Page 1: More CXR…

MORE CXR…

Helen MurphySCC Allied Health

Radiology Program

Page 2: More CXR…

Views: PA Upright/ Left Lateral Upright (already talk about)

AP upright chest

Supine chest

Cross-Table Left Lateral

AP/PA Lordotic chest

Decub Chest

Portable Chest

Dual Energy Chest (SHMC)

Page 3: More CXR…

On all CXRS…. The lung markings, diaphragms, heart borders, hilum,

greater vessels, and bony cortical outlines are sharply defined. So use high KVP (sufficient to penetrate the chest structures and provides the contrast scale necessary to visualize the lung details.) and a grid ( to absorb scatter).

Sharply defined recorded details are also obtained when patient respiration and body movements are halted(So NO MOTION!!!)

The least amount of object “image receptor distance (OID) is maintained. This why PA is better than AP (patient’s heart is closer to IR).

Use a 72-inch (183-cm) source “image receptor distance” (SID) to decrease the magnification of the heart.

Page 4: More CXR…

72 inch SID 40 inch SID

Page 5: More CXR…

The time of examination, degree of patient elevation( supine semi-upright…) and if done AP should be on the image.

Indicating the time of day on the image is especially important if the patient's progress is being followed and multiple chest images are to be taken on the same day. This is done automatically on CR and DR( but is the time right?). Knowledge of the degree of elevation helps the reviewer determine the exact amount of fluid in the patient's lungs.

CONT…

Page 6: More CXR…

Structures shown: Entire lung field

An AP projection is somewhat similar to the PA projection. AP being farther from the IR, the heart and great vessels are magnified as well as engorged, and the lung fields appear shorter because abdominal compression moves the diaphragm to a higher level. The clavicles are projected higher, and the ribs assume a more horizontal appearance

Page 7: More CXR…

PA AP

Scapulas are in the way!

Page 8: More CXR…

Upright Supine

Heart appears larger

Page 9: More CXR…

Portable Chest: AP Upright

Page 10: More CXR…

Why take a portable CXR? Patient is not stable or too sick to come

down stairs Check for position of lines and tubes Check for complications from lines and

tubes Sudden onset of chest pain and or

shortness of breath To check for a Pneumothorax (upright is

best!) Air embolism from line

Page 11: More CXR…

Another way to show air-fluid levels: To demonstrate precise air-fluid levels when pleural effusion is suspected, chest images should be taken with the patient upright and the x-ray beam horizontal. With this position, the air rises and the fluid gravitates to the lowest position, creating an air-fluid separation. This separation is identified as a decrease in density on the image wherever the dense fluid is present in the lung field. The true amount of fluid cannot be discerned on an image unless the fluid is level; in the slanted position the chest may appear to have no fluid. When the patient is supine, the fluid is evenly spread throughout the lung field, preventing visualization of fluid levels.

Page 12: More CXR…

The same patient with tube angle and without

WithoutWith 25 degrees

Page 13: More CXR…

Please: Remove all patient monitoring lines that can be

removed or shift them out of the lung field! When patient monitoring lines remain within the lung field, they may obscure lung details or the one line we are looking for.

Use a grid whenever possible, especially on larger patients. A grid is not always employed in portable imaging. Why? because it is difficult to ensure that the grid and central ray are aligned accurately, which could lead to grid cutoff. When no grid is used, a lower kVp technique is needed to prevent excessive scatter radiation from reaching the IR and hindering contrast. Although the lower kVp will sufficiently penetrate the lung field, it seldom provides enough penetration to allow visualization of structures within and behind the heart shadow

Page 14: More CXR…

Same patient demonstrating a fluid-type pathologic condition

Non gridGrid

Page 15: More CXR…

When a chest x-ray is used to evaluate the placement of apparatus positioned within the mediastinal region, the heart shadow should be penetrated. The accurate placement of these lines cannot be evaluated without heart penetration. Accomplish this penetration by increasing the kVp. The resulting image will demonstrate a penetrated heart shadow with the thoracic vertebrae, posterior ribs, and chest lines, clearly demonstrated through it. The amount of scatter radiation reaching the IR will also increase, resulting in overall lower image contrast (Gray &UGLY!!) So grid use would be helpful!!!!

Page 16: More CXR…
Page 17: More CXR…

X-Table Lateral Chest

Page 18: More CXR…

X-table Lateral

Page 19: More CXR…

Just like a lateral chest just done x-table Still need the entire lung field on Use for post pacemaker patients and

babies

Page 20: More CXR…

Pulmonary Apices Positions:o AP axial projection -Lordotic position -Lindbolm

method (pt leans backwards to the film)

o AP axial projection ( pt AP angle tube 15-20 up)

o PA axial projection ( pt PA angle tube 15-20 up)

PA axial projection- Lordotic position- Fleischer method (pt leans backwards away from film)

Not shown (no pictures in book)

Page 21: More CXR…

Structures shown:

Demonstrate the apices free from superimposion of clavicles.

Page 22: More CXR…

AP Axial Lordotic Position

Page 23: More CXR…

AP Axial

Page 24: More CXR…

PA axial

Page 25: More CXR…

Positions:

Rt lateral Decubitus (right side down)

Lt lateral Decubitus (left side down)

*For fluid: place patient on affected side

* For air: place patient on the unaffected side

*Trendelenburg helps show fluid levels better

* Can be done AP or PA

Page 26: More CXR…

What does it show:

Demonstrates the change in fluid position and reveals any previously obscured pulmonary areas or, in the case of pneumothorax (air or gas in the pleural space), the presence of any free air.

Done on pts with chest tubes to check fluid levels

Page 27: More CXR…

Right lateral decub chest xray

Page 28: More CXR…

AP projection, left lateral decubitus position .The arrows indicate the air-fluid level (air on the side up).

AP projection, right lateral decubitus position. Arrows showing a fluid level (the down side) . Note the fluid in the lung fissure (arrowhead)

I would show both sides

Page 29: More CXR…

Obliques of the Chest

RAO/LAO: 45 degree obliques demonstrate the side farthest from the IR (LAO shows right side)

RPO/LPO:45 degree obliques demonstrate the side closest from the IR

(LPO shows the left side)

Page 30: More CXR…

LAO Chest : Demonstrate the side farthest from the IR (right side)

RT side demonstrated

Page 31: More CXR…

LPO Chest: demonstrate the side closest from the IR (left side)

Left side is demonstrated

Page 32: More CXR…

Dual energy Chest

Done at SHMC on all PICC Line patients One exposure (long one) but three

images! First image normal chest Second image ribs blurred Third image just bones

Page 33: More CXR…

1 2

3

Page 34: More CXR…

No dual energy try inverting the image (image processing)

Page 35: More CXR…

Stuff in the chest

Page 36: More CXR…

Some lines found in the in chest: Central venous (CV line) catheters also known as Peripherally inserted central catheters (PICC) or Ports , pulmonary arterial line also known as a Swan-Ganz catheter

Page 37: More CXR…

o Central venous catheters are inserted into the subclavian vein or a more peripheral vein in the upper extremity are extremely useful for measurement of the central venous pressure (CVP) and for providing a conduit for the rapid infusion of fluid or chronic hyperalimentation. They allow for infusion of substances that are too toxic for peripheral infusion, such as for chemotherapy, total parenteral nutrition, dialysis, or blood transfusions.

o So that the CVP may be correctly measured, the catheter must be located with in the true central venous system, beyond all the valves, which interfere with direct transmission of the right arterial pressure to the catheter.

o We would like it to be where the brachiocephlic veins join to form the superior vena cava or within the superior vena cava itself approximately 2 to 3 cm above the right atrial junction

o Medial to the anterior border of the first rib

Page 38: More CXR…

Tip Here

Page 39: More CXR…

Images taken for CV line placement should demonstrate adequate density and penetration to visualize the line and lung conditions that may result if perforation occurs during line insertion, such as pneumothorax or hemothorax.

Page 40: More CXR…

Complications:

Pneumothorax Infections Air embolism Hemorrhage Arrhythmia

Page 41: More CXR…

pneumothorax

Page 42: More CXR…

Catheter with it’s tip in the pleural space. A right subclavian catheter, perforated the superior vena cava and eroded into the right pleural space. Note the tip of the catheter projecting beyond the right border of the mediastinum (arrow). The direct infusion of parenteral fluid into the pleural space has led to a large right hydrothorax

This is why it is very important to have a x-ray before use!

Page 43: More CXR…

Catheter in the right internal jugular vein.

Page 44: More CXR…

Broken PICC catheter. The sheared-off portion of the catheter (arrow) is located in the left lower lobe.

Page 45: More CXR…

Pulmonary arterial line (Swan-Ganz catheter): The pulmonary arterial line is similar to the Central venous line but is longer. It is used to measure atrial pressures, pulmonary artery pressure, and cardiac output. The measurements obtained are used to diagnose ventricular failure and monitor the effects of specific medication, exercise, and stress on heart function. The pulmonary arterial line is inserted in the subclavian, internal or external jugular, or femoral vein and is advanced through the right atrium into the pulmonary artery . Images taken for pulmonary arterial line placement should demonstrate adequate density and penetration to visualize the line and mediastinal structures to determine adequate placement and lung conditions that may result if perforation occurs during line insertion, such as pneumothorax or hemothorax.

Page 46: More CXR…

AP CXR demonstrating accurate pulmonary arterial(PA) line

placement.

Page 47: More CXR…

More things put into a chest: Endotracheal tube (ET): The ET is a stiff, thick-walled tube that

is used to inflate the lungs: They are inserted through the mouth into the trachea as a means of establishing or opening an air way

It should be 5-7 cm above the tracheal bifurcation(carina) Why take a cxr? To check for proper placement of tube and to

check for pneumonia or pneumothorax 20% go into right main bronchus this will cause collapse of the

left lung A tube to high will cause air to enter the stomach, this could

cause regurgitation of gastric contents into the lungs, which could lead to aspiration pneumonia.

Films taken daily to check for movement Patient could get a tension pneumothorax from the pressure of

the ventilation machine.

Page 48: More CXR…

Distal end of endotracheal tube with cuff deflated (1)

C, Pediatric endotracheal tube; note the absence of cuff.

A, Adult endotracheal tube.

side hole

end holeTracheotomy tube.(

Page 49: More CXR…

Anteroposterior mobile chest projection showing trachea (dotted line)

nasogastric tube

cardiac monitor wires

carina

endotracheal tube

Page 50: More CXR…

Carina

Page 51: More CXR…

Proper position of the endotracheal tube

5-7 cm above Carina orOld school 2 inches superior to the carina

Page 52: More CXR…

Poor placement :Intubation of right main-stem bronchus with complete occlusion of the left bronchus causing left lung atelectasis.

Page 53: More CXR…

Check for:

Head and neck in the neutral position

With flexion or extension of the neck the tube will move about 2cm

The patient is not rotated

The patient is breathing okay

Page 54: More CXR…

Still more things put into the chest: Thoracostomy tube (chest tubes)

Page 55: More CXR…

○ Thoracostomy tube (chest tube): The chest tube is a 1.25-cm diameter, thick-walled tube that is used to remove fluid or air from the pleural space when it is preventing negative pressure in the intrapleural space from aiding in lung expansion. Failure to remove the fluid or air may result in collapse of the lung. For drainage of air the tube is placed anterosuperiorly within the pleural space, typically at the level of the second or third intercostal space at the midclavicular line . For drainage of fluid the tube is typically placed within the pleural space laterally at the midaxillary line at the level of the fifth or six intercostal space. The side hole of the thoracostomy tube, which is marked by an interruption of the radiopaque identification line, should be placed within the thoracic cavity, medial to the inner ribs. Images taken for (chest tubes) thoracostomy tube placement should demonstrate adequate density and penetration to visualize the radiopaque identification line.

Page 56: More CXR…

Chest tubes radiopaque identification line

Page 57: More CXR…

Pacemaker: The pacemaker is used to regulate the heart rate by supplying electrical stimulation to the heart. This electrical signal will stimulate the heart the needed amount to maintain an effective rate and rhythm. On a PA or anteroposterior (AP) chest image the pacemaker catheter tip should be seen at the apex of the right ventricle . Care should be taken when positioning a patient whose pacemaker was inserted within 24 hours of the examination. Elevation of the left arm should not be done unless surgeon or nurse have given there okay, this is done to prevent dislodging of the pacemaker and catheter. Two views should be done to show tip placement. On a frontal radiograph, the tip often appears to be well positioned. A lateral projection is required to show that the tip is directly posterior in the coronary sinus, rather than in its proper position anterior in the right ventricle. They can be done supine (x-table lateral) if patient is still to sleepy from surgery.

Post pacemaker Left arm

Page 58: More CXR…

Fracture of a cardiac pacemaker wire (arrow)

Page 59: More CXR…

Perforation of the myocardium by an intravenous electrode usually occurs at the time of insertion or during the first few days thereafter. Perforation should be suspected when the pacemaker fails to sense or elicit a ventricular response. Plain radiographs show the electrode tip lying outside the right ventricular cavity

Page 60: More CXR…
Page 61: More CXR…