mc_961b

Upload: naranjor

Post on 02-Apr-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/27/2019 MC_961B

    1/9

    ISA-MC96.1 3469

    DO NOT PHOTOCOPY - SEE PRINTING RESTRICTIONS IN "ABOUT THIS LIBRARY"

    2.2.6 Thermocouple assembly

    A thermocouple assembly is an assembly consisting of a thermocouple element and one or moreassociated parts such as terminal block, connection head, and protecting tube.

    1) Terminal block: A terminal block is a block of insulating material that is used to supportand join the terminations of conductors. (See Figure 2.)

    Figure 2 Thermocouple element with terminal block

    2) Connection head: A connection head is a housing enclosing a terminal block for anelectrical temperature sensing device and usually provided with threaded openings forattachment to a protecting tube and for attachment of conduit. (See Figures 3 and 4.)

    .

    Figure 3 Thermocouple element with connection head

    Figure 4 Connection head

    3) Connection head extension: A connection head extension is a threaded fitting or anassembly of fittings extending between the thermowell or angle fitting and the connectionhead.

    The connection head extension length is the overall length of the connectionhead extension and is assigned the symbol N. (See Figure 11.)

    4) Protecting tube: A protecting tube is a tube designed to enclose a temperature sensingdevice and protect it from the deleterious effects of the environment. It may provide forattachment to a connection head but is not primarily designed for pressure-tightattachment to a vessel. A bushing or flange may be provided for the attachment of aprotecting tube to a vessel. (See Figures 5, 6, 7, and 8.)

    TABLE OF CONTENTS PREVIOUS PAGE

    http://about.pdf/http://mc_961a.pdf/http://mc_961a.pdf/http://mc_961a.pdf/http://about.pdf/
  • 7/27/2019 MC_961B

    2/9

    3470 ISA-MC96.1

    DO NOT PHOTOCOPY - SEE PRINTING RESTRICTIONS IN "ABOUT THIS LIBRARY"

    The protecting tube length is the overall length of a protecting tube and is as-signed the symbol P. (See Figure 5.)

    Figure 5 Protecting tube

    Figure 6 Protecting tube with mounting bushing

    Figure 7 Protecting tube with mounting flange

    Figure 8 Thermocouple element with protecting tube and connection head

    The protecting tube diameter is the outside diameter of a protecting tube and is as-signed the symbol M.

    A protecting tube has one end closed unless it is specified as open end. (See Figure 9.)

    Figure 9 Open end protecting tube

  • 7/27/2019 MC_961B

    3/9

    ISA-MC96.1 3471

    DO NOT PHOTOCOPY - SEE PRINTING RESTRICTIONS IN "ABOUT THIS LIBRARY"

    5) Thermowell: A thermowell is a pressure-tight receptacle adapted to receive a temperaturesensing element and provided with external threads or other means for pressure-tightattachment to a vessel.

    A lagging extension is that portion of a thermowell above the threads, intended toextend through the lagging of a vessel. The lagging extension length is the lengthfrom the lower end of the external threads of the well to the outer end of the por-tion intended to extend through the lagging of a vessel, less one inch allowance

    for threads, and is assigned the symbol T. (See Figure10.)

    Figure 10 Well

    Figure 11 Thermocouple assembly with thermowell

    The immersion length of a thermowell, protecting tube, or thermocouple element is the length fromthe free end to the point of immersion in the medium which is being measured and is assigned thesymbol R. (See Figure 12.)

    The insertion length of a thermowell, protecting tube or thermocouple element is the length from thefree end to, but not including, the external threads or other means of attachment to a vessel and isassigned the symbol U. (See Figures 10 and 12.)

    http://about.pdf/http://about.pdf/
  • 7/27/2019 MC_961B

    4/9

    3472 ISA-MC96.1

    DO NOT PHOTOCOPY - SEE PRINTING RESTRICTIONS IN "ABOUT THIS LIBRARY"

    .

    Figure 12 Immersion and insertion lengths for thermocoupleassembly with thermowell

    2.2.7 Angle type thermocouple assembly

    An angle type thermocouple assembly is an assembly consisting of a thermocouple element, aprotecting tube, an angle fitting, a connection head extension, and a connection head.

    2.2.8 Other forms of thermocouples and thermocouple elements

    1) Coaxial thermocouple element: A coaxial thermocouple element consists of athermoelement in wire form within a thermoelement in tube form and electrically insulated

    from the tube except at the measuring junction.

    2) Sheathed thermocouple: A sheathed thermocouple is a thermocouple having itsthermoelements, and sometimes its measuring junction, embedded in mineral oxideinsulation compacted within a metal protecting tube.

  • 7/27/2019 MC_961B

    5/9

    ISA-MC96.1 3473

    DO NOT PHOTOCOPY - SEE PRINTING RESTRICTIONS IN "ABOUT THIS LIBRARY"

    Symbols

    2.3 Wire sizes

    2.3.1 Thermocouples

    The wire sizes normally used for non-sheathed thermocouples are as follows:

    For J, K, and E: 8, 14, 20, 24, and 28 AWG*

    For T: 14, 20, 24, and 28 AWG*

    For R, S and B: 24 AWG* only

    2.3.2 Extension wires

    The wire sizes normally used for extension wire, either singly or in pairs, are 14, 16, and 20 AWG*.Sixteen (16) gage is most commonly used. Twenty (20) gage and smaller may be used when

    bundled and reinforced to provide strength for pulling. These sizes apply to all types of extensionwires.

    2.4 Upper temperature limits

    Table 7 gives the recommended upper temperature limits for the various thermocouples and wiresizes. These limits apply to protected thermocouples in conventional closed-end protecting tubes.They do not apply to sheathed thermocouples having compacted mineral oxide insulation.

    In any general recommendation of thermocouple temperature limits, it is not practical to take intoaccount special cases. In actual operation, there may be instances where the temperature limitsrecommended can be exceeded. Likewise, there may be applications where satisfactory life will notbe obtained at the recommended temperature limits. However, in general, the temperature limits

    listed are such as to provide satisfactory thermocouple life when the wires are operated continuouslyat these temperatures. Various factors affecting thermocouple life are discussed in Appendix C.

    *American Wire Gage, also known as B&S (Brown & Sharpe)

    Symbols

    Thermocouple element length A

    Thermocouple element diameter Y

    Connection head extension length N

    Protecting tube length P

    Protecting tube diameter M

    Lagging extension length T

    Immersion length R

    Insertion length U

    http://about.pdf/http://mc_961e.pdf/http://mc_961e.pdf/http://about.pdf/
  • 7/27/2019 MC_961B

    6/9

    3474 ISA-MC96.1

    DO NOT PHOTOCOPY - SEE PRINTING RESTRICTIONS IN "ABOUT THIS LIBRARY"

    Table 7 Recommended upper temperature limits for protected thermocouples. Uppertemperature limit for various wire sizes (AWG), Deg C

    2.5 Tolerance of initial calibration

    Tables 8, 9 and 10 give the standard and special tolerance of initial calibration for thermocouples andthermocouple extension wires. The tolerance of initial calibration is defined as the allowabledeviation of the thermocouple and extension wire in its initial condition as supplied by the

    manufacturer from the standard EMF-temperature tables. Once the thermocouple is in use itscalibration will change. The magnitude and direction of the change are dependent on temperature,time and environmental conditions affecting the thermocouple and may not be accurately predicted.The tolerances for each type of thermocouple apply only over the temperature range for which thewire size in question is recommended (see Table 7). These tolerances should be applied only tostandard wire sizes. The same tolerances may not be obtainable in special sizes. These tolerancesdo not include installation or system errors. See Appendix C, paragraph C4.1 for the thermocoupleinstallations and errors.

    Where tolerances are given in percent, in Table 8, the percentage applies to the temperature beingmeasured. For example, the standard tolerance of Type J over the temperature range 277 to 760Cis +3/4 percent. If the temperature being measured is 538C, the tolerance is +3/4 percent of 538, or+4.0C. To determine the tolerance in degrees Fahrenheit, multiply the tolerance in degrees Celsiustimes 1.8.

    Thermocouple

    Type

    No. 8 Gage No. 14 Gage No. 20 Gage No. 24 Gage No. 28 Gage

    B 1700

    E 870 650 540 430 430

    J 760 590 480 370 370

    K 1260 1090 980 870 870

    R & S 1480

    T 370 260 200 200

    http://mc_961e.pdf/http://mc_961e.pdf/
  • 7/27/2019 MC_961B

    7/9

    ISA-MC96.1 3475

    DO NOT PHOTOCOPY - SEE PRINTING RESTRICTIONS IN "ABOUT THIS LIBRARY"

    Table 8 Initial calibration tolerances for thermocouples

    Table 9 Initial calibration tolerances for thermocouple extension wires

    *Thermocouples and thermocouple material are normally supplied to meet the tolerances specified in the table

    for the normal specified range. The same materials, however, may not fall within the cryogenic tolerances in the

    second section of the table. If materials are required to meet the cryogenic tolerances, the purchase order must

    so state. Selection of materials usually will be required. Tolerances indicated in this table are not necessarily an

    indication of the accuracy of temperature measurements in use after initial heating of the materials.

    **Little information is available to justify establishing special tolerances for cryogenic temperatures. Limited

    experience suggests the following tolerances for Types E and T thermocouples:Type E 200 to 0C +1C or +0.5% (whichever is greater)

    Type T 200 to 0C +0.5C or +0.8% (whichever is greater)

    These tolerances are given only as guide for discussion between purchaser and supplier. Due to the characteris-

    tics of the materials, cryogenic tolerances for Type J thermocouples and special cryogenic tolerances for Type K

    thermocouples are not listed.

    Reference Junction 0C

    Thermocouple

    Type

    Temperature

    Range, C

    TOLERANCES

    Standard

    (whichever is greater)

    Special

    (whichever is greater)

    B 870 to 1700 +0.5%

    E 0 to 900 + 1.7C or +0.5% +1C or +0.4%

    J 0 to 750 +2.2C or +0.75% +1.1C or +0.4%

    K 0 to 1250 +2.2C or +0.75% +1.1C or +0.4%

    R or S 0 to 1450 +1.5C or +0.25% +0.6C or +0.1%

    T 0 to 350 +1C or +0.75% +0.5C or +0.4%

    Cryogenic Ranges

    E* 200 to 0 +1.7C or +1% **

    K* 200 to 0 +2.2C or +2% **

    T* 200 to 0 +1C or +1.5% **

    Reference Junction 0C

    Extension Wire Type Temperature Range, C

    Tolerances

    Standard Special

    EX 0 to 200C 1.7C -

    JX 0 to 200C 2.2C 1.1C

    KX 0 to 200C 2.2C -

    TX 0 to 100C 1.0C 0.5C

    http://about.pdf/http://about.pdf/
  • 7/27/2019 MC_961B

    8/9

    3476 ISA-MC96.1

    DO NOT PHOTOCOPY - SEE PRINTING RESTRICTIONS IN "ABOUT THIS LIBRARY"

    3 Non-ceramic insulation of thermocouple and extension wires

    The normal function of thermocouple and extension wire insulation is to provide electrical insulation.If this function is not provided or is compromised in any way, the indicated temperature may be inerror. Insulation of this type (non-ceramic) may be affected adversely by moisture, abrasion, flexing,

    temperature extremes, chemical attack, and nuclear radiation. Each type of insulation has its ownlimitations. A knowledge of these limitations is essential if accurate and reliable measurements are tobe made.

    A number of coatings are presently available commercially. The strong points as well as limitationsare discussed in ASTM Special Technical Publication STP-470B, "Manual on the Use ofThermocouples in Temperature Measurement."

    In summary, this type of insulation should be selected only after considering possible exposuretemperatures and heating rates, the number of temperature cycles, mechanical movement,moisture, routing of the insulated wire, and chemical deterioration.

    Table 10 Initial calibration tolerances for thermocouple compensating extension wire

    4 Temperature-EMF tables for thermocouples

    4.1 Scope and purpose

    This section applies to the temperature-EMF relationships of materials used for temperaturemeasurement thermocouples.

    Its purpose is to provide reference tables of temperature-EMF values for Type B, E, J, K, R, S, and Tthermocouples, in a form convenient for industrial and laboratory use.

    *Due to the non-linearity of the Types, R, S, and B temperature-EMF curves, the error introduced into athermocouple system by the compensating wire will be variable when expressed in degrees. The

    degree C tolerances given in parentheses are based on the following measuring junction

    temperatures:

    Type Wire Measuring Junction Temperature

    BX Greater than 1000C

    SX Greater than 870C

    **Copper (+) versus copper nickel alloy ().

    ***Copper versus copper compensating extension wire, usable to 100C with maximum errors as

    indicated, but with no significant error over 0 to 50C range. Matched proprietary alloy compensating

    wire is available for use over the range 0 to 200C with claimed tolerances of +0.033 mV. (+3.7C*).

    Reference Junction 0C

    ThermocoupleType

    CompensatingWire Type

    TemperatureRange, C

    Tolerances

    B BX*** 0 to 100

    R, S SX** 0 to 200 + .057 mV (+5C)*

    +0.000 (+0C)*

    mV

    0.033 (-3.7C)*{ }

  • 7/27/2019 MC_961B

    9/9

    ISA-MC96.1 3477

    DO NOT PHOTOCOPY - SEE PRINTING RESTRICTIONS IN "ABOUT THIS LIBRARY"

    4.2 Introduction

    The values in these tables are based upon the International Practical Temperature Scale of 1968(IPTS-68) and the U.S. legal electrical units. All the data in Tables 11 to 17 have been extracted from"Thermocouple Reference Tables Based on the IPTS-68," National Bureau of Standards Monograph125. These tables differ slightly from previous tables for the following reasons: improvedmeasurements and data analysis techniques, slight changes in commercial thermocouple materials,and also changes in the temperature scale and electrical units. The significance of these factors, as

    well as the origin of each of the tables, is discussed in the NBS reference noted above, and it shouldbe consulted for details.

    These tables give values of EMF to three decimal places (0.001 mV) for one degree Celsius (C)temperature intervals. If greater precision is required, the NBS reference noted above should beconsulted. It includes tables giving values of EMF to four decimal places (0.0001 mV) and analyticalfunctions for each thermocouple type that allow a direct and precise calculation of the EMF-temperature relationship.

    Tables for each type of thermocouple giving values of EMF as a function of temperature in degreesFahrenheit (F) can be found in ANSI/ASTM Standard E230, "Temperature-Electromotive Force(EMF) Tables for Thermocouples." Tables giving EMF-temperature values (in both C and F) forsingle-leg thermoelements referenced to platinum (NBS Pt-67) are also given in the above ANSI/ASTM standard.

    4.3 Use of temperature-EMF tables

    These temperature-EMF reference tables serve two very useful purposes in that they provide ameans for converting the generated EMF of certain thermocouple material combinations intoequivalent temperatures, and they enable the calibration and checking of thermocouples andthermocouple extension wire.

    If the reference junction is maintained at 0C, the appropriate temperature or EMF data may be readdirectly from the tables. When it is not practical to maintain the reference junction temperature at0C, these tables may still be used by applying an appropriate correction. The value of thecorrection may be obtained from these tables. An example to illustrate how to obtain and apply thiscorrection follows.

    Let us suppose a Type J thermocouple was used in an installation to determine the temperature of afluid medium and an EMF output of 18.070 mV was observed. Also, a mercury thermometer in closeproximity to the thermocouple reference junction produced a reading of 20C.

    To use the Type J Table to obtain a value for the temperature of the fluid medium, the observed EMFoutput of the thermocouple must first be corrected to compensate for the difference between thereference junction temperature actually used and 0C. The correction is the EMF value given by theType J Table at the reference junction temperature actually used (20C). As shown below, this EMFvalue (1.019 mV) is algebraically added to the observed EMF output to obtain the value of EMF thatthe thermocouple would produce if the reference junction were at 0C.

    Observed Type J Thermocouple Output: 18.070 mVCorrection Factor (Table value at reference

    temperature actually used) for

    reference junction at 20C: 1.019 mv

    Corrected Output: 19.089 mV

    The corrected output of 19.089 mV is then used to determine from the Type J Table the equivalenttemperature value of 350C.

    NEXT PAGE

    http://about.pdf/http://mc_961c.pdf/http://mc_961c.pdf/http://mc_961c.pdf/http://mc_961c.pdf/http://about.pdf/