Medical Device Regulation Conformity Routes

Page 2: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

2www.be-on-quality.com

The goal is to ensure a simple and easy to understand

regulatory environment for medical devices, which ensures the

efficient functioning of the internal market.

Page 3: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

3www.be-on-quality.com

follow-upconformity assessment procedure

risk management

usability

Classification

of the medical

product

Clinical

Evaluation

General safety

and performance

requirements

Technical

DocumentationCE marking

Post market

Surveillance

Quality

management

system

idea market

Page 4: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

4www.be-on-quality.com

Annex III Documentation on Post Market

Surveillance

Annex IV Declaration ofconformity

Annex II Technical

Documentation

Annex VIUDI - Unique

Device Identification

Annex IGeneral Safety

and Performance Requirements

Annex VIII Classification

Rules

Certification?

Page 5: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

5www.be-on-quality.com

Intended

purpose

Classification

Clinical data

Risk

management

General safety

and

performance

requirements

UDI

PMCF

PMS Common specifications

CER

Page 6: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

6www.be-on-quality.com

INTENDED PATIENT

POPULATIONThe DEMOGRAPHICS of the patient

INTENDED USERWHO should use the device

INTENDED USEHOW to use the device

PRINCIPLES OF OPERATIONHOW the device achieves its Intended Purpose

INDICATIONS FOR USEThe CLINICAL PICTURE of the patient

Contraindications

CLAIMSWHAT the device does• Prevention• Diagnosis/Prognosis (Support)• Management/Monitoring• Treatment/Alleviation• Predicate Device

Page 7: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

7www.be-on-quality.com

class I class I s / m / r class IIa class IIb class III

Annex I - General Safety and Performance RequirementsAnnex II - Technical DocumentationAnnex III - Documentation on Post Market SurveillanceAnnex VI - UDI - Unique Device Identification

Annex VIII - Classification Rules

Page 8: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

8www.be-on-quality.com

Risk

class

low medium high

Class I Class IIa Class IIb Class III

According to risk-based rules (e.g. EU) or lists (e.g. USA)

Page 9: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

9www.be-on-quality.com

Intended use of the device

Non-active medical device Active medical device Special rules

Invasive medical devices

Non-Invasive medical devices

Type and duration of invasiveness

III

II a/ b

I s m r

Page 10: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

10www.be-on-quality.com

non-invasive devices

Rule 3modifying

biological or chemical

composition

Rule 2channelling or storing blood

Rule 1

Rule 4contact with injured

skin or mucous membrane

invasive devices

Rule 8implantable

devices

Rule 7short-

term use

Rule 6transient

use

Rule 5not intended for connection to an active device or intended for

connection to class I

active devices

Rule 12intended to

administer or remove medicinal

products

Rule 11software

Rule 10devices intended for diagnosis and

monitoring

Rule 9therapeutic devices

intended to exchange energy

Rule 13

special rulesRule 18

tissues or cells of human or animal origin

Rule 16devices intended to disinfect contact

lenses

Rule 15contraception

Rule 14devices incorporating a medicinal product

Rule 17devices for

recording x-ray diagnostic images

Rules 19nanomaterial

Rule 20inhalation

Regel 21combination of substances

Rule 22closed loop systems

Page 11: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

11www.be-on-quality.com

Device + software

Not invasive

Short term

Not invasive

Radiation

Transient

Long time

Invasive

Not active

Page 12: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

12www.be-on-quality.com

QMS referred to in Article 10(9)

+

Preparation of the technical documentation according to Annex II + III

+

EU

Declaration

of

Conformity

Annex IV

Manufacturer assigns

I

Page 13: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

13www.be-on-quality.com

product conformity testing

Annex XI

Manufacturer assigns

IIa

xxxx

QMS referred to in Article 10(9)

+

Preparation of the technical documentation according to Annex II + III

+

IIa

Costs: 20k€

Page 14: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

14www.be-on-quality.com

product conformity testing

Annex XI

Manufacturer assigns

IIb + III

xxxx

type examination

Annex X

IIb + III

+

QMS referred to in Article 10(9)

+

Preparation of the technical documentation according to Annex II + III

+

Costs: 20k€

Costs: 10k€

per Batch

Page 15: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

15www.be-on-quality.com

QMS referred to in Article 10(9)

+

Preparation of the technical documentation according to Annex II + III

+

CompleteQMS

Annex IX

Manufacturer assigns

I + IIa + IIb

+ III

xxxx

Costs: 20k€

Each Scope:

15k€

Page 16: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

16www.be-on-quality.com

product conformity testing

Annex XI

Manufacturer assigns

IIa + IIb + III

xxxx

type examination

Annex X

IIb + III

+

QMS referred to in Article 10(9)

+

Preparation of the technical documentation according to Annex II + III

+ ++

+

IIa

EU

Declaration

of

Conformity

Annex IV

Manufacturer assigns

I

CompleteQMS

Annex IX

Manufacturer assigns

I + IIa + IIb

+ III

xxxx

Costs: 10k€

per Batch

Costs: 20k€ Costs: 20k€

Each Scope:

15k€

Page 17: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

17www.be-on-quality.com

product conformity testing

Annex XI

Manufacturer assigns

IIa + IIb + III

xxxx

type examination

Annex X

IIb + III

+

QMS referred to in Article 10(9)

+

Preparation of the technical documentation according to Annex II + III

+ ++

+

IIa

EU

Declaration

of

Conformity

Annex IV

Manufacturer assigns

I

CompleteQMS

Annex IX

Manufacturer assigns

I + IIa + IIb

+ III

xxxx

Costs: 10k€

per Batch

Costs: 30k€Costs: 20k€

Page 18: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

Medical Device Regulation Performing the Conformity Assesmsent

Page 19: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

19www.be-on-quality.com

Certificate

Preparation

Contact

approvals

Technical File

Assessment

Stage 1 QMS

Stage 2 Audit

Page 20: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

20www.be-on-quality.com

Roles

NotifiedBody

Authorities

EU-Commission

Manufacturer

SupplierImporter

Distributor

Marketplace

Represent-ative

Critical Suppliers

and outsourced

services only

Page 21: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

21www.be-on-quality.com

Certificate

Preparation

Contact

approvals

Technical File

Assessment

Stage 1 QMS

Stage 2 Audit

Internal Step at the

notified body

Page 22: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

22www.be-on-quality.com

Certificate

Preparation

Contact

approvals

Technical File

Assessment

Stage 1 QMS

Stage 2 Audit

Page 23: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

23www.be-on-quality.com

Section 1:

Names,

Addresses,

Identification,

Conformity

Economic

Operators

Section 2 –

Relevant

Regulations

and Technical

Standards

Standards

Common

Specifications

Section 3 –

Description of

Device

Intended

purpose

Claims

Section 4 –

Labels and

Labeling

UDI

MD-Symbol

IFU

Section 5 –

Bench Testing

and Clinical

Data

Summary

CEP, CER,

PMCF

PMS & PSUR

Section 6 –

Risk Analysis

Summary

Risk Benefit

discussion

Section 7 –

Technical

infomation

Specification

Architecture

Verification

and Validation

Page 24: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

24www.be-on-quality.com

Declaration of

conformity

ISO 13485MDR Article 10.9

Risk Management Plan

Risk analysis

Use specification

Summative Evaluation

Formative Evaluation

Development –Test

PlanSW and EMC

User Requirements

Product Requirements

SW

design

Design Release

Validation

Literature Search

Clinical Evaluation Plan

Label

Electronic and

mechanical

design

Intended PurposeGSPR

Page 25: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

25www.be-on-quality.com

Certificate

Preparation

Contact

approvals

Technical File

Assessment

Stage 1 QMS

Stage 2 Audit

Page 26: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

26www.be-on-quality.com

Step by step to QMS certification

1 • Building awareness for quality management

2 • Inventory and formulation of objectives

3 • QM documentation

4 • staff training

5 • Internal Audit

6 • management review

7 • certification

Page 27: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

27www.be-on-quality.com

Production

Thu

checkact

scheme

MDR-QMS

quality management manual

4. quality management system

5. management responsibility

6. management of resources

documentaryrequirements

Technicaldocumentation

Human Resources

7. product realization

Supplier management

External communication

Development

Clinical life cycle Risk management

software development

Identification and traceability

analysis and management evaluation

Post-market monitoring

Improvement

Complaints and feedback

Vigilance

Quality audit

Corrective and preventive measures

8. measurement, analysis and

improvement

Development changes

Page 28: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

28www.be-on-quality.com

Timeline for the creation of a QMS

Directives, laws and standards

EN ISO 13485

MDR 2017/745

QM specification documentation

lawsuits

templates

quality records

Documents (changeable)

records(not changeable)

e.g. completed protocols, reports, plans

Page 29: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

29www.be-on-quality.com

QMS setup

quality policy

process landscape

(QMH and Processes)

templates

Quality records and

documents

• "soul" of the QMS

• Q policy usually public ("corporate values")

• New employees find their way around quickly

• Consistent workflows

• "Small consultant" for document creation

• You don't always have to think of everything

yourself

• Actual documentation

• Required by the quality management system

Page 30: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

30www.be-on-quality.com

Certificate

Preparation

Contact

approvals

Technical File

Assessment

Stage 1 QMS

Stage 2 Audit

Page 31: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

31www.be-on-quality.com

Certificate

Preparation

Contact

approvals

Technical File

Assessment

Stage 1 QMS

Stage 2 Audit

Certification

check and

notified

body

internal

assessment

Page 32: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

+49 9151 865 76 [email protected]

be [email protected]+46 7 6606 1558+31 6 2398 6834

Page 33: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

MDR TD compilation

Page 34: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

34www.be-on-quality.com

Compile the perfect file

Annex II

Annex III

GSPR

Page 35: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

35www.be-on-quality.com

Remove formatting

issues

Ensure all Documents

to besigned

CompileGSPR First

Ensure all standardsused to be

valid

EnableOCR and make is

searchable

Create Annex II

and Annex III

Compilethem all intoONE BIG,

BIG document

Page 36: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

MDR Common mistakes

Page 37: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

37www.be-on-quality.com

DECLARATION OF INTEREST

(CER)

VIGILANCE PROCEDURE

PMCF NOT DETAILED

RISK BENEFIT WITHOUT CER

LABELLING

TYPE TEST

BIOLOGICAL ASSESSMENTS

STATE OF THE ART

Page 38: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

38www.be-on-quality.com

DEVICE DESCRIPTION AND SPECIFICATION,

INCLUDING VARIANTS AND ACCESSORIES

• Basic UDI-DI or other unambiguous reference allowing traceability

• Intended population, indications, contraindications, warnings

• Principle of operation of the device and mode of action

• Rationale for considering the product a medical device

• Device classification, applicable rule and rationale

• Explanation of any novel features

Page 39: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

39www.be-on-quality.com

• Description of accessories

• Description of all variants of the device

• Description of the key functional elements

• Description of all the relevant raw materials

• Technical specifications, dimensions and performance attributes

• Reference and overview of previous and similargenerations of the subject device and device market history

DEVICE DESCRIPTION AND SPECIFICATION,

INCLUDING VARIANTS AND ACCESSORIES

Page 40: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

40www.be-on-quality.com

DESIGN AND MANUFACTURING INFORMATION

• Information about the design stages

• Complete information and specifications, including the manufacturing processes and their validation, their adjuvants, thecontinuous monitoring and the final product testing

• Identification of all sites including suppliers and subcontractors; where design and manufacturing activities are performed

Page 41: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

41www.be-on-quality.com

Polymerization

Injection molding…

Material processing

like cutting, cleaning

polishing gluing

Lubricants, coating,

glue, cleaning agents

labels (glue, dye,

solvents)

Sterilization, process

condition

Temperature,

humidity, UV light,

aging

Additives like fillers,

reinforcement

composites, UV

Blocker, flame

retardents,

pigments and dyes,

stabilizers

Configuration

Raw materials

Historical data

Manufacturing

Packaging

Literature

Page 42: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

42www.be-on-quality.com

GENERAL SAFETY AND PERFORMANCE REQUIREMENTS

The documentation shall contain information forthe demonstration of conformity with thegeneral safety and performance requirementsset out in Annex I that are applicable to the devicetaking into account its intended purpose, and shallinclude a justification, validation and verification ofthe solutions adopted to meet those requirements.

→Checklist in tabular form!

Page 43: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

43www.be-on-quality.com

Risk

• Lifecycle risk

management

Benefit

• Challenge clinical

data

• User acceptance

AUDIT PITFALL:

RISK BENEFIT WITHOUT CER

Page 44: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

44www.be-on-quality.com

BENEFIT-RISK ANALYSIS AND RISK MANAGEMENT

The documentation shall contain information on:

• The benefit-risk analysis

• The solutions adopted and the results of the risk management

Page 45: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

45www.be-on-quality.com

PRODUCT VERIFICATION AND VALIDATION

• The documentation shall contain the results of all verifications and validation tests and/or studies undertaken to demonstrate conformity with the requirements of the MDR.

6.1 Pre-clinical and clinical data

• results of tests, such as engineering, laboratory, simulated use and animal tests, and evaluation of published literature applicable to the device, taking into account its intended purpose, or to similar devices, regarding the pre-clinical safety of the device and its conformity with the specifications

Page 46: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

46www.be-on-quality.com

MDR ANNEX I GSPR

Risk Management

Clinical Evaluation Biological Evaluation

Risk Analysis

Risk Management Report

Other Processes

INPUT

OUTPUT

Page 47: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

47www.be-on-quality.com

CURRENT

• REAL STANDARD

HARMONIZATION

• EXISTING

DOCUMENTS

AUDIT PITFALL:

GAP TO CURRENT STANDARD NOT EVALUATED

Page 48: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

48www.be-on-quality.com

PRODUCT VERIFICATION AND VALIDATION

Detailed discussion of test design, test protocols and reports with data analysis and conclusions in particular for the following:

• biocompatibility

• physical, chemical and microbiological characterization

• electrical safety and electromagnetic compatibility

• software verification and validation

• stability, including shelf life

• performance and safety

Page 49: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

49www.be-on-quality.com

PRODUCT VERIFICATION AND VALIDATION

• the clinical evaluation report

• the clinical evaluation plan

• the PMCF plan

• PMCF evaluation report

Page 50: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

50www.be-on-quality.com

clinical evaluation plan

clinical evaluation

report

PMCF plan

PMCF evaluation

report

FINDING PITFALL:

PMCF NOT DETAILED

FINDING PITFALL:

DECLARATION OF INTEREST (CER)

FINDING PITFALL:

MISSING INDICATIONS

Page 51: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

51www.be-on-quality.com

PRODUCT VERIFICATION AND VALIDATION

Additional information for specific cases:

• Medicinal substances

• Requirements for devices utilizing tissues or cells of human or animal origin or their derivatives

• Devices composed of substances or combinations thereof intended to be introduced into the human body that are absorbed by or locally dispersed

• ‘CMR’ and endocrine-disrupting substances

• Sterility and microbiological condition

• Measuring Function

• Devices connected to other devices

Page 52: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

52www.be-on-quality.com

Type Testing

Type Tests

Collaterals

Homecare

EmergencyAlarm

ElectricalSafety

Biological Safety

Cytotox

Chemical Character

SensitationIrritation

Sterility

Usability

Productrelevant

standards

80601-2-x

60601-2-xISO14xxx

Costs: 20k€

Costs: each

8k€

Costs: each

5k€

Costs: 5k€Costs: 20k€ - 100k€

Page 53: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal

53www.be-on-quality.com

New Approach

Collecting

data sheets is

not sufficient

Page 54: Medical Device Regulation Conformity Routes...requirements of the MDR. 6.1 Pre-clinical and clinical data • results of tests, such as engineering, laboratory, simulated use and animal