MAGNIFICATION(MACRORADIOGRAPHY/MICRORADIOGRAPHY)DISTORTION
PARTHA JYOTI DAS
MSc MIT
Contents
Macroradiography
Principles
Magnification with fixed FFD
Magnification with fixed FOD
Unsharpness with macroradiography
Applications
Microradiography
Mass miniature radiography
Distortion
References
The technique of producing an image by direct
magnification is called macroradiography.
Enables fine anatomical details than the original
and accurate diagnosis.
Macroradiograp
hy
Principles
Magnified image can be produced
by increasing the object to film
distance in which the x rays
diverging from the point source
will produce a directly magnified
image.
Relation:
The
relationship
between
size of
object and
it’s image.
FFD
FOD
OFD
Image Size
X-ray Focus
Film
Plane
Object
Plane
Principles
Magnification (M) can be calculated by the
following formula:
focus to film distance (FFD)
focus to object distance(FOD)
. The focus to object distance (or object to film
distance) is taken from the mid level or part.
M =
MAGNIFICATION WITH FIXED
FFD
Magnification is increased by bringing the object nearer to the x ray tube
FOD = FFD/Magnification.
For eg : With a fixed FFD of 100cm, if a magnification factor of 1.6 is required then;
FOD = 100/1.6 = 62.5 cm
MAGNIFICATION WITH FIXED
FOD
The required magnification is obtained by
moving the film away from the object.
The required OFD for a given magnification is
then calculated from
OFD = FOD(M – 1)
Eg. FOD is fixed at 100cm , given magnification
factor 1.6
OFD = 100(1.6 – 1) = 60cm
UNSHARPNESS WITH
MACRORADIOGRAPHY
Movement Unsharpness
Geometric Unsharpness
MOVEMENT UNSHARPNESS
Direct magnification by increased object to
film distance can be carried out if there is
complete immobilization of the patient
Any movement due to lack of immobilization or
involuntary movement will be magnified on the
radiograph due to increased OFD
Movement
unsharpness
Some steps
Use of immobilization devices, e.g. supports,
binders,sandbags and pads.
Instructions should be given to the patient to
remain still.
GEOMETRIC UNSHARPNESS
Occurs because the source of x rays is not a
point source and any distance between the
object and film will cause an image penumbra
in addition to magnifying the image
The geometrical unsharpness for given focal
spot size calculated by
Focal Spot Size x Object-Film
Distance
Ug =
Focus-Object Distance
Geometric unsharpness
How to minimize Ug:-
Focal Spot size:- By reducing Focal Spot
Size.
Object image Distance:- By reducing object
image distence.
Focus object Distance:-By increasing the
Focus object distence.
Relationship between focal spot
size and magnification
For a given focal spot size there is a limit to
magnify the image without significant loss of
details.
APPLICATIONS
Limited use of macroradiography
Dacrocystography of Skull (Lacrimal ducts after injection of contrast)
Imaging the carpal bones in cases of suspected fracture of scaphoid
Localised areas of the lung
Localised areas during angiography eg: cerebral angiography
Macroradiograph of wrist with a magnification
factor of 2
Microradiography
An imaging technique used to examine
very small objects or minute details, often
with the use of high-voltages, very small
focal spot size, and an ultrafine film
emulsion.
Ultrafine Emulsion film
Photo print made by these films performs rich
sharpness, high definition, strong scratch
resistance, as well as long lasting by pigment
Microradiography
It is also called as chest fluoroscopy.
The image on a fluorescent screen is photographed on a small film. Screen and film are separated and the image is focused on the film by a camera lens
Miniature radiographs (70mmx70mm or 100mmx100mm roll or cut films were used).
A permanent record is possible.
Mass Miniature Radiography(MMR)
In order to control tuberculosis, medical research council, London appointed their Committee on Tuberculosis in 1941.
The detection of symptomless or latent pulmonary tuberculosis by mass radiography had been the subject of much careful investigation.
An excellent instrument was designed by the council for this purpose and in 1943 the Ministry of Health announced that a limited number of mobile miniature radiography sets.
The equipment consist of
Fluorescent screen(emits blue or green light)
A mirror optics system on recording exposure
on film
Survey in progress
Control Panel
Apparatus for MMR
1.Power unit:- High tension unit
Four valve rectification
An output of from 100 to 400mA
A maximum kilovoltage of 100 kVp and 91 kVp
respectively
2. X-ray Tube
The rotating anode X-ray tube
Dual focus, Focal-spot sizes are 1mm and
2mm square.
An optical centring device.
Adjustable diaphragm
3.Fluorescent Screen and Grid:
16 by 16 inches fluorescent screen
Covered by a sheet of protective lead glass.
Yellow-blue or green type of screen
A stationary grid is also present which reduces scatter
4.The Camera.:
An electrically operated camera is linked with the X-ray exposure switch.
Lead glass protection is present on the camera side of screen
5.The lens: a number of different lenses
Developing MMR films
Special tanks meant for roll films on a spiral frame
Development time was 7min at 68 degree F
Viewing of MMR FILMS
Projector for viewing miniature radiographs
Magnifying glasses
DISTORTION
DISTORTION
Misrepresentation of the true size and
shape of an object is called distortion.
Distortion results from unequal
magnification of different parts of the
same object.
Foca
l
Spot
Film
•The object AB is
tilted with respect to
the plane of the film.
•The size and shape
of image vary with
the amount of tilting.
Distortion
A
B
A B
Distortion of the image of an object will be
different in different parts of the x-ray beam.
Distortion of thick object occurs if they are
not directly in the central part of the x-ray
beam.
When the part to be imaged does not lay parallel
with the IR (cassette).
If the Central Ray is not perpendicular to the part.
Caused by improper alignment object with relation to tube focus and film .
Distortion of an object will be different in different part of the x-ray beam.
Elongated Foreshortened Normal
GOOD RADIOGRAPHS
MAKES EVERYONE HAPPY