lorad ffdm qc procedures for medical physicist...– select a computer-generated flat-field image...

48
Lorad FFDM QC Procedures for Medical Physicist Tao Wu, Ph.D. Hologic, Inc.

Upload: others

Post on 05-May-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Lorad FFDM QC Procedures for Medical Physicist

Tao Wu, Ph.D.Hologic, Inc.

Page 2: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Lorad Selenia FFDM

Page 3: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Tests Following Lorad QC Manual

• Collimation Assessment• Artifact Evaluation• System Resolution• Phantom Image Quality Evaluation• Automatic Exposure Control Function Performance• Automatic Exposure Control Reproducibility • Signal-To-Noise and Contrast-To-Noise

Measurements• Softcopy Workstation QC

Page 4: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Acquisition Workstation Software

Page 5: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

OS LoginMedical Physicist

username: phys password: phys

Page 6: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Detector Warm-UpAfter logging in OS, a dialog showing detector warm-up status appears

If the system has been running - Dismiss

If the system has been shut down - Wait for 30-40 min (or dismiss it and start with tests in which detector is not involved, e.g. HVL, tube output)

Page 7: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Application LoginUsername:AndreaPassword: Andrea

Page 8: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Turn Off Auto-decompression

Page 9: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Send Images to Output Devices

Page 10: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Send Images to Output Devices

Page 11: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Send Images to Output Devices

Page 12: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Non-Imaging Mode

Page 13: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Create a New Patient Account

Select “No Output”

Page 14: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Create a New Patient Account

Page 15: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Create a New Patient Account

Page 16: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Selenia QC Tests

Page 17: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Tests Following Lorad QC Manual• Collimation Assessment• Artifact Evaluation• System Resolution• Phantom Image Quality Evaluation• Automatic Exposure Control Function Performance• Signal-To-Noise and Contrast-To-Noise

Measurements• Softcopy Workstation QC

Page 18: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Order of Testing

• Artifact Evaluation• Phantom Image Quality Evaluation• SNR and CNR Measurements• AEC Function Performance • Other Tests

Page 19: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Artifact EvaluationEvaluation of detector, x-ray tube, filter etc

– 4 cm thick acrylic block– No compression paddle– Four flat field images

Mo/Mo, contact mode, LFS with grid Mo/Rh, contact mode, LFS with grid Mo/Mo, mag mode, SFS without grid Mo/Rh, mag mode, SFS without grid

– Review the images on acquisition workstation monitorWindow width = 500, Level = 350

Page 20: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Artifact Evaluation

Page 21: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Artifact Evaluation

Page 22: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Artifact EvaluationEvaluation of Printer

– Select a computer-generated flat-field image– Print the image and evaluate it– Repeat the test for each printer

If a printer is connected to multiple Selenia systems, it only needs to be tested once from one Selenia.

Page 23: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Artifact Evaluation

Page 24: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Phantom Image Quality Evaluation• ACR phantom with acrylic disc• AEC position 2• Print film• Measure film densities

- Background density > 1.20, with control limit ± 0.20- Density difference over the acrylic disk > 0.40, with

control limit ± 0.05

• Phantom scoring (hard copy and softcopy station)• 5 fibers, 4 speck groups, 4 masses

Page 25: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Phantom Image Quality Evaluation

Page 26: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

SNR and CNR Measurements

Page 27: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

SNR and CNR Measurements• SNR ≥ 40• Establish the reference CNR value during

the acceptance test• CNR should stay within ±15% of the

reference

Page 28: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Automatic Exposure Control (AEC)• AEC modes

Auto-Filter: filter, kVp, mAs all automatically determined Auto-kV: filter manually selected

kVp and mAs automatically determined Auto-Time: filter and kVp manual selected

mAs automatically determined

• AEC Method- Filter and kVp are determined by breast thickness (in Auto-Filter and Auto-kV modes)- A pre-exposure is used to determine mAs

• AEC positionsAuto-AEC (The densest tissue gets enough exposure) Seven manual positions (marked on compression paddle)

Page 29: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

AEC Function Performance1. Performance at different thickness and

imaging modes 2. Performance at different compensation

levels

Page 30: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

AEC Function Performance1. Performance at different thickness and

imaging modes – Acrylic or BR-12 blocks – AEC position 2– Contact mode (LFS, grid): 2, 4, 6, 8 cm phantom

Mag mode (SFS, no grid): 4 cm phantom– Variation of target pixel values≤ 10% of the mean target pixel value

Page 31: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

AEC Function Performance

Page 32: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

AEC Function Performance2. Performance at different compensation levels– 4 cm acrylic or BR-12– Compensation levels:

Step -5 to Step +5– Use average target count at

Step 0 as reference– Compare target count at

other steps to the reference

Step -5 0.23 – 0.28Step -4 0.36 – 0.44Step -3 0.50 – 0.61Step -2 0.63 – 0.77Step -1 0.77 – 0.94Step +1 1.04 – 1.27Step +2 1.17 – 1.43Step +3 1.31 – 1.60Step +4 1.44 – 1.76Step +5 1.58 – 1.93

Page 33: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Collimation Assessment1. X-Ray Field to Light Field2. X-Ray Field to Image Receptor3. Compression Paddle to Image Receptor

Page 34: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Collimation Assessment• One test for 24×29 cm imaging mode• Three tests for 18×24 cm imaging modes

– Left imaging position– Center imaging position– Right imaging position

L RM

Page 35: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Collimation AssessmentUse “coin method” as described in ACR manual

Page 36: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Collimation Assessment1. X-Ray Field to Light Field

– Measure deviations between the edges of the light field (exterior edge of the coin) and the x-ray field at al four sides

– Sum of the magnitude of deviations on opposite sides ≤ 2% of the SID SID = 66 cm 2% of the SID = 1.32 cm

Page 37: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Collimation Assessment2. X-Ray Field to Image Receptor

– d1: measured on filmdistance from the inner edge of the coin to the edge of the x-ray field

– d2: measured on digital imagedistance from the inner edge of the coin to the edge of the detector

– (d1×1.06 ) - d2

– The x-ray field shall extend beyond the chest wall side of the image receptor

– The x-ray field must NOT extend >2% of the SID at any of the four sides of the image receptor

Page 38: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Collimation Assessment3. Compression Paddle to Image Receptor– Measure distance from the outer edge of the coin to

the edge of the image – The anterior edge of the compression paddle must be

aligned just beyond the chest wall edge of the image receptor, but ≤ 1% of the SID

Page 39: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Collimation Assessment

Page 40: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Collimation Assessment

Page 41: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Collimation Assessment

Page 42: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Evaluation of System Resolution

• Line pair phantom• 4 cm acrylic block• Line pair phantom placed

at 45º to the detector row

Page 43: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Evaluation of System Resolution

Limiting spatial resolution ≥ 7 c/mm (lp/mm)

Page 44: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Detector Ghosting (Optional)1. Create ghosting effect2. Quantify ghosting effect

Page 45: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Detector Ghosting (Optional)1. Create ghosting effect

Ghosting area

Page 46: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Detector Ghosting (Optional)2. Quantify ghosting effect

Ghosting area

Page 47: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Detector Ghosting (Optional)2. Quantify ghosting effect

ROI1 in background without ghosting

ROI2 in Al filter without ghosting

ROI3 in Al filter with ghosting

• | Ghosting factor | ≤ 0.3

Ghosting Factor =

meanregion2 - meanregion3

meanregion1 - meanregion2

Page 48: Lorad FFDM QC Procedures for Medical Physicist...– Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer

Thank you!

Tao [email protected]: 203-207-4574