www.iap.uni-jena.de Metrology and Sensing Lecture 11: Measurement of basic system properties 2018-01-11 Herbert Gross Winter term 2017

Upload: nguyentu

Post on 10-May-2018

216 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

www.iap.uni-jena.de

Metrology and Sensing

Lecture 11: Measurement of basic system properties

2018-01-11

Herbert Gross

Winter term 2017

Page 2: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

2

Preliminary Schedule

No Date Subject Detailed Content

1 19.10. Introduction Introduction, optical measurements, shape measurements, errors,

definition of the meter, sampling theorem

2 26.10. Wave optics Basics, polarization, wave aberrations, PSF, OTF

3 02.11. Sensors Introduction, basic properties, CCDs, filtering, noise

4 09.11. Fringe projection Moire principle, illumination coding, fringe projection, deflectometry

5 16.11. Interferometry I Introduction, interference, types of interferometers, miscellaneous

6 23.11. Interferometry II Examples, interferogram interpretation, fringe evaluation methods

7 30.11. Wavefront sensors Hartmann-Shack WFS, Hartmann method, miscellaneous methods

8 07.12. Geometrical methods Tactile measurement, photogrammetry, triangulation, time of flight,

Scheimpflug setup

9 14.12. Speckle methods Spatial and temporal coherence, speckle, properties, speckle metrology

10 21.12. Holography Introduction, holographic interferometry, applications, miscellaneous

11 11.01. Measurement of basic

system properties Bssic properties, knife edge, slit scan, MTF measurement

12 18.01. Phase retrieval Introduction, algorithms, practical aspects, accuracy

13 25.01. Metrology of aspheres

and freeforms Aspheres, null lens tests, CGH method, freeforms, metrology of freeforms

14 01.02. OCT Principle of OCT, tissue optics, Fourier domain OCT, miscellaneous

15 08.02. Confocal sensors Principle, resolution and PSF, microscopy, chromatical confocal method

Page 3: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

3

Content

Basic system properties

Knife edge method

Slit scan method

MTF measurement

Page 4: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Collimated incident light

Calibrated collimator with focal length fc and test chart with size y

Selection of sharp image plane

Analysis of image size

Measurement of Focal Length with Collimator

f'c f'

test chart image

y y'

collimator test optic

y

yff c

'''

4

Page 5: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Setup with distance object-image L > 4f

Known location of the principal plane P of the system

distance dP between principal planes

Selection of two system locations with sharp image

Relative axial shift D between the two setups

Measurement of Focal Length According to Gauss

H

H

dL

DdLf

44

2

F F'

L

F F'

dP

Ds

s'

P P'

P P'

position 1

position 2

f f'

5

Page 6: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Telecentric movable measurement microscope with offset y: Abbe focometer

Focusing of two different test charts with sizes y1 and y2

Determination of the focal length by

Measurement of Focal Length with Focometer

test lens

measuring

microscope

y1

Fy2

e y

f focusing

movement

u

test

scale 1

test

scale 2

e

yy

f

yu 12tan

6

Page 7: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Setup with fiber and plane mirror for autocollimation

Change of distance between test lens and fiber

Analysis of the recoupled power into the fiber (confocal) gives the focal point

Measurement of Focal Length by Confocal Setup

z

lens under

testplane mirror

autocollimation

case

recoupled

energy

lens position z

7

Page 8: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Afocal setup with sharp image plane

Measurement of long focal lengths

Insertion of test system in collimated light segment and refocussing

Applying the lens makers formula

Measurement of Focal Length with Focimeter

reference

image

plane

test

target

lens

under

test

f2f2

new image

plane

x

collimator

lens

focusing

lens f2

re-focusing

2

22

11

f

x

ff

Page 9: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Setup with collimator and two Ronchi rulings

System under test is inserted

Grating period d and azimuthal angle q between the gratings

Moire pattern is rotated by angle a, if test lens acts as focussing element

Radius of curvature R or focal length

Measurement of Focal Length by Moire Deflectometry

test

target

lens

under

test

observation

plane

collimator

lens

Ronchi

gratings

a

qMoire

pattern

d

aq tan

dR

Page 10: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Setup with Ronchi grating in collimated light gives a series of Talbot images

The Talbot planes are imaged by the system under test

Analysis of the image plane by lens formula gives the desired focal length

By use of several planes, the position of the principal plane can be eliminated

A second Ronchi grating can be used to find the accurate image planes

Measurement of Focal Length by Talbot Imaging

test

target

lens

under test

intermediate

Talbot images

collimator

lens

Ronchi

grating

-2 -1 0 +1 +2 +2 +1 0 -1 -2

Talbot

images

Page 11: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Criteria for best focus:

1. Paraxial centre of curvature for the paraxial spherical wave of an on axis object point.

2. Maximum of the Strehl ratio

3. Smallest rms-value of the wave aberration

4. Highest contrast of the modulation of an object feature of given spatial frequency

5. Highest value of the slope of an edge

6. Highest value of the entropy of the detected digital image

Requirements for focus detection procedure

1. Steep curve dependency to get high accuracy

2. Robust definition to deliver a large dynamic range

3. Suppression of side lobe effects to guarantee an unambiguous solution

4. High frequency pre-filtering to be noise insensitive

Determination of Best Focus

Page 12: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Blur of defocussed

plane

Minimum of image entropy (wj is intensity in pixel No j)

Maximum of image contrast

Determination of Best Focus

object

plane

optical

systembest

focus

plane

actual

plane

defocus

Dz

D blur

diameter

j

jj wwE 2log

Page 13: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Phase analysis by Zernike

coefficient c4

Measurement with two

Ronchi gratings

Determination of Best Focus

2

44

1NAz

nc D

object

plane

optical

systemmeasuring

planesensor

plane

defocus

Dz

Ronchi

gratings

cj

z grating

location

0

0

c4

c9

Page 14: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement by image analysis:

1. Maximum gradient of edges

2. Power of gradients

3. Laplacian

Determination of Best Focus

-4 -3 -2 -1 0 1 2 3 40

0.02

0.04

0.06

0.08

0.1

0.12

0.14

0.16

rms ( I )

Dz

[a.u.]

22

),(

y

I

x

IyxIg

dydxyxIG2

),(

dydxyxIL2

2 ),(

Page 15: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement for systems in air via the nodal planes

Imaging of a test pattern with a collimator onto a detector

Invariant lateral image location for rotated system around the nodal point

Critical: vignetting effects for large angles

Measurement of Principal Planes

test

patterncollimator

test system detector

P P'

tilt

angle

j

invariant

image

position

turning

point

movable along z

Page 16: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Setup of the test lens with different object locations: axial shift D

Analysis of the lens imaging formula

Minimizing the error of several measurements j

Measurement of Principal Planes

test

patterntest system detector

P P'

sj s'j

aj

a'jD

reference

plane

faa jj

1

'

11

D

D

2''' DD jjjjjj aaaaaaf

Page 17: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Setup with collimating auxiliary lens

Determination with measuring microscope (dynameter)

Measurement of Pupil Size

pinhole

illuminationlens under

test

image

plane

auxiliary

lens with fglass plate

with scale

u

f

D

movable for

collimation

collimated

f

Du

2tan

light

sourcecollimator test system

movable

dynameter lensexit

pupil

Page 18: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Setup with Ronchi grating

Measurement of the lateral shift of higher diffraction orders at distance z

High-NA in microscopy: NA>1

Measurement of total internal reflection

of fluorescence light

Measurement of Pupil Size

test

system

image

plane

f

Ronchi

grating

D

Dx

0th order

+1st order

-1st order

z

DExP

a

x

Dxq

Fluorescence intensity

1.5 1.0 0.5 0 0.5 1.0 1.5

rmax

rTIR

pupil illumination

TIR

NA

air uim

immersion

cover

glass

objective

lens

uair

a) b)

Page 19: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement of object sided telecentricity errors by lateral shift of image location during

defocussing

High accuracy measurement by interferometry and measurement of Zernike coefficients

c2/3.

Measurement of Telecentricity

object

space

2DzF

image

plane

exact

telecentric

telecentricity

error

jT

Dx'

Page 20: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement of reflexes at lens vertex points

Analysis of confocal signal in autocollimation

Avoiding spherical aberration induced errors by ring illumination

Measurement of Lens Position

confocal

pinhole

+M

lens under test

sensor

movable for

focusing on one

surface vertex

lenschannel 1

channel 2

laser

source

beam splitter

ring

stop

vertex

Sj

z

DI

confocal

difference signal

S1 S3 S4S2

d1

confocal pinhole -M

Page 21: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement of tilt errors (plane or spherical surface) in autocollimation

Projection of the cross

Observation of lateral shift in Fourier plane

Autocollimation

2x f j

eyepiece

illumination

objectiv

collimator

focal length freticle

measurement

scale

test

object

centered

tilt angle j

beam

splitter

f'

x

j

Page 22: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Autocollimation Principle

Spherical test surface:

- incoming and outgoing wavefront spherical

- concentric waves around center of curvature:

autocollimation

Aspherical test surface

auxiliary lensspherical test

surface

center of

curvature

wavefronts

spherical

auxiliary lens

aspherical test

surface

incoming wavefront

spherical

outcoming wavefront

aspherical

paraxial

center of

curvature

Page 23: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Projection of test marker

Autocollimation of sharp image, focal point coincides with center of curvature of surface

with radius r

Rotation of test system: tilt of surface induces a

lateral shift of the image

Problems with inner surfaces

Measurement of Centering Errors by Reflected Light

rvv M 2

q

illumination

zoom systemlight

source

adjustment of

the image

location

detector

deviation circle

test surface

C

r

tilt

rotating

Page 24: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Thin collimated beam through lens

Focussing of the beam onto detector

Measurement of wedge angle by lateral shift v

Tilt angle of lens not detectable

Not feasible for very short focal lengths

Measurement of Centering Errors in Transmission

f

vnn 21)1()1( aaqj

q

laser

light

source

lens under test

focal length f

wedge error

detector

circle of

deviation

objective

j

inclination

of the axis

rotated

Page 25: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Thin lens with wedge error

Wedge Angle of a Thin Lens

optical axis

v

C2

C1

shape axis

(edge cylinder)

v1v2

1

2

vrr

21

21

11

fn

v

)1(

Page 26: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Mechanical

Ray in transmission

Rays in reflection

Tolerances: Centering Methods

Axially fixed Chuck Pressing Chuck

Direction of

lens motion

Deflected laser beam

Chuck

Green surface after

lens rotation by Incoming laser beam

Red surface

centered

Incoming

laser beam

Chuck

Reflected laser beam

Reflected

laser beam

Incoming laser beam

Red surface

centered

Green surface after

lens rotation by

Ref.: M. Peschka

Page 27: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement of the centering of lens by inspection of the interferogram of front and

rear surface

Centering of a Lens with Interferogram

r1 r2

Z1Z2

D

source

point

Page 28: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Two equivalent forms of description:

1. Three single surface tilt angles

2. Decenter v, overall tilt angle K, tilt of cemented inner surface c , better correspondance to manufacturing steps

Centering of a Cemented Dublet

optical axis

1 / 3

vC3

C1

mechanical axis

c

K

Page 29: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Centering carrier lens

Adjusting second lens

Centering in Bonding Process

Light

source

Cross-hair

Beam splitter

Adjustable

test optics

Centering motion

of lens

Surface

being centered

Collimator

Rotating chuckEyepiece

Reading

scale

Eye / detector

Image of

Cross-hair

Adjustable

test optics

Centering motion

of lens Surface

being centered

Cross-hair

Beam splitter

Collimator

Light

source

Rotating chuck

Eyepiece

Reading

scale

Eye / detector

Image of

Cross-hair

Ref.: M. Peschka

Page 30: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

axis of lens

axis of mounting

mounting

fixating screw

Ds

Dr

a

clearance

tiltangle

Coupling optics – mechanics

Interface glass – metal

Mechanical design of mountings and

housing

Typical options:

1. Filling cylinder with fixating screw

2. Cementing, later centering

3. Lace / bordering

Critical:

- Centering tolerances

- reference surfaces

- analysis of complete geometry

(kinematic)

Tolerances and Mechanical Interface

Page 31: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Mechanicl Mounting Geometry

Filling of lenses into mounting cylinder

with spacers

Accumulation of centering errors by

transportation of reference

Definition of lens positions by:

1. mechanical play inside mounting

2. fixating ring screw

3. planarity of spacers

mounting cylinderfixating screw

spacer

Page 32: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

32

Xray CT-Tomography

Determination of

1. number of lenses

2. thicknesses

3. radii of curvature

4. refractive indices (!?)

Typical uncertainties of linear

dimensions:

0.050 mm ... 0.010 mm

Errors are scaling with size

Contrast for plastics poor

Different absorption for glass types

Cross section vs CT-projection image

Ref: B. Satzer

Page 33: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

33

Xray CT-Tomography

Various cross sections

Ref: B. Satzer

Page 34: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Reasons for reduced system transmission: 1. Absorption in the bulk material of the components

2. Scattering in the bulk materials by inclusions or finite scattering parameters

3. Absorption in the coatings of the surfaces

4. Partial reflection or transmission at the coatings at transmissive or reflective surfaces

5. Blocking of light via mechanical or diaphragm parts of the system due to vignetting

6. Scattering of light by local surface imperfections or non-perfect polished surfaces

7. Deflection of light by diffraction of the light at edges

8. Deflection of light in unwanted higher orders of diffractive elements

Usually strong dependency on:

1. field position

2. wavelength of light

3. used pupil location

4. polarization

Critical:

1. absolute values for test lens

2. influence of auxiliary components

3. change of vignetting and incidence angles

Measurement of Transmission

Page 35: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement of transmission:

Reasons for measurement errors: 1. Absorption in the component materials

2. Absorption in the coatings

3. Finite reflectivity of the coatings

4. Vignetting of the aperture bundle for oblique chief rays

5. Natural vignetting according for oblique chief rays and projection of tilted planes

6. False light from surrounding light sources, which reach the image plane

7. Scattering of light at components of the system mechanical design

8. False light due to ghost images or narcissus in infrared systems

Measurement of Transmission

a) Calibration setup b) Measurement setup

source

collimator

Ulbricht

sphere

mirror

source

collimator

system under test

Ulbricht

spherePin Pin

Pout

out

in

P

PT

Page 36: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement of unwanted light: 2 different approaches:

1. object area black, surrounded by bright source

detection of irradiance in image region

2. intensive isolated point light source in the object plane at different locations

detection of artificial distributions in the image area: glare spread function

Measurement of Ghost Images and Veiling Glare

system under testbright

illumination

field

black

screen

image

plane

irradiance

not

perfectly

black

Page 37: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement of the refractive index of a liquid

Thin film of test liquid between prisms,

adjustment of total internal reflection

Special setup with direct sight prisms, no color fringes

Abbe Refractometer

eyepieceobjectivecompensators

Amici prism without

deviation

measuring

prism

test liquid

telescope

image

direction of the

telescope axis

j

n

a

Page 38: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Standard not sufficient

Appearance:

like wood in transmission shadow image

Definition of Striae

60 mm 50 mm

Page 39: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Advantages:

1. Fast, cheap

2. Simple setup

3. Small requirements on sample surface quality

Disadvantage:

1. no quantitatve calibration until now

Measurement with Shadowgraphy

shadow

image

rotatable desk

screen

sample

quasi point-like

Hg light source

observation

camera

rotatable desk

Page 40: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Change of contrast for

rotation in experiment

Simulation of shadow image

with data of interferograms

Rotation of Sample

shadow image

interferogram

shadow image

calculated out of

interferogram data

Page 41: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement of material data by polarization

Incident linear polarized light

Reflected light elliptical polarized

Compensation of ellipticity, quantitative determination of null-test parameter

Ellipsometry for Polarization Measurement

x

yz

a1

light

source

j

test

piece

linear

polarization

elliptical

polarization

a2

receiver

linear

polarization

compensator

analyzer

variabel

variabel

Page 42: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Polarization

Reasons for polarization influence in optical systems:

1. High numerical aperture effects due to a rotation of the field components

2. Stress birefringence in the bulk materials by production or induced by the

mountings

3. Influence of coatings, this effect is mainly observed at the rim of strongly curved

surfaces due to a large variation of the incidence angle .

Especially mirror coatings are critical for these effects

4. Birefringence in components made of crystals

a) without filter b) with polarization filter

Page 43: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Wavefront

PV-, rms-value, fractional pupil area

Rayleigh-, Marechal criteria for diffraction limit

Point spread function

Strehl ratio, width, second moment, area equivalent, correlation, power in the bucket

Physical Image Quality Criteria

d) Equivalent widtha) Strehl ratio b) Standard deviation c) Light in the bucket

h) Width enclosed areae) Second moment f) Threshold width g) Correlation width

SR / Ds

STDEV

LIBEW

SM FWHM

CW

Ref WEAP=50%

Page 44: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

In principle, the complete information on the optical systems performance can be recovered

form measurements of

1. Point spread function

2. Line spread function

3. Edge spread function

The ESF and the LSF covers the spatial frequencies in the pupil only in one direction.

The complete information requires the measurements of ESF / LSF in several angle

orientations (azimuth)

Between the ESF and the LSF, a simple relationship allows a conversion

There are three symmetry classes, which can be distinguished:

1. Circular symmetry

The Abel transform allows a non-iterative calculation of the PSF from the LSF/ESF

2. Mirror symmetry with decoupling of x and y

The calculation can be performed in two separated 1D-sections

3. General case without symmetry

A complete tomographic reconstruction is necessary

PSF Measurement by Edge Spread Function

i

iESFiLSF

dx

xIdxI

)()(

Page 45: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Characteristic s-shaped curve

for the ernegy transmission

Definition of centroid by corresponding

threshold values

Knife Edge Measurement

-1.0 -0.5 0.0 0.5 1.00

20

40

60

80

100

15%

50%

85%

x

P

x

1

0.2

0.8

centroid

E

0

diameter

0.5

Page 46: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Moving a knife edge perpendicular through

the beam cross section

Relationship between power

transmission and intensity:

Abel transform for circular

symmetry

Example: geometrical spot with spherical aberration

dr

drrrIxP

x

22

)(2)(

z

beforecaustic

zone rays belownear paraxial

focus

Knife Edge Method

y

x

beam

x

knife edge

movement

46

Page 47: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Method very similar to moving knife edge

Integration of slit length must be inverted:

- inverse Radon transform

- corresponds to tomographic methods

Slit-Scan-Method

y

x

beam x

moving slit

d

47

Page 48: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Examples:

Gaussian profile

signaö

Variation of the ratio between

beam width W and slit width v:

- w / v large: width of the slit

can be neglected

- w / v small: strongly changed

profile

-8 -6 -4 -2 0 2 4 6 80

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

x

T

w = 0.3

w = 1.0

w = 2.0

w = 4.0

w = 6.0

2

2

2

2

222),( yx w

y

w

x

yx

eww

PyxI

xx

oSSCw

vxerf

w

vxerfIxI 22

2

1)(

Slit-Scan-Method

Page 49: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Measurement of an edge image

Evaluating the derivative:

Line spread function

Fourier transform:

optical transfer function

MTF-Measurement by Edge Spread Function

'

)'()'(

xd

xIdxI ESF

LSF

)'(ˆ)( xIFsH LSFOTF

x

edge

object

incoherent

illumination

Iin(x)

detector

plane

x'diffracted

light

geometrical

image

of the edge

I(x')

d I(x')

dx'

49

Page 50: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

mathematical relationships

Direct analysis of the frequency content

Problem: zero crossing points

solution: filtering, windowed calculation

MTF-Measurement by Edge Image Analysis

)(

)(')(

A

AHMTF

Page 51: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Setup:

Imaging of a grating

Possible realizations: 1. Density type grating, the sine wave is modelled by gray levels

2. Area type gratings, the sine wave is modelled by geometrical sine-shaped structures

Area coded sine grating:

MTF-Measurement by Imaging Gratings

lamp

sensor

slitlens under

test

diffusor

grating

object

51

Page 52: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

MTF Measurement

Definitions:

1. Fourier transform of PSF

2. Autocorrelation of pupil function

Overview: classification

1. Imaging of special test structures and analyzing the corresponding image contrast

behavior

1.1 If the structures are sine grating structures, a single frequency response is

determined

1.2 If the structures have a large frequency content like points, lines, edges or bar

patterns, a careful analysis of the higher frequency components and calculation

the OTF from the measurement data must be performed

2. Direct measuring of the autocorrelation function of the optical system pupil

corresponding to the Duffieux-integral formulation of the transfer function

3. Measurement of the point spread function and digital calculation of the transfer

function by performing the Fourier transform.

''','),(

''2dydxeyxIH

yxi

PSFyxOTFyx

pppp

pp

y

px

p

y

px

p

yxOTF

dydxyxP

dydxf

yf

xPf

yf

xP

H

2

*

),(

)2

',

2

'()

2

',

2

'(

),(

Page 53: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

MTF Measurement

Conditions for measuring the incoherent transfer function:

1. An object is illuminated by incoherent light.

2. The object acquires, through its structures, all relevant spatial frequencies that

have to be measured.

3. The object is imaged by the test system.

4. Spatial resolution is provided for the detection of the image intensity. As a rule this

is achieved by an adjustable slit located in front of the detector. Alternatively, the

slit can be fixed and scanning is accomplished by the imaged grating or object

structure.

5. The contrast is derived from the intensity distribution and analysed as a function of

the spatial frequency

Possible test structures of the object:

1. Point object

2. Edge object

3. Line object or slits

4. Bar pattern

5. Random transparencies

6. Sine gratings with one or several periods

7. Special test charts like the Siemens star

Page 54: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Realization by a rotating radial grating

Spatial frequency depends on azimuthal angle

MTF-Measurement by Imaging Gratings

rotation for spatial

scanning

rotation for the angle

of the slit / grating :

spatial frequency

slit

q

slit

q

g

a

grating

1/s

effective

grating period

Page 55: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Basis : Duffieux integral

MTF Measurement by Pupil Autocorrelation

laser

sourcefocusing

lens

stop

lens under

test

beam

splitter

roof

mirror

variable

shear

roof

mirror

lens

diffuser

sensor

variable

phase

pppp

pp

y

px

p

y

px

p

yxOTF

dydxyxP

dydxf

yf

xPf

yf

xP

H

2

*

),(

)2

',

2

'()

2

',

2

'(

),(

Page 56: Metrology and SensingMetrology+and... · Wavefront sensors Hartmann-Shack WFS, Hartmann method, ... Telecentric movable measurement microscope with offset y: ... Cross-hair …

Sources of Errors in MTF Measurements

Typical shortcomings of MTF measurements:

1. Mechanical tolerances of the movable parts of the setup like line scan, rotatable

edges and alignment errors

2. Application of precise correction factors for finite size slits

3. Truncation errors of the finite lengths structures of the object

4. Calibration of the spatial frequency variable, in particular for finite fields of view

with projection changes of lengths and pattern widths

5. Poorly known residual aberrations of auxiliary optical components

6. Use of incorrect spectral and coherence constraints of the illumination

7. Shortcomings of sensor performance

8. Perturbing glare and stray light