leadcare ii
TRANSCRIPT
![Page 1: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/1.jpg)
LeadCare II
Robert L. Jones, PhD
Inorganic Toxicology LaboratoryDivision of Laboratory Sciences
National Center for Environmental HealthCenters for Disease Control and Prevention
October 18th, 2006
![Page 2: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/2.jpg)
DisclaimerMention of company or Mention of company or product names does not product names does not constitute endorsement by the constitute endorsement by the National Center for National Center for Environmental Health (NCEH), Environmental Health (NCEH), Centers for Disease Control Centers for Disease Control (CDC), or the Public Health (CDC), or the Public Health Service.Service.
![Page 3: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/3.jpg)
LeadCare II: Portable Blood Lead Analyzer
www.esainc.com
![Page 4: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/4.jpg)
LeadCare to LeadCare II
1992-1997
2001-2006
![Page 5: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/5.jpg)
LeadCare II Blood Lead Testing
Sept. 18th, 2006CLIA “Waived”
![Page 6: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/6.jpg)
![Page 7: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/7.jpg)
![Page 8: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/8.jpg)
LeadCare II Issues: CLIA 88: “Waived”
Automated InstrumentMinimal risk of erroneous results
Use of fresh blood only (< 24 hours)
Blood Reagent Mixture (< 48 hours) [ambient]
Blood Reagent Mixture (< 7 days) [4°C]
![Page 9: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/9.jpg)
LeadCare II Issues: 50uL Whole Blood (capillary or venous)
LOD = 3.3 ug/dL
Range: 3.3 to 65 ug/dL
Sensor activated testing
Sensor “covered”
![Page 10: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/10.jpg)
LeadCare II Issues:
Price $2,200
Cost per test $ 6 - $8
![Page 11: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/11.jpg)
LeadCare and LeadCare II2006 PT Sample Results
LeadCare Target Value (ug/dL)0 10 20 30 40 50 60
Obs
- Ta
rget
Val
ue (u
g/dL
)
-5
-4
-3
-2
-1
0
1
2
3
4
5
2SD
-2 SD
LeadcareLeadcare II
![Page 12: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/12.jpg)
Questions
Discussions
![Page 13: LeadCare II](https://reader035.vdocuments.us/reader035/viewer/2022081614/62c8f4a1e93fda33f16ab17e/html5/thumbnails/13.jpg)
Contact InfoRobert L. Jones, PhD
Centers for Disease Control and Prevention
4770 Buford HwyMailstop F-55
Atlanta, GA 30341