EN
Manual
MADE IN GERMANY
EN
![Page 2: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/2.jpg)
![Page 3: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/3.jpg)
EN
32010/08/06 9000-618-83/30
Contents
Important Information 1. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 NoteonConformity.. . . . . . . . . . . . . . . . . 41.2 GeneralNotes. . . . . . . . . . . . . . . . . . . . . . 41.3 Appliancedisposal.. . . . . . . . . . . . . . . . . 41.4 Correct
Usage.. . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.5 Incorrect
Usage.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.6 UsingPeripheralDevices.. . . . . . . . . . . 51.7 Usingimageplates. . . . . . . . . . . . . . . . . 51.8 UseofSoftware.. . . . . . . . . . . . . . . . . . . . 5
2. Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.1 GeneralSafetyNotes.. . . . . . . . . . . . . . . 62.2 Electricalsafetyinstructions.. . . . . . . . . 6
3. Warnings and symbols . . . . . . . . . . . . . . . . 73.1 Adhesivelabel. . . . . . . . . . . . . . . . . . . . . . 73.2 Modelidentificationplate.. . . . . . . . . . . . 7
4. Delivery Contents . . . . . . . . . . . . . . . . . . . . . 84.1 Specialaccessories. . . . . . . . . . . . . . . . . 84.2 Disposablematerials. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 9
5. System requirements . . . . . . . . . . . . . . . . . 105.1 Monitor.. . . . . . . . . . . . . . . . . . . . . . . . . . 105.2. Minimumsystemrequirements. . . . . . 10
6. Technical Data . . . . . . . . . . . . . . . . . . . . . . . 116.1 CR35VET. . . . . . . . . . . . . . . . . . . . . . . . 116.2 Imageplate.. . . . . . . . . . . . . . . . . . . . . . . 11
7. Functional layout . . . . . . . . . . . . . . . . . . . . . 12 8. Functional Description . . . . . . . . . . . . . . . 13
8.1 Functionofindividualcomponents.. . 138.2 Description.. . . . . . . . . . . . . . . . . . . . . . . 14
Mounting 9. Transporting the scanner . . . . . . . . . . . . . 15 10. Carrying the scanner . . . . . . . . . . . . . . . . . 15 11. Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
11.1 Set-uplocation.. . . . . . . . . . . . . . . . . . . 1611.2 Set-upalternatives.. . . . . . . . . . . . . . . . 1611.3 Installationofthelightreducingarc..17
12. Electrical Connections . . . . . . . . . . . . . . . 17 13. Before Commissioning . . . . . . . . . . . . . . . 18
13.1 ConnectthescannertothePC.. . . . 1813.2 InstallationofVet-Exam. . . . . . . . . . . . 20
14. Commissioning .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 2114.1 ConfigurationadTesting. . . . . . . . . . . 2114.2 Systemtest.. . . . . . . . . . . . . . . . . . . . . . 2214.3 Connectingthescannerandthe
imageeditingprogram.. . . . . . . . . . . . 22
Use of the Appliance 15. Using image plates . . . . . . . . . . . . . . . . . . . 23
15.1 Workingwithimageplates.. . . . . . . . 2315.2 Storingimageplates. . . . . . . . . . . . . . 23
16. Foil cassettes/ Foil sleeve. . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
17. Light protection cover S0 - S4. . . . . . . . . 24 18. Extraoral x-ray . . . . . . . . . . . . . . . . . . . . . . . 26
18.1 PreparingtoX-ray.. . . . . . . . . . . . . . . . 2618.2 Exposingtheimageplate.. . . . . . . . . . 2718.3 Scanningtheimagedata. . . . . . . . . . 27
19. Intraoral X-ray . . . . . . . . . . . . . . . . . . . . . . . . 2819.1 PreparingtoX-ray.. . . . . . . . . . . . . . . . 2819.2 Exposingtheimageplate.. . . . . . . . . . 2819.3 Readingouttheimagedata.. . . . . . . 29
20 Erasing image data . . . . . . . . . . . . . . . . . . . 31 21. Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
21.1 Imageplates. . . . . . . . . . . . . . . . . . . . . 3221.2.Foilcassettes.. . . . . . . . . . . . . . . . . . . . 3221.3 Scanner. . . . . . . . . . . . . . . . . . . . . . . . . . 33
22. Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . 34
Trouble-shooting 23. Tips for Operators and Service
Technician . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
![Page 4: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/4.jpg)
EN
4 9000-618-83/30 2010/08/06
Important Information
1. Notes 1.1 Note on ConformityTheproducthasbeensubjectedtoaconformityacceptanceprocedureinaccordancewiththerelevantguidelinesoftheEuropeanUnionforthisapplianceandconformstothedemandedfundamentalrequirementsoftheseguidelines.
ThisclassAappliancemeetsCanadianstan-dardsICES-003.
TheappliancewastestedinaccordancewiththelimitvaluesfordigitalclassAappliancesac-cordingtopart15oftheFCC.Theselimitvaluesshouldensuresufficientprotectionagainstfaultsinthework-area.Theappliancegenerates,usesandradiatesradiofrequencyenergy.Iftheap-plianceisnotsetupandoperatedasstipulatedintheinstallationandoperatinginstructions,errorsinradiotrafficcanoccur.Operatingtheapplianceinaresidentialareacanleadtoap-plianceerrorsoccurring.Rectifyingfaultsistheresponsibilityoftheoperator.
1.2 General Notes•TheseInstallationandOperatingInstructions
formanintegralpartoftheappliance.Theymustbekeptclosetotheunitatalltimes.Ob-servingtheinstallationandoperatinginstruc-tionsisapreconditionforthedesignateduseandthecorrectoperationoftheappliance.Newemployeesmustbeinstructed.Newpersonnelmustbemadeawareofthecontents,andtheyshouldbepassedontofutureoperatingstaff.
•Safetyfortheoperatoraswellastrouble-freeoperationoftheunitareonlyensuredifuseismadeoforiginalequipmentparts.Fur-thermore,onlytheaccessoriesdescribedintheinstallationandoperatinginstructionsorapprovedbyDÜRRMEDICALforusewiththisappliancemaybeused.Ifotheraccessoriesareused,DÜRRMEDICALcanassumenowarrantyforthesafeoperationandreliablefunction.Noclaimsfordamagesresultingtherefromwillbeaccepted.
•DÜRRMEDICALisresponsiblefortheappliancesfromthepointofviewofsafety,reliabilityandfunctiononlywheninstallation,newsettings,modifications,expansionsandrepairsarecarriedoutbyDÜRRMEDICALorbyaservicecentreauthorisedbyDÜRRME-DICAL,andwhentheapplianceisusedandoperatedinaccordancewiththeinstallationandoperatinginstructions.
•TheseInstallationandOperatingInstructionsconformtotherelevantversionoftheequip-mentandtheunderlyingsafetystandardsvalidatthetimeofgoingtopress.Allcircuits,processes,names,softwareandappliancesquotedareprotectedunderindustrialpropertyrights.
•Thetranslationoftheinstallationandopera-tinginstructionshasbeenmadetothebestofknowledge.Themanufacturerbearsnoliabili-tiesinthecaseofamistakeinthetranslation.TheincludedGermanversionoftheinstallati-onandoperatinginstructionsisrelevant.
•Reprintingoftheinstallationandopera-tinginstructions,inwholeorinparts,isonlypermittedwiththewrittenapprovalofDÜRRMEDICAL.
•Retainthepackagingforpossiblereturnoftheproducttothemanufacturers.Ensurethatthepackagingisstoredawayfromchildren.Onlytheoriginalpackagingprovidesadequateprotectionduringtransportoftheunit.Iftheappliancehastobereturnedtothema-nufacturersduringthewarrantyperiod,DÜRRMEDICALwillacceptnoliabilityfordamageoccurringduringtransportresultingfrompoorpackaging!
1.3 Appliance disposal applianceTheEU-guidelines2002/96/EG-WEEE(WasteElectricandElectronicEquipment)of27thJanuary2003andthecurrentintegrationintonationallawstatesthatproductscomeun-dertheabove-mentionedguidelinesand,withintheEuropeanUnion,aresubjecttospecialdisposalmeasures.Pleaseaddressquestionsconcerningthepro-perdisposaloftheproducttoDÜRRMEDICALortothetrade.
![Page 5: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/5.jpg)
EN
52010/08/06 9000-618-83/30
image plateWhendisposingofimageplatesobservelocalandregionalrulesandregulations.Theimageplatecontainsluminescentstoragematerialandmustthereforebedisposedofasspecialwaste.•Pleaseaddressquestionsconcerning
theproperdisposaloftheproducttoDÜRRMEDICALortothetrade.
1.4 Correct Usage•Theapplianceisintendedexclusivelyforvisual
scanningandprocessingofimagedataonanimageplateinveterinarymedicine.
TheapplianceisnotsuitableforprolongedX-raydiagnosticmonitoringofanimals.
•Theapplianceisonlytobeusedinclosed,dryrooms.
•CorrectUsageoftheappliancealsoinvol-vesobservingtheInstallationandOperatingInstructionsandadheringtothoseinstructionsconcerningset-up,operatingandmainte-nance.
•Anyusagethatdiffersfromthecorrectusageleadstothelossofwarranty.
Theoperatorisliableforanydamagearisingthroughincorrectusage.
•CorrectUsagealsoinvolvesobservingalllocalandnationalregulationsconcerninghealthandsafetyatworkandradiationprevention.
1.5 Incorrect UsageTheapplianceisnotintendedforvisualscan-ningofimageplatesinhumanmedicine.Anyotheroradditionalusage,especiallyradioscopy,mammographyandapplicationsisconsiderednon-intended.Themanufacturerisnotliableforanydamagesarisingfromincorrectusage.Theoperatorbearsallrisksandliabilities.
This appliance must not be used in operating theatres or similar types of room, where there is a risk of inflam-mable materials catching fire.
Ifthedeviceisutilisedormodifiedotherthanspecifiedunderpoint14,hazardouslaserradia-tioncanbetheresult.
1.6 Using Peripheral Devices •Onlyperipheraldevices(computer,monitor,
printer)maybeconnectedtothisappliancewhichconformtostandardIEC60950-1(EN60950/1).
•Appliancesmayonlybeconnectedtogetherorconnectedtoothersystempartswherethesafetyofpatients,operatorsandalsotheenvironmentisensured.Wheretheabsolutesafetyforoperator,patientorenvironmentisnotapparentthentheope-ratorisexpectedtocallonexpertadvice,e.g.fromthemanufactureroranexpert,thatsuchconnectionsaresafebeforeproceeding.
1.7 Using image plates Thescannermayonlybeoperatedincom-binationwithimageplatesfromDürrNDTorDÜRRMEDICAL.Theimageplatesmaynotexceedamax.feedwidthof35.4cm.Useofanyotherimageplatesmustbeexpress-lyapprovedbyDÜRRMEDICAL.
1.8 Use of SoftwareX-raysoftwarethathasapprovedbyDÜRRMEDICAL,suchase.g.Vet-Exam,mustbeused.UseofanyothersoftwaremustbeexpresslyapprovedbyDÜRRMEDICAL.Furtherinformationcanbefoundontheinternetunderwww.duerr-medical.de.
Seecomputersystemminimumrequire-mentsincombinationwithDürrNDT/DürrMedical-CRsystems,orderNo.9000-608-100/..orontheInternet.
![Page 6: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/6.jpg)
EN
6 9000-618-83/30 2010/08/06
2. Safety 2.1 General Safety NotesTheappliancehasbeendevelopedanddesi-gnedinsuchawaythatdangersareeffectivelyruledoutifoperatedinaccordancewiththeintendeduse.Inspiteofthis,wefeelitisourdutytomentionthefollowingsafetymeasuresinordertopreventanypossibledanger.•Whenusingthisappliancealllocalandrele-
vantregulationsmustbeobserved!Convertingormodifyingtheapplianceinanywayisstrictlyprohibited.DÜRRMEDICALcanassumenowarrantyorliabilityformodifiedoradaptedappliances.Theoperationofmodifiedappliancescanbepunishablebylaw.Intheinterestsoftrouble-freeoperationtheoperatorisresponsibleforobservingtheseregulations.
•OpeninganappliancewhichisstillconnectedtothemainspowersupplyisdangerousandcouldleadtolaserradiationClass3B.
•Theinstallationistobeperformedbyaspe-cialistauthorisedbythemanufacturer.
•Beforeeveryusetheoperatormustcheckthefunctionalsafetyandtheconditionoftheappliance.
•Theoperatormustbeknowledgeableintheoperationoftheappliance.
2.2 Electrical safety instructions•Theappliancemayonlybeconnectedtoa
correctlyinstalledearthedsafetysocket.
•Connectingtomobilemultiplesocketoutletsisnotpermitted.
•Beforeconnectingtotheelectricitysupplytheappliancemustbeinspectedandche-ckedthatthesupplyvoltageandthesupplyfrequencycorrespondtothatofthelocalelectricalsupply.
•Useonlytheprovidedpowerunit.Unpredicta-blehazardscanbecausedbyusinganotherpowerunit.
•Beforeinitialuseandstart-uptheapplianceandallsupplylinesmustbecheckedforanysignsofdamage.Damagedsupplylinesandconnectionsmustbereplacedimmediately.
•Nevertouchpersonsandopenconnectorsoftheappliancesimultaneously.
•Whenusingtheapplianceobservealltherelevantelectricalsafetyprocedures.
Beforeopeningtheunitunplugatthemainsandremoveallpower!
![Page 7: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/7.jpg)
EN
72010/08/06 9000-618-83/30
3.2 Model identification plateThemodelidentificationplateislocatedontherearofthescanner.Thefollowingdesignationsorsymbolsareshownonthemodelidentificationplateandthescanner:
WarningLaser
Warningdangerousvoltage.
Dateofmanufacturer
Observenotesinaccompanyingdocu-mentation!
DisposeofcorrectlyaccordingtoEUdirective(2002/96/EG-WEE)
CElabelwithoutnotifiedbodynumber
ETL-acceditationCONFORMSTOULSTD61010-1CERTIFIEDTOCAN/CSASTDC22:2NO:61010-1
IP20 Fusetype
REF OrderNo./ModelNo.
SN SerialNo.
3. Warnings and symbols Intheoperatinginstructionsthefollowingwar-ningsandsymbolshavebeenused:
Information and/or mandatory regu-lations or prohibitions for the preven-tion of personal injury or substantial property damage
Specialinstructionsconcerningecono-micusageoftheapplianceandothernotes.
Unplugatthemainsandremoveallpower!
ON/OFFkey
OperationalLED,greenStandby
OperationalLED,blueDatatransmission
3.1 Adhesive label
Usetheapplianceonlywithpowerunit.
Donotmovetheappli-ancewhileitisswitchedon.
FCC-acceditation
FDA-acceditation
![Page 8: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/8.jpg)
EN
8 9000-618-83/30 2010/08/06
4. Delivery Contents CR 35 VET 2132-53CR35VET. . . . . . . . . . . . . . . . . 2132-000-53Powerunit . . . . . . . . . . . . . . . . . 2150-470-00DatacableUSB3m. . . . . . . . . 9000-119-067InstallationandOperatingInstruc-tions........................ 9000-618-83/30Lightreducingarc . . . . . . . . . . . 2131-005-01IP-CleaningWipes . . . . . . . . . . CCB351A1001
Dependingoncountryofoperation:MainscableEUR . . . . . . . . . . . 9000-119-060MainscableCH . . . . . . . . . . . . 9000-119-061MainscableGB . . . . . . . . . . . . 9000-119-062MainscableDK. . . . . . . . . . . . . 9000-119-064MainscableI . . . . . . . . . . . . . . 9000-119-068MainscableAUS . . . . . . . . . . . 9000-119-063MainscableJ. . . . . . . . . . . . . . 9000-119-084MainscableUSA . . . . . . . . . . . 9000-119-083
4.1 Special accessoriesThefollowingpartsarenotsuppliedwiththecontentsofdelivery.Pleaseorderasrequired!
Foil sleeves
Format Order No.18x24(1piece). . . . . . . . . . . . . NACS182410724x30(4pieces). . . . . . . . . . . . . 2134-024-0030x40(4pieces). . . . . . . . . . . . . 2134-033-0035x43(4pieces). . . . . . . . . . . . . 2134-034-00
Specialformatswillbeprovideduponrequest.
Solid foil cassettes
Format Order No.2x3-Size0-(1piece) . . . . . . . . . 2130-010-002x4-Size1-(1piece) . . . . . . . . . 2130-011-003x4-Size2-(1piece) . . . . . . . . . 2130-012-002.7x5.4-Size3-(1piece) . . . . . . 2130-013-005.7x7.6-Size4-(1piece) . . . . . . 2130-014-00
Flexible foil cassettes
Format Order No.18x24(1piece). . . . . . . . . . . . . . 2132-023-5024x30(1piece). . . . . . . . . . . . . . 2132-021-5030x40(1piece). . . . . . . . . . . . . . 2132-022-5035x43(1piece). . . . . . . . . . . . . . 2132-024-50
Rigid cassettesFormat Order No.18x24(1piece). . . . . . . . . . . . . KUNKA18240824x30(1piece). . . . . . . . . . . . . KUNKA24300730x40(1piece). . . . . . . . . . . . . KUNKA30401035x43(1piece). . . . . . . . . . . . . KUNKA354307
![Page 9: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/9.jpg)
EN
92010/08/06 9000-618-83/30
Image plates Normal resolution
Format Order No.18x24(1piece). . . . . . . . . . . . . . VMIP182410924x30(1piece). . . . . . . . . . . . . . VMIP243010930x40(1piece). . . . . . . . . . . . . . VMIP304011035x43(1piece). . . . . . . . . . . . . . VMIP3543109
Image plates High resolution
Format Order No.2x3(22x35mm)-Size0-(2pieces). . . . . . . . . . . . 2134-040-502x4(24x40mm)-Size1-(2pieces). . . . . . . . . . . . 2134-041-503x4(31x41mm)-Size2-(4piece). . . . . . . . . . . . . 2134-042-502.7x5.4-Size3-(2pieces) . . . . . 2134-043-505.7x7.6-Size4-(1piece) . . . . . . 2134-044-50
Protective film
Format Order No.18x24(1000pieces). . . . . . . . . PRSH182410824x30(1000pieces). . . . . . . . . PRSH243010835x43(1000pieces). . . . . . . . . PRSH3543108
light protection cover
Format Order No.2x3-Size0-(100pieces) . . . . . . 2134-070-002x4-Size1-(100pieces) . . . . . . 2134-071-003x4-Size2-(300pieces) . . . . . . 2134-072-002.7x5.4-Size3-(100pieces) . . . 2134-073-005.7x7.6-Size4-(100pieces) . . . 2134-074-00
Theusageofothercomponents,trans-formersandcablesasthosespecificallylistedunderaccessories,canleadtoeitherincreasedsignaloutputorreducetheresistancetointerferenceofthesystem.
4.2 Disposable materialsIP-CleaningWipes(100Stk.). . . CCB351A0101
OthercleaninganddisinfectantscanbeobtainedfromthecompanyOrochemie(www.orochemie.de)underthefollowingnames:QuickdisinfectionB30DisinfectantwipesB60
![Page 10: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/10.jpg)
EN
10 9000-618-83/30 2010/08/06
5. System requirementsThefollowingcomponentsarerequiredfortheoperationofthescanner.Thesearenotincludedwiththestandarddeliverycontentssuppliedwiththescanner.
SeealsotheInternetinformationunderwww.duerr-medical.de
5.1 Monitor Monitorsshouldbeusedwhicharecapableofdisplayingdigitalx-rayexposuresandwhichhaveahigherlightintensityandbroadercon-trastrangeandwhichconformtotheinterna-tionalstandardEN25580.Brightambientlight(e.g.withintheroom),directsunlightandreflec-tedlightcanallnegativelyaffectthediagnosticusefulnessoftheX-rayexposure.
5.2. Minimum system requirements
Seetheenclosedinformationsheet,ordernumber9000-608-100.
![Page 11: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/11.jpg)
EN
112010/08/06 9000-618-83/30
6. Technical Data 6.1 CR 35 VETElectrical Details power unitVoltage(V).........................................100-240Frequency(Hz).......................................50-60Current(A)............................................. max.2Currentconsumption(W)..........................<140Fusetype.................................................. IP20Protectionclass............................................... IDegreeofsoiling.............................................2
Electrical ratings and connection data Voltage(V)...................................... 24+/-10%Frequency...................................................DCCurrent(A)............................................. max.5Currentconsumption(W)..........................<120
DimensionsD=52cm,W=38cm,H=39cmFeedwidthforimageplates(cm)........................ max.35.4
Temperature ranges Applianceinoperation(°C)..............+10to+35Storageandtransport(°C)............... -20to+60
Relative humidity Applianceinoperation(%).............. 20-max.90StorageandTransport(%)................... max.10
Ambient air pressure Applianceinoperation(hPa)............750-1060heightabovesealevel(m).......................<2000Storageandtransport(hPa)............750-1060heightabovesealevel(m)....................... 16000
Weight CR35VET(kg)................................ approx.21
General valuesExtern:laserclass...................... 1(EN60825-1)
Intern:laserclass..................... 3B(EN60825-1)
Wavelengthλ635nm15mWEN60825-1:1994-03+A1:2002-07+A2:2001-03
Max.noiselevel(db(A))...............................<60Heatradiation(W).....................................<180
Conformity evaluation
06/95/EG(Low-voltageguidelines04/108/EG(EMCguidelines)
Pixel size
CR35VETcanbeselectedfromrange30µmto130µmResolutiondependingonimageplate,ca.15linepairs/mm(Lp/mm).
6.2 Image plateStorage and transportTemperature(°C).........................................<33Relativehumidity(%)...................................<80
Use of the ApplianceTemperature(°C).....................................<4518Relativehumidity(%)...................................<80
![Page 12: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/12.jpg)
EN
12 9000-618-83/30 2010/08/06
7. Functional layout
10
3
4
1
2
8
5
7
9
11
12 13
14
9
6
15
![Page 13: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/13.jpg)
EN
132010/08/06 9000-618-83/30
8. Functional Description 8.1 Function of individual compon-
ents1. Entry slot: Theimageplateisfedintothescannerthroughtheentryslotandisautomaticallytakenupbythetransportmechanism.
2. Yellow LED:LEDlightsup:•Scannerisswitchedon,butnotyetinitialised.
LEDblinks:•NoconnectionestablishedtoPC.
•Memoryisfull.
3. On / Off key:Toswitchtheapplianceonoroffholdthekeydownfor2seconds.
4. Green LED:Scannerisreadyforoperation.
5. Erasure unit:Allimagedataisautomaticallydeletedfromtheimageplateintheerasureunitafterscanning.
6. Roller end cap:Guiderollersforlow-wearimageplateexit.
7. Transport drum: InsidetheTransportdrumthereisaspecialtransportmechanism,thescanningunitandtheerasureunit.
8. Status LED:TheLEDsshowwhetheranimageplatecanbeinsertedforscanningornot.GreenLED:•Animageplatecannowbeinserted.
YellowLED:•Donotinsertanimageplateyet.
9. Light reducing arc:Thelightreducingarcpreventsanystrayam-bientlightenteringthroughtheexitslotofthetransportdrumandalsostopsthefoilsfallingoutafterscanning.
10. Rigid cassette:Therigidcassette(solidx-raycassette)canbeusedinstandardx-raycassettedrawersinme-dicalx-rayunits.Therigidcassettesalsoprotecttheimageplatesfromdamage.
11. Image plates:TheimageplateservesastheexposuremediumforX-raysandstorestheimagedata.
12. Foil sleeve:•Preventsthedeletionofimagedataonthe
imageplateandprotectstheimageplatefromdamage.
•Theimageplateisslidintothefoilcassetteandthenplacedtogetherintotherigidcasset-te.
•Theimageplateandthefoilcassettearepo-sitionedattheopeningoftheentryslottothetransportdrumandthenfedtothescannerunit.
13. Foil cassettes, flexible:•Preventsthedeletionofimagedataonthe
imageplateandprotectstheimageplatefromdamage.
•Theimageplateandthefoilcassettearepo-sitionedattheopeningoftheentryslottothetransportdrumandthenfedtothescannerunit.
14. Foil cassettes, solid:•Preventsthedeletionofimagedataonthe
imageplateandprotectstheimageplatefromdamage.
•ImageplatesS0-S4arepositionedwiththefoilcassetteattheopeningoftheentryslottothetransportdrumandthenfedsecurelytothescannerunit.
15. Guide rails:•Theextraoralimageplatesarefedevenlyby
theguiderails.
![Page 14: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/14.jpg)
EN
14 9000-618-83/30 2010/08/06
8.2 Description Thescannertakestheimagedatawhichhasbeenstoredontheimageplateandtransformsitintoadigitalimage.
X-ray:
TheimageplateisexposedtoanX-raybeaminthesamewayasastandardX-rayfilm.Thisisdoneinsidethecorrectfoilcassette.
Scanning the image data
Pressingtheon-offbutton(3)switchesonthescanner.Thentheimageprocessingsoftwareactivatesthescanningmode.Thelightprotectioncoverwiththeimageplateisthenpositionedatthemouthoftheentryslot(1)ofthetransportdrum(6)andalignedforfeeding.
WhenthestatusLEDs(7)aregreen,thentheimageplatecanbegentlyfedintotheentryslot(1).Thisisdonewiththethumbsthroughanopeninginthelightprotectioncoverorbypressingontheendoftherecesstowardthetransportdrumonlonglightprotectioncovers.
Theimageplatewillautomaticallybegintobedrawnbythetransportmechanismandwillbefedtothetransportdrum.Insidethetransportdrumtheimageplatewillbeprocessedbythescanningunit,wheretheimagedatawillbescannedandthentransmittedtothePC.Awindowonthemonitorshowstheprogressofthisprocedure.
Erasure of image data
Afterscanningtheimageplateproceedstotheerasureunit(5).Imagedatastillheldontheimageplatewillbeerased.Finallytheimageplateproceedstotheexitslotandisejectedfromthetransportdrum.TheimageplatecannowbepreparedforthenextX-rayexposure.
Ifusedcarefullyitispossibletoreusetheimageplatesseveralhundredtimes.Beforereusetheimageplatesshouldbecheckedforsignsofdirtordamage.Damagedimageplatesmustnotbereused.
imageplate
Par
abol
icm
irror
Photomultiplier
lasersource
reflectionlayer
Pen
tapr
ism
Rot
atin
gdi
sc
![Page 15: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/15.jpg)
EN
152010/08/06 9000-618-83/30
Mounting
Only suitably qualified personnel are permitted to set up, to install or to operate the appliance.
9. Transporting the scanner •Checkthatthedeliverycontentsarecomplete
andcheckforanysignsoftransportdamage.
•Emptythepackingcompletelyandstoreinadryplace.
•Donotexposethescannertoanystrongshockswhichcoulddamagetheappliance.
•Onlyusetheoriginalpackingcasesforship-ping.
10. Carrying the scanner
Always hold the scanner by its housing and grip from the rear side (fig. 1).Never carry the scanner by holding the transport drum, or the cylinder, as this can lead to sensitive parts of the scanner being damaged (fig. 2 and 3).
1
2
3
![Page 16: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/16.jpg)
EN
16 9000-618-83/30 2010/08/06
11. Set-up 11.1 Set-up location•TheVistaScanPerioshouldbesetupina
closed,dryandwell-ventilatedroom.
It is important to ensure that the scanner does not become dirty. See also section 18. Cleaning.
•Theroomtemperatureshouldnotfallbelow+10°Corexceed+35°C.
Material damage! Danger of short circuit due to build up of condensation. The system can only be put into operation once the appliance has warmed up to room temperature and it is dry.
• Installationinapurpose-builtroom,e.g.inaboilerroomorinawetroomisnotpermitted.
•Donotexposethescannertodirectsunlight.Theambientlightmustbereduced,ifneces-sary.Themaximumlightintensityis1000Lux.
•Theapplianceisdesignedforoperationinlaboratory-,testing-andmeasuring-areas.Theremustbenosourceofinterferenceintheroomforset-up(e.g.strongmagneticfield).Fieldsofinterferencecancauseerrorsintheappliance.
•Setuptheappliancesothatitisnotsituatedwithinthepathoftheraysfromthex-ray.
11.2 Set-up alternativesInordertoavoiderrorsduringscanningofimagedataitisimportanttosetupthescanneronastablesurfaceandaleversubsurface.
4
![Page 17: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/17.jpg)
EN
172010/08/06 9000-618-83/30
12. Electrical Connections Beforeoperationisbegun,makesurethatthepowersupplyandthemainsvoltagematchtheinformationonthemodelidentificationplateofthepowerunit.Theconnectionportsareintherecessesontherearsideoftheappliance.
•Connectthemainscabletotheappliancepowerunitandtothemainssupply.
•Connectthepowercabletothesocketonthescanner.
•Placethepowercableintothepoweroutletsocket.
Theappliancedoesnothaveamainpowerswitch.Theappliancemustthereforebesetupsothatthesocket-outletiseasilyaccessibleandhasamaximumpowercablelengthof3mincaseofanypossiblefaults.
11.3 Installation of the light reducing arc
Thelightreducingarcmustbeinstalledattheexitsectionofthetransportdrumtoprovideextraprotectionfromstrayambientlight.
•Engagethelightreducingarcstothetwostuds(leftandright).Thelightshieldarcpreventsthefoilfromfalling.
Ensure that transport of the foil from the exit is not hindered.
5
![Page 18: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/18.jpg)
EN
18 9000-618-83/30 2010/08/06
13. Before Commissioning 13.1 Connect the scanner to the PCConnection to USB-Port
•ConnecttheUSBcableprovidedwiththeUSB-interfaceofthePCandtheUSB-inter-faceofthescanner.
InstallationoftheCR35VETviaUSBcanonlybeeffectedwhenafunctio-ningUSB-PC-portisavailable(2.0).WindowsXPsupportsthisinterfaceasofServicePack2.WindowsVistasupportsthisinterfaceasstandard.
•Parallelinterface(*):Forservicepurposesonly.
Check USB interface
•Opensystemproperties.WindowsXP:RightclickonWorkstation,Properties.WindowsVista:RightclickonComputer,Properties.
•ChecktheversionnumberoftheServicePackandupdateifnecessary.
•OpenDevicemanager.WindowsXP:RightclickonWorkstation,Properties,Hardware.WindowsVista:RightclickonComputer,Properties,Hardware.
•AUSBportisavailablewhen„USB-Control-ler“islistedunderappliances.
* *
6
7
8
![Page 19: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/19.jpg)
EN
192010/08/06 9000-618-83/30
Install USB driver
•Switchonthescanner
•Windowswillautomaticallydetectanewappliance(CRScanUSB).
•Choose"No,notthistime".
•Clickon"Next".
•Choose"Installsoftwareautomatically(recom-mended)".
•Clickon"Next".
Thedriverisinstallednow.
•Completetheinstallationwith„Complete“.
Check the driver in device manager
•OpenDevicemanager
UnderWindowsXP:RightclickonWorkstation,Properties,Hardware,Devicemanager.UnderWindowsVista:RightclickonComputer,Properties,Devicemanager.
•ThecheckissuccessfulwhentheCRScanUSBdriverislisted.
9
10
11
![Page 20: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/20.jpg)
EN
20 9000-618-83/30 2010/08/06
USB2 Trouble-shooting
Ifscannerisdisplayedas„unidentifiedapp-liance“underdevicemanagerthenthereispossiblyaUSB2driverfault.
or:
• InstallServicepack3(XP).
•Selectthedriver„ExtendedHostController“usingtheOption[Driver]and[Updatedriver].
•Callupallthedriveralternativesinmenu,andselecttheapplicabledriverfromMicrosoftServicepack.
USB-driverscaninterferewitheachother.Thisbecomesnoticeablethroughlongerdatatransmissiontimes,orlackofcommunicationbetweenCR35andPC.
13.2 Installation of Vet-ExamIfVet-Examisnotinstalledonthecomputersprovided,performtheinstallationinaccordancewiththeinstallationinformationontheCD.
12
13
![Page 21: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/21.jpg)
EN
212010/08/06 9000-618-83/30
14. Commissioning 14.1 Configuration ad TestingProgram start
•Switchonthescanner
•Clickthefollowinginsequence[Start],[Program],[DuerrNDT],[CRScan]and[CRScanConfig].
•Click"TestCommunication".
Theseconddriverisinstallednow.Repeattheinstallationasdoneforthefirstdriver.
•Click"TestCommunication"again.Thetrans-ferrateisdisplayed.
Dialogue field [Connections]
• Ifnecessaryselectfromthemenu[Scannerisconnected]theactualport(USB)beingused.
Dialogue field [Test]
•Selectthescanmodeinthefield[selectscanmode].
•Click[Readimage].Thescannerwillnowbeactivated.CheckthatstatusoftheLEDsattheinletslotsaregreen.
•NowscananexposedX-rayimageonanimageplate.
•ThetesthasbeensuccessfulwhenanX-rayimageistransmittedtothePCmonitor.
•Closethewindow.
•SavetheX-rayexposurefordocumentationpurposes.
14
15
16
![Page 22: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/22.jpg)
EN
22 9000-618-83/30 2010/08/06
Dialogue field [Oscilloscope mode]
•Checkset-upconditions(lightexposure<1000lux).
•Setthedisplayrange(leftslidingbar)to500andtheamplificationHV(upperslidebar)to1000.
•Click[Start].
•Thetestissuccessfulwhenthemeasuredleveldoesnotexceed500.
14.2 System test•Checkandoptimisethesettingsofthe
monitorandofthegraphicsboardusingthegrey-valuetestimagesontheCD-ROM.
14.3 Connecting the scanner and the image editing program
Thescannercaneitherbeoperatedviaa".dllinterface",TWAINinterface(CRScanEasy)orviathe"VET-Exam"program,whichisprovideduponrequest.TheinterfacemustbeconnectedandsetupbyanITexpertofthesoftwaremanufacturer.
Read the data in the example Vet-Exam
SeeinstructionsVet-ExamontheCD.
17
18
![Page 23: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/23.jpg)
EN
232010/08/06 9000-618-83/30
activeside:
Use of the Appliance
15. Using image plates Thescannermayonlybeoperatedincombina-tionwithimageplatesfromDürrNDTorDÜRRMEDICAL.UseofanyotherimageplatesmustbeexpresslyapprovedbyDÜRRMEDICAL.
activeside:•Theactivesideiseitherwhiteorlightblue.
inactiveside:•Theinactivesideoftheimageplateisblack.
Themanufacturer´sinformationisprintedonthisside.
15.1 Working with image plates Imageplatesmustbevisiblycheckedforsignsofdamage.Damagedimageplates(e.g.atornouterlayer,orvisiblescratches,whichwillaffectthediagnosticassessment)mustbereplaced.
• imageplatesareflexiblelikeX-rayfilm.
•donotbendorsubjecttopressure.
•donotscratchorgetdirty.
15.2 Storing image plates •storeimageplatesinadarkplace.
•donotstoreinhotormoistconditions.
•storeimageplatesinthecorrectlysizedfoilcassette.
• ifanimageplateisstoredforlongerthan24hours,thenitshouldbeerasedagainbeforethenextX-rayexposure.
•protectfromdirectsunlightandultravioletbeams.
•WhenstoredinthevicinityofX-rayunitstheimageplatesmustbeprotectedfromthepossibilityofprematureexposuretoX-rayradiation.
inactiveside:
19
20
21
![Page 24: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/24.jpg)
EN
24 9000-618-83/30 2010/08/06
16. Foil cassettes/ Foil sleeve
•Preventimagedataontheimageplatefrombeingerased.
•Protectstheimageplatesfromdirtandconta-mination.
•Theimageplateandthefoilcassetteorthelightprotectioncoverarepositionedattheopeningoftheentryslottothetransportdrum.
•Thefoilcassetteorlightprotectioncoverisusedtoaidtheinsertionoftheimageplateintothescanner.
Foil sleeve
Ifworkinginhumidanddirtyareas,theflexiblefoilsleeveistobeused.Theimageplateisinsertedandthefoilsleeveisclosed.Thesealmustbeontheinactivesideoftheimageplate.
Foil cassette, flexible
Thefoilcassettecanbeusedinareasthatarenotcontaminated.Theinactivesideoftheimageplatemustbevisibleafterinsertioninthefoilcassette.
Foil cassette, solid (for image plates S0 - S4)
Thetongue(6c)situatedonthelever(6b)helpstopushtheimageplatefromthefoilcassetteandintotheCR35VET.
Do not remove this tongue!
Rigid cassettes
Therigidcassette(solidx-raycassette)canbeusedinstandardx-raycassettedrawersinme-dicalx-rayunits.Therigidcassettesalsoprotecttheimageplatesfromdamage.
Foil cassette, flexible
Foil sleeve
6 c
6b
Foil cassette, solid
Rigid cassette
22
23
24
![Page 25: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/25.jpg)
EN
252010/08/06 9000-618-83/30
17. Light protection cover S0 - S4
lightprotectioncover• lightprotectioncoversprotectfrominfection
bycontaminatedimageplates.
•protecttheimageplatefromcontaminationanddirt.
•protecttheimageplatesfromstraylight.
• isavailableindifferentsizesjustlikeimageplates.Choosethemostappropriatesize.
Light protection cover S0 - S4
25
![Page 26: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/26.jpg)
EN
26 9000-618-83/30 2010/08/06
18. Extraoral x-ray 18.1 Preparing to X-rayRigid cassette
Arigidcassetteistobeusediftheimageplatecanbedamagedmechanicallyduringtheexposure.
•Pushthedeletedimageplateintothefoilcas-sette,flexibleandinsertintotherigidcassette.Theactivesideoftheimageplatemustbeadjacenttotheclosedsideoftherigidcassette.
•Closethecoverontherigidcassette.
TheclosedsideoftherigidcassettemustbefacinginthedirectionoftheX-raytube.
Foil cassette, flexible / Foil sleeve
•Slidethepreviouslyerasedimageplateintothefoilcassette.Theblacksideoftheimageplatemustbevisible.
or:
•Pushthedeletedimageplateintothefoilcas-sette,flexibleinthefoilsleeveandclose.Thesealmustbeontheinactivesideoftheimageplate.
•Placethefoilcassette/foilcassetteontheobject.
TheactivesideoftheimageplatemustfacetheX-raytube.
26
27
28
![Page 27: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/27.jpg)
EN
272010/08/06 9000-618-83/30
18.2 Exposing the image plate•SetexposuretimeandsettingsontheX-ray
appliance.
•Exposetheimageplate
Possible image data loss: The image data on the image plate must be read by the scanner within a half hour!
18.3 Scanning the image data •Switchonthescanner
•UsetheimageeditingsoftwaretocalluptheX-raymoduleonthePC.
•Activatethescanner
•Choosetherequiredscanningmode.
•ThestatusLEDsofthescannermustlightupgreen.
Rigid cassette
•Removethefoilcassette,flexiblewiththeimageplatefromtherigidcassetteandputontheaccesstubeoftheCR35.
Positionthefoilcassetteproperlysothatthefoamrubberlaysevenlyagainsttheentryslotofthetransportdrum.
•Pushtheimageplateintothetransportdrumwithbothhands,itisthenautomaticallypulledintothescanner.
Foil sleeve
•Openhookandloopfastener.
•Removethefoilcassette,flexiblewiththeimageplatefromthefoilsleeveandputontheaccesstubeoftheCR35.
Positionthefoilcassetteproperlysothatthefoamrubberlaysevenlyagainsttheentryslotofthetransportdrum.
•Pushtheimageplateintothetransportdrumwithbothhands,itisthenautomaticallypulledintothescanner.
29
30
31
![Page 28: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/28.jpg)
EN
28 9000-618-83/30 2010/08/06
19. Intraoral X-ray19.1 Preparing to X-rayLight protection cover S0 - S4
•SlidetheimageplatecompletelyintothelightprotectioncoverS0-S4.Theblack(inactive)sideoftheimageplatemustbevisible.
•PeeloffthewhitestickystripandpressfirmlytoclosetheopeningofthelightprotectioncoverS0-S4.
The light protection cover S0 - S4 is a disposable article and must not be reused.
•DisinfectthelightprotectioncoverS0-S4usingadisinfectionwipe(e.g.OrochemieB60)beforepositioning.Alternatively,useadisin-fectantspray(e.g.OrochemieB30)onasoft,fluff-freeclothinsteadofadisinfectantwipe.
19.2 Exposing the image plate•SetexposuretimeandsettingsontheX-ray
appliance.
•Exposetheimageplate
Possible image data loss: The image data on the image plate must be read by the scanner within a half hour!
32
33
34
![Page 29: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/29.jpg)
EN
292010/08/06 9000-618-83/30
19.3 Reading out the image data Foil cassette, solid
•OpenthelightprotectioncoverS0-S4straight alongtheadhesiveedge(V-cut).
•Chooseafoilcassetteofthecorrectsize.
•Slidetheleveronthefoilcassettedownwardsuntilitengages.Thetonguemustbecom-pletelyvisible.
•PlacetheimageplatewithinitslightprotectioncoverS0-S4ontothistongue.
•Slidetheimageplateoutofthelightprotec-tioncoverS0-S4andimmediatelyintothefoilcassetteuntilitisfirmlyinposition.Theactivesidemustbeturnedawayandtheinactivesideupwards(visible).
•Placethefoilcassettewiththeimageplateinoneofthefreeentryslotsatthetransportdrumandengage.
35
36
37
![Page 30: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/30.jpg)
EN
30 9000-618-83/30 2010/08/06
•Slidethelevertowardthescannerapparatusuntiltheimageplateisautomaticallydrawnin.
DuringscanningthestatusLEDsattheentryslotwilllightupyellow.
Image data loss: The foil cassette may only be remo-ved from the transport drum when all 4 status LEDs are green.
•Theimageswillautomaticallybetransmittedtothesoftware.Thescanningprogresscanbefollowedinthepreviewwindow.
38
![Page 31: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/31.jpg)
EN
312010/08/06 9000-618-83/30
20. Erasing image dataAfterscanningtheimageplatepassestotheerasureunit.Imagedatastillheldontheimageplatewillbeerased.
Determine the erasure time
Theerasuretimeisbasedontheintensityofthelightsource.Ifimageinformationistobeerasedontheimageplatewithinasecond,theoreti-cally,alightintensityof2.4millionluxsec.isnecessary.Theminimumerasuretimeinsecondsisdeterminedbydividing2.4millionbytheactualintensityofthelightsourceinlux.Ifthelightintensityisnotobviousfromthedocumentationonthelightsource,aluxmetermustbeused.Typicallightintensitiesofx-rayfilmviewersarearound20,000luxforexample.Thisproducesanerasuretimeofapproximately2.4million/20,000=120s.
Dependingonthedosageused,theerasuretimecanvary.
Additional erasing of image data
Todeterminewhetherimagedatamustalsobeerasedontheimageplate:
•Rescantheimageplate.
•Evaluatethescannedx-ray.
Atthecountsinthescannedimagegreaterthan600,theimageplatemustalsobeerased.
39
![Page 32: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/32.jpg)
EN
32 9000-618-83/30 2010/08/06
21. Cleaning
OnlycleaningpreparationsapprovedbyDÜRRMEDICALmaybeused.Othercleaningagentsmaycausedamagetotheapplianceandtheaccessories.WerecommendusingDÜRRNDTIPCleaningWipes,OrderNo.CCB351A0101.
21.1 Image plates •Dirtoneithersideshouldbecleanedoffusing
asoftnon-fluffycloth.ResidualcontaminationcanberemovedbycarefullywipinginacircularmotionwithDÜRRNDTIPCleaningWipes.
•Beforereusage,theimageplatemustbecompletelydry.
40
41
![Page 33: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/33.jpg)
EN
332010/08/06 9000-618-83/30
21.2 Foil cassettes
Visible contamination must be removed before continuing use.
Withdirtorcontamination:•Largerparticlesofcontaminationistobe
removedwithdampcloth.
•Spraydisinfectant(e.g.OrochemieB30)ontoasoft,lint-freeclothoruseadisinfectionwipe(e.g.OrochemieB60)andwipetoremovecontamination.
•Beforereusage,thecassettemustbecom-pletelydry.
21.3 Scanner
Damage to the CR 35 VET: Do not spray the appliance with disinfectant! Liquid is not to infiltrate the inside of the appliance, which could damage the appliance.
•Thesurfaceoftheapplianceistobecleanedwithadampcloth.
•Disinfectthesurfacewithanon-aggressivecleaningdisinfectantwipe(e.g.OrochemieB60).
Seetheusageinstructionsforthedisinfectantandcleaningagents.Were-commendusingDÜRRNDTIPCleaningWipes,OrderNo.CCB351A0101.
42
43
44
![Page 34: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/34.jpg)
EN
34 9000-618-83/30 2010/08/06
22. MaintenanceThereisnomaintenancerequiredbytheuser.OccurringmaintenanceintervalsmustbeperformedbyaservicetechnicianthathasbeenauthorisedbyDÜRRMEDICALormustbecom-pletedintheDÜRRMEDICALplant.
![Page 35: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/35.jpg)
EN
352010/08/06 9000-618-83/30
Trouble-shooting
23. Tips for Operators and Service Technician Repairs may only be carried out by qualified personnel or by our After-Sales Service.
Before starting repairs or opening the appliance, remove the mains plug from the plug socket.
Problem Probable cause Solution1. Scanner cannot be
switched on•Nosupplyvoltage •Checkthemainscableandall
connections•Checkthemainfuseboxtobuil-
ding•Checkthepowerunit,whether
greenLEDlights
•On/Offswitchisdefective • InformyourServiceTechnician
2. Scanner switches off after a brief period
•Mainscableoroneoftheplugshasnotbeeninsertedcorrectly
•Checkthemainscableandallconnections
•Hardwaredefective • InformyourServiceTechnician
•Mainssupplyvoltagetoolow •Checkthesupplyvoltage
3. Status, error or operation LED does not light up
•LEDdefective • InformyourServiceTechnician
4. Loud noises during operation after switching on
• Ifthenoisespersistlongerthan30seconds,thenthebeambaffleisdefective
• InformyourServiceTechnician
5. Image processing software does not recognise the scanner
•Scannerisnotswitchedon •Switchonthescanner
•Connectioncablebetweenscan-nerandPClooseornotconnec-ted
•Checkallconnectioncables
•PCdoesnotrecogniseconnec-tiontoscanner
•CheckUSB-connection
•USB-drivernotinstalledorinacti-ve
•LoadoractivatethedriverfromtheCD.
•Hardwarefaultwithscanner • InformyourServiceTechnician
6. Problem on data transmission bet-ween scanner and PC. Error message: "CRC error" "Ti-meout"
•Connectioncableofwrongtypeortoolong
•Onlyuseoriginal"DürrCable"ordernumber9000-119-067
![Page 36: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/36.jpg)
EN
36 9000-618-83/30 2010/08/06
Problem Probable cause Solution7. Image plate is not
drawn into the app-liance via transport arc
•Foilcassettenotcorrectlypositio-ned
•Placethefilmcassettecorrectly
•Transportmechanismisnotactivatedandscannerisnotatstand-by
•Seesection17.3scanningimagedata
8. X-ray exposure does not appear on the monitor after scanning
• Imageplatewasinsertedupsidedownandtheinactivesidescan-ned
•Rescantheimageplateasquicklyaspossible
• Imagedatafromtheimageplatewasdeleted
•TheimagedatacontainedontheimageplatemustbescannedasquicklyaspossibleonremovalfromtheX-raycassette
•Problemwiththehardware • InformyourServiceTechnician
•X-raydeviceisfaulty • InformyourServiceTechnician
•Noimageinformationontheimageplate
•Exposetheimageplate
9. X-ray exposure is too dark
•X-rayexposureisoverexposed •SetthebrightnessoftheX-rayimageusingthesoftware
•Theimageplatewasexposedtotoomuchlightbeforeinsertionintothefilmcassette
• Insertionmustbecarriedoutwithin30secondsunderambi-entlightconditionsofunderca.20 Lux.
10. Image presented as mirror image
• Imageplatewasplacedincasset-teupsidedown.
•Placeandaligntheimageplatecorrectly
•Wrongscanningmodeselected •Choosethecorrectscanningmode(with/withoutimageinver-sion)
11. Ghost images and shadows on the X-ray exposure
• Imageinformationpresentontheimageplatewasnotsufficientlyerased
•Alwaysuserecommendederasu-retimes.
•Checkthelightintensityandtheerasuretimes
•Seesection17.4Erasureofimageplate
12. Image is very faint on the scanning window software cannot piece the data together to form a complete picture
•TheX-raydoseusedfortheimageplatewastoolow
• IncreasetheX-raydosage
•Amplificationleveltoolow(HV-value)
• Increasetheamplification(HV-value)
•Unsuitablescanningmodeused •Chooseasuitablescanningmode
•Thresholdsettingtoohigh •Reducethesettings
![Page 37: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/37.jpg)
EN
372010/08/06 9000-618-83/30
Problem Probable cause Solution13. X-ray exposure
has stripes on image
• Imageplatehasbeenexposed •Onlyusetheimageplatewithin24hoursofdataerasure.
•Partoftheimageplatewasex-posedtolightduringhandling
•Donotexposeusedimageplatestolight.Exposureandthescan-ningoftheimagedatashouldbecarriedoutwithinhalfanhour.
• Imageplatewasdirtyorscrat-ched
•Cleantheimageplateusingasoftwipeandspecialcleanerforimageplates.Allimageplateswithscratchesmustbereplacedimmediately
14. Warning mes-sage: "Too much ambient light"
•Scannerisexposedtotoomuchambientlight
•Darkentheroom•Turnthescannersothatnolight
fallsdirectlyontheentryslotarea•Changethelightprotectionbrus-
hes
15. Pale stripes in scanning window
•Toomuchstraylightenteredduringscanning
•Onlyremovethefoilcassetteafterthescanningprocesshasbeencompleted
•Darkentheroom•Turnthescannersothatnolight
fallsdirectlyontheentryslotarea•Changethelightprotectionbrus-
hes
16. As well as the correct image, narrow images with light and dark streaks are visible
•Toomuchstraylightenteredduringscanning
•Onlyremovethefoilcassetteafterthescanningprocesshasbeencompleted
•Darkentheroom•Turnthescannersothatnolight
fallsdirectlyontheentryslotarea•Changethelightprotectionbrus-
hes
17. There is a wide, pale stripe at the top border of the image
•Lightenteredafterscanningandthiswasrecognisedbythesoft-wareasbelongingtotheimage.
•Onlyremovethefoilcassetteafterthescanningprocesshasbeencompleted
•Darkentheroom•Turnthescannersothatnolight
fallsdirectlyontheentryslotarea
18. Warning message: Wrong power unit
• Incorrectpowerunithasbeenconnected
•Usethepowerunitprovided
19. Warning: exces-sive heating
•Lasertoohot•Erasureunittoohot
•Switchoffthescannerandallowtocool
20. Warning: erasure unit fault
•LEDdefective•LEDdirty
• InformyourServiceTechnician
![Page 38: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/38.jpg)
EN
38 9000-618-83/30 2010/08/06
![Page 39: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/39.jpg)
EN
392010/08/06 9000-618-83/30
![Page 40: Manual - DÜRR MEDICAL · 2019. 3. 29. · EN 6 9000-618-83/30 2010/08/06 2. Safety 2.1 General Safety Notes Theppliance a has been developed and desi-gned in such a way that dangers](https://reader035.vdocuments.us/reader035/viewer/2022081616/5ff9488929118626e4663186/html5/thumbnails/40.jpg)
Dürr NDT GmbH & Co. KGDivision DÜRR MEDICALHöpfigheimer Strasse 22 D-74321 Bietigheim-BissingenGermanyPhone: +49 7142 99381-0Fax: +49 7142 99381-299www.duerr-medical.de [email protected]