nitrate amperometric sensor in neutral ph based on pd nanoparticles on epoxy-copper electrodes
TRANSCRIPT
Accepted Manuscript
Title: Nitrate amperometric sensor in neutral pH based on Pdnanoparticles on epoxy-copper electrodes
Author: Albert Gutes Carlo Carraro Roya Maboudian
PII: S0013-4686(13)00669-5DOI: http://dx.doi.org/doi:10.1016/j.electacta.2013.03.199Reference: EA 20334
To appear in: Electrochimica Acta
Received date: 31-1-2013Revised date: 27-3-2013Accepted date: 29-3-2013
Please cite this article as: A. Gutes, C. Carraro, R. Maboudian, Nitrate amperometricsensor in neutral pH based on Pd nanoparticles on epoxy-copper electrodes,Electrochimica Acta (2013), http://dx.doi.org/10.1016/j.electacta.2013.03.199
This is a PDF file of an unedited manuscript that has been accepted for publication.As a service to our customers we are providing this early version of the manuscript.The manuscript will undergo copyediting, typesetting, and review of the resulting proofbefore it is published in its final form. Please note that during the production processerrors may be discovered which could affect the content, and all legal disclaimers thatapply to the journal pertain.
Page 1 of 19
Accep
ted
Man
uscr
ipt
Nitrate amperometric sensor in neutral pH based on Pd nanoparticles on epoxy-
copper electrodes
Albert Gutés, Carlo Carraro, Roya Maboudian*
Author for correspondence:
Roya Maboudian
201 Gilman Hall
Department of Chemical and Biomolecular Engineering
University of California
Berkeley, CA, 94720
Tel: +1 510 643 7957
Fax: +1 510 642 4778
Email: [email protected]
Website http://cheme.berkeley.edu/faculty/maboudian/
Page 2 of 19
Accep
ted
Man
uscr
ipt
1
Nitrate amperometric sensor in neutral pH based on Pd nanoparticles on epoxy-
copper electrodes
Albert Gutés, Carlo Carraro, Roya Maboudian*
Department of Chemical and Biomolecular Engineering, University of California,
Berkeley, CA, 94720, USA
Abstract
Amperometric nitrate sensing at neutral pH using an epoxy-copper electrode
modified with palladium nanoparticles is presented. After epoxy-copper is
hardened and polished, electroless deposition is employed for the deposition of
Pd nanoparticles. Scanning electron microscopy and energy-dispersive X-ray
spectroscopy reveal the morphology and composition of the Cu/Pd surface. The
effect of Pd deposition time towards nitrate electroreduction is investigated,
highlighting the importance of the bimetallic catalyst. The optimized electrode
shows a linear response at pH=7 in the range from 2 to 35 ppm of nitrate. The
simplicity and cost effectiveness of the fabrication process makes this Cu/Pd
electrode a good candidate for distributed nitrate monitoring and determination
in the field.
Keywords Nitrate sensor, electrochemistry, epoxy-copper, palladium
nanoparticles, bimetallic catalyst
1. Introduction
Nitrate ground water pollution has become a growing concern in the last few
decades due to its toxicity at ppm concentration levels [1-2]. Nitrate in
groundwater can be divided into four categories depending on its source: (i)
waste materials, (ii) natural sources, (iii) irrigation in agriculture, and (iv) row
crop agriculture [3-5]. Overuse of nitrogen-based fertilizers and subsequent
leakage of nitrates into waters have caused a tremendous increase and spread
of this water pollutant worldwide [6]. Moreover, NOx increase in the atmosphere
Page 3 of 19
Accep
ted
Man
uscr
ipt
2
has been traced to overfertilization [7]. For these reasons, nitrate detection and
monitoring is of great importance. Optical UV detection of nitrates at 220 nm
can be employed [8], but is limited to the screening of clear unpolluted water
samples with low organic contents. Other methodologies for nitrate detection
include ion chromatography and the standard cadmium reduction method [9].
The Cd reduction method is a lengthy process and generates cadmium waste
with the associated disposal issues; thus the final cost of analysis is high.
A promising method for the detection of nitrate is based on its electroreduction
in the presence of metallic or bimetallic catalyst [10-14]. One of the most
explored bimetallic catalysts with strong electrocatalytic reduction properties
towards nitrate electroreduction is Cu-Pd. Much of the literature on the use of
Cu-Pd or other metallic catalysts has been limited to nitrate electroreduction in
acidic [15-21] and basic [22-27] media but since the majority of ground and
drinking waters are close to neutral pH, the previously developed sensing
materials cannot be directly used without sample pre-treatment. The majority of
studies have focused on the determination of the nitrate electroreduction
mechanism or on the catalytic and selective effect of the Cu-Pd electrode,
showing very high conversion rates to N2. De Vooys et al. presented a
thorough study on the use of Cu-Pd electrodes for the electroreduction of
nitrate, [28] proposing mechanisms in acidic and basic media, noting the
differences as pH values change, as well as the influence of the Cu to Pd ratio
and its effects towards selectivity and activity of the electrode. In addition, a
recent review by Rosca et al. [29] presents an exhaustive compilation of the
various electrocatalytical paths for nitrogen compounds, including nitrate, but
again acidic or basic media were used in all nitrate electroreduction
measurements. A few reports on nitrate detection by electroreduction in neutral
media can be found in the literature [30], but in all cases, the detected nitrate
ranges of concentrations exceed the legislation limits [31]. Here we present a
simple and robust fabrication method for nitrate sensors with a linear response
range of 2 to 35 ppm measured at pH=7.0. To achieve the required sensitivity,
the Pd to Cu ratio had to be carefully optimized which we have achieved via
electroless deposition schemes. To the best of our knowledge, this is the first
time that electrochemical reduction of nitrate at neutral pH has been achieved in
this concentration range.
Page 4 of 19
Accep
ted
Man
uscr
ipt
3
2. Experimental
2.1 Electrode construction
Figure 1 shows schematically the fabrication of the epoxy-Cu electrodes.
Epoxy-copper (Epoxy Technologies Epotek 430) is prepared as directed by the
manufacturer. PVC tubing of 3 mm inner diameter is cut in 2 cm long pieces. On
one end, a 3 mm in diameter electrical cable is introduced 1 cm into the tubing.
Epoxy-Cu is applied to the other end until filling the remaining 1 cm cavity in the
tube. After hardening, the epoxy-Cu is polished with sandpaper of decreasing
grain size with the final polishing performed with alumina pastes of 1-, 0.3- and
0.05-m size (CHI polishing kit CHI120) until mirror-shine is achieved.
Extensive rinsing with deionized (DI) water (18 MΩ, Barnsted Nanopure Infinity)
is performed prior to drying with nitrogen flow.
Deposition of the Pd nanoparticles is performed, right after polishing, as follows.
The deposition solution consists of 1 mM K2PdCl4 (Aldrich, 99.99% purity) and
20 mM KCl (Aldrich, 99% purity) and is used for all Pd depositions. Excess of
chloride is necessary to stabilize the PdCl42- anion in solution, avoiding its
precipitation as PdCl2. Polished epoxy-Cu electrodes are immersed in the
aforementioned solution for varied time duration, ranging from 10 to 300 s, then
rinsed in DI water and dried in a nitrogen flow. As explained in the results
section, the optimal deposition time towards nitrate reduction corresponds to 2-
min deposition time.
2.2 Characterization instruments
Electrochemical measurements (cyclic voltammetry and coulometry) are
performed using a CH660D potentiostat-galvanostat (CH Instruments, USA). A
standard three-electrode configuration is used with a Ag/AgCl reference
electrode, a Pt wire auxiliary electrode and the modified epoxy-Cu/Pd
nanoparticle working electrode. Scanning electron microscopy (SEM) and
energy-dispersive X-ray (EDX) analyses are obtained using a JEOL JSM 6340F
field emission scanning electron microscope.
Page 5 of 19
Accep
ted
Man
uscr
ipt
4
2.3 Amperometric detection
All measurements are performed in a 0.1 M phosphate buffer solution (PBS,
Fluka 99.4%) with pH adjusted to 7.0. Potassium nitrate (Fluka 99.8%)standard
solution containing 1000 ppm nitrate in PBS pH = 7.0 is prepared daily. Cyclic
voltammetry is performed between +0.2 and -1.2V at a scan rate of 50 mV/s
without stirring, starting at +0.2V. 10 cycles are performed in order to reach
reproducible voltammograms. The measurements are performed in open glass
beakers without any oxygen scavenging or removal. 3-step voltammetry is
performed by applying potentials (vs. Ag/AgCl) as follows, measuring each 20
ms: +0.2V for 2 s (Cu-Pd cleaning / oxidation step) followed by 2 s at -0.2V (Cu-
Pd regeneration step) and a final measuring step at -1V for 3 s. All the 3-step
measurements are performed without stirring. The average of the final 10-
recorded currents is used as analytical data. The first two steps (cleaning and
regeneration) are found to be crucial in order to avoid surface poisoning
reported previously for nitrate electroreduction [21,30] and are selected by
direct observation on the obtained cyclic voltammograms, where maximum
currents are achieved on the positive scan at around -0.1 V due to Cu/Pd
oxidation and maximum currents are achieved on the negative scan at around -
0.2 V during Cu/Pd reduction.
3. Results and discussion
3.1 Characterization of the epoxy-Cu/Pd nanoparticle substrate
Figure 2 shows the SEM image of the polished epoxy-copper prior to palladium
nanoparticles deposition (a) and after 2 min of palladium deposition (b). As can
be observed, the electrode surface morphology changes substantially upon Pd
deposition, with the formation of Pd nanostructures and films. Figure 2c shows
the EDX spectrum of the same sample as in Figure 2b. The spectrum confirms
the Pd deposition, with Pd/Cu ratio of 1:4.6. The Al signal is attributed to the
presence of Al in the epoxy-Cu formulation as-received.
Page 6 of 19
Accep
ted
Man
uscr
ipt
5
3.2 Catalytic reduction of nitrate by cyclic voltammetry
Figure 3a shows the typical cyclic voltammograms of the bare Cu electrode in a
0.1M PBS solution at pH=7.0 and in the presence of 20 ppm of nitrate. Figure
3b shows the response obtained with the epoxy-Cu/Pd electrode (2 min of
palladium deposition) in the same PBS and nitrate dissolutions. Scan rate is 50
mV/s in both cases. As can be observed, a reduction tail is obtained in the
negative scan starting at -0.8V, corresponding to nitrate reduction on the Cu/Pd
surface. In order to minimize possible interferences of hydrogen evolution, a
less extreme potential of -1V is selected for the third step in the nitrate detection
described next.
3.3 Three-step nitrate amperometric sensing
To characterize fully the electrodes’ responses, three-step voltammetry
measurements without solution stirring are performed, following a similar
protocol as described in the literature [32] as described in section 2.3. Figure 4a
shows the typical current responses obtained for the three-step voltammetry on
the bare epoxy-Cu electrode while figure 4b shows the same response using an
epoxy-Cu electrode after a 2 min Pd deposition time. In both cases PBS pH=7.0
baseline (black line) and after nitrate is added to a final concentration of 20 ppm
(red dashed line) is shown. It can be observed that no significant difference is
obtained when no Pd is present on the surface of the electrode (Figure 4a)
while an increase in the current is obtained when the Cu/Pd electrode is used
(Figure 4b). The current differences in the first two steps when using the Cu/Pd
electrode are due to the liberation and regeneration of adsorbed nitrogenated
compounds on the electrode surface that otherwise would cause its poisoning
as reported earlier [21,30].
The currents recorded during the last second of the three-step voltammetry
measurements are averaged and used as experimental data and plotted
against the different nitrate concentrations in the electrochemical cell obtained
by nitrate addition. Stirring of the solution is performed after each addition at
600 rpm for 5 s in order to homogenize the nitrate concentration in the
electrochemical cell. Before each three-step measurement, the stirring is
Page 7 of 19
Accep
ted
Man
uscr
ipt
6
stopped. Figure 5a shows the responses obtained from the electrodes with
various Pd deposition times. As can be observed, nitrate reduction sensitivity
increases with the amount of deposited Pd showing an optimal response at a
deposition time of 120 s. After this Pd deposition time, the response signal
decreases.
The observed behavior may be understood by the analysis put forth by Gao and
Li [33] for PdCu bimetallic catalyst for nitrate reduction. Namely, for nitrate
reduction, both Cu and Pd must be available during the electron transfer
process. The need for Pd in the catalyst mixture is demonstrated in Figure 4a,
where an electrode with no deposited Pd is used and signal remained
unchanged in the presence of nitrate. The need of copper is demonstrated by
using an epoxy-silver electrode instead of an epoxy-copper electrode with a 2-
min Pd deposition following the same procedures as before [32].
Figure 5b shows the calibration plot obtained when using this electrode. As can
be observed, the data obtained in the absence of Cu in the catalyst only
presents noise. We conjecture that for Pd deposition times longer than 2 min,
Cu starts being covered by Pd to such an extent that it is no longer available to
nitrate in solution. Figure 6 shows the calibration curve for the optimal 2 min Pd
deposition time electrode on a log-log plot. The linear range for nitrate reduction
at pH=7.0 is 2 to 35 ppm, with excellent linear correlation. The achieved
detection range is well in the legislation limits of 10 ppm [31]. Reproducible
response of the electrodes is obtained over periods of months. This is attributed
to the regenerative nature of the 3-step measurements described in section
2.3.Table 1 compares the performance of the here presented Cu-Pd electrode
to those reported previously. The table clearly shows that the Cu-Pd electrode
provides an excellent response towards nitrate reduction when compared to
previous reports in the literature with the added novelty of sensing in neutral pH.
The electrochemical nitrate detection scheme presented here should not be
influenced by many ions that are usually present in natural waters (such as
HCO3-, SO4
2- Ca2+, and Mg2+) since none of these cations or anions can be
either reduced or oxidized in the potential range of our measurements. The
possibility of nitrite interference was studied using the optimal 2 min Pd
electrode by adding 1000 ppm NaNO2 (Fluka 99.8%) to 25 ml PBS buffer as
performed for nitrate measurements, with no response observed over the
Page 8 of 19
Accep
ted
Man
uscr
ipt
7
examined range of ?? to 50 ppm. Previous studies on the electroreduction of
nitrite by copper complexes [34] suggest that nitrite reduction follows a different
reaction pathway than nitrate reduction that might not be possible in the
presence of Pd.
4. Conclusions
In conclusion a facile fabrication method for a new electrochemical nitrate
sensor based on epoxy Cu and electrolessly deposited Pd is presented. The
main advantage of the presented electrode is the possibility of nitrate
determination in the legislation-relevant range at neutral pH. Nitrate response is
achieved by a three-step amperometry consisting of a surface cleaning and
regeneration followed by a nitrate electroreduction. Stable current is used as
analytical signal for the calibration of the electrodes. Optimization of the amount
of deposited Pd has been performed and a final condition of 1mM Pd
dissolution with a 2-min deposition time is found to provide the highest
sensitivity towards nitrate reduction. The linear range provided by the here
presented electrode is 2-35 ppm, a suitable range for its use in the detection of
nitrates in drinking water and crop soils among others. The simplicity of the
fabrication process, the low cost of the starting components, the high rate of
precursor utilization expected in the electroless deposition and the possibility of
using inexpensive counter electrodes, such as graphite, open potential for mass
production of the described sensor.
Acknowledgements
The authors are grateful to Mr. Peter Lobaccaro for his help with SEM and EDX
imaging. This work was supported by National Science Foundation under
Grant# EEC-0832819 (Center of Integrated Nanomechanical Systems).
References
[1] E.H. Burden, Toxicology of nitrates and nitrites with particular reference to
potability of water supplies, Analyst 102 (1961) 429.
Page 9 of 19
Accep
ted
Man
uscr
ipt
8
[2] M.H. Ward, T.M. deKok, P. Levallois, J. Brender, G. Gulis, B.T. Nolan, J.
Van Derslice, Workgroup report: Drinking-water nitrate and health-recent
findings and research needs, Environ. Health Perspect. 113 (2005) 1607.
[3] L.W. Canter, Nitrates in Groundwater, CRC Press, Boca Raton, FL, 1996.
[4] B.A. Stewart, F.G. Viets, G.L. Hutchins, Agricultures effect on nitrate
pollution of groundwater, J. Soil Water Conserv. 23 (1968) 13.
[5] M.N. Almasri, J.J. Kaluarachchi, Assessment and management of long-term
nitrate pollution of ground water in agriculture-dominated watersheds, J. Hydrol.
295 (2004) 225.
[6] J.M. Beman, K.R. Arrigo, P.A. Matson, Agricultural runoff fuels large
phytoplankton blooms in vulnerable areas of the ocean, Nature 434 (2005) 211.
[7] S. Park, P. Croteau, K.A. Boering, D.M. Etheridge, D. Ferretti, P.J. Fraser,
K.R. Kim, P.B. Krummel, R.L. Langenfelds, T.D. van Ommen, L.P. Steele, C.M.
Trudinger, Trends and seasonal cycles in the isotopic composition of nitrous
oxide since 1940, Nature Geosci. 5 (2012) 261.
[8] T. Kamiura, M. Tanaka, Determination of nitrate in suspended particulate
matter by high-performance liquid-chromatography with UV detection, Anal.
Chim. Acta 110 (1979) 117.
[9] APHA. Standard methods for the examination of water and wastewater. 18th
ed. American Public Health Association, Washington, DC, 1992.
[10] K. Fajerwerg, V. Ynam, B. Chaudret, V. Garçon, D. Thouron, M. Comtat,
An original nitrate sensor based on silver nanoparticles electrodeposited on a
gold electrode, Electrochem. Comm. 12 (2010) 1439.
[11] F. Gauthard, F. Epron, J. Barbier, Palladium and platinum-based catalysts
in the catalytic reduction of nitrate in water: effect of copper, silver, or gold
addition, J. Catal. 220 (2003) 182.
[12] G.E. Dima, A.C.A. de Vooys, M.T.M. Koper, Electrocatalytic reduction of
nitrate at low concentration on coinage and transition-metal electrodes in acid
solutions, J. Electranal. Chem. 554-555 (2003) 15.
[13] L.A. Estudillo-Wong, E.M. Arce-Estrada, N. Alonso-Vante, A. Manzo-
Robledo, Electro-reduction of nitrate species on Pt-based nanoparticles:
Surface area effects, Catal. Today 166 (2011) 201.
[14] M. J. Moorcroft, J. Davis, R. G. Compton, Detection and determination of
nitrate and nitrite: a review, Talanta 54 (2001) 785.
Page 10 of 19
Accep
ted
Man
uscr
ipt
9
[15] N. G. Carpenter, D. Pletcher, Amperometric method for the determination
of nitrate in water, Anal. Chim. Acta 317 (1995) 287.
[16] A. G. Fogg, S. P. Scullion, T. E. Edmonds, Assessment of online nitration
reactions as a means of determining nitrate by reverse flow-injection with
reductive amperometric detection at a glassy-carbon electrode, Analyst 114
(1989) 579.
[17] A. G. Fogg, S. P. Scullion, T. E. Edmonds, B. J. Birch, Direct reductive
amperometric determination of nitrate at a copper electrode formed insitu in a
capillary-fill sensor device, Analyst 116 (1991) 573.
[18] M. J. Moorcroft, L. Nei, J. Davis, R. G. Compton, Enhanced electrochemical
detection of nitrite and nitrate at a Cu-30Ni alloy electrode, Analytical Letters 33
(2000) 3127.
[19] Z. Zhao, X. Cai, Determination of trace nitrite by catalytic polarography in
ferrous iron thiocyanate medium, J. Electroanal. Chem. 252 (1988) 361.
[20] J. Davis, M. J. Moorcroft, S. J. Wilkins, R. G. Compton, M. F. Cardosi,
Electrochemical detection of nitrate and nitrite at a copper modified electrode,
Analyst 125 (2000) 737.
[21] A. O. Solak, P. Gulser, E. Gokmese, F. Gokmese, A new differential pulse
voltammetric method for the determination of nitrate at a copper plated glassy
carbon electrode, Mikrochim. Acta 134 (2000) 77.
[22] S. Cattarin, Electrochemical reduction of nitrogen oxyanions in 1-M sodium-
hyrdoxide solutions at silver, copper and CuInSe2 electrodes, J. Appl.
Electrochem. 22 (1992) 1077.
[23] H.L. Li, J.Q. Chambers, D.T. Hobbs, Electroreduction of nitrate ions in
concentrated sodium-hydroxide solutions at lead, zinc, nickel and
phtalocyaninie-modified electrodes, J. Appl. Electrochem. 18 (1988) 454.
[24] J.O. Bockris, J. Kim, Electrochemical reductions of Hg(II), ruthenium-
nitrosyl complex, chromate, and nitrate in a strong alkaline solution, J.
Electrochem. Soc. 143 (1996) 3801.
[25] J.D. Genders, D. Hartsough, D.T. Hobbs, Electrochemical reduction of
nitrates and nitrites in alkaline nuclear waste solutions, J. Appl. Electrochem. 26
(1996) 1.
Page 11 of 19
Accep
ted
Man
uscr
ipt
10
[26] M. Fedurco, P. Kedzierzawski, J. Augustynski, Effect of multivalent cations
upon reduction of nitrate ions at the Ag electrode, J. Electrochem. Soc. 146
(1999) 2569.
[27] R. Tenne, K. Patel, K. Hashimoto, A. Fujishima, Efficient electrochemical
reduction of nitrate to ammonia using conductive diamond film electrodes, J.
Electroanal. Chem. 347 (1993) 409.
[28] A.C.A. de Vooys, R.A. van Santen, J.A.R. van Veen, Electrocatalytic
reduction of NO3- on palladium/copper electrodes, J. Mol. Catal. A: Chem. 154
(2000) 203.
[29] V. Rosca, M. Duca, M.T. de Groot, M.T.M. Koper, Nitrogen Cycle
Electrocatalysis, Chem. Rev. 109 (2009) 2209.
[30] O. Ghodbane, M. Sarrazin, L. Roue, D. Belanger, Electrochemical
reduction of nitrate on pyrolytic graphite-supported Cu and Pd-Cu
electrocatalysts, J. Electrochem. Soc. 155 (2008) F117.
[31] http://water.epa.gov/drink/contaminants/basicinformation/nitrate.cfm,
consulted on October 31st 2012.
[32] A. Gutes, C. Carraro, R. Maboudian, Nonenzymatic glucose sensing based
on deposited palladium nanoparticles on epoxy-silver electrodes, Electrochim.
Acta 56 (2011) 5855.
[33] W. Gao, F. Li, Catalytic Hydrogenation of Nitrate Ions over Pd-Cu/ZSM-5
Catalyst, Advanced Materials Research 197-198 (2011) 967.
[34] J.G. Woollard-Shore, J.P. Holland, M.W. Jones, J.R. Dilworth, Nitrite
reduction by copper complexes, Dalton Trans. 39, (2010) 1576.
Page 12 of 19
Accep
ted
Man
uscr
ipt
11
Table 1.
Comparison of analytical performance of some electrochemical nitrate sensors
Electrode material Solution nitrate concentration (M) Ref.
AgNP on Au Synthetic seawater 10-5 – 10-2 [10]
Polycrystalline metal 0.5M H2SO4 or
HClO4
0.1 [12]
Cu/graphite and Pd–
Cu/graphite
1M NaCl 1 [30]
Pt-based NP 0.5M NaOH 0.001 – 1 [13]
Vitreous carbon 0.1 M Na2SO4 at
pH = 2.9
10-4 - 10-3 [15]
Glassy carbon Concentrated
H2SO4
5 ·10-4 – 5 ·10-3 [16]
Electrodeposited Cu 0.5 - 2M H2SO4 10-4 - 10-3 [17]
Epoxy-Cu-Pd-NP PBS buffer pH=7.0 3.2·10-5 – 5.6·10-4 This work
Page 13 of 19
Accep
ted
Man
uscr
ipt
12
Figure Captions:
Figure 1: Top: electrode construction diagram. A 3mm inner diameter PVC tube
(a) is cut into 2cm long pieces. Half of the tube is filled with a 3mm electrical
cable (b) and the rest is filled with epoxy-Cu and hardened (c). After hardening,
the electrode surface is polished with sandpaper and alumina slurry until a
mirror-shine surface is obtained (d). Bottom: picture of the unmodified Cu-epoxy
electrode (left) and after the 2 min Pd deposition (right).
Figure 2: SEM images of (a) polished epoxy Cu and (b) after a 2 min Pd
deposition; (c) EDX spectrum of pristine epoxy-Cu and (d) EDX spectrum of
sample shown in (b).
Figure 3: (a) Cyclic voltammograms for the bare Cu electrode in PBS (black
line) and in a PBS solution containing 20 ppm nitrate (red dashed line). (b)
Cyclic voltammograms obtained at pH=7.0 on the epoxy-Cu/Pd electrode (2 min
Pd deposition) in a 0.1M PBS solution (black line) and in a PBS solution
containing 20 ppm nitrate (red dashed line). Scan rate 50 mV/s. Scans started
at +0.2V
Figure 4: (a) Three-step voltammetry on the epoxy-Cu in PBS pH=7.0 baseline
(black line) and in the presence of 20 ppm nitrate (red dashed line). (b) Same
measurements as in (a) when Pd has been deposited on the epoxy-Cu
electrode surface for 2 min. Blue dashed line shows the applied voltage along
the 7 second measurement.
Figure 5: (a) calibration responses for the studied set of Pd deposition times.
Plotted currents correspond to the average of the currents recorded during the
last second in the three-step measurements after nitrate additions. (b)
Calibration response of an epoxy-Ag/Pd electrode with Pd deposited for 2 min.
Figure 6: Linear response of the optimal 2 min Pd deposition in the range of 2 -
35 ppm of nitrate.