mahalakshmimahalakshmiengineeringcollege.com/pdf/eee/viisem/ei2311/unit 3.pdf · differential...
TRANSCRIPT
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
MAHALAKSHMI
ENGINEERING COLLEGE
TIRUCHIRAPALLI- 621213.
QUESTION BANK
SEMESTER : VII DEPARTMENT: EEE
SUBJECT NAME: BIOMEDICAL INSTRUMENTATION SUBJECT CODE: EI2311
UNIT 3- NON – ELECTRICAL PARAMETER MEASUREMENTS
PART A (2 Marks)
1. Define cardiac output.( NOV 2007)
Cardiac output (Q or or CO ) is the volume of blood being pumped by the heart, in particular
by a left or right ventricle in the time interval of one minute. CO may be measured in many ways,
for example dm3/min (1 dm
3 equals 1 litre). Q is furthermore the combined sum of output from the
right ventricle and the output from the left ventricle during the phase of systole of the heart. An
average resting cardiac output would be 5.6 L/min for a human male and 4.9 L/min for a female
2. What is the pH value of arterial blood & venous blood? .( NOV 2007)
Arterial Blood: 7.37 – 7.44
Venous Blood : 7.35 – 7.45
3. Give the different types of oxygenators used in heart lung machine.(NOV 2011)
Bubble oxygenator
Film oxygenator
Foam oxygenator
Screen oxygenator
Blood film over sponge
Membrane oxygenator
Liquid- liquid oxygenator
4. How does the pH value determine the acidity or alkalinity in blood fluid? (NOV 2011)
The normal pH of the extracellular fluid lies in the range of 7.35 to 7.45, indicating that the body
fluid is slightly alkaline. When the pH exceeds 7.45, the body is considered to be in the state of
alkalosis. A body pH below 7.35 indicates acidosis. Both acidosis & alkalosis are disease
conditions widely encountered in clinical medicine.
5. List the applications of biotelemetry. (MAY 2012)
Bio-telemetry helps us to record the biosignals over long periods and while the patient is
engaged in his normal activities.
The medical attendant or computer can easily diagonise the nature of disease by seeing
the telemetered biosiganls without attending the patient’s room.
Patient is in his room without any mechanical or physical disturbance during recording by
means of biotelemetry.
For future reference or to study the treatment effect, the biotelemetry is the essential one.
For recording on animals, particularly for research, the biotelemetry is greatly used.
For monitoring the persons who are in action, the biotelemetry is an ideal one.
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
6. What are the advantages of MRI scan? (MAY 2012)
A conventional X ray scanner can produce an image only at right angles to the axis of the body,
whereas the MRI scanner produce the image at desired cross section, which offers a distinct
advantage to & is big boon for the radiologists. Its advantages are,
Superior contrast resolution
Direct multiplanar imaging, slices in the sagitial, coronal and oblique directions ca be
obtained easily.
No harmful radiation like X rays, Y rays, positions.
7. What is the principle of plethysmography & how many types are available? (NOV 2012)
Instrument measuring volume changes or providing output that can be related to them are called
plethysmographs & the measurement of these volume changes or phenomenon related to them
is called plethysmography. Such an instrument consists of a rigid cup or chamber placed over the
limb or digit in which the volume changes are to be measured. The cup is tightly sealed to the
member to be measured so that any changes in volume in the limb or digit reflect as pressure
changes inside the chamber. Either fluid or air can be used to fill the chamber. Types are,
capacitance plethysmograph
pseudo plethysmograph
mercury strain gauge
photoelectric plethysmograph
impedance plethysmograph
rheoencephalography
oculo pneumo plethysmography
8. List few applications of gas analysis. (NOV 2012)
To measure pH of blood
pCO2 & pO2 of blood
9. What is plethysmograph? (MAY 2013)
Same as Q7.
10. What are the automated indirect methods of blood pressure measurement? (MAY 2013)
Automatic blood pressure measuring apparatus using Korotkoff’s method
The Rheographic Method
Differential Auscultatory Technique
11. What are the limitations of capacitive sensor?
Inadequate for measuring most physiological variables because of their low frequency
components
12. What is the principle of piezoelectric sensors?
The piezoelectric material generates an electric potential when mechanically strained. Conversely
an electric potential can cause physical deformation of the materials.
13. What are the applications of piezoelectric sensors?
in cardiology
In phonocardiology
in blood pressure measurement
in measuring physiological accelerations
14. What are the different thermal sensors?
Thermocouples
Thermistors
Radiation sensors
Fiber optic detectors
15. What are the different radiation sources?
Tungsten lamp
Fluorescent lamp
LED s
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
LASERS
16. What are the different radiation sensors?
thermal sensors
Quantum sensors
Photo emissive sensors
Photo conductive cells
Photo junctions sensors
Photo voltaic sensors
17. What is a filter?
A filter is often a frequency selective circuit that passes a specified band of frequencies and
blocks or attenuated signal of frequencies outside this band.
18. List the different types of filters.
Analog or digital filters
Passive or active filters
Audio (AF) or radio (RF) filters.
19. Specify the advantages of an active filter
Gain and frequency adjustment flexibility
No loading problem
Low cost
20. Mention the factors considered while selecting a transducer.
Operating range
Sensitivity
Frequency response & resonant frequency
Environmental compatibility
Minimum sensitivity
Accuracy
Usage and ruggedness
Electrical parameters
21. What is meant by POT?
POT is a resistive potentiometer used for the purpose of voltage division. It consists for a resistive
element provided with a sliding contact called as wiper.
22. Explain the working principle of a strain gauge.
Strain gauge works on the principal that the resistance of a conductor or a semiconductor changes when
strained. This property can be used for measurement of displacement, force and pressure.
23. Name the different types of strain gauges.
Un-bonded metal strain gauge
Bonded metal wire strain gauge
Bonded metal foil strain gauge
Vacuum deposited thin metal film strain gauge
Sputter deposited thin metal strain gauge
Bonded semiconductor strain gauge
Diffused metal strain gauge.
24. Write notes on LVDT?
It is the linear variable differential transformer which is used to translate the linear motion into
electrical signals. It consists of a single primary winding and secondary winding.
25. List the advantages of LVDT?
High range of displacement measurement
Friction & electrical isolation , Ruggedness
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
Immunity from external effects
High I/p and high sensitivity
Low hysterisis & low power consumption.
26. Define operational amplifier?
It is the high gain dc differential amplifier
It is normally used in circuits that have characteristics determined by external negative
feedback networks.
27. What are the different applications of op-amp?
integrator
Differentiator
summing amplifier
Differential amplifier
Rectifier
log amplifier
28. What is heart block?
If he normal heart conduction system is disturbed, then the beat rate will be slower than the
normal rate. This state is known as heart block.
29. Classify the different types of heart block?
First degree AV block
Second degree AV block
Third degree AV block
Adam –stokes attack
Bundle block
Atria fibrillation
Ventricular fibrillations
30. Name the parts of heart conduction system?
Sino arterial node
Atria ventricular node
Bundle of his
Purkinje fibers
31. What is the color coding of the differential leads?
White – RA
Black – LA
Green - RL
Red - LL
Brown – chest
PART B (8,16 marks)
1. Draw the block diagram of automated electro sphygmomanometer for blood pressure
measurement & explain its operation. .( NOV 2007)
INDIRECT METHODS OF BLOOD PRESSURE MEASUREMENT
The classical method of making an indirect measurement of blood pressure is by the use of a cuff
over the limb containing the artery. This technique was introduced by Riva- Rocci for the
determination of systolic & diastolic pressure. Initially, the pressure in the cuff is raised to a level
well above the systolic pressure so that the flow of blood is terminated completely. Pressure in
the cuff is released at a particular rate. When it reaches a level, which is below the systolic
pressure, a brief flow occurs. If the cuff pressure is allowed to fall further, just below the diastolic
pressure value, the flow becomes normal & uninterrupted.
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
The problem here finally reduces to determining the exact instant at which the artery opens &
when it is fully opened. The method given by Korotkoff & based on the sounds produced by flow
changes is the one normally used in the conventional sphygmomanometers. The sounds firat
appears when the cuff pressure falls to just below the systolic pressure. They are produced by
the brief turbulent flow terminated by a sharp collapse of the vessel & persist as the cuff pressure
continues to fall. The sounds disappear or change in character at just below the diastolic pressure
when the flow is no longer interrupted. These sounds are picked up by using a microphone
placed over an artery distal to the cut off. The sphygmomanometric technique is an ausculatory
method, it depends upon the operator recognizing the occurrence & disappearance of the
Korotkoff sounds with variations in cuff pressure.
Automatic blood pressure measuring apparatus using Korotkoff’s method:
The method consists in putting a cuff around the upper part of the patient’s arm & applying a
microphone over the brachial artery. The compressed air required for inflating the cuff is provided
by a pumping system incorporated in the apparatus. Usually the inflating is done to a preset
pressure level, well beyond the systolic value at the rate of approximately 30mmHg/s. the
pressure in the cuff is then decreased at a relatively slow pace at the rate of 3-5mmHg/s. the cuff
is to be applied in such a way that the veins are not occluded.
While air is allowed to leak from the cuff, the korotkoff sounds are picked up by a special
piezoelectric microphone. The corresponding electrical signals are fed to a preamplifier. The
amplified signals are then passed on to a band pass filter having a bandwidth of 25 – 125 Hz.
With this pass band, a good signal to noise ratio is achieved when recording korotkoff sounds
from brachial artery beneath the lower edge of the cuff. The system is so designated that tye
appearance of the first korotkoff sound switches in the systolic manometer & locks the reading on
the indicating meter. In the similar way, diastolic value is fixed by the last korotkoff sound. The
cuff is completely deflated, automatically, after an interval of 2 – 5s after the determination of the
diastolic value.
Instruments operating on this principle are prone to error due to artefacts. On e methods of
avoiding this is to design a control system in such a way that when pressure is registered, the first
sound must be followed by the second one within a pre set interval. If this is not the case, then
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
the recorded value is automatically cancelled & the measurement starts again with subsequent
sounds.
A complete cycle of measurement consists of cuff pumping, controlled deflation, picking up &
evaluation of the korotkoff sounds, fixing of systolic & diastolic pressure & then a complete
deflation of the cuff. The cycle is initiated by the time delay & then operation is controlled by the
command pulse.
2. Explain with relevant equations the working & measurement procedure of plethysmograph.
( NOV 2007)
PLETHYSMOGRAPHY
Instrument measuring volume changes or providing output that can be related to them are called
plethysmographs & the measurement of these volume changes or phenomenon related to them
is called plethysmography. Such an instrument consists of a rigid cup or chamber placed over the
limb or digit in which the volume changes are to be measured. The cup is tightly sealed to the
member to be measured so that any changes in volume in the limb or digit reflect as pressure
changes inside the chamber. Either fluid or air can be used to fill the chamber.
Plethysmograph can be designed for constant pressure or constant volume within the chamber.
In either case some form of pressure or displacement transducer must be included to respond to
pressure changes within the chamber & to provide a signal that can be calibrated to represent the
volume of the limb or digit. This type of plethysmograph can be used in two ways. If the cuff,
placed upstream from the seal, is not inflated, the output signal is simply a sequence of
pulsations proportional to the individual volume changes with each heartbeat.
The plethysmograph illustrated above can be used to measure the total amount of blood flow into
the limb or digit being measured. By inflating the cuff to a pressure just above venous pressure,
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
arterial blood can flow past the cuff, but the venous blood cannot leave. The result is that the limb
or digit increases its volume with each heartbeat by the volume of the blood entering during that
beat. The output tracing for this is shown in the figure. The slope of the line along the peaks of
these pulsations represents the overall rate at which the blood enters the limb or the digit from the
accumulation of blood that cannot escape.
Another device quite close to the true plethysmograph is called the capacitance
plethysmograph. In this device, which is generally used either on arm or leg, the limb in which
the volume is being measured becomes one plate of the capacitor. The other plate is formed by a
fixed screen held at a small distance from the limb by an insulating layer. Pulsations of the blood
in the arm or leg cause variations in the capacitance, because the distance between the limb &
the fixed screen varies with these pulsations. Since the length of the cuff is fixed, the variations in
capacitance can be calibrated as volume changes.
Another type is the pseudo plethysmograph, which measures changes in diameter at the cross
section of a finger, toe, arm leg or other segment of the body. Since volume is related to
diameter, this type of device is accurate for many purposes.
A common method of sensing diameter changes is through use of the mercury strain gauge,
which consists of a segment of small diameter elastic tubing, long enough to wrap round the limb
or digit being measured. When the tube is filled with mercury, it provides a highly compliant strain
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
gauge that changes its resistance with changes in diameter. With each pulsation of the
blood that increases the diameter of the limb or digit, the strain gauge elongates & in stretching
becomes thinner thus increasing the resistance.
Another type is photoelectric plethysmograph. This device operates on the principle that volume
changes in a limb or digit results in changes in optical density through & just beneath the skin
over a vascular region. A light source in an opaque chamber illuminates the small area of the
finger tip or the other region to which the transducer is applied. Light scattered & transmitted
through the capillaries of the region is picked up by the photo cell, which is shielded from all other
light. As the capillaries fill with blood, the blood density increases thereby reducing the amount of
light reaching the photocell. The result causes resistance changes in the photocell that can be
measured on a Wheatstone bridge & recorded.
The more reliable type is the impedance plethysmograph, in which volume changes in a segment
of a limb or digit are reflected as impedance changes. The impedance changes are primarily due
to the changes in conductivity of the current path with each pulsation of blood. This is done by
using either a two or four electrode system. The electrodes are either conductive bands wrapped
around the limb or digit to be measured or simple conductive strips of tape attached to the skin. In
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
either case, the electrodes contact the skin through a suitable electrolyte jelly or paste to form an
electrode interface & to remove the effect of skin resistance. In two electrode system a constant
current is forced through the tissue between the two electrodes & the resulting volume changes
are measures. In the four electrode system, the constant current is forced through two outer or
current electrodes & the voltage between the two inner electrodes is measured.
Several theiry explains the actual cause of the measured impedance changes. One is that the
presence of blood filling segment of the body lowers the impedance of that segment. The second
theory is that the increase in diameter due to additional blood in a segment of the body increases
the cross sectional area of the segment’s conductive path & thereby lowers the resistance of the
path.
A special form of impedance plethysmography is rheoencephalography, measurement of
impedance changes between the electrodes placed on the scalp. This technique provides
information related to cerebral blood flow & is sometimes used to detect circulatory differences
between the two sides of the head.
Another special type of plethysmograph is the oculo pneumo plethysmography. This instrument
measures every minute volume changes that occur in the eye with each arterial blood pulsation.
A small eye cup is placed over the sclera of each eye & is connected to a transducer positioned
over the patient’s head by a short section of flexible tubing. A vaccum, which can be varied from
zero to 300mmHg is applied to hold the eye cups in place. Pulsations are recorded on two
channels of a three channel pen recorder, one for each eye. The third channel is used to record
the vacuum. By periodically allowing the vaccum to build up to -300mmHg and deplete to zero,
the instrument can also be used as recording suction opthalmodynamometer, an instrument used
for measuring arterial blood pressure within the eye.
3. What are the methods for measuring blood pressure? Draw a typical set up & explain. (NOV
2011)
Same as Q1
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
4. Explain the measurement of heart sound with suitable diagram. (NOV 2011)
MEASUREMENT OF HEART SOUND
The technique of listening to sounds produced by the organs & vessels of the body is called
auscultation. The heart sound heard by the physician through his stethoscope occurs at the time
of closure of major valves in the heart. This timing could easily lead to false assumption that the
sounds which are heard are primarily caused by the snapping together of the vanes of these
valves. The principal cause for heart sound is due to vibrations set up in the blood inside the
heart by sudden closure of the valves. These vibrations together with eddy currents induced in
the blood as it is forces through the closing valves, produce vibrations in the walls of the heart
chambers & in adjoining blood vessels.
With each heart beat, the normal heart produces two distinct sounds that are audible in the
stethoscope, often described as LUB- DUB. The lub is caused by the closure of atrioventricular
valves. Normally this is called first heart sound 7 it occurs approximately at the time of the QRS
complex of the ECG & just before ventricular systole. The dub part of the heart sound is called
the second heart sound & is caused by the closing of semilunar valves. These valves close at the
end of systole, just before atrioventricular valves reopen. This second heart sound occurs about
the time of the end of the T wave of the ECG.
A third heart sound is sometimes heard in young adults. This sound occurs for 0.1 – 0.2 sec after
the second sound due t the rush of blood from the atria into the ventricles. This sound actually
preceeds atrial contraction, which means that the inrush of blood to the ventricles causing the
sound is passive, pushed only by the venous pressure at the inlets to the atria.
An atrial heart sound, which is not audible but visible on graphic recording occurs when the atria
actually do contract, squeezing the remainder of the blood into the ventricles. The inaudibility of
this heart sound is a result of the low amplitude & loe frequency of the vibrations.
The fig shows the time relationships between the first, second,& third heart sound with respect to
ECG.
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
In abnormal hearts, additional sounds called murmurs are heard between the normal heart
sounds. Murmurs are generally caused either by improper opening of valves or by regurgitation,
which results when the valves do not close completely & allow some backward flow of blood. The
sound is due to high velocity blood flow through small opening. Another cause of murmurs can be
due to small opening in the septum, which separates the left side & right side of the heart. In this
case the pressure difference between the two sides of the heart force blood through the opening,
usually from the left ventricle into the right ventricle, bypassing systemic circulation.
Normal heart sound are quite short in duration, approximately on eteh=nth of a second for each,
while murmurs extend between the normal sounds.
There is also a difference in frequency range between normal & abnormal heart sounds. The first
sound is composed primarily of energy in the 30-45 Hz range. The second heart sound is usually
of higher in pitch than the first with maximum energy in the 50 – 70 Hz range. The third heart
sound is extremely weaker in vibration, with most of the energy below 30 Hz. Murmurs on the
other hand produce much higher pitched sounds with energy in 100 – 600 Hz. Although
asculcation is the method of detecting 7 analyzing heart sounds, the graphic recording of heart
sounds is called phonocardiogram.
An entirely different waveform is produced by the vibrations of the heart against the thoracic
cavity. The vibrations of the side of the heart as it thumps against the chest wall form the
vibrocardiogram, whereas the tip or apex of the heart hitting the rib cage produces the apex
cardiogram.
Sounds & pulsations can also be detected & measured at various locations in the systemic
arterial circulation system where the major arteries approach the surface of the body. The most
common one is the pulse, which can be felt with finger tips at certain points on the major arteries.
The waveform of this pulse can also be measured 7 recorded. In addition, when an artery is
partially occluded so that the blood velocity through the constriction is increased sufficiently,
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
identifiable sounds can be heard downstream through the stethoscope. These sounds are called
as Korotkoff sounds commonly used in measurement of blood pressure. Ballistocardiogram is
related to the measures of direct result of dynamic forces of the heart as it beats & pumps into the
major arteries.
5. Explain the measurement of blood PO2 & PCO2. (NOV 2011)
BLOOD GAS ANALYZERS
Blood gas analyzers are used to measure the pH, partial pressure of CO2 (pCO2), partial pressure
of O2(O2) of the body fluids with special reference to human blood. The measurements of these
parameters are essential to determine the acid base balance.
The normal pH of the extracellular fluid lies in the range of 7.35 to 7.45, indicating that the body
fluid is slightly alkaline. When the pH exceeds 7.45, the body is considered to be in the state of
alkalosis. A body pH below 7.35 indicates acidosis. Both acidosis & alkalosis are disease
conditions widely encountered in clinical medicine. Any tendency of the pH of the blood to deviate
towards these conditions is dealt with the following three physiological mechanisms: (i) buffering
by chemical means, (ii) respiration and (iii) excretion, into the urine by kidneys.
The blood & tissue fluids contain chemical buffers, which react with added acids & bases &
minimize the resultant changes in hydrogen ions & they respond to changes in carbon dioxide
concentration in seconds. Respiration can adjust sudden changes in carbon dioxide back to
normal in few seconds. Kidney requires many hours to readjust hydrogen ion concentration by
excreting highly acidic or alkaline urine to enable body conditions to return to normal. In order to
maintain pO2, pCO2 & pH within normal limits, the rate & depth of respiration vary automatically
with changes in metabolism.
pCO2 Measurement:
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
the emf generated by a pCO2 electrode is a direct logarithmic function of pCO2. It is observed
that a ten fold change in pCO2 causes the potential to change by 58 ±2mV. The pH versus log
pCO2 relationship is linear within ±0.002pH unit from 1 to 100% carbon dioxide. Since 0.01 unit
pH change corresponds to a 2.5% change in pCO2 or 1 mmHg in 40mmHg, for achieving an
accuracy of 0.1mmHg, it is desirable to read 0.001pH unit. This order of accuracy can be read
only on a digital readout type pH meter or on an analog meter with expanded scale. The
instrument should have a very high degree of stability & a very low drift amplifier. It is essential to
maintain the temperature of the electrode of the electrode assembly constant within close limits.
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
6. Explain the Rheographic method of blood pressure measurement. (MAY 2012)
The Rheographic Method:
A fully automated apparatus for measuring systolic & diastolic blood pressure has been
developed using the ordinary Riva- Rocci cuff & the principle of rheographic detection of an
arterial pulse. Here, the change in impedance at the two points under the occluding cuffs forms
the basis of detection of the diastolic pressure
In this method, a set of three electrodes which are attached to the cuff are places in contact with
the skin. A good contact is essential to reduce the skin electrode contact impedance. Electrode B
which acts as a common electrode is placed slightly distal from the midline of the cuff. Electrode
B& C are placed at a certain distance from the electrode B, one distally & the other proximally. A
high frequency current source of 100KHz is connected to the electrodes A& C. When we
measure the impedance between any two electrodes before pressurizing the cuffs, it shows
modulation in accordance with the blood flow pulsations in the artery. Therefore the arterial
pulses can be detected by the demodulation & amplification of the modulation
When the cuff is inflated above the systolic value, no pulse is detected by the electrode A. the
pulse appears when the cuff pressure is just below the systolic level. The appearance of the first
distal arterial pulse results in an electrical signal, which operates a valve to fix a manometer value
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
on the systolic value. As long as the pressure in the cuff is between the systolic & diastolic
values, differential signal exists between electrode A & C. this is because the blood flow is
impeded underneath the occluding cuff & the pulse appearing at the electrode A is time delayed
from the pulse appearing at electrode C. when the cuff pressure reaches the diastolic pressure,
the arterial blood flow is no longer impeded & the differential signal disappears. A command
signal is then initiated & the diastolic pressure is indicated on the manometer.
7. Explain with functional diagram the working of spirometer. (MAY 2012)
Same as Q12
8. Explain the measurement methods of Galvanic skin response (GSR) & Basal Skin Resistance
(BSR). (MAY 2012)
9. Explain in detail with neat diagram, differential ausculatory technique of blood pressure
measurement. (MAY 2012)
Differential Auscultatory Technique:
The differential auscultatory technique is a non invasive method for accurately measuring blood
pressure. A special cuff mounted sensor consisting of a pair of pressure sensitive elements,
isolates the signal created each time the artery is forced open.
The figure shows how high frequency pulses are created each time, the intra arterial pressure
exceeds the cuff pressure. As long as the cuff pressure exceeds the pressure in the artery, the
artery is held closed & no pulse is generated. However, as soon as the intra arterial pressure
rises to a value, which momentarily exceeds the cuff pressure, the artery snaps open & a pulse
is created. Once this artery is open, blood flows through it giving rise to the low frequency
pressure wave signal, which lasts until the arterial pressure again drops below the cuff pressure.
The process is repeated until the cuff pressure drops to a value below the diastolic.
The figure is a cut way view of an arm with the partially occluding the brachial artery. Each time
the artery opens, the signal shown in the fig is created. The signal consists of a slow rising. Low
frequency component with a fast pulse superimposed on it., this signal is marked by arrows
marked A in the fig transmitted from the artery to both the sensor & the airbag in the cuff.
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
Due to air bag characteristics, the high frequency component is g=highly attenuated, leaving only
the low frequency signal is transmitted to the side of the sensor facing the airbag, as denoted by
arrow marked B in the fig.
The systolic pressure is determined as the pressure at which the first opening of the artery
occurs, shown by the first pulse in the figure. Because this pulse is created the first time the
artery is forced open by intra arterial pressure. Similarly, diastolic value is determined as the
pressure at which the differential signal essentially disappears, because this corresponds to the
last time the artery is forced open. The differential sensor subtracts side B signal from the side A
signal, therby cancelling out the pressure wave component & the motion artifact signals, 7 the
higher frequency Korotkoff signals are isolated.
10. Explain the fick’s method for cardiac output measurements. (NOV 2012)
11. Describe the need for blood pH measurement. (NOV 2012)
BLOOD GAS ANALYZERS
Blood gas analyzers are used to measure the pH, partial pressure of CO2 (pCO2), partial pressure
of O2(O2) of the body fluids with special reference to human blood. The measurements of these
parameters are essential to determine the acid base balance.
The normal pH of the extracellular fluid lies in the range of 7.35 to 7.45, indicating that the body
fluid is slightly alkaline. When the pH exceeds 7.45, the body is considered to be in the state of
alkalosis. A body pH below 7.35 indicates acidosis. Both acidosis & alkalosis are disease
conditions widely encountered in clinical medicine. Any tendency of the pH of the blood to deviate
towards these conditions is dealt with the following three physiological mechanisms: (i) buffering
by chemical means, (ii) respiration and (iii) excretion, into the urine by kidneys.
The blood & tissue fluids contain chemical buffers, which react with added acids & bases &
minimize the resultant changes in hydrogen ions & they respond to changes in carbon dioxide
concentration in seconds. Respiration can adjust sudden changes in carbon dioxide back to
normal in few seconds. Kidney requires many hours to readjust hydrogen ion concentration by
excreting highly acidic or alkaline urine to enable body conditions to return to normal. In order to
maintain pO2, pCO2 & pH within normal limits, the rate & depth of respiration vary automatically
with changes in metabolism.
Blood pH measurement:
The acidity or alkalinity of a solution depends on the concentration of hydrogen ions. The
increased concentration of hydrogen ions makes the solution more acidic & decreased
concentration makes it more alkaline. pH is the measure of hydrogen ion concentration
expressed logarithmically. It is the negative exponent (log) of the hydrogen ion concentration.
pH= -log(H*)
if the number 10-7
represents the concentration of hydrogen ions in a certain solution, then its pH
would be 7. As hydrogen ion increases, pH falls because logarithm gets smaller & vice versa. pH
of +7 is considered as neutral solution, pH of +6 represents acid & pH of +8 represents alkali.
Concentration of ions is expressed in mols/ litre.
Electrochemical pH determination uses the difference in potential occurring between solutions of
different pH separated by special glass membrane. The device used to this effect is called glass
electrode.
Glass Electrode:
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
The potential of the glass electrode can be written by means of Nernst equation:
Where, Eo=standard potential, R=gas constant, T=absolute temperature, F=faraday constant,
∆pH= pH value deviation from 7.
The above equation shows that the emf developed in the electro chemical pH cell is a linear
function of ∆pH. Change of pH of one unit=58.2mV @20oC=62.2mV @40
oC
The factor -2.3036 RT/F is called the slope factor & it depends on the solution temperature. With
one degree change in temperature, emf changes by 0.2mV. the figure shows the relation
between pH & emf at different temperatures. The reference electrode provides a constant
potential against which the potential af the indicator or the glass electrode is measured. Almost
universally employed reference electrode is the saturated calomel electrode.
pH Measurement:
For making pH measurements, the solution is taken in a beaker. A pair of electrodes, one glass
or indicating electrode & the other reference calomel electrode is immersed in the solution. The
voltage developed across the electrodes is applied to an electronic amplifier, which transmits the
amplified signal to the display. The pH meter has controls for calibration & temperature
compensation. The glass electrode exhibits a high degree of electrical resisitance on the orser of
100 – 1000Mohms. The emf measurement necessitates the use of measuring circuits with high
input impedance.
Electrodes for blood pH measurement:
The common type used is the syringe electrode, preferred for taking small samples of blood an
aerobically. The small dead space between the electrode bulb & the inner surface of the syringe
barrel is usually filled with dilute heparin solution to prevent blood coagulation.
Microcapillary glass electrodes are preferred when it is required to monitor pH continuously, for
example during surgery
Microelectrode for clinical applications requires only 20 – 25 microlitre of capillary blood for the
determination on pH. The electrode is enclosed in a water jacket with circulating water at a
constant temperature of 38 degrees. The water contains 1% NaCl for shielding against static
interference. The capillary is protected with polyethylene tubing. The internal reference electrode
is silver/ silver chloride & the calomel reference electrode is connected to a small pool of
saturated KCl, through a porous pin. An accuracy of 0.001pH can be obtained with this electrode
against a constant buffer. The figure shows the constructional details of a typical blood pH
electrode & the measurement set up used in practice.
Glass electrodes deteriorate if allowed to remain in contact with the bl;ood for a long time. This
results in the change of emf – pH slope. Therefore , as a precautionary measure, in an apparatus
where blood necessarily remains in contact around the electrode for long time, the response must
be checked frequently against the buffer solution. The poisoning effect can be reduced by putting
the electrode in pepsin & 0.1 N Hcl followed by careful wiping with the tissue paper.
The pH of blood is found to change linearly with temperature in the range of 18 to 38 degree C.
the temperature co efficient for the pH of blood is 0.0147 pH unit per degree C.
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
Buffer solutions are primarily used for (i) creation & maintenance of a desired, stabilized pH in a
solution & (ii) standardization of electrode chains for pH measurements. The presence of a buffer
in a solution is to counteract pH changes in the solution caused by the addition or removal of
hydrogen ions.
12. Explain the working of spirometer. (NOV 2012)
SPIROMETER
The instrument used to measure lung capacity & volume is called spirometer. The record
obtained from this is called spirogram. Spirometers are calibrated containers that collect gas &
make measurements of lung volume or capacity that can be expired. The gas analyzer combined
with a spirometer makes a complete pulmonary testing system.
Basic Spirometer:
Respiratory measurements are carried out by using classic water sealed spirometer. This
consists of an upright water filled cylinder containing an inverted counter weighted bell. Breathing
into the bell changes the volume of gases trapped inside, and the change in volume is translated
into vertical motion, which is recorded on the moving drum of a kymograph. The excursion of the
bell will be proportional to the tidal volume. For most purpose, the bell has a capacity of the order
of 6 – 81. The spirometer cannot respond to rapid breathing unless a light weight bell is
provided.frequency response of the spirometer must be adequate for the measurement of the
forced expiratory volume. The instrument should have no hysteresis. The same volume should be
reached whether the spirometer is being filled or being emptied to that volume.
As the water filled spirometer includes moving masses in the form of the bell 7 conterweights, this
leads to the usual problems of inertia & possible oscillation of the bell. This can lead to
overestimation of expiratory volume. So a spirometer having a bell of large diameter & which
closely fits into the central core is used. So the area of water covered by the bell is small in
relation to that of water tank.
The spirometer is a mechanical integrator since the input is airflow & the output is volume
displacement. An electrical signal proportional to volume displacement can be obtained by using
a linear potentiometer connected to the pulley portion of the spirometer. The spirometer is
aheavily damped device so that small changes in inspired & expired air volumes are not
recorded. The spirometer can be fitted with a linear motion potentiometer, which directly converts
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
spirometer volume changes into an electrical signal. The signal may be used to feed a flow
volume differentiator for the evaluation & recording of data. The response is usually ±-1% to 2Hz
±10% to 10Hz..
Tests made in spirometer are not analytical & not completely objective becoz the results are
dependent on the cooperation of the patient & the coaching efforts of good respiratory technician.
Transducers have been designed to transform the movement of the bell, bellows or piston of
volume spirometers into electrical signals. These are then used to compute the numerical value
electronically.
Wedge Spirometer:
A wedge spirometer consists of two square pane. Parallel to each other & hinged along one
edge. The first pan is permanently attached to the wedge casting stand & contains a pair of 5cm
inlet tubes. The other pan swings freely along its hinge with respect to the fixed pan. A space
existing between the two pans is sealed airtight with vinyl bellows. The bellow is flexibile in
direction of pan motion but offers high resistance to ballooning or inward & outward expansion
from the spirometer. So when a pressure gradient exists between the interior of the wedge & the
atmosphere, there will only be a negligible distortion of bellows.
As gas enters or leaves the wedge, the moving pan will change position in compensation for this
change in volume. The construction of the wedge is such that the moving pan will respond to very
slight changes in volume. Under normal conditions the pressure gradient that exists between the
wedge & the atmosphere amounts to a fraction of millimeter of water. Volume & floew signals of
the wedge are obtained independently from two linear transducers. The transducer are attached
to the fixed frame & are coupled to the edge of the moving pan. One transducer produces a dc
signal proportional to displacement ( volume), while the other has the dc output proportional to
the velocity (flow).
The transducer out[ut are connected to an electronic circuit, which contains the power supply, an
amplifier & built in calibration networks.
A pointer attached to the moving pan & a scale affixed to the frame, combine to provide a
mechanical read out for determining the approximate volume position of the spirometer. When
one open to the atmosphere standing upright, the wedge will empty itself due to the force of
gravity acting on a moving member. An adjustsble tilt mechanism provides the means for
changing the resting point of the moving pan to any desired volume poin. An adjustable magnetic
stop insures a more highly defined resting position.
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
Thus the relatively small forces due to gravity & magnetic stop are overcome by a negligible rise
in pressure in patient’s lungs. When the gravitational return of the moving pan to the resting
position is undesirable, the wedge may be turned on its side so that at any point the pan will be in
state of equilibrium. The wedge may be calibrated with a selector switch, which determines the
magnitude of the calibration signal. The flow calibration signals for each particular wedge are
adjusted using special fixtures.
Ultrasonic Spirometer:
Ultrasonic transducers depend for their action on transmitting ultrasound between a pair of
transducers & measuring changes in transit time caused velocity of intervening fluid medium.
They employ piezoelectric transducers & are operated at their characteristic resonant frequency
for high efficiency. Gas floe meters generally operate in the range from about 40 – 200KHz. At
frequencies higher than 200KHz, the absorption losses in the gas are very high.
Ultrasonic spirometers utilize a pair of ultrasonic transducers mounted on opposite sides of a flow
tube. The transducers are capable of both transmitting & receiving ultrasonic pulses. In
conventional ultrasonic flow meters, pulses are transmitted through the liquid or gas in the flow
tube, against & then with the direction of flow. The pulse transit time upstream t1 and downstream
t2 can be expressed,
and
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
Where D is the distance between the transducers, C is the velocity of sound propagation in the
fluid & v’ is the fluid velocity vector along the path of the pulses. The average gas vel;ocity v
through the flow tube is a vector v’ so that,
V=v cosθ
The velocity of sound C does not appear in the final equation. Thus, the output accuracy is
unaffected by fluid density, temperature or viscosity. In gas flow measurements, pulmonary
function tubes larger than 3cm in diameter must be used. The single frequency systems that
measure time delay directly must be able to resolve nanoseconds since the total transit delay, t is
usually measured in microseconds. This technique is not easily implemented because of the
difficulty in measuring these small time differences.
13. Discuss a finger tip oxymeter. (NOV 2012)
OXIMETERS:
Oximeters are used to measure the percentage of oxygen saturation of the circulating arterial
blood.
Where [HbO2] is the concentration of oxygenated hemoglobin & [Hb] is the concentration of
deoxygenated hemoglobin. The measurement of oxygen saturation is mainly used in the
diagnosis of cardio respiratory functions. It is known that blood consists of red blood cells(RBC)
with density 4.62 – 6.2 millions/ microlitre & size 6.8 – 7.5 micrometer, white blood cells(WBC)
with density 0.004 – 0.011 millions/ microlitre & size 6 – 18 micrometer, platel;ets with density
0.15 – 0.40 millions/ microlitre & size 2 – 4 micrometer in liquid plasma. In plasma the oxygen is
dissolved about 0.3% & therefore the plasma is a very poor carrier of oxygen. Actually the oxygen
is carried by the red blood cells through the hemoglobin. Hemoglobin in the RBC combines with
oxygen & forms a compound called oxyhemoglobin which contains more oxygen. The amount of
oxygen combining with blood depends on the partial pressure of oxygen.
EI 2311 / BIOMEDICAL INSTRUMENTATION VII SEM EEE
Oximeter is divided into two types, vitro oximeter & vivo oximeter. In vitro oximetry, the blood is
taken out & measurement for oxygen saturation is made at later time in the laboratory. In vivo
oximetry the oxygen saturation is measured while the blood is flowing through the circulatory
system.
14. Explain the working of spirometer with the help of functional diagram. (MAY 2013)
Same as Q12
15. Explain the rheographic method of blood pressure measurement. (MAY 2013)
Same as Q 6