measurement of the absolute efficiency, with a precision better than 2%,

22
Measurement of the absolute efficiency, with a precision better than 2%, of a PMT working in single photoelectron mode Philippe Gorodetzky APC lab, Paris, France VLVT09, Athens, October14, 2009

Upload: carmelita-aguilar

Post on 03-Jan-2016

26 views

Category:

Documents


0 download

DESCRIPTION

Measurement of the absolute efficiency, with a precision better than 2%, of a PMT working in single photoelectron mode. Philippe Gorodetzky APC lab, Paris, France. VLV  T09, Athens, October14, 2009. How to calibrate ?. Comparison to a reference. Calibrated source. Calibrated - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Measurement of the absolute efficiency,  with a precision better than 2%,

Measurement of the absolute efficiency, with a precision better than 2%,

of a PMT working in single photoelectron mode

Philippe GorodetzkyAPC lab, Paris, France

VLVT09, Athens, October14, 2009

Page 2: Measurement of the absolute efficiency,  with a precision better than 2%,

How to calibrate ?

Comparison to a reference

Calibratedsource

Calibrateddetector

(photodiode)

Page 3: Measurement of the absolute efficiency,  with a precision better than 2%,

1) Calibrated Source

• GOOD : Only 1 measurement– Time variations of no importance

• BAD : Control of the spatial variations of the source. ==> IMPOSSIBLE– Angle & surface of emission (Liouville)

Page 4: Measurement of the absolute efficiency,  with a precision better than 2%,

2) Calibrated Detector

• GOOD : Spatial variations of no importance if comparison made in same conditions

• BAD : Need 2 measurements=> time variations important, but can

be controlled through a third detector ==> POSSIBLE

Page 5: Measurement of the absolute efficiency,  with a precision better than 2%,

Needs

• NIST photodiodes have a gain of about 0.5. So the light flux has to be reduced ~106 times

• Avoid geometry problems in illumination=> exactly same geometry for PMT and

calibrated detector

Page 6: Measurement of the absolute efficiency,  with a precision better than 2%,

How (is the calibration done)2 steps

• Mapping of the photocathodesRELATIVE

• Comparison to a NIST photodiode @ 1 position

ABSOLUTE

Page 7: Measurement of the absolute efficiency,  with a precision better than 2%,

Mapping of the photocathodes

Page 8: Measurement of the absolute efficiency,  with a precision better than 2%,

Very few photons

We illuminate with an LED of the good wavelength, and pulse at one kHz in order for the ADC to follow. To make a single photoelectron spectrum, while we acquire on the ADC, we lower the quantity of photons sent per pulse until the obtained peak has a stable position*. Then the number of events in that peak lowers while the number of events in the pedestal increases. When do we stop lowering the light?

The spectrum is mainly a one pe (the bump), but there is still a « peak » at 2 pe, very weak, which will be troublesome when a discriminator is set between the pedestal and the 1 pe, and we count with a scaler: each 2 pe counts double, and the result will be wrong.

scaler

scaler

•One can also use an oscilloscope and watch when the base-line under the pulse begins to fill

SINGLE PHOTO-ELECTRON

Page 9: Measurement of the absolute efficiency,  with a precision better than 2%,

We use Poisson: and P0, P1, P2 are the respective populations of the

pedestal, of the 1 pe and of the 2 pe

P0 = (m0 / 0!)e-m = e-m

P1 = (m1 / 1!) e-m = m e-m = m P0

P2 = (m2 / 2!) e-m = (m2 / 2) e-m = (m / 2)P1 = (m2 / 2)P0

If one wants that P2 = 1% x P1, (m / 2)P1 = 0.01P1, then m / 2 = 0.01, and m = 0.02

Now, the ratio between P0 and P1: P0 / P1 = P0 / (mP0) = 1 / m will be 1 / 0.02 = 50

In our case, as soon as the pedestal is 50 times more important than the 1 pe, the 2 pe will be less than 1% of the 1 pe

Usually, one takes: pedestal = 100 times 1 pe. Then we are sure not to pollute the measurement. Now we can set the discriminator threshold to be in the bottom between the pedestal and the 1 pe (at 0.25 of the 1 pe), and we just have to count in two scalers the pulses sent to the LED and the discriminator output. Exit the ADC: one can pulse until 100 kHz, which allows comfortable statistics in a few minutes.

Also, the threshold being in a valley, the measurement will not be very sensitive to a small variation of the threshold, or of the gain due to HV small changes.

Px =mx

x!e−m

SINGLE PHOTO-ELECTRON

Page 10: Measurement of the absolute efficiency,  with a precision better than 2%,

Mapping of the photocathodes

Reduce the light per pulse & adjust the gain

Optical fiber

•One can also use an oscilloscope and watch when the base-line under the pulse begins to fill

Page 11: Measurement of the absolute efficiency,  with a precision better than 2%,

Mapping of the photocathodes

In red: Coïncidences between generator & PMT discriminator

Page 12: Measurement of the absolute efficiency,  with a precision better than 2%,

Mapping : « PMT-JY »

The photocathode is naked ( = 51 mm)

Page 13: Measurement of the absolute efficiency,  with a precision better than 2%,

Mapping of the photocathodes.Here, absolute

Better efficiency if we use only the central part=> diaphragm of 20 mm

Full pmt (40 mm diameter)

Page 14: Measurement of the absolute efficiency,  with a precision better than 2%,

Absolute measurement

• PMT and 1 photodiode at the same time

• BUT : very different gains : 1 vs 107

=> how to divide a light flux by 107 ?

Page 15: Measurement of the absolute efficiency,  with a precision better than 2%,

Absolute measurement

• Use of integrating spheres to reduce light

• Measurement of the light flux reduction

• Measure the PMT efficiency

Page 16: Measurement of the absolute efficiency,  with a precision better than 2%,

http://www.labsphere.com/data/userFiles/A%20Guide%20to%20Integrating%20Sphere%20Theory&Apps.pdf

SINGLE PHOTO-ELECTRON

Page 17: Measurement of the absolute efficiency,  with a precision better than 2%,

If one measures 1 nW in the second diode (noise = 1 pW) and 0.775 mW in the first :

Ratio = 7.75 105

Calibration of the system

Page 18: Measurement of the absolute efficiency,  with a precision better than 2%,

ANALYSIS

100 kHz: 14.425 nW in NIST

As the ratio = 7.75 105

One sends on the PMT: 14.425 nW / 7.75 105 = 1.861 10-5 nW

Energy of a photon @ 378 nm: E = h = hc/

E = 6.026 10-34 x 3 108 / 378 10-9 = 4.783 10-19 J

One knows that 1 J = 2.090 1018 photons

So 1.861 10-5 nW ==> 1.861 10-14 x 2.090 1018

= 38912 photons / sec.

In one measurement of 100 sec, we have sent on the PMT: 3891200 ph

We have measured 686797 pe ==> efficiency (discri) = 686797 / 3891200 = 17.65%

We have to add 8.8% (discri) so efficiency = 19.2 % at 377 nm (PMT center), and 15.8% for full pmt, instead of 22% given by Photonis

Discri

Calibration of the PMT

Page 19: Measurement of the absolute efficiency,  with a precision better than 2%,

Absolute measurement

Uncertainties :

• Flux reduction (ratio) : 3 % (2 NIST diodes)

ΔR/R = (ΔI/I + Δα/α)udt + (ΔI/I + Δα/α)o1

• Efficiency measurement : 1.7 % (1st NIST cancels out)

Δε/ε = ΔR/R - (ΔI/I + Δα/α)udt

Page 20: Measurement of the absolute efficiency,  with a precision better than 2%,

3 Leds

NIST Photodiode

trans-impedance amp.

Integrating sphere

4 cmAmplification of TTL pulses in 40 V pulses with a risetime of 2 ns collimator

If one wants a more collimated photon beaminstead of Lambertian distribution

Another way to look at the set-up: the first sphere is a "perfect" splitter (to the NIST and the first diaphragm)followed by a very stable light reducer.

Page 21: Measurement of the absolute efficiency,  with a precision better than 2%,

One application: AntaresAnd why not NESTOR, or km3 ?

They calibrate their system with atmospheric muons,but do not know very well (!!!) the efficiency in the back of the tube

X,Y,Z, , movement in a black box

The light source:

- integrating sphere

- collimator

Page 22: Measurement of the absolute efficiency,  with a precision better than 2%,

Another application: JEM-EUSO 36 pixel Hamamatsu PMT

Efficiency of pixel 22, at 378 nm, and looking only at anode 22 = (17.9 ± 0.32)%

Then, the sum of all 9 pixels gives the full pixel 22 absolute efficiency at 378 nm when looking at all 9 anodes: (29.8 ± 0.54)% where the relative uncertainty has the same causes than above, and is 1.8%.

Assuming a collection efficiency of 70% (Hamamatsu), one gets a quantumefficiency of 40%Less than 1% of the counts are in coincidence in any combination of 2 pixels

So, it is not a cross-talk, but a point spread function of 5.8 mm diameter, twice thediameter of the PSF of the lenses, that is 4 times its surface.

Hence, Hamamatsu is designing a new PMT, with a better focus (a 64 pixels)

We illuminate the center of pixel 22 with a spot of 1 mm size.