Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics New and Improved AMS Facilities Hans-Arno Synal Laboratory of Ion Beam Physics 8093 Zurich Switzerland HIAT'09 June, 10 th 2009

Upload: others

Post on 01-Aug-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

New and Improved AMS Facilities

Hans-Arno Synal

Laboratory of Ion Beam Physics8093 Zurich Switzerland

HIAT'09 June, 10th 2009

Page 2: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Discovery of AMS in 1977AMS-Heros

A.E. LitherlandK.H. PurserH.E. Gove

R.P. BeukensR.P. Clover

W.E. SondheimR.B. LiebertC.L. Bennet

first 14C spectraThe Rochester MP Tandem accelerator

Page 3: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Long-lived / AMS radionuclides

14C 5730 yrsArchaeology, Art History, Oceanography, Earth &

Environmental Sciences

10Be 1.5 MyrsOceanography, Geology, Geophysics, Earth &

Environmental Sciences...

26Al 716 kyrsGeology, Geophysics, Life sciences, Biomedical

Applications...

36Cl 301 kyrsPolar Ice Cores, Cosmic Ray, Solar Activity,

Hydrology..

41Ca 103 kyrsLife Sciences, Biomedical Applications

129I 16 MyrsEarth & Environmental Sciences, Radioecology,

Nuclear safeguards,...

Actinides (Pu, U, Pa)Radioecology, Nuclear Safeguards,Oceanography...

1012

109

106

103

1

10-6

10-3

Hal

f-life

/ yr

s

age of the universeage of the solar system

grain of sand

Page 4: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Why does AMS reach unparalleled sensitivity?

• Isobar separation– Negative ion formation

• 14C(14N); 26Al(26Mg);• 10Be(10B); 36Cl(36S); 41Ca(41K); • 10BeF/BeBaF(10BF); 41CaH3(41KH3); 41CaF3(41KF3)

– Single ion detection• Abundance sensitivity (1:1015)

– Suppression of neighboring isotopes– Multi-step mass filtering process

• Reliable normalization– Reproducible isotope ratio measurements– High ion optical transmission

• Eliminate mass interferences– Molecule destruction

– Charge state >3+ molecule dissociation by coulomb force– Charge state 1+ in multiple ion gas collisions

Stripper

gas / foil

negative ions positive ions

Page 5: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

AMS: principles of operation

3. Filtermolecule destruction

3. Filtermolecule destruction

2. Filtermass analysis

(low-energy end)

2. Filtermass analysis

(low-energy end)

mass spectrometer charge exchange

5. Filterparticle identification

5. Filterparticle identification

4. Filtermass analysis

(high-energy end)

4. Filtermass analysis

(high-energy end)

massspectrometer

electrostaticdeflector

1.Filternegative ion formation

1.Filternegative ion formation

13C / 12C14C / 12C

(Stripping process)

13CH-

12CH2-

14C-

argon stripper gasnegative ions positive ions

14CΦ+

13CΦ+

12CΦ+

H+

10-9

1

1

:

:

7Li2-

12C

13C

Page 6: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

The Golden AMS Rule: q > 3+ !!!

Strippernegative ions positive ions

atomic ions

molecularfragments

Coulomb disintegration (q≥3+)

e-

e-

e-

e-

13C3+

H

maximum yield (14C):high energies and large accelerators

q=3+ 2.5 MVq=4+ 6.5 MV

atomic ions

molecules

13CH-

12CH2-

14C-

argon stripper gasnegative ions positive ions

14CΦ+

13CΦ+

12CΦ+

H+

10-9

1

1

:

:

7Li2-

Page 7: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Charge state yield of 14C ions in Ar gas

0 1+ 2+ 3+ 4+

Traditional AMS3 MV - 6 MV

Next AMS Generation

200 - 300 kV

Multiple ion gas collisions Coulomb disintegration

Compact AMS0.5 MV - 1 MV

Page 8: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Breaking the Rule: AMS with q = 1+, 2+ Ions

Page 9: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Breaking the Rule: AMS with q = 1+, 2+ Ions

Page 10: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Compact AMS systems

The first compact AMS system (1998) using charge state 1+ (ETH/PSI-NEC Collaboration)

Commercial systems are now on the market from NEC and HVEE25 out of the 86 world-wide operating AMS spectrometers

are based on this principle!

Page 11: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Commercial AMS systems

National Electrostatics CorporationMiddleton, Wisconsin, USA

5-6 MV

3 MV 1 MV

5-6 MV

3 MV

500 kV

250 kV SSAMS

Page 12: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

1 MV Multi-nuclide AMS system

Supplier: High Voltage Corp., Amersfoort, The Netherlands

AMS facility, at CNA, Seville, Spain quadrupole triplet to match ion optics of different charge states

1 MV Tandetronaccelerator

≈ 4.5 m

Page 13: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

NEC-Single-Stage-AMS system

6.55 m

5.25

m

National Electrostatics Corp. (USA)Single-Stage-AMS system (2002)

Page 14: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Compact lab-sized instrument– Designed for operator safety

– No open high voltages– Easy to operate

– Easy to tune– Fully automated

– Web based monitoring– Fail-safe sample handling

– High precision– High throughput

– Reasonable investment costs

Figure of merit a of next generation AMS systems

2.5 m3.0 m

Rather a “mass spectrometer”than an accelerator based system

Page 15: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Low energy limits?

0

10

20

30

40

50

60

70

80

0 200 400 600 800Energy / keV

68%80%95%

5

5.5

6

6.5

7

7.5

8

8.5

9

0 200 400 600 800

Energy / keV

13CH+12CH2+12CH2++

Cross sections of molecule destruction processes

Molecular species

Transmitted beamintensities

• Cross sections are comparable to molecular sizes

• Only weak energy dependence

• @ 230 keVcross sections are about 10 % lower

Energy dependence of angular straggling

New concepts can be applied at stripping energies below 250 keV!!

To deal with ion beams of large divergence

2 µg /cm2

stripper gas

Page 16: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

MIni CArbon DAting System: Prototype (2004)

High energy end: achromatic

mass spectrometer

Injection magnet: fast beam switching

system

200 kV acceleration unit:vacuum insulated

high voltage platform

Sputter ion source:spherical ionizermulti cathode

sample changer

Detector:high resolutiongas ionization

detector

2.5 m

3.0 m

Page 17: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Acceleration unit details

200 kV corona gap inside acceleration unit

Stripper housing on insulator

Page 18: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Routine Performance of ETH-MICADAS AMS system

Routine operation for all radiocarbon measurements at ETH since May 2008

- 3500 graphite targets analyzed - 700 CO2 sample (<50μg)

- runs unattended with low maintenance

- dates samples back to 45 000 years

- allows high-precision measurements

Blank variation: < 0.5 ‰

Statistic uncertainty: 1.0 - 3.0 ‰

Reproducibility of samples: 0.5 - 3.0 ‰

Standard normalization: 0.5 - 1.5 ‰

Over all uncertainty: 1.5 - 5.0 ‰

ETH-MICADAS AMS system

Page 19: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Development of a novel Biomed AMS system

have entered into a collaborative agreement:

To build a novel instrument for biomedical applications using the most advanced AMS technology

• PSI will provide expertise for design and construction of a BIO-MICADAS machine specifically optimized for biomedical AMS applications

• Vitalea Science will provide expertise and experience as a biomedical AMS service provider.

Paul Scherrer Institut

Page 20: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Why AMS for Bio Sciences?

Page 21: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Micro dosing to accelerate drug development

+0 200 400 600 800

100

200

300

400

500

Fit: A*e-x/c1+B*e-x/c2+C Gas

PMC

Time after Labeling/h

labeled substance

Administration at μ−dosewithout pharmaceutic or toxic effect

14C

Absorption

Distribution

Metabolism

Excretion

Page 22: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

BioMICADAS: Ready to go

Page 23: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Dual load lock system

Vacuum Lock 2

Vacuum Lock 1

(invisible)

Measuring

chamber

Turbopump Gate valves

Samplein measuring

position

Sample transfer unit

random access

to all samples

continuous operation for high throughput

Page 24: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Installation of BioMICADAS

Shipping: 6260 kg

with air freightDavis, CA Zurich

June, 19th 2008

BioMICADASArrival:June, 26th, 2008

Laboratory of Ion Beam Physics

Page 25: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Characteristic / specifications (BioMICADAS)

• Beam parameter status– Typical beam 12C currents: ≈ 40-80 μA– Transmission: 41-42%– 14C counting rate for modern sample: ≈ 200 cps

• Measuring parameter– Typical measurement time per sample: 300 s– Repeatability: ≈ 2 ‰– Cover the full range of Biomed AMS 10-9- 3 x 10-15

• Operation parameter– Continuous measurements of samples– Throughput per running day (24h): more than 200 samples / day– 20’000 - 50’000 annual measurement capacity

• Shipping the instrument to Vitalea, Davis USA– Shipping June 19-26, 2008– Installation and commissioning June 28 - July 8, 2008– Final acceptance tests July 17-21, 2008

Page 26: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

High precision radiocarbon dating

Dating MICADAS

Developed and built atOperated in Mannheim (spring 2010)

Page 27: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Frontiers in radiocarbon dating

The latest generation of radiocarbon AMS systems (MICADAS and DatingMICADAS)

• rather mass spectrometer than accelerator based systems• improved radiocarbon measurement capabilities

What can we do with this?

Page 28: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Test of high precision capabilities

4 individually processed Oxa II standards during high precision measurements

Each standard was analysed 3 hours, acquiring a total counting statistic of ≈ 1‰

Page 29: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Sampling the historic document

Cutting sample material from the parchment document

Taking fibers of the seal cord

GoldeneHandfeste

Berne: an Imperial City of the Holy Roman Empire of the German Nation

Signed: 15. April 1218Aproved by

Rudolf v. Habsburg 1274

Friedrich II (Emperor 1194–1250

Page 30: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

"Goldene Handfeste“ results

Sample Radiocarbon age δ13C Date ofnumber Type Years (BP) (‰) measurement

ETH-36716.1 Parchment 888 ± 20 -22.6 ± 1.1 18. Dec. 08ETH-36716.2 Parchment 878 ± 19 -20.7 ± 1.1 18. Dec. 08ETH-36716.3 Parchment 882 ± 19 -23.9 ± 1.1 27. Jan. 09ETH-36716.4 Parchment 875 ± 19 -22.3 ± 1.1 27. Jan. 09ETH-36716 Parchment 881 ± 10 -22.4 ± 0.6 mean

Sample Radiocarbon age δ13C Date ofnumber Type Years (BP) (‰) measurement

ETH-36717.1 Seal cord 800 ± 20 -24.0 ± 1.1 18. Dec. 08ETH-36717.2 Seal cord 808 ± 19 -29.1 ± 1.1 27. Jan. 09ETH-36717.3 Seal cord 833 ± 18 -25.5 ± 1.1 27. Jan. 09ETH-36717.4 Seal cord 808 ± 18 -27.1 ± 1.1 27. Jan. 09ETH-36717.5 Seal cord 800 ± 17 -27.7 ± 1.1 27. Jan. 09ETH-36717 Seal cord 809 ± 8 -26.7 ± 0.9 mean

Page 31: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Absolute age calibration

Page 32: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Refining the calibration curve

parchment

seal cord

Historic date

Page 33: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Refining the calibration curve

parchment

seal cord

Historic date

Page 34: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Ion source developments

Physical constrains:• beam emittance• ionization efficiency• isotope fractionation

Practical constrains:• reliability of operation• sample handling• maintenance and performance• hybrid mode operation (solid and gas targets)

The ion source is the most critical element of an AMS system!

Page 35: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

New ion source concepts

Primary design goals– No open high voltage potentials

– Precise/reproducible positioning of targets– Good vacuum conditions

– Hybrid operation graphite / gaseous samples– Continuous operation during exchange of

magazines

Ionizer assembly in operation: view through ion source window

1.30 m

ETH ion source

Page 36: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Hybrid source for direct CO2 analysis

Page 37: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Gas feed operation (Ultra small samples <50μg)

Target positionIn source

Page 38: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Gas feed operation (Ultra small samples <50μg)

�capillary to ion source

Syringe

CO2 trap

Ampoule cracker

Target positionIn source

Page 39: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Performance in routine operation

Measured samples (12 month): 700 direct CO2Limit of quantification: 35'000 yrsIon source currents: 3-5μATypical uncertainty: 8-15 ‰Sample size: 5-50 μg

Sample size / μg

Page 40: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Dating Alpine ice cores (Collé Gnifetti)

Collaboration University of Berne: H.Gäggeler, M.Schwikowski,...

Method:Dating of organic carbon (OC)

• 5-20 μg samples

Page 41: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

The other nuclides?

10Be, 26Al, 41Ca, 129I, Pu,..

Has to be explored at very low beam energies.

36Cl remains a "pièce de resistance"But there might be alternatives at eV energies!

Accelerators of about 500 kV,are interesting alternatives!

Page 42: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Revolution in low energy ion detection

Si3N4 detector window

30 nm

Detector with CoolFETpre-amplifiers

Anodes arrangement

Resolution of total energy measurements

Page 43: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Gas ionization detector

Nuclide separation: specificStopping power characteristics

-100V

e-

dxdx

dEdxdE

restEE

⎟⎠⎞

⎜⎝⎛−⎟

⎠⎞

⎜⎝⎛= ∫

21δ

10Be

10B

ERest

ΔE ERestΔE

10B

10Be

Detector position / mm

dE /

dx /

(keV

/ m

m)

ions

detector window

• Ionization

• Charge collection

• Charge sensitive Pre-Amplifier

• Energy loss measurements

Bethe-Bloch

Page 44: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Energy-loss straggling

0.01 0.1 1 100

1

2

3

4

12C

56Fe

32S27Al

16OYa

ng /

Bohr

Stra

gglin

g

E / Z3/2 / (MeV /amu)

7Li

Bohr:Ω2 ∝ ⋅Z1

2 ⋅ Z2 ⋅N

Chu/Yang:

⎟⎠⎞

⎜⎝⎛

ΩΔΩ+⎟

⎠⎞

⎜⎝⎛

ΩΩ=⎟

⎠⎞

⎜⎝⎛

ΩΩ

22

2

22

22

BB

Chu

IonBγ

Charge state fluctuations

Z1 > 2

Energy straggling semi empirical model form Chu, γ: mean charge of ion

Bohr value

Page 45: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

• Energy straggling is less than expected• Pulse height defect helps for isobar suppression• Focusing of straggling at left side of Bragg peak (Tschalär und Schmidt-Böcking)

Focusing of energy straggling

FT =

S(E0 −ΔE)S(E0)

S(E) : Stopping power at energy EE : initial ion energyΔE : energy loss

E0−ΔEE0Region of focusing

Scaling for energy loss straggling

Page 46: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Routine 129I analysis @ 2 MeV

Measurement parameter:Target AgI/AgCurrent 5-10 μACharge state 3+Stripper Ar (0.22 μg/cm2)Terminal 0.5 MV -> 2 MeVTransmisson 8%Machine blank 9.10-14

20.10-12 sample 1.5% in 2 min

Detector parameter:2 Anodes 3 cm eachGas IsobutaneSiN window 4x4 mm2, 50 nm

≈ 700 129I samples analysed per year (unattended operation)

Blank:129I/I <10-13Standard:129I/I ≈10-11

Page 47: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Routine 129I analysis @ 2 MeV

Measurement parameter:Target AgI/AgCurrent 5-10 μACharge state 3+Stripper Ar (0.22 μg/cm2)Terminal 0.5 MV -> 2 MeVTransmisson 8%Machine blank 9.10-14

20.10-12 sample 1.5% in 2 min

Detector parameter:2 Anodes 3 cm eachGas IsobutaneSiN window 4x4 mm2, 50 nm

≈ 700 129I samples analysed per year (unattended operation)

Blank:129I/I <10-13Standard:129I/I ≈10-11

Page 48: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

10BeO- and SiN degrader

2600 40 80 120 160 200

Eres in channels240

300

340

380

420

460

500

10B

10Be gate

2D - Spectrum of the ETH low level standardafter passing the 2nd HE-magnet

ETH- 600 kV Pelletron System

10Be gate

9Be

Page 49: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

10BeO- and SiN degrader

2600 40 80 120 160 200

Eres in channels240

300

340

380

420

460

500

10B

10Be gate

2D - Spectrum of the ETH low level standardafter passing the 2nd HE-magnet

SiN degrader foil

ETH- 600 kV Pelletron System

10Be gate

9Be

Page 50: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

10BeO- and SiN degrader

2600 40 80 120 160 200

Eres in channels240

300

340

380

420

460

500

10B

10Be gate

2D - Spectrum of the ETH low level standardafter passing the 2nd HE-magnet

SiN degrader foil

ETH- 600 kV Pelletron System

10Be gate

9Be

Page 51: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

10BeO- and SiN degrader

2600 40 80 120 160 200

Eres in channels240

300

340

380

420

460

500

10B

10Be gate

2D - Spectrum of the ETH low level standardafter passing the 2nd HE-magnet

SiN degrader foil

achromatic mass analysis

ETH- 600 kV Pelletron System

Page 52: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

10BeO- and SiN degrader

2600 40 80 120 160 200

Eres in channels240

300

340

380

420

460

500

10B

10Be gate

2D - Spectrum of the ETH low level standardafter passing the 2nd HE-magnet

SiN degrader foil

achromatic mass analysis

ETH- 600 kV Pelletron System

Page 53: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Extension of HE-mass spectrometer

MC SNICSIon Source

LE-Magnet(90°)

600 kV Pelletron Accelerator

1. HE-Magnet(90°)

Stripper

2. HE-Magnet(135°) Electrostatic

Analyzer (90°)

Degrader Foil

Page 54: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

Improved HE-spectrometer setup for 600kV system

Page 55: New and Improved AMS Facilities · Why AMS for Bio Sciences? Eidgenössische Technische Hochschule Zürich Swiss Federal Institute of Technology Laboratory of Ion Beam Physics Micro

Eidgenössische Technische Hochschule ZürichSwiss Federal Institute of Technology

Laboratory of Ion Beam Physics

6 MV EN-Tandem 600 kV PSI/ETH system 200 kV system for 14C

Multi isotope system

Evolution of AMS over the last 10 years

High precision 14C dating

15 m 2.5 m 1 m

Biomedical 14C applications

TANDY BioMICADASMICADAS/DatingMICADAS