mpd-600-anp-10036-enu

11
© OMICRON Page 1 of 11 Application Note Method statement for partial discharge measurement Author Dr.-Ing. Kay Rethmeier | [email protected] Date Oct, 2007 Related OMICRON Product CAL 542, MPD 600, MCU 502 Application Area Partial Discharge Measurements Version v1.0 Document ID ANP_10036_ENU

Upload: insan-aziz

Post on 13-Sep-2015

43 views

Category:

Documents


1 download

DESCRIPTION

MPD

TRANSCRIPT

  • OMICRON Page 1 of 11

    Application Note

    Method statement for partial discharge measurement

    Author Dr.-Ing. Kay Rethmeier | [email protected] Date Oct, 2007

    Related OMICRON Product CAL 542, MPD 600, MCU 502

    Application Area Partial Discharge Measurements

    Version v1.0

    Document ID ANP_10036_ENU

  • OMICRON 2013 Page 2 of 11

    Content

    1 Principle measurement cicuit ...............................................................................................................3

    2 Connection (decoupling) possibilities ................................................................................................4

    3 Test equipment ......................................................................................................................................6

    4 Test procedure .......................................................................................................................................7

    5 Calibration ..............................................................................................................................................8

    6 Measurement ..........................................................................................................................................9

    Please use this note only in combination with the related product manual which contains several important safety instructions. The user is responsible for every application that makes use of an OMICRON product. OMICRON electronics GmbH including all international branch offices is henceforth referred to as OMICRON. OMICRON 2010. All rights reserved. This application note is a publication of OMICRON.

    All rights including translation reserved. Reproduction of any kind, for example, photocopying, microfilming, optical character recognition and/or storage in electronic data processing systems, requires the explicit consent of OMICRON. Reprinting, wholly or in part, is not permitted.

    The product information, specifications, and technical data embodied in this application note represent the technical status at the time of writing and are subject to change without prior notice.

    We have done our best to ensure that the information given in this application note is useful, accurate and entirely reliable. However, OMICRON does not assume responsibility for any inaccuracies which may be present. OMICRON translates this application note from the source language English into a number of other languages. Any translation of this document is done for local requirements, and in the event of a dispute between the English and a non-English version, the English version of this note shall govern.

  • OMICRON 2013 Page 3 of 11

    1 Principle measurement cicuit The principle measurement circuit is described in the block diagram below. The coupling capacitor Ck is only one possibility to decouple the PD signal. Other ways of decoupling are shown in chapter 2.

    Figure 1: principle measurement circuit

    HV filter: High voltage filter to suppress noise / interferences from the high voltage source (e.g. resonance voltage source). Recommended for sources with high interferences

    Test object: object under test, e.g. cable, transformer, Ck: Coupling capacitor for decoupling the PD signal. CPL542: Measuring impedance for splitting up the voltage and the PD measurement path (incl.

    additional protection circuit) MPD600: Acquisition unit (measuring channel) USB502: fiber optical controller The connection(s) between the test object and the acquisition unit MPD 600 should be as short as possible to reduce unwanted interference. The length of the connection between the MPD 600 and the USB 502 can be as long as necessary (up to 2 km) because no additional interferences will be added due to the fiber optical transmission. If more than one measurement is needed, the channels can be easily cascaded via the optical bus.

  • OMICRON 2013 Page 4 of 11

    2 Connection (decoupling) possibilities The decoupling of the PD signal can be done on different ways depending on the test object and the onsite conditions. The most common ways are shown in the pictures below:

    Figure 2: connection to capacitive measurement taps of the transformer

    Figure 3: connection through coupling capacitors

  • OMICRON 2013 Page 5 of 11

    Figure 4: decoupling through RFCT's at the Figure 5: decoupling through RFCT's at the cable cross bonding box cable screen connection

    Figure 6: decoupling with capacitive sensor in the cable joint

  • OMICRON 2013 Page 6 of 11

    3 Test equipment Beside the decoupling equipment mentioned before, following items are necessary to perform the PD test:

    Figure 7: MPD 600 test equipment

    The charge range of the calibrator has to be selected in the range of the expected (allowed) charge level for the test object.

  • OMICRON 2013 Page 7 of 11

    4 Test procedure

    Figure 8: Test procedure

    Preparation: The prior information about the test object is necessary to select the right measurement equipment (calibrator, quadripole/measuring impedance, fiber optical cable length). The system will be set up the following way:

    > Connect the fiber optics controller to the Notebook Computer using a standard USB cable (USB 2.0).

    > Connect the MPD 600 Acquisition unit to the fiber optics controller using fiber optical cables.

    See wiring diagram on next page for details. Please observe that TX (transmit) is always

    connected to RX (receive) and vice versa.

    > Connect the power supply (or battery) to the MPD 600 Acquisition unit. The red LED starts flashing indicating that the acquisition unit is ready for operation.

    > Connect the external coupling unit CPL 542, also referred to as quadripole or measuring impedance.

    Use two short BNC cables and connect the PD and V outputs of the coupling unit

    CPL 542 to the PD and V inputs of the acquisition unit MPD 600.

    > Connect the coupling unit CPL 542 to a high voltage coupling capacitor (or RFCT, measuring tap,

    other Sensor). Always remember to keep the cable(s) as short as possible.

    > Connect the Notebook PCs power supply and switch it on. Perform test: Select an appropriate measurement frequency and bandwidth (in the range of the applicable measurement standard). This is to optimize the signal to noise ratio and to reduce external interferences. In case of high external disturbances a gaiting channel with antenna can be connected to the measuring system. The "perform test" steps are described in the next chapters. Reporting/assessment The reporting will be done based on the stored data. This can be done either with the exported graphics or with the stored data stream.

  • OMICRON 2013 Page 8 of 11

    5 Calibration After setting up the system a calibration of the complete test setup is essential:

    > Once the computer has started, double-click the Omicron Icon to start the software:

    > Use the mouse or keyboard to select the main menu item Mode in the top menu bar. Then

    select Basic for the systems operating mode.

    > Make sure that the Settings tab is selected. Find the combo box labeled Trigger source and

    select unit1.1 as the trigger source. If you are performing a calibration you may also select line

    as trigger source. The system then uses a light-sensitive sensor to synchronize itself to

    fluorescent light sources. Do not forget to change the trigger source back to unit 1.1 before

    performing the actual measurement.

    > Select the fCenter (center frequency) box and enter a value of 350 kHz.

    > Select a bandwidth of 300 kHz. This center frequency and bandwidth configuration sets up the

    measurement system for an IEC 602 70-compliant measurement. Other settings might be

    necessary to receive a better signal to noise ration

    > Make sure that the Input Protection option is checked. You may uncheck this option for very

    sensitive measurements but the PD input of the acquisition unit is then somewhat more sensitive

    to damage by electrical surges.

    > Go to the Settings tab. For Mode select Unipolar Logarithmic.

    > Set QMax (maximum charge displayed) to 1 nC.

    > Set QMin (minimum charge registered and displayed) to 1 pC.

    > Optionally, check the Show individual PD events option. Individual PD events will then be shown

    as phase and amplitude resolved spikes in real time. Now change to the calibration settings by selecting the Calibration tab.

  • OMICRON 2013 Page 9 of 11

    Figure 9: Omicron Software, Calibration Tab

    > Uncheck the Enable test generator option.

    > Check the External quadripole configuration option.

    Connect the calibrator to the test object. Select a charge of 100 pC. Pulses should clearly extend from the background noise as spikes. If the noise floor covers PD pulses, select a higher charge on the calibrator (depending on your test object and available charge calibrator you may use another charge level).

    > Enter the calibrator charge (e.g. 100 pC) into the Calibrate to field in the PD section.

    > Click on the Compute button in the PD section. The display value for QIEC should now

    correspond to the calibrator charge. Now the charge value is calibrated. The second step is to calibrate the test voltage value:

    > Disconnect the calibrator.

    > Secure the high voltage test set-up observing safety procedures.

    > Turn on the high-voltage supply. Set voltage to a known level (e.g. 2 kV).

    > Enter voltage level (e.g. 2 kV) into the Calibrate to field in the Voltage section.

    > Click on the Compute button in the voltage section. The display value for voltage should now

    correspond to the actual voltage.

    6 Measurement After the calibration is finished, the PD measurement can take place. The measurement should be performed according to the applicable standard for this type of test and test object. Main issues of the standards are test voltage levels, test duration and maximum PD levels. To get a so called "pattern" of the partial discharge, the creation of a histogram is useful:

  • OMICRON 2013 Page 10 of 11

    > Start the test voltage run (of the test voltage source)

    > Click on the Go button in the "Settings" tab. A histogram will be accumulated over a specified

    time period. The progress bar at the bottom will indicate the progress of this operation.

    > Once a histogram has been accumulated, the image may be saved to disk by right-clicking on the

    scope display. A context-menu will pop up and offer to save the image to disk as a PNG file. PNG

    is a popular image format providing lossless compression.

    > All of the data gathered during a measurement may be recorded for later analysis by clicking on

    the Record button (marked by a large red dot). The software will display the hard disk space

    consumed during this operation and the remaining disk space. The stored data can be used afterwards to assess the test object, to create a test report and to give recommendations.

  • OMICRON is an international company serving the electrical power

    industry with innovative testing and diagnostic solutions. The application of

    OMICRON products provides users with the highest level of confidence in

    the condition assessment of primary and secondary equipment on their

    systems. Services offered in the area of consulting, commissioning,

    testing, diagnosis, and training make the product range complete.

    Customers in more than 130 countries rely on the company's ability to

    supply leading edge technology of excellent quality. Broad application

    knowledge and extraordinary customer support provided by offices in

    North America, Europe, South and East Asia, and the Middle East,

    together with a worldwide network of distributors and representatives,

    make the company a market leader in its sector.

    Europe, Middle East, Africa

    OMICRON electronics GmbH

    Oberes Ried 1

    6833 Klaus, Austria

    Phone: +43 5523 507-0

    Fax: +43 5523 507-999

    [email protected]

    Asia-Pacific

    OMICRON electronics Asia Limited

    Suite 2006, 20/F, Tower 2

    The Gateway, Harbour City

    Kowloon, Hong Kong S.A.R.

    Phone: +852 2634 0377

    Fax: +852 2634 0390

    [email protected]

    Americas

    OMICRON electronics Corp. USA

    12 Greenway Plaza, Suite 1510

    Houston, TX 77046, USA

    Phone: +1 713 830-4660

    +1 800-OMICRON

    Fax: +1 713 830-4661

    [email protected]

    For addresses of OMICRON offices with customer service

    centers, regional sales offices or offices for training,

    consulting and commissioning please visit our web site. www.omicron.at www.omicronusa.com