Mass Spectrometry and Proteomics- Lecture 3 -

Matthias TrostNewcastle University

[email protected]

Page 2: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Previously:

68

• Mass analysers: • Quadrupole• Ion Traps• TOF• Orbitrap• FT-ICR

• MS/MS experiments• Parent/Product Ion Scan• Neutral Loss Scan• Precursor Ion Scan• Selected Reaction Monitoring

Page 3: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Lecture 3

• Peptide fragmentation• Fragmentation Techniques• Hybrid Instruments

69

Page 4: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

70

Proton affinities of amino acids

• Not all peptides have same Proton affinity – basic residues

Page 5: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Gas phase basicity affects ionisation efficiency in MALDI

71

Brancia et al, RCMS, 2000

Equimolar mixture of AFLDASKAFLDASR

Page 6: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Ion suppression• Ion suppression describes the adverse effect on detector

response due to reduced ionisation efficiency for analyte(s) of interest, resulting from the presence of species in the sample matrix which compete for ionisation or inhibit efficient ionisation in other ways.

• Solution: de-complexify the sample.

72

Page 7: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

73

Fragmentation techniques• Collision-Induced Dissociation (CID) (also called

Collisionally-activated dissociation (CAD)• Electron-Transfer Dissociation (ETD)• Higher-energy Collision (C-trap) Dissociation (HCD)

not covered:• Pulsed-Q (Collision induced) Dissociation (PQD)• Electron-capture dissociation (ECD)• Infrared Multiphoton Dissociation (IRMPD)• Blackbody Infrared Radiative Dissociation (BIRD)

Page 8: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

74

Peptide Fragmentation in MS/MS

Page 9: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

75

Mobile Proton Model

http://www.lamondlab.com/MSResource/LCMS/MassSpectrometry/mobileProtonModel.php

• N-terminal proton can locate to amide-bonds and destabilise this bond.

• during collision-induced activation, these bonds break

Page 10: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Amino acid masses

76

Page 11: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Collision-induced Dissociation (CID)

• CID is realised by passing an ion beam (>1 keV) through a collision cell with a partial pressure of collision gas (He, N2, Ar).

• Very fast process: ~10-15 s.• Leads to vibrational excitation and preferred breakage of

weakest bond. • Due to mobile protons, backbone can fragment at

various positions.• Results preferentially in b & y ions. • Weak bonds such as phosphorylation and glycosylations

break first.

77

Page 12: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

CID fragmentation: formation of b & y-ions

78

Paizs et al, 2005

Page 13: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Fragmentation of a peptide

Page 14: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

P E P T I D E Kb-ions

b-ions

m/z

inte

nsity

Page 15: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

K E D I T P E Py-ions

y-ions

m/z

inte

nsity

Page 16: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

K E D I T P E P-H2O

m/z

Some ions lose H2O, CO or NH3in

tens

ity

Page 17: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Theoretical MS/MS spectrum

m/z

inte

nsity

Page 18: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Fragment intensities vary

m/z

inte

nsity

Page 19: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

noise

Noise & co-fragmenting peptides

m/z

inte

nsity

Page 20: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Fragmentation of a peptide

m/z

Page 21: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Identifying the mass differences between peaks that correspond to Amino Acids

m/z

inte

nsity

K

E

D

I

T

P

E P

Page 22: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

88

CID Peptide Fragmentation in MS/MS

Rioli et al, 2003

Page 23: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Electron Transfer Dissociation (ETD)

• “Stable” radical anions are obtained by electron capture of fluoranthene or anthracene.

• The radical anion reagent reacts with peptides very fast in a nonergodic manner (so fast that there is no internal equilibration of energy prior to dissociation).

• Results predominately in c & z ions. • Leaves weak bonds intact.

89

Page 24: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

ETD mechanism

90

Page 25: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Electron Transfer Dissociation (ETD)

• “Stable” radical anions are obtained by electron capture of fluoranthene or anthracene.

• The radical anion reagent reacts with peptides very fast in a nonergodic manner (so fast that there is no internal equilibration of energy prior to dissociation).

• Results predominately in c & z ions. • Leaves weak bonds intact.• Low fragmentation efficiency (~25%), thus lower sensitivity.• Works best for higher charges ≥3+.• Works well on small intact proteins (Top-down proteomics).

91

Page 26: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Higher-energy Collision Dissociation (HCD)

• Originally only possible in Orbitrap-type instruments. • Now also available for ion traps.• Ions are transferred into a HCD cell in which they

undergo few, but higher-energy collisions with inert gas. These lead to slightly different fragmentations than CID.

• As the ions are produced in a separate cell, low m/z ions are not lost like CID in ion traps.

• Produces preferentially b & y ions, good for isobarically tagged peptides.

• HCD in Orbitraps leads to high resolution MS/MS spectra.

92

Page 27: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

93

CID/HCD Peptide Fragmentation in MS/MS

MS/MS spectra resulting from either iHCD (A), CAD (B), PQD (C), or HCD (D) of triply protonated angiotensin cations.

McAlister G C et al. Mol Cell Proteomics 2011;10:O111.009456

Page 28: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

New Fragmentation techniquesRecently developed were• EThcD (ETD +HCD at the same time, leading to c,z,b &y

ions) (Frese et al, 2013).

• iHCD: ion trap HCD (McAlister et al, 2011)

94

Page 29: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Mass spectrometry traces

95

• Ion Chromatogram: shows intensity of all ions detected by mass spectrometer in time (A).

• Base peak chromatogram: shows the intensity of the highest peak at any time over the chromatogram.

• MS spectrum/precursor ion scan: shows the masses detected at a certain moment of the chromatogram (B).

• MS/MS spectrum: Dependent scan –fragmentation of an ion detected in B leading to a spectrum from which the peptide sequence can be derived (C).

Steen and Mann, 2004

Page 30: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

96

Hybrid Instruments

• Triple Quadrupole (QqQ)• TOF/TOF• Q-TOF• TIMS-TOF• IT/Orbitrap (Orbitrap Elite, Orbitrap Velos)• Q-Orbitrap (Q Exactive)• Q/IT/Orbitrap (Orbitrap Fusion Tribrid)• IT-FT-ICR

Page 31: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

97

Triple Quadrupole (QqQ)• Simplest MS/MS

capable instrument.• Low resolution

(<5000).• High linear dynamic

range.• Relatively sensitive. • Mostly used for

SRM type experiments.

• £

Page 32: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

98

TOF/TOF• Only really used

in MALDI instruments.

• <30k resolution. • Good for intact

proteins.• Only 2 high-end

vendors (Bruker and ABSciex)

• £££

Page 33: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

99

Q-TOF• Various vendors.• <30k resolution• Up to 30 Hz MS/MS.• High resolution

MS/MS.• Can be coupled with

ETD or Ion mobility (Waters) or SWATH (ABSciex).

• ££-£££

Page 34: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

TIMS QuadrupoleFilter

Time of FlightMS

Mass scan

110 µs

9.1 kHz

Mass Selection

2.5 – 10 ms

100 – 350 precursors/s

Ion Mobility Separation

20 - 100 ms

10 - 50 Hz8

Trapped Ion Mobility Spectrometry (TIMS) TOF

• <50,000 resolution• Utilises ion mobility• Very sensitive• Very fast MS/MS• £££

Page 35: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

timsTOF Pro

Extremely high speed @ high resolution

Usage of all ions (100% duty cycle)

High sensitivity

Page 36: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

TIMS QuadrupoleFilter

Time of FlightMSHPLC

Mass scan

110 µs

9.1 kHz

Mass Selection

2.5 – 10 ms

100 – 350 precursors/s

Ion Mobility Separation

20 - 100 ms

10 - 50 Hz

Chromatography

FWHM~ 7 s

8

TIMS design

Additional benefit: improved robustness due to orthogonal ESIand long TIMS tunnel

Page 37: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Dual TIMS analyzers enable a 100% duty cycle 

to massanalyzer

Entrance Funnel Analyzer 2 Exit FunnelCap.Exit

Def

lect

ion

Plat

e

Gas

Ions

from

spr

aych

ambe

r

Gas

Accumulate Trap Elute

E

Z

Analyzer 1

Parallel accumulation capability

9

Page 38: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Parallel Accumulation Serial Fragmentation (PASEF)

accumulation(100 ms)

fromSource

12

Page 39: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

transfer(2 ms)

13

Parallel Accumulation Serial Fragmentation (PASEF)

Page 40: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

parallelaccumulation

m/z

TIMS scan(100 ms)

TIMS MS Heat Map1/k0[Vs/cm2]

14

TIMS MS Scan

Page 41: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

parallelaccumulation

m/z

TIMS scan

precursor selection1/k0[Vs/cm2]

15

TIMS MS Scan

Page 42: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

PASEF MS/MS Scan

parallelaccumulation PASEF scan

PASEF MS/MS Heat Map

isolation fragmentation(~3 ms / precursor)

m/z

TIMS MS Heat Map1/k0[Vs/cm2]

1/k0[Vs/cm2]

16

Page 43: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

TIMS ‐ PASEF alignment

1/k0[Vs/cm2]

PASEF MS/MS Heat Map

m/z

TIMS MS Heat Map

m/z

1/k0[Vs/cm2]

17

Page 44: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

1.0

1.2

1.4

Mobility1/K0

200 400 600 800 1000 1200 1400 1600 m/z

PASEF MS/MS Heat Map

1.0

1.2

1.4

Mobility1/K0

200 400 600 800 1000 1200 1400 1600 m/z

TIMS MS Heat Map

PASEF MS/MS Heat Map

m/z

TIMS MS Heat Map1/k0[Vs/cm2]

1/k0[Vs/cm2]

18

TIMS ‐ PASEF alignment

Page 45: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

111

IT/Orbitrap (Orbitrap Elite)

• Linear Ion trap/ Orbitrap configuration.• Very fast. 15 Hz CID, 10Hz HCD MS/MS• Up to 500k resolution in MS.• Can have ETD• £££££

Page 46: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

112

Q/Orbitrap (Q Exactive)

• Quadrupol/Orbitrap combination.

• Very fast (~10Hz MS/MS).• MS resolution <240,000• All MS/MS in HCD.• SRM-type experiments:

Product Reaction Monitoring (PRM)

• ££-£££

Page 47: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

113

Q/IT/Orbitrap (Fusion Tribrid)• Q/IT/Orbitrap

combination.• Most sensitive

Orbitrap.• Very fast (~15Hz

MS/MS).• MS resolution

500,000• ETD, CID, HCD

capable.• Very flexible

experiments possible.• £££££

Page 48: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

IT-FT-ICR• Highest resolution

MS (<106)• Expensive for

purchase and in usage.

• Requires superconducting magnet.

• Up to 12 Tesla.• Allows gas phase

reactions in ICR.• ££££££

114

Page 49: Mass Spectrometry and Proteomics - Lecture 3€¦ · Mass Spectrometry and Proteomics - Lecture 3 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk. Previously: 68 •

Comparison of instruments

115

Quadrupole QIT LIT TOF Orbitrap FT-ICR

Mass Range 4000 4000 4000 Theor. unlimited

4000 >104

Resolving Power

<5000 <10K <10K <30k-50k <500k >106

Mass accuracy (ppm)

100 50 50 5-20 1-5 1-5

Dynamic Range 107 <105 <105 <106 <105 <105

Cost £ £ ££ ££-£££ ££-££££ £££££