Bone Conserving Hip Replacement Surgical technique MiniHip™

Upload: others

Post on 14-Oct-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

Bone Conserving Hip ReplacementSurgical technique

MiniHip™

Page 2: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

a

MiniHip™

2 |

Contents

Operative summary 4

Overview 5

Pre-operative templating 6

Operative technique 7

Neck resection 7

Femoral canal preparation 7

Trial reduction 8

Implantation of the stem 8

Final reduction 9

Appendix: Stem extraction 9

Sizing guide 10

Description 11

Indications 11

Contraindications 11

Ordering information 11

Page 3: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

a

Bone conserving hip replacement

| 3

Page 4: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

a

MiniHip™

4 |

Operative summary

a. Curved awl b. Broaching c. Trial reduction d . Implantation

Page 5: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

a

| 5

Overview

The MiniHip bone preserving, physiological hip replacement has been specifically designed to address the clinical needs of active men and women of all ages who would usually be appropriate for a standard uncemented hip replacement.

Anatomic restoration can be achieved by utilizing an optimal neck resection (Figure 1). The dedicated short stem design of the MiniHip combines the bone preserving benefits of hip resurfacing with the stability of traditional total hip replacement, providing a solution for uncemented patients.

Designed to ‘fit and fill’ the proximal neck, the MiniHip may promote immediate and long-term proximal stability, which is further enhanced with a hydroxyapatite coating for stable cementless fixation (Bi-coat™). The initial press fit coating may facilitate primary stability to help promote bone integration1.

The specialized instrumentation designed for all approaches, including minimally invasive, combined with a practical operative technique, assists in facilitating:

■ Bone preservation

■ Natural physiological load transfer2

■ Proven Bi-coat fixation1

■ Reproducible anatomical restoration3,4

■ Efficient three stage procedure

Note: MiniHip is suitable for Dorr type A and B bone quality. MiniHip is not suitable for type C, osteoporotic stove femurs.

Figure 1. Reproducibility through anatomical restoration

138º ≤

127º

118º ≥

Page 6: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

a

MiniHip™

6 |

Pre-operative templating

When using the MiniHip system, pre-operative planning is critical to determine the optimal size, neck angle and offset of the implant. Templating will help determine the level of neck resection and optimal cup position. The MiniHip X-ray templates are available in four different magnifications (100%, 110%, 115% and 120%). The 115% magnification is provided as standard.

A/P and M/L X-rays should be used in combination with the templates to determine the correct size and positioning of the implant. The resection guide line or demarcation line is defined by using a medium length head (0mm) on the template overlaid onto the center of rotation. The template is then rotated to achieve fit and fill of the neck superiorly and laterally whilst also maximizing coverage of the medial calcar and having contact with the distal lateral cortices. When using the M/L X-ray, proximal contact should be achieved posteriorly and anteriorly, and distal contact should be achieved posteriorly.

The demarcation line is where the cementless coating ends proximally.

To achieve anatomical restoration the following four key stages need to be implemented:

■ Define center of rotation

■ Fit and fill in the neck

■ Maximal medial calcar contact

■ Distal lateral contact/fill (posterior, anterior, posterior- PAP)

The neck resection/demarcation line of the stem should always be parallel to the head-neck junction as shown by the red line in Figure 2.

Figure 2. Pre-operative templating

Page 7: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

a

| 7

Step 3. Femoral canal preparation

The stem is designed to be supported by compacted cancellous bone thereby gaining mechanical stability and fixation in the neck and metaphysis. Always start with the smallest broach, progressively using larger broaches until the pre-templated size is achieved.

Step 1. Neck resection

When defining the neck resection in situ, use the piriformis fossa as a constant landmark to determine the level of resection.

Operative technique

Step 2. Curved starter awl

The entry point of the starter awl should be 3-4mm superior to the center of the neck to create a pathway into the top of the diaphysis. This will help prevent both undersizing of the stem and misalignment of the broaches.

The curved starter awl is used for further femoral preparation accommodating the natural curvature of the proximal femur allowing the compaction broaches to follow a sequential pathway.

Page 8: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

a

MiniHip™

8 |

A stable fit is achieved when the broach fits and fills the proximal femur and the face of the final broach sits flush with the resection line, as detailed in the pre-operative templating on page 6.

In order to ensure that the femur will accommodate the planned size of implant, a 1mm cancellous ring should be maintained around the broach for sizes 3-4 and a 2mm ring for sizes 5-9. Usually a 2-4mm ring of cancellous bone around the medial calcar is also observed.

Step 4. Trial reduction

A trial reduction may then be performed using a trial head and the broach.

Intraoperative X-ray

To assess the trial orientation and sizing, it is recommended that an intraoperative X-ray using a C-arm be taken whilst the broach, trial neck and head are in still in situ prior to implantation of the definitive implants.

Step 5. Implantation of the stem

The final broach indicates the size of the definitive implant, which is inserted by hand.

If the definitive stem seats by more than 10mm proud above the resection line, remove the stem and re-broach with the finishing broach 2-4mm below the resection line.

Page 9: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

a

| 9

Once the stem is seated by hand, the stem impactor can be used to fully seat the implant until the edge of the Bi-coat (demarcation line) is flush with the resection line.

Step 6. Final reduction

Once the acetabular cup is implanted, a full trial reduction can be performed allowing for fine tuning of the head offset. The definitive head is impacted onto the stem ensuring the trunnion is free from debris. The hip can then be reduced and closure performed.

Stem extraction

In the unlikely event that the stem needs to be removed, the stem extractor set can be used. The extractor device is placed over the 12/14 trunnion and the slap hammer is rotated until the trunnion is securely clamped. Do not implant the stem with this instrument.

Note: Do not overtighten the thread as this will cause damage to the plastic collar on the extractor. It should be tight enough to form an interference fit around the base of the neck to aid removal. The stem can then be extracted in conjunction with revision ancillary instruments.

If the stem extractor is required, it should be ordered separately prior to the operation.

Appendix

Page 10: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

a

MiniHip™

10 |

stem length

offset

neck length

stem length (mm)

neck length (mm)

offset (mm)

Size

89.0

31.5

34.5

3

93.8

32.5

35.6

4

98.5

33.5

36.8

5

103.3

34.5

37.9

6

108.0

35.5

39.1

7

112.8

36.5

40.2

8

117.5

37.5

41.4

9

Sizing guide

Page 11: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

a

| 11

stem length (mm)

neck length (mm)

offset (mm)

Size

89.0

31.5

34.5

3

93.8

32.5

35.6

4

98.5

33.5

36.8

5

103.3

34.5

37.9

6

108.0

35.5

39.1

7

112.8

36.5

40.2

8

117.5

37.5

41.4

9

Description

The MiniHip is a titanium femoral stem (Ti-6Al-4V) coated with a layer of hydroxyapatite applied over a layer of pure titanium. The distal section of the stem is polished. The device is available in a range of seven sizes each providing a 130o CCD neck angle. The device is intended to be used with 12/14 modular taper heads.

The MiniHip is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Indications

The indications for the MiniHip stem as a total hip arthroplasty include:

■ Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis

■ Rheumatoid arthritis

■ Correction of functional deformity

■ Developmental dysplasia/congenital dislocation of the hip

The MiniHip stem is indicated for cementless use only.

Contraindications

■ Infection

■ Osteomyelitis

■ Sepsis

■ Osteomalacia

■ Distant foci of infections

■ Osteoporosis

■ Marked bone loss or bone resorption

■ Metabolic disorders which may impair bone formation

■ Vascular insufficiency

■ Muscular atrophy or neuromuscular disease

■ Allergy to implant material

■ Severe deformity

Due to components not yet available in the USA, the MiniHip stem should not be used with modular head components which result in an effective offset of greater than +4mm.

Stem sizes 1 and 2 are not currently available for use in the United States.

Ordering information

Implants

580.0003 Size 3 Standard stem

580.0004 Size 4 Standard stem

580.0005 Size 5 Standard stem

580.0006 Size 6 Standard stem

580.0007 Size 7 Standard stem

580.0008 Size 8 Standard stem

580.0009 Size 9 Standard stem

Instruments

580.9991 MiniHip cementless stem system

X-Ray templates

580.9190 MiniHip X-ray template 115%

580.9090 MiniHip X-ray template 100%

Stem extractor set

580.9410 Stem extractor set

Page 12: MiniHip™€¦ · damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Indications The indications for

©2012 Corin P No I887 Rev3 02/2012 ECR 11672

Corin USA10500 University Center Drive Suite 190, Tampa, Florida 33612 USAt: 813-977-4469f: 813-979-0042e: [email protected]

www.coringroup.com

References:

1. Collier CG. The assessment of early osteointegration, as a function of coating. August 2002. Report held on file.

2. Short stems are less likely to lead to bone resorption; bone remodeling following THR. Yeoman M, Cizinauskas A, Lowry C, Vincent G, Collins SN, Simpson DJ, Continuum Blue, UK; 2 – Corin Ltd, UK; 3 – Imorphics, UK. Data held on file at Corin

3. Short stem total hip replacement: are you being conservative enough? Simpson DJ, Lowry C, Yeoman M, Cizinauskas A, Vincent G, Collins SN, Corin, UK; 2 - Continuum Blue, UK; 3 - Imorphics, UK. Data held on file at Corin

4. Presented data at the Baden Baden 58th Annual VSO meeting, (Tetnang, Germany) 2010. Dr Christian Grasselli discussed the multicentre clinical data for 250 bone conserving MiniHips.