Module 11

Compliance with established quality

assurance requirements

1

Page 2: Module 11 Compliance with established quality assurance requirements 1

Learning objectives

At the end of this module you will be able to:explain the three main components of the QA system and their importance;adhere to the QC and EQA procedures.

2

Page 3: Module 11 Compliance with established quality assurance requirements 1

Content outline

• Quality assurance (QA) in mycobacteriology:– internal quality assurance (IQA) or quality control

QC– external quality assessment (EQA)– quality improvement (QI)

3

Page 4: Module 11 Compliance with established quality assurance requirements 1

Definitions

• Quality assurance (QA) System designed to continuously improve the

reliability and efficiency of laboratory services. Includes internal quality control, external quality assessment, and quality improvement.

• Quality control (QC)Includes all means by which the TB laboratory

controls operations (instrument checks, checking new lots of staining solutions, reagents, media).

4

Page 5: Module 11 Compliance with established quality assurance requirements 1

Definitions

• External quality assessment (EQA):– participating laboratories assess their

capabilities through panel testing; – includes on-site evaluation of the laboratory.

5

Page 6: Module 11 Compliance with established quality assurance requirements 1

Definitions

• Quality improvement (QI): – to permanently remove obstacles to success;– involves continued monitoring, identifying defects;– followed by remedial action (e.g. retraining);

– often relies on effective on-site evaluation visits.

6

Page 7: Module 11 Compliance with established quality assurance requirements 1

System designed to improve:– reliability– efficiency– useof laboratory services in order to achievethe required technical quality in laboratorydiagnosis.

Quality assurance in mycobacteriology

7

Page 8: Module 11 Compliance with established quality assurance requirements 1

The process of quality assurance should be continuous and monitored

• Central laboratories

• Intermediate laboratories

• Peripheral laboratories

Organized according to existing laboratory network levels:

8

Page 9: Module 11 Compliance with established quality assurance requirements 1

Quality assurance

• Process of effective and systematic monitoring of all laboratory activities– establishes limits of acceptable standard of

test performance;– ensures laboratory data accuracy, reliability,

reproducibility, comparability;– not for punitive purposes!

QA is the responsibility of all laboratory technicians and supervisors.

9

Page 10: Module 11 Compliance with established quality assurance requirements 1

Quality assurance should be:

• Practical

• Workable

• Comprehensive

10

Page 11: Module 11 Compliance with established quality assurance requirements 1

QA should be applied to:

• Laboratory arrangement• Human resources (training, health control)• Laboratory equipment• Collection and transport of specimens• Handling of specimens• Reagents and media• Culture procedures• Reporting of results

11

Page 12: Module 11 Compliance with established quality assurance requirements 1

QA – laboratory arrangement

• BSL2/BSL3• Doors in the laboratory to be kept

closed.• Work areas, equipment and supplies

arranged for logical and efficient work-flow (clean to dirty).

• Benches swabbed at least once a day with an appropriate disinfectant.

12

Page 13: Module 11 Compliance with established quality assurance requirements 1

QA – laboratory administration

• Use Standard Operating Procedures (SOPs).

• Keep written procedures in the laboratory for easy reference.

• Any changes must be dated and initialled by the laboratory supervisor.

• Retain all records for 2 years.

13

Page 14: Module 11 Compliance with established quality assurance requirements 1

QA – human resources

• Training (retraining)– before starting– refreshment – in case of failure

• Adequate number

14

Page 15: Module 11 Compliance with established quality assurance requirements 1

Training form (example for

NTP)

15

Page 16: Module 11 Compliance with established quality assurance requirements 1

QA – laboratory equipment

• Equipment should meet the manufacturer’s claims and specifications.

• Equipment should be calibrated.• Dated service records must be kept for all

equipment.• Equipment must be monitored regularly to

ensure constant accuracy and necessary precision.

16

Page 17: Module 11 Compliance with established quality assurance requirements 1

QC – biological safety cabinet

Daily checks:•Minimal airflow: 0.38–0.51 metre/second (75–100 feet/minute) depending on (sub)-classification. •Check the magnehelic gauge in the exhaust duct for any pressure drop across the filters;  replace the filters when the gauge indicates that the airflow across the front opening has dropped below optimal levels.

17

Page 18: Module 11 Compliance with established quality assurance requirements 1

QC – other equipment• Refrigerator 2–8 ºC / Freezer – 20°C

– record temperature daily;– clean monthly;– defrost or check refrigerator and freezer

compartment every 3 months.

• Glassware: – discard chipped or etched glassware;– ensure that glassware is free of detergents;– do not store sterile glassware for more than

3 weeks before it is used.

18

Page 19: Module 11 Compliance with established quality assurance requirements 1

QC model for

monitoring temperature of equipment

Trimester : Year :

Month Temp. Visa Month Temp. Visa Month Temp. Visa

°C (Initials) °C (Initials) °C (Initials)

1 1 1

2 2 2

3 3 3

4 4 4

5 5 5

6 6 6

7 7 7

8 8 8

9 9 9

10 10 10

11 11 11

12 12 12

13 13 13

14 14 14

15 15 15

16 16 16

17 17 17

18 18 18

19 19 19

20 20 20

21 21 21

22 22 22

23 23 23

24 24 24

25 25 25

26 26 26

27 27 27

28 28 28

29 29 29

30 30 30

31 31 31

19

Page 20: Module 11 Compliance with established quality assurance requirements 1

QA – collection and transport of specimens

• Process specimens only in response to written requests – do not consider oral requests.

• Insist on specimen request forms being kept separate from the specimens.

• Insist on  adequately completed request forms and proper labelling of specimens.

• Differentiate specimens for diagnosis from those for control of treatment. Lack of this information rules out the use of monitoring of bacteriological results as a method of IQA.

20

Page 21: Module 11 Compliance with established quality assurance requirements 1

QC – handling of specimens

• Process  sputum specimens in batches according to centrifuge capacity.

• CAUTION : If delivery of specimens is delayed, contamination rate will increase

21

Page 22: Module 11 Compliance with established quality assurance requirements 1

QC – reagents and media

• Discard media that are discoloured or have bubbles following inspissation.

• Check all batches of media for sterility by incubation at 35–37 ºC for 48 hours.

• Keep all media in the dark in the refrigerator and discard unused drug-containing media after 4 weeks.

• For culture and DST media: check with M. tuberculosis H37Rv.

• IQC for identification test with M. terrae and M. tuberculosis H37Rv

22

Page 23: Module 11 Compliance with established quality assurance requirements 1

QC – preparation of egg-based media

Sensitivity

• The numbers of colonies obtained by inoculating 5 tubes from a previous batch of media and 5 tubes from a newly prepared/purchased batch should be comparable.

Sterility

• No bacterial growth should be present 48 hours after the incubation at 37 ºC of each new batch.

Storage

• Avoid the accumulation of expired media.

• Estimate the number of tubes to be prepared or purchased to cover culturing needs for about 2 months of work.

• Store media in a cool and humid place, preferably in a refrigerator at 4 ºC.

23

Page 24: Module 11 Compliance with established quality assurance requirements 1

Media register

Batchno.

Volume(ml)

Date of: Sterility control

Sensitivity control

Preparationor

purchasedelivery

Start of use

End of

use

ContaminationdetectedYes/No

Strain/ specimen inoculated

Growth detected at20 days

(no. of colonies)

Growth detected at60 days

(no. of colonies)

Control batch

Present batch

Control batch

Present batch

24

Page 25: Module 11 Compliance with established quality assurance requirements 1

QC – media for DST

H37Rv

MIC expected

INH

0.06 µg/ml

RIF

4.0

µg/ml

EMB

0.5

µg/ml

SM

2.0

µg/ml

Signature

MIC

Observed

Date of batch production: __________ Date:___________

25

Additional concentrations tested (µg/ml)INH 0.025•0.05•0.10 RIF 2.5•5.0•10.0 EMB 0.125•0.25•0.5SM 0.5•1.0•2.0

Page 26: Module 11 Compliance with established quality assurance requirements 1

QC – culture procedures: decontamination

• Use aliquoted reagent solutions.• Do not keep all tubes open at the same time.• Do not reuse the reagent solution aliquots opened

during the work day. • The total time of contact of the sample with the

decontaminant must be strictly controlled. Too short a time results in high contamination rates, too long a time causes loss of viability of the bacilli.

• Be suspicious of several successive positive specimens or of cultures with few colonies following a heavily positive culture.

26

Page 27: Module 11 Compliance with established quality assurance requirements 1

QC – reporting of results

• Always check for agreement between the laboratory register number and the number written on tubes/slants.

• Use WHO forms for reporting results.• A copy of the completed final report

should be retained in the laboratory for 2 years.

27

Page 28: Module 11 Compliance with established quality assurance requirements 1

Reporting – turn-around time• Smear from specimen (fluorochrome or ZN) –

report positive or negative and the staining method used, preferably with no more than 24 hours delay from the moment of receipt of specimen in the laboratory.

• Culture contaminated: report immediately and a repeat specimen requested.

• Culture-positive and identification of M. tuberculosis: report immediately.

• Upon completion of the drug susceptibility test – susceptible or resistant to each test drug.

28

Page 29: Module 11 Compliance with established quality assurance requirements 1

Reporting – turn-around time

• Culture-negative – upon completion of the incubation protocol (6 weeks for liquid, 8 weeks for solid).

• For solid media:- at 4 weeks, send an interim report on all negative stating that another report will be issued in the event of the specimen becoming positive later on- at 8 weeks, issue a final report

29

Page 30: Module 11 Compliance with established quality assurance requirements 1

Daily monitoring – alarm signals

Smear-positive/culture-negative specimens

• What are the possible explanations?• What are the corrective measures?

Contaminated specimens

• What are the possible explanations?• What are the corrective measures?

30

Page 31: Module 11 Compliance with established quality assurance requirements 1

Alarms and remedial actionsSmear-positive/culture-negative specimens • Follow-up?• If for diagnosis:

– Does this type of result recur?– Check the quality of the media being used.

– Concentration of decontaminating solution? Time of contact with

specimens? Temperature of incubator? Contaminated specimens• Time between specimen collection and processing too long?

– Corrective measures needed in specimen transportation system.

• Contamination of specimens from the same patient?– Use a harsher decontamination procedure for testing further

specimens from the patient. 31

Page 32: Module 11 Compliance with established quality assurance requirements 1

Daily monitoring – Alarm signals

Recurrent contamination

• What are the possible explanations?• What are the corrective measures?

Clustering of culture-positive specimens

• What are the possible explanations?• What are the corrective measures?

32

Page 33: Module 11 Compliance with established quality assurance requirements 1

Alarms and remedial actionsRecurrent contamination• Check sterility of decontamination reagent solutions.

• Check the whole decontamination process.

• If the problem is traced to the technician performing the procedure, he or she should be immediately retrained.

Clustering of culture-positive specimens• Check contamination of decontamination solution

• Check that:– aliquoted reagent solutions are discarded after single use;– processing sequence of specimens is maintained, i.e.

smear-positive specimens processed last;– tubes are not uncapped immediately after removal from

centrifuge;– supernatants are discarded carefully. 33

Page 34: Module 11 Compliance with established quality assurance requirements 1

Periodic monitoring

• EQA for whole culture process NOT feasible.

• Necessary to evaluate indicators monthly.

• DO NOT inoculate H37Rv in specimens to evaluate the culture process.

34

Page 35: Module 11 Compliance with established quality assurance requirements 1

Periodic monitoring – indicators

Classification of specimens from adult pulmonary TB patients investigated for diagnosis (not for follow-up):

a. Smear-positive and culture-positive

b. Smear-positive and culture not done

c. Smear-negative and culture-positive

d. Smear-positive and culture-negative

e. Smear-positive and culture contaminated

f. Smear not done and culture-positive

35

Page 36: Module 11 Compliance with established quality assurance requirements 1

Periodic monitoring – indicators

c + fContribution of culture to diagnosis = ---------------------------- x 100

a + b + c + d + e + f

Expected contribution: 20%

Diagnosis of adult pulmonary cases – for a specific setting

36

Page 37: Module 11 Compliance with established quality assurance requirements 1

Periodic monitoring – indicators

c Contribution of culture to diagnosis = ------------------------- x 100over microscopy a + c + d + e

Expected contribution: >20%

(higher with liquid culture)

Diagnosis of adult pulmonary cases – for a specific setting

37

Page 38: Module 11 Compliance with established quality assurance requirements 1

Periodic monitoring – indicators

dPercentage of smear-positive = --------------------- x 100and culture-negative cases a + c + d + e

Expected: 2–3%

38

Page 39: Module 11 Compliance with established quality assurance requirements 1

Periodic monitoring – indicators

contaminated tubesPercentage contamination = ----------------------------- x 100

all inoculated tubes

Expected: < 5% for solid

<10% for liquid

39

Page 40: Module 11 Compliance with established quality assurance requirements 1

Alarms: what to investigate

Indicators of culture performance

Normal value(%)

Much higher:

investigate

Much lower: investigate

Contribution of culture to bacteriological diagnosis of tuberculosis

20 A B and C

Percentage of smear-positive/ culture-negative specimens

2–3 C and D Not a problem

Percentage of contaminated tubes < 5% (solid)< 10% (liquid)

E F

40

Page 41: Module 11 Compliance with established quality assurance requirements 1

DST QC and indicators

– IQC: Before use, new batches of media should be tested with a reference strain

– Alarms:

– change in patterns of resistance;

– unusual resistance rate to a drug e.g. monoresistance to RMP higher than monoresistance to INH

41

Page 42: Module 11 Compliance with established quality assurance requirements 1

EQA for culture :• EQA for culture procedure:

Indirect external supervision based on alarm signals, see above (slide 40) for QC

• EQA of culture media

If low sensitivity in two consecutive controls:The laboratory is requested not to use or distribute culture media until the deficiencies are overcome.

Culture media are offered so that the laboratory can continue its activity

Retraining in culture media preparation procedures should also be offered, if needed.

42

Page 43: Module 11 Compliance with established quality assurance requirements 1

International EQA for DST

• EQA of NRLs.

• Blinded test of a panel of 20 strains from SRLs.

• Parameters evaluated: sensitivity, specificity, reproducibility for each drug.

• Minimum agreement of 90% for at least isoniazid and rifampicin.

• If not reached, re-evaluate crucial points: drugs, media, procedures, supervision visit.

43

Page 44: Module 11 Compliance with established quality assurance requirements 1

National EQA for DST

• EQA of national laboratories performing culture.

• Blinded test of the SRL panel of 20 strains from the NRL.

• Parameters evaluated: sensitivity, specificity, reproducibility for each drug.

• Minimum agreement of 90% for at least isoniazid and rifampicin.

• If not reached, re-evaluate crucial points: drugs, media, procedures, supervision visit.

44

Page 45: Module 11 Compliance with established quality assurance requirements 1

Quality improvement

It is not sufficient in IQC simply toidentify errors or weaknesses in laboratoryservices; remedial action must be taken to

permanently eliminate them.

This implies continuous monitoring ofperformance, an IQC programme, as well as

a direct supervision programme.

45

Page 46: Module 11 Compliance with established quality assurance requirements 1

Keys to successful quality control

• Adequately trained, interested and committed staff.

• Common-sense use of practical procedures.

• Willingness to admit and rectify mistakes.

• Effective communication.

46

Page 47: Module 11 Compliance with established quality assurance requirements 1

True and false take-home messages

1. Quality assurance is a system designed to discover and punish people who don’t work well.

2. It is useful periodically to calculate indicators to evaluate the performance of the laboratory.

47

Page 48: Module 11 Compliance with established quality assurance requirements 1

Module review: take-home messages

A quality assurance (QA) programme is essential for improving the reliability, efficiency and use of laboratory services in order to achieve the required technical quality in laboratory diagnosis.

The process of quality assurance should be continuous and monitored

Quality assurance is the responsibility of all laboratory technicians and supervisors.

It is useful periodically to calculate indicators to evaluate the performance of the laboratory.

It is not sufficient in QC simply to identify errors or weaknesses in laboratory services; remedial action must be taken to permanently remove them.

48

Page 49: Module 11 Compliance with established quality assurance requirements 1

• Explain the three main components of the QA system.

• Which components of the laboratory workflow should be checked for QC?

Self-assessment

49