LIP - LABORATORY OF INSTRUMENTATION AND EXPERIMENTAL PARTICLE PHYSICS Lisbon, PORTUGAL THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 2010 The ClearPEM Breast Imaging Scanner Jorge A. NEVES On behalf of the ClearPEM Collaboration PEM Collaboration

Upload: gudrun

Post on 11-Jan-2016

20 views

Category:

Documents


0 download

DESCRIPTION

THE 12 TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 2010. The ClearPEM Breast Imaging Scanner. Jorge A. NEVES On behalf of the ClearPEM Collaboration. PEM. Collaboration. LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL. 2 /. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

LIP - LABORATORY OF INSTRUMENTATION AND EXPERIMENTAL PARTICLE PHYSICSLisbon, PORTUGAL

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION

15-20 FEV 2010

The ClearPEM Breast Imaging Scanner

Jorge A. NEVESOn behalf of the ClearPEM Collaboration

PEMCollaboration

Page 2: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

Motivation

Breast Cancer and Positron Emission Tomography

01

10

11 np2/ ν

γ 511keVγ 511keV e+ e-

Breast cancer is the most common cancer among women mean incidence rate of 1.2 million females per year worldwide

PET is a functional imaging technique that has demonstrated large potential for breast cancer detection - Positron Emission Mammography (PEM)

Patient is injected with 18F-FDG radiotracer that fix in tumor cells and decays by positron emission. The 2 γ photons resulting by the positron-electron annihilation are detected in temporal coincidence to imaging the biodistribution of the radiotracer.

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 20102/18

Page 3: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

The ClearPEM Scanner

Detector Calibration

Energy and Time Resolution

Image Reconstruction

Conclusions

Outline

The ClearPEM Scanner

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 20103/18

Page 4: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

Good spatial resolution (~1.5 mm in whole FoV) Fine crystal segmentation (2x2 mm)

Dual APD readout of individual crystal pixels

DoI measurements with good resolution (FWHM ~2 mm)

High Sensitivity Long LYSO:Ce crystal (20 mm)

Two detector plates with large active area (17x15cm2 FOV)

Reduced Random Background (~30%) Large flux of single photons (up to 10 MHz)

Coincidence time resolution of ~4 ns FWHM

The ClearPEM Scanner

The ClearPEM Characteristics

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 20104/18

Page 5: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

ClearPEM MOVIE

The ClearPEM Scanner

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 20105/18

Page 6: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

The ClearPEM Scanner

The ClearPEM Detector ModulesTwo Detector Plates

160x180 mm2 active area

6144 scintillation crystals LYSO:Ce (emit visible light when high energy photons interact with them)

12288 APD pixel channels ( Highly sensible photo-detector. Generate pulses in response to scintilation light)

Double readout of crystal pixels for Depth-of-Interaction measurent (to minimize parallax effect)

Water cooling system (18.0 ± 0.1 °C)

Hamamatsu S8550 Avalanche Photo Diode

LYSO:Ce 2x2x20 mm3 crystal

4x8 crystal matrices made of BaSO4 walls

384 APD arrays, Operating Voltage 350-450V

ε ~ 82% @ 511 keV

7.4 g.cm-3

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 20106/18

Page 7: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

The ClearPEM Scanner

FrontEnd ASIC

ASIC or APD signal processing

Charge Amplifier Characteristics Technology: AMS 0.35 μm CMOS, 70 mm2 Area

Input: 192 channels

Output: 2 highest channels (192:2 mux)

-> readout Compton events

Max Input Charge: 90 fC

Noise: ENC ~1300 e- (Baseline RMS = 2.2 ADC counts = 5 keV)

Shaping: 40 ns

Analog Memories: 10 pulse samples

Clock Frequency: 50-100 MHz

Power: 3.6 mW/channel

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 20107/18

Page 8: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

12 cm

4.5 cm

ASICs (2x192 channels)

FrontEnd Board

Modules (12x32 crystals, 24 APDs)

Super Module 2 FrontEnd Boards and 12 Detector Modules Processes 768 APD channels

FrontEnd Board Contains 2 ASICs for signal selection 2 High-speed dual ADCs (10bits, 100MHz) 1 LVDS Channel Link Transmitter (600Mbps)

Detector Head 8 Super Modules (16 for both DHs) 1 Service Board (HV & LV distribution,

temperature monitoring) 2 water cooling plates

The ClearPEM Scanner

FrontEnd Electronics

192 Detector Modules (96 per DH)

192 Detector Modules (96 per DH)

HV matrix

Service Board

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 20108/18

Page 9: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

The ClearPEM Scanner

Off-Detector and Data Acquisition Electronics

DAE System – L1 Trigger/DAQ4 DAQ Boards (Slave)

8 XlinxTM FPGA First data filtering to identify usefull data (find Top-

Bottom crystal coincidences) Check signal integrity calculating basic parameters Send relevant data to TGR/DCC Board

1 TGR/DCC – Trigger & Data Concentrator Board (Master) 1 XilinxTM FPGA DAQ Board’s arbitration System’s Sync and Reset Responsible for the identification of coincidence

between detector heads Sends relevant data to Acquisition Server (S-Link

Bridge)

Data Transfer Bandwidth

6.4 Gbps Trigger/DCC Acquisition Server

800 MBps (520) @ S-Link FedKit (PCI)

60 MBps (42) @ USB 2.0

400 MBps Storage Rate

Coincidence Triggering Rate 800 kHz

19’’ crate 2cPCI backplanes

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 20109/18

Page 10: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

The ClearPEM Scanner

Scanner Integration @ IPO

Portuguese Institute of Oncology - Porto

Examination BedDetector Heads

Robotic structure

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 201010/18

Page 11: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

511 keV

APD pixel

APD pixel

LYSO:Ce Crystal

Relative Gain

Distributio

n of Energ

y Calib

rations C

onstants

3 Calibration constants per crystal (Top and Bottom readout)

Detector Calibration

Absolute Gain

Distribution of pulse peak timeRequires > 4%/mm for DOI resolution < 2 mm FWHM

DOI Calibration

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 201011/18

Page 12: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

Tmax

Tpeak

Photon time is extracted from the pulse samples fitted by the function:

Energy and Time Resolution

Time Measurements

Typical pulse 50 MHz sampling

The coincidence time resolution of the whole scanner is 5.2 ns FWHM

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 201012/18

Time C

alibratio

n

Page 13: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

Energy and Time Resolution

Energy MeasurementsAverage energy resolution at 511 keV for the

full scanner is 16.0 %

22Na spectra for all crystals

Photopeak measurements

511 keV photopeakcompton

Good energy linearity Energy resolution and photopeak position not dependent of DOI

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 201013/18

Page 14: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

Without DOI Information

With DOI Information

1.2 mm FWHM

Image Reconstruction

ClearPEM Spatial ResolutionPoint Source Imaging

22Na point source in a grid with 5 mm pitch

Energy window 400-600 keV

Sinograms of 16 source positions are added

3D-OSEM/STIR Reconstruction

Spatial Resolution Transaxial 1.2 mm FWHM

(corrected by source size ~ 1 mm)

DOI Effect Images without using DOI information

show considerable blurring

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 201014/18

Page 15: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

Image Reconstruction

ClearPEM Spatial ResolutionDerenzo Phantom Imaging

Sealed phantom with 22Na gel

20 μCi activity (T1/2 = 2.6y)

Active area: 35 mm Ø x 38.1 mm length

Phantom Draw

3.0 mm

2.5 mm

2.0 mm

1.5 mm

1.2 mm

Dist. = 150 mmTakes = 4 x 20 min

450-600 keV energy window6 ns time window

3.0 mm

2.5 mm

2.0 mm

1.5 mm

1.2 mm

3996 0

3D-OSEM

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 201015/18

Page 16: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

Image Reconstruction

Initial Clinical TestsExample os a typical exam

dose 7.6 mCi 150 mm detector plate opening 4 angular orientations coincidence windom ± 4 ns energy window 400 - 650 keV low coincidences rate ~1.2 kHz fraction of randoms in FoV IS 35%

Reconstruction 3D-OSEM simple normalization correction randoms, attenuation and scatter correction not applied 22Na source data added to sinogram, emulating lesion (L/B~4 for 3 mm lesion)

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 201016/18

Page 17: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

ClearPEM-Sonic

Conclusions

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 201017/18

ClearPEM electronics is one of the most innovative systems available for APD-based PET systems

Excellent detector performance

Time Resolution: 5.2 ns FWHM

Energy Resolution: 16 %

Spatial Resolution: 1.2 mm FWHM

Initial clinical trials have been started

Needs and efforts on image corrections

Page 18: LIP - L ABORATORY OF I NSTRUMENTATION AND EXPERIMENTAL P ARTICLE PHYSICS Lisbon, PORTUGAL

ClearPEM-Sonic

The ClearPEM Breast Imaging Scanner Jorge A. NEVES

THE 12TH VIENNA CONFERENCE ON INSTRUMENTATION 15-20 FEV 201018/18

E. Albuquerque1, F. G. Almeida2,13, P. Almeida3, E. Auffray10, J. Barbosa2, A. L. Bastos9, V. Bexiga1, R. Bugalho4, C. Cardoso4, S. Carmona8, J.F. Carneiro2, B. Carriço4, C. S. Ferreira4, N. C. Ferreira5, M. Ferreira4, M. Frade4, F. Gonçalves1, C. Guerreiro5, P. Lecoq10, C. Leong1, P. Lousã6, P. Machado1, M. V. Martins3, M. C. Martins6, N. Matela3, R. Moura4, J.A.Neves4, P. Neves6, N. Oliveira3, C. Ortigão4, F. Piedade6, J. F. Pinheiro4, P. Relvas6, A. Rivetti10 , P. Rodrigues4, I. Rolo4, M. Rolo4, A. I. Santos8, J. Santos2, M. M. Silva1, S. Tavernier11, I. C. Teixeira1,9, J. P. Teixeira1,9, J. C. Silva4,10, R. Silva4, A. Trindade4, J. Varela4, 10