Lowering the Dose: mA and kVp Marilyn J. Siegel, M.D Mallinckrodt Institute of Radiology Washington University Medical Center St. Louis, MO

Upload: others

Post on 14-Jul-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Lowering the Dose: mA and kVp

Marilyn J. Siegel, M.D

Mallinckrodt Institute of Radiology

Washington University Medical Center

St. Louis, MO

Page 2: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Disclosure of Commercial Interest

I have a financial relationship with a

commercial organization that may have a direct

or indirect interest in the content as follows:

• Siemens Medical Solutions: consultant,

speakers bureau

Page 3: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Objective

• Describe role of mA and kVp in

lowering CT in an adult population

Page 4: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Major CT Parameters Affecting Dose

• Tube current (mA)

• Tube voltage (kV)

• Pitch

• Collimation

Page 5: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Effect of CT Adaptations on Dose

• 50% decrease in mA = 50% dose decrease

• 80 kVp = 30% to 50% dose decrease

• 50% increase in pitch = 50% dose decrease

• Thicker slices = dose decrease

Page 6: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

mA and kV

• Tube milliamperage (mA)

– rate of x-ray production

• Kilovoltage (kV)

–number of x-rays produced

Page 7: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

mA

Tube Current Reduction

Page 8: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Best Approach for mA Reduction

• Automated technology introduced in 1990s

• Superior to technique charts (no guess work)

• Exceptions: screening chest CT

• Other exception is head CT where beam

attenuation comes from bone formation and

this process is age dependent so mA is age

dependent-use fixed tube current

Graser AJR 2006; 659

Herzog et al. AJR 2008, 1232

Mulkens Radiology 2005; 213

Page 9: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Automated mA Reduction

• mA is adapted to body parts, not patient weight,

thinner parts need less radiation

• mA based on attenuation data from scout image

• Dose modulation done in x, y, z planes

Tube

current

Z-position

Page 10: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Ex: shoulders to pelvis Ref value 150 mA

Page 11: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Automated mA Modulation Mean Dose Reduction

• Stone disease 64%

– Mulkens AJR 2007; 188: 553

• Cardiac CT 60%

– Herzog, AJR 2008; 190:1232

• Colonography 35%

– Graser, AJR 2006; 187:695

Page 12: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Lung Cancer Screening

• Fixed tube current

• mA of 20 to 40 rather than standard 150

• Low attenuation background of air-filled

lungs offers a high contrast setting which

can be used favorably to assess lung

abnormalities

Page 13: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Older woman breast cancer low dose vs. standard CT

Similar image quality

150 mA 20 mA

Page 14: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

National Lung Cancer Screening Trial

• 26,724 participants

• Scanned with 20-40 mA

• Average effective dose

–2 mSv for low-dose CT vs.

–7-8 mSv for standard chest CT

• Dose >> females than males (2.4/1.6)

Larke et al. AJR Nov 2011; 197:1165-1169

Page 15: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

A Tale of Two Techniques

kV Modulation

Page 16: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Major Untapped Potential has been kV

?

Distribution of scans in clinical practice

Page 17: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Why use low kV---Rationale

• Lowers radiation dose

• Improves contrast

• Greatest benefit in contrast CT exams

Siegel MJ Radiology 2004; 233:515

Funama, et al., Radiology 2005

Nakayama, et al., Radiology 2005

Huda, et al., Med Phys 2004

Nakayama, et al. AJR 2006 80kV 120kV

Page 18: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Low kV-Effect on Dose: Phantom Study

Siegel MJ: Radiology 2004; 233: 515

0

5

10

15

20

25

30

35

40

45

0 4 8 12 16 20 24 28 32 36

Phantom diameter, cm

Do

se,

mG

y

140 kV

120 kV

100 kV

80 kV

•Reducing kV reduces dose

Page 19: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

0

50

100

150

200

250

300

350

400

0 4 8 12 16 20 24 28 32 36

Phantom diameter, cm

Iod

ine

co

ntr

as

t(H

U)

80 kV

100 kV

120 kV

140 kV

Low kVp: Effect on Contrast

•Reducing kV increases Iodine contrast (HU)

HU

Siegel MJ: Radiology 2004; 233: 515

Page 20: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Low kVp

• Decreases dose & increases contrast

• Rationale: K-edge of iodine 32 keV

• Mean photon energy

– 80 kVp 44 keV

–100 kVp 52 keV

–120 kVp 57 keV

–140 kVp 62 keV

Huda W, et al. Radiology 2000; 217:430

Page 21: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Low kVp: Effect on Image Noise

0

10

20

30

40

50

60

0 4 8 12 16 20 24 28 32 36

Phantom diameter, cm

No

ise

(HU

) 80 kV

100 kV

120 kV

140 kV

•Reducing kV increases image noise in large phantoms

Noise

Page 22: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

kV Modulation-Key Point

• The use of lower-kV is highly dependent

on patient size and diagnostic task

• Increases in contrast are seen ONLY

with high atomic number substances

–Iodine

–NOT water, soft tissue, calcium

• Benefit greatest in children and small

sized adults and contrast examinations

Page 23: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Approaches for kV Reduction?

• Conventionally done using technique charts

• Recently automatic selection tools have

become available (2011)

• Most experience with kV reduction is in PE

and cardiac CTA

Page 24: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Low kV Coronary CTA

• 100 patients (≤ 85 kg)

–Dual Source 64 CT

–Retrospective gating

–120 kVp / 330 mAs: 12 mSv

–100 kVp / 330 mAs: < 8 mSv

• 39% decrease in radiation exposure

Pflederer T, et al. AJR 2009; 192:1045–1050

Page 25: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Low kVp and High Pitch CT

• Dose < 1 mSv

• Total scan time 0.25-0.27s

100 kV and 0.8 mSv

Achenbach et al Eur Heart J. 2010;31:340-6

Leschka et al. Eur Radiol. 2009;19:2896-903

Page 26: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Low kV Pulmonary Embolism CTA

• Park et al. (Korean J Radiol 2009; 10:235)

–dose reduction of 24%

–10.1 to 7.8 mSv (120 to 100 kV)

–Smaller patients (BMI < 25)

• Matsuoka et al. (AJR 2009; 1651-1656)

–40 to 50% dose reduction (120 to 100 kV)

Page 27: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Low kV—Pulmonary Embolism CTA

• Lower dose, better contrast

• Mean enhancement > with 100 kVp (492 vs. 345 HU)

• Sensitivity 100% for detecting PE

120 kV 80 kV

Szucs-Farkas AJR 2011: 197:852

Page 28: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Automated kV Selection Technology

• A tool which automatically adjusts kV for body size

determined from topogram and the exam type

• Up to 60% dose reduction

Page 29: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Auto kV User Interface

• With auto kV off: choose exam type as you would routinely do, screen shows mA, kV & dose -

• Turn auto kV on: take scout CT-program selects best kV to lower dose and maintain image quality

Small adult

Contrast

chest CT

Page 30: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Auto kV

• Used in conjunction with auto mA

• kV decreases AND mA increases

• But overall dose is lower than with fixed kV

Page 31: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Auto kV contrast abdomen, pelvis

• 15 year old boy

• REF: 120kV / 75mA, CTDIvol = 5.77 mGy

• 80 kV / 177mA CTDIvol = 3.2 mGy

Page 32: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Adults--Automated kV 2011

Automated kV selection based on the attenuation

profile of the topogram is feasible, provides a

diagnostic image quality for body CTA, and reduces

overall radiation dose by 25% as compared with a

standard protocol with 120 kV

Investigative Radiology 2011

Page 33: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Auto kV vs. Image Quality

Investigative Radiology 2011

Page 34: Lowering the Dose: mA and kVp - Society of Computed Body ... · •Exceptions: screening chest CT •Other exception is head CT where beam attenuation comes from bone formation and

Summary

• Beautiful pictures do NOT need to come

at the cost of higher radiation dose

• Adjusting mA and kV can dial down the

dose and drive up imaging quality-ALARA