manual klimet a30 englisch v1_30
TRANSCRIPT
-
8/11/2019 Manual Klimet A30 Englisch V1_30
1/14
Hofstrasse 92 Tel +41 1 934 40 40CH-8620 Wetzikon 1 Fax +41 1 934 40 99Internet: www.meteolabor.ch E-Mail: [email protected]
Operating instructionsand interface description
KLIMET A30
2. November 2004Software Ver 1.30
-
8/11/2019 Manual Klimet A30 Englisch V1_30
2/14
KLIMET A30 Operating instructions Page 2 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
Contents1 Transport, storage ..........................................................................................................................................32 Installation, setting up ....................................................................................................................................33 Technical data .................................................................................................................................................3
3.1 Installed loads...........................................................................................................................................................33.2 General ...................................................................................................................................................................... 33.3
Measurement cycle...................................................................................................................................................3
3.4 Standard deviation....................................................................................................................................................43.4.1 Air temperature.............................................. ..................................................... ............................................. 43.4.2 Temperature differences..................................................................................................................................43.4.3 Dew point temperature.....................................................................................................................................43.4.4 Relative humidity..............................................................................................................................................43.4.5 Atmospheric pressure................................................ ..................................................... ................................. 4
4 Maintenance.....................................................................................................................................................55 Indication of special operating conditions...................................................................................................56 Calibration........................................................................................................................................................5
6.1 Calibration of the barometer KLA30-P35 ................................................................................................................ 5Command desciption of barometer KLA30 P35.......................................................... ............................................... 5Simple offset adjustment.............................................................................................................................................6Offset and span adjustment ................................................... ................................................................ ..................... 7
6.2
Calibration of the barometer KLA30-P (Serial Numbers 0001.. 0021)...................................................................7
Simple offset adjustment.............................................................................................................................................7Offset and span adjustment ................................................... ................................................................ ..................... 7
6.3 Calibration of the thermometer................................................................................................................................76.4 Calibration of the dew-point temperature...............................................................................................................8
7 Data format ......................................................................................................................................................87.1 General ...................................................................................................................................................................... 87.2 Data communication cable ......................................................................................................................................8
8 Command set...................................................................................................................................................98.1 Standard commands ................................................................................................................................................ 9
8.1.1 AQ, measured value query ................................................ ................................................................... ...........98.1.2 AP, atmospheric pressure......................................... ........................................................ ............................... 98.1.3 AT, temperature....... ........................................................ .................................................................... ............98.1.4 AU, dew point temperature .................................................. ................................................................. ...........98.1.5
AZ, set/read time................................................................................... ......................................................... 10
8.1.6 AI, identification................................................................ ......................................................................... .....108.1.7 AS, status request......................................................................................... ................................................. 108.1.8 AA, send buffer again ...................................................... .......................................................................... ....108.1.9 AC, cleaning mirror ............................................ .................................................... ........................................ 108.1.10AM, select measurement input .................................................. .................................................................... 118.1.11AD, measurement interval ...................................................... ....................................................................... 118.1.12AR, remote resetting.............. ........................................................ ................................................................ 12
8.2 Commands for calibration ..................................................................................................................................... 128.2.1 AO, set offset calibration values ................................................... ................................................................. 128.2.2 AH, set correction factor calibration values............................. .................................................................. .....128.2.3 AE, write to EEPROM............................. .................................................... ................................................... 128.2.4 AL, read calibration values..................................................................... ........................................................ 12
8.3 Commands for addressed operation .................................................................................................................... 13
8.3.1
AG, set sensor address ...................................................... ....................................................................... ....148.3.2 AK, Kiss protocol........................................................................................ .................................................... 14
-
8/11/2019 Manual Klimet A30 Englisch V1_30
3/14
KLIMET A30 Operating instructions Page 3 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
1 Transport, storage
Storage temperature: 5C (41F) to 50C (104F).Atmospheric humidity: 10 to 80% relative atmospheric humidity, noncondensing.Atmospheric pressure: 800...1060 hPa (transport in aircraft only in pressurized cabin; if the lower and upper specified limits
of the atmospheric pressure are violated, recalibration is necessary).Weight: 7.9 kgDimensions: width 450 mm, height 152 mm (19 inch standard housing, 3 height modules), depth 350 mm.
The instrument should remain connected to the power supply and switched on even when not in use.
2 Installation, setting up
Important: in order to measure the correct atmospheric pressure, the instrument must be positioned at the height at whichthe pressure should be measured, connection of a hose is not sufficient! Reference point for the calibration is the height ofthe instrument. The instrument must be set up horizontally.
If the dew-point sensor is operated via a hose, the fan must be switched to full power; to do this, short diode ZD1 (3.3 V) onthe board with the temperature inputs (KLA30 T4) by closing the soldering jumper LB1. If necessary, the fan can also beswitched over to continuous operation by closing the soldiering jumper LB2.
3 Technical data
3.1 Installed loads
Operating voltage: Selectable 200...250 V or 100...130 V, 50/60 HzPower consumption: Typically 3.8 VA, max: 10 VA
3.2 General
The specified accuracy is first reached after a warm-up time of min. 90 minutes. The instrument should always remainswitched on.
The specifications apply to a calibration interval of 12 months.
3.3 Measurement cycle
The measurement cycle is 5 minutes, measurement is always performed according to the following scheme:Time Measurement0:00...3:00 test of dew-point mirror, clean if necessary3:00...3:40 fan on, adjust mirror system3:40...4:50 measurement4:50...4:52 evaluation, output new measured value/new measured value ready4:52...5:00 wait until time x:x0 or x:x5
-
8/11/2019 Manual Klimet A30 Englisch V1_30
4/14
KLIMET A30 Operating instructions Page 4 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
3.4 Standard deviation
The standard deviation of the values sampled during the measurement time (70 s) is specified.
3.4.1 Air temperature
Resolution: 0.001 C
Measurement range: 15...25 C KLIMET A30-I: -100...+100 CAccuracy: +/-0.05 K KLIMET A30-I: -20...+60 C +/- 0.08 K-100...+100 C +/- 0.10 K
Standard deviation at constant temperature:
-
8/11/2019 Manual Klimet A30 Englisch V1_30
5/14
KLIMET A30 Operating instructions Page 5 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
4 Maintenance
The instrument and its sensors require no maintenance. The dew-point mirror is also maintained automatically and needs nomanual cleaning.
5 Indication of special operating conditions
The KLIMET A30 data message contains a status value which indicates certain malfunctions.Several status messages can appear at the same time, the corresponding bits in the status byte are then added. Status
messages
ccc Checksum (must have 3 digits)
The "Frame"A97......ccc which surrounds the commands has a fixed format, there must always be a 2 digit address
and a 3 digit checksum. By switching off the checksum test, you might enter any 3 digits as the checksum.
-
8/11/2019 Manual Klimet A30 Englisch V1_30
6/14
KLIMET A30 Operating instructions Page 6 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
The most important commands
A97Eccc Write data to EEPROM
Write individual data to EEPROM.
Example:
A97Eccc
A97Lccc Read calibraton value
Example:
A97L20ccc Read calibration value 20
A97Rccc Remote reset
The software will wait in a loop, letting the watchdog reset circuit performing a hardware reset.
Example: A97Rccc
A97W:ccc Send adjustement value to barometer
The calibration value is stored in the RAM of the barometer. Attention: After successful adjustment, you must enter the
corresponding command, to copy the values to EEPROM.
Nr.: Number of the calibration value
EW: calibration value
Examples:
A97W20:-0.10ccc calibration value [20] = -0.10
A97W21:1.009ccc calibration value [21] = 1.009
Simple offset adjustment
For this adjustment, a constant value EW[20] will be added to the measured air pressure. The offset is derived from acalibrated reference cell.
POffset= PRef PThe value will be stored as EW[20].
Example:
A97W20,-0.10ccc
Now the KLIMET A30 shows in column 2 the corrected air pressure PK and in column 1 the air pressure before correction.
The corrected air pressure must now be equal to the reverence cell.
If the values are corresponding, then the calibration values must be stored in EEPROM by invoking the command
A97Eccc.
By invoking the commandA00Rccc the KLIMET A30 and also the barometer returns in normal operation.
-
8/11/2019 Manual Klimet A30 Englisch V1_30
7/14
KLIMET A30 Operating instructions Page 7 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
Offset and span adjustment
For an offset and span correction, min 2 measuring points are needed. By means of the linear regression the offset b and
the slope m (span) is calculated. To start calculation you need the uncorrected values P displayed in the first column.
The software calculates the corrected air pressure PKfrom the formula:
PK= m * P + bThe offset will be stored in AVW[20] die slope m (span) as AVW[21].
Example:A97W20:-0.10ccc
A97W21:1.009ccc
To double-check if the adjustment is correct a second measuring is needed.
If the values are corresponding, then the calibration values must be stored in EEPROM by invoking the command
A97Eccc.
By invoking the commandA00Rccc the KLIMET A30 and also the Barometer returns in normal operation.
6.2 Calibration of the barometer KLA30-P (Serial Numbers 0001.. 0021)
The barometer must not be calibrated over the entire measurement range, but only over the range applicable to the intendedmeasurement location (usually 850...1060 hPa). After the calibration, the atmospheric pressure must never exceed or dropbelow the selected range. If the barometer is transported in an aircraft, it must be kept in a pressurized cabin.
For calibration of the barometer, the measurement cycle can be accelerated: Enter commands "AVQ" (fast measurementcycle) and "AD10" (10 s integration time) to output a measured value every 10 s.
Simple offset adjustment
For this adjustment, a constant value EW[19] will be added to the measured air pressure. The offset is derived from a
calibrated reference cell.
POffset= PRef PThe value will be stored as AVW[19].
Example:
AVW19,-0.10
-
8/11/2019 Manual Klimet A30 Englisch V1_30
8/14
KLIMET A30 Operating instructions Page 8 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
6.4 Calibration of the dew-point temperature
It is essential to perform calibration of the dew-point sensor using the standard measurement procedure. The measurementtime is 70 s in this case. The built-in fan is switched on approx. 40 s before the measurement is started.For this calibration, the sensor can be removed from the housing and placed in a test chamber or the air can be introducedusing a hose. If more than 1 m hose is attached, the air must be actively pumped through the system (throughput > 1liter/min).
7 Data format
7.1 General
The data interchange between KLIMET A30 and a PC takes place via an RS232 interface.The data are transferred in ASCII format.
Each command starts with the letter "A" and is closed with "CR" (carriage return, ASCII 13).Any line feed characters (ASCII 10) are ignored.
All commands must be entered in UPPERCASE LETTERS.
Setting of the interface: 2400 baud, 8 bits, no parity.
7.2 Data communication cablePinout of the interface connector:
Pin signal signal direction2 RXD from Computer to KLIMET A303 TXD from KLIMET A30 to Computer5 GND
Note: Use a Nullmodem cable for connection to a PC.
Nullmodem cable for PC with DB9 connector:
Nullmodem cable for PC with DB9 connector:
Klimet A30
7
9
8
DB9
4
3
2
1
6
5 PC
7
9
8
DB9
4
3
2
1
6
5
Klimet A30
7
9
8
DB91
2
4
3
5
6
22
23
25
24
7
14
21
20
19
18
17
16
15
10
11
12
13
9
8
DB25
4
3
2
1
6
5
PC
-
8/11/2019 Manual Klimet A30 Englisch V1_30
9/14
KLIMET A30 Operating instructions Page 9 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
8 Command set
Command, PC to sensor
A {parameter}"CR LF"(The LF characters are optional, i .e. they will be ignored by the system).
8.1 Standard commands
8.1.1 AQ, measured value query
AQ KLIMET A30 sends a complete set of measured values to the PC
Example: Q..,::,,,,,,,std dev. dewpoint>,, "CR LF"
Example: Q0,14.3.96,11:40:00,952.023,23.934,15.433,89.53,0.045,0.003,0.010,0.035Status: 0Date: 14.3.96Time: 11:40:00Pressure: 952.023 hPa
Temperature: 23.934 CDew point: 15.433 CHumidity: 89.53 %Standard dev. pressure: 0.045 hPaStandard dev. temperature: 0.003 KStandard dev. dew point: 0.010 KStandard dev. humidity: 0.035 %
Only the measured values selected with the command "AM" are outputted. (Default: as shown in example.)
8.1.2 AP, atmospheric pressure
AP output current atmospheric pressure
Example: P 951.344 = atmospheric pressure 951.344 hPa
8.1.3 AT, temperature
AT{sensor number} output current temperature
Example: T23.456 = temperature 23.456C
8.1.4 AU, dew point temperature
AU output current dew-point temperature
Example: U20.456 = dew point temperature 20.456C
-
8/11/2019 Manual Klimet A30 Englisch V1_30
10/14
KLIMET A30 Operating instructions Page 10 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
8.1.5 AZ, set/read time
AZ..,::Set KLIMET A30 clock
Example: AZ31.1.96, 12:31:5 = 31st Jan. 96, 12:31:05
- Four figures can be entered for the year if desired.- Entry of the seconds is optional, without any entry the clock will be started with xx:xx:00.
AZ Read clock
Example: 31:1:96,12:31:50 = 31st Jan. 96, 12:31:05
8.1.6 AI, identification
AI KLIMET A30 sends an identification message to the PC.
Example: KLIMET A30, Ver 1.1, 00, SN 000002Instrument, software version, address, serial number
8.1.7 AS, status request
AS KLIMET A30 sends a status message to the PC.
Example: Intervall: 0 s Setting of the interval (0 = off)pwr on since 201 s Time since instrument switched onlast cleaning: 1.3.96, 14:45:00 Date and time of last mirror cleaningcleaning intervall: 460 cycles Cleaning intervalready
8.1.8 AA, send buffer again
AA The complete content of the RS232 buffer will be sent again to the PC.(This function may be used to measure off-line).
8.1.9 AC, cleaning mirror
AC In the next cycle a mirror cleaning is performed, the dew-point mirror is cleanedautomatically if dirty or at the latest after 5 days
-
8/11/2019 Manual Klimet A30 Englisch V1_30
11/14
KLIMET A30 Operating instructions Page 11 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
8.1.10 AM, select measurement input
AM{channel},{on/off} On/off switching of measurement inputs (1 = on, 0 = off)
Channel No Default Function1 on temperature 1
2 on temperature 23 on temperature 34 on temperature 45 on dew point temperature6 off internal compensation channel7 on atmospheric pressure8 on relative humidity12 off differential measurement, sensor 1 - sensor 213 off differential measurement, sensor 1 - sensor 314 off differential measurement, sensor 1 - sensor 434 off differential measurement, sensor 3 - sensor 4
Example: AM1,1temperature 1 on, AM8,0humidity off
AM Request what channels are switched on: The instrument responds with:
Bit Value Function0 1 temperature 11 2 temperature 22 4 temperature 33 8 temperature 44 16 dew point temperature5 32 compensation channel(always on)6 64 atmospheric pressure7 128 relative humidity
Example: 255 inputsall inputs on
AM{no} Alternatively, the numberof the attached temperature sensor can be entered.
Entry of the number of temperature sensors: 90 = no temperature sensor, 91 = 1 sensor
Example: AM93 scan 3 temperature sensors
8.1.11 AD, measurement interval
AD{measurement interval} Output of the data every xx seconds. Primarily of interest for adjustment and forrecording of the climate with a terminal program.0: No interval, output only on request by PC1: Measurement interval 1 second (i.e. as soon as a new measurement value isavailable, = every 5 minutes)With very short interval times, the data are not sent until new measurement data areavailable. The minimummeasurement time is 5 minutes.
If the interval selected is 300, 600, 900, 1200, 1800, 3600 or 7200 seconds (5, 10, 15,20, 30, 60 or 120 Minutes) the output would be synchronized with the clock, i.e. Outputall 2 hours shortly before the full hour. On the other interval times the output will beasynchrony, just after the expiration of the selected time.
Example: AD600 = switch to 10' interval.The "query" data record is outputted every 10 minutes
-
8/11/2019 Manual Klimet A30 Englisch V1_30
12/14
KLIMET A30 Operating instructions Page 12 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
8.1.12 AR, remote resetting
AR The program remains in a wait loop until the watchdog circuit resets the entire hardware.
Example: KLIMET A30, Ver 1.0, 00, SN 000002
8.2 Commands for calibration
8.2.1 AO, set offset calibration values
AO{number of the channel},{calibration value}Transfer calibration values to KLIMET A30.
Example: AO1,0.04567 calibration value 1 = 0.04567
Warning:As calibration values are first changed in RAM, they must be saved with the "AE" command to ensure the changeremains permanently stored.
If an offset has already been entered, this is taken into account. In other words, the new offset is added to the existing one.
8.2.2 AH, set correction factor calibration values
AH{number of the channel},{calibration value}Transfer calibration values to KLIMET A30
Example: AH1,10001567 calibration value 1 = 1.0001567
Warning:As calibration values are first changed in RAM, they must be saved with the "AE" command to ensure the changeremains permanently stored.
If a correction factor has already been entered, this is taken into account. In other words, the new entry is multiplied by theexisting correction. Entry of the calibration value "0" deletes the correction, i.e. it is set to 1.0000.
8.2.3 AE, write to EEPROM
AE Copy calibration values from RAM to EEPROM, the copying operation lasts severalseconds and is documented on the screen.
8.2.4 AL, read calibration values
AL Read calibration values of KLIMET A30.
Meaning of the calibration values:
AVW1 Date of the last calibration
AVW6 Correction factor for temperature sensor 1 (the result is multiplied by this factor)AVW7 Correction offset for temperature sensor 1 (added to the result)
AVW8 Correction factor for temperature sensor 2 (the result is multiplied by this factor)AVW9 Correction offset for temperature sensor 2 (added to the result)
AVW10 Correction factor for temperature sensor 3 (the result is multiplied by this factor)AVW11 Correction offset for temperature sensor 3 (added to the result)
AVW12 Correction factor for temperature sensor 4 (the result is multiplied by this factor)AVW13 Correction offset for temperature sensor 4 (added to the result)
AVW14 Correction factor for dew point temperature (the result is multiplied by this factor)AVW15 Correction offset for dew point temperature (added to the result)
AVW18 Correction factor for barometer (the result is multiplied by this factor)AVW19 Correction offset for barometer (added to the result)
-
8/11/2019 Manual Klimet A30 Englisch V1_30
13/14
KLIMET A30 Operating instructions Page 13 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
AVW37 Correction factor for differential temperature measurement 1-2 (the result is multiplied by this factor)AVW38 Correction offset for differential temperature measurement 1-2 (added to the result)
AVW39 Correction factor for differential temperature measurement 1-3 (the result is multiplied by this factor)AVW40 Correction offset for differential temperature measurement 1-3 (added to the result)
AVW41 Correction factor for differential temperature measurement 1-4 (the result is multiplied by this factor)AVW42 Correction offset for differential temperature measurement 1-4 (added to the result)
AVW43 Correction factor for differential temperature measurement 3-4 (the result is multiplied by this factor)AVW44 Correction offset for differential temperature measurement 3-4 (added to the result)
AVW50 Measuring interval fast, if AVW50 is bigger then 0 the instrument is started just after powering on.
AVW51 Default value for choosing the measuring inputs (AM1..8) Default is 224
AVW52 Default value for choosing differential measuring (AM1..8) Default is 00
8.3 Commands for addressed operation
"CR LF"
Designates the type of measured value:
P: atmospheric pressure, T: temperature, U: dew point, F: rel. humidityQ: PTUF combined (depending on preselection of measurement channels
: (01...99: free sensor addresses)
informative data
sum of all character values modulo 1000
Sensor address 00: All sensors are addressed with address 0001...99: Free sensor addresses
Instruments with address 00 are addressed by all addresses.If address 00 is selected, the checksum is ignored.
As the "frame" ("A") in which the information is packaged has a fixed data format, 2characters must always be sent for the sensor address and 3 characters for the checksum.
-
8/11/2019 Manual Klimet A30 Englisch V1_30
14/14
KLIMET A30 Operating instructions Page 14 (14)
meteolabor ag CH-8620 Wetzikon November/2/2004File: Manual Englisch V1.30.doc
8.3.1 AG, set sensor address
AG{sensor address} Set new sensor address and switch on addressing mode.
The addressing and checksum is normally "bypassed" thus making the interface somewhat simpler (so-called Kiss mode(Keep It Simple and Stupid), The Kiss mode is activated by switching on the instrument or by entry of A00Kxxx.
The addressing mode is activated as soon as a valid address is entered.
Example: AG23 = New sensor address 23AG00 = General sensor address, instrument responds to all addresses.
8.3.2 AK, Kiss protocol
A{sensor address}K{checksum} Selection of the Kiss protocol, deactivates the addressing as well as the generation andtesting of the checksum.
Example: A00K236 (this command must always be sent with address and checksum: A00K236)