Mechanical Design, Instrumentation and Measurements from a Hemoacoustic Cardiac Phantom Hani Bakhshaee+, Guillaume Garreau*, Gaspar Tognetti*, Kourosh Shoele+, Ronann Carrero+, Thomas Kilmar+, Chi Zhu+ William R. Thompson++, Jung Hee Seo+, Rajat Mittal+, Andreas G. Andreou* *Department of Electrical and Computer Engineering, Johns Hopkins University, ++Department of Pediatrics, Johns Hopkins University School of Medicine +Department of Mechanical Engineering, Johns Hopkins University,

Upload: others

Post on 08-Nov-2021

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Mechanical Design, Instrumentation and Measurements from …

Mechanical Design, Instrumentation andMeasurements from a Hemoacoustic

Cardiac PhantomHani Bakhshaee+, Guillaume Garreau*, Gaspar Tognetti*, Kourosh Shoele+, Ronann

Carrero+, Thomas Kilmar+, Chi Zhu+William R. Thompson++, Jung Hee Seo+, Rajat Mittal+, Andreas G. Andreou*

*Department of Electrical and Computer Engineering, Johns Hopkins University, ++Department of Pediatrics, Johns Hopkins University School of Medicine

+Department of Mechanical Engineering, Johns Hopkins University,

Page 2: Mechanical Design, Instrumentation and Measurements from …

Outline

• Brief overview

• Phantom design

• Hemoacoustic computational simulations

• Comparison between experimental and computational results

• Future versions of the phantom

2

Page 3: Mechanical Design, Instrumentation and Measurements from …

Overview

• Heart disease:

– Most consequential disease in the industrialized world

– Annual US expenditure on heart disease exceeds half a trillion dollars

• Cardiac auscultation:

– Been around for 200 years

– Limitations: subjective, inaccurate

• Automated cardiac auscultation via a wearable acoustic array (the"StethoVest"):

– Expensive Cost-effective

– Reactive-Proactive,

– Hospital centric Patient centric

3

Page 4: Mechanical Design, Instrumentation and Measurements from …

Developing the thoracic phantom

• The phantom will be used to validate the codes and to examine the sensors

• To design the phantom the following items should be considered :

– Tissue mimicking homogeneous material and characterization

– Murmur generating embedded fluid-circuit

– Measurements : Variety of acoustic sensors

4

Page 5: Mechanical Design, Instrumentation and Measurements from …

Material selection

• Acoustic and mechanical properties should be similar

• Examples of previous tissue-mimicking materials in the literature

– Agar

– Silicone

– Polyvinyl alcohol gel (PVA) and

– Polyacrylamide gel (PAA)

5

Page 6: Mechanical Design, Instrumentation and Measurements from …

Acoustic property of the material

6 Acoustical properties of selected tissue phantom materials for ultrasound imaging, K Zell et al , 2007

Page 7: Mechanical Design, Instrumentation and Measurements from …

Mechanical Property

7 Elastic properties of synthetic materials for soft tissue modeling. Mansy Et al, 2007

Page 8: Mechanical Design, Instrumentation and Measurements from …

Silicone rubber

• Silicone rubber, Ecoflex 010 (Smooth-on)

– Easy to produce ,

– extremely stable

– non-toxic and

– negligible shrinkage

• Procedure to make :

– Mixing Part A part B,

– Adding Silicon thinner,

– Degassing for 3-4 min in (-29 in Hg) to remove air bubbles

8

Page 9: Mechanical Design, Instrumentation and Measurements from …

Material characterization

9

Page 10: Mechanical Design, Instrumentation and Measurements from …

Speed of sound

Speed of sound : 993-1043m/s

10

Ultrasound Probe

Gel sample

Thanks to Dr. Emad Boctor and Fereshte Alamifar

The ratio of sound speeds are equal to the inverse ratio of the depth seen in the ultrasound image

Page 11: Mechanical Design, Instrumentation and Measurements from …

Murmur generating

11

3D printed Cast

Page 12: Mechanical Design, Instrumentation and Measurements from …

Fluid Flow Circuit

12

Page 13: Mechanical Design, Instrumentation and Measurements from …

13

Bipac sensor attached to the Micromanipulaor

HP sensor attached to the Micromanipulaor

Page 14: Mechanical Design, Instrumentation and Measurements from …

Micromanipulators

14

Page 15: Mechanical Design, Instrumentation and Measurements from …

Measurements

• Different acoustic sensors used in the phantom tests.

– A: Commercially available electronic stethoscope.

– B: Accelerometers

– C: HP 21050A sensor mounted on a micromanipulator.

– D: Biopac sensors

15

Page 16: Mechanical Design, Instrumentation and Measurements from …

Sensor selection …

16

• Pump was turned on and off • Clear difference between two diagrams

for HP and Biopac• Poor SNR for stethoscope and the

accelerometer

Page 17: Mechanical Design, Instrumentation and Measurements from …

Effect of Indentation

17

To compare the effect of indentation :• Reference position :Sensors touching

the sample • Gradual increase in the indentation • Indentation = 0, 0.03, 0.06, 0.09, 0.12

and 0.15 in ~ 0 : 0.76 :3.81 mm• After 0.12, no differences were

observed

Page 18: Mechanical Design, Instrumentation and Measurements from …

Distance after the constriction

18

Page 19: Mechanical Design, Instrumentation and Measurements from …

Computational model

By Dr. Jung-Hee Seo

19

Page 20: Mechanical Design, Instrumentation and Measurements from …

Hemoacoustic Simulation

20

HemodynamicsIBM, Incompressible N-S

U=0.25 m/s Structural wave eq.For viscoelastic material

Generalized Hooke’s lawKelvin-Voigt model

i ij jPn p n

4~ (10 )iu O

0

1

ij jk iij

k j i

ij ji i

j j j i

p uu u

t x x x

p uu u

t x x x x

0iu

0iu

High-order IBM,6th order Compact Finite Difference Scheme,4 stage Runge-Kutta method

Page 21: Mechanical Design, Instrumentation and Measurements from …

Freq [Hz]

PS

D

100

101

102

103

10-7

10-6

10-5

10-4

10-3

10-2

10-1

100

0.5D

1D

1.5D

2D

2.5D

3D

3.5D

4D

4.5D

Hemodynamic Simulation Results

Pressure

Vorticity

3D vortical structure

D

ReD=4000

Pressure fluctuation is responsible for the murmur generation

Wall pressure spectrum

2D0

Strong pressure fluctuations are observed beyond 2D downstream of the stenosis

Page 22: Mechanical Design, Instrumentation and Measurements from …

3D Elastic Wave Simulation

gelB

L=7D

DT=6.22D

Radial velocity fluctuation

•200x200x320 (12.8 M)•Compression wave speed is reduced by an order (100 m/s)•Shear wave speed remains the same (4.2 m/s)•200 hrs with 256 cores for real time 0.8 sec

Page 23: Mechanical Design, Instrumentation and Measurements from …

Surface Velocity Fluctuations

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7-5

0

5x 10

-7

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7-1

0

1x 10

-6

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7-1

0

1x 10

-6

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7-1

0

1x 10

-6

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7-1

0

1x 10

-6

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7-1

0

1x 10

-6

ur/

c p

t [sec]

-1D

1D

2D

3D

4D

0

0

2D

Page 24: Mechanical Design, Instrumentation and Measurements from …

Surface Acceleration Spectrum

0

2D

Frequency [Hz]R

ad

iala

cce

lera

tio

n[m

/se

c2]

100

101

102

103

10-9

10-8

10-7

10-6

10-5

10-4

10-3

10-2

-1D

0

1D

2D

3D

4D

Page 25: Mechanical Design, Instrumentation and Measurements from …

Comparison

25

• The frequency spectrum of themeasured acceleration at the downstream location is plotted along with the computational ones

Page 26: Mechanical Design, Instrumentation and Measurements from …

Future versions

26

• Adding lung to the phantom • Foam is used to model the lung• Non-axisymmetric model

Page 27: Mechanical Design, Instrumentation and Measurements from …

Summary

• Different steps to make the Cardiothoracic phantom were explained

– Material selection and characterization

– Murmur generating embedded fluid-circuit

– Measurements options

• Hemoacoustic simulation results were presented and compared with those from experiment

– Good agreement was seen based on the preliminary results

27

Page 28: Mechanical Design, Instrumentation and Measurements from …

Acknowledgment

• Dr. Emad Boctor and Fereshteh Alamifar

• NSF for funding

28

Page 29: Mechanical Design, Instrumentation and Measurements from …

Thank you

29

Page 30: Mechanical Design, Instrumentation and Measurements from …

Sensor selection

Frequency [Hz]

PS

D

101

102

103

10-5

10-4

10-3

10-2

w/ Flow

30

Frequency [Hz]

PS

D

101

102

103

10-5

10-4

10-3

10-2

w/ Flow

no Flow

A: Electronic stethoscope B: Accelerometers

Page 31: Mechanical Design, Instrumentation and Measurements from …

Model for the Aortic Stenosis Murmur

U

Stenosis

Flow fluctuation

Thoracic phantom (silicone gel)

D

Wave propagation

Re=UD/=4000St=fD/U

For the joint computational/experimental studyEcoFlex-10

=1040 kg/m3

K=1.04 GPa (cb=1000.0 m/s)G=18.39 kPa (cs=4.2 m/s)=14 Pa s

U=0.25 m/sD=1.5875 cmDT=9.84 cm (gelA), 16.51 cm (gelB)

c.f.Biological soft tissue:K=2.25 GPa (cb=1500 m/s)G=0.1 MPa (cs=10 m/s)=0.5 Pa s

DT

L=7D