methods for ltq orbitrap a guided tour with examples
DESCRIPTION
Methods for LTQ Orbitrap A Guided Tour with Examples. Dr. Michaela Scigelova. Outline. Data acquisition strategies Method setup and examples OT_LTQ_Big6 MS in Orbi + 6 MS/MS in ion trap OT_3OT MS in Orbi + 3 MS/MS in Orbi NL_MS3_AccMass Phosphopeptides with accurate NL - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/1.jpg)
The world leader in serving science
Methods for LTQ OrbitrapA Guided Tour with Examples
Dr. Michaela Scigelova
![Page 2: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/2.jpg)
2
Outline
Data acquisition strategies Method setup and examples
• OT_LTQ_Big6• MS in Orbi + 6 MS/MS in ion trap
• OT_3OT• MS in Orbi + 3 MS/MS in Orbi
• NL_MS3_AccMass• Phosphopeptides with accurate NL • MS in Orbi, MS2 in Orbi, and MS3 in ion trap
• OT_MSA• Phosphopeptides• ‘composite’ MS2 and MS3 spectrum = MSA• MS in Orbi, MSA in LTQ
![Page 3: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/3.jpg)
3
What is the objective of the analysis?
One or more of these:• Maximise the protein ID• Boost the sequence coverage• Phosphorylation• De novo sequencing• Quantitation
![Page 4: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/4.jpg)
4
Data Acquisition Strategies
Parallel mode• Survey MS in Orbitrap• MS/MS in LTQ
For highest protein ID and coverage
assuming very complex mixtures
Serial mode• Survey MS in Orbitrap • MS/MS in Orbitrap
For de novo sequencing, PTM discovery
![Page 5: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/5.jpg)
5
Mass Accuracy and Resolution Settings
Resolution Mass Accuracy
Survey MS (OT) 15,000 3 ppm
MS/MS (OT) 7,500 3 ppm
Resolution Mass Accuracy
Survey MS (OT) 60,000 3 ppm
MS/MS (LTQ) Unit ~200 ppm
Parallel Mode
Serial Mode
0.9 s
0.1 s
0.35 s
0.2 s
![Page 6: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/6.jpg)
6
AGC settings LTQ
Full MS 3.00e+04 IT 50ms SIM 1.00e+04 IT 100ms MSn 1.00e+04 IT 100ms
![Page 7: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/7.jpg)
7
AGC settings orbitrap
Full MS 5.00e+05 - 1.00e+06 IT 500ms SIM 1.00e+05 - 2.00e+05 IT 500ms MSn 1.00e+05 - 2.0e+05 IT 500ms
![Page 8: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/8.jpg)
8
Settings for HCD, Lock mass
Act.Q 0.13-0.14 Normalized Collision Energy 60-70
specify the FT lock mass abundance (%) Diagnostics/Set Device Default value 10%
![Page 9: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/9.jpg)
The world leader in serving science
OT_LTQ_Big6
MS in Orbi + 6 MS/MS in ion trap
![Page 10: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/10.jpg)
10
Parallel Mode – Method Setup
Step-by-step guide Method BIG 6 Tune settings
• Max fill time 100 ms• Number of microscans 1
![Page 11: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/11.jpg)
11
Parallel Method – starting the method
![Page 12: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/12.jpg)
12
Parallel Method – Survey MS
![Page 13: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/13.jpg)
13
Parallel Method – Add another 6 scan events
![Page 14: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/14.jpg)
14
Parallel Method – BIG 6
Save yourself the typing: use copy and paste!
![Page 15: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/15.jpg)
15
Parallel Method – DDA Scan event 2
Can be used for directing MS/MS to a preferred region – e.g. 800-1300 in glycopeptide analysis
![Page 16: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/16.jpg)
16
Parallel Method – DDA Scan event 2
![Page 17: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/17.jpg)
17
Parallel Method – DDA Scan event 2
Ignore
![Page 18: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/18.jpg)
18
Parallel Method – DDA Scan event 2
This is the key to PARALLEL mode
![Page 19: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/19.jpg)
19
Parallel Method – DDA Scan event 2
![Page 20: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/20.jpg)
20
Parallel Method – DDA Scan event 2
Import lists of target parent ions (parent list) or contaminants (reject list)
![Page 21: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/21.jpg)
21
Parallel Method – DDA Scan event 2
Reject ‘1+’ and ‘unassigned’ for max number of PROTEIN ID
![Page 22: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/22.jpg)
22
Parallel Method – DDA Scan event 2
Leave unticked if going for max sequence COVERAGE
![Page 23: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/23.jpg)
23
Parallel Method – DDA Scan event 2
Even better COVERAGE: if you inject sample 3 times and set the following:
![Page 24: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/24.jpg)
24
Parallel Method – DDA Scan event 2
![Page 25: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/25.jpg)
25
Parallel Method – DDA Scan event 3
Default settings:
WRONG!!!!
!
WRONG!!!!
!
![Page 26: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/26.jpg)
26
Parallel Method – DDA Scan event 3
![Page 27: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/27.jpg)
27
Parallel Method – DDA Scan event 4
![Page 28: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/28.jpg)
28
Parallel Method – DDA Scan event 5
![Page 29: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/29.jpg)
29
Parallel Method – DDA Scan event 6
![Page 30: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/30.jpg)
30
Parallel Method – DDA Scan event 7
![Page 31: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/31.jpg)
31
When you have built your method.. Check it!
![Page 32: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/32.jpg)
32
Who Is Working this Way?
John Yates (Scribbs Inst.) Ruedi Aebersold (ETH Basel) David Goodlet (ISB Washington)
Yates et al., AC 2006, 78, 493-450
![Page 33: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/33.jpg)
The world leader in serving science
OT_3OT
MS in Orbi + 3 MS/MS fragmented in LTQ and measured in OT
![Page 34: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/34.jpg)
34
Serial Method Setup
Full scan in Orbi + 3 MS/MS in Orbi Tune settings
• Max fill time 100 ms• Number of microscans 2 to improve S/N in MSn
Resolution Mass Accuracy
Survey MS (OT) 15,000 3 ppm
MS/MS (OT) 7,500 3 ppm
Can resolve 4+ ions
0.35 s
0.2 s
![Page 35: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/35.jpg)
35
Serial Method – Full scan MS
![Page 36: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/36.jpg)
36
Serial Method – MS + 3 MS/MS
Save yourself the typing: use copy and paste!
![Page 37: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/37.jpg)
37
Serial Method – Major change in ‘Current Segment’
There is NO box ticked!
![Page 38: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/38.jpg)
38
Serial Method – DDA Scan event 2
![Page 39: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/39.jpg)
39
Serial Method – DDA Scan event 3
![Page 40: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/40.jpg)
40
Serial Method – DDA Scan event 4
![Page 41: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/41.jpg)
41
Who Is Doing It This Way?
Matthias Mann (MPI Martinsreid)• Lock mass ‘trick’• Divide the mass range is sections
Roman Zubarev (Uppsala)
Olsen et al., MPC 2005, 4, 2010-2021
![Page 42: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/42.jpg)
42
Want ppb Mass Accuracy?
Use Lock mass in the method
Olsen et al., MPC 2005, 4, 2010-2021
![Page 43: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/43.jpg)
43
Lock Mass - Caveats
Lock mass should be present in your analysis all the time • Can be a background ion• Or spike in a lock mass solution via a Y-connector (P-773,17 nL
swept volume) with fused silica tubing attached to a syringe pump Use multiple lock masses For calibrating MS/MS spectra the lock mass closest to your
parent mass of interest is taken (good for peptides) Once you fill in the lock mass dialog box, you can export it, save
it, and import it to other methods
![Page 44: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/44.jpg)
The world leader in serving science
NL_MS3_AccMass
MS in Orbi + MS/MS in Orbi + MS3 in LTQ
![Page 45: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/45.jpg)
45
Strategy for Phosphopeptides
Enrichment!!! Use prominent neutral loss of phosphate in MS2 to trigger MS3 Accurately defined neutral loss
• Use Orbi for MS and MS2• Cuts down on false positive MS3 triggers
MS3 is highly informative for peptide sequence and PTM position Data review is fast and easy – search MS3 scans only Thorough search (MS2 + MS3) provides high confidence ID
Olsen and Mann, PNAS 2004, 101, 13417-22
![Page 46: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/46.jpg)
46
NL MS3 Method – starting the method
![Page 47: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/47.jpg)
47
NL MS3 Method – Survey MS
![Page 48: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/48.jpg)
48
NL MS3 Method – MS2
![Page 49: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/49.jpg)
49
NOTE: phosphopeptide analysis
NL MS3 Method – MS2
Accurate NL determination avoids false triggers of MS3 more efficient analysis
![Page 50: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/50.jpg)
50
NOTE: phosphopeptide analysis
NL MS3 Method – MS2
![Page 51: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/51.jpg)
51
NOTE: phosphopeptide analysis
NL MS3 Method – MS2
![Page 52: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/52.jpg)
52
NOTE: phosphopeptide analysis
The system recalculates NL for all relevant charge states
NL MS3 Method – MS2
![Page 53: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/53.jpg)
53
NOTE: phosphopeptide analysis
NL MS3 Method – MS2
![Page 54: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/54.jpg)
54
NL MS3 Method – MS3
![Page 55: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/55.jpg)
55
NL MS3 with accurate mass support – MS3
![Page 56: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/56.jpg)
The world leader in serving science
NL_MSA
MS in Orbi + MSA in LTQ
![Page 57: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/57.jpg)
57
Multistage Activation - MSA
To improve the signal, it is advantageous to lump together the fragments from both MS2 and MS3 scans and search them as one ‘strong’ scan
![Page 58: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/58.jpg)
58
CID MS² CID MS³ [M+2H-98]2+
Phosphopeptide analysis with MSA
APPDNLPSPGGpSR 5 fmol
MSA
Data courtesy of Karl Mechtler
![Page 59: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/59.jpg)
59
MSA Method – starting the method
![Page 60: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/60.jpg)
60
MSA Method – Survey MS
![Page 61: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/61.jpg)
61
MSA – data dependent MS2
![Page 62: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/62.jpg)
62
Dynamic exclusion
![Page 63: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/63.jpg)
63
Neutral Loss – measured in LTQ
![Page 64: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/64.jpg)
64
Current Segment
![Page 65: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/65.jpg)
65
MSn Settings
![Page 66: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/66.jpg)
66
Monoisotopic precursor selection
![Page 67: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/67.jpg)
67
Neutral Loss – no need to spell it our for different charge states
![Page 68: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/68.jpg)
68
Multistage Activation - ON
![Page 69: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/69.jpg)
69
MSA - Conclusions
MSA enables to activate ions from MS2 that show neutral loss from parent, while still keeping the rest of the MS2 fragments in the trap. The MS3 fragments from activated NL candidates are ‘added’ to the rest of the MS2 fragments
Data can be processed by SEQUEST/BioWorks or Mascot
![Page 70: Methods for LTQ Orbitrap A Guided Tour with Examples](https://reader033.vdocuments.us/reader033/viewer/2022061503/568145b7550346895db2bd38/html5/thumbnails/70.jpg)
70
Who is Doing it this Way
Donald Hunt (Uni Virginia) Karl Mechtler (IMP, Vienna) Andy West, Peter Francis (GSK, Stevenage)
Schoeder et al, AC2004, 76, 3590-3598.