measuring principle

Upload: kamalnyn

Post on 29-May-2018

219 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/8/2019 Measuring PriNciple

    1/36

    Electrolyte Analyzer

    MEASURING PRINCIPLES

    1. ISE THEORY

    The analyzer utilizes Ion Selective Electrode (ISE) technology. Ion SelectiveElectrode is a type of electrochemical sensor. It converts the ion activity to theelectric potential of the electrode. The relation conforms to the NERNST equation,that the Logarithm of the ion activity has a linear relation with the electrode potential.In addition, different electrode is sensitive to different ions, for example, sodium

    electrode is only sensitive to Na ions, and potassium electrode is only sensitive to Kions. If potassium electrode, sodium electrode, and chloride electrode are beingcombined together, then K ions, Na ions, and Chloride ions in the sample can bemeasured at the same time.

    The key part of the electrode is the sensitive membrane. ON one side, it is incontact with the sample, responds to the change of the concentration of certain ionsin the sample. On the other side, it is in contact with the internal filling solution, andconverts the ionic conduction to the electronic conduction through a silver thread i.e.internal electrode. In addition, there is a reference electrode providing thread i.einternal electrode. In addition there ia reference electrode providing referencepotential and forming a complete measuring circuit. Inside the reference electrodethere is also an internal electrode. Its potential remains constant when theconcentration of the solution changes, so it provides a reference point to measurethe potential differences.

    2. MEASURING PRINCIPLE

    The Instrument measures the electrode potentials, and the data is processed by themicroprocessor to obtain the concentration of a given ion. The measure method iscalled standard comparison It uses two kind of standard solutions, one for the

    calibration of the base point, and the other for the calibration of the slope. The resultis obtained from the potentials of the sample and two standard solutions.Following are the equations:

    C x =C A * EXP [(E X-E A) / S] (1)

    S = EB E ALog ( CB / Ca ) (2)

    Note :

    CX, EX: the concentration and potential of the sampleCA EA: the concentration and potential standard A

  • 8/8/2019 Measuring PriNciple

    2/36

    S: the slope of electrodeIn order to improve the precision, the contents of the standard solution should be

    similar with the blood samples as much as possible..

    _

    3. MANOMETRIC ME;THOD (FOR HCO3 )

    Add certain quantity of blood serum and regent (Lactic acid) into the sealed reactionchamber, the HCO3 ions in the serum will participate into the reaction and releaseCO2 as a result; the gas pressure inside the reaction chamber will be increasedaccordingly. The pressure sensor detects the changes and sends the signals to themicroprocessor to determine the amount of HCO3 ion of serum and then the amountcould be displayed and printed. The instrument uses AB (Actual bicarbonate) standfor HCO3 ion.

  • 8/8/2019 Measuring PriNciple

    3/36

    1.1 System introduction

    1.1.1 Front panel

    1 Reagent Std A Std B and waste container and Std R

  • 8/8/2019 Measuring PriNciple

    4/36

    3. Pressure sensor and tube

    4. Mixing chamber assembly

    Aspiration pump assembly (Pump head, Pump Tube, Pump Tube

    Bracket, Pump Tube Connector

    5. Lactic acid Pump assembly ((Pump head, Pump Tube,

    Pump Tube Bracket, Pump Tube Connector)

    6. liquid outlet valve and waste tube

    7. Air outlet valve and tubing

    8. Sample probe assembly (Including Auto lifting

    mechanism)

    9. wash block

    10. Localizer

    11. Amplifier board

    12. Electrode assembly

    13. Touch screen

    14. Printer

  • 8/8/2019 Measuring PriNciple

    5/36

    1.1.2 Rear panel

    1. Power switch

    2. Power socket (including fuse holder)

    3. RS-232 serial port

    4. Grounding pole

    5. Machine label

    - 2 -

  • 8/8/2019 Measuring PriNciple

    6/36

    1.2 Environment requirements

    1. The instrument should be installed on a stable and solid platform that is free of mechanical vibration

    and away from vibration source.

    2. The environment should be as free as possible from dust, corrosive gas, loud noises and electrical

    interference.

    3. Avoid placing the instrument in direct sunlight or in front of a source of heat or vent.

    4. Ambient temperature: 10~35, relative humidity:20%-85%.

    5. The power supply should be AC220/110V10%, 50/60 Hz

    6. Power supply and grounding should be connected correctly.

    7. No strong electric field and magnetic field interference.

    1.3 External Connections

    1. Power cable: should be connected to the power socket firmly and reliably without loose or poor

    connection.

    2. Grounding: the grounding pole should be connected to ground properly. The grounding wire should

    be as short as possible and the grounding resistance must be less than 10

    3. If the power supply to the instrument is not stable, or there are high-power equipments on the same

    line, then a power stabilizer (output 100W) is strongly suggested.

    4. Avoid operating the instrument under high humidity environment for long time, otherwise the

    instrument performance may be affected. Use desiccant or air conditioner in case of necessary. Do

    not place the instrument near a water pool.

    1.4 Internal Connections

    Press the front door to open it. The tubing and electrodes have been installed at the factory, the

    operator still need to check them before using.

    1. Pump tube installation: put the aspirating pump tube and the lactic pump tube on the pump head

    according the illustration figure of front panel, fix the tube connectors to the bracket, adjust the tube

    to the middle of the pump head

  • 8/8/2019 Measuring PriNciple

    7/36

    2. Check the electrode assembly: take out the electrode assembly from the accessory kit. Check if the

    setting screws are tightened, inlet and outlet port are clean, the filling solution of each electrode is

    sufficient. If there are bubbles on the electrode probe, tap the electrode slightly to let the bubbles

    disappear.

    3. Installation of the electrode assembly: insert the whole assembly to the mounting base, connect the

    lead wire of each electrode (KNaCaPHCLRef) to the plug of amplifier board correctly.

    Connect the tube to the inlet and outlet port of the assembly.

    4. Check the liquid positioner: replug the positioner and make sure it is well connected.

    5. Check the tubing: make sure all the tubings are connected tightly and there is no leaking.

    6. Connect to the reagent pack: connect the Std A tube to port A of the reagent pack. Connect the Std B

    tube to port B. Connect Std R tube to port R Connect the waste tube to port W.

    7. Installation of printing paper: open the printer cover, insert the printing paper into the guide slot (the

    thermal side face down), feed the paper out from the slot on the printer cover.

    Note:

    z Make sure all tubings are connected firmly and there is no leakage.

    z Make sure the positioner is connected firmly.

    z Make sure all the lead wires of the electrodes are connected firmly.

    z Make sure there is no bubble on the electrode probes.

    z The printing paper is using out when you see two red lines on it. Change the printing paper

    in time.

  • 8/8/2019 Measuring PriNciple

    8/36

    Power-on check

    2.1 Self-test and adjustment

    When switch on the power to the instrument, system self-test will be performed automatically. The

    installation personnel or operator may be required to do some adjustments to ensure the instrument under

    good conditions.

    Operation Function properly Possible problems

    Switch on the power, the - External power supply, power - External power supply, power

    LCD light on. switch, internal power supply switch connections or fuse is

    works properly not good

    - LCD and cables works properly - Internal power supply doesnt

    work properly

    Detect the positioner, - The positioner works properly - The positioner loose.

    printer and auto sampler - The printer works properly - The printer cables or internal

    (if have)- Auto sampler works properly power supply is not good

    - The system will not give

    alarm or stop even if errors

    found on the positioner orprinter. The operator should

    turn off the instrument

    manually and solve the

    problems.

    Feed paper and print-out - The power supply and paper - Internal power supply is not

    the system version feeding of the printer is good good

    - Data transmission of the printer is - The printing paper is not

    good installed correctly.

    - The printer cables are loose

    or not good.

    Sample probe moves up - The lifting motor, limit switch and - The lifting mechanism is not

    until it touches the cables work properly. good

    position limit switch, Check if the sample probe is - Adjust or replace the sample

    then moves down to the bended or out of position during probe if necessary

    wash block the movement.

    - 5 -

  • 8/8/2019 Measuring PriNciple

    9/36

    Calibrating - The liquid distribution valve and - The system gives alarm and

    optical coupler, lifting stops if errors found on

    mechanism, switches and liquid distribution valve,

    software work properly lifting mechanism or

    switches

    The instrument aspirates - Pump motor works properly - Pump motor is malfunction

    Std B and Std A in turn - The tubing system is good if Std - The liquid can not reach

    for test and display the B and Std A can be aspirated into correct position, showing the

    voltages the distribution valve correctly tubing system is leaking

    - Details of voltage display and - The inlet tubes for Std A and

    print please refer to the operation B are connected to the

    manual wrong ports

    - The operator can turn off the - Std A or B can not be

    function of voltage display or aspirated

    print by extended program

    Screen display after - All electrodes, Std A, Std B, - If slope is not stable, then it

    calibration:tubing connections, pump, valve, shows shift

    Slope signal detection and data - If slope is abnormal, then itK 54.5 OK

    processing work normally shows XNa 52.3 OK - Refer to trouble-shooting forCl 51.6 OK

    Ca 25.5 OK the reason of unstable or

    pH 55.6 OK abnormal slope.

    The system enters into Main Menu

    main menu after self-test Sample Cal. W. ListSTD Service Results

    If there is no operation - the internal clock works normally - The internal clock is

    for more than 2 minutes, malfunction if the system

    the system will enters can not enter into standby

    into standby mode to mode

    maintain the electrodes.

    Note:

    z Try the other side of the printing paper or change the whole paper roll if there is no printing

    on the paper. Adjust the position of the paper if it is misalignment.

    z If the sample probe is found distortion, misaligned or bended, the operator should adjust or

    change the probe and make sure it can move into the wash block correctly.

    z If the aspiration is abnormal, the operator should check the tubing connections, reagent

    connections and electrode connections and make sure there is no blockage or leakage.

    z Activate the electrode for 10-30 minutes before using if the electrode has been stored or not

    used for long time. Run calibration when the electrode is stable.

  • 8/8/2019 Measuring PriNciple

    10/36

    2.2 Prompts during self-test and calibration

    2.2.1 Display Prompts

    Code Error message

    1 # The optical coupler of liquid distribution valve failed

    2 # Elevator switch failed

    2.2.2 Print Prompts

    Code Error message

    Error 0# Liquid positioning failed

    Error 1# Liquid detecting failed

    Warn 2# Bubbles detected

    Error 3# Too much or less sample aspirated

    Error 4# Slope abnormal

    Error 5# Slope unstable

    Error 6# AB Slope abnormal

    Error 7# AB Slope unstable

    For the solutions please refer to Trouble-shooting

    - 7 -

  • 8/8/2019 Measuring PriNciple

    11/36

    Function description and operation

    3.1 Flow chart of the operation

    Please refer to the operation manual.

    3.2 System functions

    Please refer to the operation manual.

    3.3 Extended programs:

    Press Service in the main menu, press Time and enter 983, press Exit, then the system will enter into

    extended program menu, the screen displays:

    Pr. Vol Pr. PH Pr. TCa

    Aj. PH Standby Valve

    Show mv Range Pr. clea

    Exit

    (Attention: the extended programs are not open for the end-users to avoid wrong setting)

    3.3.1 Print voltage

    Press Pr. Vol , the button turns white. The voltages of the liquid positioner and all electrodes

    will be printed out during calibration or sample measurement.

    Press Pr. Vol again, white color disappear from the button. The voltages will not be printed out

  • 8/8/2019 Measuring PriNciple

    12/36

    during calibration or sample measurement. This is the default setting.

    3.3.2 Print PH value

    Press Pr. PH , the button turns white. The pH value will be printed out when the sample

    measurement finish.

    Press Pr. PH again, the button turns dark. The pH value will not be printed out when the sample

    measurement finish. This is the default setting.

    3.3.3 Print TCa

    TCa is a calculated parameter.

    Press Pr. TCa , the button turns white. The value of iCa and TCa will be printed out when the

    sample measurement finish.

    Press Pr. TCa again, the button turns white. Only the value of iCa will be printed out when the

    sample measurement finish. This is the default setting.

    3.3.4 Adjust pH

    This function has been fixed at factory, no need to adjust it again.

    3.3.5 Standby mode setting

    The instrument will enter into Standby mode if there is no operation in 2 minutes. In standby mode,

    the operator should press Wake Up to activate the instrument to the working mode. If the instrument

    stays in standby mode for 4 hours, it will maintain the electrodes automatically. When wake up the

    instrument, it will carry out a calibration before returning to the working mode.

    The operator can change the waiting time for entering into standby mode. The suggested waiting time

    is 2-60 minutes.

    To change the waiting time, press Standby ,input the value, then press Yes to confirm.

    3.3.6 Test Valve

    The liquid distribution valve may be blocked or leaking after using for a long time. The measuring

    results could be affected. Therefore it is necessary to check the valve periodically.

    Touch Valve, the system will aspirate Std A and Std B automatically. Observe the aspiration volume,

  • 8/8/2019 Measuring PriNciple

    13/36

    - 9 -

  • 8/8/2019 Measuring PriNciple

    14/36

  • 8/8/2019 Measuring PriNciple

    15/36

    (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling them in the

    laboratory.

    4. The reagents are harmful to human body. If the reagents accidentally spill on your skin, wash them

    off with plenty of water and if necessary, go to see a doctor.

    5. If the reagents accidentally spill into your eyes, wash them off with plenty of water and

    immediately go to see a doctor

    3.4.2 Sample Collection and handling

    Sample collection and handling must be carried out by the professionals. Always avoid the

    Hematolysis. In addition, the following points should be noted:

    1. The serum or plasma can be stored in the refrigerator, but they must be warmed up to the room

    temperature before test.

    2. When preparing the blood serum samples, do not add any materials like the surface active agent

    that may interfere in the measurement or even damage the sensor.

    3. Hemolysis can affect K result significantly as well as other parameters. Special remark should be

    made for hemolysis sample, and the results are for reference to the lab only.

    4. If the serum does not been separated completely, the fibrins in the sample may affect the aspiration,

    and the proteins may deposit inside the tubing thus cause contamination or blockage. In such case,

    the results will be affected, so the operator should clean the tubing in time.

    5. The sample should be tested within 2 hours, otherwise the result may be not accurate.

    3.4.3 Operation precautions

    1. The analyzer is designed to work continuously for 24 hours a day. No need to shut down the

    machine everyday.

    2. Do not use the standard solutions for flame luminosity. They include strong acid and other

    supplements that may damage the electrodes.

    3. Not all commercial controls are suitable for ISE measurement. Some of them contain too much

    chemical additives that may interfere in the measurement.

    4. The bubbles cannot be aspirated during the sample aspiration, otherwise the results will be

    unreliable.

    5. When the sample reaches the checking point, make sure there are no bubbles inside, otherwise

  • 8/8/2019 Measuring PriNciple

    16/36

    the sample should be measured again.

    6. Never aspirate the sample directly from the blood separating tube, because it is very easy to block

    the tubing system. Always move the serum to a small and clean tube or cup with a pipette

    before testing on the instrument.

    7. If the ambient temperature fluctuates for more than 10, the instrument should be calibrated

    again.

    8. The pH value of the standard solutions and the samples should be within 6-9, otherwise it will

    interfere in the measurement of sodium ions.

    9. Discard the reagent if mildew or deposition found.

    10. Perform the routine maintenance according to the manual.

    11. Every electrode has a serial number. Keep it on the electrode. The warranty to the electrode will

    become invalid if the number can not be recognized.

    12. When remove the sample tube after aspiration, be careful not to hit the sample probe, otherwise

    the probe will bend or misaligned and can not move into the wash block properly.

    13. Do not open the front door during calibration or sample measurement, otherwise the result will be

    affected.

    14. Check and make sure the grounding is good if static interference is found.

    3.4.4 Daily operation

    Please refer to the operation manual.

  • 8/8/2019 Measuring PriNciple

    17/36

    Routine maintenance

    4.1 Daily maintenance

    Keep the instrument clean, especially the sampling unit.

    4.2 Weekly maintenance

    1. Check the electrodes and make sure the filling solutions are sufficient. Add filling solution if the

    volume is less than 2/3.

    2. Check if there is salty crystal on the internal electrode, if so, clean it with wet tampon.

    3. Check and clean the rubber in the wash block. Clean the sample probe.

    4. Run deprotein program in service menu. (If less than 5 samples per day, then just run deprotein

    program every 2 weeks) Check and clean the tubing system.

    5. Run Na adjust program if the slope of Na electrode is less than 50.

    4.3 Monthly maintenance

    1. Check the tubing system and make sure there is no blockage or leakage.

    2. Check the internal electrode and make sure it doesnt blanch or its coating doesnt flake off. Change

    the internal electrode if necessary. Check the filling solution, change it if it become feculent.

    3. Observe the voltage of the electrode. If the voltage of KNaCl are all less than 20 mV, then

    change the reference filling solution first. If it doesnt improve, change the reference membrane.

    4. Run calibration if the slopes are low.

    4.4 Check tubing system

    Check the tubing system if the aspiration volume or flow speed is abnormal, or bubbles found. Observe

    the flow speed, liquid position and bubble status during sample test or calibration.

    1. Blockage will cause insufficient aspiration volume for sample, Std A and Std B, or no aspiration at

    all if the blockage is very bad.

  • 8/8/2019 Measuring PriNciple

    18/36

    2. Leakage will cause bubbles or no aspiration.

    3. A segment of air will appear in the tube when the liquid distribution valve switches between Std

    A and Std B. When aspirating Std A, if small bubbles found at the outlet of distribution valve,

    then the flow path inside the distribution valve for Std A is leaking. If bubbles always stay at the

    inlet of Std A and can not be excluded, then the flow path for Std A is blocked slightly. Check

    flow path for Std B by the same way.

    4. The blockage in the distribution valve is usually caused by floccule in Std A or Std B. It is

    important to keep the reagents under proper temperature.

    5. The bubbles will appear near the connector if it is leaking. Tighten the tube or connect it again.

    Cut a small piece of tube if necessary.

    6. If leakage found between the electrodes, then the O ring between the electrodes may miss or out

    of position, or the setting screws are not tightened.

    7. If the pump tube conglutinated, distortion, leaking or aged, it will cause no aspiration,

    insufficient aspiration volume or incorrect liquid position. Change the pump tube if necessary.

    8. Wash the tubing system regularly to prevent the blockage caused by protein deposition. Clean

    the external of the sample probe. Clean the wash block.

    When the instrument can not aspirate, check the tubing system with Valve program in Service

    menu.

    Pull out the liquid outlet tube from the mixing chamber and put it into a waste container

    temporarily. Check if there is blockage and leakage at the segment in front of mixing chamber.

    (1) Take off the pump tube, check if it is conglutinated, distortion, leaking or aged. Check if the

    tubing connector is blocked. Check the connection between pump tube and waste tube. Put the

    waste tube into water, disconnect the connecting tube from the outlet of electrode assembly, and

    rotate the pump head clockwise by hand. If the water can be aspirated normally, then the pump

    is working well. Otherwise check and change the pump if necessary.

    (2) Put the waste tube and connecting tube back. Disconnect the aspirating tube from the sample

    probe and put it into water. Rotate the pump head clockwise by hand. If the water can be

    aspirated normally, then the electrode assembly is working well. Otherwise disassemble the

    electrode assembly for checking.

    (3) Put the aspirating tube back. Take off the tube from the outlet of the distribution valve and put it

  • 8/8/2019 Measuring PriNciple

    19/36

    into water. Rotate the pump head clockwise by hand. If the water can be aspirated normally,

    then the sample probe is working well. Otherwise check the sample probe.

    (4) If the results of the above checking are good, then most likely the distribution valve is

    malfunction. Change the valve if necessary.

    (5) If the segment in front of mixing chamber is normal, then the blockage and leakage should be

    happened in the mixing chamber.

    Blockage is usually in the outlet at the bottom of mixing chamber or the stainless steel tube for

    sample injection.

    Leakage is usually caused by

    - waste tube of mixing chamber broken

    - liquid outlet valve closed unfirmly

    - press tube of outlet valve distortion or broken

    - Four stainless steel tubes on the cover of mixing chamber not stuck well and leakage

    - Air outlet valve broken

    - The tube connected to four stainless steel tubes on the cover of mixing chamber broken.

    (7) Cleaning the mixing chamber

    Check all the tubes every 1-3 months. Replace the tube which is broken. The breakage is usually

    caused by the aged tube or incorrect cleaning and repairing.

    Blockage is usually caused by the serum protein deposit. Regular cleaning is necessary.

    Clean the mixing chamber by the following steps.

    a. Pull out the white tube connected to reagent R, which is in the middle of the mixing chamber, then

    inject 3ml sodium hypochlorous solution into the chamber, wait for 5 minutes, put the tube back

    b. Pull out four tubes on the mixing chamber cover straightly upward, loose the three setting nuts, take

    out three bolts and three nuts. Pull out mixing chamber cover, clean the chamber and all the protein

    deposit on the outlet. Clean the protein deposit under the cover, clean all four stainless steel tubes,

    cleaning mixer. Put the mixer back with small side downward, insert the mixing chamber cover (pay

    attention to the direction of four stainless steel tubes), tighten three nuts. Connect the Std R tube to the

    inlet in the middle of the reaction chamber, then connect the liquid outlet tube (smallest tube), sensor

    connection tube, and air outlet tube. Do not damage the tube.

  • 8/8/2019 Measuring PriNciple

    20/36

    Warning:

    z Pay attention to the biological pollution by the sample when connect to the external waste bottle

    temporarily. Dispose the waste instantly . Or add antiseptic into waste bottle in advance.

    z

    Wear the protection glove when clean the mixing chamber. Do not touch serum protein deposit.

    Wash and disinfect your hand after cleaning the chamber.

    1. liquid Storage Chamber 7. Liquid outlet tube

    2. Pressure Sensor 8. Gas outlet tube

    3. Pressure Sensor connection tube 9. Lactic Acid inlet of mixing chamber

    4. Sample inlet tube 10. Liquid exhaust valve

    5. Mixing Chamber cover 11. Gas exhaust valve

    6. Mixing Chamber 12. Gas outlet

    [ Illustration figure ofMixing chamber assembly]

  • 8/8/2019 Measuring PriNciple

    21/36

    4.5 Check the sample probe

    1. Adjust the sample probe if it is distortion, misaligned or bended. Make sure the sample probe

    can move into the wash block smoothly and correctly.

    2. If the sample probe is blocked, move it up to top position, take the probe out, and use a thin

    needle to remove the blockage from the probe.

    3. If the wash block is blocked, lift up the sample probe, remove the cover of the wash block, take

    out the rubber and clean it.

    4.6 Electrode maintenance and replacement

    4.6.1 Structure of the electrode

    4.6.2 Replacement of internal electrode and filling solution

    Follow the steps below to replace the internal electrode or filling solution:

    1. Pull out all electrode wires from the plugs. Remove the tubes from the inlet and outlet of the

    electrode assembly.

    2. Loose the fixing nuts of the electrode assembly.

    3. Disassemble the whole electrode assembly.

    4. Discard the electrode to be replaced.

    5. Take a new electrode and add the filling solution.

    6. Assemble the electrodes in correct order. Make sure that the rubber gaskets are in right position.

  • 8/8/2019 Measuring PriNciple

    22/36

    7. Tighten the fixing nuts of the electrode assembly.

    8. Connect the tubing to inlet and outlet. Insert the electrode wires back to the plugs.

    9. Run calibration again.

    If the internal electrode blanches or its coating flakes off, then remove the old filling solution, add freshfilling solution and install a new internal electrode.

    Change the filling solution if it is insufficient or the slope of the electrode is too low. Take out the internal

    electrode (do not touch any metal materials), remove the old filling solution, add fresh solution to 85% of

    total volume, then put the internal electrode back and tighten it.

    K filling solution is for K electrode only, Ca filling solution is for Ca electrode only, and reference filling

    solution is for reference electrode only. Na/Cl filling solution is for Na, Cl and pH electrodes.

    4.6.3 Replace the reference membrane

    Follow the steps below to replace the reference membrane:1. Pull out all electrode wires from the plugs. Remove the tubes from the inlet and outlet of the

    electrode assembly.

    2. Loose the fixing nuts of the electrode assembly.

    3. Take out the reference electrode.

    4. Loose the screw cap, take out the internal electrode, remove the filling solution, O-ring and old

    reference membrane (see the following figure).

    5. Soak the new reference membrane in distilled water, then place it evenly on the internal interface

    and put on the O-ring.

    6. Refill the reference filling solution and tighten the screw cap

    7. Clean and dry the electrode.

    8. Assemble the electrodes in correct order. Make sure that the rubber gaskets are in right position.

    9. Connect the tubing to inlet and outlet. Insert the electrode wires back to the plugs.

    10. Run calibration again.

  • 8/8/2019 Measuring PriNciple

    23/36

    - 18 -

  • 8/8/2019 Measuring PriNciple

    24/36

    4.7 Maintenance of pump, valve and lifting unit

    4.7.1 Maintenance of aspirating pump and Lactic acid pump

    After long time of using, the pump tube will get aged and may cause insufficient aspiration volume,

    therefore the accuracy of the results will become poor. If so, it is necessary to change the pump tube.

    If there is big noise when the pump head rotating, disconnect the pump tube, pull out the pump head,

    loose three setting screws, take out the plate, add a little lubrication oil to the ball bearing, then assemble

    the pump head back. If the pump head is damaged, then change it with a new one.

    4.7.2 Maintenance of distribution valve

    The function of the distribution valve is to switch the flow path to Std A or Std B according to the control

    program. The valve status is detected by an optical coupler. The performance of the optical coupler can be

    affected by dust or foreign objects, and may cause incorrect switch between Std A and B. In such

    case,1# 2# or 3# error message will show on the screen and be printed out. Clean the optical

    coupler periodically.

    Disassemble the valve:

    Loose four setting screws on the valve plate, pull the whole valve out a little, loose four nuts, take out the

    valve plate and four copper standings, loose the setting screws for the optical coupler. Take out the optical

    coupler. Turn the valve head with a tweezers and take it out. Loose two screws on the valve cap and take

    the cap out. Clean five O rings on the valve cap. Clean the valve body and rotor.

    Assemble the valve:

    Install the rotor and cap to the valve body, tighten two screws. Pay attention to the five O rings on the cap.

    Install the optical coupler to center position. Put on the valve head, tighten the screws on the valve motor

    axis. Put the copper standings back, tighten the valve body to the plate. Tighten the valve plate to

    instrument.

    Note:

    z Pay attention to the position of the O-rings, copper standings, optical coupler and connectors to

  • 8/8/2019 Measuring PriNciple

    25/36

    Std A and B.

    Maintain the valve once every year. Keep the optical coupler clean and check it periodically.

    4.7.3 Lifting unit

    The maintenance of the lifting unit is to remove the dust, foreign objects and oil dirt from it. Remove the

    cover of the sample probe, remove the sample probe, clean the axis and add a little grease. Let the sample

    probe move up and down for several times, make sure it moves smoothly and low noise.

    Note:

    z The electrode should be sealed and stored under dry environment. Do not add filling solution

    to it.

    z The tubing and O-rings should be sealed and stored under dry environment.

    z The internal electrode should be sealed and stored under dry environment. Do not store with

    metal materials.

  • 8/8/2019 Measuring PriNciple

    26/36

    . Hardware

    5.1 Connector and Cable list

    Connector Cable Connector Cable

    J101 Optical coupler cable J203 Positioner cable

    J102 Elevator switch cable J204 Signal cable

    J103 Auto sampler home position J205 RS232 cableoptical couplercable

    J104 Auto sampler counter optical J206 Pressure sensor cable

    coupler cable

    J107 Transformer output cable J207PRT Printer data cable

    J108 Printer power cable J207Key Touch screen cable

    J109 Air outlet valve connection

    cable J208 LCD cable

    J110 Liquid outlet valve connection

    cableNBQ LCD inverter power cable

    J111 Mixing motor power supply

    cable

    J112 Auto sampler motor cable

    J114 Lifting motor cable

    J115 Valve motor cable

    J116 Aspirating Pump motor cable

    J117 Lactic acid pump motor cable

    5.2 PCB errors

    1. Because of vibration or dust cumulating, the connectors or IC chips may have cold connections.

    Disconnect the power cable, open the rear cover, and check the cables and IC chips.

  • 8/8/2019 Measuring PriNciple

    27/36

    2. When switch on the power, the LCD doesnt light on, the lifting unit doesnt work, the pump motor

    doesnt work: check 9-pin DC power cable and transformer output cable J107.

    3. When switch on the power, the LCD doesnt light on, but the lifting unit work normally: Check

    U102 (LM7812)9-pin DC power cable, NBQ connector and 12V voltage. The LCD inverter may

    be damaged.

    4. The lifting unit can not move, the valve doesnt work: Check U103 (LM7812)9-pin DC power

    cableU106U107

    5. The pump motor doesnt work: Check U110LM78129-pin DC power cableU108U109

    6. The signals of calibration or measurement are confused or all zero: check cable J204.

    7. If the program works incorrect, clean the dust cumulating around U301 IC chip and check. It is

    recommended to clean with PCB washing solution (pure alcohol) and blow to dry

    8Auto sampler doesnt move, check U104, 74HC373,U105, ULN2803A.

    9No AB test result, check J206 connector and U209, LM324.

    10No print. Check J108, J207(PRT) connectors and the connectors between printer head and

    print control board. Replace the printer head if it is broken.

    Note:

    Only authorized personals are allowed to repair the PCB.

    5.3 Description of RS-232 interface

    The analyzer can send or receive command/data via RS-232 interface.

    5.3.1 Features

    Electric feature comply with EIA RS-232C

    Transmission mode asynchronism

    Stop bit1 digit

    Data bit8 digit

    Parity bitnone

    Transmission speed 19200 baud

  • 8/8/2019 Measuring PriNciple

    28/36

    5.3.2 Pin description

    Pin no. Signal Function Note

    1 NC none

    2 RXD Receive data Input

    3 TXD Transmit data Output

    4 NC none

    5 GND signal ground

    6 NC none

    7 NC none

    8 NC none

    9 NC none

    5.3.3 Data format

    The instrument communicates with external PC through RS-232 serial port. After each test finish, the data

    will be automatically sent out during the time of printing. The content of the transmitting data is almost

    same as the printing data. The operator can also send out the data manually by Send program in

    Service menu.

    Example:

    001 111111 063 5.23 151.2 111.4 1.44 7.62 12.0

    004 112224 063 5.25 151.6 110.8 1.40 7.59 42.0

    Data format:

    Sample number + 2 blank + ID + 2 blank + symbol byte + blank + k result + blank + Na result + blank +

    Cl result + blank + Ca result + blank + PH result + blank + AB result + newline

    Note:

    Symbol byte: this byte has 8 digits h g f e d c b a

    a =1: K electrode calibration failed, a =0: K electrode calibration OK

    b =1: Na electrode calibration failed, b =0: Na electrode calibration OK

    c =1: Cl electrode calibration failed, c =0: Cl electrode calibration OK

  • 8/8/2019 Measuring PriNciple

    29/36

    d =1: Ca electrode calibration failed, d =0: Ca electrode calibration OK

    e =1: pH electrode calibration failed, e =0: pH electrode calibration OK

    f =1: AB calibration failed, f =0: AB calibration OK

    h =1: warning information of this result (e.g. aspiration abnormal, bubbles detected, etc.)

    Trouble-shooting

    6.1 Error prompts

    Display prompts

    Code Error Solution

    1# Optical coupler error Check J101, clean optical coupler, adjust the position of

    optical coupler

    2# Elevator switch error Check J102, J114, elevator switch, U103(LM7812)

    U107(ULN2803)U106(74HC373) 9-pin DC power

    cable

    Printing prompts

    Code Error Solution

    Error 0# Liquid positioning Connect the positioner again

    failed

    Error 1# Liquid detecting failed Check the optical coupler and tubing connection of the

    distribution valve

    Error 2# Bubbles detected Check the positioner, optical coupler, distribution valve,

    check if there is any leakage in the tubing sstem

    Error 3# Too much or less Check the distribution valve and positioner

    sample aspirated

    Error 4# Slope abnormal Refer to 6.5 Slope abnormal

    Error 5# Slope unstable Refer to 6.4 Slope unstable

  • 8/8/2019 Measuring PriNciple

    30/36

    6.2 Possible causes of errors

    (1) Power supply not good

    May causethe instrument works abnormally; the slopes become abnormal during calibration; thetesting result not stable

    (2) Grounding not good

    May causeresult not stable during testing or calibration, especially for Na.

    (3) Interfere by electromagnetism

    May causethe instrument can not work normally; result not stable during testing or calibration

    (4) Interfere by static and dry

    May cause: instrument works unstable, the deviation of Na, Cl is too big

    (5) interfere by moisture

    May cause: instrument works unstable, the deviation of Na is too big

    (6) interfere by tempreture

    May cause: instrument works unstable.

    (7) Positioner not good

    May causeliquid position not accurate, prompt bubbles frequently, slope abnormal or

    unstable, testing result unstable

    (8)Flow path in the electrode assembly contaminated

    May causeslope abnormal or too low

    (9) Aspirating tube contaminated

    May causeslope abnormal, especially for K electrode

    (10) Std A or B contaminated

    May causeslope not stable or too low

    (11) Reference electrode not good

    May causeslope not stable, testing result not

    stable (12)Flow channel not good

    Check segment by segment.

  • 8/8/2019 Measuring PriNciple

    31/36

    6.3 Trouble-shooting of positioner and electrode

    Positioner:

    First, make sure the connection of the positioner is good. Then check the positioners voltage with

    Voltage program in Service menu. Rotate the pump manually and let the liquid pass through the

    positioner. The positioners voltage should become small when there is liquid or big when there is no

    liquid. The difference should be more than 500.

    Electrode:

    Check the voltage of the electrode by Voltage program in Service menu. If the voltage is low, then

    the electrode should be replaced as soon as possible.

    (3) If aspiration alarm appeared or no aspiration at all, replace a new reagent pack (Std A, Std B, Std

    R). Restart the instrument and run self-test.

    6.4. Slope unstable

    Cause Solution

    Unreliable grounding Check the connection of the

    grounding wire

    No standard A or standard B Check and replace Standard A or B;

    aspirated Check the tubing connection

    Incorrect positioning Adjust positioning again

    The reference filling solution or reference Replace when necessary

    membrane not working

    The internal electrode turns gray Replace when necessary

    Poor connection of the electrode wire Check and connect again

    Electrode membrane leaks Replace when necessary

    Power supply voltage fluctuates Use UPS or power stabilizer

    Humidity too high Lower the humidity or move the instrument to a dry place

    Bubbles inside the tubing Check the tubing system

    Liquid leakage inside the valve Replace gasket or valve when necessary

    The electrode is not activated or the Activate the electrode first

    activating time is insufficient

    Mixing Chamber Blocked .Clean Mixing Chamber

    26

  • 8/8/2019 Measuring PriNciple

    32/36

    6.5 Slope abnormal

    Cause Solution

    Too many organic deposits on the Wash it with de-protein

    electrode membrane SolutionReagent contaminated Replace the reagent

    Insufficient filling solution Refill the filling solution

    Electrode does not work Replace the failed electrode

    Dust or moisture around the electrode or plug Clean and dry the electrode and plug

    tubing loose reconnected

    Pump tube stuck recover the tube

    Pump tube broken replace the tube

    Tubing blockage clean the tubing

    Gasket between the electrodes wrong Re assemble itplaced or missing

    Leakage between the electrode Reinstalled firmly

    Optical coupler too dirty or broken Clean or replace it

    Mixing Chamber Blocked Clean Mixing Chamber

    6.6 AB slope and result abnormal

    Cause Solution

    Mixing Chamber Blocked Clean Mixing Chamber

    Mixing chamber leakage Check tubing connection and chamber cover

    condenser condensing Liquid outlet blocked

    AB slope is zero No Std R, pressure sensor cable loose or damaged.

    Mixer dont move Reconnect J111 , replace mixing motor

    Liquid outlet valve dont work Reconnect J111 , replace liquid outlet valve

    Air outlet dont work, leakage, AB slope is low Reconnect J109, replace air outlet valve

    AB slope abnormal Std R invalid, leaked or blocked

    6.7 Auto sampler trouble shooting.

    Cause Solution

    Auto Sampler can not be detected Check J112 connector and cable,

    replace Auto sampler motor

    The initial position of auto sampler is not on Check J103 connectors, clean the dust on the optical

    the clean position coupler for auto sampler home position, replace the

    optical coupler

    - 27 -

  • 8/8/2019 Measuring PriNciple

    33/36

    Auto sampler on incorrect position Check J104 connector and cable, clean the dust on

    the counting optical coupler,

    replace the optical coupler

    6.8 Aspiration abnormal

    Cause Solution

    Aspirating tube loose or broken Connect again or replace it

    Pump tube sticks Restore the tube

    Pump tube broken Replace the tube

    Pump tube blocked Clear the blockage

    The gasket between the electrodes does Place the gasket properly

    not placed properly or missing

    The electrode assembly leaks Tighten the assembly again

    Dust in the positioner, or Clean or replace the positioner

    positioner damaged

  • 8/8/2019 Measuring PriNciple

    34/36

    BLOOD GAS ANALYSER

    The Blood Gas Analyzer have the following essential components.

    Essential Components

    1. Electrodes

    The system should be able to measure accurately the following parameters.

    pH

    pCO2

    pO2

    Haematocrit and Hemoglobin

    Electrolytes, Sodium, Potassium and Chloride Lactate

    Calcium

    magnesium

    The equipment should possess electrodes with long life of atleast 2 years

    Facility for regular quality assessment of the instrument should be provided by thecompany.

    2. Analyzer controller with soft ware.

    The Instrument should provide the following calculated parameters.

    Bicarbonate (HCO3)

    Standard HCO3 (standard HCO3)

    Base Excess of Blood (BE)

    Base Excess of Extra cellular fluid (BE-Ecf)

    Oxygen Content (O2Ct)

    Oxygen Saturation (SO2%)

    Total carbon dioxide (TCO2)

    Alveolar to Arterial oxygen tension gradient (AaDO2)

    Arterial Alveolar oxygen tension Ratio (a/A)

    Oxygen carrying capacity (O2 Cap)

    PO2/FIO2 Ratio

    P 5O

    Respiratory index

    Anion Gap

    Plasma Osmolatity

  • 8/8/2019 Measuring PriNciple

    35/36

    pH/pCO2/pO2 Corrected to patient temperature

    All results should be available within 3 min.

    Reagents remaining status should be available on the main screen for easy monitoring and

    replacement.

    The instrument should have facilities like monitor screen, external keyboard, mouse, and barcode

    reader

    All results should be microprocessor controlled and of latest technology version

    The instrument should have the capability to interface a computer and a computer

    should be supplied for data acquisition and patient record with recommended

    software. The system should have RS232 serial port.

    3. Recording Devices:

    High end colour inkjet printer. Refilling of cartridges should be possible compatible with ABG machine.

    4. Sampler

    The sample volume for measuring all parameters should exceed 200 ul

    The instrument should accept heparinized whole blood, serum or plasma, arterial, mixed venousand capillary sample

    It should also provide the facility to measure the above parameters in gases, cerebrospinal fluid,

    dialysate, pleural fluid and urine.

    5. Reagents

    The instrument should use liquid calibration for calibration of all measure parameters without useof any gas cylinders or humidifiers

    The company should supply reagents with the analyzer to run the machine

    satisfactorily for two years. 25 samples / day.

  • 8/8/2019 Measuring PriNciple

    36/36

    The equipment should be incorporated to perform regular calibration satisfactorily

    The instrument should have a standby mode(economy mode) with auto wake up facility to

    optimize the reagent utilization

    .

    6. Waste bag

    The Waste container should be sealed to prevent operator biohazard

    7. Back up power supply: 1 KV UPS systems for blood gas analyzer for a minimum one hour back

    up.

    8. Power requirements: 220-240 V, 50/60 Hz

    B. Optional accessories:

    P.C.Computer with facility of data transfer for mass storage of data to interface with the

    ABG analyzer

    a) P.C. Monitor 15 Flatb) Processor: Intel Pentium 4, 2.8 GHz with HT

    c) Mother board: Intel 915 Express Chipset mother board or higher (Compatible

    with the processor)

    d) Memory: 512 MB DDR SD RAM or better

    e) 80 GB Hard Disk

    f) 144 MB FDD

    g) DVD &CD Reader & writer for having back up data

    h) 2 USB, One Parallel & Serial port

    i) Latest configuration available

    j) Optical mouse

    k) Multi media key board

    Software: a) Windows XP operating system

    b) compatible software to transfer and analyse data from the ABG machine1. All spares parts and consumables should be available with the supplier or principals

    for the period of 10 years.