Anyone who has ever visited the clinic of a physiotherapist or a sports masseur will have been amazed at the large amount of complex-looking equipment to be found there. In this area, as in many others, modern technology (and in particular electronics technology) has become indis- pensable in a relatively short time. One type of equipment that is most often used by these people is a muscle stimulator. There are all sorts of different models available with a variety of names, but they all work in essentially the same manner. Two electrodes are attached to some part of the body a small ways apart, and a pulsating current is passed between the electrodes. The objective of this is to stimulate the muscle located between the electrodes. This usually starts with loosening up a cramped muscle and promoting the blood circulation in the muscle tissue. A GENERAL INTEREST 12 Elektor Electronics 4/2000 Low-Impact Muscle Stimulator L.I.M.S. with treatment programs Design by K. Walraven This is a DIY version of a device that can be found in various forms in the clinics of physiotherapists and massage institutes. It produces a series of electrical shocks that stimulate the muscles. The intensity and duration of the shocks are adjustable, and we have given great attention to the safety of this device. Use this equipment only with bat- tery power. Never use this equipment in the vicinity of the heart! It should preferably be used only on the arms and legs. Always set the amplitude control to the minimum level when start- ing a treatment.

Upload: dangliem

Post on 29-Jun-2018

231 views

Category:

Documents


8 download

TRANSCRIPT

Page 1: Low-Impact Muscle Stimulator - ToulonuXtoulonux.org/documents/electrostim/e004012_low-impact muscle... · Low-Impact Muscle Stimulator L.I.M.S. with treatment programs ... electrical

Anyone who has ever visited the clinic ofa physiotherapist or a sports masseur willhave been amazed at the large amount ofcomplex-looking equipment to be foundthere. In this area, as in many others,modern technology (and in particularelectronics technology) has become indis-pensable in a relatively short time.

One type of equipment that is mostoften used by these people is a musclestimulator. There are all sorts of differentmodels available with a variety of names,but they all work in essentially the samemanner. Two electrodes are attached tosome part of the body a small ways apart,and a pulsating current is passedbetween the electrodes. The objective ofthis is to stimulate the muscle locatedbetween the electrodes.

This usually starts with loosening upa cramped muscle and promoting theblood circulation in the muscle tissue. A

GENERALINTEREST

12 Elektor Electronics 4/2000

Low-Impact Muscle StimulatorL.I.M.S.

with treatment programs

Design by K. Walraven

This is a DIY version of a device that can be found in various forms in the clinics ofphysiotherapists and massage institutes. It produces a series of electrical shocksthat stimulate the muscles. The intensity and duration of the shocks are adjustable,and we have given great attention to the safety of this device.

Use this equipment only with bat-tery power.Never use this equipment in thevicinity of the heart! It shouldpreferably be used only on thearms and legs.Always set the amplitude controlto the minimum level when start-ing a treatment.

Page 2: Low-Impact Muscle Stimulator - ToulonuXtoulonux.org/documents/electrostim/e004012_low-impact muscle... · Low-Impact Muscle Stimulator L.I.M.S. with treatment programs ... electrical

to leave this electronic stimulator alone and insteadask you spouse or another family member to helpyou. After all, there’s nothing wrong with a normalhand massage! Then at least you do not have toworry about possible harmful effects.

DesignThe complete schematic diagram of the stimulatoris shown in Figure 1. We can distinguish betweentwo things in terms of its operation: the generationof the electrical pulses and the implementation ofa few special functions.

The first of these is remarkably simple. Micro-processor IC1 (an 89C2051) is programmed to gen-erate pulses on its RxD and TxD outputs after theStart/Stop button S1 has been pressed. Thesepulses drive MOSFETs T1 and T2 alternately intosaturation. Transformer Tr1 boosts the voltage ofthese pulses by a factor of around 10, and they arethen applied to the electrodes I and II. That’s allthere is to it. The strength (amplitude) of the pulses

low-intensity pulsating current is ade-quate for this. With some therapeutictreatments, the only objective is to createa general sense of wellbeing, which canbe achieved using even weaker shocks. Ifon the other hand the equipment is usedto strengthen the muscles, the pulsesthat are used are significantly stronger.

However, we don’t want say too muchabout this, since you might otherwise getthe false impression that we are expertson the subject. Just to avoid any misun-derstanding, let us say that this isabsolutely not the case. All that we aredoing is making the necessary equipmentavailable, in response to repeatedrequests. From what we have been ableto find out, the technology used in thissort of equipment is in fact fairly simple.We have translated this technology into aDIY design that is as safe and reliable as

possible. That is our job and our skill.We would rather not say anything

about how to use the equipment. Instead,we strongly recommend that you consulta doctor or other specialist. You definitelyshould not use the muscle stimulator ifyou have not first informed yourself aboutwhat can and cannot be achieved withelectrical muscle stimulation. Further-more, do not regard this as a kind of toy.Even though it is powered by a battery, itcan without question produce veryintense current pulses, and these couldhave undesirable effects if the equipmentis not used properly. Depending on theskin resistance, the peak voltage acrossthe electrodes is 35 to 45 V; with no load itcan easily reach 50 V!If all that you are interested in is looseningup certain muscles or enjoying a relaxingmassage, we would actually advise you

GENERALINTEREST

134/2000 Elektor Electronics

LD1

7 a6

b4

c2

d1

e9

f10

g

3 8

5

dp

CA CA

HD1131 O HD1131 O

LD2

10dp

CA CA 7a6

b4

c2

d1

e9

f

g

3 8

5

INT0/P3.2

INT1/P3.3

P1.0/IN+

P1.1/IN–

RXD/P3.0

TXD/P3.1

89C2051

T0/P3.4

T1/P3.5

IC1

P1.2

P1.3

P1.4

P1.5

P1.6

P1.7P3.7

RST

X1 X2

20

10

12

13

14

15

16

17

18

1911

5 4

2

3

1

6

7

8

9

R51k

R61k

R71k

R41k

R81k

R91k

R101k

R12

1k

R11

1k

X1

12MHz

C3

33p

C4

33p

S4S3S2S1

C1

100n

R3

4k

7

R2

4k

7

BUZ11

SG

D

T2T1

BUZ11D7

39V400mW

D8

39V400mW

D1

TIME

D2

PROG

R1

1k

D6

1N4148

D3D4

R14

4k

7

Tr1

R16

1k

R15

1k

C5

47µ25V

T3

BD140

R13

22

k

2xINTENSITY

MAX

MIN

CW

START/STOP UP DOWN PROG/TIME

7805

IC2

C6

100µ16V

C7

100µ16V

D5

1N4001

S5 5V

5V

5V

BE

C

BD140P1

50klin

C2

10µ 63V

R17

10

k

000041 - 11

20W

0 4 16

9V

+U

+U ON/OFF

Figure 1. The use of a pre-programmed microcontroller allows the hardware to be kept to a minimum.

Page 3: Low-Impact Muscle Stimulator - ToulonuXtoulonux.org/documents/electrostim/e004012_low-impact muscle... · Low-Impact Muscle Stimulator L.I.M.S. with treatment programs ... electrical

is controlled by simple current source formed by T3,which can be adjusted using P1. The current sourcecharges electrolytic capacitor C5, and the voltageon this capacitor is used for the output pulses. Thisapproach also ensures that only a certain limitedamount of energy reaches the transformer for eachtime interval, which is very important with regard tosafety. LEDs D3 and D4 act as a simple amplitudeindicator for the output pulses.

This takes care of the pulse generation. Now youmay be wondering why we have used a micro-processor for the pulse generator, since this couldbe done just as well with a couple of transistors anda 555 timer IC. The presence of the processor isentirely due to a number of extra functions thathave been included in the stimulator. These extrafunctions also explain the presence of the two-digitdisplay (LD1 and LD2), the two indicator LEDs D1and D2 and the control buttons S1 through S4.

The buttons can be used to select a specifictreatment time and a particular treatment program.You will see shortly how this is done. To avoid anymisunderstanding, we should say right away thatoutput pulses can be produced only after both atime and a program have been selected. Until thishas taken place, pressing the start/stop button S1has no effect. This has been done for safety rea-sons, in order to prevent the stimulator from unex-pectedly delivering a continuous current at anunguarded moment.

Before we go any further into the details of theprogram options of the stimulator, we must firstdivert our attention to a more detailed descriptionof the pulse signals that it produces.

PulsesAfter immersing ourselves deeply into theavailable literature, we came to the clearconclusion that the developers of suchdevices are remarkably unanimous withregard to the width and repetition rate ofthe pulses. Based on this, we havedecided on alternating positive and neg-ative pulses with a width of 256 µs and arepetition rate of 109 Hz. This corre-sponds to an idle time of 4.61 ms.

The pulses do not appear continuouslyat the output, but rather in packets orbursts with a duration of around one sec-ond. A typical burst is illustrated in Fig-ure 2. This naturally does not show allthe pulses in the burst, since we’d needa lot bigger page for that.

The interval between successivebursts can be set between 0.5 and 4.5seconds. The resulting signal on the elec-trodes appears as shown in Figure 2b.The intensity of the shocks dependsalmost as much on the repetition intervalas on the pulse amplitude set by P1. Thevarious treatment programs take advan-tage of this fact.

Program and timeAs already noted, two things mustselected before the circuit can start deliv-ering pulses: the treatment time and thetreatment program.

The time is set by selecting the ‘time’mode using button S4. The ‘time’ indi-cator LED D1 is on when the unit is inthe ‘time’ mode. After this, the time can

be set between 1 and 19 minutes usingthe up and down buttons S2 and S3.The time is shown on the display. Thepulses will stop after the selected timehas expired.

If the unit is switched to the ‘program’mode by pressing button S4, the indica-tor LED D2 lights up. In this mode, one ofa total of fifteen programs can beselected. The programs labelled 1 thor-ough 9 are the simplest ones, since theywork with a constant intensity levelbased on selecting the burst intervaltime. When the display shows a ‘9’, theintensity is at maximum, since the mini-mum burst interval of 0.5 s is used. Pro-gram 8 corresponds to an interval time of1 s, ‘7’ represents an interval time of 1.5 sand so on. This ‘illogical’ reversesequence has been chosen intentionallyso that the strongest program has thehighest number.

The programs indicated by ‘A’ through‘F’ are more sophisticated, since variousintensity levels are used during thecourse of the treatment session. There areactually only three different programs,each in a weak and a strong version.

Figure 3 shows how programs Athrough F are set up. The intensity isdivided into the same levels as for pro-grams 1 through 9. This means that theburst interval is the shortest, and theintensity the greatest, at level 9. The hor-izontal axis shows the treatment time —which as you know can vary.

In programs A and B, the intensity ofthe pulses is increased to its maximum

GENERALINTEREST

14 Elektor Electronics 4/2000

256 µs

1 s 0.5 ... 4.5 s

adjustable

000041 - 12

4.61 ms

ca.1 s

a

b

1

0.06 0.12 0.18 0.24 0.30 0.36 0.42 0.48 0.54 0.60

2

3

4

5

6

7

8

9

inte

nsity

time

= A= B

1

0.06 0.12 0.18 0.24 0.30 0.36 0.42 0.48 0.54 0.60

2

3

4

5

6

7

8

9

inte

nsity

time

= C= D

= E= F

1

0.06 0.12 0.18 0.24 0.30 0.36 0.42 0.48 0.54 0.60

000041 - 13

2

3

4

5

6

7

8

9

inte

nsity

time

Figure 2. The pulse signal produced by the muscle stimulator.

Figure 3. There is a choice of three differentarrangements for the automatic treatmentprograms, each with two intensity levels.

Page 4: Low-Impact Muscle Stimulator - ToulonuXtoulonux.org/documents/electrostim/e004012_low-impact muscle... · Low-Impact Muscle Stimulator L.I.M.S. with treatment programs ... electrical

value in three steps, following which it remains atthis level for a reasonable length of time and thendrops back down in steps. Program A is the weakversion, and program B is the strong version.

In programs C and D, the treatment time isdivided into three equal intervals, and the intensityis gradually increased in each of these intervals.Once again, C is the weak version and D is thestrong version.

In programs E and F, the treatment time isdivided into five intervals, and the intensity alter-nates between two levels in each of these intervals.Version F is the stronger of the two.

The durations of the individual phases of pro-grams A through F naturally depends on the

GENERALINTEREST

154/2000 Elektor Electronics

Project softwareDisk # 000041-1 supplied through Readers Services

LIMS 2.734 21-10-00 10:49 limsLIMS LST 86341 21-01-00 10:49 lims.txtLIMS HEX 5.881 21-01-00 10:49 lims.HEXLIMS PRJ 2.112 21-01-00 11:33 lims.prjCONTENTS TXT 0 08-02-00 11.49 contents.txtLIMS M51 4.200 21-01-00 10:49 lims.m51LIMS C 16.984 08-02-00 11:38 Lims.CLIMS OBJ 5.683 21-01-00 10:49 Lims.OBJREADME TXT 220 08-02-00 11:43 Readme.txt

(Sophos data integrity and fingerprint files not listed)

SoftwareThe software contained in the microprocessorconsists of a main routine and an interrupt routine.The interrupt routine handles all real-time pro-cessing, such as keeping track of the time, generat-ing the pulses and scanning the keypad (withdebouncing). Pulses are generated as soon as themain routine has set a flag. In the reverse direction,the interrupt routine sets flags whenever it detectsthat a valid button has been pushed. The time (inthe form of a counter with a resolution of half asecond) is also made available to the main routine.The main routine responds to the pushbuttons,drives the display and determines when outputpulses must occur.The processor works with a 12-MHz crystal. Theinternal Timer 0 works in mode 3, which meansthat it divides by 256 (8-bit counter) and has aninput frequency of (12 MHz ÷ 12) = 1 MHz.Timer 0 thus generates an interrupt every 256 µs.The number of interrupts is counted and used tocontrol the circuit. For example, output P30 is sethigh when the interrupt count is 2 and it is set lowagain when the count is 3. Output P31 is set highwhen the count is 20 and again low when thecount is 21.At a count of 37, the counter resets itself (to 1)and a new cycle begins. The time between twopulses is thus 18 times 256 µs, or 4608 µs. A fullcycle lasts 36 times 256 µs, which is 9216 µs. Eachcycle is always completed once it is started. Time-keeping is based on counting 1964 interrupts forevery half second. Here a full second is not exactlyone second, but instead 1.000448 s.The main program actually consists of a statemachine. In order to prevent the accidental gener-ation of pulses, there are two parts, one with andthe other without pulses. Only after all of the pre-requisites have been satisfied is it possible to trans-fer to the part that has pulses.In the idle state, the software checks to see if a button is pressed. Until this happens, the program runs in an endless loop. As soon as abutton is pressed, the program takes action. It switches over to the part of the program with output pulses only after a program has beenselected, the treatment time has been set and the start button has been pressed.Based on the treatment time and the selected treatment program, the main program then periodically sets a flag that causes the interruptroutine to generate pulses. When the stop button is pressed or the time has expired, the other half of the program is again called.The ‘running 8’ display is an exception, in that it is generated in the interrupt routine rather than in the main program.

interruptcounter: = interruptcounter + 1

Interrupt routine of the Low Impact Muscle stimulator (LIMS)This routine executes on every overflow of the 8 bit timer 0 (timer 0 is used in Mode 3)

The timer counts at CLOCKFREQ/12, so the interrupt routine executes every 256 µs (@12 MHz crystal)

are all input-key values the sameas the previous interrupt cycle ?

output counter = 0 ?

outputcounter = outputcounter + 1

on outputcounter = 2 activate positive out

on outputcounter = 3 deactivate positive out

on outputcounter = 18 activate negative out

on outputcounter = 19 deactivate negative out

on outputcounter = 37 outputcounter: = 0

000041 - 14

outputcounter =outputcounter + 1

keep positive andnegative outputs high

output flag = 1 ?

"running segments" displaymode flag = 1 ?

interruptcounter: = 0

display next segment every 1/16 second

treatment time: =treatment time - 1

HalfSecCounter: = HalfSecCounter + 1

HalfSecondsCounterMOD120=0 ?

interruptcounter > number of interrupts that fit in 1/2 sec.AND treatment is running ?

true

true false

false

true

true

execute code for key debounce and key repeatThis code sets "pressed keys flags" or pressed keys

to 1 if these keys are pressed during a numberof interrupt cycles (about 40 cycles).

This code also sets the "keypressed" flags after alarger number of interrupt-cycles (about 2500).

In this way we create a key-repeat effect.

Save key values, overwriting the values of theprevious interrupt cycle

false

true

true

false

false

Page 5: Low-Impact Muscle Stimulator - ToulonuXtoulonux.org/documents/electrostim/e004012_low-impact muscle... · Low-Impact Muscle Stimulator L.I.M.S. with treatment programs ... electrical

selected treatment time. The microprocessoralways calculates the correct distribution.

Starting and stoppingAs soon as the start/stop button S1 is pressed afterthe time and program have been selected, the stim-ulator starts to generate pulses and a ‘running 8’appears on the display. As long as pulses are beinggenerated, it is not possible to change the selec-tions, and only the start/stop button can be used.If any of the other buttons is pressed, the displayshows the number of the selected program or theremaining treatment time. The display automati-cally returns to showing a running ‘8’ after aboutfive seconds.

The program can be interrupted by pressing onthe start/stop button. This amounts to a sort ofpause function, since the program continues if thebutton is pressed again. While the program ispaused, the program number or the remaining time

is displayed, and it is possible to changethe settings.

After the set time has expired, pulseproduction stops automatically and thedefault values are shown on the display:‘–‘ for the program and ‘0’ for the time.New values for both of these must beselected before a new treatment cyclecan be started.

Safety and powerWith a circuit such as this, safety musthave the highest priority. We have thusgiven it the proper amount of attention.

An important point is that pulses cannever be produced ‘by accident’ if thestart/stop button is pressed at the wrongtime. It is always necessary to intention-ally select a program and a time beforethis button has any effect.

The benefit provided by the currentsource T3 has already been mentioned. Itlimits the amount of energy that is sup-plied to the transformer. If a voltage reg-ulator had been used instead of a currentsource, the voltage would try to remainconstant, with the consequence thatoccasionally much more current wouldbe delivered than is actually necessary.This could lead to dangerous situations.

However, using a current source alsohas an unavoidable disadvantage, whichis that when there are no output pulsesthere is no current consumption, and con-sequently capacitor C5 will be charged tothe maximum voltage. The first pulsesafter this could thus have a dangerouslyhigh intensity. This is countered by hav-ing the microprocessor drive both tran-sistors (T1 and T2) at the same time when

GENERALINTEREST

16 Elektor Electronics 4/2000

(C) ELEKTOR

000041-1

C1

C2

C3

C4

C5

C6 C7

D1

D2

D3

D4

D5

D6

D7 D8

H1

H2 H3

H4

IC1

IC2

LD1

LD2

P1

PC1

PC2

PC3

PC4

R1

R2

R3

R4R5R6R7R8R9

R10R11R12

R13

R14 R15R16

R17

S1

S2

S3

S4

S5

T1

T2

T3

TR1

X1

00

00

41

-1

20W

16440

+ 0

(C) E

LEK

TOR

00

00

41

-1

Figure 4. Aside from the processor, the transformer and the displays, there aren’t a lot of components on the circuit board.

Page 6: Low-Impact Muscle Stimulator - ToulonuXtoulonux.org/documents/electrostim/e004012_low-impact muscle... · Low-Impact Muscle Stimulator L.I.M.S. with treatment programs ... electrical

circuit board and have carefully inspected it (payparticular attention to the quality of the solder jointsand watch out for possible shorts!), you can look fora suitable enclosure. The type shown in the com-ponents list is attractively compact, but it’s a bit onthe small side. If your transformer turns out to besomewhat larger than the one used in our prototypemodel, the whole assembly may not fit in this enclo-sure, and you will have to look for something bigger.

As indicated on the component layout drawing,a number of components (such as the transistorsand the voltage regulator) should be mounted flaton the board in order to limit the height of theassembled board. The photograph of the board inFigure 5 clearly illustrates this.

The board is mounted using four small screwsthat are glued to the rear side of the front panel ofthe enclosure. However, it’s better to make the firsttests without the front panel in place, since it’s eas-ier to access the board in this case. If there are anyproblems, it’s then easier to check whether the sup-ply voltage is present at certain key points.

We designed a front panel for our prototype thatmatches the enclosure shown in the componentslist, as shown in Figure 6. Of course, you are freeto choose something different if you want to.

ElectrodesIn our experiments, we were able to successfullyuse homemade electrodes consisting of smallpieces of unetched printed-circuit board materialaround 2 by 3 cm in size. These were clamped toan arm or leg using elastics bands. To make a goodcontact, place a small piece of paper that is gener-ously moistened with ordinary tap water between

no pulses are being generated. This dis-charges C5, and the first pulses will actu-ally be weaker than normal. This is apleasant side effect, since it results in asort of ‘soft start’ effect for each cycle.

The choice of the power supply is alsoa critical issue with regard to safety. Thiscircuit must be powered only from a bat-tery, and absolutely not from a mainsadapter. If you use rechargeable batteries,the battery charger must never be con-nected to the stimulator while it is in use.

The power source can be either sixpenlight cells or a 9-volt battery, asdesired. The current consumption is

around 50 mA, which is not all that low.However since a treatment session lastsat most 19 minutes, even a 9-V batterywill last for at least 30 sessions. If youreplace the voltage regulator (IC2) witha low-drop type (LM2940-5 or 4805), thebattery can be used until it is dis-charged to 6 V, and the number of ses-sions that you can get from a single bat-tery is even greater.

ConstructionThe printed circuit board for the musclestimulator is shown in Figure 4. Sincethere are actually only very few compo-nents in this circuit, we hardly need tosay much about building it. If you use thecomponents specified in the componentslist and mount them as shown in the lay-out drawing, there’s not much that cango wrong.

Transformer Tr1 is what is called a‘100-V’ output transformer. Even thoughoutput transformers are not all that pop-ular any more, it is reasonably easy toobtain. Both Conrad Electronics andMonacor (or Monarch in some countries)still have this type of transformer in theircatalogues. We need a type that has0/4/16-Ω outputs and can deliver 20 W.Although it’s true that we do not needanything like 20 watts here, this is theonly transformer that has the correctwinding ratio (1:10).

After you have finished assembling the

GENERALINTEREST

174/2000 Elektor Electronics

COMPONENTS LIST

Resistors:R1,R4-R12,R15,R16 = 1kΩR2,R3,R14 = 4kΩ7R13 = 22kΩR17 = 10kΩP1 = 50kΩ linear potentiometer

Capacitors:C1 = 100nFC2 = 10µF 63 V radialC3,C4 = 33pFC5 = 47µF 25V radialC6,C7 = 100µF 16V radial

Semiconductors:D1-D4 = high-efficiency-LED, red

(3 mm)D5 = 1N4001D6 = 1N4148D7,D8 = zener diode 39V/0.4WT1,T2 = BUZ11 (BUZ10,

BUZ100)T3 = BD140IC1 = 89C2051-12PC

(programmed), order code000041-41

IC2 = 7805 (or LM2940T-5, 4805)

Miscellaneous:S1-S4 = Digitast pushbutton

(ITT/Schadow)S5 = on/off switchTr1 = 100-V line output

transformer 0-4-16 Ω/20 W(Conrad Electronics orMonacor/Monarch)

X1 = 12 MHz quartz crystal LD1,LD2 = HD1131OEnclosure: e.g., TEKO 160x95x61

mmPC1-PC4 = miniature wander

socketsDisk, project software (source

code and hex code): order code000041-11 (see ReadersServices)

Figure 5. The transistors and the voltage regulator are mounted flat on the board, in orderto limit the height of the assembly.

Page 7: Low-Impact Muscle Stimulator - ToulonuXtoulonux.org/documents/electrostim/e004012_low-impact muscle... · Low-Impact Muscle Stimulator L.I.M.S. with treatment programs ... electrical

the electrode surface and the skin (don’t use dis-tilled water, since it doesn’t conduct electricity).The separation of the two electrodes must be deter-mined experimentally; it will usually lie somewherebetween 5 and 30 cm.

If you plan to use the stimulator regularly, it isbetter to use a more durable material for the elec-trodes. You might consider purchasing ‘real’ elec-trodes with Velcro fastenings. These can usually beobtained in specialist medical equipment shops.

The connection between the electrodes and thestimulator can best be made using lengths of flex-ible wire (test lead cable) with miniature bananaplugs at each end.

Initial operationTo check out the stimulator once you have built it,proceed as follows:– Connect a 9-V battery, turn P1 completely to theleft and switch on the device with S5. The ‘pro-gram’ LED D2 must now be on, and ‘Lims’(for ‘Low-impact muscle stimulator’) should appearon the display, followed by a static dash. The dashmeans that no program has been selected. If thisdisplay indication appears, you can be practicallycertain that the processor part of the circuit is work-ing properly.– Press on the ‘up’ button. A ‘1’ should appear onthe display. Each time that the ‘up’ button ispressed again, the program number will beincreased by one, while pressing on the ‘down’ but-

ton will reduce it by one.– Press the ‘program/time’ button S4. A‘0’ will appear on the display, and the‘time’ LED will be lit. Choose a timebetween 1 and 19 minutes using the upand down buttons.– Press on the ‘start/stop’ button. A linesegment will start to run around on thedisplay in the form of an ‘8’, as a sign thatthe circuit is active.– Turn the amplitude control P1 slowly tothe right. LEDs D3 and D4 should light upat a certain point, and they shouldbecome brighter as P1 is turned further tothe right.– If everything is OK up to now, turn P1back to its minimum position (left) andswitch off the stimulator with S5.– Connect two electrodes to the stimula-tor, and clamp them to your leg (for exam-ple) about 10 cm apart. Switch on thestimulator, and select program 8 and atime of 2 or 3 minutes.– Check that P1 is turned to the mini-mum level, and then press the start/stopbutton. You will not feel anything at thispoint.– Turn P1 slowly to the right. At a certainpoint, you should feel a slight tingling.Cautiously turn P1 further until the desiredintensity is reached. Don’t overdo it!

The circuit also has a test program forchecking that it is working properly. Thiscan be run by holding the up and downbuttons pressed while you switch the uniton with S5. If you then press thestart/stop button, all segments of the dis-play should light up and LEDs D1 and D2should blink. You can exit the test pro-gram by pressing on the program/timebutton, which causes the software num-ber to appear on the display.

(000041-1)

Text (Dutch original): S. van Rooij

GENERALINTEREST

18 Elektor Electronics 4/2000

Low Impact Muscle Stimulator

L I M S

000041 - F

UPINTENSITY

TIME

PROG

DOWN

POWER

PROG/TIME

START/STOP

MAX

OFF

ON

MIN

Figure 6. This front panel design matches the TEKO enclosure noted in the components list.