mabpac rp columns - assets.thermofisher.com · analytical format mabpac rp columns in 2.1 mm id and...

27
For Research Use Only. Not for use in diagnostic procedures. MAbPac RP Columns 065625 Revision 03 March 2017 User Manual

Upload: nguyendiep

Post on 08-Aug-2019

213 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

For Research Use Only. Not for use in diagnostic procedures.

MAbPac RP Columns

065625 Revision 03 • March 2017

User M

anual

Page 2: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

Thermo Scientific Product Manual for MAbPac RP Columns Page 2 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

Product Manual

for

MAbPac RP Analytical and Guard Columns 3 × 100 mm, (Item # 088644)

3 × 50 mm, (Item # 088645) 3 × 10 mm, (Item # 088646) 2.1 × 100 mm, (Item # 088647)

2.1 × 50 mm, (Item # 088648) 2.1 × 10 mm, (Item # 088649)

1.0 × 150 mm, (Item # 303184) 1.0 × 100 mm, (Item # 303183) 1.0 × 50 mm, (Item # 303182)

Page 3: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

Thermo Scientific Product Manual for MAbPac RP Columns Page 3 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

© 2017 Thermo Fisher Scientific Inc. All rights reserved.

All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.

Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is

strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.

The contents of this document are subject to change without notice. All technical information in this document is

for reference purposes only. System configurations and specifications in this document supersede all previous

information received by the purchaser.

Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might result from

any use of this document, even if the information in the document is followed properly.

This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This

document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of

Sale shall govern all conflicting information between the two documents.

For Research Use Only. Not for Use in Diagnostic Procedures.

Revision History:

Revision 2, March, 2017, Added information for 1 mm column format.

Revision 3, October, 2017, Revised information for 1 mm column format.

Page 4: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

Thermo Scientific Product Manual for MAbPac RP Columns Page 4 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

Safety and Special Notices

Make sure you follow the precautionary statements presented in this guide. The safety and other

special notices appear in boxes.

Safety and special notices include the following:

Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

Indicates a potentially hazardous situation which, if not avoided, could result in damage to equipment.

Indicates a potentially hazardous situation which, if not avoided, may result in minor or

moderate injury. Also used to identify a situation or practice that may seriously damage the instrument, but will not cause injury.

Indicates information of general interest.

IMPORTANT Highlights information necessary to prevent damage to software, loss of data, or invalid test results; or might contain information that is critical for optimal performance of the

system.

Tip Highlights helpful information that can make a task easier.

SAFETY

!

WARNING

!

CAUTION

!

NOTE

!

Page 5: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

Contents

Thermo Scientific Product Manual for MAbPac RP Columns Page 5 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

Contents

1. Introduction ...................................................................................................................... 7

1.1 Introduction to the MAbPac RP Column ........................................................................... 7

1.2 MAbPac RP Operating Limits and Specifications ............................................................... 8

1.2.1 Operating Conditions .................................................................................................................................. 8

1.2.2 Physical Characteristics .............................................................................................................................. 8

1.3 Formats of the MAbPac RP Columns ............................................................................... 9

2. Getting Started; Step-By-Step Procedure...................................................................... 10

2.1 Step 1 – Visually inspect the column ...............................................................................10

2.2 Step 2 – Mobile Phase Selection .....................................................................................10

2.3 Step 3 – Set up the LC system ........................................................................................10

2.4 Step 4 – Condition the column ........................................................................................11

2.5 Step 5 – Verify the performance of the column..................................................................11

2.6 Step 6 – Real sample analysis .........................................................................................11

3. Column Care................................................................................................................... 12

3.1 Column storage ............................................................................................................12

3.2 Operating pH range: pH 0 to 14 .....................................................................................12

3.3 Operating temperature limit: up to 110 ºC ........................................................................12

3.4 Pressure limit: 4000 psi .................................................................................................12

3.5 Flow rate .....................................................................................................................12

3.6 Loading capacity ..........................................................................................................13

3.7 Ruggedness .................................................................................................................14

3.8 Column washing procedure ............................................................................................15

4. Example Applications ..................................................................................................... 16

4.1 Optimization for protein separation .................................................................................16

4.2 Separation of mAb and mAb fragments............................................................................17

4.3 LC/MS analysis of mAb ................................................................................................18

4.4 LC/MS analysis of mAb fragments..................................................................................19

4.5 LC/MS analysis of oxidized mAb ...................................................................................20

4.6 Analysis of Antibody-Drug Conjugate (ADC)...................................................................23

4.7 Top-down LC/MS/MS Analysis of Intact Protein Mix ........................................................24

Page 6: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

Contents

Thermo Scientific Product Manual for MAbPac RP Columns Page 6 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

5. Frequently Asked Questions ........................................................................................... 26

5.1 What factors do I need to consider for developing a high throughput method

using MAbPac RP ........................................................................................................26

5.2 What is the recommended operating temperature for MAbPac RP? ......................................26

5.3 What is the carryover on MAbPac RP? ............................................................................26

5.4 What is the recommended maximum flow rate? ................................................................26

5.5 What is the recommended TFA concentration for protein separation? ...................................26

5.6 How to mitigate carryover issue? ....................................................................................26

5.7 How to optimize the HPLC system configuration to achieve the best separation

using the 1.0 mm ID MAbPac RP column?.......................................................................27

Page 7: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

1 – Introduction

Thermo Scientific Product Manual for MAbPac RP Columns Page 7 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

1. Introduction

1.1 Introduction to the MAbPac RP Column

MAbPac RP is a reverse phase (RP) column specifically designed for separation of intact monoclonal antibodies (mAbs) and mAb fragments (figure 1). The stationary phase is designed

to be compatible with mass spectrometry friendly organic solvent such as acetonitrile and isopropanol and low pH eluents containing trifluoroacetic acid or formic acid . The MAbPac RP

is based on supermacroporous 4 µm polymer particles that are stable at extreme pH (0 – 14) and high temperature (up to 110 °C). Particles are inherently hydrophobic so there is no bonded phase, alkyl ligand required for reversed-phase separations. The large pore size polymeric resin enables

efficient separation of protein molecules with very low carry over.

Figure 1 – Structure of IgG and typical forms of heterogeneity

Antigen-bindingsiteHeavy Chain

Glycosylation

Fc

Fab

Reduction of disulfide bondor disulfide shuffling

Light Chain Asparagine deamidation

Methionine oxidation

Lysine truncationK K

Aspartic acid isomerization

Pyroglutamate formationfrom glutamine or glutamic acid

Page 8: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

1 – Introduction

Thermo Scientific Product Manual for MAbPac RP Columns Page 8 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

1.2 MAbPac RP Operating Limits and Specifications

1.2.1 Operating Conditions

Parameter Recommendation

HPLC systems Thermo Scientific™ Vanquish™ Horizon UHPLC system with a 35 µL gradient mixer and 75 µm I.D. connection tubing

for the 1.0 mm I.D. columns Thermo Scientific™ Vanquish™ Horizon UHPLC system or

Thermo Scientific™ Vanquish™ Flex system for the 2.1 mm and 3.0 mm I.D. columns

Flow Rate Range (recommended)

500 – 1,000 µL/min for the 3.0 mm I.D. columns

300 – 600 µL/min for the 2.1 mm I.D. columns

75 – 150 µL/min for the 1.0 mm I.D. columns

Shipping Solution / Long Term Storage Solution MeCN/H2O (50:50 v/v)

Typical buffers

LC/UV experiment Mobile phase A: H2O/TFA (99.9:0.1 v/v) Mobile phase B: MeCN/H2O/TFA (90:9.9:0.1 v/v/v)

LC/MS experiment Mobile phase A: H2O/FA/TFA (99.88:0.1:0.02 v/v/v) Mobile phase B: MeCN/H2O/FA/TFA (90:9.88:0.1:0.02

v/v/v/v)

Solvents Compatibility Up to 100% acetonitrile, isopropanol, methanol

Temperature Range up to 110 °C

Pressure Limit 4,000 psi

pH Range 0 – 14 for 3.0 mm and 2.1 mm I.D. columns, 1 – 7 for 1 mm I.D. columns

Do NOT flow hydrofluoric acid into the 1 mm I.D. columns.

Do NOT flow hot phosphoric acid into the 1 mm I.D. columns.

Assistance is available for any problem during the shipment or operation of Thermo

Scientific columns at [email protected]

1.2.2 Physical Characteristics

Substrate: Supermacroporous polymer

Particle size: 4 µm

WARNING

!

NOTE

!

Page 9: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

1 – Introduction

Thermo Scientific Product Manual for MAbPac RP Columns Page 9 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

1.3 Formats of the MAbPac RP Columns

Currently, MAbPac RP size exclusion columns are available in 3.0 mm, 2.1 mm, and 1.0 mm diameter formats.

Product Description Part Number

MAbPac RP, 4µm, Analytical column 3.0 × 100 mm 088644

MAbPac RP, 4µm, Analytical column 3.0 × 50 mm 088645

MAbPac RP, 4µm, Guard 3.0 × 10 mm 088646

MAbPac RP, 4µm, Analytical column 2.1 × 100 mm 088647

MAbPac RP, 4µm, Analytical column 2.1 × 50 mm 088648

MAbPac RP, 4µm, Guard 2.1 × 10 mm 088649

MAbPac RP, 4µm, Analytical column 1.0 × 150 mm 303184

MAbPac RP, 4µm, Analytical column 1.0 × 100 mm 303183

MAbPac RP, 4µm, Analytical column 1.0 × 50 mm 303182

Page 10: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

2 – Getting Started; Step-By-Step Procedure

Thermo Scientific Product Manual for MAbPac RP Columns Page 10 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

2. Getting Started;

Step-By-Step Procedure

Thermo Fisher Scientific recommends that you perform an efficiency test on your MAbPac RP

column before use. The purpose of column performance validation is to ensure no damage has occurred during shipping. Steps 1 – 5 below outline the necessary steps to perform this validation test. Test the column using the conditions described on the Quality Assurance (QA) report

enclosed in the column box. Repeat the test periodically to track the column performance over time. Note that slight variations may be found on two different HPLC systems due to system electronic, hardware, plumbing, operating environment, reagent quality, column conditioning,

and operator technique.

2.1 Step 1 – Visually inspect the column

Report any visible damage upon receiving the column to Thermo Fisher Scientific immediately.

Depending upon the nature of the damage, we may request that you return the damaged column

back to us for a replacement column.

2.2 Step 2 – Mobile Phase Selection

The MAbPac RP column can be used with a variety of mass spectrometry friendly organic

solvents such as acetonitrile and isopropanol. In general we recommend adding 0.1% TFA to the mobile phase for good separation. When the HPLC system is coupled to mass spectrometry, it is

recommended to reduce the TFA content to 0.02% to avoid ion suppression.

2.3 Step 3 – Set up the LC system

Use a standard LC system equipped with a LC pump, a column oven, a UV detector (210 - 220 nm and/or 280 nm) and an injector (or an autosampler). It is highly recommended that the system

be optimized for low dead volume; usage of small internal diameter tubing (such as 100 µm) and

a proper detector flow cell (such as the 2.5 µL semi-micro flow cell) is required for best results. The system should be thoroughly primed before use. It is recommended the column is run at high

temperature (70 to 80 °C) to achieve better separation of mAb and mAb fragments.

Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure mixing pump (HPG) or low pressure mixing

pump (LPG) to achieve high resolution separation. Low flow MAbPac RP columns in 1.0 mm ID are normally operated in the flow rate range of 75 to 150 µL/min and therefore should be operated with a high pressure mixing pump such as the Vanquish Horizon UHPLC system to avoid

excessive gradient delay. The 1.0 mm ID MAbPac RP is designed to be used with high resolution mass spectrometer such as Thermo Scientific™ Q Exactive™ BioPharma for accurate intact mass

detection and top-down protein sequencing.

Page 11: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

2 – Getting Started; Step-By-Step Procedure

Thermo Scientific Product Manual for MAbPac RP Columns Page 11 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

2.4 Step 4 – Condition the column

Set the pressure limit on the pump to ≤ 4000 psi (276 bar). Slowly ramp up the flow rate: 0.5 mL/min for 3.0 mm ID column, 0.3 ml/min for 2.1 mm ID column, or 0.1 ml/min for 1.0 mm ID

column. If possible, set 1-minute flow ramps up and down. Wash the column with mobile phase

for 20 minutes.

2.5 Step 5 – Verify the performance of the column

Perform the column performance test using the conditions described in the Quality Assurance

Report and compare the result with the one in the report. After the column is fully equilibrated,

multiple injections of protein samples should be made until the reproducible results are obtained.

Due to various reasons, such as differences in LC systems, mobile phases, etc, you may observe somewhat different separation from that in the report.

2.6 Step 6 – Real sample analysis

Once the column performance is satisfactorily confirmed in Step 1-5, the column is ready for real

sample analysis. Equilibrate the column with the desired mobile phase before sample analysis.

It is recommended that the column performance test be performed periodically to monitor the condition of the column.

NOTE

!

NOTE

!

Page 12: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

3 – Column Care

Thermo Scientific Product Manual for MAbPac RP Columns Page 12 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

3. Column Care

3.1 Column storage The column can be stored in the mobile phase for short-term storage. For long-term storage (more than 5 days), it is recommended to store the column in a solution containing 90% acetonitrile in

deionized water.

3.2 Operating pH range: pH 0 to 14 for 3.0 mm and 2.1 mm I.D. columns, pH 1 to 7 for 1.0 mm I.D. columns

Do NOT flow hydrofluoric acid into the 1 mm I.D. columns. Do NOT flow hot phosphoric acid into the 1 mm I.D. columns.

3.3 Operating temperature limit: up to 110 ºC MAbPac RP column is stable at high temperature up to 110 ºC. The typical operating temperature

for mAb and mAb fragment separation is between 70 ºC to 80 ºC.

3.4 Pressure limit: 4000 psi The back pressure of the column is strongly correlated to the column temperature and flow rate.

3.5 Flow rate Please refer to operating condition table (section 1.2) for recommended flow rate at 80 ºC.

WARNING

!

Page 13: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

3 – Column Care

Thermo Scientific Product Manual for MAbPac RP Columns Page 13 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

3.6 Loading capacity The loading capacity of MAbPac RP column spans at least three orders of magnitude. Figure 2

shows a mAb analyzed on MAbPac RP at four different sample loadings: 0.02 µg, 0.2 µg, 2 µg,

and 20 µg. The peak area to sample load is plotted and linear fit yields a R2 equal to 1. This

correlation allows the MAbPac RP column to be used for quantitation of mAb over a wide range

of concentrations.

Figure 2 – Loadability

Column: MAbPac RP, 4 µmFormat: 2.1 ×50 mmMobile phase A: H2O/TFA (99.9 : 0.1 v/v)

Mobile phase B: MeCN/ H2O/TFA (90: 9.9 :0.1

v/v/v/)Gradient:

Time (min) %A %B-1 85 15

0.0 85 15

2.5 50 502.7 50 50

2.8 85 153.0 85 15

Temperature: 80 ºC

Flow rate: 0.6 mL/minInj. volume: 1 µL

Detection: UV (280 nm)Sample: a. mAb (0.02 mg/mL)

b. mAb (0.2 mg/mL)

c. mAb (2 mg/mL)d. mAb (20 mg/mL)

-0.50

1.25

2.50

-2.0

10.0

18.0

-20

140

1.00 1.50 2.00 2.50 3.00 3.50 4.00

-100

500

900

Retention Time (min)

(a)

(d)

(c)

(b)

Page 14: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

3 – Column Care

Thermo Scientific Product Manual for MAbPac RP Columns Page 14 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

3.7 Ruggedness Column ruggedness is a critical characteristic for accurate and reproducible results, as well as good column lifetime. MAbPac RP columns are packed using a carefully developed packing

protocol to ensure excellent packed bed stability, column efficiency and peak asymmetry. Figure 3 demonstrates that the excellent performance of the MAbPac RP is maintained throughout 1,000 runs at 80 ºC providing consistent retention time, peak shape, and peak efficiency, with minimal

increase in column backpressure. The RSDs of retention time from four protein peaks are

tabulated in Figure 3.

Figure 3 – Excellent reproducibility

Column: MAbPac RP, 4 µmFormat: 2.1 ×50 mmMobile phase A: H2O/TFA (99.9 : 0.1 v/v)

Mobile phase B: MeCN/ H2O/TFA (90: 9.9 :0.1

v/v/v)Gradient:

Time (min) %A %B0.0 80 20

2.5 50 50

2.7 50 502.8 80 20

4.0 80 20

Temperature: 80 ºC

Flow rate: 0.6 mL/minInj. volume: 1 µL

Detection: UV (280 nm)Sample: 1. Ribonuclease A (0.5 mg/mL)

2. Cytochrome C (0.5 mg/mL)

3. Lysozyme (0.5 mg/mL)4. mAb (1 mg/mL)

0.00 0.50 1.00 1.50 2.00 2.50 3.00

-80

-60

-40

-20

0

20

40

60

80

100

120

#1101

#901

#801

#701

#651

#501

#401

#301

#201

#101

Retention Time (min)

1

2

3

4

Peaks 1 2 3 4

RSD of

Retention

Time 0.535 0.262 0.225 0.123

Page 15: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

3 – Column Care

Thermo Scientific Product Manual for MAbPac RP Columns Page 15 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

3.8 Column washing procedure To elute very hydrophobic/sticky protein, increase the mobile phase organic content up 90% (such as 90% acetonitrile) and wash the column for 10-20 min. MAbPac RP chemistry (PS-DVB) is stable under a wide pH range. If needed, the 3.0 and 2.1 mm I.D. columns can be washed with a

mixture of organic solvent and high pH eluent. Figure 4 shows that MAbPac RP, 2.1x50 mm column maintains its performance after 6 hours of wash with 0.8 M NaOH and 18% acetonitrile

at 80 ºC. Compared to silica based RP columns, the MAbPac RP has chemical stability, especially under alkaline condition and provides a great advantage for column cleaning and removal of protein carryover. However for the 1 mm I.D. column, do NOT wash the column with high pH

eluent.

To eliminate metal contamination, wash the column in the following sequence:

1. 100 mM NH4OAc (pH 5.0-5.4) for 10 column volumes

2. 100 mM Sodium Pyrophosphate solution (pH 6.5; adjust pH using phosphoric acid) for

100 column volumes

3. 100 mM NH4OAc (pH 5.0-5.4) for 10 column volumes

4. MeCN/100 mM NH4OAc (90:10; v/v) for 20 column volumes

5. MeCN/H2O (50:50; v/v) for 10 column volumes

Figure 4 – Superior chemical stability

Column: MAbPac RP, 4 µmFormat: 2.1 ×50 mmMobile phase A: H2O/TFA (99.9 : 0.1 v/v)

Mobile phase B: MeCN/ H2O/TFA (90: 9.9 :0.1

v/v/v)Gradient:

Time (min) %A %B0.0 85 15

2.5 50 50

2.7 50 502.8 85 15

5.0 85 15

Temperature: 80 ºC

Flow rate: 0.6 mL/minInj. volume: 1 µL

Detection: UV (280 nm)Sample: 1. Ribonuclease A (0.5 mg/mL)

2. Cytochrome C (0.5 mg/mL)

3. Lysozyme (0.5 mg/mL)4. mAb (1 mg/mL)

0.50 1.00 1.50 2.00 2.50 3.00 3.50

-20

0

13

25

38

50

63

75

88

100

113

125

140

After 6 hrs of 0.8 M NaOH

and 18% acetonitrile

wash at 80 C

Before NaOH wash

1

2

3

4

Retention Time (min)

Page 16: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

4 – Example Applications

Thermo Scientific Product Manual for MAbPac RP Columns Page 16 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

4. Example Applications

4.1 Optimization for protein separation

The MAbPac RP column is designed for high resolution and fast separation of proteins. The speed and resolution of the separation can be optimized by adjusting both gradient slope and flow rate. The supermacroporous nature of the

resin provides low back pressure characteristic and enables high flow rate. Figure 5 illustrates the separation of four standard proteins: ribonuclease A, cytochrome C, lysozyme, and BSA. In figure 5a, the four proteins were baseline

separated using a 10-min gradient method with a gradient slope of 10% B per min. The total run time was 15 minutes. By doubling the flow rate, a 4-min gradient method with a gradient slope of 25% B per min (figure 5b) can separate the same four proteins, although the resolution is slightly lower than the separation achieved by the 10-min gradient

method. By doubling the flow rate and reducing the total gradient range (10% B to 60% B), slightly better separation was achieved using a 4-min gradient method with a gradient slope of 12.5% B per min (figure 5c).

Figure 5 – Impact of flow rate and gradient slope on protein separation

-50

125

250

375

500

-50

100

200

350

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0

-50

100

250

Retention Time (min)

Column: MAbPac RP, 4 µm

Format: 2.1 ×50 mmMobile phase A: H2O/TFA (99.9:0.1 v/v)

Mobile phase B: MeCN/H2O/TFA (90:9.9:0.1

v/v/v)Gradient slope: (a) 10% B/min

(b) 25% B/min(c) 12.5% B/min

Flow rate: (a) 0.3 mL/min

(b) 0.6 mL/min(c) 0.6 mL/min

Temperature: 80 ºCInj. volume: 2 µL

Detection: UV (280 nm)

Sample: 1. Ribonuclease A (2.5 mg/mL)2. Cytochrome C (2.5 mg/mL)

3. Lysozyme (2.5 mg/mL)4. BSA (5 mg/mL)

1

2

3

4

1

2

3

4

1

2

3

4

(a)

(b)

(c)

Time (min) %A %B

0.0 100 0

1.0 100 0

11.0 0 100

12.0 0 100

14.0 100 0

15.0 100 0

Time (min) %A %B

0.0 100 0

0.5 100 0

4.5 0 100

5.0 0 100

5.5 100 0

6.0 100 0

Time (min) %A %B

0.0 90 10

0.5 90 10

4.5 40 60

5.0 40 60

5.5 90 10

6.0 90 10

Page 17: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

4 – Example Applications

Thermo Scientific Product Manual for MAbPac RP Columns Page 17 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

4.2 Separation of mAb and mAb fragments

Monoclonal antibodies are heterogeneous (Figure 1). Comprehensive analysis of mAb post-translational modifications, such as deamidation, C-terminal lysine truncation, N-terminal pyroglutamation, methionine (Met) oxidation, and

glycosylation, requires complete digestion of the mAbs and sequencing of all the peptides. However, “peptide mapping” is time consuming. A simpler and direct way to analyze the mAb variants and locate the modifications is to measure mAb fragments. Light chain (LC) and heavy chain (HC) are generated by the reduction of mAb, Fc and Fab

fragments are generated by papain digestion. scFc and F(ab)2 fragments are generated by IdeS digestion. Figure 6 shows the analysis of Trastuzumab intact molecule and fragments. LC and HC (Figure 6b), Fc and Fab (Figure 6c),

scFc and F(ab)2 (Figure 6d) are baseline separated using a MAbPac RP column with a 10-min gradient. Similar

experiments have been carried out for Rituximab, Infliximab, and Bevacizumab. In all cases, mAb fragments have been successfully separated (data not shown).

Figure 6 – Analysis of Trastuzumab and fragments using MAbPac RP.

(a) Trastuzumab; (b) Trastuzumab LC and HC; (c) Trastuzumab Fc and Fab fragments; (d) Trastuzumab scFc and

F(ab)2 fragments.

Column: MAbPac RP, 4 µmFormat: 3 ×50 mmMobile phase A: H2O/FA/TFA (99.88 : 0.1:0.02 v/v/v)

Mobile phase B: MeCN/ H2O/FA/TFA (90: 9.88

:0.1:0.02 v/v/v/v)Gradient:

Time (min) %A %B0.0 80 20

1.0 80 20

11.0 55 4512.0 55 45

14.0 80 2015.0 80 20

Temperature: 80 ºCFlow rate: 0.5 mL/min

Inj. volume: 5 µL Detection: UV (280 nm)

Sample: (a) Trastuzumab (5 mg/mL)

(b) Trastuzumab + DTT (4 mg/mL)(c) Trastuzumab + Papain (2 mg/mL)

(d) Trastuzumab + IdeS (2 mg/mL)

-20

140

-10.0

50.0

80.0

-10

50

90

1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0

-10

50

100

Retention Time (min)

mAb

LC

HC

Fc

Fab

scFc

F(ab)2

(b) Trastuzumab + DTT

(a) Trastuzumab

(c) Trastuzumab + Papain

(d) Trastuzumab + IdeS

Page 18: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

4 – Example Applications

Thermo Scientific Product Manual for MAbPac RP Columns Page 18 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

4.3 LC/MS analysis of mAb

High resolution mass spectrometers such as Thermo QExactive provide accurate mass information of large biologic molecules such as mAbs. The MAbPac RP column can be directly coupled to the mass spectrometer for MS

detection of mAb and mAb fragments. While trifluoroacetic acid (TFA) as ion-pairing reagent provides excellent separation results, TFA can suppress ionization in the LC-MS interface, causing a drop in signal. This can be mitigated by reducing the TFA concentration to 0.02% and additional use 0.1% Formic acid (FA). Figure 7 shows

the intact mass detection of Trastuzumab. The top trace shows the total ion-current chromatogram. The middle trace shows the MS spectrum of Trastuzumab in the mass range of 2,000 to 4,000 m/z. The bottom trace shows a zoom-in spectrum of Trastuzumab with 50+ charges. The cluster shows the glycosylation profile of Trastuzumab .

Figure 7 – LC/MS analysis of Trastuzumab on MAbPac RP

Column: MAbPac RP, 4 µmFormat: 3 ×50 mmMobile phase A: H2O/FA/TFA (99.88:0.1:0.02

v/v/v)

Mobile phase B: MeCN/ H2O/FA/TFA (90:9.88 :0.1:0.02 v/v/v/v)

Gradient:Time (min) %A %B

0.0 80 20

1.0 80 2011.0 55 45

12.0 55 4514.0 80 20

15.0 80 20

Temperature: 80 ºC

Flow rate: 0.5 mL/minInj. volume: 1 µL

MS Detection: Positive-ion mode

Mass Spec: Q Exactive™ PlusSample: Trastuzumab (5 mg/mL)

TIC

MS spectrum

m/z at 50+

4.5 5 .0 5.5 6.0 6.5 7.0 7.5 8 .0 8.5 9.0 9.5 10.0 10.5 11.0 11.5 12.0 12.5 13.0Tim e (m in)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lati

ve

Ab

unda

nce

8 .26

8.44

12.748.6310.5310.339.888.97 9.729.377.395.935.685.43 6.095.134.47 6.996.74 7.64

2000 2200 2400 2600 2800 3000 3200 3400 3600 3800 4000

m /z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lati

ve

Ab

unda

nce

3223.20

3088.92

3294.80 3369.643154.64

3025.91

2965.42 3447.98

2907.27

3530.09

3616.14

2851.39

3706.502797.55

3801.532745.82

2695.91

2647.76

3901.54

2601.37

2513.18

2471.32

2430.80

2394.25

2283.772183.13

2057.49

3917.63

3994.21

2930 2935 2940 2945 2950 2955 2960 2965 2970 2975 2980 2985 2990 2995 3000

m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rela

tive A

bundance

2965.42

2968.61

2962.22

2971.87

2959.34

2974.96

2996.302953.05 2993.072936.54 2977.49

2956.45

2933.53 2980.99 2986.822939.75 3002.662946.14

Page 19: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

4 – Example Applications

Thermo Scientific Product Manual for MAbPac RP Columns Page 19 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

4.4 LC/MS analysis of mAb fragments

During characterization, mAb is often reduced to LC and HC. Mass spectrometry analysis of these fragments can quickly review and localize the modifications. Figure 8 shows the separation of Trastuzumab LC and HC on a 3 × 50

mm MAbPac RP column. Total ion chromatogram (TIC) and UV spectrum show identical retention time of mAb fragments. The mass spectrum of LC shows multiple charge states of a single polypeptide chain while the mass spectrum of the HC shows multiple glycosylation forms of the heavy chain.

Figure 8 – LC/MS analysis of Trastuzumab light chain (LC) and heavy chain (HC) on MAbPac RP

Column: MAbPac RP, 4 µmFormat: 3 ×50 mmMobile phase A: H2O/FA/TFA (99.88:0.1:0.02 v/v/v)

Mobile phase B: MeCN/ H2O/FA/TFA (90:9.88

:0.1:0.02 v/v/v/v)Gradient:

Time (min) %A %B0.0 80 20

1.0 80 20

11.0 55 4512.0 55 45

14.0 80 2015.0 80 20

Temperature: 80 ºCFlow rate: 0.5 mL/min

Inj. volume: 1 µL UV Detection: 280 nm

MS Detection: Positive-ion mode

Mass Spec: Q Exactive™ PlusSample: Reduced trastuzumab (4 mg/mL)

TIC

UV

LC

HC

LC HC

31+

1200 1300 1400 1500 1600 1700 1800 1900 2000 2100 2200 2300 2400

m/z

0

10

20

30

40

50

60

70

80

90

100

0

10

20

30

40

50

60

70

80

90

100

Re

lativ

e A

bu

nd

an

ce

1954.161803.88

2131.63

1675.14

1563.47 2344.74

1465.80

1379.68

1303.031967.87 2146.62

1234.471687.01 2361.04

2233.121575.62

1819.28

1478.76

1762.96

2095.001392.86

1638.291539.03

1692.831493.85

1751.18

1446.471813.69

1410.901880.81

1953.111368.36

2031.28

2115.811336.67

2200.69

2308.14 2418.021269.96

1238.962472.972253.37

1180.58

2358.102158.66

3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0

Time (min)

0

50000

100000

150000

200000

250000

300000

350000

400000

450000

500000

550000

600000

uA

U

0

10

20

30

40

50

60

70

80

90

100

Rela

tive A

bundance

8.28

7.43

8.58 8.8610.7710.373.943.63 9.114.793.23 10.024.34 9.765.20 5.50 5.95 6.10 7.956.41 6.96

8.28

7.42

8.859.107.736.806.636.35

T:

1600 1605 1610 1615 1620 1625 1630 1635 1640 1645 1650 1655 1660 1665 1670m /z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lati

ve

Ab

unda

nce

1638.29

1633.01

1643.60

1631.64 1640.081634.91

1628.281625.611646.93

1671.371624.16 1661.331605.32 1608.89 1666.591653.791620.161600.00 1615.28

Page 20: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

4 – Example Applications

Thermo Scientific Product Manual for MAbPac RP Columns Page 20 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

4.5 LC/MS analysis of oxidized mAb

Methionine (Met) oxidation is one of the critical quality attributes required to be closely monitored. The two Met residues in the CH2-CH3 domain interface of recombinant humanized and fully human IgG1 antibodies were found

susceptible to oxidation. It is desirable to monitor the progress of the Met oxidation without complete digestion of mAb. A workflow was designed to first reduce mAb and then further digest it with IdeS resulting in smaller (25 kDa) fragments (Figure 9). Figure 10a shows that oxidized HC and non-oxidized HC can be barely separated

chromatographically, with 0.02% TFA in the mobile phase. However, the high resolution orbitrap instrument can clearly resolve the oxidized (Figure 10b) and non-oxidized HC (Figure 10c) at m/z 1633.54 and 1633.06 respectively.

Further digestion of the HC by IdeS resulted in two smaller fragments: scFc and Fd. Figure 11a shows the baseline

separation of scFc, LC, and Fd. In addition, oxidized and non-oxidized scFc fragments are better separated than the oxidized and non-oxidized HC. The +10 charge state of the oxidized scFc and non-oxidized scFc are shown in Figure 11b (at m/z 2525.60) and in Figure 11c (at m/z 2524.08). Separation of these oxidized and non-oxidized fragments can

be improved by increasing the mobile phase TFA concentration from 0.02% to 0.1% (Figure 12).

Figure 9 – mAb reduction and IdeS digestion flowchart.

Page 21: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

4 – Example Applications

Thermo Scientific Product Manual for MAbPac RP Columns Page 21 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

Figure 10 – LC/MS analysis of Trastuzumab LC and HC. (a) total ion current (TIC); (b) mass spectrum of oxidized HC; (c) mass spectrum of non-oxidized HC.

Column: MAbPac RP, 4 µmFormat: 3 ×50 mmMobile phase A: H2O/FA/TFA (99.88:0.1:0.02

v/v/v)

Mobile phase B: MeCN/ H2O/FA/TFA (90:9.88 :0.1:0.02 v/v/v/v)

Gradient:Time (min) %A %B

0.0 75 25

1.0 75 2511.0 63 37

12.0 63 3714.0 75 25

15.0 75 25

Temperature: 80 ºC

Flow rate: 0.5 mL/minInj. volume: 2 µL

MS Detection: positive-ion mode

Mass spec: Q Exactive™ PlusSample: oxidized trastuzumab, reduced

by DTT (2 mg/mL)

(a) TIC

(b) Oxidized HC

(c) HC

LCoxidized HC

HC

+31

RT: 2.00 - 12.00

2 3 4 5 6 7 8 9 10 11 12

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rela

tive A

bundance

8.33

10.2710.21

10.78 11.445.134.93 5.493.873.37 4.07 5.69 8.936.09 6.393.02 6.75 7.15 7.352.81 9.537.702.51

1605 1610 1615 1620 1625 1630 1635 1640 1645 1650 1655 1660 1665

m/z

0

10

20

30

40

50

60

70

80

90

100

0

10

20

30

40

50

60

70

80

90

100

Re

lativ

e A

bu

nd

an

ce

1638.76

1633.54

1644.001628.94

1636.88 1640.111632.23

1649.171652.801626.281612.881610.041605.43 1660.831617.35 1665.641622.36

1633.05

1638.26

1643.51

1631.541636.69

1639.651610.59 1614.85 1625.56 1653.751647.531607.87 1622.58 1657.35 1664.30

Page 22: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

4 – Example Applications

Thermo Scientific Product Manual for MAbPac RP Columns Page 22 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

Figure 11 – LC/MS analysis of Trastuzumab scFc, LC, and Fd .

(a) total ion current (TIC); (b) mass spectrum of oxidized scFc; (c) mass spectrum of non-oxidized scFc.

Figure 12 – Separation of Trastuzumab fragments with mobile phases containing 0.1% TFA.

(a) Trastuzumab LC and HC; (b) Trastuzumab scFc, LC, and Fd.

Column: MAbPac RP, 4 µmFormat: 3 ×50 mmMobile phase A: H2O/FA/TFA (99.88 : 0.1:0.02

v/v/v)

Mobile phase B: MeCN/ H2O/FA/TFA (90: 9.88 :0.1:0.02 v/v/v/v)

Gradient:Time (min) %A %B

0.0 75 25

1.0 75 2511.0 63 37

12.0 63 3714.0 75 25

15.0 75 25

Temperature: 80 ºC

Flow rate: 0.5 mL/minInj. volume: 2 µL

MS Detection: positive-ion mode

Mass spec: Q Exactive™ PlusSample: Oxidized trastuzumab, reduced

by DTT and digested by IdeS (1 mg/mL)

(a) TIC

(b) Oxidized scFc

(c) scFc

LCFd

Oxidizied scFc

scFc

+102525.60

2524.08

2 3 4 5 6 7 8 9 10 11 12

Tim e (m in)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

Re

lati

ve

Ab

unda

nce

10 .42

8.35

7.37

7.16

10.10

10.7911.90

5.074.87 6.285.825.524.213.91 9.633.45 6.483.25 9.388.922.852.50 7.79

2460 2480 2500 2520 2540 2560 2580 2600 2620

m/z

0

10

20

30

40

50

60

70

80

90

100

0

10

20

30

40

50

60

70

80

90

100

Re

lativ

e A

bu

nd

an

ce

2541.99

2525.60

2557.972529.45

2510.98 2545.66

2560.15

2521.752502.76 2566.26

2524.08

2540.38

2509.45 2527.982544.22 2556.93

2501.48 2520.11 2560.78 2585.84

-2.0

2.5

5.0

7.5

12.0

1.0 2.5 3.8 5.0 6.3 7.5 8.8 10.0 11.3 12.5 13.8 15.0

-1.0

2.0

4.0

6.0

9.0

Retention Time (min)

Column: MAbPac RP, 4 µmFormat: 3 ×50 mmMobile phase A: H2O/TFA (99.9:0.1 v/v)

Mobile phase B: MeCN/H2O/TFA (90:9.9:0.1 v/v/v)

Gradient:Time (min) %A %B

0.0 70 301.0 70 30

11.0 60 40

12.0 60 4014.0 70 30

15.0 70 30

Temperature: 80 ºC

Flow rate: 0.5 mL/minInj. volume: 2 µL

Detection: UV (280 nm)Sample: (a) Trastuzumab + DTT (2 mg/mL)

(b) Trastuzumab + DTT + IdeS (1

mg/mL)

LC

scFc

LC

HC

Oxidized scFc

Fd

Oxidized HC(a)

(b)

Page 23: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

4 – Example Applications

Thermo Scientific Product Manual for MAbPac RP Columns Page 23 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

4.6 Analysis of Antibody-Drug Conjugate (ADC)

Antibody-drug conjugates have proved to be a very effective cancer therapy. Due to the heterogeneous nature of the ADC, it is critical to characterize its multiple Drug-to-Antibody Ratio (DAR) forms. The MAbPac RP column can be

utilized in the separation of mAb and its conjugates. In Figure 13, ADCs were prepared by enzymatically activating mAb Fc domain glycans with azides using the mutant beta-galactosyltransferase enzyme. The azide-activated antibodies were then conjugated with dibenzocyclooctyne6 (DIBO) -activated Val-Cit-PAB-Monomethyl Auristatin E

(MMAE) toxin in a copperless click reaction, resulting in a mixture of drug-loaded antibody species with 0 to 4 MMAE molecules. The unmodified mAb and ADCs with DAR values ranging from 0 to 4 are well resolved on the MAbPac RP column (Figure 14).

Figure 13 – Site-selective antibody-drug conjugates (ADCs)

Figure 14 – Site-selective antibody-drug conjugates (ADCs)

DIBO-MMAE

25 C, ON

Add in

GalT(Y289L)

UDP-GalNAzβ-1,4-Galactosidase

37 C, 5 hr

Azide-Activated Ab

(stable for long-term storage)

Antibody drug conjugate (ADC)Unlabeled Ab Cleave Terminal Gal

37 C, ON

N3

N3

N3

N3

MMAE

MMAE

MMAE

MMAE

Column: MAbPac RP, 4 µm

Format: 2.1 ×50 mm

Mobile phase A: H2O/TFA (99.9 : 0.1 v/v)

Mobile phase B: MeCN/ H2O/TFA (90: 9.9 :0.1

v/v/v)

Gradient:

Time (min) %A %B

0.0 65 35

0.5 65 35

4.5 45 55

5.0 45 55

5.5 65 35

6.0 65 35

Flow rate: 0.6 mL/min

Temperature: 80 ºC

Inj. volume: 2 µL

Detection: UV (280 nm)

Sample: Trastuzumab-MMAE

1.00 1.50 2.00 2.50 3.00 3.50 4.00

-5.0

0.0

5.0

10.0

15.0

20.0

25.0

Retention Time (min)

Page 24: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

4 – Example Applications

Thermo Scientific Product Manual for MAbPac RP Columns Page 24 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

4.7 Top-down LC/MS/MS Analysis of Intact Protein Mix

ThermoFisher Pierce Intact Protein Mix consists of six recombinant proteins: IGF-I LR3 (9 kDa), Thioredoxin (12 kDa), Protein G (21 kDa), Carbonic Anhydrase II (29 kDa), Protein AF (51 kDa), and Exo Klenow (68 kDa). These

proteins are selected to satisfy the following criteria: 1) evenly covering a MW range of 10kD – 66kD, 2) presenting mostly clean, modification and adduct-free ESI spectra, and 3) having ESI charge state distributions covered a wide m/z range from 500-2,000. 1 mm ID MAbPac RP column is chosen to analyze these proteins because it provides

higher sensitivity. Figure 15 shows baseline separation of all six protein using an acetonitrile/H2O/formic acid mobile phase. Top-down MS/MS spectra were acquired using Top 3-5 DDA methods. OT MS1 data was acquired at resolution settings of 15 at m/z 200 and OTMS2 at a resolution of 120K at m/z 200. Top panel shows deconvolution

results from Protein Deconvolution 4.0 software and Bottom panel shows top down results from ProSight PD 1.1 node in Proteome Discoverer 2.1 software.

Figure 15 – LC/MS/MS Analysis of Intact Protein Mix

Column: MAbPac RP, 4 µmFormat: 1.0 × 150 mm

Mobile phase A: H2O/FA (99.9 : 0.1 v/v)

Mobile phase B: MeCN/FA (99.0: 0.1 v/v)

Gradient:

Time (min) %A %B

0.0 90 10

1.0 80 20

15.0 55 45

16.0 10 90

20.0 10 90

20.1 90 10

25.0 90 10

Temperature: 60 ºC

Flow rate: 0.1 mL/min

Inj. volume: 1 µL

MS Detection: positive-ion mode

Mass Spec: Q Exactive™ HF

Sample: Pierce Intact Protein Standard Mix

(500 ng/µL)

1. IGF-I LR3 (9 kDa)

2. Thioredoxin (12 kDa)

3. Protein G (21 kDa)

4. Carbonic Anhydrase II (29 kDa)

5. Protein AF (51 kDa)

6. Exo Klenow (68 kDa)6 8 10 12 14 16 18

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Ab

un

da

nce

5

2

1

3

6

4

Page 25: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

4 – Example Applications

Thermo Scientific Product Manual for MAbPac RP Columns Page 25 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

Protein G

CA II

Klenow

Protein AG

Thioredoxin

IGF-LR3

Page 26: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

5 – Frequently Asked Questions

Thermo Scientific Product Manual for MAbPac RP Columns Page 26 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

5. Frequently Asked Questions

5.1 What factors do I need to consider for developing a high throughput method using MAbPac RP

Flow rate, column operating temperature, and gradient slope should be considered (see section

4.1). MAbPac RP is a polymer column which is stable at extreme high temperature and extreme pH. If higher flow rate is desired, operating temperature can be increased to decrease the column

back pressure.

5.2 What is the recommended operating temperature for MAbPac RP?

For small proteins such as cytochrome C and lysozyme, 30 °C is sufficient. For large proteins

such as mAb, 70 °C to 80 °C is recommended in order to decrease the strength of secondary

interactions between mAb and the stationary phase and obtain decent peak shape.

5.3 What is the carryover on MAbPac RP?

In general, the carryover is less than 0.7% for mAb. For example, mAb peak area is determined

by injecting 25 µg of mAb onto a 3.0 × 50 mm MAbPac column; the carry over peak area is determined from the blank run immediately following the mAb injection. The carryover is

calculated using the following equation: Areacarryover/AreamAb x 100%.

mAbs are large and very hydrophobic biomolecules. In order to achieve low carryover for mAb

analysis, such as less than 1%, the pore size of the column should be larger than the conventional RP column and the stationary phase is optimized. MAbPac RP column is designed for low

carryover and high resolution mAb analysis.

5.4 What is the recommended maximum flow rate?

At 80 °C, the maximum flow rate for the 1.0 × 50 mm ID column is 0.15 ml/min.

At 80 °C, the maximum flow rate for the 2.1 × 50 mm ID column is 0.6 ml/min.

At 80 °C, the maximum flow rate for the 3.0 × 50 mm ID column is 1 ml/min.

The column pressure should not exceed 4,000 psi in order to avoid compressing the resin bed .

5.5 What is the recommended TFA concentration for protein separation?

0.1% TFA if the column is not coupled to the mass spectrometer and 0.02% for LC/MS

experiment. While trifluoroacetic acid (TFA) as ion-pairing reagent provides excellent separation results, TFA can suppress ionization in the LC-MS interface, causing a drop in signal. This can be mitigated by reducing the TFA concentration to 0.02% and additional use 0.1% Formic acid

(FA).

5.6 How to mitigate carryover issue?

The system may not be clean or previous sample may not have completely eluted from the column. Wash the system and the column with high organic eluent, such as 90% acetonitrile. If needed,

100 mM NaOH can be added to the high organic wash (for 3.0 mm and 2.1 mm I.D. columns. Do

Page 27: MAbPac RP Columns - assets.thermofisher.com · Analytical format MAbPac RP columns in 2.1 mm ID and 3.0 mm ID format can be used with HPLC system equipped with either high pressure

5 – Frequently Asked Questions

Thermo Scientific Product Manual for MAbPac RP Columns Page 27 of 27 065625-03 For Research Use Only. Not for use in diagnostic procedures.

not use NaOH solution for 1 mm I.D. columns). In addition, a blank can be added in between

samples.

5.7 How to optimize the HPLC system configuration to achieve the best separation using the 1.0 mm ID MAbPac RP column?

The recommended flow rate for the 1.0 mm ID column is between 75 µL to 150 µL. In order to

avoid excessive gradient delay, the high pressure pump Vanquish Horizon system with the default 35 µl gradient mixer (25 µL capillary mixer + 10 µL static mixer) is recommended. If Vanquish Horizon system is not available, Vanquish Flex Binary system with 200 µL gradient mixer (50

µL capillary mixer + 150 µL static mixer) will be the alternative choice. In addition, 75 µm I.D. connection tubing could reduce the system dead volume. Figure 16 shows that excellent

separation of the four protein standards is achieved on a 1 × 100 mm MAbPac RP using a default configuration of Vanquish™ Horizon UHPLC system. The gradient delayed observed here is ~

2.5 min.

Figure 16 – Separation of intact proteins/mAb using 1 mm ID format column

HPLC: Thermo Scientific™ Vanquish™

Horizon UHPLC system

Column: MAbPac RP, 4 µmFormat: 1.0 × 100 mm

Mobile phase A: H2O/TFA (99.9 : 0.1 v/v)

Mobile phase B: MeCN/ H2O/TFA (90: 9.9 :0.1

v/v/v/)

Gradient:

Time (min) %A %B

0.0 80 20

1.0 80 20

11.0 50 50

13.0 50 50

13.1 80 20

25.0 80 20

Temperature: 80 ºC

Flow rate: 0.1 mL/min

Inj. volume: 0.5 µL

Detection: UV (280 nm)

Sample: 1. Ribonuclease A (0.5 mg/mL)

2. Cytochrome C (0.5 mg/mL)

3. Lysozyme (0.5 mg/mL)

4. mAb (1 mg/mL)

Retention Time (min)

1.0 2.5 5.0 7.5 10.0 12.5 15.0 16.0-20

0

100

120

1

2

3

4