Miniaturized ultra high-mass resolution time-of-flight mass spectrometer “infiTOF”

and applications for environmental analysis

Graduate school of engineering, Osaka University

Shuichi Shimma, PhD

Page 2: Miniaturized ultra high-mass resolution time-of-flight

(Gao L et al., Anal. Chem., 2008, 80, 7198-7205)

(HAPSITE system brochure)

(TORION Technologies HP)

Miniaturized mass spectrometersTRIDION9 GC-TMS

(TORION Technologies Inc)Mini 11

(Purdue Univ. )

HAPSITE system (Inficon)

38.1 cm x 39.4 cm x 22.9 cm (14.5 kg)Purdue Mini-12 (QuantIon source + Mini 11)

Page 3: Miniaturized ultra high-mass resolution time-of-flight

(Ouyang Z et al., Anal. Chem., 2009, 81, 2421-2425)

Summary of reported miniaturized instruments.

Q1. High mass resolution is unnecessary for on-site measurement ? Q2. High mass resolution cannot be available in miniMS ?

Page 4: Miniaturized ultra high-mass resolution time-of-flight

What kinds of instruments are suitable for high mass resolution ? (1) Magnetic sector, (2) FT instruments (FT-ICR and Orbitrap), (3) Time-of-Flight instruments.�

Page 5: Miniaturized ultra high-mass resolution time-of-flight

High mass resolution

Maximize flight time T

Maximize flight path length

Minimize peak width ΔT

Mass Resolution of TOF instruments

Mass Resolution: mΔ m = T

2 ΔT

Page 6: Miniaturized ultra high-mass resolution time-of-flight

Mass Resolution of TOF instruments

Mass Resolution: mΔ m = T

2 ΔT

❏ To extend the flight time T : Long flight path length

❏ To minimize the peak-width ΔT (FWHM)(1) Two-stage acceleration and time-lag focusing (Delayed Extraction) (2) Orthogonal acceleration (3) Ion mirrors (Reflectron) (4) Electric sectors

Straight Forward Way

(maXis http://www.bdal.com)

Dimensions: 800 x 1320 x 2850 mm Weight: 500 kg Mass resolution: 60,000 (these values from maXis brochure)

Page 7: Miniaturized ultra high-mass resolution time-of-flight

Photographs of “MULTUM-S II” system

Ion Source

MULTUM

Detector

45 cm 23 cm 64 cm35kg

TMPTMP

DiaphragmPump

DiaphragmPump

20 cm

20 c

m

Orbiting Sector

Orbiting Sector

InjectionSector

EjectionSector

Ion Selector

(a) (b)

Ion source: electron ionization Mass range: m/z 1 - 1000 (depend on ionization methods) Pressure in analyzer: typically 5x10-5 Pa (with diaphragm pump and TMP 70 Ls-1 (N2))

Shimma S et al. Anal. Chem., 82, 8456 (2010)

Commercialized as “infiTOF” (KANOMAX)

Page 8: Miniaturized ultra high-mass resolution time-of-flight

Operation of “MULTUM-S II” system (Zooming mode)

• Injection and ejection electrodes are synchronized with the ion source pulsing triggering.

• Injection electrode has to be in the on state while ions enter the analyzer, then has to be turned off before first ion (lightest ions) returns to it.

• Orbiting electrodes are constant, so orbiting ion can be held until ejection electrode is ON.

injectionejection

detector

Page 9: Miniaturized ultra high-mass resolution time-of-flight

(C)

Performance evaluations

(A) Mass resolution and peak width (N2), Mass resolution is linearly increased by changing number of cycles.

(B) Ion transmittance (N2), 98% ions survived in each cycle.

(C) Linear signal response (PFTBA) The dynamic range was 3 orders of magnitude.

Shimma S et al. Anal. Chem., 82, 8456 (2010)

Mass resolution

Page 10: Miniaturized ultra high-mass resolution time-of-flight

4He+. D2+. 4He+.

D2+.

Separation of doublets (direct injection of standard gas)

O+. or (O22+.)

CH4+.

N2+.

O2+.

Ar+.

O+. or (O22++..)

CH4+.

He and D2

O and CH4

Resolution 1000 at 10 cycle Low resolution

Low resolutionResolution 3200

at 6 cycle

Page 11: Miniaturized ultra high-mass resolution time-of-flight

43.5 43.6 43.7 43.8 43.9 44 44.1 44.2 44.3 44.4m/z

%

100

80

60

40

20

0

10 cycles

43.5 43.6 43.7 43.8 43.9 44 44.1 44.2 44.3 44.4m/z

%

100

80

60

40

20

0

43.99220 cycles

43.5 43.6 43.7 43.8 43.9 44 44.1 44.2 44.3 44.4m/z

%

100

80

60

40

20

0

43.990100 cycles

Mass resolution> 30,000

Separation of CO2&N2O doublet

CO2&N2O

N2OCO2

N2O

CO2

MULTUM-S II has a tunable mass resolution, and can work as a high performance gas monitor system.

Shimma S et al. Anal. Chem., 82, 8456 (2010)

CO2 and N2O are known as greenhouse gases.

Real-time monitoring of CO2 and N2O is required to elucidate the genera t ing mechan ism and investigate its trend of spread.

m/z of CO2 43.9898 N2O 44.0011 Δm: 0.0113

High mass resolution is required.

Page 12: Miniaturized ultra high-mass resolution time-of-flight

PCBs analysis using “GC-infiTOF”

Polychlorinated biphenyls: a class of organic compounds with 2 to 10 chlorine atoms attached to biphenyl. 209 congeners. Toxic compounds.

Conventional analysis methods for PCB congeners: 1. Gas chromatograph (GC)/ECD 2. GC/MS/MS, GC/QMS, GC/negative ion chemical ionization (NICI) 3. GC/high-resolution MS (resolution > 10, 000. Magnetic sector instruments)

(Cl)n(Cl)n

Structure of PCBs

GC-infitOF system (GC: Agilent 6890N)

Point of this application: GC-MULTUM-S II can be alternative to GC-magnetic sector system??

For GC/HRMS and GC/NICI, sample preparation is simple. ONLY dilution with n-hexane.

Page 13: Miniaturized ultra high-mass resolution time-of-flight

3,200

2,400

1,600

800

1000 1500 2000 2500 3000Scan1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0Min.

Total ion chromatogram of 66 PCBs mixture.

By fast-GC technique, PCBs mixture can be analyzed within 5.5 min.

(Shimma S et al. J Env Moni 2012)

ConditionsColumn: Forte HT8 (12 m x 0.22 id, film 0.25 µm) Inlet temp.: 320°C Temp. program: 100°C (0.5min)->40°C min-1

320°C

Carrier gas: helium (70 cm s-1) Sample: 66 PCB mixture (mono, di-PCBs: 2 ppm others: 1 ppm)

High mass resolution mode (R > 10,000)

Page 14: Miniaturized ultra high-mass resolution time-of-flight

Hepta-CBTIC

Mass Chromatogramm/z 393.80

2050 2100 2150 2200 2250 2300 2350 2400

1 ppb was

detectable.

Limit of detection at 20 cycles

1 ppb of hepta-CB in the high mass resolution mode (R > 10,000)

Page 15: Miniaturized ultra high-mass resolution time-of-flight

Dielectric coolant analysis (hexane dilution)

In the low mass resolution mass spectra, signal peak and matrix peak were overlapped. Therefore, matrix interference were observed as a background noise in each mass chromatogram.

Mass resolution 300- Comparison of low- and high-resolution mode-

Page 16: Miniaturized ultra high-mass resolution time-of-flight

In the high mass resolution mass spectra, signal peak and matrix peak were clearly separated. High mass selectivity provided mass chromatogram without matrix interference even in diluted dielectric oil sample.

Dielectric coolant analysis (hexane dilution)

Mass resolution 10 000- Comparison of low- and high-resolution mode-

Page 17: Miniaturized ultra high-mass resolution time-of-flight

Summary

1. On-site mass spectrometry is the future application in MS. 2. To develop miniaturized MS instruments are essential. 3. High performance miniaturized TOF instrument

MULTUM-S II was developed. 4. Obtained mass resolution was > 30,000. 5. MULTUM-S II can work as the high performance

simultaneous gas monitoring system. 6. In the PCB analysis, MULTUM-S II is the alternative to

conventional HRMS instruments.