lifeport ops manual - rev ir

47
LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 1 1610 Heritage Street Woodland, WA 98674 USA Phone (360) 225-1212 Fax (360) 225-1458 E-mail: [email protected] Title: Operators Manual for EMS Litter System in the Boeing 737 Series Aircraft Document No: CSR-737-00 Rev Date Description of Revision By Approval IR 05-14-05 Initial Release McDaniel McDaniel

Upload: nckad1

Post on 06-Apr-2015

1.569 views

Category:

Documents


46 download

TRANSCRIPT

Page 1: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 1

1610 Heritage Street Woodland, WA 98674 USA

Phone (360) 225-1212 Fax (360) 225-1458

E-mail: [email protected]

Title: Operators Manual for EMS Litter System in the Boeing 737 Series Aircraft Document No: CSR-737-00

Rev Date Description of Revision By Approval

IR 05-14-05 Initial Release McDaniel McDaniel

Page 2: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 2

NOTICE OF CONFIDENTIALITY “This document/data is the property of LifePort, Inc. You may not possess, use, copy, or disclose this document/data or any information contained therein, for any purpose, including without limitation to design, manufacture, or repair parts, or obtain FAA or any other government approval to do so, without, LifePort, Inc. express written permission. Neither receipt nor possession of this document alone, from any source, constitutes such permission. Possession, use, copying, or disclosing by anyone without LifePort, Inc. express written permission is not authorized and may result in criminal and/or civil liability. This document/data remains the property of LifePort, Inc. This document/data is not releasable under the Freedom of Information Act.”

© Copyright LifePort, Inc.

TECHNICAL DATA EXPORT NOTICE This data is exported pursuant to the requirements of the United States Government Export Administration Act of 1969, as amended, and promulgated by the export administration regulations as issued by the U.S. Department of Commerce. The data may not be reproduced and shall not, without the written permission of LifePort, Incorporated, be used for purposes of manufacture, or shall it be disclosed, re-exported nor transmitted directly or indirectly from the importing foreign country to any person, government, governmental entity or institution of another foreign government. It is understood and agreed that the use of this data shall be limited to the following purposes: (i) use by Support Service Contractors 9except for manufacture), (ii) emergency repair or overhaul work, (iii) receiving inspection of hardware, (iv) evaluation of a bid or proposal. By acknowledgement of receipt of data containing this legend, importer agrees to comply thereto.

PMA PART NOTICE

THE USE OF NON-LIFEPORT, INCORPORATED PARTS WILL VOID ANY AND ALL WARRANTIES. LIFEPORT, INCORPORATED WILL NOT ACCEPT RESPONSIBILITY OR LIABILITY FOR ANY PMA PARTS OTHER THAN LIFEPORT, INCORPORTED PARTS; NOR WILL LIFEPORT, INCORPORATED ACCEPT RESPONSIBILTY OR LIABLITY FOR ANY OF THE LIFEPORT INCORPORATED END ITEM PRODUCTS THESE PARTS ARE INSTALLED ON.

Page 3: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 3

List of Active Pages

Pg # Rev i IR ii IR iii IR iv IR 1 IR 2 IR 3 IR 4 IR 5 IR 6 IR 7 IR 8 IR 9 IR 10 IR 11 IR 12 IR 13 IR 14 IR 15 IR 16 IR 17 IR 18 IR 19 IR 20 IR 21 IR 22 IR 23 IR 24 IR 25 IR 26 IR 27 IR 28 IR 29 IR 30 IR 31 IR 32 IR 33 IR 34 IR 35 IR 36 IR 37 IR 38 IR

Page 4: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 4

Table of Contents

NOTICE OF CONFIDENTIALITY .......................................................................................................... 2 TECHNICAL DATA EXPORT NOTICE ................................................................................................. 2 PMA PART NOTICE ................................................................................................................................ 2 LIST OF ACTIVE PAGES.............................................................................................................................. 3

Table of Contents ................................................................................................................................... 4 1.0 Introduction............................................................................................................................ 6 1.1 Purpose of the Manual ........................................................................................................... 6 1.1 Purpose of Equipment............................................................................................................ 6 1.2 Description of (ALS) Equipment........................................................................................... 6

Figure 1.0.1 Base Unit- Left Hand.............................................................................................. 7 Figure 1.0.2 AeroSled TS, Mattress and ARCH attached – Right Side Shown ......................... 7

2.0 Description of Operation for the Base Unit ........................................................................... 8 Figure 3.0.1 AeroSled TS Assembly .......................................................................................... 9

3.0 Description of Operation for the AeroSled TS ...................................................................... 9 3.1 Lever Release......................................................................................................................... 9 3.2 Backrest ................................................................................................................................. 9

Figure 3.0.2 Arch Assembly ..................................................................................................... 10 4.0 ARCH Assembly ................................................................................................................. 11 5.0 Description of (BLS) Equipment ......................................................................................... 12

Figure 5.0.1 (BLS) System ....................................................................................................... 12 6.0 Sled Stands........................................................................................................................... 13

Figure 6.0.1 Sled Stand............................................................................................................. 13 7.0 (ALS) Pneumatic Components ............................................................................................ 14 7.1 Oxygen System .................................................................................................................... 14

Figure 7.0.1 Oxygen Drawer .................................................................................................... 15 Figure 7.0.2 Oxygen Drawer-Sub Components........................................................................ 15

7.2 Air System ........................................................................................................................... 15 Figure 7.0.3 Pump Plate Assembly........................................................................................... 17 Figure 7.0.4 Pump Plate Location............................................................................................. 18

7.3 Vacuum System ................................................................................................................... 18 7.4 Control and Outlet Panel...................................................................................................... 18

Figure 7.0.5 Control and Outlet Panel ...................................................................................... 19 7.5 Components ......................................................................................................................... 19 8.0 Installation and Removal of Base Unit(s)............................................................................ 20 8.1 Installation Instructions for Base Unit(s)............................................................................. 20 8.2 Removal Instructions for Base Unit(s) ................................................................................ 21

Figure 8.0.1 Station location for base units .............................................................................. 21 Figure 8.0.2 Installation Configuration for (ALS) Base units .................................................. 22 Figure 9.0.1 Sled Stand Installation spacing............................................................................. 23

9.0 Installation and Removal of Sled Stand(s) .......................................................................... 23 9.1 Installation Instructions for Sled Stand(s) ........................................................................... 24

Figure 9.0.2 Sled Stand Removal ............................................................................................. 25 9.2 Removal Instructions for Sled Stand(s) ............................................................................... 25 9.3 Removal Instructions for Air/Vacuum Pump ...................................................................... 25

Page 5: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 5

9.4 Installation Instructions for Air/Vacuum Pump................................................................... 25 Figure 9.0.3 Air and Vacuum Pump ......................................................................................... 26 Figure 9.0.4 Air/Water Separator.............................................................................................. 26

9.5 Removal Instructions for Air Regulator/ Water Separator .................................................. 27 9.6 Installation Instructions for Air Regulator/ Water Separator .............................................. 27 9.7 Replacement Instructions for Air Regulator/ Water Separator filter................................... 27 9.8 Removal Instructions for Air Accumulator ......................................................................... 27 9.9 Installation Instructions for Air Accumulator...................................................................... 27 9.10 Removal Instructions for Air Pressure or Suction Gauge.................................................... 28

Figure 9.0.5 Pressure Gauge, Outlet Panel ............................................................................... 28 9.11 Installation Instructions for Air Pressure or Suction Gauge................................................ 29 9.12 Removal Instructions for Switch ......................................................................................... 29

Figure 9.0.6 Switch Removal/Installation ................................................................................ 29 9.13 Installation Instructions for Switch...................................................................................... 29 9.14 Removal Instructions for Circuit Breakers .......................................................................... 31 9.15 Installation Instructions for Circuit Breakers ...................................................................... 31 9.16 Removal Instructions for Converter .................................................................................... 31

Figure 9.0.7 Typical Converter Location.................................................................................. 31 9.17 Installation Instructions for Converter................................................................................. 32

Figure 9.0.8 Oxygen Drawer .................................................................................................... 32 9.18 Removal of Oxygen Drawer ................................................................................................ 32 9.19 Installation of Oxygen Drawer ............................................................................................ 34 9.20 Removal of Oxygen Cylinder .............................................................................................. 34

Figure 9.0.9 Bottle Valve.......................................................................................................... 34 Figure 9.0.10 Drawer Components ........................................................................................... 35

9.21 Installation of Oxygen Cylinder .......................................................................................... 35 9.22 Oxygen System Leak Check................................................................................................ 36 9.23 Removal of Oxygen Regulator ............................................................................................ 36 9.24 Installation of Oxygen Regulator......................................................................................... 36 9.25 Removal of Oxygen Fill Port............................................................................................... 37 9.26 Installation of Oxygen Fill Port ........................................................................................... 37 10.0 Maintenance......................................................................................................................... 38

Figure 10.0.1 Air Intake Location............................................................................................. 38 11.0 System Care ......................................................................................................................... 39 12.0 Oxygen Servicing ................................................................................................................ 40 13.0 Trouble Shooting ................................................................................................................. 41 13.1 Oxygen System .................................................................................................................... 41 13.2 Electrical System ................................................................................................................. 41 13.3 AeroSled and Base............................................................................................................... 42 14.0 Appendix A System Diagrams – Electrical & Plumbing..................................................... 43

Figure 14.0.1 Oxygen Drawer Plumbing Schematic (For components within the oxygen drawer) 43 Figure 14.0.2 Oxygen Drawer Plumbing Routing (For components within the oxygen drawer) 44 Figure 14.0.3 Base Unit Wire Diagram .................................................................................... 45 Figure 14.0.4 Base Unit Wire Diagram Components ............................................................... 46

Page 6: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 6

Figure 14.0.5 Control Panel and Converter Wire Diagram & Components ............................. 47 1.0 Introduction 1.1 Purpose of the Manual This manual provides maintenance and operation information for the LifePort EMS Litter system installed under MDL-747. 1.1 Purpose of Equipment The purpose of this equipment is to provide Advance Life Support (ALS) systems for up to six patients, or Basic Life Support (BLS) for up to six patients, in a Boeing 737 series aircraft. ALS and BLS Systems maybe interchangeable as mission profiles require. 1.2 Description of (ALS) Equipment Each patient position is made of four basic components a base unit, AeroSled TS, AeroSled TS Mattress and an ARCH.

Base Unit. Installs on existing seat track with Brownline studs.

AeroSled TS guide

Latch Station

Storage Drawers

AeroSled TS stop

Oxygen Drawer with Control Panel

Page 7: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 7

Figure 1.0.1 Base Unit- Left Hand

AeroSled ARCH Assembly

AeroSled TS with moveable backrest

AeroSled TS Mattress

Figure 1.0.2 AeroSled TS, Mattress and ARCH attached – Right Side Shown

WARNING Patient’s Head must always be in the direction of Flight, and Backrest must be in the down position for Taxi, Take-off, and Landing.

Page 8: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 8 2.0 Description of Operation for the Base Unit The base carries two Med M oxygen cylinders. These provide oxygen to the outlet panel as well as components mounted on the ARCH assembly. There is a CGA-540 type fill port and gage mounted in the drawer with the cylinders. The base unit also contains an air pump and a vacuum pump. These provide air and suction to outlets on the control panel on this unit and components mounted on the ARCH assembly. This unit receives 115 VAC or 230 VAC from the aircraft and powers outlets on the control panel. This unit also has three storage drawers. The AeroSled TS attaches to the base units with retractable pins. It has a gas spring assisted backrest, a mattress, and restraint belts. An Arch assembly can also be installed on the AeroSled that spans across the patient’s legs and provides mounting for medical equipment.

Page 9: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 9

Backrest Release Cable

Gas Strut

AeroSled TS Release Handles

Forward Latch Plates and Release Pins

Backrest Rack and Pinion assembly

AeroSled TS frame

Figure 3.0.1 AeroSled TS Assembly

3.0 Description of Operation for the AeroSled TS 3.1 Lever Release There are two Lever Release handles located on the AeroSled, one on each side. By pulling out on the Handle Release you pull in on four pins (two per Latch Station) via the Rack and Pinion assembly. Once the pins are retracted you may remove the AeroSled from the Base or Stand assemblies. 3.2 Backrest

The cable at the head of the patient activates the gas spring to raise the backrest. The backrest rises to 60°. The gas spring only assists. Depending on the size of the patient some lifting may be required. Always support the backrest when releasing the gas spring. Caution: Backrest will rise quickly with no patient.

Page 10: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 10

Adjustable IV pole with IV Hooks and mounting capability for up to 12 lbs.

Mount rail for up to 25lbs of storage at a “2.0” C.G

IV Pole lock ring

Support Bar

Figure 3.0.2 Arch Assembly

Page 11: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 11 4.0 ARCH Assembly The ARCH Assembly connects to the AeroSled TS with two Mount Blocks. The mount block (one on each side) has two halves. Each half goes around the tube of the AeroSled TS and secure together with AN bolts in the desired location. The top of the ARCH has a rail that spans across for mounting up to 25lbs with a 2.0” C.G. for items such as Monitors, Flow regulators, Suction canisters, etc. The IV Pole has two hooks for IV bags. The Pole is also rated for an additional 12lbs of storage at a 4.0” C.G off the pole.

Page 12: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 12 5.0 Description of (BLS) Equipment The (BLS) system is comprised of two Sled Stands and the AeroSled TS per patient. The Arch maybe used with the AeroSled if required.

The Sled Stands mount directly to existing seat track with Brownline studs same as the base unit(s).

AeroSled TS

Sled Stands

Figure 5.0.1 (BLS) System

Page 13: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 13 6.0 Sled Stands Each Sled Stand is equipped with a Latch Station and an EMS Emergency Release. The Latch Station is what the AeroSled TS uses to secure itself to the Stand. The EMS Emergency Release is used to release the AeroSled TS if the normal Handle Release becomes faulty or fails while the AeroSled is locked into the Latch Station. To operate the EMS Emergency Release, simply push in both of the red buttons (plungers) at the same time and lift up on the side of the AeroSled TS disengaging the Life Lock Pins.

Latch Station

EMS Emergency Release

Sled Stand

Figure 6.0.1 Sled Stand

Page 14: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 14 7.0 (ALS) Pneumatic Components 7.1 Oxygen System The oxygen drawer is equipped with a CGA-540 fill port and gage for filling the cylinders in place. The following chart was supplied by the manufacturer of the "M" cylinders. It refers only to the Medical "M" cylinder, and should be used as a reference only to gauge the contents of the oxygen system consisting of a single Medical "M" cylinder at different pressures.

Reference Only Sys. Pressure Capacity Capacity Sys. Pressure Capacity Capacity

(PSIG) (cu. ft.) (liters) (PSIG) (cu. ft.) (liters) 2216 122.57 3471 ║ 1200 65.57 1857 2200 121.69 3446 ║ 1100 59.99 1699 2100 116.11 3288 ║ 1000 54.43 1541 2000 110.53 3130 ║ 900 48.89 1385 1900 104.96 2972 ║ 800 43.39 1229 1800 99.31 2812 ║ 700 37.92 1074 1700 93.66 2652 ║ 600 32.48 920 1600 88.08 2494 ║ 500 27.09 767 1500 82.43 2334 ║ 400 21.73 615 1400 76.78 2174 ║ 300 16.42 465 1300 71.18 2016 ║ 200 11.15 316

In all cases, pressure at the outlet is regulated to approximately 55 PSI. LifePort Inc. uses a two-stage regulator. The first stage brings bottle pressure down to a manageable level (around 375 PSI). The second stage regulates to approximately 60 PSI. Care should be taken not to let the bottle pressure run down too low, however, as low pressure can cause fluctuation at the outlet. Never let the bottle pressure go below 400 PSIG if at all possible. Oxygen will still be delivered but flows may vary at the outlet.

Oxygen Gauge

Oxygen Bottles

Control Panel with Flow Regulator

Oxygen Drawer Assembly

Page 15: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 15

Figure 7.0.1 Oxygen Drawer

Air and Suction Gauge

Bottle Stop

Fill Port

Oxygen Gauge

Latch Assemblies

Oxygen Regulator

Bottle Clamp Oxygen Manifold

Drawer Interconnect fittings for Oxygen, Suction, and Air

Figure 7.0.2 Oxygen Drawer-Sub Components 7.2 Air System Compressed air is provided through an air pump. The air pump is an oil-less, non-lube piston pump and cylinder. The pump is rated at 120 PSI (with pressure relief at 100 PSI), but LifePort sets it to 60 PSI. The reason for this is that most medical devices are calibrated to work at 50 PSI, and once air begins to be drawn from the system the pressure will drop. In-house testing has shown a setting of 60 PSI produces the maximum output from the pump when the pressure reduces to 50 PSI. This pressure is adjustable should you need more or less for the optimum performance of your medical devices (see below). The air regulator should only be adjusted by a trained medical technician and tested with the actual medical device with which the air system will be used. In the event of a momentary interruption of the pump, an accumulator provides even pressure and flow at the outlets. An accumulator is simply a cylinder that contains a volume of air (in this case 26 cu. inch as the standard) at the pumps maximum pressure. Also plumbed into the air system is an air - water separator / regulator. This allows the delivery pressure of the compressed air system to be adjusted if need be. It also removes moisture from the air prior to delivery. The separators come standard with a pulse drain, which will automatically expel any moisture so that manual draining is

Page 16: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 16 not necessary. As air is drawn into the system and the pressure fluctuates it causes the drain to retreat from its seat, allowing any accumulated moisture to escape.

Page 17: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 17 There is also a 5-micron filter attached inside of this unit. The filter will further purify the air prior to delivery to the patient. Compressed air delivery is provided though the outlets on the side of the base unit. Outlet types vary depending on customer requirements. All types are checked and will not dispense air until the connection is made with the mating connector. Likewise, each system is keyed so as not to allow connection to the wrong gas source, such as attaching a compressed air line into the oxygen source. Below are the approximate flow rates of the compressed air systems with one pump. These figures are minimums based on LifePort testing and should be taken as reference figures only, as there are many variables that affect the actual flow of air (length of plumbing lines, number of fittings, atmospheric conditions, etc.). LifePort, Inc. recommends that tests be conducted using the operator’s particular brand of ventilator to determine the maximum setting achievable using the compressed air system before actual use on a patient. PRESS. (PSIG) FLOW (CFM) FLOW (LPM) PRESS. (PSIG) FLOW (CFM) FLOW (LPM)

0 .8 21 60 .4 10 10 .7 18 70 .3 9 20 .6 16 80 .3 8 30 .5 14 90 .3 7 40 .5 13 100 .2 7 50 .4 11

As you can see the pressure greatly affects the flow, therefore the allowable pressure differential between the oxygen and compressed air system (when using a customer supplied blending device) also becomes a factor. Again, testing with your ventilator at different settings to determine the maximum allowable settings is strongly advised.

Air & Vacuum Pump

Vibration absorber mount

Air/Water Separator and Accumulator location

Figure 7.0.3 Pump Plate Assembly

Page 18: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 18

Pump Plate location

Figure 7.0.4 Pump Plate Location

7.3 Vacuum System There are two vacuum pumps is installed for medical suction. LifePort’s vacuum pump is a diaphragm pump capable of producing between 20" and 22" -Hg of suction (with free flow). The pump is plumbed directly to the suction outlets on the face panel and arch and is intended to be hooked to a vacuum canister. The same types of gas outlets are available as in the compressed air system. Pumps used in LifePort, Inc. equipment are tested at the factory for flow. Our baseline acceptable output is 20 lpm at 5" -Hg. 7.4 Control and Outlet Panel All the systems are accessed and controlled on the control and outlet panel on the aft end of the base unit.

Page 19: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 19

Air, Suction, and Oxygen Outlets Power

“On-Off” switches and Circuit

Figure 7.0.5 Control and Outlet Panel

7.5 Components Description P/N Figure

1) Base Unit PL560-001-“X’ 2) Oxygen Drawer Rack OX2-003-“X” 3) AeroSled TS AS1-001 4) AeroSled Arch 100-4049 5) AeroSled Mattress FB2-010 6) Sled Stand SN2-006-“X” 7) Ramp Bay Assembly ST2-081

Note: (-“X”) denotes a dash number required for specific configurations.

Page 20: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 20 8.0 Installation and Removal of Base Unit(s) The following information applies to the Base Unit only -- the AeroSled is installed at the time of flight. After the initial installation has been performed and signed off, additional removals and installations require only a log entry, and weight and balance change by a properly rated technician. If the aircraft is to be operated in more than one configuration (i.e. aeromedical and corporate or cargo, etc.), this will facilitate changes from configuration-to-configuration by maintaining a weight and balance form for each configuration used. 8.1 Installation Instructions for Base Unit(s) The following instructions are for installing and/or removing a Base Unit that had, initially, been installed per the STC data, and is reinstalled, unchanged, in the same location.

Step 1: Remove seats as required. Step 2: Install 6 brownline nuts (PN 900-1034) per unit into the aircraft brownline tracks.

Brownline spacing fore/aft, assuming 1st set of brownline studs are 0.0 dimension, 2nd brownline stud set are located at 25.00 inches aft of 1st (for –01 configuration, -02 configuration is located at 26.00 inches aft of 1st), and 3rd brownline stud set are located at 51.00 inches aft of 1st.

Step 3: Lock brownline studs into aircraft seat tracks by rotating the brownline nut. Step 4: Lift Base Unit over brownline studs aligning holes in bottom of weldment. Lower Base

Unit onto brownline studs. Step 5: Install and secure 6 washers (NAS1149F0632P) and 6 nuts (MS21044N6) over the

brownline studs. Step 6: Verify that the Base Unit is securely attached. Step 7: Install drawers. Oxygen drawer installation requires that the Base Unit hoses for air,

vacuum, and oxygen be connected to the outlets located on the outboard side of the drawer. The oxygen drawer also requires the Base Unit electrical connector be connected to the connector located on the outboard side of the drawer.

Step 8: Verify that the oxygen tank is serviced and that the shutoff valve at the head of the bottle is “on.”

Step 9: Update the aircraft weight and balance record and equipment list. The weight and balance update should use data from the initial installation.

Step 10: Make the required log entries.

Page 21: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 21 8.2 Removal Instructions for Base Unit(s)

Step 1: Verify that the oxygen tank shutoff valve at the head of the bottle is “off” and remove seats as required.

Step 2: Remove drawers. Oxygen drawer removal requires that the Base Unit hoses for air, vacuum, and oxygen be disconnected from the outlets located on the outboard side of the drawer. The oxygen drawer also requires the Base Unit electrical connector be disconnected from the connector located on the outboard side of the drawer.

Step 3: Remove 6 washers (NAS1149F0632P) and 6 nuts (MS21044N6) per unit located over the brownline studs.

Step 4: Lift and remove Base Unit off brownline studs. Step 5: Unlock brownline studs into aircraft seat tracks by rotating the brownline nut. Remove

brownline studs from aircraft seat tracks. Step 8: Update the aircraft weight and balance record and equipment list. The weight and balance

update should use data from the initial installation, or per Section 3.4. Step 9: Make the required log entries.

Figure 8.0.1 Station location for base units

Page 22: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 22

Figure 8.0.2 Installation Configuration for (ALS) Base units

Page 23: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 23

Figure 9.0.1 Sled Stand Installation spacing 9.0 Installation and Removal of Sled Stand(s) The following information applies to the Sled Stands only -- the AeroSled is installed at the time of flight. After the initial installation has been performed and signed off, additional removals and installations require only a log entry, and weight and balance change by a properly rated technician. If the aircraft is to be operated in more than one configuration (i.e. aeromedical and corporate or cargo, etc.), this will facilitate changes from configuration-to-configuration by maintaining a weight and balance form for each configuration used.

Page 24: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 24 9.1 Installation Instructions for Sled Stand(s) The following instructions are for installing and/or removing a set of Sled Stands that had, initially, been installed per the STC data, and is reinstalled, unchanged, in the same location. Step 1: Remove seats as required. Step 2: Place Sled Stands at STA 570, 655, and/or 740 (all +/- 10.0). Install first stand and measure 52.38

from the aft portion of the first latch station plate to the aft portion of the latch station located on aft Sled Stand. Install 4 brownline nuts (PN 900-1034), 4 per Sled Stand, into the aircraft brownline tracks.

Step 3: Lock brownline studs into aircraft seat tracks by rotating the (1ea.) brownline nut. Step 4: Lift Sled Stand(s) over brownline studs aligning holes in bottom of weldment. Lower Stand onto

brownline studs. Step 5: Install and secure (8ea.) brownline nuts over the brownline studs. Step 6: Verify that the Sled Stand(s) are securely attached. Step 7: Install AeroSled TS on top of Sled Stands and lock into place. Step 8: Update the aircraft weight and balance record and equipment list. The weight and balance update

should use data from the initial installation. Step 9: Make the required log entries.

Brownline Stud

Knurled Nuts

Page 25: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 25

Figure 9.0.2 Sled Stand Removal 9.2 Removal Instructions for Sled Stand(s)

Step 1: Remove AeroSled TS from Sled Stands if installed. Step 2: Remove (8) brownline nuts from the brownline studs that secure Sled Stands to the brownline

track. Step 3: Lift Sled Stands up and off of the studs and track. Step 4: Remove brownline stud locking nut and studs from track. Step 5: Reinstall seats as required in accordance with manufacturers approved data. Step 6: Update the aircraft weight and balance record and equipment list. The weight and balance update

should use data from the initial installation. Step 7: Make the required log entries. 9.3 Removal Instructions for Air/Vacuum Pump

Step 1: Remove power from Base Unit(s). This may be accomplished by removing connector labeled

"EMS POWER". Step 2: Remove PLUS from aircraft. Step 3: Disconnect wires to pump. Step 4: Bleed off pressure and disconnect plumbing at pump. Cap open lines. Step 5: Remove nuts, bolts or screws attaching pump. Step 6: Remove pump(s). 9.4 Installation Instructions for Air/Vacuum Pump Step 1: Install nuts, bolts or screws to attach pump to PLUS module (3 each MS24694-S5, 3 each AN960-

08). Step 2: Install with Loctite 222 (Loctite Corp, 705 N Mountain Rd, Newington, CT 06111).

CAUTION: Verify that all threads on the pump fittings and respective gas lines are free of foreign material. Step 3: Connect gas lines to pump fittings. Step 4: Connect wires to pump noting polarity of wiring. Step 5: Connect power to Base Unit(s) by installing connector labeled as "EMS POWER". Step 6: Install Base Unit(s) into aircraft. Step 7: Check operation and leak check the system per section

Page 26: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 26

.

Figure 9.0.3 Air and Vacuum Pump

.

Air Pump

Suction Pump

Accumulator

Air/Water Separator and filter

Pressure Relief

Pump Plate

Muffler

Figure 9.0.4 Air/Water Separator

Page 27: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 27

9.5 Removal Instructions for Air Regulator/ Water Separator Step 1: Remove Base Unit(s) from aircraft per section Step 2: Disconnect hoses from Air Regulator / Water Separator. Step 3: Disconnect wires to pump. Step 4: Remove nuts, bolts or screws attaching Air Regulator / Water Separator. Step 5: Remove Air Regulator / Water Separator. 9.6 Installation Instructions for Air Regulator/ Water Separator Step 1: Install nuts, bolts or screws to attach Air Regulator / Water Separator to PLUS module. Step 2: Connect hoses to Air Regulator / Water Separator. CAUTION: Verify that all threads on the Air Regulator / Water Separator fittings and respective gas lines are free of foreign material. Step 3: Connect power to Base Unit(s) by installing connector labeled as "EMS POWER". Step 4: Install Base Unit(s) into aircraft. Step 5: Check operation and leak check the system. 9.7 Replacement Instructions for Air Regulator/ Water Separator filter If the filter is discolored, Step 1: Unscrew and remove the clear plastic housing Step 2: Unscrew the retainer Step 3: Slide the filter off the shaft Step 4: The new filter (P/N 1539B5P Parker Inc. Richland, Michigan or 29660-1 The Aro Co.

Bryan, Ohio) may be installed by reversing this process. 9.8 Removal Instructions for Air Accumulator Step 1: Remove Air Regulator / Water Separator. Step 2: Disconnect hoses from Accumulator. Step 3: Remove fasters and Adel clamps holding the Air Accumulator. 9.9 Installation Instructions for Air Accumulator Step 1: Install fasters and Adel clamps holding the Air Accumulator. Step 2: Connect hoses to Accumulator. Step 3: Install Air Regulator / Water Separator.

Page 28: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 28 9.10 Removal Instructions for Air Pressure or Suction Gauge

Step 1: Remove Control Panel Step 2: From the backside of the control panel, simultaneously push in the two retaining clips.

Step 3: While holding in the retaining clips, push the gauge through the panel. Step 4: Using a wrench, retain the Hose Attach Fitting and rotate the gauge until it is free of the

threaded fitting. Step 5: Cap-n-Stow any open gas lines.

Pressure GaugePanel

AA

View A-A

Hose Attach Fitting

Hose

Retaining Clip

Panel

Figure 9.0.5 Pressure Gauge, Outlet Panel

Page 29: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 29 9.11 Installation Instructions for Air Pressure or Suction Gauge Step 1: Thread the gauge onto the hose attach-fitting. The hose should extend out of the panel cutout. Step 2: Push the gauge into the panel cutout until both retaining clips lock into place. Step 3: Install Control Panel per Section 6.0.4. Step 4: Return power to PLUS by installing connector labeled "EMS POWER". 9.12 Removal Instructions for Switch Step 1: Disconnect power to the Base Unit(s) by removing the "EMS POWER". Step 2: Remove Control Panel. Step 3: Remove the two Spade Terminals on the back of the switch. Step 4: Push in the Retaining Clips and remove the switch from the front of the panel.

NOTE: Switch removal may damage the switch retaining clips; contact LifePort for a purchase of a new switch (P/N 000-4033).

A

A

ON

OFF

Panel

Switch

View A-A

Spade Terminal

Retaining ClipPanel

Figure 9.0.6 Switch Removal/Installation 9.13 Installation Instructions for Switch Step 1: CAUTION: Verify that power has been removed from Base Unit(s). Step 2: Push the switch into the panel until the retaining clips lock into place. Step 3: Install Control Panel.

Page 30: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 30 Step 4: Return power to Base Unit(s) by installing connector labeled "EMS POWER".

Page 31: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 31 9.14 Removal Instructions for Circuit Breakers Step 1: Remove power from Base Unit(s). This may be accomplished by removing connector labeled

"EMS POWER". Step 2: Remove (1) fastener that attaches the circuit breaker to copper bus bar. Step 3: Remove (1) fastener that attaches the wire to circuit breaker Step 4: Remove (1) nut on face of panel that attaches circuit breaker to panel. Step 5: Remove circuit breaker from panel, bag and tag fastener with circuit breaker. 9.15 Installation Instructions for Circuit Breakers Step 1: CAUTION: Verify that power has been removed Base Unit. Step 2: Install (1) nut that attaches circuit breaker to panel. Step 3: Attach wire to circuit breaker using (1) fastener. Step 4: Attach circuit breaker fastener to copper buss bar using (1) fastener. Step 5: Connect power to Base Unit(s) by installing connector labeled as "EMS POWER". 9.16 Removal Instructions for Converter Step 1: Remove power from Base Unit(s). This may be accomplished by removing connector labeled

"EMS POWER". Step 2: Remove Base Unit(s) from aircraft. Step 3: Remove Converter from outboard side of Base Unit(s) by removing four bolts. Step 4: Remove cannon plug from end of converter. Step 5: Remove converter.

Converter location

Figure 9.0.7 Typical Converter Location

Page 32: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 32

9.17 Installation Instructions for Converter Step 1: CAUTION: Verify that power has been removed Base Unit. Step 2: Place converter in required location within Base Unit. Step 3: Reinstall mounting hardware in (4ea.) locations. Step 4: Reinstall cannon plug on end of converter. Step 5: Reinstall Base Unit(s) from aircraft.

Pneumatic lines

Drawer slider Electrical connector

Drawer locks

Figure 9.0.8 Oxygen Drawer

9.18 Removal of Oxygen Drawer Step 1: CAUTION: Drawer is heavy and a two-person lift is advised with bottles installed. Step 2: Ensure Oxygen Bottle valves are closed. Step 3: Remove power from Base Unit(s). This may be accomplished by removing connector labeled "EMS POWER". Step 4: Turn drawer locks on front face of base unit to the “down” position. Step 5: Slide fingers under drawer latches and pull out drawer. Step 6: Disconnect cannon plug from electrical connector found in back of drawer. Step 7: Disconnect pneumatic hand-tight QD’s found in back of drawer and cap.

Page 33: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 33 Step 8: Push in on retainer in drawer sliders and pull drawer out of slider rail.

Page 34: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 34 9.19 Installation of Oxygen Drawer Step 1: CAUTION: Verify that power has been removed Base Unit. Step 2: Slide drawer into drawer sliders and push drawer in until it locks. Step 3: Pull drawer back out to make sure it locks when pulled all of the way out and does not allow the

drawer to come out of the slider rail. Step 4: Remove cap and connect cannon plug to the electrical connector found in back of drawer. Step 5: Remove caps from lines and connect them to the pneumatic hand-tight QD’s found in back of

drawer. Step 6: Open Oxygen Bottle valves. Step 7: Slide drawer closed and turn drawer locks to the “up” position. 9.20 Removal of Oxygen Cylinder

WARNING! Oil, grease or other lubricants, in contact with high-pressure oxygen, can create an extreme fire hazard and any such contact must be avoided. Do NOT permit smoking or open flame of any kind near aircraft or PLUS Utility System for Air Ambulance Operations while work is performed on the oxygen system. Step 1: Remove Base Unit(s) from aircraft Step 2: Close the valve on Oxygen bottle(s). Step 3: Slowly loosen high-pressure line at bottle valve at "CGA" nut and allow lines to discharge. Step 4: Disconnect high-pressure line from bottle valve at "CGA" nut. Step 5: Cap off open oxygen lines. Step 6: Loosen and open band clamps that secure the bottle. Step 7: Remove Oxygen bottle.

Bottle Valve

“CGA” t

Figure 9.0.9 Bottle Valve

Page 35: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 35

CGA Fill Port Bottle Clamps

Regulator

Pneumatic connections Electrical Connector

Figure 9.0.10 Drawer Components 9.21 Installation of Oxygen Cylinder

WARNING! Oil, grease or other lubricants, in contact with high-pressure oxygen, can create an extreme fire hazard and any such contact must be avoided. Do NOT permit smoking or open flame of any kind near aircraft or PLUS Utility System for Air Ambulance Operations while work is performed on the oxygen system. Step 1: Place Oxygen bottle(s) into the Base Unit. Step 2: Wrap Band Clamps around Bottle. Step 3: Secure Bottle with Band Clamps and tighten retaining nut(s). Step 4: CAUTION: Verify that all fittings are clean and free of contaminates. Step 5: Connect high-pressure line to Oxygen bottle valve at large "CGA" nut. Step 6: CAUTION: Verify that all high-pressure lines are connected. Step 7: Open the Oxygen valve on the bottle and leak check the system.

Page 36: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 36 9.22 Oxygen System Leak Check

WARNING! Oil, grease or other lubricants, in contact with high-pressure oxygen, can create an extreme fire hazard and any such contact must be avoided. Do NOT permit smoking or open flame of any kind near aircraft or PLUS while work is performed on the oxygen system. Step 1: Charge the system by opening the valve on the oxygen bottle. Step 2: Connect a low pressure gauge (0-100psi) to the outlet panel oxygen fitting. Step 3: Close the valve on the oxygen bottle. Step 4: Record the pressure of the low pressure gauge and the pressure on the high pressure gauge. This

may be accomplished by placing a piece of masking tape on the cover of each gauge so that the edge of the tape is aligned with the needle, and the indicated amount of pressure.

Step 5: Verify that the pressure in either gauge does not decrease by more than 5% in five minutes. Step 6: Remove pressure gauges, and tape from gauge cover. 9.23 Removal of Oxygen Regulator

WARNING! Oil, grease or other lubricants, in contact with high-pressure oxygen, can create an extreme fire hazard and any such contact must be avoided. Do NOT permit smoking or open flame of any kind near aircraft or PLUS Utility System for Air Ambulance Operations while work is performed on the oxygen system. Step 1: Disconnect power to the Base Unit by removing the "EMS POWER". Step 2: Close the valve on Oxygen bottle(s). Step 3: Open slowly the fitting on the high-pressure side of the Oxygen regulator and allow any Oxygen

to escape. (The High and Low side is marked on the Oxygen regulator and Oxygen solenoid housing.) After pressure has been released, fully remove the high-pressure side fitting from the Oxygen regulator.

Step 4: Cap-n-Stow an open Oxygen lines. Step 5: Remove Adel clamps and hardware securing regulator. Step 6: Remove the regulator from the low-pressure fitting. Step 7: Remove the regulator from the Base Unit 9.24 Installation of Oxygen Regulator

WARNING! Oil, grease or other lubricants, in contact with high-pressure oxygen, can create an extreme fire hazard and any such contact must be avoided. Do NOT permit smoking or open flame of any kind near aircraft or PLUS Utility System for Air Ambulance Operations while work is performed on the oxygen system. Step 1: Install the Oxygen regulator onto the Base Unit using the appropriate hardware. Step 2: Connect the high and low-pressure lines. Step 3: Open the Oxygen valve on the bottle and leak check the system.

Page 37: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 37 9.25 Removal of Oxygen Fill Port Step 1: Close Oxygen bottle valve. Step 2: Discharge oxygen system by venting CGA nut at bottle valve. Step 3: Remove high-pressure line from fill port. Step 4: Cap off all open oxygen lines. Step 5: Remove screws supporting fill port. Step 6: Remove fill port from the PLUS. 9.26 Installation of Oxygen Fill Port Step 1: Install supporting fasteners for Fill Port, using (3 each AN525-10R8). Step 2: Remove temporary cap from Oxygen line. Step 3: CAUTION: Verify that all threads on the Oxygen filler port fitting and respective Oxygen lines

are free of foreign material. Step 4: Install Oxygen line onto fill port, tighten CGA nut at bottle valve. Step 5: Open the Oxygen valve on the bottle and leak check the system.

Page 38: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 38 10.0 Maintenance The Instructions for Continued Airworthiness includes the required inspection schedule. The only regular maintenance requirements concern the oxygen bottles and the filters. The oxygen bottle must be hydrostatically tested (ref CFR 173.34) at time intervals not to exceed those specified by DOT. For the 3AL aluminum pressure vessels, the time limit is 5 years. LifePort, Inc. requires that the oxygen regulator be replaced with a new unit at the time of the oxygen bottle hydrostatic check. The regulator is not considered a repairable item. The air - water separator has a filter that should be replaced when discolored. There is an additional filter on the side of the air pump for the intake. This filter is a white fiber pad about 1” in diameter. Replace filter when dirty, retain with safety wire or a cotter pin.

Pressure Relief Filter location

Figure 10.0.1 Air Intake Location

Page 39: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 39 11.0 System Care To disinfect the medical system, isopropyl alcohol or a 5% bleach and water mix is recommended. Commercially available products to disinfect may also be used as long as they don’t contain other chemicals. ****************************************************CAUTION***************************************************

NEVER USE AN OIL OR CHEMICAL BASED CLEANER. The front panels of the base unit are made of Lexan plastic. Interaction with some chemicals causes “crazing”, or radical cracking. Oil or oil-based products should never be used in the vicinity of oxygen. The mattress has a stain resistant coating on it. Use soap and water or a very mild detergent for cleaning. Chemical based or abrasive cleaners can remove the stain resistant properties of the material and therefore should not be used. Moving parts, such as the back rest hinge, knee lift hinges, bed arch hinge and bed release pins, should be periodically lubricated for optimum performance. Any dry type lubricant (such as graphite or silicon lubricant) can be used. When the system is removed from the aircraft it should be stored so that it is protected from the elements, such as rain, direct sunlight, sand and salt spray. Extreme temperatures should be avoided. System components should be kept free of oil and grease due to the oxygen. The bed and cargo pallet have plastic covers on he feet that can pick up debris. These should only be set on a clean surface. All open connectors and gas fittings should be capped.

Page 40: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 40 12.0 Oxygen Servicing The oxygen system is serviced through a fill port located in the large drawer in the 4’ base unit. The location of the fill port is clearly marked.

WARNING! Oil, grease or other lubricants, in contact with high-pressure oxygen, can create an extreme fire hazard and any such contact must be avoided. Do NOT permit smoking or open flame of any kind near airplane or system while work is performed on the oxygen system.

1. Use breathing oxygen conforming to MIL-O-27210 or other specification approved by the medical personnel for

medical use. 2. Check oxygen system high pressure gauge. 3. Check that oxygen bottle valve is on. 4. Do not charge system if fittings or filler valve are corroded, contaminated or damaged. Clean filler valve, if

contaminated, with stabilized trichloroethylene and let air-dry. Do not use solvent. 5. If cylinder is completely empty, do not charge until system has been purged. To purge, charge system to

approximately 500 psi and vent. 6. Connect filler valve to external filling equipment. 7. Fill bottle per instructions associated with the filling apparatus, whether a large vessel or cascade system. 8. Fill to the pressure in the following table based on the ambient temperature.

Temp. Press. Temp. Press. (°F) (PSIG) (°C) (KPa)

0 1975 -20 13467 10 2035 -15 13807 20 2065 -10 14147 30 2125 -5 14488 40 2185 0 14828 50 2244 5 15168 60 2304 10 15508 70 2364 15 15849 80 2394 20 16189 90 2454 25 16529 100 2514 30 16869 110 2574 35 17210 120 2633 40 17550

45 17890 50 18230

Note: When filling the bottle on a 60°F day, fill to 2304 PSI. Upon cooling, the pressure should stabilize to 2216 PSI.

Page 41: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 41 13.0 Trouble Shooting Some of the more common failures and their remedies are listed here. This is not intended as a comprehensive list, but as a general guide 13.1 Oxygen System Trouble Probable Cause Remedy No indication on oxygen Bottle depleted Fill Bottle pressure gauge. Gage line crimped Replace line Bad Gage Replace gage Bottle valve off Turn valve on Oxygen duration too short Leak in system Repair leak Regulated pressure too high Replace regulator Oxygen flow inadequate Regulator defective Replace regulator Line Crimped Replace line Leak Repair leak 13.2 Electrical System Trouble Probable Cause Remedy No 115/230 VAC at outlet Circuit breaker tripped Reset circuit breaker EMS switch in Turn switch on cockpit "OFF" Defective wiring Replace defective wiring Defective Outlet Replace Outlet No air pressure at outlet Circuit breaker tripped Reset circuit breaker Switch "OFF" Turn "ON" Defective wiring Replace defective wiring Pump defective Replace pump Leak Repair leak Regulator defective Replace regulator

Page 42: Lifeport Ops Manual - Rev IR

LifePort, Inc.

EMS Litter Operators Manual CSR-737-00 Rev IR

Page 42

13.2 Electrical System (Continued) Trouble Probable Cause Remedy No vacuum at outlet Circuit breaker tripped Reset circuit breaker Switch "OFF" Turn "ON" Defective wiring Replace defective wiring Pump defective Replace pump Leak Repair leak 13.3 AeroSled and Base Trouble Probable Cause Remedy AeroSled does not Dirt in latch mechanism Clean mechanism engage (All four pins) Bent latch pin Replace pin Foreign object in latch Clear receptacle station receptacle AeroSled does not release Obstruction in release Clear mechanism from Base mechanism Lap belt fails to engage Defective latch Replace entire belt assembly AeroSled backrest Broken Cable Replace cable does not release Cable out of rig Rig cable Gas strut defective Replace strut AeroSled backrest Broken Cable Replace cable does not lock Cable out of rig Rig cable Gas strut defective Replace strut

Page 43: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 43 14.0 Appendix A System Diagrams – Electrical & Plumbing

1 900-1015 TUBING - STAINLESS - 1/8"2 HS2-002-( ) ASSEMBLY - HOSE - BRAIDED/END3 984-0006-01 SS BRAIDED HOSE - HIGH PSI

Figure 14.0.1 Oxygen Drawer Plumbing Schematic (For components within the oxygen drawer)

Page 44: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 44

Figure 14.0.2 Oxygen Drawer Plumbing Routing (For components within the oxygen drawer)

Page 45: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 45

Figure 14.0.3 Base Unit Wire Diagram

Page 46: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 46

Figure 14.0.4 Base Unit Wire Diagram Components

Page 47: Lifeport Ops Manual - Rev IR

LifePort, Inc. EMS Litter Operators Manual CSR-737-00 Rev IR Page 47

Figure 14.0.5 Control Panel and Converter Wire Diagram & Components