber promote the

    development of a cellular lining.

    IPLVAS can provide flow rates of

    up to 12 liters per minute.

    The system can operate in three modes: automatic mode, fixed-rate

    and external (synchronous) mode.

  • 8/3/2019 Non Pharmachological Treatment of HF

    24/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    Implantable Non Pulsatile

    Thoratec HeartMate II LVAS

    High-speed, axial flow, rotary blood pump producing non pulsatile action

    12 ounces (about 375 grams) 1.5 inches (4 cm) in diameter and

    2.5 inches (6 cm) long, so it is suitable for a wider range of patients,

    including small adults and children.

    The internal pump surfaces are asmooth, polished titanium.

    A magnet is within the rotor

    The pump speed can vary from

    6,000 rpm to 15,000 rpm

    (10 liters per minute).

    The pump can run in two modes:In fixed-speed mode: constant speed,

    which can be adjusted via the

    system monitor.

    In auto-speed mode: the pump speed varies in response to different levels of patient or

    cardiac activity.

  • 8/3/2019 Non Pharmachological Treatment of HF

    25/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    Implantable Non Pulsatile

    Jarvik 2000 Heart

    The device is a valveless, electrically powered axial flow pump that fits

    directly into the left ventricle and continuously pushes oxygen-rich

    blood throughout the body

    The Jarvik 2000 Heart is an axial flow blood pump that uses electrical power to rotate a vaned

    impeller

    The device is 2.5 cm wide, 5.5 cm long,and weighs 85 grams.

    The impeller is a

    neodymium-iron-boron magnet, which

    is housed inside a welded titanium

    shell.

    A small cable, which exits the body

    through the abdominal wall,delivers power to the impeller.

    All of the blood-contacting surfaces

    are made of highly polished titanium.

    The normal operating range for the control system is 8,000 to 12,000 revolutions per minute,

    which will generate an average pump flow rate of 5 liters per minute.

  • 8/3/2019 Non Pharmachological Treatment of HF

    26/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    Patients who develop heart failureafter heart surgery

    and who have not responded to standard

    medical therapy.

    The system is designed to provide

    immediate support and restore

    hemodynamic stability for a period of up

    to 7 days.The microaxial pump gives up to 4.5 liters per minute at a speed of 33,000 rpm.

    Miniaturized impeller pump located

    within a catheter.

    IMPELLA

    Venezia, 5-6 November 2009

  • 8/3/2019 Non Pharmachological Treatment of HF

    27/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    AbioCor implantable replacement heart is the first completely self-

    contained total artificial heart

    Criteria

    end-stage heart failure.life-expectancy of less than 30 days.

    not eligible for a natural heart transplant.

    no other viable treatment options.

    Patients are not tethered to a large, air-pumping console nor do they have wires or

    tubes piercing their skin.

    Venezia, 5-6 November 2009

  • 8/3/2019 Non Pharmachological Treatment of HF

    28/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    COMPLICATIONS

    BleedingPerioperative hemorrhage due to hepatic dysfunction, poor nutritional status, and

    antibiotic therapy; cardiopulmonary-bypassinduced thrombocytopenia and platelet

    dysfunctionA recent study documented the presence of sustained thrombin generation and

    fibrinolysis in patients with low-grade disseminated intravascular coagulopathy.

    Absorption of tissue macrophages that express tissue factor by the surface of thedevice may be the trigger that generates this systemic procoagulant state. This may

    account for reduced risk of clinically important thromboembolism.Thromboembolismandairembolism

    Rightsidedheartfailure

    Usually cytokine mediated (IL1, IL6, TNF, IL10)

    Associated with perioperative hemorrhage and the need for blood transfusion

    Infection

    NOSOCOMIALpatients prolonged hospitalization, immobilization, endotracheal

    intubation, suboptimal nutritional status, and need for multiple intravascular and bladder

    catheters

    DEVICE-RELATEDAbdominal pocket, leads exit site

  • 8/3/2019 Non Pharmachological Treatment of HF

    29/31

    Ventricular Assist Device

    Venezia, 5-6 November 2009

    Devices can be operated either in

    FIXED RATE MODE

    AUTOMATIC MODE

    more closely resembles normal physiologic conditions.

    the device ejects when the pump is 90 percent full or when it senses a

    decreased rate of filling.

    As the patients activity increases,the pump fills faster and the rate (or stroke

    volume) automatically increases, resulting in an increase in pump output. With a

    decrease in activity pump filling and output decrease.

    Because the aortic valve rarely opens when the heart is being supported by a

    left ventricular assist device, pump output is synonymous with cardiac output

    The development of long-term implantable devices that allow patients to be

    discharged to their homes has imposed the need to assess a patients ability to

    manage the device before implantation can be offered. The presence of a

    constant companion, though desirable, is no longer required.

    Interactions between the Patientandthe Device

  • 8/3/2019 Non Pharmachological Treatment of HF

    30/31

    Heart Transplantation

    Last therapeutical option

    Long waiting list

    Number of patients whoreceive heart transplants is

    still relatively low (around2,200 each year)

    85 percent of patients livefor more than a year aftertheir operations

    Venezia, 5-6 November 2009

  • 8/3/2019 Non Pharmachological Treatment of HF

    31/31

    Venezia, 5-6 November 2009

    Thankyouforyourattention!