module 1,, part 4: qc-related validation slide 1 of 28 © who – edm – 1/2002 validation part 4:...
Post on 22-Dec-2015
247 views
TRANSCRIPT
Module 1,, Part 4: QC-related validation Slide 1 of 28 © WHO – EDM – 1/2002
Validation Part 4:
QC-related validation
Supplementary Training Modules on Good Manufacturing Practice
Module 1,, Part 4: QC-related validation Slide 2 of 28 © WHO – EDM – 1/2002
Validation
Introduction
Why is analytical monitoring necessary?
What is the purpose of analytical validation?
Module 1,, Part 4: QC-related validation Slide 3 of 28 © WHO – EDM – 1/2002
Validation
Objectives
To introduce the concepts of : Protocol development
Instrument qualification
Analytical procedure
Extent of validation
Method transfer
Chemical and physical, biological, and microbiological test validation
Module 1,, Part 4: QC-related validation Slide 4 of 28 © WHO – EDM – 1/2002
Validation of analytical procedures requires: Qualified and calibrated instruments
Documented methods
Reliable reference standards
Qualified analysts
Sample integrity
Validation
Module 1,, Part 4: QC-related validation Slide 5 of 28 © WHO – EDM – 1/2002
Validation protocol for analytical method Statement of purpose and scope Responsibilities Documented test method List of materials and equipment Procedure for the experiments for each parameter Statistical analysis Acceptance criteria for each performance parameter
Validation
Module 1,, Part 4: QC-related validation Slide 6 of 28 © WHO – EDM – 1/2002
Qualification of the instrument
Make, model and maker’s manual
Modifications
Installation and operational qualification
Calibration programs
Maintenance schedules
Validation
Module 1,, Part 4: QC-related validation Slide 7 of 28 © WHO – EDM – 1/2002
Characteristics of analytical procedures (1) Accuracy
Precision
Repeatability
Reproducibility
Validation
Module 1,, Part 4: QC-related validation Slide 8 of 28 © WHO – EDM – 1/2002
Inaccurate &imprecise
Inaccurate butprecise
Accurate butimprecise
Validation
Relationship between accuracy and precision
Accurate AND Precise
Module 1,, Part 4: QC-related validation Slide 9 of 28 © WHO – EDM – 1/2002
Characteristics of analytical procedures (2)
Ruggedness Robustness Variability caused by:
Day-to-day variations Analyst-to-analyst Laboratory-to-laboratory Instrument-to-instrument Chromatographic column-to-column Reagent kit-to-kit Instability of analytical reagents
Validation
Module 1,, Part 4: QC-related validation Slide 10 of 28
© WHO – EDM – 1/2002
Characteristics of analytical procedures (3)
Linearity and range
Specificity
Sensitivity
Limit of detection
Limit of quantitation
Validation
Module 1,, Part 4: QC-related validation Slide 11 of 28
© WHO – EDM – 1/2002
Linearity of an analyte in a material
0.010
0.015
0.020
0.025
0.030
0.035
0.040
0.01 0.015 0.02 0.025 0.03 0.035 0.04
Reference material mg/mlCalculated analyte in mg/mL
Table of values (x,y)
x y # Reference
material mg/mlCalculated
mg/ml
1 0.0100 0.0101
2 0.0150 0.0145
3 0.0200 0.0210
4 0.0250 0.0260
5 0.0300 0.0294
6 0.0400 0.0410
Validation
Module 1,, Part 4: QC-related validation Slide 12 of 28
© WHO – EDM – 1/2002
Linearity Statistics Intercept -0.0002
Limit of Linearity and Range 0.005 – 0.040 mg/mL
Slope 1.0237
Correlation coefficient Pearson 0.9978 Olkin and Pratt 0.9985
Relative procedure standard deviation 3.4%
Validation
Module 1,, Part 4: QC-related validation Slide 13 of 28
© WHO – EDM – 1/2002
LOQ, LOD and SNR
Limit of Quantitation
Limit of Detection
Signal to Noise Ratio
noise
Peak ALOD
Peak BLOQ
Baseline
Validation
Module 1,, Part 4: QC-related validation Slide 14 of 28
© WHO – EDM – 1/2002
Different classes of analytical tests
Class A: To establish identity
Class B: To detect and quantitate impurities
Class C: To determine quantitatively the concentration
Class D: To assess the characteristics
Validation
Module 1,, Part 4: QC-related validation Slide 15 of 28
© WHO – EDM – 1/2002
* A degree of bias may be allowed
Characteristic A B quant.
B Limit test
C D
Accuracy
X X X*
Precision X X X
Robustness X X X X X
Linearity and range
X X X
Specificity X X X X X
Limit of detection X
Limit of quantitation
X
Validation
Module 1,, Part 4: QC-related validation Slide 16 of 28
© WHO – EDM – 1/2002
Extent of validation
New methods require complete validation
Pharmacopoeial methods require partial validation (or verification)
Significant changes mean partial revalidation equipment changes formula changed changed suppliers of critical reagents
Validation
Module 1,, Part 4: QC-related validation Slide 17 of 28
© WHO – EDM – 1/2002
Analytical method transfer Method transfer protocol and procedure
precision accuracy ruggedness
Written and approved specific test method
Proficiency check
Formal acceptance by new laboratory
Validation
Module 1,, Part 4: QC-related validation Slide 18 of 28
© WHO – EDM – 1/2002
Chemical laboratory validation requirements (1)
Balances Chromatography
HPLC, HPTLC, GC, TLC Dissolution or disintegration apparatus Karl Fischer moisture determination Melting, softening or freezing point apparatus Ovens, refrigerators, incubators
Validation
Module 1,, Part 4: QC-related validation Slide 19 of 28
© WHO – EDM – 1/2002
Chemical laboratory validation requirements (2) pH meter
Polarimeter - optical rotation
Refractometer
Spectrophotometer UV/Vis, IR, FTIR, Raman, AA
Timers
Viscometer
Volumetric equipment
Validation
Module 1,, Part 4: QC-related validation Slide 20 of 28
© WHO – EDM – 1/2002
Validation
Typical validation of HPCL assay (1)
System suitability (performance check) system precision column efficiency symmetry factor capacity factor
Module 1,, Part 4: QC-related validation Slide 21 of 28
© WHO – EDM – 1/2002
Validation
Typical validation of HPLC assay (2) Method validation
specificity accuracy precision linearity robustness
Module 1,, Part 4: QC-related validation Slide 22 of 28
© WHO – EDM – 1/2002
Biological assays Can be difficult to "validate"
"Validity" on a case by case basis
Strictly adhere to the Biological Testing monographs in pharmacopoeias
Validation
Module 1,, Part 4: QC-related validation Slide 23 of 28
© WHO – EDM – 1/2002
Microbiological testing requiring validation
Microbial limit testing
Microbial count
Sterility testing
Preservative effectiveness testing
Environmental monitoring program
Biological testing
Validation
Module 1,, Part 4: QC-related validation Slide 24 of 28
© WHO – EDM – 1/2002
Validation of microbial test procedures (1)
Virtually impossible to completely validate test procedures for every microorganism
Neutralize /inactivate inhibitory substances, or dilute
Periodic media challenge
Media QC
Reliable methods
Validation
Module 1,, Part 4: QC-related validation Slide 25 of 28
© WHO – EDM – 1/2002
Validation of microbial test procedures (2)
Incubation temperature and time
Media may not grow all microorganisms
Variations in media may affect recovery
Inhibitory disinfectants or preservatives
Sample
procedures
handling, storage, transport
Validation
Module 1,, Part 4: QC-related validation Slide 26 of 28
© WHO – EDM – 1/2002
Microbiological viable count method validation (1) Methods
pour plate / spread plate membrane filtration Most Probable Number
Sample size
Test dilution
Inoculation size
Validation
Module 1,, Part 4: QC-related validation Slide 27 of 28
© WHO – EDM – 1/2002
Microbiological viable count method validation (2)
Membrane filtration conditions
Incubation conditions
Acceptance criteria
Validation
Module 1,, Part 4: QC-related validation Slide 28 of 28
© WHO – EDM – 1/2002
Sterility testing validation
requirements
Media growth promotion, sterility, pH
Product validation
Stasis testing
Environmental monitoring
Negative controls
Challenge organisms
Validation