Model EQ-99X-FC LDLS™
Laser-Driven Light Source
Operation and Maintenance Manual
Revision 1 January 2014
Part Number DOC-6522
Copyright © 2014 Energetiq Technology Inc. All rights reserved.
Energetiq products are covered by US and foreign patents. All technical information, including
drawings, schematics and specifications contained in this manual are the property of Energetiq
and shall not be reproduced in whole or in part without the written consent of Energetiq. The
content of this manual is subject to change without notice.
Energetiq Technology Inc.
7 Constitution Way, Woburn, MA 01801 USA
Tel. +1 (781) 939-0763
Fax +1 (781) 939-0769
E-mail: [email protected]
http://www.energetiq.com
Declaration of Conformity
We, the manufacturers
Energetiq Technology Inc. 7 Constitution Way Woburn, Massachusetts 01801 USA
hereby declare that the product family LDLS™ Laser-Driven Light Source Model EQ-9X Series High Brightness Broadband Light Source is in conformity with the requirements of the following standards
EN 61010-1:2010 Safety Requirements for Electrical Equipment for Measurement, Control and (3rd Edition) Laboratory Use: Part 1 – General Requirements CISPR 11:2009 Industrial, scientific and medical equipment - Radio-frequency disturbance +A1:2010 characteristics - Limits and methods of measurement IEC61000-4-2:2001 Electrostatic Discharge Immunity IEC61000-4-3:2010 Radiated Electromagnetic Field Immunity IEC61000-4-4:2012 Electrical Fast Transient Burst Immunity IEC61000-4-6:2008 Radio Frequency Common Mode Immunity IEC61000-4-8:2009 Power Frequency Magnetic Field Immunity EN61326-1:2006 Electrical equipment for measurement, control and laboratory use.
EMC requirements. General requirements EN60825-1 (2007) Safety of laser products - Part 1: Equipment classification and
requirements in accordance with the provisions of
2006/95/EC EU Low Voltage Directive 2004/108/EC EU Electromagnetic Compatability Directive.
Signed, 28 January 2014 Paul Blackborow, CEO Date at Woburn, Massachusetts USA
CE Authorized Person in the Community:
Paul Laidler, Business Director (Machinery) TÜV SÜD Product Service Belasis Business Centre Coxwold Way Billingham TS23 4EA, UK Tel. +44 (0) 1642 345637
Table of Contents
Chapter 1 .................................................................................................................................1 General Information....................................................................................................1
Safety............................................................................................................................................... 1 Chapter 2 .................................................................................................................................8
Description ....................................................................................................................8 General ........................................................................................................................................... 8 System Description .................................................................................................................... 11 Power Supply Controller........................................................................................................... 12 Lamp House................................................................................................................................ 14
Installation................................................................................................................... 16 Unpacking.................................................................................................................................... 16 Connections................................................................................................................................. 16 Installation Procedure................................................................................................................ 22
Chapter 4 .............................................................................................................................. 24 Operation.................................................................................................................... 24
Starting.......................................................................................................................................... 24 Stopping ....................................................................................................................................... 24
Chapter 5 .............................................................................................................................. 26 EQ-99 SMA Fiber Cleaning Process...................................................................................... 26 Troubleshooting ......................................................................................................................... 30 Lamp Replacement .................................................................................................................... 32
EQ-99X-FC Operation Manual Rev. 1 1
Chapter 1
GENERAL INFORMATION
Safety
WARNING
CAUTION
The EQ-99 emits dangerous levels of UV
radiation. Even short exposures to skin or eyes
may cause burns. Ensure that only authorized
personnel are in the vicinity of source during
operation. Personnel in vicinity of operating
source should wear protective eyewear, clothing,
and gloves. Lighted UV warning lights and signs
posted on doors to lab areas may help prevent
accidental exposure.
This unit emits ultraviolet (UV) radiation that is
harmful to humans. Avoid exposure to the
direct or reflected output beam. Make certain
that the appropriate output beam shields and
optics are in place prior to energizing the unit.
All interlocks must be satisfied prior to
operation; failure to do so may lead to
hazardous conditions.
2 EQ99X-FC Operation Manual Rev. 1
WARNING
General Precautions
The EQ-99X-FC should only be operated with a fiber connected to the broadband output and
connected using an FC connector. Operation without an FC terminated fiber in place will
result in hazardous broadband radiation. The far end of the output fiber should be terminated
into a closed beam tube. Due to the possibility of generating ozone when ambient oxygen is
exposed to short wavelength light, the beam should always be enclosed in an appropriate beam
pipe, tube, or enclosed space. We suggest purging any beam transport space with dry nitrogen
gas.
The EQ-99X-FC source must also be cabled correctly and connected to a socket with a
protective earth ground prior to operation.
Refer to the Installation section of this manual (Chapter 3) for details of the facilities
connections.
Other than a bulb replacement, there are no user-serviceable parts inside the EQ-99X-FC.
For any problems encountered during operation, please contact Energetiq Technology for
assistance. If there is a component failure, do not attempt to open the Power Supply
Controller or Lamp House enclosure of the EQ-99X-FC.
The EQ-99 controller utilizes an internal Class 4
IR laser capable of causing severe injury to eyes
or skin. Do not open or attempt to service this
unit. Contact Energetiq regarding any problems
with the unit.
EQ-99X-FC Operation Manual Rev. 1 3
The EQ-99X-FC utilizes a quartz lamp containing a high-pressure gas fill. Explosion
of the lamp and possible injury from flying fragments can occur if the lamp is
mishandled.
Do not open the enclosure of either the Lamp House enclosure or the Power Supply
Controller. Dangerous invisible infrared laser beams and hazardous voltages exist
inside the units. Opening the chassis both voids the warranty and exposes the user to
dangerous radiation and hazardous voltages.
CAUTION
Laser Information
The EQ-99X-FC uses a patented (U.S. Patent #7435982, 7786455, 7989786, others pending)
laser drive system to excite a plasma that radiates in the UV as well as the visible bands. A
class 4 laser is located in the Power Supply Controller enclosure. Laser energy is delivered via
an armored fiber to the Lamp House enclosure and connected with an SMA-type connector.
Safety interlocks shut down the laser power if the SMA connector is removed from the Lamp
House enclosure. The optical configuration of the Lamp House ensures that the direct laser
beam can not exit the unit. The EQ-99X-FC laser product is designated as Class 1 during all
normal operation.
Use of controls or adjustments or performance of procedures
other than those specified herein may result in hazardous
radiation exposure.
4 EQ99X-FC Operation Manual Rev. 1
The parameters of the non-accessible internal laser are given below in Table 1.
Wavelength 974 nm
Emission Type CW
Laser Power for classification <1.38mW through a 7mm aperture
at 70mm
Beam Diameter ~3 mm at aperture
Divergence >100 mRad
Transverse Beam Mode Diffuse
Table 1: Embedded Laser Parameters
No regular service is required for the EQ-99X-FC. Any service to the system must be
performed only by factory authorized and trained technicians. To avoid injury, under no
circumstances should the user open or modify the Lamp House or Power Supply Controller
enclosure.
The unit must not be operated if the covers are removed or it is defective in any way. Contact
Energetiq if any problems with the equipment are suspected.
EQ-99X-FC Operation Manual Rev. 1 5
Labels and Safety Notification
The following safety labels appear on the product. Figure 1 shows the location of each label
on the EQ-99X-FC system.
Output Connector Warning Label – indicates hazardous levels
of broadband light are present. Unit should only be operated
with an output fiber installed to the EQ-99X-FC and the far
end of the fiber terminated so that the output beam remains
enclosed.
UV Hazard warning label – indicates hazardous levels of UV
light are present.
Manufacturer’s identification label – gives the manufacturer’s
name and address, and the model, serial number, and date of
manufacture of the equipment.
Explanatory label – states the classification of the laser product.
Class 1 is the lowest hazard level classification.
Certification label – states that the equipment has been tested
and verified to meet the standards indicated.
Non-interlocked housing label – notifies of a potential hazard
when covers are removed.
6 EQ99X-FC Operation Manual Rev. 1
Figure 1: Safety Label Locations
Safety Interlocks
The EQ-99X-FC is equipped with interlocks to prevent operation of the device when any of
the following conditions are present:
1. Bulb is not properly installed into the Lamp House enclosure
2. The laser fiber is not properly connected to the Lamp House enclosure
3. An external interlock is open
Certification Label
Explanatory Label
Non-interlocked Housing Label
Manufacturer’s Identification label
UV Hazard Warning Label
Output Connector Warning Label
Manufacturer’s ID Label
Non-Interlocked Housing Label
EQ-99X-FC Operation Manual Rev. 1 7
External Interlock
External interlock pins are provided for the customer’s use (see Chapter 3 for connection
details). Any suitable normally-open contact or solid-state switch can operate the interlock
circuit.
The interlock circuit must be connected to enable the operation of the unit. Should the
interlock connection open during operation or standby, the source is immediately disabled, and
all light output from the aperture ceases.
8 EQ99X-FC Operation Manual Rev. 1
Chapter 2
DESCRIPTION
General
The EQ-99X-FC is a broad-band lamp system for use in a wide variety of applications. The
lamp produces high brightness, broad-band light from DUV wavelengths through visible and
beyond. The output is very stable, and has a long lifetime before any service is required. A
simple control interface ensures ease of use.
Some of the advantages of the EQ-99X-FC include:
• Very high brightness across complete spectrum – 170nm through visible and beyond
• Eliminates need for multiple lamps (replaces D2/Tungsten/Xenon Arc) – Simplified optical system
• Excellent spatial stability – Repeatable measurements
• Superior short and long term power stability – Repeatable measurements
• Electrodeless operation for long life – Reduced consumable costs – Minimal recalibration of instrument
The EQ-99X-FC system consists of a Power Supply Controller unit, Lamp House unit, and
interconnecting cables. Connection to DC power is required for operation. Connection to
nitrogen purge gas is optional, but recommended for best performance. See Chapter 3 for
connection details.
EQ-99X-FC Operation Manual Rev. 1 9
Specifications
Optical Performance
• Typical output spectrum: see Figure 2.
Figure 2: Typical Output Spectrum
EQ-99X-FC Spectral Power Fiber Output230um Energetiq Fiber, 1m long
1
10
100
1000
180 280 380 480 580 680 780 880 980
Wavelength (nm)
uW/n
m
MeasuredEstimated
Physical Specifications
Dimensions (H x W x D) • Lamp House: 82 x 86 x 76 mm (3.2 x 3.4 x 3.0 in) • Power Supply Controller: 113 x 111 x 299 mm (4.4 x 4.4 x 11.8 in)
Weight
• Lamp House: 0.7 kg (1.5 lbs) • Power Supply Controller: 1.4 kg (3.0 lbs)
10 EQ99X-FC Operation Manual Rev. 1
Utility Requirements
• Electrical: 12VDC, 140W • Cooling: natural convection and internal fan, no auxiliary cooling necessary • Purge gas (optional): clean dry nitrogen, filtered to 5um 20 psig (0.14 MPa) supply
pressure
Remote Interface
Digital Inputs • Type: Optocoupler LED • Logic: Active High • Input voltage: 5VDC • Input current: 8mA
Digital Outputs
• Type: Open collector to ground (digital common) • Logic: Active Low • Voltage: 30VDC max. • Sink current: 30mA max.
User Power
• Voltage: 5VDC, referenced to digital common • Current: 50mA maximum
Environmental Requirements
Operating • Ambient temperature: 15–35°C • Relative Humidity: non-condensing, 80% max. for temperatures up to 31°C,
decreasing linearly to 50% max. at 40°C. • Pollution Degree 2 (normally only non-conductive pollution; occasional,
temporary condensation possible) • Installation Category II • Indoor use only
Transport • Temperature: -5–95°C • Relative Humidity: non-condensing, 95% max.
EQ-99X-FC Operation Manual Rev. 1 11
System Description
As shown in Figure 3 the EQ-99X-FC system consists of a Power Supply Controller unit, Lamp House, laser fiber optic cable, and Lamp House signal cable (not shown). Power and I/O interface connections (also not shown) are provided by the user. The following sections provide descriptions of the system components and controls, and gives
an overview of their functions. Refer to the “Installation” section of this manual (Chapter 3)
for more detailed information.
Figure 3: EQ-99X-FC Lamp System
Status Indicators (on rear panel)
Lamp House
Fault Status Indicators
Optical Output Fiber
I/O Interface Connector (rear panel) Power Supply
Controller
12 VDC Input Connector (rear panel)
Status Indicators
Armored Laser Fiber
12 EQ99X-FC Operation Manual Rev. 1
Power Supply Controller
The Power Supply Controller contains: • IR Diode Laser • Laser power supply • Thermo-electric cooler for laser • Permanently attached, armored laser fiber optic cable • Control electronics • Status indicator LEDs • Interface connectors
External features (refer to Figure 3):
Status Indicator LEDs
These five LEDs indicate the system status. The function of these indicators is shown below
in Table 2.
LED Label Meaning (when lit)
POWER ON DC power is connected to the EQ-99X-FC Power Supply Controller
LAMP ON UV Light is on
LASER ON Laser power is ON and laser light is being delivered to the Lamp House
CONTROLLER FAULT One of the following has occurred in the Power Supply Controller: 1. External interlock open 2. Controller internal temperature too high 3. Laser power not reaching setpoint 4. Laser Temperature Fault
LAMP MODULE FAULT One of the following has occurred in the Lamp House module: 1. Control Cable not connected properly 2. Lamphouse internal termperature too high 3. Laser fiber not correctly connected 4. Bulb not correctly installed 5. Ignition Failure
Table 2: Status Indicator LED Functions
EQ-99X-FC Operation Manual Rev. 1 13
Input/Output (I/O) Connector
Provides access to control and status signals. See Chapter 3 for pin assignments and functions.
This is the only operator interface to the EQ-99X-FC – there are no local controls. Energetiq
offers the EQ-99-RC Remote Control Module which connects to the Input/Output
connector and provides a means of local control. Contact Energetiq for additional
information.
Power Input Connector
This is a latching input connector for 12 VDC power. Power can be provided using the
optional PS00018 12VDC power supply. Alternatively, the EQ-99X-FC can be powered
directly from a customer provided 12VDC power supply. See Chapter 3 for detailed
information.
Lamp House Signal Connector (mini D sub – 9-pin)
Provides various power and control signals to/from the Lamp House module. No other
connector or cable may be used with the EQ-99X-FC other than the one supplied.
Armored Laser Fiber
The laser light is delivered from the Power Supply Controller to the Lamp House via a fiber
optic cable with armored protection. The fiber is permanently attached to the Power Supply
Controller, and connected to the Lamp House by an SMA-type connector.
It is critical that this armored fiber be treated with care and inspected for any abnormalities
prior to operation. Avoid sharp bends, which will permanently damage the fiber. Minimum
bend radius is 30mm (1.18 in). Avoid crimping or compressing the fiber. See Chapter 3 for
more information on installation and end face cleaning.
14 EQ99X-FC Operation Manual Rev. 1
Lamp House
The Lamp House assembly contains: • Lamp • Igniter • IR pumping optics • Output window • Laser ON indicator • Interface connectors
Figure 4: Lamp House Assembly
FC Output Connector
Laser Input (SMA-type)
Laser ON Indicator
Purge Gas Port
Signal Connector (to Controller)
Lamp mounting flange
Optical fiber output cable
EQ-99X-FC Operation Manual Rev. 1 15
External features (refer to Figure 4):
FC Output Connector
Provides broadband optical output using a standard FC connector. Energetiq recommends the
use of an Energetiq-supplied fiber for the broadband output. These fibers are able to
withstand the very high power density of deep UV radiation that is generated by the EQ-99X-
FC, thus giving best longevity and performance.
Nitrogen Purge Inlet
This is the inlet fitting for nitrogen purge gas. Purge gas is optional but is recommended for
best performance. With no purge, ozone will form from atmospheric oxygen and attenuate
the light output in the 220 – 280nm band. In addition, atmospheric oxygen and water vapor
will attenuate the output below 200nm.
There is no return fitting for the purge nitrogen. The purge flow normally escapes within the
Lamp House enclosure, and then to atmosphere.
Laser Input
This is the inlet connector for the armored fiber from the Power Supply Controller. It
contains a set of interlock pins that disable the laser when the SMA connector is not properly
in place.
Laser On Indicator
This LED is illuminated when the laser is ON.
J4 Signal Connector (mini D sub – 9-pin)
Provides various power and control signals to/from the Power Supply Controller. No other
connector or cable may be used with the EQ-99X-FC other than the one supplied.
16 EQ99X-FC Operation Manual Rev. 1
Chapter 3
INSTALLATION
Unpacking
Upon arrival, start by inspecting all parts of the system for completeness and any damage
incurred in shipping. The EQ-99X-FC shipping box should contain:
1) EQ-99X-FC Power Supply Controller unit
1) EQ-99X-FC Lamp House unit
1) Black interconnecting cable from Lamp House to Power Supply Controller (9
pin mini D-sub).
OPTIONAL
• EQ-99-RC Remote Control Module with interlock connector and I/O cable
• PS00018 12 VDC Power Supply (universal input voltage)
• UVFIBER-115-1M-FC 115um Output Fiber, 1m long
• UVFIBER-115-2M-FC 115um Output Fiber, 2m long
• UVFIBER-230-1M-FC 230um Output Fiber, 1m long
• UVFIBER-230-2M-FC 230um Output Fiber, 2m long
• UVFIBER-450-1M-FC 450um Output Fiber, 1m long
• UVFIBER-450-2M-FC 450um Output Fiber, 2m long
Use care when unpacking to avoid damaging the armored fiber optic cable.
If any part is missing or appears damaged, contact Energetiq immediately. Do not attempt to
substitute any parts. There are no user-serviceable parts inside the EQ-99X-FC Lamp House
or Power Supply Controller unit.
Connections
Installation of the EQ-99X-FC consists of connecting electrical and gas supplies, and
connecting the Lamp House module to the user’s equipment.
EQ-99X-FC Operation Manual Rev. 1 17
Electrical Power
The EQ-99X-FC requires 12VDC at 11.7A minimum (140W rating). Power consumption is
approximately 100W during normal operation.
Power can be provided using the optional PS00018 12VDC power supply. Alternatively,
power can be provided directly via a customer provided 12VDC power supply. The connector
on the EQ-99 is a latching connector. Once fully inserted, the connector will not release unless
the body of the connector is pulled first. This protects from accidental removal of power if the
power cable is pulled. Connect to a 12VDC source as follows:
Connector Kycon KPPX-4P
Pins 3 & 4 +12VDC
Pins 1 & 2 12VDC return
Purge Gas
Purge gas is optional but is recommended for best performance. With no purge, ozone will
form from atmospheric oxygen and attenuate the light output in the 220 – 280nm band. In
addition, atmospheric oxygen and water vapor will attenuate the output below 200nm.
If required, connect a source of nitrogen purge gas to the port on the Lamp House. The
fitting is a push-to-connect type, sized for 4 mm tubing.
Clean and dry nitrogen from either a dewar or research-grade N2 bottle is recommended. Do
not use any other purge gas. Grade 6 or better gas purity is recommended to maintain
cleanliness of the optics, and the gas should be filtered to < 5um. Supply pressure should be
20 psig (0.14 MPa). With a 20 psig inlet pressure, the EQ-99X-FC will consume approximately
1 slm of flow.
There is no return fitting for the purge nitrogen. The purge flow normally escapes within the
Lamp House enclosure, and then to atmosphere.
18 EQ99X-FC Operation Manual Rev. 1
Optical Interface
The lamp window accommodates a light output of 0.47 NA. A pair of pins and a 6-32
threaded hole are available for mounting. An internally-threaded SM1 adapter is provided for
connection of optical hardware. See Figure 5 below for mechanical layout of the Lamp
House.
Figure 5: Lamp House mechanical layout
EQ-99X-FC Operation Manual Rev. 1 19
Fig
ure
6:
Po
wer
Su
pply
Cont
rolle
r m
echa
nica
l la y
out
20 EQ99X-FC Operation Manual Rev. 1
Signal Connections
The EQ-99X-FC is controlled through the remote I/O connector.
Table 3 gives the pin assignments and functions for this interface. Connect to the user’s
control system using a suitable cable. Mating connector is a standard high-density 15-pin d-
sub male (for example, Amp part no. 748364-1 with contacts 1658670-2).
Optionally, connect a model EQ-99-RC remote control module to the I/O connector using
the supplied cable
User I/O can be powered either by the EQ-99 internal isolated power supply, or an external
supply.
Figure 7 shows connection schematics for both configurations.
Description Pin # Details
Commands (Inputs)
LAMP OPERATE 12 OPERATE REQUEST, apply +5V (referenced to digital common) to initiate ignition
EXTERNAL INTERLOCK 13 EXTERNAL INTERLOCK, apply +5V (referenced to digital common) to close interlock and allow operation
Status Indicators (Outputs) LAMP ON 1 Pulled to digital common when ON
LASER ON 2 Pulled to digital common when ON
LAMP MODULE FAULT 3 Pulled to digital common when OK, float on FAULT
CONTROLLER FAULT 4 Pulled to digital common when OK, float on FAULT
ISOLATED +5V SUPPLY 5 50mA maximum, referenced to digital common
DIGITAL COMMON 6,7,8,9 Galvanically isolated from system
RESERVED 10, 11 Do not connect
RESERVED 14,15 Do not connect
Table 3: I/O Connector Pin Assignments
EQ-99X-FC Operation Manual Rev. 1 21
0.1uF
100
Pins 1, 2, 3, 4
Pins 6, 7, 8, 9
Pin 5
499 Pins 11, 12, 13
Pins 6, 7, 8, 9
499
+-
0.1uF
0.1uF
Pins 11, 12, 13
Pin 5
+-
100
0.1uF
Pins 1, 2, 3, 4
+-
Inputs
Outputs
Isolated5VDC supply
5V Power
5V Return
Contact or
Load
solid-state switch
Outputs
Inputs
Isolated5VDC supply
5V Power
5V Return
Contact or
Load
solid-state switch
5VDC supplyUser
N/C
N/C
Figure 7: Remote Interface Schematic
22 EQ99X-FC Operation Manual Rev. 1
Installation Procedure
1. Mount the Power Supply Controller unit rigidly to either an optical breadboard plate or another suitable mounting structure using the supplied tabs on the bottom of the chassis. The holes are sized to accept standard ¼-20 optical bench hardware, and spaced to be compatible with a standard 1” grid mounting hole pattern. See Figure 6 for dimensional and mounting details.
Do not block the air vents of the Power Supply Controller.
2. Connect the Lamp House unit output to the user equipment. The output should always be connected to a fiber optic cable. The output end of the fiber optic cable should always be either directly coupled to another fiber optic cable, or enclosed in an appropriate beam pipe, tube, or enclosed space and purged with nitrogen. Operating the source without any output target or beam transport is not recommended, and may lead to unsafe operating conditions. Consult Energetiq for applications information and suggested configurations.
The Lamp House should be mounted in the orientation shown in Figure 3, with the output connector on the top and the flat surface of the housing on the bottom. The lamp has been factory aligned in this position. Mounting the lamp in a different orientation will cause the plasma position inside the bulb to shift slightly. This will result in slight misalignment and a resulting drop in output power.
3. Setup the Lamp House unit with appropriate ultraviolet safety measures and laser light safety measures in place. It is recommended that any enclosure or aperture-blocking hardware utilize switches wired to the EQ-99X-FC external interlock circuit.
4. Connect the black 9 pin mini D-sub interconnect cable from the Power Supply Controller (labeled TO LAMP) to the Lamp House (labeled J4).
5. The EQ-99X-FC is delivered with the armored laser fiber connected to the lamp house. This is done to minimize the possibility of debris or particles contaminating the end of the laser fiber. Energetiq strongly recommends leaving the laser fiber connected, unless it is necessary to disconnect it for installation or routing of the laser fiber. If the laser fiber must be disconnected, apply SMA caps immediately to both ends, and follow directions on inspection and cleaning of the laser fiber described in Chapter 5.
Use care when handling the fiber optic cable to avoid sharp bends, which will permanently damage the cable. Minimum bend radius for the cable is 30mm (1.18 in).
6. If necessary, connect the SMA connector to the Lamp House laser input and tighten. The SMA connector should be oriented such that the indicator line on the laser fiber is facing upwards (towards the bulb).
7. If required, connect nitrogen purge gas to the Lamp House. Refer to “Facilities Requirements” above.
EQ-99X-FC Operation Manual Rev. 1 23
8. Connect user’s control system to the I/O interface connector per Table 3 and Figure 7.
9. Alternately, if using the EQ-99-RC Remote Control Module, place it on a clean rigid surface. Install the supplied 15-pin cable from the Power Supply Controller to the EQ-99-RC.
10. Connect 12VDC input power source to the Power Supply Controller.
The system is now ready to operate.
24 EQ99X-FC Operation Manual Rev. 1
Chapter 4
OPERATION
Starting
Once the lamp is set up properly, verify that all personnel that will be in contact with the lamp
system are aware of the potential hazards involved. It is the responsibility of the user to verify
that the lamp is being used safely.
This example assumes the use of the EQ-99-RC Remote Control Module to provide local
control. If using a custom control system, substitute the appropriate digital input and output
lines from Table 3 for the switches and LEDs described below.
1. With the EQ-99-RC Remote Control Module connected properly, review the status LEDs on the Power Supply Controller. The POWER ON LED should be lit, and neither the CONTROLLER FAULT nor LAMP MODULE FAULT LEDs should be lit.
2. Turn on the OPERATE Switch (place switch in UP position).
3. Within several seconds the LASER ON LED will light. Laser light is now present in the Lamp House.
4. In approximately 20-150 seconds the igniter will be turned on automatically and the plasma will ignite. The LAMP ON LED will be lit. The duration of the warm up time (20-150 seconds) will depend on the temperature and previous operating condition of the EQ-99X-FC. The EQ-99X-FC will automatically detect when the unit has reached the optimum conditions for ignition.
5. If a bulb fails to ignite, 150 seconds after the OPERATE switch was activated, the LASER ON LED will go out, the LAMP FAULT LED will be lit, and LAMP ON LED will remain off. This is very unusual. However, if this occurs, turn the OPERATE switch to the OFF position (down) and begin at Step 1 again. If this occurs multiple times, contact Energetiq service.
Stopping
To turn the LAMP off, simply turn the OPERATE Switch to the OFF position. If the lamp
will not be used for some time, the 12VDC supply can be turned off.
EQ-99X-FC Operation Manual Rev. 1 25
To minimize wear on the ignition components, it is best to avoid frequently starting and
stopping the lamp. It is recommended to run the lamp continuously if long off-periods are
not required.
26 EQ99X-FC Operation Manual Rev. 1
Chapter 5
EQ-99 SMA Fiber Cleaning Process
Fiber Cleaning Basics:
• The LDLS laser fiber connector is carefully inspected and is clean prior to shipment. Energetiq strongly recommends leaving the laser fiber connected, unless necessary for installation or routing of the laser fiber.
• Operating the LDLS with a contaminated fiber introduces the risk of decreased performance or damage to the unit.
• Leave Laser Fiber SMA connected whenever possible.
• A fiber is considered clean when there are no particles on the fiber core or cladding that are detectable with >200X magnification.
• Follow the fiber inspection and cleaning process (Figure 8) before making a laser fiber SMA connection. As shown in the figure, inspecting and cleaning is an iterative process. The final step of the cleaning should always be a successful inspection. Sometimes the cleaning action redistributes particles from an area that is not visible with the inspection scope (and not problematic) to area that is visible (and is problematic). Because of this, we recommend that the first step is to remove any loose particles or debris with pressurized N2.
Fiber Caps:
• If the SMA fiber must be disconnected, always use a metal cap on the laser fiber end, and on the lamp head.
• Store caps open side down when possible.
• The threads on the caps can generate and store metal particles, which are particularly problematic to the laser fiber. Blow out caps with CDA (Clean Dry Air) or N2 prior to use.
• Never leave the fiber disconnected without a cap.
EQ-99X-FC Operation Manual Rev. 1 27
Fiber Inspection and Cleaning Process
CAUTION
Figure 8: Fiber Cleaning Process Flowchart
The fiber endface may not have any particles on the core or cladding that are detectable with >200X magnification (see Figure 4).
Disconnect power from the EQ-99 Power Supply Controller unit before performing fiber inspection.
•Is fiber end
•Connector
•Clean Fiber•end with•Clean Dry Air
•Is fiber end
•Install Laser SMA•Connector
•Clean Fiber•end with•Cleto
•Is fiber end
•Connector
•Isopropyl•Alcoho
Yes
No
Yes
No
Inspect Fiber end
Is fiber end contaminated?
Connector
Clean Fiber end with N2 or CDA
Inspect Fiber end
Is fiber end
contaminated?
Install Laser SMA Connector
Clean Fiber end with Cletop
Inspect Fiber end
Is fiber end contaminated?
Install Laser SMA Connector
Clean Fiber Lint Free Wipe and Isopropyl Alcohol
Yes
No
Start
Install Laser SMA
Blow off Ferrule and Nut threads with N2 or CDA
Blow off Receptacle Threads with N2 or CDA
28 EQ99X-FC Operation Manual Rev. 1
Fiber Inspection Tools
Figure 9: CL-200 Handheld Fiber Inspection Scope Figure 10: JDS Uniphase / Westover FVD-2400 Benchtop USB/PC Operated Fiber Viewer
Note – A special SMA adapter must be used with either the CL-200 handheld inspection scope or the Westover Digital inspection scopes. This custom adapter allows the scope to be used with the EQ-99 laser fiber and integral interlock ring. The custom adapter may not be compatible with other fiber inspection equipment.
The following fiber inspection equipment can be purchased from Energetiq: Energetiq Part Number Description Use
EQ-FIBER-INSPECT CL-200 Fiber Inspection Scope with Custom SMA Adapter
Standalone Fiber Inspection
MPD-5265 Custom Adapter from CL-200 or Westover to EQ-99 Fiber
For use with an existing CL-200 or Westover Digital Microscope
EQ-99X-FC Operation Manual Rev. 1 29
Examples of Fiber Images
Figure 11: Fiber End Images
Fiber Cladding Fiber Core Area
Fiber Cleaning Tools
1. Clean Dry Air or Pressurized Nitrogen
2. Dust Off (or similar clean, compressed air)
3. Cletop (www.cletop.com)
4. Lint Free Wipe with Isopropyl Alcohol
Figure 11. Cletop Fiber Cleaning Tool To Clean with the Cletop or Wipe, gently press the fiber end face into the cleaning cloth or wipe, rotate in place several times, then reduce pressure and drag the fiber 1-2 cm across the cleaning cloth or wipe.
Cletop Type A SMA cleaner with Blue tape
14100700
14100500
Box of 6 Replacement Blue Tape Reels
30 EQ99X-FC Operation Manual Rev. 1
Troubleshooting
Fault Indicator Block Diagram:
Lamp Fault Asserted
(pin 3 on remote)
DiagnosticSafety
Laser SMA Not Connected
Bulb Not Installed Properly
Lamp Head Overtemp > 85C
Control Cable Not Connected
Plasma Extinguished (after ignition)
Controller Fault Asserted
(pin 4 on remote)
Laser Overtemp (>66C)
Controller PCB Overtemp (>82C)
Laser Current Regulation Error
External Interlock Open
Laser Temperature Regulation Error
Signal Available on User Interface Internal Logic That Generates Signal
Lamp On De-Asserted (pin 1 on remote)
Laser On De-Asserted 150 sec
after Operate Command
(pin 2 on remote)
Fail to Ignite
Failed Ignition
Laser Fiber Failure
EQ-99X-FC Operation Manual Rev. 1 31
See below if any problems are encountered in operating the EQ-99X-FC.
Condition:
Controller Fault and/or Lamp House Fault LED(s) are ON.
Action:
• Always begin operation of the source by verifying the interlocks.
• Confirm that the external interlock contact is closed (or that the black jumper plug
is fully inserted into the back of the EQ-99-RC Remote Control Box)
• Check that the bulb and laser fiber are properly connected at the Lamp House
unit.
• To reset or clear a Fault condition, the actuate the “Operate” switch from the ON
position to the OFF position. If a fault was generated while the “Operate” switch
was in the OFF position, first actuate the switch to the ON position, then to the
OFF position. The unit will not turn on if a fault condition exists.
• If all of the interlocks are OK and either the lamp or controller interlock faults
will not clear, please contact the factory.
Condition:
Lamp fails to ignite after several tries.
Action:
Contact Energetiq.
32 EQ99X-FC Operation Manual Rev. 1
Condition:
Optical output shows degradation in the deep UV (i.e. <300nm)
Action:
The optical output fiber may have degraded due to the high UV power density
generated by the EQ-99X-FC. Replace the output fiber with a new one to determine if the
original performance is restored. All fibers will degrade over time, although fibers supplied
by Energetiq have been designed for the EQ-99X-FC.
Lamp Replacement
Contact Energetiq if a bad lamp is suspected.
Copyright © 2014 Energetiq Technology Inc. All rights reserved.
Energetiq products are covered by US and foreign patents. All technical information, including
drawings, schematics and specifications contained in this manual are the property of Energetiq
and shall not be reproduced in whole or in part without the written consent of Energetiq. The
content of this manual is subject to change without notice.
Energetiq Technology Inc.
7 Constitution Way, Woburn, MA 01801 USA
Tel. +1 (781) 939-0763
Fax +1 (781) 939-0769
E-mail: [email protected]
http://www.energetiq.com