NEW METHOD TO PRODUCE LASER CALIBRATION BEAMS IN GASEOUS DETECTORS.
Alexei Lebedev, Alexei Lebedev, Brookhaven National Laboratory, Upton, NY, 11973, Brookhaven National Laboratory, Upton, NY, 11973, USAUSA
Introduction
Narrow laser beams widely used in many experiments to simulate charge particles and align detectors[1]. For LHC or future linear collider the requirements to correct the position and to measure coordinates of particles are about 10-20m for whole detector with the size ~ 10 meters. Different approaches are proposed to create laser beams [2-4]. The direct way for alignment is directing of narrow laser beam going through all detectors. The ability to resolve the location of beam is proportional to the beam radius and it is attractive to use calibration beams with less diameters. Current approach is to use focusing telescopes or diffraction through hole(fig.1). There are physical limitations due to diffraction limit. Divergence of the beam ~0.64 /w, where -wavelength and w- waist of the beam. It is obvious that it is impossible to create long beams with sub millimeter diameter.
It is known the creation of bright spot, called Poisson spot(fig.2), when wide wave plane illuminates an opaque sphere. This line created in the shadow of sphere and could be used to monitor positioning of different objects, because the divergence of Poisson line is significantly smaller ~ /d, where d is diameter of the ball [5,6].
We propose to use “Poisson beam” to simulate straight tracks in gaseous detectors. This beam could create ionization due to two-photon ionization process. Additional diaphragm installed after the ball or annular diaphragms eliminates background ionization from bright field. We present results with different balls and diaphragm sizes to form very narrow laser beams with 100-500m diameter and 3-10 meters long with an ionization exceeding relativistic particle(fig.5). Note that our beam size measurements were made on base of beam profile(fig. 4). Special optics with cone lens proposed to create a system of these beams (fig.6).
References:
1. W. Blum, L. Rolandi. Particle detection with Drift Chambers, Springer-Verlag, 1994.
2.RASNIK for ATLAS http://www.nikhef.nl/pub/departments/et/experiments/atlas/rasnik/atlas_rasnik/atlas_rasnik.html
3. W. Blum, H. Kroha, P. Widman, NIM, A 267 (1995) 413-417.
4. T.Akesson et al. An alignment method for the ATLAS end-cup TRD detector using a narrow monochromatic X-ray beam, NIM A463 (2001), 129-141.
5. L.V.Griffith et al. Magnetic alignment and the Poisson alignment reference system. Rev. Sci. Instr. 61(8), Aug 1990, 2138-2154.
6. A.Lebedev et al, A laser calibration system for the STAR TPC, NIM A478 (2002) 163, NIM A499 (2003), 692.
VCI 2007
ABSTRACT
A new method utilizing diffraction of UV laser beams on annular diaphragms provides very narrow laser beams with full diameter 100-400 m, divergence ~0.05mrad and effective length up to 10 meters, which exceeds existing methods with focusing optics. The characteristics of laser beams and linear ionization created with different diaphragm sizes are present. Optics schemes proposed to create a system with new beams.
1. Focusing through telescope
w
w, ~.18mrad, 266nm
2. Diffraction through hole- Airy disk
d~1mm d, ~.32mrad, 266nm
Fig. 1 Principles to create laser beams
Fig. 2 Principles to create Poisson beams
Laser D
D1L
Fe55 Drift chamber
Microscope
or
screen with scale
d
Fig. 3. Experimental setup
Fig 4. Beam profile for different
ball sizes and wavelengths.
d mrad, m, mmHe-Ne laser, 633nm
d mrad, m, mmUV laser, 266nm
Fig. 5 Electron linear density
and beam profile for 3.12 and
5.0 mm balls..
widelaserbeam
micromirror
cone lens
micromirror’sbundle
annular beam
Fig. 6. Proposed optics to create annular beams.
UV laser beam creates
electrons via 2-photon
ionization