mdch lab poster bonita3

1
PBDE Analysis in Human Blood Serum by Gerstel Twister© Stir-bar Sorptive Extraction Technique and GC/MS-ECNI-SIM Quality Control Materials Calibrators are prepared in pre-screened blank fetal bovine serum Three method controls are added to each batch at blank, low, and high levels. These are Blank: consisting of prescreened Fetal bovine serum, QL: NIST Human Serum 1957, and QH: NIST Human Serum 1958. The NIST 1958 also contained normal human population ranges of chlorinated pesticides and chlorinated biphenyls to demonstrate the selectivity of the method. A serum surrogate containing known concentrations of analytes similar to the compounds of interest, but not commonly found in the sample matrix tested (PBB-155) is spiked into every calibrator, unknown and QC sample tested in order to monitor extraction efficiency and analyte recovery. Internal standard consists of difluoro-BDE_047 (FBDE-47) and BDE156 are spiked into each sample. Stir-bar Sorptive Extraction (SBSE) Extraction procedure is outlined in Figure 1. 0.25 ml of serum is diluted 1:1 with water and denatured using equal volume concentrated formic acid. Twister stir bars are added and allowed to stir at room temperature for 2 hours prior to analysis. Analytes are desorbed from the stirbar in a thermo-desorption unit (Gerstel TDU) and cryo-trapped using a Gerstel CIS4, see Figure 2. A Custom BDE Mixture (1ppm for each congener, AccuStandard Inc.) was serially diluted using isopropanol to prepare the spiking standard. Standard was spiked into screened fetal bovine serum to prepare a 6-point calibration curve (0.064 ppb to 3.20 ppb). Quality control materials run with each batch consisted of blank fetal bovine serum (QB), NIST Human serum SRM-1957 (QL) and Human serum SRM-1958(QH). Contact Information Authors: Colin Johnson, Paul R. Loconto, Michael O’Keefe, Bonita Taffe www.michigan.gov/mdchlab “Providing Quality Laboratory Science for Healthier People and Communities” Analysis Stir Bar Conditioning/Cleaning Results Conclusion References Acknowledgements Standards and Limit of Detection Methods Michigan Department of Community Health - Bureau of Laboratories 3350 N. Martin Luther King Blvd. P.O. Box 30035 Lansing, MI 48909 Office: (517)- 335-9490 E-Mail: [email protected] E-Mail: [email protected] The quantification of PBDEs in human serum is carried out using capillary gas chromatographic (C-GC) separation with electron capture negative ion mass selective detection (ECNI-SIM, where SIM stands for selective ion monitoring). In this method, ten BDE congeners and two PBB congeners (see Figure 2) are baseline separated and all but BDE- 183 is quantitated against difluoro-BDE_047. BDE-183 is quantitated against BDE-156 due to comparable retention time. This methodology detects eluting analytes as bromine ions at m/z 79, with the abundance m/z 81 serving as a qualifier ion. All targeted analyte peaks must meet the qualifier ion criteria (abundance of m/z 81 100% relative to the abundance of quant ion m/z 79). A Multi-Purpose Sampler Single Rail, Single Injector (MPS) was used [Gerstel GmbH & Co. KG]. The MPS is mounted atop the GC-MSD (Agilent 6890 GC /5973N MSD). The MPS is integrated with ChemStation® [Agilent Technologies, Inc.] and is programmed using Maestro software [Gerstel GmbH & Co. KG]. The stir-bar is placed in a conditioned Twister desorption liner and automatically transferred from the tray holder to the Thermal Desorption Unit (TDU), see Figure 3. Analytes sorbed on Twister are thermally desorbed in the TDU and then are cryotrapped onto a glass liner inside a Cooled Inlet System (CIS). Analytes are then thermally desorbed in the CIS into a DB-XLB (J&W and Agilent Technologies, Inc.) capillary column. Background levels of BDE47 and BDE99 from dust are common problems with trace analysis for these compounds due to the ubiquitous use in building materials. Steps taken to lower background in this method includes special washing of all glassware with dust free storage and the following procedure to condition the stirbars: Soak Twisters (50:50 MeOH:MeCl, 2h) Remove and dry under nitrogen Soak in Toluene overnight Remove and dry under nitrogen Place Twisters in glass tubes, bake at 300oC for 2 hrs under dry N2 Remove Twisters and allow to cool in holder Increase temperature on conditioner to 350oC, bake for 5 minutes and remove Store in clean, dust free vial until next use. Good separation of congeners was achieved with this method and background levels were acceptable as shown in Figure 5 This method uses a 6 point calibration curve in serum for analysis of BDE congeners, R2 0.99. Quality control limits using NIST SRM 1958 for QH and are shown in Figure 6 The LOD for each congener measured is less than 0.2 ppb No interference with extraction of BDEs by the SBSE was detected from other analytes in the NIST SRM 1958 (which contains PCBs, pesticides and other organochlorine metabolites) Measured analytes were averaged and compared with NIST reported levels, Figure 7. All BDEs except BDE 85 were within 2SD of reported values; BDE 85 was within 3SD. Measured values for PBB 153 were lower than NIST reported values. Accuracy and Precision were measured using replicates of the Arctic Monitoring and Proficiency Program (AMAP) samples as well as the NIST SRM and compared to proficiency program participant results. See Figure 8 A rapid analytical method which uses 0.25 ml of human serum has been validated for PBDEs This method uses Gerstel Twister SBSE extraction. A procedure for enhanced BDE background elimination and cleanup for stirbar re-use has been developed. The method has been demonstrated to be both accurate and precise at concentration levels found in human subpopulations PBDE-47 and 99 are generally demonstrated in analysis. This study demonstrates ways to reduce high environmental background levels without extreme measures (clean room preparation). The limit of detection for this method for each congener is less than 0.2 ppb. This method could be used for population screening to identifying more highly exposed subpopulations. 1. ATSDR. Toxicological Profile for Polybrominated Biphenyls and Polybrominated Diphenyl Ethers (PBBs and PBDEs); Agency for Toxic Substance and Disease Registry: Atlanta, GA, 2004. 2. Betts KS 2008. Unwelcome Guest: PBDEs in Indoor Dust. Environ Health Perspect 116:A202-A208. 3. Report on Human Biomonitoring of environmental Chemicals in Canada, Measures Survey Cycle 1 (2007), Health Canada/ Laboratorie de Toxicologie, Institut National de Sante Publique, Quebec, CA, www. healthcanada.gc.ca 4. Sjödin A, Wong LY, Jones RS, Park A, Zhang Y, Hodge C, DiPietro E, McClure C, Turner W, Needham LL and Patterson, DG, “Serum Concentrations of Polybrominated Diphenyl Ethers (PBDEs) and Polybrominated Biphenyl (PBB) in the United States Population: 2003–2004”, Environ. Sci. Technol., 2008, 42 (4), pp 1377–1384. 5. Johnson, PI, Stapleton, HM, Sjodin, A, Meeker, JD, “Relationships between Polybrominated Diphenyl Ether Concentrations in House Dust and Serum”, Environ Sci Technol, 44, 5627-32, 2010. 6. Loconto, P.R., “Evaluation of automated stir bar sorptive extraction-thermal desorpton-gas chromatography electron capture negative ion mass spectrometry for the analysis of PBDEs and PBBs in sheep and human serum” Journal of Chromatographic Science, 47 (2009) 656-669. 7. Loconto,P.R. “Selectivity and sensitivity improvements for selected polybrominated diphenyl ethers and polybrominated biphenyls using capillary gas chromatography/electron capture neagative ion mass selective detection: a cost effective approach to biomonitoring” LC-GC North America 26(11) (2008) 1118- 1130. 8. J.K. Taylor. Quality Assurance of Chemical Measurements. Lewis Publishers, Boca Raton, FL 1987 IUPAC name Abbreviation LOD 1 ppb 2,2’,4,4’,5,5’-Hexabromobiphenyl PBB-153 2,2’,4,4’,6,6’-Hexabromobiphenyl PBB-155 2,4,4’-Tribromodiphenyl Ether BDE_028 0.122 2,2’,4,4’-Tetrabromodiphenyl Ether BDE_047 0.165 2,3’,4,4’-Tetrabromodiphenyl Ether BDE_066 0.112 2,2’,3,4,4’-Pentabromodiphenyl Ether BDE_085 0.160 2,2’,4,4’,5-Pentabromodiphenyl Ether BDE_099 0.160 2,2’,4,4’,6-Pentabromodiphenyl Ether BDE_100 0.190 2,2’,4,4’,5,5’-Hexabromodiphenyl Ether BDE_153 0.182 2,2’,4,4’,5,6’-Hexabromodiphenyl Ether BDE_154 0.168 2,2’,3,4,4’,5’,6-Heptabromodiphenyl Ether BDE_183 0.192 5,5’-difluoro-2,2’,4,4’-tetrabromodiphenyl ether F-BDE_047, IS 2,3,3’,4,4’,5-Hexabromodiphenyl ether BDE_156, IS LOD was determined by calculating the standard deviation at each standard concentration following repeated measurements (n=15-20) of the lowest 3 concentration standards in serum. The absolute values of the standard deviations were then plotted versus concentration. The intercept of the least squares fit of this line equals So, with 3So being the LOD. J.K. Taylor. Quality Assurance of Chemical Measurements. Lewis Publishers, Boca Raton, FL 1987. Figure 2 Sample Preparation for Twister Stir-bar Sorptive Extraction in Human Serum Matrix Rinse a 5mL headspace vial 3x with dichloromethane. Place a previously cleaned and conditioned Twister [10mm long and 0.5mm film thickness] into the headspace vial. To 0.25mL of unspiked or spiked human serum is added 0.25mL of HPLC Grade H 2 O. Then add 0.50mL of 5% acetonitrile in concentrated formic acid. Stir at 500 rpm for exactly two hours then removed the Twister and place in HPLC Grade H 2 O to remove biological material. Dry the Twister with a Kim-Wipe or equivalent and place the Twister in a clean desorption liner. Carefully place the transport adapter atop the desorption liner that contains the Twister and place in the TDU tray holder. Figure 1 Figure 3 Thermal Desorption Unit liner with Twister stirbar in place Thermal Desorption Unit (TDU) on Agilent GC Figure 4 Figure 6: QC CHARTS F11-PBDEs PBDE_028_QH Individ.: cl : 0.467143 ucl: 0.541331 lcl: 0. 392955 * Rule violation Subgrp Size 1 O O O O O O O O O O O O O O O O O O Individ. cl lcl ucl 0.4 0.45 0.5 0.55 2 4 6 8 10 12 14 16 18 20 F11-PBDEs PBDE_047_QH Individ.: cl : 0.609333 ucl: 0.921229 lcl: 0. 297438 * Rule violation Subgrp Size 1 O O O O O O O O O O O O O O Individ. cl lcl ucl 0.4 0.6 0.8 2 4 6 8 10 12 14 16 18 20 F11-PBDEs PBDE_100_QH Individ.: cl : 0.335294 ucl: 0.407457 lcl: 0. 263131 * Rule violation Subgrp Size 1 O O O O O O O O O O O O O O O O O Individ. cl lcl ucl 0.25 0.3 0.35 0.4 2 4 6 8 10 12 14 16 18 20 F11-PBDEs PBDE_099_QH Individ.: cl : 0.364444 ucl: 0.567712 lcl: 0. 161177 * Rule violation Subgrp Size 1 O O O O O O O O O O O O O O O Individ. cl lcl ucl 0.2 0.3 0.4 0.5 0.6 2 4 6 8 10 12 14 16 18 20 F11-PBDEs PBDE_154_QH Individ.: cl : 0.268421 ucl: 0.398075 lcl: 0. 138767 * Rule violation Subgrp Size 1 O O O O O O O O O O O O O O O O O O Individ. cl lcl ucl 0.15 0.2 0.25 0.3 0.35 0.4 2 4 6 8 10 12 14 16 18 20 F11-PBDEs PBDE_153_QH Individ.: cl : 0.298333 ucl: 0.413582 lcl: 0. 183085 * Rule violation Subgrp Size 1 O O O O O O O O O O O O O O O O O Individ. cl lcl ucl 0.2 0.25 0.3 0.35 0.4 2 4 6 8 10 12 14 16 18 20 F11-PBDEs PBDE_183_QH Individ.: cl : 0.38 ucl: 0.549917 lcl: 0. 210083 * Rule violation Subgrp Size 1 O O O O O O O O O O O O O O O O O O Individ. cl lcl ucl 0.2 0.3 0.4 0.5 2 4 6 8 10 12 14 16 18 20 NIST 1958 (QH) Serum Chromatogram showing baseline separation of congeners. Time, Minutes Blank Fetal Bovine Serum spiked with FBDE-47 and BDE-156 internal standards. This figure illustrates a stable baseline, indicating a clean solid-phase on the stir-bar as well as a clean blank serum for calibrators. BDE/BB Congener Retention Time (min) BDE-28 8.780 BDE-47 10.035 BDE-66 10.495 BDE-100 12.353 BDE-99 13.181 BDE-85 14.190 BB-153 15.504 BDE-153 16.659 BDE-183 19.383 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 500 1000 1500 2000 2500 3000 3500 4.369 9.813 11.182 20.923 Abundance FBDE-47 BDE-156 Time, Minutes 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 500 1000 1500 2000 2500 3000 3500 4000 4500 5000 5500 6000 6500 7000 7500 8000 8500 3.676 3.730 3.839 3.909 3.952 4.024 4.222 4.310 4.349 4.457 4.528 5.251 5.388 5.485 6.491 7.150 7.547 8.780 9.832 10.035 10.495 12.220 12.353 13.181 14.910 15.232 15.504 16.659 19.383 Abundance Figure 8: AMAP Recovery AMAP W0906 0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 BDE-28 BDE-47 BDE-100 BDE-99 BDE-154 BDE-153 BDE-183 PPB Inter-Lab Twister AMAP W0909 0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 BDE-28 BDE-47 BDE-100 BDE-99 BDE-154 BDE-153 BDE-183 PPB Inter-Lab Twister AMAP W1004 0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 BDE-28 BDE-47 BDE-100 BDE-99 BDE-154 BDE-153 BDE-183 PPB Inter-Lab Twister AMAP W1005 0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 BDE-28 BDE-47 BDE-100 BDE-99 BDE-154 BDE-153 BDE-183 PPB Inter-Lab Twister AMAP W1007 0.00 0.10 0.20 0.30 0.40 0.50 0.60 BDE-28 BDE-47 BDE-100 BDE-99 BDE-154 BDE-153 BDE-183 PPB Inter-Lab Twister AMAP W1008 0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 BDE-28 BDE-47 BDE-100 BDE-99 BDE-154 BDE-153 BDE-183 PPB Inter-Lab Twister Abstract Introduction Polybrominated diphenyl ethers (PBDEs), widely used fire retardants, are currently under investigation as agents interfering with immunological, embryonic development and endocrine functions. Low population background levels have favored analysis by high resolution mass spectrometry (HRMS), limiting this analysis to highly specialized laboratories. This presentation describes a rapid solvent free extraction of PBDEs from a small serum sample (0.25 ml), with processing time of 2-3 hours using Twister® stirbar sorptive extraction. Gas chromatographic separation with electron capture negative ion mass selective detection (GC/ MSD-ECNI-SIM) produces quantitative results with detection limits fit for purpose using instrumentation readily available in most analytical laboratories. Polybrominated diphenyl ethers (PBDEs) have been widely used as fire retardants in many consumer items including polyurethane foam found in furniture and carpeting, home electronics and other textiles. In 2004 Penta-BDEs and Octa-BDEs were banned by the European Union and US production was phased out at this time, however items containing these chemicals persist. The BDEs consist of 10 homolog groups (1 to 10 bromine atoms) with 209 possible congeners. BDEs are not chemically bound to products and can be mobilized by leaching or degradation of treated materials over time. Concentrations of BDEs in indoor air and dust have been shown to be significantly higher than outdoor concentrations. A recent study of the relationship between house dust and serum concentrations of BDEs in adults shows high correlation5. Young children are considered to be at highest risk of exposure from house dust due to their full body contact with the floor and carpets and hand to mouth activity. BDEs are classed as potential endocrine disruptors and the mechanism of action is currently under investigation in animal models with respect to immunological, embryonic development and endocrine functions1,2. Further epidemiological research to evaluate human response to these agents is needed. Low general population background levels and background environmental contamination by these analytes has favored quantitative analysis by high resolution mass spectrometry (HRMS), which is costly, time consuming and limits this analysis to highly specialized laboratories3,4. Building on previous work in this laboratory6,7 this presentation describes the characterization and validation of a rapid solvent free extraction of a small serum sample (0.25 ml), with processing time of 2-3 hours to analysis using solid phase extraction of analytes from matrix (Twister® stir bar sorptive extraction). This method involves solvent-less extraction of hydrophobic organics with a 10mm long and 0.5mm thick Twister® stir-bar coated with a non-polar polydimethylsiloxane (PDMS). The stir-bar is dropped into an aqueous matrix containing hydrophobic POPs and stirred for a period of time. Partitioning into the PDMS coating is related to the octanol-water partition coefficient for the analyte. Analysis by capillary gas chromatographic separation with electron capture negative ion mass selective detection (GC/ MSD-ECNI- SIM), produces quantitative results with detection limits fit for purpose using instrumentation more readily available to most analytical laboratories. Validation parameters demonstrated in this method include linearity, accuracy, detection limits, and precision (evaluating intra and inter assay variation). Figure 5 Figure 6 NIST Human Serum 1958 0.0000 0.2000 0.4000 0.6000 0.8000 1.0000 1.2000 PBDE-028 PBDE-047 PBDE-066 PBDE-100 PBDE-099 PBDE-085 PBDE-153 PBDE-154 PBDE-183 PBB-153 PPB NIST ± SD Measured Figure 7 Figure 8 This poster was supported by: The Michigan Department of Community Health with funding from the CDC Public Health Emergency Preparedness (PHEP) Cooperative Agreement # 2U90TP517018-11 An appointment to the Environmental Public Health (EPH) Fellowship Program (CJ) administered by the Association of Public Health Laboratories (APHL) and funded by the Centers for Disease Control and Prevention (CDC). GERSTEL Global Analytical Solutions with training of CJ and instrumental support during this project.

Upload: colin-daniel-johnson

Post on 17-Feb-2017

66 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: mdch lab poster  bonita3

PBDE Analysis in Human Blood Serum by Gerstel Twister© Stir-bar Sorptive Extraction Technique and GC/MS-ECNI-SIM

Quality Control Materials

• Calibratorsarepreparedinpre-screenedblankfetalbovineserum• Threemethodcontrolsareaddedtoeachbatchatblank,low,andhighlevels.TheseareBlank:

consistingofprescreenedFetalbovineserum,QL:NISTHumanSerum1957,andQH:NISTHumanSerum1958.TheNIST1958alsocontainednormalhumanpopulationrangesofchlorinatedpesticidesandchlorinatedbiphenylstodemonstratetheselectivityofthemethod.

• Aserumsurrogatecontainingknownconcentrationsofanalytessimilartothecompoundsofinterest,butnotcommonlyfoundinthesamplematrixtested(PBB-155)isspikedintoeverycalibrator,unknownandQCsampletestedinordertomonitorextractionefficiencyandanalyterecovery.

• Internalstandardconsistsofdifluoro-BDE_047(FBDE-47)andBDE156arespikedintoeachsample.

Stir-bar Sorptive Extraction (SBSE)

Extraction procedure is outlined in Figure 1. 0.25 ml of serum is diluted 1:1 with water and denatured using equal volume concentrated formic acid. Twister stir bars are added and allowed to stir at room temperature for 2 hours prior to analysis. Analytes are desorbed from the stirbar in a thermo-desorption unit (Gerstel TDU) and cryo-trapped using a Gerstel CIS4, see Figure 2. A Custom BDE Mixture (1ppm for each congener, AccuStandard Inc.) was serially diluted using isopropanol to prepare the spiking standard. Standard was spiked into screened fetal bovine serum to prepare a 6-point calibration curve (0.064 ppb to 3.20 ppb). Quality control materials run with each batch consisted of blank fetal bovine serum (QB), NIST Human serum SRM-1957 (QL) and Human serum SRM-1958(QH).

Contact Information

Authors: Colin Johnson, Paul R. Loconto, Michael O’Keefe, Bonita Taffe www.michigan.gov/mdchlab

“Providing Quality Laboratory Science for Healthier People and Communities”

Analysis

Stir Bar Conditioning/Cleaning

Results Conclusion

References

Acknowledgements

Standards and Limit of Detection

Methods

Michigan Department of Community Health - Bureau of Laboratories

3350 N. Martin Luther King Blvd. P.O. Box 30035

Lansing, MI 48909Office: (517)- 335-9490

E-Mail: [email protected]: [email protected]

• ThequantificationofPBDEsinhumanserumiscarriedoutusingcapillarygaschromatographic(C-GC)separationwithelectroncapturenegativeionmassselectivedetection(ECNI-SIM,whereSIMstandsforselectiveionmonitoring).Inthismethod,tenBDEcongenersandtwoPBBcongeners(seeFigure2)arebaselineseparatedandallbutBDE-183isquantitatedagainstdifluoro-BDE_047.BDE-183isquantitatedagainstBDE-156duetocomparableretentiontime.Thismethodologydetectselutinganalytesasbromineionsatm/z79,withtheabundancem/z81servingasaqualifierion.Alltargetedanalytepeaksmustmeetthequalifierioncriteria(abundanceofm/z81100%relativetotheabundanceofquantionm/z79).

• AMulti-PurposeSamplerSingleRail,SingleInjector(MPS)wasused[GerstelGmbH&Co.KG].TheMPSismountedatoptheGC-MSD(Agilent6890GC/5973NMSD).TheMPSisintegratedwithChemStation®[AgilentTechnologies,Inc.]andisprogrammedusingMaestrosoftware[GerstelGmbH&Co.KG].

• Thestir-barisplacedinaconditionedTwisterdesorptionlinerandautomaticallytransferredfromthetrayholdertotheThermalDesorptionUnit(TDU),seeFigure3.AnalytessorbedonTwisterarethermallydesorbedintheTDUandthenarecryotrappedontoaglasslinerinsideaCooledInletSystem(CIS).AnalytesarethenthermallydesorbedintheCISintoaDB-XLB(J&WandAgilentTechnologies,Inc.)capillarycolumn.

Background levels of BDE47 and BDE99 from dust are common problems with trace analysis for these compounds due to the ubiquitous use in building materials. Steps taken to lower background in this method includes special washing of all glassware with dust free storage and the following procedure to condition the stirbars:

• SoakTwisters(50:50MeOH:MeCl,2h)• Removeanddryundernitrogen• SoakinTolueneovernight• Removeanddryundernitrogen• PlaceTwistersinglasstubes,bakeat

300oCfor2hrsunderdryN2• RemoveTwistersandallowtocoolin

holder• Increasetemperatureonconditionerto

350oC,bakefor5minutesandremove

• Storeinclean,dustfreevialuntilnextuse.

• GoodseparationofcongenerswasachievedwiththismethodandbackgroundlevelswereacceptableasshowninFigure5

• Thismethodusesa6pointcalibrationcurveinserumforanalysisofBDEcongeners,R2≥0.99.QualitycontrollimitsusingNISTSRM1958forQHandareshowninFigure6

• TheLODforeachcongenermeasuredislessthan0.2ppb• NointerferencewithextractionofBDEsbytheSBSEwasdetectedfromotheranalytesinthe

NISTSRM1958(whichcontainsPCBs,pesticidesandotherorganochlorinemetabolites)• MeasuredanalyteswereaveragedandcomparedwithNISTreportedlevels,Figure7.AllBDEs

exceptBDE85werewithin2SDofreportedvalues;BDE85waswithin3SD.MeasuredvaluesforPBB153werelowerthanNISTreportedvalues.

• AccuracyandPrecisionweremeasuredusingreplicatesoftheArcticMonitoringandProficiencyProgram(AMAP)samplesaswellastheNISTSRMandcomparedtoproficiencyprogramparticipantresults.SeeFigure8

• Arapidanalyticalmethodwhichuses0.25mlofhumanserumhasbeenvalidatedforPBDEs

• ThismethodusesGerstelTwisterSBSEextraction.AprocedureforenhancedBDEbackgroundeliminationandcleanupforstirbarre-usehasbeendeveloped.

• Themethodhasbeendemonstratedtobebothaccurateandpreciseatconcentrationlevelsfoundinhumansubpopulations

• PBDE-47and99aregenerallydemonstratedinanalysis.Thisstudydemonstrateswaystoreducehighenvironmentalbackgroundlevelswithoutextrememeasures(cleanroompreparation).

• Thelimitofdetectionforthismethodforeachcongenerislessthan0.2ppb.Thismethodcouldbeusedforpopulationscreeningtoidentifyingmorehighlyexposedsubpopulations.

1. ATSDR.ToxicologicalProfileforPolybrominatedBiphenylsandPolybrominatedDiphenylEthers(PBBsandPBDEs);AgencyforToxicSubstanceandDiseaseRegistry:Atlanta,GA,2004.

2. BettsKS2008.UnwelcomeGuest:PBDEsinIndoorDust.EnvironHealthPerspect116:A202-A208.

3. ReportonHumanBiomonitoringofenvironmentalChemicalsinCanada,MeasuresSurveyCycle1(2007),HealthCanada/LaboratoriedeToxicologie,InstitutNationaldeSantePublique,Quebec,CA,www.healthcanada.gc.ca

4. SjödinA,WongLY,JonesRS,ParkA,ZhangY,HodgeC,DiPietroE,McClureC,TurnerW,NeedhamLLandPatterson,DG,“SerumConcentrationsofPolybrominatedDiphenylEthers(PBDEs)andPolybrominatedBiphenyl(PBB)intheUnitedStatesPopulation:2003–2004”,Environ.Sci.Technol.,2008,42(4),pp1377–1384.

5. Johnson,PI,Stapleton,HM,Sjodin,A,Meeker,JD,“RelationshipsbetweenPolybrominatedDiphenylEtherConcentrationsinHouseDustandSerum”,EnvironSciTechnol,44,5627-32,2010.

6. Loconto,P.R.,“Evaluationofautomatedstirbarsorptiveextraction-thermaldesorpton-gaschromatographyelectroncapturenegativeionmassspectrometryfortheanalysisofPBDEsandPBBsinsheepandhumanserum”JournalofChromatographicScience,47(2009)656-669.

7. Loconto,P.R.“Selectivityandsensitivityimprovementsforselectedpolybrominateddiphenylethersandpolybrominatedbiphenylsusingcapillarygaschromatography/electroncaptureneagativeionmassselectivedetection:acosteffectiveapproachtobiomonitoring”LC-GCNorthAmerica26(11)(2008)1118-1130.

8. J.K.Taylor.QualityAssuranceofChemicalMeasurements.LewisPublishers,BocaRaton,FL1987

IUPAC name Abbreviation LOD1 ppb

• 2,2’,4,4’,5,5’-Hexabromobiphenyl PBB-153 • 2,2’,4,4’,6,6’-Hexabromobiphenyl PBB-155 • 2,4,4’-Tribromodiphenyl Ether BDE_028 0.122 • 2,2’,4,4’-Tetrabromodiphenyl Ether BDE_047 0.165 • 2,3’,4,4’-Tetrabromodiphenyl Ether BDE_066 0.112 • 2,2’,3,4,4’-Pentabromodiphenyl Ether BDE_085 0.160 • 2,2’,4,4’,5-Pentabromodiphenyl Ether BDE_099 0.160 • 2,2’,4,4’,6-Pentabromodiphenyl Ether BDE_100 0.190 • 2,2’,4,4’,5,5’-Hexabromodiphenyl Ether BDE_153 0.182 • 2,2’,4,4’,5,6’-Hexabromodiphenyl Ether BDE_154 0.168 • 2,2’,3,4,4’,5’,6-Heptabromodiphenyl Ether BDE_183 0.192 • 5,5’-difluoro-2,2’,4,4’-tetrabromodiphenyl ether F-BDE_047, IS • 2,3,3’,4,4’,5-Hexabromodiphenyl ether BDE_156, IS

LOD was determined by calculating the standard deviation at each standard concentration following repeated measurements (n=15-20) of the lowest 3 concentration standards in serum. The absolute values of the standard deviations were then plotted versus concentration. The intercept of the least squares fit of this line equals So, with 3So being the LOD. J.K. Taylor. Quality Assurance of Chemical Measurements. Lewis Publishers, Boca Raton, FL 1987.

Figure 2

Sample Preparation for Twister Stir-bar Sorptive Extraction in Human Serum Matrix

Rinse a 5mL headspace vial 3x with dichloromethane.

Place a previously cleaned and conditioned Twister [10mm long and 0.5mm film thickness] into the headspace vial.

To 0.25mL of unspiked or spiked human serum is added 0.25mL of HPLC Grade H2O.

Then add 0.50mL of 5% acetonitrile in concentrated formic acid.

Stir at 500 rpm for exactly two hours then removed the Twister and

place in HPLC Grade H2O to remove biological material.

Dry the Twister with a Kim-Wipe or equivalent and place the Twister in a clean desorption liner.

Carefully place the transport adapter atop the desorption liner that

contains the Twister and place in the TDU tray holder.

Figure 1

Figure 3

Thermal Desorption Unit liner with Twister stirbar in place

Thermal Desorption Unit (TDU) on Agilent GC

Figure 4

Figure 6: QC CHARTS

F11-PBDEsPBDE_028_QH

Individ.: cl : 0.467143 ucl: 0.541331 lcl: 0.392955 * Rule violationSubgrp Size 1

O

O

O O

O

OO

O

O O

O

O

O

O

O

O

O

O

Individ.

cl

lcl

ucl

0.4

0.45

0.5

0.55

2 4 6 8 10 12 14 16 18 20

F11-PBDEsPBDE_047_QH

Individ.: cl : 0.609333 ucl: 0.921229 lcl: 0.297438 * Rule violationSubgrp Size 1

O

O OO O

O

O

O

O

O

OO

O

O

Individ.

cl

lcl

ucl

0.4

0.6

0.8

2 4 6 8 10 12 14 16 18 20

F11-PBDEsPBDE_100_QH

Individ.: cl : 0.335294 ucl: 0.407457 lcl: 0.263131 * Rule violationSubgrp Size 1

O

OO

O

OO

O

O

OO

OO

O

O O

O

O

Individ.

cl

lcl

ucl

0.25

0.3

0.35

0.4

2 4 6 8 10 12 14 16 18 20

F11-PBDEsPBDE_099_QH

Individ.: cl : 0.364444 ucl: 0.567712 lcl: 0.161177 * Rule violationSubgrp Size 1

O

O

O

O

O

O

O

O

O

O O

O O O

O

Individ.

cl

lcl

ucl

0.2

0.3

0.4

0.5

0.6

2 4 6 8 10 12 14 16 18 20

F11-PBDEsPBDE_154_QH

Individ.: cl : 0.268421 ucl: 0.398075 lcl: 0.138767 * Rule violationSubgrp Size 1

O

O

O

O

O

OO

O OO O

O

O O

OO

O

O

Individ.

cl

lcl

ucl

0.15

0.2

0.25

0.3

0.35

0.4

2 4 6 8 10 12 14 16 18 20

F11-PBDEsPBDE_153_QH

Individ.: cl : 0.298333 ucl: 0.413582 lcl: 0.183085 * Rule violationSubgrp Size 1

OO

O

O

O

O

O

O

O

O O O O

O

O

O

O

Individ.

cl

lcl

ucl

0.2

0.25

0.3

0.35

0.4

2 4 6 8 10 12 14 16 18 20

F11-PBDEsPBDE_183_QH

Individ.: cl : 0.38 ucl: 0.549917 lcl: 0.210083 * Rule violationSubgrp Size 1

O

O

O

O

O

OO

O

O

O

O

OO

O O

O

O

O

Individ.

cl

lcl

ucl

0.2

0.3

0.4

0.5

2 4 6 8 10 12 14 16 18 20

Figure 5 NIST 1958 (QH) Serum Chromatogram showing baseline separation of congeners.

Time, Minutes

Blank Fetal Bovine Serum spiked with FBDE-47 and BDE-156 internal standards. This figure illustrates a stable baseline, indicating a clean solid-phase on the stir-bar as well as a clean blank serum for calibrators.

BDE/BB Congener Retention Time (min) BDE-28 8.780 BDE-47 10.035 BDE-66 10.495 BDE-100 12.353 BDE-99 13.181 BDE-85 14.190 BB-153 15.504

BDE-153 16.659 BDE-183 19.383

4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00

500

1000

1500

2000

2500

3000

3500

4.369

9.813

11.182

20.923 Abu

ndan

ce

FBDE-47

BDE-156

Time, Minutes 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00

500 1000 1500 2000 2500 3000 3500 4000 4500 5000 5500 6000 6500 7000 7500 8000 8500

3.676

3.730

3.839

3.909

3.952

4.024

4.222

4.310

4.349 4.457

4.528

5.251

5.388

5.485

6.491

7.150 7.547

8.780

9.832

10.035

10.495 12.220

12.353

13.181

14.910 15.232 15.504

16.659 19.383

Abu

ndan

ce

Figure 8: AMAP Recovery

AMAP W0906

0.00

0.10

0.20

0.30

0.40

0.50

0.60

0.70

BDE-28

BDE-47

BDE-100

BDE-99

BDE-154

BDE-153

BDE-183

PPB

Inter-LabTwister

AMAP W0909

0.00

0.10

0.20

0.30

0.40

0.50

0.60

0.70

0.80

0.90

BDE-28

BDE-47

BDE-100

BDE-99

BDE-154

BDE-153

BDE-183

PPB

Inter-LabTwister

AMAP W1004

0.00

0.05

0.10

0.15

0.20

0.25

0.30

0.35

0.40

0.45

BDE-28

BDE-47

BDE-100

BDE-99

BDE-154

BDE-153

BDE-183

PPB

Inter-LabTwister

AMAP W1005

0.00

0.20

0.40

0.60

0.80

1.00

1.20

1.40

BDE-28

BDE-47

BDE-100

BDE-99

BDE-154

BDE-153

BDE-183

PPB

Inter-LabTwister

AMAP W1007

0.00

0.10

0.20

0.30

0.40

0.50

0.60

BDE-28

BDE-47

BDE-100

BDE-99

BDE-154

BDE-153

BDE-183

PPB

Inter-LabTwister

AMAP W1008

0.00

0.05

0.10

0.15

0.20

0.25

0.30

0.35

0.40

0.45

BDE-28

BDE-47

BDE-100

BDE-99

BDE-154

BDE-153

BDE-183

PPB

Inter-LabTwister

Abstract

Introduction

Polybrominated diphenyl ethers (PBDEs), widely used fire retardants, are currently under investigation as agents interfering with immunological, embryonic development and endocrine functions. Low population background levels have favored analysis by high resolution mass spectrometry (HRMS), limiting this analysis to highly specialized laboratories. This presentation describes a rapid solvent free extraction of PBDEs from a small serum sample (0.25 ml), with processing time of 2-3 hours using Twister® stirbar sorptive extraction. Gas chromatographic separation with electron capture negative ion mass selective detection (GC/ MSD-ECNI-SIM) produces quantitative results with detection limits fit for purpose using instrumentation readily available in most analytical laboratories.

Polybrominated diphenyl ethers (PBDEs) have been widely used as fire retardants in many consumer items including polyurethane foam found in furniture and carpeting, home electronics and other textiles. In 2004 Penta-BDEs and Octa-BDEs were banned by the European Union and US production was phased out at this time, however items containing these chemicals persist. The BDEs consist of 10 homolog groups (1 to 10 bromine atoms) with 209 possible congeners. BDEs are not chemically bound to products and can be mobilized by leaching or degradation of treated materials over time.

Concentrations of BDEs in indoor air and dust have been shown to be significantly higher than outdoor concentrations. A recent study of the relationship between house dust and serum concentrations of BDEs in adults shows high correlation5. Young children are considered to be at highest risk of exposure from house dust due to their full body contact with the floor and carpets and hand to mouth activity. BDEs are classed as potential endocrine disruptors and the mechanism of action is currently under investigation in animal models with respect to immunological, embryonic development and endocrine functions1,2. Further epidemiological research to evaluate human response to these agents is needed.

Low general population background levels and background environmental contamination by these analytes has favored quantitative analysis by high resolution mass spectrometry (HRMS), which is costly, time consuming and limits this analysis to highly specialized laboratories3,4. Building on previous work in this laboratory6,7 this presentation describes the characterization and validation of a rapid solvent free extraction of a small serum sample (0.25 ml), with processing time of 2-3 hours to analysis using solid phase extraction of analytes from matrix (Twister® stir bar sorptive extraction). This method involves solvent-less extraction of hydrophobic organics with a 10mm long and 0.5mm thick Twister® stir-bar coated with a non-polar polydimethylsiloxane (PDMS). The stir-bar is dropped into an aqueous matrix containing hydrophobic POPs and stirred for a period of time. Partitioning into the PDMS coating is related to the octanol-water partition coefficient for the analyte.

Analysis by capillary gas chromatographic separation with electron capture negative ion mass selective detection (GC/ MSD-ECNI-SIM), produces quantitative results with detection limits fit for purpose using instrumentation more readily available to most analytical laboratories. Validation parameters demonstrated in this method include linearity, accuracy, detection limits, and precision (evaluating intra and inter assay variation).

Figure 5

Figure 6

NIST Human Serum 1958

0.0000

0.2000

0.4000

0.6000

0.8000

1.0000

1.2000

PBDE-028

PBDE-047

PBDE-066

PBDE-100

PBDE-099

PBDE-085

PBDE-153

PBDE-154

PBDE-183

PBB-153

PPB

NIST ± SDMeasured

Figure 7

Figure 8

Thisposterwassupportedby:• TheMichiganDepartmentofCommunityHealthwithfundingfrom

theCDCPublicHealthEmergencyPreparedness(PHEP)CooperativeAgreement#2U90TP517018-11

• AnappointmenttotheEnvironmentalPublicHealth(EPH)FellowshipProgram(CJ)administeredbytheAssociationofPublicHealthLaboratories(APHL)andfundedbytheCentersforDiseaseControlandPrevention(CDC).

• GERSTELGlobalAnalyticalSolutionswithtrainingofCJandinstrumentalsupportduringthisproject.