Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

A MINIATURE MASS SPECTROMETER AS A MELTING PROBE PAYLOAD

INSTRUMENT

S. SheridanPlanetary and Space Sciences Research

Institute, Open University, Milton Keynes,

MK7 6AA, U.K. [email protected]

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

Talk overview

Context: PSSRI research and instrumentation• Previous missions• Mass spectrometry background

Proposed payload for a melting probe

Results

Future work and conclusions

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

PSSRI instrument heritageFrom sample analysis

• Lunar samples• Meteorites• Stardust, Genesis…

…to in-situ instruments• Cassini-Huygens

– SSP, HASI, CDA• Beagle 2

– ESS, GAP• Rosetta

– Ptolemy, MUPUS, GIADA

And also enabling research:• Medical/life sciences • Next generation instruments• MEMS

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

The Gap instrument

GAP on Beagle2 (6 kg)

High precision (<<1%)

•Solid sample inlet (carousel, MPAe)

•Chemical sample processing

•Chemical / getter pumps

•Onboard reference material

•Magnetic sector mass spectrometer (2kg, 5W)

•VERY complex sample preparation

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

The Ptolemy instrument

Ptolemy on Rosetta lander (<5 kg)

Medium precision (1 to 5 %)•Solid sample inlet (carousel, MPAe)

•Chemical sample processing

•3-4 channel gas chromatograph

•Ion trap mass spectrometer (500 g, 1 W)• MEMS field emission electron source

(CCLRC RAL)• Digital ion counting (no noise)• m/z 10-200 • C, H, N, O, (S?)

•VERY complex sample preparationQM Post Vibration Test: E-Box Panels Removed to Reveal PCBs

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

•Simple device:

• Low mass, low power & rugged• 3 electrodes

– 3D trapping field

– r.f. (approximately 1 MHz)

• Electron multiplier detector• Ion source

•Fast scanning (m/z 10-250 in 100 ms)

•Operates under poor vacuum (1x10-3 mbar)

•Software control of one parameter (r.f.)

Limitations:

•Requires sophisticated drive electronics

The Ion trap Mass Spectrometer

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

PSSRI Ion trap interests

•Melting probe instrument• terrestrial (Lake vostock)• Europa, Mars poles

•EVITA for an Instrumented mole

•Active EDLS – measuring atmospheric properties (pressure, density, composition during descent e.g. to Mars)

•Monitoring breathing air

•Venus Entry Probe, balloon, • 100 gram microprobes

– A batch of MEMS traps (due in next few weeks)

•PhD student (1 Oct 2006-) sub-1 kg MS for planetary missions

•Ongoing work in support of the Rosetta mission

•Wellcome trust project - Breath analyser

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

The EVITA instrument

•EVITA Evolved Volatiles Ion Trap Analyser

•Initially developed for instrumented mole deployment

• Modular system• Mission specific sampling method• Airless bodies:

– Bepicolombo– NEO missions– Luna missions

• Mars missions– Miniature Turbo pump (130 grams)

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

Ion trap on a Melting ProbeInstrument proposal:

•Miniature ion trap (EVITA type) mass spectrometer• m/z of 10-200 amu

• measurement of – either gaseous volatile compounds (if sampling in a non-aqueous environment)– or dissolved organic compounds (if sampling in an aqueous environment)– the above measurements to PPM levels

•Use technology derived from previous projects•Low mass (<500 g) and low power (<1 W)

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

Sample inlet / preparation (1)

•Direct sampling of the environment •Feedback enables constant pressure (flow)

> 1x10-10 to 1x10-4 mbar l sec-1

•Very simple - a variable capillary leak•Pressure tested to 150 bar (Helium)

> Used on the Ptolemy instrument•Will allow evacuation during cruse phase

Limitations:

•no selective control of sample species> Lets everything into the mass spectrometer

•problem with aqueous environment> large gas load - Pumping requirement

(1) Thermally actuated proportional valve

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

Sample inlet / preparation (2)

•Membrane is the interface between the liquid and the vacuum of the mass spectrometer •very low gas load in aqueous environments

>minimal pumping required •Rugged / simple devices•Negate extensive sample preparation•Continuous monitoring possible

Limitations:

•Sensitivity depends on hydrophobicity and volatility of the analyte.

2) Membrane inlet

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

Ion trap for a Melting Probe

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

The prototype ion trapPrototype electronics Prototype hardware

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

Future work• More representative system

• A vacuum vessel has been designed and is under construction to allow testing of the Ion trap with a MIMS and a direct inlet system

– Sampling from liquids

– investigation of pressure effects / length of operation with the mini chemical getter pump

• Replace the Filament used in the prototype ion trap:• A batch of MEMS nanotip electron sources is being manufactured by RAL

– testing and characterisation of these devices with the ion trap system

• A alternative batch of miniature, rugged and low power electron sources has been procured.

– A testing programme is currently underway to evaluate these devices with the Ion trap

• Development of new electron multiplier devices

• Further development of the Ion trap drive electronics is now required:• Funds required to enable a study to evaluate / design and manufacture an ASIC.

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review

Conclusions• PSSRI has an active, ongoing programme of missions and technology

development involving small ion trap mass spectrometers and the associated inlet systems

• A prototype Ion trap instrument has been developed which is low mass, small, flexible in operation and could be carried by the proposed melting probe

• Two sampling methods have been discussed which are compatible with the envisaged environment in which the melting probe will operate

• Ongoing characterisation / development of the system is currently underway• Further development of the electronics, with the aim of designing and the

eventual manufacture of an ASIC solution (maybe part of a common electronics solution???)

• Looking for funding opportunities to achieve this

Simon SheridanMelting Probe Review

Simon SheridanGraz 11 April 2006

Melting Probe Review