medical optics
TRANSCRIPT
-
8/2/2019 Medical Optics
1/254
Bio Medical Optics
Jayakumar D Swamy M.Sc., M.Tech.,
Optical Engineer
-
8/2/2019 Medical Optics
2/254
Definition of cataract Opacity of the lens, which occurs when fluid gathers between
the lens fibers.When eyes work properly:
Light passes through the cornea and the pupil to the lens.
The lens focuses light & producingclear, sharp images on theretina.
As a cataract develops, the lens becomes clouded, whichscatters the light and prevents a sharply defined image fromreaching retina. As a result, vision becomes blurred.
-
8/2/2019 Medical Optics
3/254
-
8/2/2019 Medical Optics
4/254
Causes of cataract
Old age (commonest)
Ocular & systemic diseases
DM
Uveitis
Previous ocular surgery
Systemic medication
Steroids
Phenothiazines
Trauma & intraocular foreign
bodies Ionizing radiation
X-ray
UV
Congenital
Dominant
Sporadic
Part of a syndrome
Abnormal galactosemetabolism
Hypoglycemia
Inherited abnormality
Myotonic dystrophy
Marfans syndrom
Rubella
High myopia
-
8/2/2019 Medical Optics
5/254
Cataract
Divided to :
Acquired cataract
Age - related cataract
Presenile cataract
Traumatic cataract
Drug induced cataract
Secondary cataract
Congenital Cataract
Systemic associationNon-systemic association
-
8/2/2019 Medical Optics
6/254
Age -related cataract
It is the Most commonly occurred.Classified according to:
Morphological Classification
Nuclear
Cortical
Subcapsular
Christmas tree uncommon
Maturity classification
Immature Cataract
Mature Cataract
Hypermature Cataract
-
8/2/2019 Medical Optics
7/254
Nuclear cataract Most common type
Age-related
Occur in the center of the lens. In its early stages, as the lens changes theway it focuses light, patient may becomemore nearsighted or even experience atemporary improvement in reading vision.Some people actually stop needing their
glasses. Unfortunately, this so-called 2nd sightdisappears as the lens gradually turns moredensely yellow & further clouds vision.
As the cataract progresses, the lens mayeven turn brown. Advanced discoloration
can lead to difficulty distinguishing betweenshades of blue & purple.
-
8/2/2019 Medical Optics
8/254
Cortical cataract
Occur on the outer edge of the lens (cortex).
Begins as whitish, wedge-shaped opacities or streaks.
Its slowly progresses, the streaks extend to the center and
interfere with light passing through the center of the lens.
Problems with glare are common with this type of cataract.
-
8/2/2019 Medical Optics
9/254
Symptoms
A cataract usually developsslowly, so:
Causes no pain.
Cloudiness may affect only asmall part of the lens
People may be unaware of anyvision loss.
Over time, however, as thecataract grows larger, it:
Clouds more the lens
Distorts the light passingthrough the lens.
Impairs vision
Reduced visual acuity (near
and distant object)
Glare in sunshine or with
street/car lights.
Distortion of lines.
Monocular diplopia.
Altered colours ( white
objects appear yellowish) Not associated with pain,
discharge or redness of the
eye
-
8/2/2019 Medical Optics
10/254
Signs
Reduced acuity.
An abnormally dim red reflex is seen when the eye is viewedwith an ophthalmoscope.
Reduced contrast sensitivity can be measured by the
ophthalmologist. Only sever dense cataracts causing severely impaired visioncause a white pupil.
After pupils have been dilated, slit lamp examination shows thetype of cataract.
-
8/2/2019 Medical Optics
11/254
Gradual loss of vision
DDX:
1. Cataract
2. Glaucoma
3. Diabetic retinopathy4. Hypertensive retinopathy
5. Age related macular degeneration
6. Retinitis pigmentosa
7. Trachoma8. Onchocerciasis (river blindness)
9. Vitamin A deficiency
-
8/2/2019 Medical Optics
12/254
Treatment
Glasses: Cataract alters the refractive power of the natural lens
so glasses may allow good vision to be maintained.
Surgical removal: when visual acuity can't be improved with
glasses.
Surgical techniques
Phacoemulsification method.
Extracapsular method. Intracapsular method
-
8/2/2019 Medical Optics
13/254
Pre-op assesments
General health evaluation including blood pressure check
Assessment of patients ability to co-operate with theprocedure and lie reasonably flat during surgery
Instruction on eye drop instillation
The eyes should have a normal pressure, or any pre-existingglaucoma should be adequately controlled on medications.
An operating microscope is needed, in order to reach the lens,a small corneal incision is made close to the limbus for the
phaco-probe. It is important to appreciate anterior chamber depth and to
keep all instruments away from the corneal endothelium in theplane of the iris.
-
8/2/2019 Medical Optics
14/254
-
8/2/2019 Medical Optics
15/254
-
8/2/2019 Medical Optics
16/254
Pupil conjugate plane
Lamp
Aperture
45 mirror
Practical Retinal Illumination System
-
8/2/2019 Medical Optics
17/254
Retinal Imaging System
-
8/2/2019 Medical Optics
18/254
stop in pupil planeconjugate
circular aperturein stop
opticaxis
SUBJECT
-
8/2/2019 Medical Optics
19/254
opticaxis
observers pupil inconjugate pupil plane
SUBJECT
OBSERVER
-
8/2/2019 Medical Optics
20/254
FIRST ATTEMPT AT BINOCULAR VIEW
Obs. L eye
Obs. R eye
Ss eye
Combine L and R eye views
Observers eyes have to be too close
-
8/2/2019 Medical Optics
21/254
-
8/2/2019 Medical Optics
22/254
-
8/2/2019 Medical Optics
23/254GTT 05
OPHTHALMOSCOPE MAGNIFICATION
-
8/2/2019 Medical Optics
24/254
20 DlensRI
60 Deye
OPHTHALMOSCOPE MAGNIFICATION
Mag of RI
Peye
Plens=
60 D
20 D
= 3.0M =
-
8/2/2019 Medical Optics
25/254
42
40 mm
50 mm
20 D
1 mm dia exit pupil
2.0 mm
MONOCULAR FIELD OF VIEW
-
8/2/2019 Medical Optics
26/254
20 D
40
Area of binocular view
BINOCULAR FIELD OF VIEW
GTT 04
-
8/2/2019 Medical Optics
27/254
Clear Aperture: CLAP
Working Distance: WD CLAP2
WD
54.72 mm
51.04 mm
= 24 mm
47 mm
CLAPWD= tan-1 ( (
= 23.7 FOV = 47.4 = 25.2 FOV = 50.3 = 25.2 FOV = 50.3
Example: OI Maxlight 20 DCLAP = 48 mmWD = 47 mmFOV = 50
ESTIMATING FIELD OF VIEW
-
8/2/2019 Medical Optics
28/254
Complete binocular
Indirect
ophthalmoscope
-
8/2/2019 Medical Optics
29/254
GTT 05
-
8/2/2019 Medical Optics
30/254
hole in 45o mirror
-
8/2/2019 Medical Optics
31/254
camera or CCD
Fundus camera
-
8/2/2019 Medical Optics
32/254
3. Optical Coherence Tomography (OCT)
h f h
-
8/2/2019 Medical Optics
33/254
coherent incoherent
partially coherent
Coherence of Light Waves
-
8/2/2019 Medical Optics
34/254
Laser Beam Coherence
Laser
coherence
length
-
8/2/2019 Medical Optics
35/254
fixedmirror
movable
mirror
laser
negative lens
screen
interferencefringes
beam-splittingprism
L1
L1
L2
Michelson Interferometer
reference arm
sample arm
-
8/2/2019 Medical Optics
36/254
screen
plane wavesfrom fixedmirror
plane wavesfrom movablemirror
Interference Fringes in Michelson Interferometer
-
8/2/2019 Medical Optics
37/254
low coherence length
long coherence length
Mi h l I t f t O ti l C h T h
-
8/2/2019 Medical Optics
38/254
movable
mirror
laser
negative lens
fixedmirror
Michelson Interferometer Optical Coherence Tomography
photodetector
electronics
video monitor
-
8/2/2019 Medical Optics
39/254
lateral (X)scanningmirror
negative lens
axial(Z-axis)
scan
photodetector
sample
video monitor
electronics
reference arm
sample arm
-
8/2/2019 Medical Optics
40/254
Fringes form when reference mirror path length matches path
length of a reflective piece in the tissue in the sample arm.
Fringes only form when the path difference is within the
coherence length of the light source.
IN MICHELSON INTERFEROMETER
-
8/2/2019 Medical Optics
41/254
lateral (X)scanningmirror
negative lens
axial(Z-axis)
scan
photodetector
video monitor
electronics
A SCAN
B SCAN
OCT using fiber optics
-
8/2/2019 Medical Optics
42/254
electronics
photodetector SLD
sample
reference
-
8/2/2019 Medical Optics
43/254
GTT/98
-
8/2/2019 Medical Optics
44/254
MICROSCOPES
ANGULAR MAGNIFICATION
-
8/2/2019 Medical Optics
45/254
Apparent size of object depends on angle it subtends at eye.
100 m
100 m
10 m
10 m
-
8/2/2019 Medical Optics
46/254
25 cm
On average, an object cannot be closer than 25 cm from the eye to be seen clearly.
Average distance of
most distinct vision
ANGULAR MAGNIFICATION
-
8/2/2019 Medical Optics
47/254
25 cm
f
h
25 cm
f
h
tanh
25
tan hf
Angular Magnification =tan
tan h25
h
f= =
25
f
virtual image
cm
(cm)
-
8/2/2019 Medical Optics
48/254
Eye
Eyepiece
Objective
Object f
Real imagef
objective
eyepiece
BASIC MICROSCOPE
magnifier
real image
magnification
-
8/2/2019 Medical Optics
49/254
M1
=Im
Ob
M2
=25
f
Mtotal
=Im
Ob
25
f
X
MICROSCOPE MAGNIFICATION
OBJECTIVES
-
8/2/2019 Medical Optics
50/254
anw.d.
D
NA = sinn a
a= 14
n = 1.00 (air)
EXAMPLE
NA = 1.00 x sin(14 )
NA = 0.24
OBJECTIVESNumerical Aperture (NA)
Light gathering ability Resolution
OBJECTIVES
-
8/2/2019 Medical Optics
51/254
a = 28n =1.00
NA = 0.46
a = 35n =1.00
NA = 0.57
a = 60n =1.52 (oil)
NA = 1.32
OBJECTIVES
N.A. Examples
EYEPIECES
-
8/2/2019 Medical Optics
52/254
converging
rays from
objective
Real image
Real image
parallel rays from
eyepiece
Huygens RamsdenEYEPIECES
(OCULARS)
D
-
8/2/2019 Medical Optics
53/254
Specimen
Objective
Real image
REAL MICROSCOPE
EXPERIMENT 4
-
8/2/2019 Medical Optics
54/254
EXPERIMENT 4
Basic Microscope
onion skin
real image
on card
f
Produce real image of onion skin on card.
Mark distance of real image on base.
irisdiaphragm
EXPERIMENT 4--CONTINUED
-
8/2/2019 Medical Optics
55/254
View real image with magnifier (eyepiece)
real
image
plane
f
Adjust iris diaphragm. How does image change?
What is the total magnification? Mtotal=
Im
Ob
25
fX
-
8/2/2019 Medical Optics
56/254
Slit-lamp Biomicroscope
-
8/2/2019 Medical Optics
57/254
The slit-lamp biomicroscope
begins with a microscope.
Objective
Specimen
Eyepiece
turned on its side
-
8/2/2019 Medical Optics
58/254
.turned on its side
subjectobjective
Huygens
eyepiece
.fundamental slit-lamp biomicroscope
.change specimen, objective & eyepiece
image
plane
B ild i ifi ti h ith t h i ki
-
8/2/2019 Medical Optics
59/254
Build in magnification change without changing working
distance
fobj
Galilean
telescope to
change mag
working
distance
no image in
image plane
B ild i ifi ti h ith t h i ki
-
8/2/2019 Medical Optics
60/254
Build in magnification change without changing working
distance
fobj
Galilean
telescope to
change mag
working
distance
no image in
image planeD
-
8/2/2019 Medical Optics
61/254
..add lens to form image in eyepiece image plane
astronomical
telescope
D
Porro* prism
-
8/2/2019 Medical Optics
62/254
F
F
p
2 right-angle prisms
1800 image rotation
reduce length of
telescope
displace image
horizontally
Porro -Abbe
-
8/2/2019 Medical Optics
63/254
Slit-lamp with folded optical path
D
-
8/2/2019 Medical Optics
64/254
D
-
8/2/2019 Medical Optics
65/254
binocular slit-lamp viewing system
-
8/2/2019 Medical Optics
66/254
Anatomy of the ApodizedDiffractive Technology
Central 3.6 mm apodized
diffractive structure
Step heightsdecrease peripherally from
1.3 0.2 microns
A +4 D at lens plane
equaling +3.2 at spectacle
plane
Outer refractive zone
Anterior
aspheric
optic
-
8/2/2019 Medical Optics
67/254
Anatomy of the ApodizedDiffractive Technology
13.0 mm
Anterior
Apodized
Diffractive Optic
6.0 mm6.0 mm
Symmetric
Biconvex
Anterior Aspheric
Optic
-
8/2/2019 Medical Optics
68/254
Gradual reduction or blendingof the diffractive step heights.
Optimally manages light energy
delivered to the retina as itdistributes the appropriateamount of light to near anddistant focal points, regardlessof the lighting situation.
Designed to improve imagequality while minimizing visualdisturbances.
Apodization
1.3 micron
step
-
8/2/2019 Medical Optics
69/254
Technology of the AcrySofReSTORIOL in Human Terms
Thickness of a Human Hair= 60 microns
Thickness of a Red Blood Cell= 7 microns
Step Height at periphery ofthe diffractive portion of theAcrySof ReSTOR AsphericIOL = 0.2 microns
-
8/2/2019 Medical Optics
70/254
Design considerations for the AcrySof
ReSTOR Aspheric IOL:
Induce negative Spherical Aberrationswith the lens to compensate for positive
corneal Spherical Aberrations
Design Objective
-
8/2/2019 Medical Optics
71/254
Spherical aberration occurs when light rays areover-refracted at the periphery of a lens system,resulting in a region of defocused light which candecrease image quality.
The Problem Spherical Optics
Spherical
Aberration
Marginal Rays
Paraxial Rays
Light Rays
Spherical IOL
*Smith, G., Atchinson D.A., (1997) The Eye and Visual Optical Instruments. Cambridge University Press, Cambridge, United Kingdom, pp. 667.
-
8/2/2019 Medical Optics
72/254
Aspheric
Surface
Aspheric optics align the light rays to compensate for positivecorneal spherical aberration, resulting in enhanced imagequality.
The Solution Aspheric Optics
*Smith, G., Atchinson D.A., (1997) The Eye and Visual Optical Instruments. Cambridge University Press, Cambridge, United Kingdom, pp. 667.
Light Rays
Aspheric IOL
-
8/2/2019 Medical Optics
73/254
73
AcrySof ReSTOR Aspheric IOL(SN6AD3) Specifications
13.0 mm
6.0 mm
OPTICS
Multifocal Apodized Diffractive Optic
Compensation for PositiveCorneal Spherical Aberration
Aspheric Optic
Optic Type Proprietary Symmetric Biconvex
Optic Diameter 6.0 mmOverall Length 13.0 mm
MATERIAL
Optic/Haptic Material AcrySofHydrophobic Acrylic
Light Filtration UV and High-Energy Blue
DESIGN
IOL Design Single-Piece
Haptic Design STABLEFORCEModified-L
SPECIFICATIONS
Diopter Range +10 D +30 D in 0.5 D increments+31 D +34 D in 1.0 D increments
-
8/2/2019 Medical Optics
74/254
Surgical Loupe
-
8/2/2019 Medical Optics
75/254
OCT: Basic Principles
Three-dimensional imaging technique with highspatial resolution and large penetration depth even
in highly scattering media
Based on measurements of the reflected light fromtissue discontinuities
e.g. the epidermis-dermis junction.
Based on interferometry
interference between the reflected light and the reference
beam is used as a coherence gate to isolate light from
specific depth.
-
8/2/2019 Medical Optics
76/254
1 mm 1 cm 10 cm
Penetration depth (log)
1 mm
10 mm
100 mm
1 mm
Resolution (log)
OCTConfocal
microscopy
Ultrasound
Standard
clinical
High
frequency
OCT vs. standard imaging
OCT in non invasive
-
8/2/2019 Medical Optics
77/254
OCT in non-invasive
diagnostics
Ophthalmology diagnosing retinal diseases.
Dermatology skin diseases,
early detection of skin cancers.
Cardio-vascular diseases vulnerable plaque detection.
Endoscopy (fiber-optic devices) gastroenterology
gynecology
Embryology/Developmentalbiology
Functional imaging Doppler OCT (blood flow)
spectroscopic OCT (absorption, highspeed)
optical properties
Polarization Sensitive-OCT(birefringence).
Guided surgery
delicate procedures
brain surgery,
knee surgery
OCT: Principle of operation
-
8/2/2019 Medical Optics
78/254
OCT: Principle of operation
OCT is analogous to ultrasound imaging
Uses infrared light instead of sound
Interferometry
is used to measuresmall time delays
of scattered photons
Human skin
5 mm wide x 1.6 mm deepSpatialResolution: 10-30 m
Time resolution: 30fs!!!
Speed of sound ~ 1480 m/sec (in water)
Speed of light 3x108 m/sec
-
8/2/2019 Medical Optics
79/254
-
8/2/2019 Medical Optics
80/254
-
8/2/2019 Medical Optics
81/254
Good OCT sources have small coherence length and large bandwidth
Axial resolution
-
8/2/2019 Medical Optics
82/254
Axial resolution
The axial resolution is
notice that D is the 3dB-bandwidth!
The broader the bandwidth the shorter the
coherence length and the higher the resolution
2 2
0 02 ln 2 1 2 ln 2
0.44c
cl
D D D
-
8/2/2019 Medical Optics
83/254
Lateral resolution: Decoupled from
-
8/2/2019 Medical Optics
84/254
axial resolution
xD
2 xD4 f
xd
D
Lateral resolution
Dz
Dz
Dz
High NA
Low NA
xDb Dz
Lateral resolution similar to that in a standard microscope
f=focal length
d= lens diameter
Light sources for OCT
-
8/2/2019 Medical Optics
85/254
Light sources for OCT
Continuous sources SLD/LED/superfluorescent fibers,
center wavelength;
800 nm (SLD),
1300 nm (SLD, LED),
1550 nm, (LED, fiber), power: 1 to 10 mW (c.w.) is sufficient,
coherence length;
10 to 15 mm (typically),
Example
25 nm bandwidth @ 800 nm12 mm coherence length (in air).
S l i t di d (SLD )
-
8/2/2019 Medical Optics
86/254
Superluminescent diodes (SLDs)Definition: broadband semiconductor light sources based on
superluminescence(Acronym: SLD)
Superluminescent diodes (also sometimes called superluminescence diodes or
superluminescent LEDs) are optoelectronic semiconductor devices which are
emitting broadband optical radiation based on superluminescence. They are
similar to laser diodes, containing an electrically driven p-n junction and an
optical waveguide, but lack optical feedback, so that no laser action can occur.Optical feedback, which could lead to the formation ofcavity modes and thus
to pronounced structures in the spectrum and/or to spectral narrowing, is
suppressed by means of tilting the output facet relative to the waveguide, and
can be suppressed further with anti-reflection coatings.
Superluminescence: amplified spontaneous emission
http://www.rp-photonics.com/superluminescent_diodes.html
Light sources for OCT
http://www.rp-photonics.com/superluminescence.htmlhttp://www.rp-photonics.com/laser_diodes.htmlhttp://www.rp-photonics.com/waveguides.htmlhttp://www.rp-photonics.com/lasers.htmlhttp://www.rp-photonics.com/cavity_modes.htmlhttp://www.rp-photonics.com/anti_reflection_coatings.htmlhttp://www.rp-photonics.com/anti_reflection_coatings.htmlhttp://www.rp-photonics.com/anti_reflection_coatings.htmlhttp://www.rp-photonics.com/anti_reflection_coatings.htmlhttp://www.rp-photonics.com/cavity_modes.htmlhttp://www.rp-photonics.com/lasers.htmlhttp://www.rp-photonics.com/waveguides.htmlhttp://www.rp-photonics.com/laser_diodes.htmlhttp://www.rp-photonics.com/superluminescence.html -
8/2/2019 Medical Optics
87/254
Light sources for OCT
Pulsed lasers mode-locked Ti:Al2O3 (800 nm),
3 micron axial resolution (or less).
Scanning sources tune narrow-width wavelength over entire spectrum,
resolution similar to other sources,
advantage that reference arm is not scanned,
advantage that fast scanning is feasible.
Construction of image
-
8/2/2019 Medical Optics
88/254
Construction of image
Source of contrast: refractive
index variations
Image reconstructed by
scanning
Applications in ophthalmology
-
8/2/2019 Medical Optics
89/254
Normal patient
Patient with impaired vision (20/80):
The cause is a macular hole
Patients other eye (vision 20/25):
Impending macular hole, which can be
treated
http://rleweb.mit.edu/Publications/currents/cur11-2/11-2oct.htm
Applications in cancer detection
-
8/2/2019 Medical Optics
90/254
Applications in cancer detection
Loss of organization
Columnar epithelium: crypts
Squamous epithelium
http://rleweb.mit.edu/Publications/currents/cur11-2/11-2oct.htm
Applications in developmental biology
-
8/2/2019 Medical Optics
91/254
pp p gy
Ey=eye; ea=ear; m=dedulla; g=gills; h=heart; i=intestine
Ultra high resolution OCT
-
8/2/2019 Medical Optics
92/254
Ultra-high resolution OCT
Image through the skin of a living frog tadpole
Resolution: 3 mm
http://rleweb.mit.edu/Publications/currents/cur11-2/11-2oct.htm
Ultra-high-resolution-OCT
-
8/2/2019 Medical Optics
93/254
mm
mm
versus commercial OCT
W. Drexler et al., Ultrahigh-resolution ophthalmic optical coherence
tomography, Nature Medicine 7, 502-507 (2001)
3-D Reconstruction: In vivo images of human eyeusing spectral-domain OCT
-
8/2/2019 Medical Optics
94/254
using spectral domain OCT
RPE
NFL
I
T
N
S
I
S
TN
N. A. Nassifet al., Opt. Express 12, 367-376 (2004)
-
8/2/2019 Medical Optics
95/254
Applanation tonometry
Theoretical Basis
-
8/2/2019 Medical Optics
96/254
Applanation tonometry is based on the Inbert-Fick
principle.
Which states that for an ideal sphere the pressure (P)
inside the sphere is equal to the force (F) required to
applanate (flatten) its surface, divided by the area (A)
of flattening:P = F/A or F = PA.
The ideal sphere is dry, thin-walled, and readily
flexible.
The cornea, which is not even a true sphere, is none of
these three. Because of this, there are two other
significant forces at work.
-
8/2/2019 Medical Optics
97/254
The force of capillary attraction (T) between the
tonometer head and the tear film is additive tothe external force.
In addition, a force (C), independent of IOP, is
required to flatten the relatively inflexible
cornea. Thus,
F = PA , becomes
F + T = PA + C , orP =( F + T - C) / A
The A is actually on the interior surface of the
-
8/2/2019 Medical Optics
98/254
The A, is actually on the interior surface of the
cornea.
The Goldmann applanator is designed so that Ais equal to 7.35 mm 2.
To achieve this, the diameter of flattening of the
cornea is 3.06 mm.
With this value for A, the opposing forces of
capillary attraction and corneal inflexibility
cancel out.P = F / 7.35 mm 2
-
8/2/2019 Medical Optics
99/254
In addition, with this value for A the IOP in
millimeters of mercury (mmHg) is equal to tentimes the force applied to the cornea in grams,
which is a convenient conversion.
Since only 0.5 m{mu} is displaced from the eyeand the additional increase in pressure induced
in the eye from its steady state by the
tonometer tip is negligible, applanation
tonometery is not significantly affected by
ocular rigidity.
Slit Lamp Imaging
-
8/2/2019 Medical Optics
100/254
Slit Lamp Imaging
-
8/2/2019 Medical Optics
101/254
UC Berkeley Retinal Reading Program
-
8/2/2019 Medical Optics
102/254
Diabetic Retinopathy Screening with EyePACS
Program Manual
10/07
EyePACS
Anatomy of the Retina
-
8/2/2019 Medical Optics
103/254
Optic Disc
Macula
Superior temporal artery and vein
Inferior temporal artery and vein
MaculaSuperior temporal
artery and vein
Inferior temporal artery
and vein
RIGHT EYE
RIGHT EYE
-
8/2/2019 Medical Optics
104/254
GLAUCOMA
Sneak Thief of Sight
GLAUCOMA
-
8/2/2019 Medical Optics
105/254
GLAUCOMA
13% of the blind in India have been
blinded due to Glaucoma
It may be
-
8/2/2019 Medical Optics
106/254
YOU
-
8/2/2019 Medical Optics
107/254
GLAUCOMA
PATIENT EDUCATION
Glaucoma: The Disease
-
8/2/2019 Medical Optics
108/254
Glaucoma: The Disease
Caused by increased pressure of fluid in theeye
The fluid is known as aqueous humor.
Aqueous humor is not same as tears, whichbathe the outside of the eye.
Aqueous humor maintains the normal shapeof the eyeball and nourishes its internalstructure
Glaucoma: The Disease
-
8/2/2019 Medical Optics
109/254
Normal Drainage
Picture
The Trabecular meshwork is
the eyes drain
The Ciliary Bodyis the eyes faucet
or tap where fluid is made
When this drainage of the fluid gets blocked, excess
pressure is formed leading to Glaucoma
Lens
Glaucoma: Symptoms
-
8/2/2019 Medical Optics
110/254
y p
Most patients with Glaucoma, especially Primary Open Angle Glaucoma
are asymptomatic i.e. without any symptoms, until late in the course ofdisease. However, certain patients may have symptoms such as pain,redness, halo vision, blurred vision.
-
8/2/2019 Medical Optics
111/254
Tunnelvision
Red eye, pain in theeye,
Halo around lights
Blurred vision
Visionloss
SYMPTOMS
http://www26.brinkster.com/cowjam/photos/red%20eye.jpg -
8/2/2019 Medical Optics
112/254
Glaucoma: Symptoms
-
8/2/2019 Medical Optics
113/254
Glaucoma: Symptoms
Normal vision
Reduced side vision,
central vision intactTunnel
vision
How common is glaucoma and who gets it?
-
8/2/2019 Medical Optics
114/254
Any one at any age can get glaucoma,
but the older you are the more likely
you are to get it.
People above 45 years are more likely
to get Glaucoma.
Who is most likely to get glaucoma?
-
8/2/2019 Medical Optics
115/254
y g g
You are more likely to get glaucoma if you:
Have family members with glaucoma
Are over 45 years of age
Have poor vision
Have diabetes
Take steroid medication Previous eye injury.
Glaucoma: Types of Glaucoma
-
8/2/2019 Medical Optics
116/254
Chronic or
Primary Open
Angle Glaucoma
Clogged Drainage holes
The angle between the iris and the cornea is normal,
but the drainage holes get clogged from the inside.
Lens
Glaucoma: Types of Glaucoma
-
8/2/2019 Medical Optics
117/254
Acute Angle
Closure Glaucoma
Blocked Drainage holes
The angle is narrower than normal. If fluid cant flow
easily through the opening in the pupil, the iris pushes
forward and blocks the drainage holes.
Lens
How can I find out if I have glaucoma?
-
8/2/2019 Medical Optics
118/254
A series of test performed by your eye doctor
will help to determine whether you have
glaucoma or likely to develop glaucoma.
Glaucoma: Diagnosis
-
8/2/2019 Medical Optics
119/254
1st History and General Examination
Glaucoma: Diagnosis
-
8/2/2019 Medical Optics
120/254
5th PerimetryActual measurement of visual field looking for any
dark areas in field of vision
Treatment Options
-
8/2/2019 Medical Optics
121/254
Eye drops
Pills
Laser surgery Eye operations
Combination
method
Glaucoma: Treatment
-
8/2/2019 Medical Optics
122/254
MEDICAL
Drugs increase conventional outflow.
Lens
Glaucoma: Treatment
-
8/2/2019 Medical Optics
123/254
Drugs reduce production of fluid in the eye.
LensMEDICAL
-
8/2/2019 Medical Optics
124/254
The purpose of treatment is to
prevent further loss of vision.
This is important because loss of
vision due to glaucoma is
irreversible.
Glaucoma: Treatment
-
8/2/2019 Medical Optics
125/254
SURGERY
Trabeculectomy
New Drain
Sclera
Lens
Glaucoma: Treatment
-
8/2/2019 Medical Optics
126/254
SURGERY
Trabeculoplasty
Lens
Open Drainage hole
Laser
Glaucoma: Treatment
-
8/2/2019 Medical Optics
127/254
SURGERY
Iridotomy
Laser
Lens
Making a tiny opening in theiris with a laser allows fluid to
drain freely
Will I go blind because of glaucoma?
-
8/2/2019 Medical Optics
128/254
If glaucoma is left untreated, damage increases,
which may eventually lead to blindness.
Therefore, you should have
-
8/2/2019 Medical Optics
129/254
Regular eye examinations Regular intake of medications as
instructed by the ophthalmologist
Dos and Donts of Glaucoma
-
8/2/2019 Medical Optics
130/254
Stock the medicine before they run out; it isimportant to continue medication on schedule
If more than one drug is used, wait for 10 minutes
between drops
Do not increase the number or amount of
medication taken at one time.
Do not stop taking medication just because you
have no obvious symptoms
Dos and Donts of Glaucoma
-
8/2/2019 Medical Optics
131/254
Take all prescribed doses.
Remember to take medications with you when
you travel
Learn how to take eye drops properly ask yourdoctor for help
Maintain a record with your medication schedule
and lists of treatments and doctors
Remember
-
8/2/2019 Medical Optics
132/254
Treatment for glaucoma requires a team made up of bothyou and your doctor (ophthalmologist)
Your doctor can prescribe treatment for glaucoma, but only
you can make sure to put your eye drops regularly
Do not stop taking or changing your medications withoutfirst consulting your doctor (ophthalmologist)
Frequent eye examinations and tests are critical to monitor
your eyes for any changes
Remember,
It is your vision, and therefore its your responsibility to maintain it
Diagram of the Canon CR6-45NM
-
8/2/2019 Medical Optics
133/254
FocusKnob
Monitor
Setting
Switches
Power
Lamp
Shutter Button
JoystickHeight Adjusting Dial
View SwitchingButtonFixation Target Button
Platform LockingKnob
Diagram of the Canon CR6-45NM
-
8/2/2019 Medical Optics
134/254
Small Pupil KnobInfrared Ray Knob
Lamp Knob
Power SwitchFuse Holders
Power Connector
Forehead Rest
HeightAdjustment Mark
Objective Lens
Chin RestHeight AdjustmentRing
RS422A Connector
Focus Knob
High Correction Sleeve
Fundus Reflex Photographs and 3 Standard FieldsFirst Set: The Right Eye Invert if using Canon DGi
External: Fundus Reflex
-
8/2/2019 Medical Optics
135/254
Field 1: Macula
Field 2: DiscPosition the optic disc at the center. Use the
Fixation Target to position the optic disc (2 stops left
from default).
Position the macula to the far left of the optic
disc.Field 3: Temporal to Macula
Position the optic disc to the far right until it has
disappeared from the screen. Use the Fixation Target
to position the optic disc off to the left (3 stops right
from default).
Position the macula a little lower and a little off to
the right of the center.
High correction sleeve on right side of camera is pulled out;
set S.P. setting to on; F-stop set to F-1; head is positioned 1
inch from head rest bar; focus on iris detail.
High correction sleeve on right side of camera is pushed in;
set S.P. setting to off , F-stop set to ~F-4 F-5
The Optic Disc and Macula should be about equal
distances from the center. This is the default position of
the camera when it is turned on.
-
8/2/2019 Medical Optics
136/254
Using EyeScapeOnce finished taking photos click End Procedure
-
8/2/2019 Medical Optics
137/254
g p
Review your photos. Discard any poor quality photos by highlighting the photo tab and
clicking the Delete button. Once you have the best photos in order, click the Save all button.
Upload Instructions(continued)Step Five: Uploading is not complete until you see the View Case Details page with thumbnail images
of the retinal images you captured
-
8/2/2019 Medical Optics
138/254
of the retinal images you captured.
Depth of focus
-
8/2/2019 Medical Optics
139/254
D Wilson 2002
The effects of a smaller pupil
O'
O''
O'''
I'''
I''
I'
Blur
circles
Screen
p
The pupil & aberrations
-
8/2/2019 Medical Optics
140/254
p p
Spherical aberration and coma are reduced by the
eyes pupil
D Wilson 2002
E
E (entrance pupil)
The pupil & spectacle
magnification
-
8/2/2019 Medical Optics
141/254
magnification
Spectacle magnification will occur at all positions
except at the entrance pupil
D Wilson 2002
E
E (entrance pupil)
The pinhole & myopia
-
8/2/2019 Medical Optics
142/254
Lewis Williams 2002
Myopia
With pinhole
Without pinhole
Pinhole
aperture
Blur circles
The pinhole & hyperopia
-
8/2/2019 Medical Optics
143/254
Lewis Williams 2002
Blur circles
With pinhole
Without pinhole
Hyperopia
Pinhole
aperture
The effects of the pinhole
-
8/2/2019 Medical Optics
144/254
Without pinhole aperture
With pinhole aperture
F or P?
Lewis Williams 2001
The future of vision
-
8/2/2019 Medical Optics
145/254
Super
vision
The future of vision
-
8/2/2019 Medical Optics
146/254
Can we improve on creation?Yes
The future of vision
-
8/2/2019 Medical Optics
147/254
Aberrations and the eye
The eye is subject to wavefront aberrations
These affect all eyes but are more significant incases of:
Keratoconus
Larger pupils
Corneal surgery (eg refractive surgery)
The future of vision
-
8/2/2019 Medical Optics
148/254
It is possible to correct these aberrations in the
laboratory but not yet clinically
Work is currently underway to correct the
aberrations for real subjects using:
The excimer laser to custom ablate (wavefront
guidance)Customised, aberration correcting, contact lenses
but they may change the corneal curvature
The future of vision
-
8/2/2019 Medical Optics
149/254
What are the optical limits?
Diffraction
Changing aberrations with changing accommodation
Changing aberrations with changing direction of gaze
Changing aberrations with the ageing eye
Chromatic aberration
The future of vision
-
8/2/2019 Medical Optics
150/254
Are there other limits?
Image transmission
The cones are 0.5' of arc apart, meaning we have digital
vision!
So, the image may be too detailed for the receptors
creating the familiar TV tweed coat effect
The future of vision
-
8/2/2019 Medical Optics
151/254
Restoration of natural vision
Insert for ametropia
Accommodating
IOL
-
8/2/2019 Medical Optics
152/254
-
8/2/2019 Medical Optics
153/254
-
8/2/2019 Medical Optics
154/254
Radiation WavelengthsSURYA
-
8/2/2019 Medical Optics
155/254
193 nm - Excimer (Cornea)
488 - 514 nm - Argon (Retina)
694.3 nm - Ruby
780 - 840 nm - Diode
1064 nm - Nd Yag (Capsule)
10,600 nm - Carbon dioxide (Skin)
UsesSURYA
-
8/2/2019 Medical Optics
156/254
Diagnostic Therapeutic
Diagnostic UsesSURYA
-
8/2/2019 Medical Optics
157/254
Laser Fluorescence Spectroscopy
Scanning Laser Ophthalmoscopy
Laser Interferometry
Fundus Fluorescein Angiography
Ffa PIC
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
158/254
Widely Used -
Extra-ocular adnexae Anterior Segment
Posterior Segment
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
159/254
LASIK
Suction Ring Microkeratome Flap Removed
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
160/254
LASIK
LASIK Flap Replaced Post - Op.
Therapeutic Uses
B Anterior SegmentSURYA
-
8/2/2019 Medical Optics
161/254
B. Anterior Segment
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
162/254
What is Glaucoma ?
Therapeutic Uses
B Anterior SegmentSURYA
-
8/2/2019 Medical Optics
163/254
iv. Reopen failedfiltering blebs
v. Iridoplasty,
Gonioplastyvi. Iris cyst,Pupilloplasty
B. Anterior Segment
Therapeutic Uses
B. Anterior SegmentSURYA
-
8/2/2019 Medical Optics
164/254
vii. Posterior Capsular Opacification
B. Anterior Segment
Therapeutic UsesC. Posterior Segment
SURYA
-
8/2/2019 Medical Optics
165/254
C. Posterior Segment
What is Diabetic Retinopathy ?
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
166/254
p y
What is Diabetic Retinopathy ?
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
167/254
What is Diabetic Retinopathy ?
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
168/254
What is Diabetic Retinopathy ?
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
169/254
What is Diabetic Retinopathy ?
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
170/254
What is Diabetic Retinopathy ?
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
171/254
i. Diabetic Retinopathy
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
172/254
i. Diabetic Retinopathy
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
173/254
Focal Grid Panretinal
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
174/254
ii. Retinal Haemorrhage
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
175/254
iii. Retinal Breaks or Tears
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
176/254
iv. Subretinal neovascularisation
v. Central serous retinopathy
C. Posterior Segment
Therapeutic Uses
C. Posterior SegmentSURYA
-
8/2/2019 Medical Optics
177/254
vi. Vitreolysis in cystoid macular edema
g
Therapeutic Uses
C. Posterior SegmentSURYA
-
8/2/2019 Medical Optics
178/254
vii. Vitreous traction bands, to freeencapsulated foreign bodies
g
Therapeutic Uses
C Posterior SegmentSURYA
-
8/2/2019 Medical Optics
179/254
viii. Drainage of subretinal fluid / haem.
C. Posterior Segment
Therapeutic UsesC. Posterior Segment SURYA
-
8/2/2019 Medical Optics
180/254
ix. Intraocular tumors (RB)
Therapeutic UsesC. Posterior Segment SURYA
-
8/2/2019 Medical Optics
181/254
ix. Intraocular tumors (Choroidal Melanoma)
Therapeutic UsesSURYA
-
8/2/2019 Medical Optics
182/254
x. Laser scleral buckling
C. Posterior Segment
Therapeutic Uses
D Mi ll U
SURYA
-
8/2/2019 Medical Optics
183/254
i. Neovascular stimulation
ii. Aseptic phototherapy for pre-op
preparationiii. Laser asepsis for diagnosed infectious
corneal ulcers
iv. Endonasal DCR
D. Miscellaneous Uses
What is the Latest ?
PDT (Photo Dynamic Therapy)SURYA
-
8/2/2019 Medical Optics
184/254
PDT (Photo Dynamic Therapy)
TTT (Transpupillary Thermo Therapy)
What is ARMD ?SURYA
-
8/2/2019 Medical Optics
185/254
Two types -
Dry ARMD
Wet ARMD
What is ARMD ?SURYA
W t ARMD
-
8/2/2019 Medical Optics
186/254
Wet ARMD -
Rare
More devastating
Drusen
SRNVM
What is ARMD ?SURYA
Drusen
-
8/2/2019 Medical Optics
187/254
Drusen
What is ARMD ?SURYA
SRNVM
-
8/2/2019 Medical Optics
188/254
SRNVM
What is ARMD ?SURYA
Vision in ARMD
-
8/2/2019 Medical Optics
189/254
Vision in ARMD
What is PDT ?SURYA
-
8/2/2019 Medical Optics
190/254
Visudyne (Verteporfin) Smart Bomb for wet ARMD
Selective Damage
of SRNVM
Costly
Rear Mirror
NEODYMIUM YAG LASER
-
8/2/2019 Medical Optics
191/254
Rear Mirror
Adjustment Knobs
Safety Shutter Polarizer Assembly (optional)
Coolant
BeamTube
AdjustmentKnob
OutputMirror
BeamBeam Tube
Harmonic
Generator (optional)
Laser Cavity
PumpCavity
Flashlamps
Nd:YAGLaser Rod
Q-switch(optional)
Laser-Professionals.com
Light Detection - The Retina
-
8/2/2019 Medical Optics
192/254
Layer of light sensitive cells on inner surface
of the eye
Fovea
Retina
Blind Spot
Optic Nerve
Light Detection - The Retina
There are two types of photoreceptor cells in the retina:
-
8/2/2019 Medical Optics
193/254
There are two types ofphotoreceptor cells in the retina:
cones and rods. The cones are responsible for sharpcolour vision in daylight. The rods provide vision in dim
light.
Near the centre of the retina is a small depressionabout 0.3 mm in diameter which is called thefovea. It
consists entirely ofcones packed closely together. Each
coneis about 2min diameter. Most detailed vision is
obtained on the part of the image that is projected onthefovea. When the eye scans a scene, it projects the
region of greatest interest onto thefovea.
Light Detection - The Retina
The region around thefovea contains both cones and
-
8/2/2019 Medical Optics
194/254
g f
rods. The structure of the retina becomes more coarseaway from thefovea. The proportion ofconesdecreases until, near the edge, the retina is composedentirely ofrods.
In thefovea, each cone has its own path to the opticnerve. This allows the perception of details in theimage projected on thefovea.
Away from thefovea, a number ofreceptors are
attached to the same nerve path. Hence the resolutiondecreases, but the sensitivity to light and movementincreases.
Light Detection - The RetinaWith the structure of the retina in mind, let us examinehow we view a scene from a distance of about 2 m
-
8/2/2019 Medical Optics
195/254
how we view a scene from a distance of about 2 m.
From this distance, at anyone instant, we can see mostdistinctly an object only about 4 cm in diameter. Anobject of this size is projected into an image about thesize of the fovea. Objects about 20 cm in diameter areseen clearly but not with complete sharpness. The
periphery of large objects appears progressively lessdistinct.
Thus, for example, if we focus on a person's face 2 maway, we can see clearly the facial details, but we can
pick out most clearly only a subsection about the sizeof the mouth. At the same time, we are aware of thepersons arms and legs, but we cannot detect, forexample, details about the persons shoes.
Light Detection - The Retina
-
8/2/2019 Medical Optics
196/254
-
8/2/2019 Medical Optics
197/254
Receptors
-
8/2/2019 Medical Optics
198/254
Sensitivity:
-
8/2/2019 Medical Optics
199/254
Summary of Properties of Cones
-
8/2/2019 Medical Optics
200/254
Cones Colour receptors (three types red, blue & green)
Respond in high illumination (daylight)
About 6.5 million per eye, concentrated at thefovea (i.e., high resolution in this region)
In the fovea, each cone connects to one nerve
fibre. Elsewhere, several to one fibre
Overall peak response at ~ 550 nm
Summary of Properties of Rods
-
8/2/2019 Medical Optics
201/254
Rods Respond to intensity only (monochrome)
Respond to low illumination (night vision)
About 120 million per eye, their highest concentration is at about 20 from the fovea
Hundreds of rods connect to each nerve fibre, hence low resolution
Peak response at 510 nm
Resolution of the Eye
So far in our discussion of image formation we
-
8/2/2019 Medical Optics
202/254
have used geometric optics, which neglects thediffraction of light.
Geometric optics assumes that light from a pointsource is focused into a point image. This is not
the case. When light passes through an aperturesuch as the iris, diffraction occurs, and the wavespreads around the edges of the aperture.
As a result, light is not focused into a sharp pointbut into a diffraction pattern consisting of a disksurrounded by rings of diminishing intensity.
Diffraction from a circular aperture (Airy disc)
-
8/2/2019 Medical Optics
203/254
Diffraction Intensity from a square aperture
-
8/2/2019 Medical Optics
204/254
Resolution of the Eye
If light originates from two point sources that are closetogether their image diffraction disks may overlap
-
8/2/2019 Medical Optics
205/254
together, their image diffraction disks may overlap,making it impossible to distinguish the two points.An optical system can resolve two points if theircorresponding diffraction patterns are distinguishable.
This criterion alone predicts that two points are
resolvable if the angular separation between the linesjoining the points to the centre of the lens is equal to orgreater than a critical value given by sin = 1.22/dwhere is the wavelength of light and d the diameter ofthe aperture.
For an iris diameter of 0.5 cm and green light (500nm),
= 1.22x10-4 radians.
Resolution of the Eye
Experiments have shown that the eye does not
-
8/2/2019 Medical Optics
206/254
p y
perform this well.Most people cannot resolve two points with anangular separation of less than 5 x10-4 radians.
Clearly there are other factors that limit theresolution of the eye.
Imperfections in the lens system of the eyecertainly impede the resolution. But perhapseven more important are the limitationsimposed by the structure of the retina.
Resolution of the EyeThe cones in the closely packedfovea are about 2 m indiameter. To resolve two points, the light from each pointmust be focused on a different cone and the excited cones
-
8/2/2019 Medical Optics
207/254
must be focused on a differentcone and the excited cones
must be separated from each other by at least one conethat is not excited.Thus at the retina, the images of two resolved points areseparated by at least 4 m. A single unexcited conebetween points of excitation implies an angular resolution
of about 3 x 10-4
radians (using nodal point 15mm fromretina).Some people with acute vision do resolve points with thisseparation, but most people do not. We can explain thelimits of resolution demonstrated by most normal eyes if
we assume that, to perceive distinct point images, theremust be three unexcited cones between the areas ofexcitation. The angular resolution is then, as observed, 5 x10-4 radians.
Resolution of the Eye
Let us now calculate the size of the smallest detail that
-
8/2/2019 Medical Optics
208/254
the unaided eye can resolve. To observe the smallestdetail, the object must be brought to the closest pointon which the eye can focus. Assuming that thisdistance is 20 cm from the eye, the angle subtendedby two points separated by a distancexis:
tan-1(/2) = (x/2)/20.
If is very small, this becomes = x/20. Because thesmallest resolvable angle is 5 x 10-4 radians the
smallest resolvable detail x is 0.1 mm (5 x 10-4
x 20).Using the same approach facial features such as thewhites of the eye are resolvable from as far as 20m.
Sensitivity of the Eye
-
8/2/2019 Medical Optics
209/254
The sensation of vision occurs when light is absorbedby the photosensitive rods and cones.
At low levels of light, the main photoreceptors are
the rods. Light produces chemical changes in the
photoreceptors which reduce their sensitivity.For maximum sensitivity the eye must be kept in the
dark (dark adapted) for about 30 minutes to restore
the composition of the photoreceptors.
Sensitivity of the Eye
Under optimum conditions, the eye is a very sensitive
-
8/2/2019 Medical Optics
210/254
detector of light.The human eye, for example, responds to light from a
candle as far away as 20 km.
At the threshold of vision, the light intensity is so small
that we must describe it in terms of photons.
Experiments indicate that an individual photoreceptor
(rod) is sensitive to 1 quantum of light. This, however,
does not mean that the eye can see a single photonincident on the cornea. At such low levels of light, the
process of vision is statistical.
Sensitivity of the Eye
In fact, measurements show that about 60quanta must arrive at the cornea for the eye to
-
8/2/2019 Medical Optics
211/254
quanta must arrive at the cornea for the eye toperceive a flash.
Approximately half the light is absorbed orreflected by the ocular medium.
The 30 or so photons reaching the retina arespread over an area containing about 500 rods.
It is estimated that only 5 of these photons areactually absorbed by the rods.
It seems, therefore, that at least 5photoreceptors must be stimulated to perceivelight.
Sensitivity of the Eye
The energy in a single photon is very small.
-
8/2/2019 Medical Optics
212/254
gy g p y
For green light at 500 nm, it is (using E=hc/)
4 x 10-19 Joules
This amount of energy, however, is sufficient to
initiate a chemical change in a single molecule
which then triggers the sequence of events that
leads to the generation of the nerve impulse
Vision
Vision cannot be explained entirely by the physicaloptics of the eye.
-
8/2/2019 Medical Optics
213/254
optics of the eye.
There are many more photoreceptors in the retina thanfibres in the optic nerve. It is, therefore, evident thatthe image projected on the retina is not simplytransmitted point by point to the brain.
A considerable amount of signal processing occurs inthe neural network of the retina before the signals aretransmitted to the brain.
The neural network "decides" which aspects of theimage are most important and stresses thetransmission of those features
-
8/2/2019 Medical Optics
214/254
Limits of Detection
-
8/2/2019 Medical Optics
215/254
Lower limit of illumination about 30 photons spread over about 500 rods
Resolution
Diffraction effects and structure of retina limitresolution to about 8mm under optimal conditions
Blind Spot
Caused by region where nerve fibres enter the
optic nerve - edited out by the brain
-
8/2/2019 Medical Optics
216/254
Understanding VisionDistanceVision
The Human Eye = The Camera
-
8/2/2019 Medical Optics
217/254
The cornea & lens work together tofocus images in the eye
Cornea
Lens
FocalPoint
Understanding VisionYour eye focuses on what you are looking directly at
Central vision is sharp & clear peripheral vision blurred
-
8/2/2019 Medical Optics
218/254
Your eye is continually refocusing as you look from far to near Near vision focusing is called ACCOMMODATION
Near
Normal AccommodationIntermediate Vision
-
8/2/2019 Medical Optics
219/254
Lens shape fordistance vision
When looking at arms length, the lens changesshape & moves forward to focus images
Near imagesfocus behind retina
Intermediate vision is clearDistance & near vision out of focusLens changesshape & position
Intermediate imagesfocus on retina
Normal AccommodationNear Vision
Near vision is clearDistance vision
-
8/2/2019 Medical Optics
220/254
Lens shape fordistance vision
When looking at near objects, the lens continues tochange shape & move forward to focus image
Near imagesfocus behind retina
Lens changesshape & position
Near imagesfocus on retina
out of focus
The Ageing EyeNear Vision
-
8/2/2019 Medical Optics
221/254
The ageing lens loses its ability to change shapeReading glasses or bifocals are required
Loss of Accommodation is called PRESBYOPIA
Lens unableto focus image
FocalPoint
IF YOU HAVE A CATARACT, YOURE NOT
ALONE
-
8/2/2019 Medical Optics
222/254
2.5 million cataract surgeries per year
Number-one therapeutic surgical procedure
for Americans over 65
TODAYS CATARACT SURGERY
-
8/2/2019 Medical Optics
223/254
Greatly improved technology
Usually no hospital stay or long recovery
period
Safer, faster and more comfortable
than ever
WHAT IS A CATARACT?
-
8/2/2019 Medical Optics
224/254
The lens focuses light on the retina
As we age, the lens hardens and cant
focus at close distancesAs we continue to age, the lens may
become cloudy
The cloudiness is the cataract
HOW DOES A CATARACT AFFECTVISION?
-
8/2/2019 Medical Optics
225/254
A cataract scatters light in the eye
instead of focusing it
The cloudier the lens,
the more light isscattered
HOW DOES A CATARACT AFFECTVISION?
Simulated Cataract Vision Simulated Normal
-
8/2/2019 Medical Optics
226/254
Vision
Photos courtesy of the NationalEye Institute
Gradually, visionbecomes dimmerObjects lose their
color
PEOPLE WITH CATARACTS HAVE DIFFICULTY:
-
8/2/2019 Medical Optics
227/254
Seeing in the distance or reading
Distinguishing road signs at dusk
Recognizing colors
Recognizing friends and family at
a distance
Driving at night
WHO GETS CATARACTS?
-
8/2/2019 Medical Optics
228/254
Almost everyone sooner or later
Half of all people between the ages
of 52 and 64*
Younger people, due to injury, excessive
sunlight, metabolic changes, or drugs
*American Academy of Ophthalmology
DETECTING A CATARACT
-
8/2/2019 Medical Optics
229/254
Visual acuity testSlitlamp examination
Glare test
TREATING A CATARACT
-
8/2/2019 Medical Optics
230/254
Is vision impaired?
Is quality of life affected?
The eye care practitioner andthe patient decide:
TODAYS CATARACT SURGERY
-
8/2/2019 Medical Optics
231/254
A marvel of medical technology
Outpatient procedure
Local anesthesia
Tiny incision heals rapidly
Little of no discomfort
THE CATARACT PROCEDURE
-
8/2/2019 Medical Optics
232/254
The clouded natural lens is removed
A man-made lens is inserted
The new lens is an intraocular lens
(IOL)
CHOICES FOR RESTORING VISION
-
8/2/2019 Medical Optics
233/254
Todays technology offers two
different types of intraocular
lenses (IOLs)
Monofocal ReZoom
MONOFOCAL
-
8/2/2019 Medical Optics
234/254
MONOFOCAL
-
8/2/2019 Medical Optics
235/254
Good vision at one distance
usually far
Most people need glasses for close-upactivities like reading or crafts
Good vision when you go to aballgame or read road signs
The ReZoom Multifocal Lens
-
8/2/2019 Medical Optics
236/254
The ReZoom and Crystalens IOL
-
8/2/2019 Medical Optics
237/254
Designed for gooddistance vision
and near vision
Can reduce the need forglasses in activities like
reading, watching
television, or watching a
movie
ReZoomRANGE OF VISION EXAMPLES
-
8/2/2019 Medical Optics
238/254
If youre golfing, you may be able tosee where your drive lands, sink your
putt, and write down the score,
without glasses
When shopping, you may be able to
read the aisle signs and the package
labels, and count your change,without glasses
EQUAL SAFETY FOR BOTH IOL TYPES
-
8/2/2019 Medical Optics
239/254
Identical surgical procedures
The real difference is the type of vision
ReZoomBalanced View Optics
Technology
-
8/2/2019 Medical Optics
240/254
Pictorial representation
ReZoomTM IOL Spectacle Independence
%
-
8/2/2019 Medical Optics
241/254
Source: Product labeling.
93% 93%
81%
0%
20%
40%
60%
80%
100%
Distance Intermediate Near
ReZoom IOL
ReZoom IOL Visual Outcomes
-
8/2/2019 Medical Optics
242/254
PatientBrochure
-
8/2/2019 Medical Optics
243/254
Which Lens Is Best For You?We will help you decide which lens is the bestalternative for your specific refractive need
-
8/2/2019 Medical Optics
244/254
Not everyone is a candidate for CrystaLens VisionEnhancement Surgery
Lifestyle
Expectations
A thorough examination will be performedMultiple diagnostic tests will be performed
Expectations
All surgery involves risk
-
8/2/2019 Medical Optics
245/254
Not everyone responds to the surgery in the sameway
Other medical & eye diseases may influence yourability to seeclearly &/or accommodate
Vision After Accommodative Surgery
1 to 10 days after surgery
-
8/2/2019 Medical Optics
246/254
Distance vision is typically excellent in the
majority of patients
Near & intermediate vision may be excellentafter surgery, however varies from patient to
patient
Typically continues to improve over time
Vision After Accommodative Surgery
One year after surgery*
-
8/2/2019 Medical Optics
247/254
98.4% were able to drive, watch TV, participate in sports &
perform normal activites
98.0% were able to work on their computer, read product
labels & read the speedometer
98.4% were able read newspapers, magazines, recipes; sew
& dial their cell phone
*FDA Clinical Study
Without Glasses73.5% do not depend on glasses atall or wear them only occasionally
Are You A Candidate Forcrystalens Vision Enhancement?
Schedule a eye examination
-
8/2/2019 Medical Optics
248/254
Talk with people who have had cataract surgery
Research lens replacement after lens removal surgery
Find a qualified & certified crystalens
VisionEnhancement Surgeon
Non- Contact Tonometers
-
8/2/2019 Medical Optics
249/254
- Invented by Dr. Bernie Grolman in the 1960s (American Optical)
- To enable ODs in the USA to perform tonometry
- Introduced in 1971
- Uses rapid air pulse technology
- Easy to use
- Strong Goldmann correlation
- Objective: no operator bias
- No anesthetic required
- No risk of cross-contaminationModern NCT - AT555
NCT Traditional Method of Operation
-
8/2/2019 Medical Optics
250/254
Method of Operation
-
8/2/2019 Medical Optics
251/254
Applanation Signal Plot
-
8/2/2019 Medical Optics
252/254
Definitions
H t i
-
8/2/2019 Medical Optics
253/254
HysteresisThe phenomenon was identified, and the term coined, by
Sir James Alfred Ewing in 1890.
Hysteresis is a property of physical systems that do not
instantly follow the forces applied to them, but reactslowly, or do not return completely to their original state.
Corneal Hysteresis
The difference in the inward and outward pressure
values obtained during the dynamic bi-directionalapplanation process employed in the Ocular Response
Analyzer, as a result of viscous damping in the cornea.
-
8/2/2019 Medical Optics
254/254